Right ventricular (RV) dysfunction due to pulmonary hypertension, acute myocardial infarction, and left ventricular assist device-induced hemodynamic changes has limited the effectiveness of mechanical circulatory support therapy in heart failure patients. Right ventricular (RV) dysfunction can result as a sequelae of pulmonary hypertension, myocardial infarction, and acute/chronic volume or pressure overload conditions. Mechanical circulatory support (MCS) devices, specifically left ventricular assist devices (LVADs), have extended the lives of many adults suffering from end-stage congestive heart failure (HF). However, LVAD-induced right heart dysfunction is a problem that has limited the effectiveness of MCS therapy in the HF population (Dang et al., J Heart Lung Transplant. 2006, 25(1):1-6). Some researchers have reported up to 30-40% of heart failure patients with LVADs have developed some degree of right heart dysfunction regardless of the type of device used (pulsatile versus continuous flow; Patel et al., Ann Thorac Surg. 2008, 86(3):832-840). The majority of patients that develop right heart failure are relegated to drug therapy. Several groups have used currently available LVADs to support the RV with mixed results (Bernhardt et al., Eur J Cardiothorac Surg. 2015, 48(1):158-162; Potapov et al., ASAIO J. 2012, 58(1):15-18). Despite the potential of RVAD therapy, the development of right ventricular assist devices (RVADs) has lagged significantly compared to LVAD technology. An RVAD that can be deployed without a sternotomy and that can provide safe pulmonary circulatory support would be ideal for patients that develop right heart failure.
Percutaneous intravascular devices offer the potential to support the failing RV without the need for an extensive surgical procedure. Percutaneous blood pumps are devices that can be implanted via catheter-based procedures and are typically used clinically to provide partial support (2.5-5 L/min) for patients with acute HF. Recently, Stretch et al. showed that the use of percutaneous intravascular devices for short-term MCS in patients with acute HF has increased approximately 10 times between 2007 and 2011 (Stretch et al., Journal of the American College of Cardiology. 2014, 64(14):1407-1415). During this period, hospitals saw a decrease in mortality and morbidity and a decrease in hospital costs. Percutaneous pumps have already been successfully used as right ventricular assist devices in the setting of acute right ventricular failure (Kapur et al. The Journal of Heart and Lung Transplantation. 2011, 30(12):1360-1367; Cheung et al., J Heart Lung Transplant. 2014, 33(8):794-799). In these clinical studies, the percutaneous pump provided increased patient cardiac index, reduced patient central venous pressure, and mediated recovery of RV function. All of these effects facilitated RV recovery and eventual device explantation in some patients. Computer simulation studies also suggest that RVADs, in most circumstances, only need to provide a modest 1.5 to 2 L/min in additional flow to benefit patient hemodynamics (Punnoose et al., Progress in cardiovascular diseases. 2012, 55(2):234-243.e232). Despite this potential paradigm shift in RV dysfunction therapy, percutaneous pump technology is still limited to short-term use (a few hours) because of the need for a driveline to power the device and the need for purge sealing system that cools the pump motor and provides a seal between the motor-shaft and impeller interface (Butler et al., IEEE Trans Biomed Eng. 1990, 37(2):193-196; Rosarius et al., Artif Organs. 1994, 18(7):512-516; Siess et al., Artif Organs. 2001, 25(5):414-421). Together, the purging fluid line and driveline exit the patient's vasculature and limits patient mobility.
Thus, there is a need in the art for novel right ventricular assist devices (RVADs), in particular RVADs that can be deployed without a sternotomy and that can provide safe pulmonary circulatory support for patients that develop right heart failure. There is also a need in the art for novel RVADs featuring axial magnetic couplings which can help to eliminate the seal, and sealing system, typically needed to isolate the motor and bearings from blood contact. The present invention satisfies these unmet needs.
