Embodiments of the present disclosure relate to automatic Doppler measurement, and more specifically, to a workflow for automatic measurement of Doppler.
According to embodiments of the present disclosure, methods of and computer program products for automatic Doppler measurement are provided. A plurality of frames of a medical video are read. A mode label indicative of a mode of each of the plurality of frames is determined. At least one of the plurality of frames is provided to a trained feature generator. The at least one of the plurality of frames have the same mode label. At least one feature vector is obtained from the trained feature generator corresponding to the at least one of the plurality of frames. At least one feature vector is provided to a trained classifier. A valve label indicative of a valve is obtained from the trained classifier corresponding to the at least one of the plurality of frames. One or more measurement is extracted indicative of a disease condition from those of the at least one of the plurality of frames matching a predetermined valve label.
With the growth of big data through large electronic health records (EHR), there is an opportunity to leverage medical image analysis in combination with other modality data in EHR to impact the quality of care to patients in a significant way.
Despite the opportunity to record the data in digital electronic records, data entry human errors are quite common in workflow-based electronic health record systems. A billing operator may fail to capture a relevant billing code for a disease by looking at the problem list in an EMR system. A clinician or nurse assistant may fail to put the disease in the problem list of the EMR system by transcribing from the exam's clinical reports. A data entry operator in an imaging facility may fail to transcribe all measurements from the technician into the report. A technician may fail to capture the measurements in the diagnostic study either because he/she was not told to look for the evidence for a disease (possibly because the ordering clinician did not specify this), or because he/she simply forgot to save the measurements or screens on an imaging study in PACS. Any of these errors can ultimately result in a patient not being flagged for a disease. If the disease is one in which sudden death is possible, such as in aortic stenosis, catching such data entry errors through a systematic peer review process can save lives. The present disclosure provides systems and methods for systematically identifying such discrepancies in order to notify the corresponding data entry operator. In various embodiments, this is done by detecting evidence for a disease from multiple information sources, and may include spotting mentions of the disease name in reports, noting the measurements already made by operators that were not flagged, or making new measurements directly from imaging studies.
Accordingly, the present disclosure provides systems and methods for disease detection based on various pre-existing patient data drawn from a variety of different data sources. In particular, the present disclosure provides for the use of medical image analysis in combination with textual and other multimodal data analysis for purposes of identifying patient cohorts at risk for serious diseases such as aortic stenosis.
A Picture Archiving and Communication System (PACS) is a medical imaging system that provides storage and access to images from multiple modalities. In many healthcare environments, electronic images and reports are transmitted digitally via PACS, thus eliminating the need to manually file, retrieve, or transport film jackets. A standard format for PACS image storage and transfer is DICOM (Digital Imaging and Communications in Medicine). Non-image data, such as scanned documents, may be incorporated using various standard formats such as PDF (Portable Document Format) encapsulated in DICOM.
An electronic health record (EHR), or electronic medical record (EMR), may refer to the systematized collection of patient and population electronically-stored health information in a digital format. These records can be shared across different health care settings and may extend beyond the information available in a PACS discussed above. Records may be shared through network-connected, enterprise-wide information systems or other information networks and exchanges. EHRs may include a range of data, including demographics, medical history, medication and allergies, immunization status, laboratory test results, radiology reports, radiology images, vital signs, personal statistics like age and weight, and billing information.
EHR systems may be designed to store data and capture the state of a patient across time. In this way, the need to track down a patient's previous paper medical records is eliminated. In addition, an EHR system may assist in ensuring that data is accurate and legible. It may reduce risk of data replication as the data is centralized. Due to the digital information being searchable, EMRs may be more effective when extracting medical data for the examination of possible trends and long term changes in a patient. Population-based studies of medical records may also be facilitated by the widespread adoption of EHRs and EMRs.
Health Level-7 or HL7 refers to a set of international standards for transfer of clinical and administrative data between software applications used by various healthcare providers. These standards focus on the application layer, which is layer 7 in the OSI model. Hospitals and other healthcare provider organizations may have many different computer systems used for everything from billing records to patient tracking. Ideally, all of these systems may communicate with each other when they receive new information or when they wish to retrieve information, but adoption of such approaches is not widespread. These data standards are meant to allow healthcare organizations to easily share clinical information. This ability to exchange information may help to minimize variability in medical care and the tendency for medical care to be geographically isolated.
