The present invention relates generally to a surgical imaging system including a CT scanner that automatically determines a volume of interest of a patient.
It is sometimes desirable to be able to take a CT scan of a patient during surgery. For example, a surgeon may want to check the progress of the surgery (e.g., to determine whether a problem has been completely corrected or whether a tumor has been completely removed, etc).
For image-guided surgery, it is sometimes desirable to periodically update a pre-operative CT scan of the patient. The relevant volume of the patient may have shifted between the time the pre-operative image was taken and the time of surgery. This is especially true once surgery has begun. For example, in cranial surgery, a shape of an intracranial cavity changes as the surgeon gains access. Changes in the pre-operative image and the actual surgical subject introduce variations into the surgical process. In matters like intracranial surgery, the tolerance for variations is low, thus even small changes between the image and actual subject may cause problems and make the surgery less effective.
To solve this problem, a new, partial CT scan may be taken during surgery to update the previously received information. It is known that a baseline, pre-operative CT scan can be updated with a partial CT scan of a volume of interest in which an x-ray source is collimated to scan only the volume of interest. The partial CT scan is used in conjunction with the pre-operative CT scan (which includes volumes that have presumably not changed) to obtain a full CT image.
However, selecting the volume of interest can be time-consuming if the surgeon is required to find the volume of interest on the pre-operative CT scan. Also, if the patient and/or the CT scanner have moved, the relative locations of the CT scanner and patient must be determined and registered with the image guided surgical system before the location of the volume of interest can be determined.
A CT scanner automatically determines a volume of change of a patient based upon anatomical changes that can be determined by the CT scanner in one or more (but significantly less than a full set) of frames. During surgery, when an updated CT scan is requested, the CT scanner begins performing a scan of a patient using a full field of view. The CT scanner takes a series of two-dimensional initial images (or “frames”) of the patient from a plurality of angularly spaced positions about the patient. When a sufficient number of initial images have been obtained, the CT scanner first registers its location relative to the patient and the previous CT images based upon the initial images. In other words, based on the initial images, the CT scanner determines its position relative to the current position of the patient and the current position of the patient relative to the previous CT scan(s).
When operating in an automatic mode, the CT scanner then compares the initial images to a previous CT image (either a pre-operative CT scan, a previously-updated CT scan or generic data). Based upon the comparison, the CT scanner determines where changes to the anatomy have occurred (e.g., because of the surgery so far). The volume where changes have occurred, plus some defined margin, becomes the volume of interest. The CT scanner then collimates an x-ray source to perform an intra-operative updated CT scan of only the volume of interest. The CT scanner updates the previous CT scan(s) with the new images of the volume of interest from the intra-operative updated CT scan to create a fully updated CT image, reducing x-ray exposure of the patient.
Alternatively, the surgeon can manually select the volume of interest on a previously-stored CT image using a computer user interface (graphic user interface or voice-activated, etc). In this example, the volume of change has been automatically determined by the CT scanner. The CT scanner then determines its location relative to the patient based upon the initial images. Then, with its location registered relative to the patient, the CT scanner collimates the x-ray source and focuses on the manually-defined volume of interest and completes the intra-operative updated CT scan of the volume of interest.
Preferably, the CT scanner provides a user interface that offers the surgeon the option of performing either the fully automatic volume of interest determination update scan or a surgeon-selected volume of interest determination update scan with automatic registration of the volume of change by the CT scanner.
Alternately, the pre-operative data and the initial images can be taken using a lower resolution and optionally, a low x-ray dosage. The CT scanner generates lower resolution images by downsampling the information from a detector. After the volume of interest is determined, the x-ray source is collimated and a CT scan of the volume of interest is taken at a higher resolution than the pre-operative data and the initial images.
As shown schematically in
Returning to
During surgery, the CT scanner 10 takes intra-operative CT scans of the volume of interest 59 within the patient P so that the surgeon (or a dentist) can determine the current progress of the surgery (e.g., has a tumor been completely removed or a sinus cavity been completely repaired?) The CT scanner 10 only performs a complete CT scan of the volume of interest 59, which is the volume (or volumes) where the surgeon is working. The volume of interest 59 is defined as the volume of change 57 in the anatomy of the patient P as detected, plus some margin.
The computer 30 uses the pre-operative data surrounding the volume of interest 59 in conjunction with the new information from the intra-operative CT scans to create a fully updated three-dimensional CT image. Therefore, a new, full intra-operative CT scan is not required to form a CT image of the volume of interest 59. The smaller scan of the volume of interest 59 also reduces the dosage of x-rays experienced by the patient P.
When a surgeon determines that an updated CT image is needed, a fully automatic updated CT scan or a manually designated updated CT scan of the volume of interest 59 can be requested. The user interface offers the surgeon the option of performing either the fully automatic volume of interest determination update scan or an updated CT scan of a manually selected volume of interest, both with automatic registration by the CT scanner 10. The request can be made by either using a graphical or voice-activated user interface on the computer 30.
When a fully automatic updated CT scan is requested, the CT scanner 10 takes a series of full field of view, two-dimensional CT images (initial images) from a plurality of angularly separated positions about the patient P (as shown in
The initial images provide two primary purposes: 1) to determine the position of the CT scanner 10 relative to the patient P (who may have been moved during the surgery); and 2) to determine the location of the volume of interest 59. Additionally, the initial images (or portions of them) may be used to perform the update.
