The technology discussed below relates generally to optical spectroscopy including diffuse reflectance and transmission spectroscopy, and in particular, to mechanisms to effectively increase the spectrometer collection for better performance with inhomogeneous samples and increased sensitivity.
Diffuse reflectance spectroscopy may be utilized to study the molecular structure of a given material based on its spectral response. In diffuse reflectance spectroscopy, a light source (e.g., a wide band light source) shines light onto the material under test. The incident light interacts with the material such that part of the light is transmitted, another part of the light is reflected, and another part of the light is scattered. The scattered portion is affected by the sample absorption spectrum and can be used to identify the material based on its spectral print. Diffuse reflectance spectroscopy can be used with different forms of the material, such as solids, powders, and liquids.
Although the absorption spectrum is mainly determined by the material itself, measurement errors can start to appear if the scattered light does not fully represent the measured sample. This can be attributed to many reasons. For example, one reason is related to using an inhomogeneous sample. This means that the material has different regions where the absorbance spectrum varies depending on the measurement position. Another reason is related to the size and the shape of the material macroscopic particle size. Various materials can exhibit a wide range of macroscopic particle forms, spanning from fine particles found in powders to large spherical grains with diameters measuring a few millimeters. Materials like grains, e.g., corn kernels, demonstrate both inhomogeneity and particle size variations. For example, corn can have kernels with irregular shapes with lengths that vary from 5 mm to 20 mm. In this example, the scattered light is dominated by the part of the kernel that is aligned with the spectrometer's field of view. This makes a single measurement representative for only a small part of the corn kernel. Typically, corn kernels may have different absorption spectrums depending on the measurement position. In addition, it is difficult to measure multiple kernels in the same measurement due to the limited spectrometer field of view. Furthermore, this shape irregularity can affect the scattered light that can be collected into the spectrometer, which affects the signal-to-noise ratio (SNR) of the spectrometer measurement.
Conventional spectrometers may overcome these issues through an initial step of sample preparation. Such a sample preparation step usually involves grinding the material to form a homogenous powder. Such a sample preparation process, while feasible in the laboratory, may not be an easy task in the field for a portable spectrometer or in a production line for quality control. The evolution of portable and handheld spectrometers in the last decade has resulted in improvements in the sample interface to overcome the above-mentioned issues. For example, one way to overcome the sample inhomogeneity issue is to increase the spectrometer field of view. This helps average the sample response across the different inhomogeneous regions and ensures that the different regions are well represented in the scattered light and collected light inside the spectrometer.
However, increasing the spectrometer field of view may not be practical in some cases, especially when there are limitations in the acceptance angle and the aperture size of the spectrometer. This can be the case of handheld and portable spectrometers. Many of these spectrometers are based on MEMS (Micro-Electro-Mechanical Systems) components that typically have smaller optical surfaces which accordingly limit the optical throughput of the whole spectrometer. Such limitations impose constraints on the achieved field of view with enough optical coupled power. For example, increasing the area viewed by the spectrometer with limited optical throughput may decrease the coupled optical power coupled, which may result in a reduction in the SNR. A degraded SNR can greatly affect the material analysis and introduce errors in the chemo-metrics models.
The following presents a summary of one or more aspects of the present disclosure, in order to provide a basic understanding of such aspects. This summary is not an extensive overview of all contemplated features of the disclosure, and is intended neither to identify key or critical elements of all aspects of the disclosure nor to delineate the scope of any or all aspects of the disclosure. Its sole purpose is to present some concepts of one or more aspects of the disclosure in a form as a prelude to the more detailed description that is presented later.
In an example, an optical device is disclosed including a reflective surface positioned apart from a sample and configured to receive a first portion of scattered light from the sample and to redirect the first portion of the scattered light back to one or more discrete spots on the sample in a non-random manner to produce redirected scattered light from the sample. The optical device further includes a spectrometer configured to receive coupled light from the sample at an input thereof and to obtain a spectrum of the sample based on the coupled light. The coupled light includes at least a portion of the redirected scattered light.
Another example provides a method for increasing collection of a spectrometer. The method includes receiving a first portion of scattered light from a sample at a reflective surface positioned apart from the sample, redirecting the first portion of the scattered light back to one or more discrete spots on the sample in a non-random manner to produce redirected scattered light from the sample, and receiving coupled light from the sample at an input of the spectrometer to obtain a spectrum of the sample based on the coupled light. The coupled light includes at least a portion of the redirected scattered light.