In various systems, connections between a PACS, Electronic Medical Record (EMR), Hospital Information System (HIS), Radiology Information System (RIS), or report repository are provided. In this way, records and reports form the EMR may be ingested for analysis. For example, in addition to ingesting and storing HL7 orders and results messages, ADT messages may be used, or an EMR, RIS, or report repository may be queried directly via product specific mechanisms. Such mechanisms include Fast Health Interoperability Resources (FHIR) for relevant clinical information. Clinical data may also be obtained via receipt of various HL7 CDA documents such as a Continuity of Care Document (CCD). Various additional proprietary or site-customized query methods may also be employed in addition to the standard methods.
According to various embodiments of the present disclosure, automated systems and methods for retrospectively predicting patients likely to have disease conditions such as aortic stenosis are provided. In various embodiments, medical image analysis of Doppler patterns is combined with textual content analysis of imaging and reports in a multimodal learning framework. Specifically, evidence of disease conditions such as aortic stenosis may be extracted from sources, such as billable diagnosis, significant problems from EHR, echocardiogram reports, measurements shown on echocardiography video frames, or CW Doppler patterns in echocardiography videos. In some embodiments, disease concepts are identified in echocardiogram reports using a concept extraction algorithm to detect UMLS concept vocabularies and their relevant associated measurements. In some embodiments, measurements captured by echocardiogaphers are reliably extracted through selective image processing and optical character recognition in tabular regions on echocardiogram video frames. In some embodiments, diagnostically relevant measurements for aortic stenosis are automatically extracted from Doppler envelopes using a three step process of relevant Doppler frame identification, envelope tracing and measurement extraction. The frame identification includes classification using convolutional neural network (CNN)-based learned features from Doppler regions. The envelop extraction is made robust by incorporating echocardiographer tracings. In some embodiments, the disease-specific features extracted from each multimodal source of information are combined using a random forest learning formulation to predict patients that are likely to have aortic valve disease.
Referring to
The present disclosure provides for identifying patients at risk of various disease conditions such as aortic stenosis by combining medical text and image analysis of medical data such as echocardiagram studies. Such multimodal information may be used for cohort identification. Several embodiments provided herein overlap three inter-disciplinary fields of text analysis, optical character recognition (OCR), and medical image analysis. In various embodiments extraction of clinical concepts from text is provided. In various embodiments, measurements are extracted in addition to disease name mentions for detection of disease conditions. In various embodiments, OCR extraction of clinical measurements from text screens of studies such as echocardiogram studies is provided. Some such embodiments do not rely on manual creation of templates for various manufacturer's echo screens. In various embodiments, reliable extraction of Doppler envelopes is provided, even in the presence of electrocardiogram (ECG) fluctuations during arrhythmia and overlay artifacts in Doppler spectra. In various embodiments, the automatic selection of Doppler frames depicting aortic valves is provided.
Various exemplary embodiments provided herein are given with regard to aortic stenosis. However, it will be apparent that the present disclosure is suitable for application to any other disease condition that may be tied to clinical data or observations. Aortic stenosis (AS) is a common heart disease that can result in sudden death. It can be diagnosed through the Doppler patterns in echocardiogram studies such as the exemplary study shown in
Disease Extraction from Reports
In various embodiments of the present disclosure, extraction of evidence of a disease condition from clinical reports is provided. In some exemplary embodiments, the disease condition is aortic stenosis and the clinical report is an echocardiogram report.
To extract evidence of aortic stenosis from echocardiogram reports, a large knowledge graph is generated of over 5.6 million concept terms by combining over 70 reference vocabularies such as SNOMED CT, ICD9, ICD10, RadLex, RxNorm, and LOINC, where its concept nodes are used as vocabulary phrases. The detections of clinical concepts within sentences of the clinical reports uses the longest common subfix (LCF) algorithm. To detect evidence of stenosis, tuples of Di, Sj, Ak, Vl, are found where Di are disease name indicators (e.g., “aortic valve disorders,” “aortic valve stenosis,” etc.), Sj are specific symptoms associated with the disease such as “chest pain,” Ak are anatomical abnormalities such as “thickened,” “calcified,” and Vl are qualifiers such as “mild,” “moderate,” “severe.” These detections are done within neighboring sentences in selected paragraphs where the aortic valve is described in echocardiogram reports.
