The production of ultrathin cryosections (dry cutting) of usually, but not exclusively, biological specimens has proved a critical phase in the examination of objects in transmission electron microscopy (TEM), which determines the success of the process. In TEM, only very thin solid bodies can be examined. A reasonable thickness of a biological specimen is less than 200 nm. In order to convert a biological specimen into a solid body, in the traditional process the specimen is fixed, normally by aldehyde, the water is replaced with solvent and the specimen is embedded in a polymer by substitution of the solvent, and then cut ultrathin on water. With this method, the specimen loses all water, i.e. 60% to 80% of the mass of a biological specimen is lost and artefacts are necessarily induced. The alternative is to vitrify the specimens, i.e. freeze the specimens rapidly so that the water in the biological structures does not form ice crystals during cooling, and from this produce cryosections which are examined on a TEM cryoholder at around 100° K. (Al-Amoudi et al., EMBO J. 2004, 23(18):3583-8).
According to this process, ultrathin cryosections are produced as follows: The specimen must be vitrified, i.e. be extremely viscous or quasi-solid at 100° K., and free from ice crystals, because specimens which contain ice crystals are brittle and almost impossible to cut. For preparation, the specimen is frozen in or on a specimen holder or reliably fixed by other means. The specimen which is prepared in this way is fixed in a pre-cooled cryo-ultramicrotome (usually at 110° K.) for cryosectioning. The microtome arm is introduced into the cryostat where it collects the vitrified specimen and controls the advance and cutting movement.
The specimen is then trimmed with a special trimming diamond knife. The result of trimming is normally a truncated pyramid which consists of the biological material of interest. The top square of the truncated pyramid has a side length of around 150 μm and protrudes by approximately 30 μm.
After trimming, the specimen is normally cut ultrathin at 110° K. using a diamond knife with a nominal advance of around 50 nm (layer thickness 50-70 nm), with a cutting speed of for example 1 mm s−1. This cryosectioning causes a compression of the sections of up to 50% in the cut direction, i.e. the volume of the section remains constant, the edge length of the cut square lying in the cutting direction is reduced by half, and the section thickness is twice as thick. By ionizing the cutting environment (distance from knife edge around 15-20 mm) with a commercially available ionizer (for example CRION for Leica EM FC 7 from Leica-Microsystems or Static-Line Ionizer from Diatome US), the compression can be reduced but is not totally prevented.
The vitrified specimens form section ribbons on the diamond knife edge, i.e. the first section adheres to the second section and is pushed down by this on the knife surface. When the short section ribbon succeeds in adsorbing the first sections (which is not normally achievable permanently) by mounting on a fibre tip, for example a human eyelash, or an animal hair (for example a Dalmatian dog fur hair or guinea pig fur hair), a longer section ribbon with a plurality of individual sections can be produced if the section ribbon is guided away from the cut edge by this fibre. Alternatively, the fibre can catch on an often self-forming loop of the short section ribbon. This too does not give a permanent connection between the fibre and the section ribbon. Both these approaches are very demanding, difficult steps in the production of cryosections since the fibre is guided manually, which requires great concentration with fine motor control and can usually be performed successfully only by a few operators.
Alternatively, the fibre can be attached to the end of a suitably dimensioned moulding (e.g. rod) of heat-insulating material, which in turn can be clamped in the arm of a commercial micromanipulator (for example Micromanipulator-M by Leica Microsystems GmbH). This allows the fibre to be moved with precision control in the microregion using the manipulator (M. S. Ladinsky et al., Journal of Microscopy 224 (2006), 130). Fixing the first section to the fibre remains difficult. A permanent connection between the fibre and the section ribbon is not possible.
The section ribbon which is prepared in this way is then attached to a specimen holder which is suitable for TEM, usually with a grid with a mesh width of 50-100 μm. This step is usually carried out manually: the operator guides the grid, for example a 200-mesh molybdenum-based grid coated with a carbon film (for example EM Sciences, Hatfield, USA) below the section ribbon, which is still connected to the diamond knife edge, and makes contact with this by moving the grid vertically from bottom to top.