These and other aspects of the invention will become more fully understood upon a review of the detailed description, which follows. Other aspects, features, and embodiments of the present invention will become apparent to those of ordinary skill in the art, upon reviewing the following description of specific, exemplary embodiments of the present invention in conjunction with the accompanying figures. While features of the present invention may be discussed relative to certain embodiments and figures below, all embodiments of the present invention can include one or more of the advantageous features discussed herein. In other words, while one or more embodiments may be discussed as having certain advantageous features, one or more of such features may also be used in accordance with the various embodiments of the invention discussed herein. In similar fashion, while exemplary embodiments may be discussed below as device, system, or method embodiments it should be understood that such exemplary embodiments can be implemented in various devices, systems, and methods.
The detailed description set forth below in connection with the appended drawings is intended as a description of various configurations and is not intended to represent the only configurations in which the concepts described herein may be practiced. The detailed description includes specific details for the purpose of providing a thorough understanding of various concepts. However, it will be apparent to those skilled in the art that these concepts may be practiced without these specific details. In some instances, well known structures and components are shown in block diagram form in order to avoid obscuring such concepts.
Spectroscopy may be utilized, for example, to characterize the chemical content of solids, liquids and gases. Spectrometers analyze and interact with the sample under test using various mechanisms, such as transmission or diffuse reflectance sampling. Transmission is commonly used with liquid samples or gases. In this configuration, the light passes through the sample under test and then is coupled into the spectrometer after interacting with the molecules of the sample. In diffuse reflectance sampling, the light is incident on the sample under test and the scattered light from the sample is collected and coupled into the spectrometer. Diffuse reflectance is mostly used for solids.
In both cases, the light collected by the spectrometer only interacts with a limited portion of the sample. In some applications, the sample is not homogenous and can have a large particle size. Measuring a small spot size in this case may not be representative of the sample spectral properties. This can be observed in the measured spectrum as a large variation across the different locations on the sample. For example, in grain analysis, the particle size can vary from 5-20 mm, and accordingly, the optical system of the spectrometer should be able to accommodate this large spot size. In most of the current systems available in the market, the sample is measured several times at different positions and then the readings are averaged together. In this disclosure, a technique is provided to increase the spot size of the spectrometer and ultimately increase the accuracy of the chemo-metrics models to simultaneously couple the light from the different locations from the sample at the same time.
In particular, aspects are directed to techniques to redirect scattered light from a sample to one or more spots on the sample in a non-random manner to increase the effective spot size area. In some examples, a portion of the scattered light from a spot on the sample may be directly coupled into the spectrometer, while a remaining portion of the scattered light may be redirected back to the spot on the sample to increase the effective size of the spot and to further increase the optical power of the optical device. In other examples, the scattered light from a first spot on the sample that is out of the field-of-view of the spectrometer may be redirected to a second spot on the sample that is within the field-of-view of the spectrometer, where the first and second spots collectively form an extended spot area on the sample.
The spectrometer 104 as an optical system can be simplified as an aperture 106 with certain dimensions (e.g., diameter) and an acceptance angle θ. The acceptance angle θ represents an angle of a head of a cone of light optically coupled into the spectrometer 104 from an illuminated spot 102 on a sample. For example, light passing through the sample or reflected from (e.g., scattered from) the sample may be optically coupled from the illuminated spot 102 towards the spectrometer 104. The throughput of the spectrometer 104 shown in
Throughput∝aperture diameter2×sin2(θ). (Equation 1)
The distance between the sample interface and the spectrometer input surface may be represented as X, as shown in
Spot diameter≈2×tan(θ). (Equation 2)
Based on Equations 1 and 2 above, the key limitation in spectrometer optical systems in terms of the collection spot size may be considered the acceptance angle θ of the system. In miniaturized spectrometers where the input aperture 106 has a small area and acceptance angle, the performance of such spectrometers may be limited to homogeneous samples that have almost constant spectral response across different locations on the sample.
In various aspects of the disclosure, techniques are provided to redirect the useless scattered light 208 from the sample back to the sample surface to increase the light interaction with the sample and collect more information from the sample, hence averaging the measured spectra, which results in an enhancement in coupling light to the spectrometer 204. Various aspects provide different mechanisms to extend the coupled spot size and effectively achieve a larger field of view. In some examples, the spot can be coupled from different locations on the sample at the same time, which may decrease the measurement time if the sample is inhomogeneous since less averaging is needed to cover the whole sample.