Next, key measurement names are selected indicating aortic stenosis, such as peak velocity, mean pressure gradient, and aortic valve area. Using their values ranges and units, a measurement name-value pair detector is developed. As the spoken utterances of these names vary in echocardiograms, n-gram analysis is performed of a corpus of over 50,000 reports in a data collection to identify all such significant variants of the measurement names. To detect occurrences of measurement names and their associated values within the context of a detected sentence, the pattern of their occurrences in a sentence is analyzed using part-of-speech (POS) tagging, and dependency graph parsing. For each root concept (e.g., ‘gradient’), a chain of its modifiers (in the form of nouns or adjectives, e.g., ‘mean trans aortic’) are automatically identified from a sentence using an automatic POS tagger. In some embodiments, the automatic POS tagger comprises the Stanford POS tagger. By analyzing thousands of sentences containing the occurrences of measurement vocabulary terms in connection with measurement values and units, regular expression patterns are formed, such a pattern ABC where A is any disease indicating phrase A: {aorta, aortic, AV, AS}, B is any measurement term {gradient, velocity, area}, and C is no negation terms of the kind {no, not, without, neither, none}. Once the pattern is matched, numeric values are located following the measurement names in the same sentence that were juxtaposed with names of relevant units. An example of aortic stenosis measurement extraction is illustrated in the paragraph below.
“Aortic Valve: The aortic valve is thickened and calcified. Severe aortic stenosis is present. The aortic valve peak velocity is 6.18 m/s, the peak gradient is 152.8 mmHg, and the mean gradient is 84.9 mmHg. The aortic valve area is estimated to be 0.28 cm2.”
In general, the text-based aortic stenosis detection is fairly stable with very few false positives as indicated in Table 1. Only 3 errors were observed among thousands of patients after a thorough analysis of the detected cases, as listed in the third column. Table 1 illustrates the false discovery rate (FDR) of disease (AS) and measurement (peak velocity and mean gradient) detection.
Extracting Echocardiographer Measurements
Referring to
The evidence for aortic stenosis can be extracted from the measurements made by the echocardiographer captured as text-only screens such as the one shown in
An optical character recognition algorithm is used to extract text. In some embodiment, the DataCap OCR engine is used while in other embodiments, the Tesseract OCR engine is used. However, it will be appreciated that any number of OCR engines are suitable for use according to the present disclosure. The document layout templates of device manufacturer's screens is learned automatically. The template learning is focused per anatomical region and exploits the invariance in topological layout of the measurement name value pairs in the tabular regions. Once the templates are learned, they are matched to any given text only screen to read off the expected measurement names.
The images are processed within the text regions through an image enhancement process to increase the robustness of OCR.
Referring to
According to various embodiments, the training mode takes in an initial sample of images drawn from a larger corpus of medical imagery. In some embodiments, this initial sample comprises ˜100-200 randomly selected images. In addition to generating a template as set forth herein, a list of common typos is also compiled based on the OCR module output. This list is useful to supplement the knowledge base used to increase measurement accuracy. In order to simplify and speed up training, the training images are ranked for presentation to a user for verification, as discussed above. The ranking algorithm allows presentation of images with the most information to the user in the earliest stages of training. In this way, the initial determinations made by the system are reinforced. Via a user interface, a user can correct for mistakes or missed measurements. By optimizing presentation of screens for verification, a user is required to evaluate fewer images and to spend less time overall on validation and correction.
As noted above, in various embodiments, a graphical user interface is provided for training. Given a pool of new machine images N, they are clustered into k different groups G. The images are ranked based on the set cover algorithm given below. A user is presented with one or more images from each group G in combination with an initial determination, which in some embodiments is based on rule-based approach, for every measurement attribute-value-unit pair and its section header. In some embodiments, the rule based approach comprises applying predetermined templates based on vendor and device information associated with the image. For example, a given vendor may always place certain measurements at a predetermined location in a frame, while other vendors may have significant variation among product lines. Such a rule based approach is used to detect table structure to locate section names, table header, measurement names and units.
Through the UI, a user can annotate the image, i.e., by correcting misclassified pairs, removing non-relevant words, or adding user-defined naming. Subsequent images may then be updated based on the current corrections. This may reduce by half the amount of clicking in following annotations.
The above process may be repeated until all images in the training set have been annotated. Image templates of each group Gi:ti, . . . , tj are merged to generate a cluster-specific/rank-specific template Ti. After training, the machine-specific template comprises a set of cluster-templates/rank-templates, i.e., T{T1, . . . , Tk}, which can be used to extract measurements from any test image at run time.