Alternatively, after reaching the desired length, the section ribbon which is guided by a micromanipulator with the fibre can be brought into contact with the grid by lowering in the vertical direction with the micromanipulator. A second fibre is then used to slide the section ribbon onto the grid, or press this as required, and finally it is separated from the rest of the section ribbon as closely as possible to the diamond knife (Ladinsky et al., ibid., page 130 and 132, FIG. 2B).
According to Ladinsky et al., the adhesion of the section ribbon to the grid is achieved by mechanical pressing. This method is not very efficient; often sections are lost. Pierson et al. (2010, J Struct Biol. 1692: 220-221, in particular FIG. 1 page 221) therefore attach the sections to the grid using an ionizer (DE publication 10 2008 059 284 of Jul. 30, 2009 and CRION for Leica EM FC 7, both from Leica-Microsystems GmbH, or Static-Line Ionizer by Diatome US). In normal ionization (discharge mode), the electrode which is installed close to the knife edge causes a reduction in the electrostatic effect (which for example is responsible for the often strong adsorption of the sections which are formed on the knife edge to the surface of the knife) and thus improves the sliding of the sections during or after the cutting process (Leica EM FC7 product information, Leica Microsystems GmbH 2013, page 7).
Pierson et al. found that section ribbons could be applied firmly to an electrically conductive, carbon-coated grid by ionization. For this, normal ionization (discharge mode) was switched off and an earthed grid was brought from below into contact with the section ribbon. At the time of contact of the section ribbon and grid, using the same ionizer in charge mode, a brief (less than 1 sec.) increased ionization pulse was triggered, causing the section ribbon to adhere irreversibly to the grid. In practice therefore, simply by switching on the ionizer (charge mode), preferably by means of a foot pedal, the operator can start the ionization on the surface of the section ribbon and hence attach the section ribbon to the grid (Pierson et al. 2010, J Struct Biol. 1692: 220-221, in particular FIG. 1 page 221, Leica EM FC 7 product information, pages 11/12).
The object of the present invention is to simplify and improve the handling of ultrathin section ribbons which are produced in ultramicrotomes.
Also using a micromanipulator, the application of the first section of a section ribbon to the fibre (hair) introduced by Ladinsky et al. is usually only possible with additional manual support. In comparison with this difficult manual preparation on the fibre, the object of the invention is therefore to reduce the manual requirements for the performing operator and thus ensure that even averagely-trained and skilled technical personnel are able to prepare perfect specimens of section ribbons in great quantities.
In relation to micromanipulation with a non-electrically-conductive fibre of a dielectric such as animal or human hair (eyelash) or a non-conductive plastic (nylon etc.), the object of the invention is to increase the speed and reproducibility of the preparation of section ribbons and integrate all necessary instruments into one easy-to-operate device.
This object is achieved by a device according to the features of claim 1 and by a method for its use according to the features of claim 21.
The device according to the invention substantially increases the reproducibility of the results of the preparation and practically completely eliminates the rejection level. The work sequences for preparation using the device according to the invention are simplified such that the preparation can be carried out reliably by averagely-trained technical personnel with average manual dexterity.
The method of handling section ribbons is greatly simplified with the section ribbon manipulator according to the invention and the grid holder which is also easy to manipulate. This manipulator precisely guides a tip, which is mounted to be easily exchanged on the end of the manipulator. This tip (fibre) for example consists of an animal or human hair (eyelash, bristle) or a fibre of a suitable dielectric material (for example plastic or glass). The surface of this fibre is made electrically conductive by coating (for example by vapour deposition in high vacuum or by treatment with a sputter source) with a suitable, electrically conductive, surface layer (preferably gold, another metal or carbon) and can thus be earthed. The fibre should be as soft as possible so that any contact with the delicate cutting edge during manipulation does not damage it. The use of a fine metal wire which increases in hardness and stiffness in the cold is therefore of only limited suitability.
The electrically conductive tip of the fibre is positioned behind the section ribbon which is forming on the knife of the microtome. Then using a commercially available ionizer (see above, CRION for Leica EM FC 7 by Leica Microsystems or Static-Line Ionizer by Diatome US), the section ribbon is attached to the earthed tip by increasing the ionization. The distance between the tip of the ionizer and the tip (fibre) of the section ribbon manipulator is preferably around 1.5 cm.