In one aspect, the invention relates to an implantable device for transferring a bodily fluid between two anatomically distinct locations in a subject, comprising: a pump unit having an inflow port and an outflow port; at least one anchoring structure associated with the pump unit; and a conduit having first and second ends, the first end connected to the outflow port of the pump unit, and the second end having an outflow port. In one embodiment, the pump unit has a substantially cylindrical cross section, and a diameter between about 1 mm and about 20 mm. In another embodiment, the pump unit comprises a motor having a motor shaft, an impeller, a casing, and a diffuser. In one embodiment, the impeller is attached to the motor shaft. In another embodiment, the device further comprises a drive magnet and a following magnet, wherein the drive magnet is connected to the motor shaft, and the following magnet is connected to the impeller. In one embodiment, the device is a catheter-deliverable cavo-arterial pump (CAP). In another embodiment, the device is a catheter-deliverable right ventricular assist device (RVAD). In one embodiment, the anchoring structure comprises at least a strut comprising a nonferromagnetic flexible material. In another embodiment, the pump unit comprises a cable for transfer of power and data to and from the device. In another embodiment, the conduit comprises an optional cannula.
In another aspect, the invention relates to a method of assisting right ventricular circulation in a subject, comprising: placing the device of claim 1 in the vasculature of the subject, wherein the pump unit is anchored to the wall of the inferior vena cava (IVC) of the subject, and the outflow port of the conduit is placed in the main pulmonary artery of the subject; and directing blood flow through the device, from the IVC of the subject to the main pulmonary artery of the subject. In one embodiment, the conduit passes through the right atrium and the right ventricle of the subject. In one embodiment, the pump unit comprises a motor having a motor shaft, an impeller, a casing, and a diffuser, wherein the impeller is attached to the motor shaft. In another embodiment, the pump unit comprises a motor having a motor shaft, an impeller, a casing, a diffuser, a drive magnet, and a following magnet, wherein the drive magnet is connected to the motor shaft and the following magnet is connected to the impeller. In one embodiment, the pump unit comprises a cable for transfer of power and data to and from the device. In another embodiment, the conduit comprises an optional cannula. In another embodiment, the anchoring structure comprises at least a strut comprising a nonferromagnetic flexible material. In one embodiment, the blood flow is between about 0 and about 5 L/min. In another embodiment, the pressure head is between about 5 mmHg and about 100 mmHg. In another embodiment, the impeller speed is between about 5 kRPM and about 30 kRPM.
The following detailed description of preferred embodiments of the invention will be better understood when read in conjunction with the appended drawings. For the purpose of illustrating the invention, there are shown in the drawings embodiments which are presently preferred. It should be understood, however, that the invention is not limited to the precise arrangements and instrumentalities of the embodiments shown in the drawings.
The invention relates in part to a cavo-arterial pump (CAP), functioning as a right ventricular assist device (RVAD), which is an intravascular blood pump designed to provide pulmonary circulatory support for patients that develop RV dysfunction, in particular LVAD-induced RV dysfunction. The pump features either a direct drive pump mechanism, or a magnetic drive pump mechanism. The magnetic drive mechanism eliminates the need for an external purge seal line by utilizing permanent magnet magnetic bearings or magnetic couplings, enabling the development of fully implantable intravascular pumps.
An intravascular pump of the invention can provide sufficient pulmonary support, i.e., up to about 2.25 L/min. In a magnetic drive pump of the invention, including magnets with an about 90 degree offset separated by an about 2.5 mm gap, the coupling can provide up to about 6 mNm of torque. The couplings can be spaced up to 4 mm apart before torque transmission falls below the motor output. The magnetic drive CAP was able to operate at up to about 18.5 kRPM, and produce a maximum flow rate of about 1.35 L/min and a maximum pressure head of about 40 mm Hg. In addition, computational fluid dynamic (CFD) simulations show that the pump can provide flow between 1.4-3 L/min of flow at venous pressures (0-30 mmHg) when the motor is run between 10 kRPM and 22 kRPM.
Unless defined otherwise, all technical and scientific terms used herein have the same meaning as commonly understood by one of ordinary skill in the art to which this invention belongs. Although any methods and materials similar or equivalent to those described herein can be used in the practice or testing of the present invention, the preferred methods and materials are described.
As used herein, each of the following terms has the meaning associated with it in this section.
The articles “a” and “an” are used herein to refer to one or to more than one (i.e., to at least one) of the grammatical object of the article. By way of example, “an element” means one element or more than one element.
“About” as used herein when referring to a measurable value such as an amount, a temporal duration, and the like, is meant to encompass variations of ±20%, ±10%, ±5%, ±1%, or ±0.1% from the specified value, as such variations are appropriate to perform the disclosed methods.