Referring to
Referring to
In the runtime stage, a new image is assigned to a cluster based on a similarity score. The cluster-template is retrieved and used to extract measurements from the test image. If there is no close template, the default rule-based approach is used to extract the measurements. T is searched extensively to see if any measurements are matched to the test image. Images that need retraining will be set aside or flagged so user can rerun the training phase to create a new template from the unclassified test image.
Disease Extraction from Doppler Image Analysis
In Doppler echocardiography images, the clinically relevant region is known to be within the Doppler spectrum, contained in a rectangular region of interest as shown in
In various embodiments, a frame of interest is read, and the mode reflected in the frame is determined in advance of valve detection. For example, a frame may reflect B-Mode, M-Mode, CW-Doppler, PW-Doppler, Text-Panel, Color-Doppler, Color-Doppler M-Mode. A frame, or a region of interest therein is provided to a pre-trained convolutional neural network (CNN). The CNN is used as a feature generator. In some embodiments, a feature vector of size 4096 is harvested at the output of the first fully connected layer of the network and the images classified using a support vector machine (SVM) classifier. To train the SVM, a dataset of imagery is used reflecting a plurality of modes, each labeled with the appropriate mode. In embodiments including mode detection, the valve detection stage may take as input the mode label. In some embodiments, multiple valve detection stages are maintained, corresponding to each possible mode label, and the image is routed to the appropriate valve detection stage according to the valve label.
Extraction of Doppler Patterns
Extracting of Doppler spectrum in some embodiments uses pre-processing steps of region of interest detection, ECG extraction, and periodicity detection. In addition, various embodiments exploit the tracings of echocardiographers as shown in
Measurement Extraction from Doppler Patterns
Maximum jet velocity (Vmax) is defined as the peak velocity in the negative direction for the Doppler pattern for aortic stenosis. Since the Doppler envelope traces are available, the pixel value of the negative peak in the Doppler spectra can be easily noted. To convert the imaging-based measurement to a physical velocity value, the text calibration markers on the vertical axis in the ROI are analyzed using an OCR engine to read off the velocity value. The maximum value of velocity during systole within each cycle is a candidate for the Vmax. The second measurement indicative of aortic stenosis is mean pressure gradient (MPG). MPG is calculated from velocity information following the estimation in Equation 2, where N is the number of pixels within the QT interval of ECG, and V is the velocity.
Disease Prediction Using Multimodal Learning
In some embodiments, after collecting all the measurements derived from each modality processing, a feature vector is formed as follows in Equation 3, where the ‘b’ is for billable diagnosis, ‘s’ for significant problems, ‘t’ for textual reports, ‘o’ for video text, and ‘i’ for image analysis features. The first 3 features are binary while the rest are actual measurements made in the respective modalities. To train the predictor, a set of patients may be used with known aortic stenosis (independently validated clinically), and learn the correlation between feature values and the disease label (aortic stenosis) using a random forests learner. In some embodiments, the random forests are constructed with T trees (e.g., T=4, T=100), with each tree having a minimum node size n (e.g., n=4, n=10), and maximum depth of 10. It will be appreciated that the parameters can be tuned depending on the task of interest and the size of the available training data.
Fp={V1b,V2s,V3t,V4t,V5t,V6o,V7o,V8i,V9i} Equation 3
Given determination of a disease label as described above, prior studies may be surveyed to determine when such a diagnosis first occurred. In this way, gaps or discrepancies in the record may be located. When a gap or discrepancy is detected, a notification may be dispatched to a technician or other personnel such as a medical coder.
According to various embodiments, when reading a current study for a patient the past study records for that patient are processed (both reports NLP and image cognition) and compared to the current study both as prior positive findings to reconfirm and as incidental findings (e.g., a mass was seen but dismissed as benign). In some embodiments, prior inconsistent results may be included in a PACS worklist. In this manner, a re-read or review process may be triggered.
Clinical Study Results
A retrospective clinical study was conducted on a large patient data set acquired from a nearby hospital. The experimental context was to evaluate if there were missed diagnosis of aortic stenosis in their records when in fact evidence could be found from the underlying clinical data. Specifically, the analysis was restricted to patients for which 4 modalities of information were available, namely, billable diagnosis, significant problems, and echocardiogram reports and imaging studies giving rise to a total of 991 patients with 1,226 reports and 121,811 Doppler images. These studies were independently validated clinically and 395 patients were found to have aortic stenosis serving as the ground truth.