Based upon the initial images, the CT scanner 10 first registers its location relative to the patient P (who may have moved during surgery) and the previous CT scans. This can be done by locating and orienting a known structure in this part of the patient's P anatomy (e.g., part of the skull) in the initial images. The CT scanner 10 then compares the initial images to a previous CT image (the pre-operative scan, a previously-updated scan or generic data). Based upon the comparison, the CT scanner 10 determines where changes to the patient's P anatomy have occurred (e.g., because of the surgery so far). The volume of change 57, plus some defined margin, becomes the volume of interest 59.
The CT scanner 10 then collimates the x-ray source 20 and takes a plurality of images at a plurality of angularly-spaced positions to perform the intra-operative updated CT scan of the volume of interest 59, as shown in
Alternatively, the surgeon can select “manual designation” to manually select the volume of interest 59 on the previously-stored CT image using software on the computer 30. After the CT scanner 10 locates the volume of change 57 and displays it to the surgeon, the surgeon can circle the volume of interest 59 on a three dimensional CT image on the computer 30 by using the mouse 38 or the keyboard 40. That is, the CT scanner 10 determines the volume of change 57, and the surgeon selects the size of the volume of interest 59. The CT scanner 10 then determines its location relative to the patient P based upon the initial images. Then, with its location registered relative to the patient P, the CT scanner 10 focuses in on the volume of interest 59 and completes the updated CT scan.
If more than one volume of change 57 is detected in the initial images, the CT scanner 10 could present the surgeon with the option of choosing one or more of the volumes of change 57 as the volume of interest(s) 59. If more than one volume of interest 59 is selected, the CT scanner 10 could then collimate to obtain images for each of the volumes of interest 59 in alternating frames as the CT scanner 10 rotates around the patient P. The CT scanner 10 could also perform multiple rotations about the patient P (or multiple 180 degree scans, or anywhere between 180 and 360 degrees for each scan).
The CT scanner 10 then automatically (i.e., without further prompting or input) displays the volume of interest 59 on the display 36. If more than one volume of interest 59 was selected, the CT scanner 10 marks the locations of the volumes of interest 59 so the surgeon can easily toggle or scroll between the volumes of interest 59.
The feature of registering the location of the CT scanner 10 relative to the current patient location could be used independently of the automatic determination of the volume of interest 59 feature, and vice versa.
The intra-operative updated CT scan of the volume of interest 59 could be performed at a different resolution than the pre-operative scan or data and the initial images. The pre-operative data is used only for background information and calculations required in creating a new image and has lesser importance in the new image than does the intra-operative data. The pre-operative CT scan may use a lower resolution and optionally, a low x-ray dosage, than the intra-operative scan, resulting in a safer pre-operative scan for the patient P and a cost savings in obtaining the pre-operative scan. The pre-operative CT scan is also a smaller computer file and therefore takes up less space and uses less processing power.
When desired, the CT scanner 10 generates lower resolution images by downsampling the information from the detector 22. Downsampling includes any of several methods for reducing the resolution of the information from the detector 22. In one example, a certain percentage of the pixels are ignored. For example, every other pixel or every third pixel, etc. is sampled Another way of downsampling is to average together the signal from adjacent pixels, such as an adjacent pair or a small array of four or more pixels, and then to treat it as a single pixel of information. Information from adjacent pixels can be statistically combined in many different ways besides averaging. The amount of downsampling (or not downsampling at all) can be varied by the CT scanner 10. The amount of downsampling (the resolution of the image) can even be varied within an image, as controlled by the computer 30, such that selected volumes of the image are at a higher resolution than the remainder of the image. In this manner, for example, a volume of interest 59 within the patient P can be recorded at a high resolution, without unnecessarily increasing the image file size for the entire image. Varying the resolution of the image can be used in several different ways, as will be explained below.
For example, the CT scanner 10 takes a plurality of downsampled (low resolution) initial images at the full field of view to determine the volume of interest 59. After the volume of interest 59 is located on the downsampled initial images, the x-ray source 20 is collimated, and a CT scan of the volume of interest 59 is taken at a higher resolution (or higher dosage) than the pre-operative images and the initial images.
Alternatively, with the x-ray source 20 collimated to the volume of interest 59, the volume of change 57 within the volume of interest 59 where changes have occurred could be scanned at a high resolution and the remainder of the volume of interest 59 could be scanned at low resolution. As the x-ray source 20 is collimated, the patient P is exposed to less x-rays. The CT scanner 10 takes a plurality of images at a plurality of angularly-spaced positions while the x-ray source 20 collimated.
In another example, the dataset required to form a multi-resolution image could be taken during a single scan by adapting the x-ray flux spatially (i.e., a low dose for lower resolution volumes and a higher dose for higher resolutions volume). The CT scanner 10 would employ automatic volume of interest 59 localization.
The foregoing description is only exemplary of the principles of the invention. Many modifications and variations are possible in light of the above teachings. It is, therefore, to be understood that within the scope of the appended claims, the invention may be practiced otherwise than using the example embodiments which have been specifically described. For that reason the following claims should be studied to determine the true scope and content of this invention.
This application claims priority to U.S. Provisional Application Nos. 60/786,638 filed Mar. 28, 2006 and 60/851,196 filed Oct. 12, 2006.
Number | Date | Country | |
---|---|---|---|
60786638 | Mar 2006 | US | |
60851196 | Oct 2006 | US |