Ranges: throughout this disclosure, various aspects of the invention can be presented in a range format. It should be understood that the description in range format is merely for convenience and brevity and should not be construed as an inflexible limitation on the scope of the invention. Accordingly, the description of a range should be considered to have specifically disclosed all the possible subranges as well as individual numerical values within that range. For example, description of a range such as from 1 to 6 should be considered to have specifically disclosed subranges such as from 1 to 3, from 1 to 4, from 1 to 5, from 2 to 4, from 2 to 6, from 3 to 6 etc., as well as individual numbers within that range, for example, 1, 2, 2.7, 3, 4, 5, 5.3, and 6. This applies regardless of the breadth of the range.
In one aspect, the invention relates to a minimally-invasive cavo-arterial pump device that can be positioned within the body of a subject to aid in the movement or pumping of a bodily fluid. For example, in certain instances, the device provides for movement of blood, urine, sweat, air, and the like. In a particular embodiment, the device aids or replaces ventricle function of the heart by moving blood past the right or left ventricle into the pulmonary or systemic circulation, respectively. In one embodiment, the placement of device 101 is as depicted in
In one embodiment, the invention provides right ventricular assist devices (RVADs) configured for minimally-invasive percutaneous delivery to the implantation site. The devices are capable of providing long-term support with overall hemodynamic performance and durability superior and comparable to current conventional therapeutic approaches to right ventricular assistance. The devices are constructed of durable materials allowing for long-term use. A device of the invention generally has dimensions that allow for its insertion and guidance through a blood vessel.
As shown in
As depicted in
Further attached to pump 201 is fluid conduit 206, for example a flexible tube, for blood transport, conduit 206 ending with outflow port 207 for blood delivery into main pulmonary artery 105. The flexible tube can have optional terminal cannula 208 also having outflow port 209. In one embodiment, pump 201 is about 10 mm in diameter and about 65 mm in length.
It should be appreciated that the casing, tubing 206, outflow port 207, and optional cannula 208 can be composed of any material, such as medical grade alloys or polymers. In some embodiments anchoring structures 205 can be composed of a nonferromagnetic, flexible, shape memory material, such as nitinol, a composite of nickel and titanium known for its superelasticity and ability to expand to a different shape. For example, in one embodiment, the struts are configured to expand at a temperature threshold at or near body temperature. It should be appreciated that any rigid, yet flexible material may be used, such as a medical grade alloy or polymer, so that struts can be compressed inwardly toward the body of pump 201 in a compressed state, creating an expanding bias which can be restrained for example by a delivery catheter. Once the restraint is removed, for example by removing the body of pump 201 from the delivery catheter, the expanding bias forces struts to return to their relaxed, expanded state. The medical grade materials described herein may also include an anti-thrombogenic coating or admixture to reduce the incidence of thrombus buildup, promoting hemocompatibility and the maintenance of high blood flow rates pass the pump. The material may also include a coating comprising an immunosuppressant, e.g., rapamycin (sirolimus).
The cavo-arterial pump of the invention can be designed to work with either a direct drive mechanism pump (
In either the direct drive mechanism pump (
As shown in the exploded view in
In a preferred embodiment, a pump of the invention includes an axial magnetic coupling utilizing permanent magnets, for example neodymium permanent magnets. An axial magnetic coupling offers the potential to eliminate the purge seal needed in intravascular pumps previously known in the art. As shown in the exploded view in
Similarly to direct drive pump, impeller 404 and diffuser 405 in the magnetic drive pump have a sapphire hemisphere and cup as axial bearings, respectively, but as readily apparent, any suitable type of hemisphere and cup bearings can be used. As readily apparent, the magnetic drive pump operates by the magnetic field coupling of magnets 402 and 410. When motor 401 rotates, drive magnet 402 will engage following magnet 410 through a magnetic field, and as a result following magnet 410 will rotate attached impeller 404. As readily apparent, the gap distance between drive magnet 402 and impeller following magnet 410 can vary, and is generally between about 0.05 mm to about 20 mm. In one embodiment, the gap distance between drive magnet 402 and impeller following magnet 410 is about 1 mm. In another embodiment, the gap distance between drive magnet 402 and impeller following magnet 410 is about 2.5 mm. In another embodiment, the gap between impeller magnet 410 and drive magnet 402 is reduced to the minimum limit allowed by fabrication tolerances.