In comparison with a non-electrically-conductive fibre which is guided by the micromanipulator, the device according to the invention with the electrically conductive surface of the fibre has the advantage that the end of the section ribbon can be affixed to the earthed tip by the ionizer easily and substantially more quickly and more precisely, and gives a permanent attachment of the section ribbon to the tip of the fibre. The permanent connection between the conductive fibre and the section ribbon can only be undone by destruction, i.e. tearing off the sections which are not connected to the fibre from the sections which are attached to the fibre. With a fibre of dielectric material, no comparable permanent connection is possible.
This fixed connection between the conductive fibre and the section ribbon is of great benefit. For example, the isolation intensity in discharge mode, which can cause the section ribbon to oscillate, is not as critical as when the section ribbon is attached loosely with a non-conductive fibre. Also, with a permanent connection between the conductive fibre and section ribbon, it is possible to counter sudden section deformations which can for example be attributed to a not totally homogenous specimen, in that for example the tension of the fibre on the section ribbon can vary or its direction can change slightly. If the fibre is firmly connected with the short section ribbon, an unsuccessful guidance of the section ribbon which is formed by the sections produced by the microtome is almost excluded, since the micromanipulator with which the fibre tip is guided easily executes the necessary movements. If the fibre is not electrically conductive and cannot therefore be permanently connected with the section ribbon, the section ribbon can be detached from this non-conductive tip, which makes manipulation correspondingly more difficult and time-consuming.
When the resulting short section ribbon is firmly connected to the fibre tip by way of the electrostatic attraction, it is guided on the earthed tip of the section ribbon manipulator using a conventional micromanipulator so as to give a cohesive section ribbon as long as possible. The section ribbon usually has a length of slightly more than 3 mm and thus extends over the entire diameter of a conventional TEM grid.
A second micromanipulator is arranged on the opposite edge of the device, which can be rotated away around a vertical axis and carries a grid holder (see DE publication 10 2008 059 284 dated Jul. 30, 2009, FIGS. 3 and 4). The TEM grid which is fixed to this grid holder is first moved into the desired position vertically below the section ribbon, then brought upwardly from below into contact with this section ribbon, and then the section ribbon is connected to the grid by an ionization stroke in the “charge” mode. The grid holder allows the grid with the attached sections to be transferred very quickly into a commercial storage container (Gatan) which for example is prepared in liquid nitrogen.
The adhesion and guidance of the section ribbon with a non-conductive dielectric fibre (hair), which is difficult both manually and with the use of a micromanipulator, and the transfer of the section ribbon using the fibre onto the grid, which is carried out manually or with a micromanipulator, are substantially simplified by the device according to the invention. The advantage lies above all in the simplified and improved fixing of the cryosection ribbons to an electrically conductive tip, i.e. the creation of a permanent connection between the tip and the section ribbon. For selection and training of the operators, this gives the advantage that no highly developed fine motor control is required to produce cryosections successfully, since all demanding steps are carried out by micromanipulators which are easy to operate. Also, the fixing and exchange of the grid of a material suitable for TEM are simple and efficient. Since the grid can also be changed and manipulated simply and easily, a further advantage of the device according to the invention results.
When the section ribbon is applied to the TEM grid, it is transferred into liquid nitrogen as quickly as possible in order to minimise contamination with ice crystals. The result is a substantially lower contamination with ice crystals than with the conventional manual method. This is decisive for the success of the method since electrons cannot propagate through small ice crystals (diameter<0.5 microns). The grid holder transfers the grid with the section ribbon into a commercial storage container in a very simple manner.
As a result, the method according to the invention massively increases the chances of success and the operating reliability of the production of section ribbons, in particular of cryosection ribbons, and thus also increases the chances of success of examination of the section ribbons produced in this way by means of TEM.
The section ribbon manipulator according to the invention also allows easier handling after cutting of dry-cut specimens which are embedded in polymer, wherein the method in this case is carried out at room temperature without cooling.