A 10 fold cross-validation was done by randomly splitting the data into 10 folds, 9 for training and 1 for testing. Table 3 shows the precision, recall, F-score, and overall accuracy of the baseline and random forests with different combinations of features, including a fusion of image and OCR features—referred to as min(I,O). Selecting the minimum of these two values gave a more conservative estimate of the severity of the disease. Out of the 395 patients manually identified by experts, 99 were newly discovered patients from our multimodal analysis giving rise to over 25% new discoveries.
Table 3 illustrates comparative performance of rule-based baseline and random forest with features extracted from structured information, reports, images, and OCR text. min(I,O) refers to the fusion of image and OCR features by taking the minimum of the two for each individual feature/parameter.
Comparison Against Baseline
The baseline was a rule-based model, which returned all patients with at least one piece of evidence from any of five sources. Here the evidence was either the presence of disease mentions or exceeding the normal ranges for Vmax and Mg according to the AHA guidelines. The best-performing model was a random forest with features from all the different sources, achieving 96% precision that is 12% higher than the baseline. Combining features using random forests compensates for potential errors in individual modality detections, making its precision higher than the baseline method. The higher precision will reduce unnecessarily flagging of patients which would have otherwise have lowered the confidence in such prediction system for practical uses.
Referring to
A PACS may handle images from various medical imaging instruments, such as X-ray plain film (PF), ultrasound (US), magnetic resonance (MR), Nuclear Medicine imaging, positron emission tomography (PET), computed tomography (CT), endoscopy (ES), mammograms (MG), digital radiography (DR), computed radiography (CR), Histopathology, or ophthalmology. However, a PACS is not limited to a predetermined list of images, and supports clinical areas beyond conventional sources of imaging such as radiology, cardiology, oncology, or gastroenterology.
Different users may have a different view into the overall PACS system. For example, while a radiologist may typically access a viewing station, a technologist may typically access a QA workstation.
In some implementations, the PACS Gateway 811 comprises a quality assurance (QA) workstation. The QA workstation provides a checkpoint to make sure patient demographics are correct as well as other important attributes of a study. If the study information is correct the images are passed to the archive 812 for storage. The central storage device, archive 812, stores images and in some implementations, reports, measurements and other information that resides with the images.
Once images are stored to archive 812, they may be accessed from reading workstations 821 . . . 829. The reading workstation is where a radiologist reviews the patient's study and formulates their diagnosis. In some implementations, a reporting package is tied to the reading workstation to assist the radiologist with dictating a final report. A variety of reporting systems may be integrated with the PACS, including those that rely upon traditional dictation. In some implementations, CD or DVD authoring software is included in workstations 821 . . . 829 to burn patient studies for distribution to patients or referring physicians.
In some implementations, a PACS includes web-based interfaces for workstations 821 . . . 829. Such web interfaces may be accessed via the internet or a Wide Area Network (WAN). In some implementations, connection security is provided by a VPN (Virtual Private Network) or SSL (Secure Sockets Layer). The clients side software may comprise ActiveX, JavaScript, or a Java Applet. PACS clients may also be full applications which utilize the full resources of the computer they are executing on outside of the web environment.
Communication within PACS is generally provided via Digital Imaging and Communications in Medicine (DICOM). DICOM provides a standard for handling, storing, printing, and transmitting information in medical imaging. It includes a file format definition and a network communications protocol. The communication protocol is an application protocol that uses TCP/IP to communicate between systems. DICOM files can be exchanged between two entities that are capable of receiving image and patient data in DICOM format.
DICOM groups information into data sets. For example, a file containing a particular image, generally contains a patient ID within the file, so that the image can never be separated from this information by mistake. A DICOM data object consists of a number of attributes, including items such as name and patient ID, as well as a special attribute containing the image pixel data. Thus, the main object has no header as such, but instead comprises a list of attributes, including the pixel data. A DICOM object containing pixel data may correspond to a single image, or may contain multiple frames, allowing storage of cine loops or other multi-frame data. DICOM supports three- or four-dimensional data encapsulated in a single DICOM object. Pixel data may be compressed using a variety of standards, including JPEG, Lossless JPEG, JPEG 2000, and Run-length encoding (RLE). LZW (zip) compression may be used for the whole data set or just the pixel data.