In some aspects, a reflective surface is used to redirect the scattered light 208 that is typically missed in the input aperture and acceptance angle of the spectrometer back to the sample. This redirected light interacts with the sample again and the scattered back part of this scattered light is coupled into the spectrometer. The coupled power into the spectrometer then includes the contribution from the first sample reflection and the second sample reflection. Depending on the reflective surface and material of the sample under test characteristics, light may be trapped between the reflective surface and the sample. In this example, light may be coupled after multiple reflections from the sample.
In the example shown in
Therefore, in aspects of the disclosure, as shown in
In some examples, the redirected scattered light may produce a scattering pattern that covers the half sphere solid angle. Part of these scattered rays (e.g., indirectly coupled redirected scattered light 316) can be coupled into the spectrometer acceptance angle, while the remainder of the redirected scattered light (e.g., additional missed scattered light) can again be collected by the reflective surface 308 and redirected once more to the sample 306 to produce additional redirected scattered light that may be coupled into the spectrometer 304. The same process can occur multiple times with multiple redirections of the missed scattered light; however, less power may be coupled each time due to light leakage through the sample 306 and the coupling hole 318 to the spectrometer 304. Therefore, the coupled light (coupled power) to the spectrometer 304 may be represented as:
Coupled power=Directly coupled scattered light from first interaction+Indirectly coupled Redirected Scattered Light from second interaction (Equation 3)
More generally though, the spectrometer 304 is configured to receive coupled light including the second portion of the scattered light (e.g., directly coupled scattered light 310) and at least a portion of the redirected scattered light (e.g., indirectly coupled redirected scattered light 316), which can include redirected scattered light from multiple subsequent interactions with the sample.
In addition, in the example shown in
Light scattered from other spots (e.g., spots 502b-502e) on the sample 506 may be trapped between the sample 506 and a reflective surface 508 for redirection of the scattered light towards a spot (e.g., spot 502a) on the sample 506 within the field of view of the spectrometer 504. For example, the reflective surface 508 may have a curvature configured to collect scattered light (e.g., scattered light 512) from an illuminated spot (e.g., spot 502d) outside the field of view of the spectrometer 504 and to redirect (reflect) the scattered light as reflected scattered light 514 to another spot (e.g., spot 502a) on the sample 506 within the field of view of the spectrometer 504, where the light is scattered for a second time to produce redirected scattered light that may be coupled into the spectrometer 504 as indirectly coupled redirected scattered light 516. In this example, the first spot 502d and the second spot 502a form an extended spot area on the sample 506. The reflective surface 508 may further have a hole 518 therein for coupling of scattered light including the directly coupled scattered light 510 and the indirectly coupled redirected scattered light 516 into the spectrometer 504.
In some examples, light may be redirected multiple times. For example, the reflective surface 508 may be configured to receive scattered light (e.g., scattered light 512) from a first illuminated spot (e.g., spot 502e) outside the field of view of the spectrometer 504 and to redirect (reflect) the scattered light as reflected scattered light 514 to a second spot (e.g., spot 502d) on the sample 506 that is also outside (or partially outside) the field of view of the spectrometer 504. The light may then be scattered for a second time to produce redirected scattered light 512 that may be redirected again by the reflective surface 508 to a third spot (e.g., spot 502a) within the field of view of the spectrometer 504, where the light is scattered again for a third time to produce redirected scattered light, at least a portion of which may be coupled into the spectrometer 504 as indirectly coupled redirected scattered light 516.
The reflective surface 608 may include a first section 608a adjacent to the illumination system 616 having a first curvature configured to couple the input light 618 to the spot 602 on the sample 606 and a second section 608b having a second curvature different from the first curvature and configured to collect scattered light (e.g., scattered light 612) from the illuminated spot 602 and to redirect (reflect) the scattered light as reflected scattered light 614 to the spot 602 on the sample 606, thereby increasing the effective spot size of the spot 602 (e.g., as a result of different angles of light being incident on the sample). Here, the light is scattered for a second time to produce redirected scattered light that may be coupled into the spectrometer 604 as indirectly coupled redirected scattered light 624 to increase the illuminated power in the throughput-limited effective area that gets coupled into the spectrometer 604. In this example, the first section 608a includes two outside sections 622a and 622b and the illumination system includes two light sources 620a and 620b, each positioned adjacent to one of the outside sections 622a and 622b. The reflective surface 608 may further have a hole 626 therein for coupling of scattered light including the directly coupled scattered light 610 and the indirectly coupled redirected scattered light 624 into the spectrometer 604. It should be understood that the present disclosure is not limited to any particular number or configuration of the light sources 620a and 620b or the number or configuration of the outside sections 622a and 622b. For example, the position and number of light sources 620a and 620b and the outside sections 622a and 622b may vary, based on the configuration of the optical device 600.