In one embodiment, the pump of the invention has a motor capable of achieving various rotation speeds between 5 and 30 kRPM (thousands of rotations per minute). In various embodiments, the motor can rotate at 10.7 kRPM, 11 kRPM, 14.5 kRPM, 14.7 kRPM, 16 kRPM, 16.7 kRPM, 17.5 kRPM, 20 kRPM, and 24 kRPM, or any other suitable speed. As readily apparent, in a direct drive mechanism pump, the rotational speed of the impeller is identical to the rotational speed of the motor shaft.
For the magnetic drive mechanism pump, the rotational speed of the impeller is equal or less than the rotational speed of the motor shaft, and the relationship between the rotational speed of the impeller and the rotational speed of the motor is influenced by the gap distance between the drive magnet and the following magnet, and the physical properties of the liquid being pumped. For example, for a 3 mm separation, while pumping water, the impeller speed matches the motor rotational speed up to 21 kRPM, and above 21 kRPM the impeller rotational speed decreases with increasing motor shaft speed. Similarly, while pumping water, the maximum rotational speed in which the impeller matches the motor shaft speed in a magnetic drive pump is 20.3, 18.5, and 14.3 kRPM for 4 mm, 5 mm, and 6 mm gap distance magnet separation, respectively (
In either the direct drive mechanism pump (
In one embodiment, the device of the invention includes a power cable operably connected to the motor. In certain embodiments, the power cable, lead, or line, can be externalized from the device to outside of the body using known techniques. For example, in one embodiment, the power cable can be guided from the device through the superior vena cava and into the subclavian vein to an area over the right or left side of the chest, where a small incision can be made to retrieve the cable. In one embodiment, the pump can be powered via a transfemoral lead that exits the patient's femoral artery. In another embodiment, the power line exits the brachiocephalic vein, while the controller, backup battery, and a wireless powering coil resides in the infra-clavicular pocket.
The device of the invention can be operated with both wired and/or transcutaneous energy transfer (TET) power delivery systems. For the implementation of TET power delivery, a small superficial pocket is created just underneath the skin where a TET coil can be placed and connected to the power cable of the device. Exemplary TET power delivery systems, including systems that wirelessly deliver power to implantable devices, are described in U.S. patent application Ser. Nos. 13/843,884 and 14/213,256, each of which are incorporated by reference in their entirety.
In certain embodiments, the device of the invention is operably connected to a pump controller. The pump controller may be located exterior to a patient, or implanted within the patient. In certain embodiments, the pump controller delivers and receives signals from the device relating to function of the pump unit of the device. For example, the controller may provide signals relating to the control of pump speed, desired flow rate, type of flow produced (pulsatile vs. continuous), and the like. The controller may be directly wired to the device of the invention or may communicate wirelessly to the device.
In some embodiments, the device of the invention is controlled by an implantable controller that is sized and shaped to be implanted within the body of the user. The controller may comprise a power supply, or alternatively may be powered externally by a separate wired or wireless power source positioned outside the body of the user. In one embodiment, the invention may be powered by a wireless power system, such as a system as described in U.S. Pat. No. 8,299,652; U.S. Patent Application Publication No. 2013/0310630; Sample et al., 2011, IEEE Transactions, 58(2): 544-554; and Waters et al., 2012, Proceedings of the IEEE, 100(1): 138-149, the entire disclosures of which is incorporated by reference herein in their entireties.
In some embodiments, the controller is communicatively connected to an external control unit, which may comprise a smartphone, a desktop, a tablet, a wristwatch, or any suitable computing device known in the art. In addition to exercising control over the various functions of the CAP, the controller may receive data from one or more sensors. Examples of such data include an EKG signal, the current pump speed of the CAP, the current flow rate within the CAP, the power consumption of the CAP, pulse oximetry, or any other information relevant to the function of the CAP. In some embodiments, some or all of the collected data is presented as part of a user interface (UI) of the external control unit. In some embodiments, the UI may provide the user with the ability to modify the function of the controller, display information related to historical or real-time functionality of the CAP, and/or display historical or real-time information related to the user's cardiac function.
As would be understood by those skilled in the art, the external control unit may be directly connected via wires or wirelessly connected via any suitable radio-frequency, optical, or other wireless communication standard. In some embodiments, the external control unit may be physically far removed from the CAP and only in indirect communication with the CAP and/or the implantable controller, connected via one or more wireless networks, Ethernet switches, or the Internet. In some embodiments, control signals transmitted from the external control unit to the implantable controller are encrypted.