Preferred embodiments of the device according to the invention and their use are explained below with reference to drawings which are given as examples, wherein the invention is not restricted to the particular embodiments shown. In the drawings:
a to 9b show the different phases I to VI of the method according to the invention for manipulation of section ribbons using the device according to the invention as an example. These show:
FIG. 7Aa phase IV of the section manipulation in the cryostat: positioning of the grid below the section ribbon,
FIG. 8Aa phase V, grid with section ribbon applied, after removal of the section ribbon manipulator with the fibre tip, in top view,
FIG. 8Bb phase V in side view,
A first micromanipulator 3 is attached exchangeably to the cover 2 of the boundary of a cryostat (cryostat 1) or an uncooled ultramicrotome apparatus, and has an apparatus for adjustment in the three spatial axes x, y, z (
The opening 8 in the cover 2 is a docking point made of magnetizable material for a commercially available ionization device 9 (
A second micromanipulator 10 is attached to an arm 11 of the cover 2 which can be rotated about a vertical axis (
The entire manipulation of both the sections ribbons and the TEM grid takes place using these two micromanipulators by an operator from outside the working chamber for the cutting process in the microtome apparatus.
The section ribbon manipulator 4 shown in
The assembled tip 5 is inserted in the angled part of the interior of the tube 13 (
Before insertion, the surface of the fibre 16 is made electrically conductive by coating (for example by coating with a sputter source or vapour deposition under rotation in a vacuum) with an electrically conductive surface layer of a metal or carbon. The material of this surface layer is preferably selected from one or more of the following materials: gold, silver, copper, nickel, palladium, platinum, cobalt, but also carbon. This electrically conductive surface layer allows the tip 5 to be earthed during use of the method, where necessary for handling the section ribbons. Before coating, the fibre 16 is cleaned using solvent which is suitable for degreasing of hairs, preferably acetone, and then in the manner shown attached to the tube 15, for example with the conductive carbon already mentioned or another electrically conductive adhesive.
A plurality of tubes with the fibres 16 attached thereto are then coated together in the same work process. The layer thickness of the conductive surface layer, for metals, is preferably between 30 and 50 nm. The metal layers are sputtered onto the surface. This is achieved using a glow discharge with an argon gas pressure of preferably 0.1 mbar, by bombarding a target of the desired metal, which for example is produced in dish form. The metal clusters ejected from the target are distributed in the recipient, covering this and also the object to be coated according to the sputter energy applied and the duration of the coating process.
The carbon is vapour-deposited, preferably in a vacuum (p<10−5 mbar) (so-called electron bombardment or resistance vapour deposition). Experiments have shown that the optimum layer thickness of the carbon is between 10 and 20 nm. Thicker layers can easily break and detach from the surface of the coated fibre 16.
The fibres 16 which are coated with metal or carbon in this way can be used repeatedly with careful handling. If the tip of the fibre is covered with section residue, it can be removed from the cryostat for short periods together with the section ribbon manipulator and thawed at room temperature. Conductive fibres are damaged under excessive deformation. This can occur if the fibre accidentally “rides over” the cutting edge of the knife, or the section ribbon manipulator is positioned unsuitably in the cryostat or removed carelessly after the end of work.
The connecting piece 12 and metal tube 13 are connected together electrically conductively. As shown in
As shown in
A spring mechanism 21 is provided at the upper end of the tube 19, which allows a thin wire or rod 22 to be pushed in the axial direction in the interior of the tube 19. Since this wire 22 is bent or angled at its end, as shown in enlarged detail in
As shown in enlarged detail B in
A grid 23, as used as a specimen carrier in transmission electron microscopy, is clamped in the grid holder 17 in this way. Such grids 23, which are known in themselves, have different mesh widths for use in TEM, the preferred diameter is 3 mm, the preferred layer thickness 30 μm. This grid 23 consists at least partly of an electrically conductive material such as copper, gold or molybdenum, which is usually coated with a thin (around 10 nm thick) electrically conductive layer, preferably a carbon layer. Preferably, commercially available grids, such as for example C-Flat™ (Electron Microscopy Sciences, Hatfield, Pa., US) are used. These are copper grids which are coated with a carbon layer around 10 nm thick, and suitably have defined holes (hole diameter for example between 1 and 2 μm) in a regular arrangement.