The reflective surface 708 may include a first section 708a adjacent to the illumination system 716 having a first curvature configured to couple the input light 718 to the spot 702 on the sample 706 and a second section 708b having a second curvature different from the first curvature and configured to collect scattered light (e.g., scattered light 712) from the illuminated spot 702 and to redirect (reflect) the scattered light as reflected scattered light 714 to the spot 702 on the sample 706. Here, the light is scattered for a second time to produce redirected scattered light that may be coupled into the spectrometer 704 as indirectly coupled redirected scattered light 724 to increase the effective spot size of the spot 702 (e.g., as a result of different angles of light being incident on the sample) and to increase the illuminated power in the throughput effective area coupled into the spectrometer 704. In this example, the first section 708a includes two outside sections 722a and 722b and the illumination system includes two light sources 720a and 720b, each positioned adjacent to one of the outside sections 722a and 722b. The reflective surface 708 may further have a hole 730 therein for coupling of scattered light including the directly coupled scattered light 710 and the indirectly coupled redirected scattered light 724 into the spectrometer 704. It should be understood that the present disclosure is not limited to any particular number or configuration of the light sources 720a and 720b or the number or configuration of the outside sections 722a and 722b. For example, the position and number of light sources 720a and 720b and the outside sections 722a and 722b may vary, based on the configuration of the optical device 700.
In addition, in the example shown in
In addition, in the example shown in
The first section 808a is adjacent to the illumination system 816 and has a first curvature configured to couple the input light 818 to the spot 802 on the sample 806. For example, the first section 808a includes two outside sections 826a and 826b and the illumination system includes two light sources 824a and 824b, each positioned adjacent to one of the outside sections 826a and 826b (e.g., between the reflective surface 808 of the outside sections 826a and 826b and the sample 806). The second section 808b is positioned between the two outside sections 826a and 826b of the first section 808a in front of the sample 806 (e.g., on a same side as the spectrometer 804) and has a second curvature different from the first curvature. The third section 808c is positioned between the two outside sections 826a and 826b behind the sample 806 (e.g., on an opposite side as the spectrometer 804) and has a third curvature different from the first and second curvatures.
The second section 808b and third section 808c are each configured to collect scattered light (e.g., scattered light 812) from the spot 802 on the sample 806 or from other spots on the sample 806 (e.g., other spots radiated by the light sources 824a and 824b) in all directions and to redirect (reflect) the scattered light as reflected light 814 to the spot 802 on the sample 806 to increase the effective spot size of the spot 802 (e.g., as a result of different angles of light being incident on the sample). Here, the light is scattered for a second time to produce redirected scattered light that may be coupled into the spectrometer 804 as indirectly coupled redirected scattered light 822. The reflective surface 808 may further have a hole 828 therein for coupling of scattered light including the directly coupled scattered light 810 and the indirectly coupled redirected scattered light 822 into the spectrometer 804. It should be understood that the present disclosure is not limited to any particular number or configuration of the light sources 824a and 824b or the number or configuration of the outside sections 826a and 826b. For example, the position and number of light sources 824a and 824b and the outside sections 826a and 826b may vary, based on the configuration of the optical device 800. Moreover, the shape and curvature of the sections 808a, 808b, and 808c of the reflective surface 808 may vary depending on the desired coupling into the spectrometer 804.
The optical device 800 may further include a sample holder 820 configured to hold the sample 806. In the example shown in
In addition, in the example shown in
The first section 908a is adjacent to the illumination system 916 and has a first curvature configured to couple the input light 918 to the spot 902 on the sample 906. For example, the first section 908a includes two outside sections 926a and 926b and the illumination system includes two light sources 924a and 924b, each positioned adjacent to one of the outside sections 926a and 926b (e.g., between the reflective surface 908 of the outside sections 926a and 926b and the sample 906). The second section 908b is positioned between the two outside sections 926a and 926b of the first section 908a in front of the sample 906 (e.g., on a same side as the spectrometer 904) and has a second curvature different from the first curvature. The third section 908c is positioned between the two outside sections 926a and 926b on the sides of the sample 906 (e.g., on an opposite side as the spectrometer 904) and has a third curvature different from the first and second curvatures.