Referring to
Referring to
Referring to
Referring to
Referring to
Referring to
In various embodiments, determining the mode label includes providing the plurality of frames to a second trained feature generator. A plurality of feature vectors corresponding to the plurality of frames are obtained from the second trained feature generator. The plurality of feature vectors is provided to a second trained classifier. The mode label is obtained from the second trained classifier. In such embodiments, a set of features discriminating between mode labels are learned by the classifier, which may be a deep learning network, using a set of prior chosen training images. These features are then extracted from incoming images and then classified.
In various embodiments, an incoming imaging study is processed to first select frames depicting CW Doppler pattern. A plurality of frames of a medical video are read. A mode label indicative of a mode of each of the plurality of frames is determined. A set of features discriminating between mode labels are learned using a deep learning network using a set of prior chosen training images. These features are then extracted from incoming images and classified. The images classified into CW Doppler mode labels are then retained. From among the frames depicting CW Doppler patterns, a set of frames are selected that depict the valve of interest. A region of interest is extracted in each CW Doppler image and features discriminating between different heart valves are then learned using another deep learning network using prior chosen training CW Doppler images. New CW Doppler images are classified using the learned network and those images that are classified as containing the target valve of interest are retained. A Doppler envelope is then extracted from the selected frame. Based on the frame and the Doppler envelope, one or more measurements is extracted indicative of a disease condition from those of the at least one of the plurality of frames matching a predetermined valve label.
Referring to
In various embodiments, retrospective review includes searching for electronic medical records relevant to the disease condition. This includes problem list and recorded diagnosis, clinical reports and imaging studies. Given a target disease for whom the discrepancy needs to be detected, and a time period over which the discrepancy needs to be detected, a set of encounters are selected from the patient records corresponding to the specified time period. The imaging studies, and their associated reports are then drawn. The problem list and recorded diagnosis associated with these encounters are drawn. The disease mention in the problem list and recorded diagnosis is noted. Disease-specific features are extracted from imaging studies and reports. The disease specific features along with indications for the disease from problem lists are fed to the classifier to predict patients that are likely to have the disease. From the list of patients identified by the predictor classifier, their encounter records are now analyzed for discrepancy. Evidence from each step in the workflow is compared to the next step with respect to the indications for the disease. In spotting a discrepancy the positive occurrence of disease mention is used. Similarly in spotting a discrepancy in modality data (imaging), established guidelines for exceeding the values for normal ranges of measurements are utilized. The order of analysis of discrepancy in the workflow begins with the earliest evidence in imaging, then to technician measurements embedded in imaging, followed by imaging reports and then problem list and finally the recorded diagnosis in the system. With this order, the earliest place in the electronic medical record where this discrepancy has occurred is noted. The definition of a discrepancy is simply that the earlier stage in the workflow says ‘yes’ to the disease while the next consecutive stage in the workflow says ‘no’.
Referring now to
In computing node 10 there is a computer system/server 12, which is operational with numerous other general purpose or special purpose computing system environments or configurations. Examples of well-known computing systems, environments, and/or configurations that may be suitable for use with computer system/server 12 include, but are not limited to, personal computer systems, server computer systems, thin clients, thick clients, handheld or laptop devices, multiprocessor systems, microprocessor-based systems, set top boxes, programmable consumer electronics, network PCs, minicomputer systems, mainframe computer systems, and distributed cloud computing environments that include any of the above systems or devices, and the like.
Computer system/server 12 may be described in the general context of computer system-executable instructions, such as program modules, being executed by a computer system. Generally, program modules may include routines, programs, objects, components, logic, data structures, and so on that perform particular tasks or implement particular abstract data types. Computer system/server 12 may be practiced in distributed cloud computing environments where tasks are performed by remote processing devices that are linked through a communications network. In a distributed cloud computing environment, program modules may be located in both local and remote computer system storage media including memory storage devices.
As shown in
Bus 18 represents one or more of any of several types of bus structures, including a memory bus or memory controller, a peripheral bus, an accelerated graphics port, and a processor or local bus using any of a variety of bus architectures. By way of example, and not limitation, such architectures include Industry Standard Architecture (ISA) bus, Micro Channel Architecture (MCA) bus, Enhanced ISA (EISA) bus, Video Electronics Standards Association (VESA) local bus, and Peripheral Component Interconnect (PCI) bus.
Computer system/server 12 typically includes a variety of computer system readable media. Such media may be any available media that is accessible by computer system/server 12, and it includes both volatile and non-volatile media, removable and non-removable media.