The present invention comprises a method of promoting the movement or flow of a body fluid. The method may be used to aid in the movement or pumping of any body fluid in any location within the body. For example, in certain embodiments, the method comprises delivery and implantation of the device described herein into the IVC to promote pumping of blood to the pulmonary artery. The device thereby provides long term RVAD function. In one embodiment, the method comprises inserting the device into the vasculature, and guiding the device through the vasculature to the implantation site. In some embodiments, the method comprises inserting a delivery catheter, loaded with the device of the invention, into the vasculature, and guiding the catheter and device to the implantation site. In one embodiment, the method comprises releasing the device from the delivery catheter at the implantation sit. In one embodiment, releasing the device from the catheter allows for one or more anchoring structures to expand into its relaxed state to allow for engagement of the vessel wall. In one embodiment, the method comprises anchoring the pump unit in the vessel wall in the terminal portion of the IVC. In one embodiment, the method comprises guiding the fluid conduit to the right atrium via the cavo-atrial opening, to the right ventricle via the tricuspid valve, and to the pulmonary artery via the pulmonary valve, such that the outflow port resides in the lower portion of the pulmonary artery. However, the device may be inserted at any suitable access site.
In one embodiment, the method comprises sending a signal to the pump unit of the device to start the motor and set the rotation speed. In another embodiment, the method comprises setting the rotation speed based on a set of sensor inputs measured by the device or other implanted or external devices. In one embodiment, the method comprises sending a signal to the pump unit of the device to stop pumping based on one or more sensor inputs. In one embodiment, the method comprises intermittently starting or stopping the rotation of the pump. In some embodiments, the method comprises running the pump continuously, but varying the speed of the pump over time according to a pre-determined pattern. The method of the present invention may further comprise adjusting the speed of the pump motor based on sensor data related to the performance of the pump. For example, in response to a measured impeller rotation rate provided by a hall effect sensor or other rotation speed sensor, a controller might adjust the driven speed of the motor in order to optimize efficiency.
The invention is further described in detail by reference to the following experimental examples. These examples are provided for purposes of illustration only, and are not intended to be limiting unless otherwise specified. Thus, the invention should in no way be construed as being limited to the following examples, but rather, should be construed to encompass any and all variations which become evident as a result of the teaching provided herein.
Without further description, it is believed that one of ordinary skill in the art can, using the preceding description and the following illustrative examples, make and utilize the present invention and practice the claimed methods. The following working examples therefore, specifically point out the preferred embodiments of the present invention, and are not to be construed as limiting in any way the remainder of the disclosure.
The intravascular pump designed is intended to provide partial circulatory support (2.5-3 L/min) to patients with LVAD-induced right ventricular dysfunction. The intravascular pump 101, which is called the cavo-arterial pump (CAP), would sit in the inferior vena cava 104 and propel venous blood to the main pulmonary artery 105 (
where y is the specific work, ΔP is the pressure head across the pump, and p is the density of blood. The specific work in this design was calculated to be 3.9 m2/s2.
The specific diameter was also calculated using:
where Ds is the impeller specific diameter, d is the impeller diameter, and Q is the desired flow rate. For a diameter of 7.5 mm and a flow rate of 2.5 L/min, the specific diameter calculated is 1.72. Using a Cordier diagram, the specific speed, Ns, was found to be 2 for this design.
Lastly, the rotational speed required to produce 2.5 L/min against a 30 mm Hg pressure head using a 7.5 mm impeller was calculated using:
Thus, the impeller speed required to produce 2.5 L/min against a 30 mm Hg pressure head is 24 kRPM. An AC motor capable of achieving rotational speeds above 24 kRPM was chosen for device fabrication.
Two pump prototypes were designed and fabricated. A direct drive pump, in which the impeller was attached directly to the motor shaft was fabricated. In addition, the same design was adapted to use a magnetic drive mechanism. Both designs consist of a brushless 10 mm diameter in-runner motor (Turnigy 1015, Hobby King USA LLC, Lakewood, Wash., USA), 4-blade impeller and diffuser, and 10 mm outer diameter pump housing with 4 side inlets and one outlet. The impeller and diffuser were designed on ANSYS® BladeModeler and converted to three-dimensional models in ANSYS® DesignModeler.