The second section 908b and third section 908c are each configured to collect scattered light (e.g., scattered light 912) from the spot 902 on the sample 906 or from other spots on the sample 906 (e.g., other spots radiated by the light sources 924a and 924b or by reflected light) in all directions and to redirect (reflect) the scattered light as reflected light 914 to the spot 902 on the sample 906 to effectively increase the spot size of the spot 902 (e.g., as a result of different angles of light being incident on the sample). Here, the light is scattered for a second time to produce redirected scattered light that may be coupled into the spectrometer 904 as indirectly coupled redirected scattered light 922. The reflective surface 908 may further have a hole 928 therein for coupling of scattered light including the directly coupled scattered light 910 and the indirectly coupled redirected scattered light 922 into the spectrometer 904. It should be understood that the present disclosure is not limited to any particular number or configuration of the light sources 924a and 924b or the number or configuration of the outside sections 926a and 926b. For example, the position and number of light sources 924a and 924b and the outside sections 926a and 926b may vary, based on the configuration of the optical device 900. Moreover, the shape and curvature of the sections 908a, 908b, and 908c of the reflective surface 908 may vary depending on the desired coupling into the spectrometer 904.
The optical device 900 may further include a sample holder 920 configured to hold the sample 906. In the example shown in
In the example shown in
For example, the reflective surface 1008 may include a first section 1008a adjacent to the illumination system 1016 having a first curvature configured to couple the input light 1018 to the respective redirected spot 1002b on the sample 1006. In this example, the first section 1008a includes two outside sections 1020a and 1020b and the illumination system includes two light sources 1024a and 1024b, each positioned adjacent to one of the outside sections 1020a and 1020b. The reflective surface 1008 may further include a second section 1008b having a second curvature different from the first curvature and configured to collect scattered light (e.g., scattered light 1012) from the illuminated redirected spots 1002b outside or partially outside the field of view of the spectrometer 1004 and to redirect (reflect) the scattered light as reflected scattered light 1014 to the collection spot 1002a (e.g., a collection area) on the sample 1006 that is within the field of view of the spectrometer 1004. Here, the light is scattered for a second time to produce redirected scattered light 1010 that may be coupled into the spectrometer 1004. In this example, the redirected scattered light 1010 may correspond to the directly coupled scattered light. The reflective surface 1008 may further have a hole 1022 therein for coupling of scattered light 1010 into the spectrometer 1004.
Missed light scattered outside the acceptance numerical aperture of the spectrometer 1004 may further be redirected again towards the collection spot 1002a on the sample 1006 via the second section 1008b of the reflective surface 1008, and scattered a third time to produce additional redirected scattered light 1010, at least a portion of which may be coupled into the spectrometer 1004 as indirectly coupled redirected scattered light. It should be understood that the number of bounces may be increased, depending on the configuration of the optical device 1000. For example, scattered light from redirected spots 1002b may be directed to additional spots (not shown) on the sample 1006 outside or partially outside the field of view of the spectrometer 1004 prior to being directed to the collection spot 1002a within the field of view of the spectrometer 1004. The multiple-bounce configuration shown in
In some examples, the measured power spectral density PSD by the spectrometer 1004, in presence of the sample, Ss, is given by:
S
s
=R
s
2
S
o, (Equation 1)
where So is the reference PSD in presence of an ideal diffuse reflectance (e.g., Spectralon) and Rs is the sample reflectance. The squaring of the Rs is based on the light interaction with the sample twice before being coupled to the spectrometer (as shown in the configuration of
For non-homogeneous samples where the spectral characteristics may differ depending on measurement location on the sample, the light scattered from the different spots may experience different reflectance. Accordingly, in the configuration shown in
where Rs1 is the sample reflectance from the right-hand spot, Rs2 is the sample reflectance from the left-hand spot, and Rs0 is the sample reflectance from the middle spot, as shown in
Thus, it is clear that the extracted reflectance is the geometrical mean of middle-region reflectance (e.g., collection spot 1002a) and the mathematical mean of the left and right regions (e.g., redirected spots 1002b). Assuming random noise effect due to non-homogeneity across the sample, the reflectance of each specific spot can be expressed as:
R
sk
=R
s,H
+n
sk, (Equation 4)
where Rsk represents the reflectance from spot k, Rs,H is the homogenous reflectance of the sample, and nsk is the reflectance noise due to non-homogeneity from different regions/spots, the extracted sample reflectance of a non-homogenous sample, Rs,NH, can be expressed as:
For Rs,H>>ns, the third term under the square root of 1/Rs,H2 can be ignored, and the expression can be approximated as:
The above illustrates that noise due to sample non-homogeneity is reduced by around 39%, which is close to the case of tripling the spot size, as if non-homogeneity noise is averaged from three different spots, leading to noise degradation by around 42%. Extending this concept to M light sources illuminating the sample at M distinct spots around the main collection region (collection spot), which is re-directed back to the main collection region, the extracted reflectance is expressed as:
As indicated in the above equation, the expected non-homogeneity noise reduction is saturating at 50%, and therefore, increasing the number of bulbs and consequently illumination spots mainly provides an increase in the collected signal level.