System memory 28 can include computer system readable media in the form of volatile memory, such as random access memory (RAM) 30 and/or cache memory 32. Computer system/server 12 may further include other removable/non-removable, volatile/non-volatile computer system storage media. By way of example only, storage system 34 can be provided for reading from and writing to a non-removable, non-volatile magnetic media (not shown and typically called a “hard drive”). Although not shown, a magnetic disk drive for reading from and writing to a removable, non-volatile magnetic disk (e.g., a “floppy disk”), and an optical disk drive for reading from or writing to a removable, non-volatile optical disk such as a CD-ROM, DVD-ROM or other optical media can be provided. In such instances, each can be connected to bus 18 by one or more data media interfaces. As will be further depicted and described below, memory 28 may include at least one program product having a set (e.g., at least one) of program modules that are configured to carry out the functions of embodiments of the invention.
Program/utility 40, having a set (at least one) of program modules 42, may be stored in memory 28 by way of example, and not limitation, as well as an operating system, one or more application programs, other program modules, and program data. Each of the operating system, one or more application programs, other program modules, and program data or some combination thereof, may include an implementation of a networking environment. Program modules 42 generally carry out the functions and/or methodologies of embodiments of the invention as described herein.
Computer system/server 12 may also communicate with one or more external devices 14 such as a keyboard, a pointing device, a display 24, etc.; one or more devices that enable a user to interact with computer system/server 12; and/or any devices (e.g., network card, modem, etc.) that enable computer system/server 12 to communicate with one or more other computing devices. Such communication can occur via Input/Output (I/O) interfaces 22. Still yet, computer system/server 12 can communicate with one or more networks such as a local area network (LAN), a general wide area network (WAN), and/or a public network (e.g., the Internet) via network adapter 20. As depicted, network adapter 20 communicates with the other components of computer system/server 12 via bus 18. It should be understood that although not shown, other hardware and/or software components could be used in conjunction with computer system/server 12. Examples, include, but are not limited to: microcode, device drivers, redundant processing units, external disk drive arrays, RAID systems, tape drives, and data archival storage systems, etc.
The present invention may be a system, a method, and/or a computer program product. The computer program product may include a computer readable storage medium (or media) having computer readable program instructions thereon for causing a processor to carry out aspects of the present invention.
The computer readable storage medium can be a tangible device that can retain and store instructions for use by an instruction execution device. The computer readable storage medium may be, for example, but is not limited to, an electronic storage device, a magnetic storage device, an optical storage device, an electromagnetic storage device, a semiconductor storage device, or any suitable combination of the foregoing. A non-exhaustive list of more specific examples of the computer readable storage medium includes the following: a portable computer diskette, a hard disk, a random access memory (RAM), a read-only memory (ROM), an erasable programmable read-only memory (EPROM or Flash memory), a static random access memory (SRAM), a portable compact disc read-only memory (CD-ROM), a digital versatile disk (DVD), a memory stick, a floppy disk, a mechanically encoded device such as punch-cards or raised structures in a groove having instructions recorded thereon, and any suitable combination of the foregoing. A computer readable storage medium, as used herein, is not to be construed as being transitory signals per se, such as radio waves or other freely propagating electromagnetic waves, electromagnetic waves propagating through a waveguide or other transmission media (e.g., light pulses passing through a fiber-optic cable), or electrical signals transmitted through a wire.
Computer readable program instructions described herein can be downloaded to respective computing/processing devices from a computer readable storage medium or to an external computer or external storage device via a network, for example, the Internet, a local area network, a wide area network and/or a wireless network. The network may comprise copper transmission cables, optical transmission fibers, wireless transmission, routers, firewalls, switches, gateway computers and/or edge servers. A network adapter card or network interface in each computing/processing device receives computer readable program instructions from the network and forwards the computer readable program instructions for storage in a computer readable storage medium within the respective computing/processing device.
Computer readable program instructions for carrying out operations of the present invention may be assembler instructions, instruction-set-architecture (ISA) instructions, machine instructions, machine dependent instructions, microcode, firmware instructions, state-setting data, or either source code or object code written in any combination of one or more programming languages, including an object oriented programming language such as Smalltalk, C++ or the like, and conventional procedural programming languages, such as the “C” programming language or similar programming languages. The computer readable program instructions may execute entirely on the user's computer, partly on the user's computer, as a stand-alone software package, partly on the user's computer and partly on a remote computer or entirely on the remote computer or server. In the latter scenario, the remote computer may be connected to the user's computer through any type of network, including a local area network (LAN) or a wide area network (WAN), or the connection may be made to an external computer (for example, through the Internet using an Internet Service Provider). In some embodiments, electronic circuitry including, for example, programmable logic circuitry, field-programmable gate arrays (FPGA), or programmable logic arrays (PLA) may execute the computer readable program instructions by utilizing state information of the computer readable program instructions to personalize the electronic circuitry, in order to perform aspects of the present invention.