The computer aided design (CAD) model of the direct drive pump is shown in
A CAD model of the magnetic drive pump is shown in
A finite element model of two permanent magnet couplings was created on COMSOL Multiphysics® software (Burlington, Mass., USA) to estimate the range of torque values needed to rotate the impeller across an air gap. The model, shown in
M
x
=M
r cos(θ)
M
y
=M
r sin(θ)
A parametric sweep was carried out in which the gap between the magnets were changed from 3 mm to 6 mm (in 1 mm steps) and the angle, θ, was varied from 0° to 360° in 22.5° steps. The torque from the following magnet to the drive magnet was calculated for all these parameters. The torque calculations represent the maximum torque that can be transmitted by the magnetic couplings across the gap. Air was used as the surrounding medium. The wall separating the magnets and the working fluid were not taken into consideration in this model. The model consisted of 344,000 mesh elements.
The torque magnitude calculated from the finite element model at various angles and gap distances is shown in
The direct drive CAP was tested on a bench-top flow loop to test the performance. The flow loop, shown in
The performance of the direct drive CAP at different speeds in water is shown in
The direct drive CAP is capable of producing sufficient partial circulatory support in the pulmonary circulation of a right heart failure patient. As seen in
A second setup was fabricated to test the CAP driven with axial magnetic couplings. The magnetic drive CAP was tested on a bench-top flow loop similar to the direct drive CAP. The flow loop, shown in
The effectiveness of the magnetic coupling in the magnetically-driven CAP in water is shown in
The maximum flow rate produced by the magnetically-driven CAP as a function of the motor shaft speed for different air gaps is shown in
The performance of the magnetic drive CAP at different speeds in water is shown in
Axial magnetic couplings utilizing neodymium permanent magnets offer the potential to eliminate the purge seal needed in intravascular pumps like the Impella® RP, 2.5 and 5.0 pumps. This advances intravascular pump technology one step closer to fully implantable systems.
Even though magnetic couplings facilitate contactless torque transmission across narrow gaps, mechanical bearings are still needed to support the rotating impeller on both the inlet and outlet ends. Thus, careful consideration is needed in utilizing mechanical bearings that can support both high rotational impeller speeds and the attractive force produced between the coupling magnets. In addition, utilizing magnetic bearings necessitates small gaps between the pump housing and the rotating impeller. These narrow pathways may increase shear stress on the circulating blood, which may lead to hemolysis. Intravascular pump designs (those which are near animal testing and commercialization) that utilize magnetic couplings should be aimed at ensuring that these narrow gaps and the use of mechanical bearings do not promote blood damage. This can be studied by merging the magnetic finite element model presented in this paper with some fluid dynamics physics to estimate shear and axial forces on blood-like fluid. In addition, extensive hemolysis testing should be carried out when a pump design is nearly finalized.
While providing contactless torque transmission is necessary to eliminate the purge seal of intravascular pumps, the motor driveline still limits the use of intravascular pumps for long-term therapy. Researchers have proposed a technique to power an intravascular pump by providing a transfemoral lead that exits the patient's femoral artery (Clifton et al., The Journal of Heart and Lung Transplantation. 34(4):S177). While this technique has proven to be safe in animals, it still has the potential to lead to bleeding, infection, and thrombotic events that are traditionally associated with MCS drivelines. In addition, the study is statistically limited in the number of animals for which this method was tested. Thus, a roadmap for improving intravascular pump implantability by eliminating the purging seal system was provided. It is envisaged that the power line would exit the brachiocephalic vein with controller, backup battery and a wireless powering coil will reside in the infra-clavicular pocket, thus leveraging our prior work on wirelessly powered systems (Waters et al., ASAIO journal (American Society for Artificial Internal Organs: 1992) 2014, 60(1):31-37).
The disclosures of each and every patent, patent application, and publication cited herein are hereby incorporated herein by reference in their entirety. While this invention has been disclosed with reference to specific embodiments, it is apparent that other embodiments and variations of this invention may be devised by others skilled in the art without departing from the true spirit and scope of the invention. The appended claims are intended to be construed to include all such embodiments and equivalent variations.
This application claims priority to U.S. Provisional Patent Application No. 62/342,301, filed on May 27, 2016, the contents of which are incorporated by reference herein in its entirety.
Number | Date | Country | |
---|---|---|---|
62342301 | May 2016 | US |