In some examples, the spectrometer signal to noise ratio (SNR) can be analyzed, considering the spectrometer system noise apart from noise due to sample non-homogeneity, and coupled signal level. The multiple-incidence configuration shown in FIG. 10AError! Reference source not found. can then be compared to the case of a direct collection of a spot with three times larger diameter 3Ds, assuming the same light sources are used. For the direct collection case, according to throughput limitation, coupling efficiency is inversely proportional to spot area, and therefore, the coupling efficiency is reduced nine times due to three times increase in diameter.
For the multiple-incidence configuration shown in
The example shown in
For example, the reflective surface 1108 may include a first section 1108a adjacent to the illumination system 1116 having a first curvature configured to couple the input light 1118 to the respective redirected spot 1102b on the sample 1106. In this example, the first section 1108a includes two outside sections 1120a and 1120b and the illumination system includes two light sources 1132a and 1132b, each positioned adjacent to one of the outside sections 1120a and 1120b. The reflective surface 1108 may further include a second section 1108b having a second curvature different from the first curvature and configured to collect scattered light (e.g., scattered light 1112) from the illuminated redirected spots 1102b outside or partially outside the field of view of the spectrometer 1104 and to redirect (reflect) the scattered light as reflected scattered light 1114 to the collection spot 1102a (e.g., a collection area) on the sample 1106 that is within the field of view of the spectrometer 1104. Here, the light is scattered for a second time to produce redirected scattered light 1110 that may be coupled into the spectrometer 1104. In this example, the redirected scattered light 1110 may correspond to the directly coupled scattered light. The reflective surface 1108 may further have a hole 1122 therein for coupling of scattered light 1110 into the spectrometer 1104.
The reflective surface 1108 may further include a third section 1108c (e.g., an inside section) having a third curvature different than the first curvature and the second curvature and configured to collect missed scattered light 1124 outside the acceptance numerical aperture of the spectrometer 1104 and to redirect the missed scattered light 1124 again towards the collection spot 1102a on the sample 1106 as missed reflected light 1126. Here, the missed reflected light may be scattered a third time to produce additional redirected scattered light 1128, at least a portion of which may be coupled into the spectrometer 1104 as indirectly coupled redirected scattered light. As shown in
The example shown in
For example, the reflective surface 1208 may include a first section 1208a adjacent to the illumination system 1216 having a first curvature configured to couple the input light 1218 to the respective redirected spot 1202 on the sample 1206. In this example, the first section 1208a includes two outside sections 1222a and 1222b and the illumination system includes two light sources 1232a and 1232b, each positioned adjacent to one of the outside sections 1222a and 1222b. The reflective surface 1208 may further include a second section 1208b having a second curvature different from the first curvature and configured to collect scattered light (e.g., scattered light 1212) from the illuminated redirected spots 1202 outside or partially outside the field of view of the spectrometer 1204 and to redirect (reflect) the scattered light as reflected scattered light 1214 to the collection area 1220 (e.g., diffuse reflective material) on the sample 1206 that is within the field of view of the spectrometer 1204. Here, the light is reflected to produce redirected scattered light 1210 that may be coupled into the spectrometer 1204. In this example, the redirected scattered light 1210 may correspond to the directly coupled scattered light. The reflective surface 1208 may further have a hole 1236 therein for coupling of scattered light 1210 into the spectrometer 1204.