The method for use of the device according to the invention comprises six individual steps which are designated phases I to VI:
Phase I: In the starting position, the first sections are on the microtome knife 6 and the leading fibre tip 16 of the section ribbon manipulator 4 does not yet touch the still short section ribbon 24 (
Phase II: The electrically conductive and earthed tip 16 is brought between the knife 6 and the first sections of the forming section ribbon 24. Using the first micromanipulator 3, it is brought from below into contact with the section ribbon 24 emerging from the microtome, and by a brief increase in ionization (charge mode) and the resulting electrostatic attraction, is firmly connected to the section ribbon 24 (
Phase III: After this fixing, the fibre tip 16 with the section ribbon adhering thereto is moved as far as possible away from the knife 6 of the microtome. By production of new sections of the specimen by microtomisation, the section ribbon 24 is extended and must be stretched horizontally by drawing back the fibre tip 16 of the section ribbon manipulator 4 using the micromanipulator 3. After emergence of the section ribbon from the microtome, only a minimum tension is exerted on the section ribbon which is usually mechanically very fragile, and by retracting the fibre tip 16 in the horizontal direction, the sag of the section ribbon following the microtome output is prevented.
Phase IV: Positioning of the TEM grid, consisting of a metal film coated with a thin (approx. 10 nm thick) carbon layer, below the section ribbon. As shown in FIG. 7Aa, the grid 23 is positioned below the section ribbon 24, using the grid holder 17 which is guided by the second micromanipulator 10, when the section ribbon 24 cut from the specimen has reached the desired length. The application and correct positioning of the grid 23 on the grid holder 17 using the spring mechanism 21 are simple and efficient. Thus the grid 23 can quickly be exchanged and manipulated.
Phase V: Fixing of the section ribbon on the grid; separation from the fibre tip. When the section ribbon 24 has reached the desired length and the grid 23 is placed in the desired oblique position below the section ribbon 24 using the second micromanipulator 10, by moving the grid 23 in the vertical direction the section ribbon 24 is brought into contact with the grid 23 on the edge facing the knife, and at the same time the fibre tip 16 is lowered with the first micromanipulator 3 so that the section ribbon 24 comes to lie closely on the surface of the grid 23 (
Phase VI: Cutting or tearing of the section ribbon on the grid, separation from the ultramicrotome: After fixing the section ribbon 24 on the grid 23 the fibre tip is moved away from the section ribbon 24 and hence the connection between the two is broken (
Finally, by suitable movement of the section ribbon 24 in its longitudinal axis, the part of the section ribbon 24′ adhering to the grid 23 is separated from the rest. Now the part of the section ribbon 24′ adhering to the grid 23 is removed from the device using the second grid holder 17 guided by the second micromanipulator 10 (
The advantage of this method is that no highly developed fine motor skills are required by the operator in order to produce cryosections successfully, since all demanding steps are carried out by micromanipulators which are easy to operate. These are designed so that the function of the magnifying glass, mounted swivelably above the cryostat for observation in commercial microtomes, is not obstructed (over-large manipulators prevent the unhindered swiveling of the magnifying glass). Furthermore, the rotatability of the grid micromanipulator, shown in
An important advantage of the invention is the rapid manipulation of the grid 23 in the case of cryosectioning after successful fixing of the sections on the grid 23. Ice crystals are always present in the cryostat. If these adsorb on the sections in the cold nitrogen gas atmosphere, the structures below are no longer suitable for examination in the TEM since electrons cannot propagate through small ice crystals (diameter<0.5 microns) irradiated in the electron microscope. Under liquid nitrogen however no ice crystals adsorb onto the sections. Rapid transfer of the section ribbons 24 adsorbed on the grid 23, as enabled by the device according to the invention, is therefore of great preparative benefit.
The method for transfer of the section ribbons emerging from the microtome to a TEM carrier can be used both for specimens of biological origin (tissue specimens) and for specimens of other materials (for example high-polymer plastics). It is equally suitable for working at room temperature and at temperatures in the cryoregion between 190° K. and 110° K., preferably around 120° K.
Number | Date | Country | Kind |
---|---|---|---|
13405063.2 | May 2013 | EP | regional |