Aspects of the present invention are described herein with reference to flowchart illustrations and/or block diagrams of methods, apparatus (systems), and computer program products according to embodiments of the invention. It will be understood that each block of the flowchart illustrations and/or block diagrams, and combinations of blocks in the flowchart illustrations and/or block diagrams, can be implemented by computer readable program instructions.
These computer readable program instructions may be provided to a processor of a general purpose computer, special purpose computer, or other programmable data processing apparatus to produce a machine, such that the instructions, which execute via the processor of the computer or other programmable data processing apparatus, create means for implementing the functions/acts specified in the flowchart and/or block diagram block or blocks. These computer readable program instructions may also be stored in a computer readable storage medium that can direct a computer, a programmable data processing apparatus, and/or other devices to function in a particular manner, such that the computer readable storage medium having instructions stored therein comprises an article of manufacture including instructions which implement aspects of the function/act specified in the flowchart and/or block diagram block or blocks.
The computer readable program instructions may also be loaded onto a computer, other programmable data processing apparatus, or other device to cause a series of operational steps to be performed on the computer, other programmable apparatus or other device to produce a computer implemented process, such that the instructions which execute on the computer, other programmable apparatus, or other device implement the functions/acts specified in the flowchart and/or block diagram block or blocks.
The flowchart and block diagrams in the Figures illustrate the architecture, functionality, and operation of possible implementations of systems, methods, and computer program products according to various embodiments of the present invention. In this regard, each block in the flowchart or block diagrams may represent a module, segment, or portion of instructions, which comprises one or more executable instructions for implementing the specified logical function(s). In some alternative implementations, the functions noted in the block may occur out of the order noted in the figures. For example, two blocks shown in succession may, in fact, be executed substantially concurrently, or the blocks may sometimes be executed in the reverse order, depending upon the functionality involved. It will also be noted that each block of the block diagrams and/or flowchart illustration, and combinations of blocks in the block diagrams and/or flowchart illustration, can be implemented by special purpose hardware-based systems that perform the specified functions or acts or carry out combinations of special purpose hardware and computer instructions.
The descriptions of the various embodiments of the present invention have been presented for purposes of illustration, but are not intended to be exhaustive or limited to the embodiments disclosed. Many modifications and variations will be apparent to those of ordinary skill in the art without departing from the scope and spirit of the described embodiments. The terminology used herein was chosen to best explain the principles of the embodiments, the practical application or technical improvement over technologies found in the marketplace, or to enable others of ordinary skill in the art to understand the embodiments disclosed herein.
Number | Name | Date | Kind |
---|---|---|---|
8280750 | Krishnan et al. | Oct 2012 | B2 |
20070150025 | Dilorenzo et al. | Jun 2007 | A1 |
20080242996 | Hall | Oct 2008 | A1 |
20120250957 | Syeda-Mahmood et al. | Oct 2012 | A1 |
20130262357 | Amarasingham et al. | Oct 2013 | A1 |
20130290231 | Chbat | Oct 2013 | A1 |
20130345563 | Stuebe | Dec 2013 | A1 |
20140233826 | Agaian | Aug 2014 | A1 |
20160026915 | Delp et al. | Jan 2016 | A1 |
20160232658 | Syeda-Mahmood | Aug 2016 | A1 |
20180225820 | Liang et al. | Aug 2018 | A1 |
20180259608 | Golden et al. | Sep 2018 | A1 |
Number | Date | Country |
---|---|---|
105574337 | May 2016 | CN |
2628113 | Aug 2013 | EP |
Entry |
---|
T. Syeda-Mahmood, et al., Shape-based Similarity Retrieval of Doppler Images for Clinical Decision Support, 2010, IEEE (Year: 2010). |
Y. Zhang, S. Nitter-Hauge, Determination of the mean pressure gradient in aortic stenosis by Doppler echocardiography, 1985, 6, 999-1005 (Year: 1985). |
Number | Date | Country | |
---|---|---|---|
20180107787 A1 | Apr 2018 | US |