The reflective surface 1208 may further include a third section 1208c (e.g., an inside section) having a third curvature different than the first curvature and the second curvature and configured to collect missed scattered light 1224 outside the acceptance numerical aperture of the spectrometer 1204 and to redirect the missed scattered light 1224 again towards the collection area 1220 on the sample 1206 as missed reflected light 1226. Here, the missed reflected light may be reflected to produce additional redirected scattered light 1228, at least a portion of which may be coupled into the spectrometer 1204 as indirectly coupled redirected scattered light. As shown in
The light source 1304 is configured to illuminate the sample 1302 with input light 1316 via the corrugated window 1308. For example, the sample 1302 may be positioned on the outer half period of the corrugation making an incident angle of the input light 1316 centered around 45°. The input light 1316 incident on the sample 1302 will interact with the sample 1302 at spot 1328 to produce first scattered light 1324. Part of the first scattered light 1324 will hit the sample 1302 at spot 1330 placed on the other half of the first corrugation period and interact with the sample 1302 to produce second scattered light 1326. This second scattered light 1326 carries sample spectral information from the two halves of the corrugation period. A corner cube mirror 1314 may be positioned to receive the second scattered light 1326 after the two sample hits and redirect the light 1326 to the next corrugation period. In addition, one or more lenses (e.g., lens 1312) may be included and configured to increase the collection of the scattered light 1326 to the corner cube mirror 1314 and from the corner cube mirror 1314 to the sample 1302. The scattered light 1326 hitting the next corrugation period interacts with the sample 1302 at spot 1332 for the third time on the first half of this new period and for the fourth time on the second half at spot 1334 to produce redirected scattered light 1318. In this example, the redirected scattered light 1318 may include the directly coupled scattered light. The redirected scattered light 1318 may then be coupled into the spectrometer 1306. Each of spots 1328, 1330, 1332, and 113 collectively form an extended spot are on the sample 1302.
In addition, similar to the configuration shown in
In addition, in the example shown in
Collection optics (e.g., a lens 1518) is positioned to couple the redirected scattered light 1532 after the second interaction with the sample 1502. In this example, the redirected scattered light 1532 may include the directly coupled scattered light. The redirected scattered light 1532 may then be coupled into the spectrometer 1506. As a result, this configuration enables spectral measurement from the two sides 1528 and 1530 of the sample 1502. In addition, in this configuration, the parallel rays 1524 produced by the off-axis parabolic mirror 1510 enable longer propagation distances, which can be advantageous for thick samples 1502.
At block 1602, the optical device may receive a first portion of scattered light from a sample at a reflective surface positioned apart from the sample. In some examples, the reflective surface may include a half-sphere or a sphere.
At block 1604, the optical device may redirect the first portion of the scattered light back to one or more discrete spots on the sample in a non-random manner to produce redirected scattered light from the sample. In addition, at block 1606, the optical device may receive coupled light from the sample at an input of the spectrometer to obtain a spectrum of the sample based on the coupled light. The coupled light may include at least a portion of the redirected scattered light.
In some examples, the optical device may include an illumination system positioned between the reflective surface and the sample for illuminating the sample with input light. The input light may be scattered from the sample as the scattered light. In some examples, the coupled light includes a second portion of the scattered light directly coupled from the sample into the spectrometer.
In some examples, the one or more spots include a single spot within the field of view of the spectrometer. The single spot may have an extended spot size based on the redirected scattered light. In this example, the optical device may further be configured to receive another portion of the redirected scattered light at the reflective surface and may further redirect the other portion of the redirected scattered light back to the single spot on the sample to produce additional redirected scattered light. Here, the coupled light may further include at least a portion of the additional redirected scattered light. In some examples, the optical device may further reflect at least a portion of the input light back to the reflective surface via a diffuse reflective material on each side of the sample for redirection of the input light towards the single spot on the sample.
In some examples, the optical device may receive the first portion of the scattered light from a first spot on the sample having a spot area at least partially outside of a field of view of the spectrometer and may further redirect the first portion of the scattered light to a second spot on the sample within the field of view of the spectrometer. In this example, the first spot and the second spot form an extended spot area on the sample. In some examples, the optical device may further redirect the first portion of the scattered light to the second spot on the sample via a third spot on the sample that is at least partially outside of the field of view of the spectrometer. In this example, the extended spot area further includes the third spot.
In some examples, the illumination system may include at least two light sources, each configured to direct a respective portion of the input light to a respective spot on the sample. In this example, the optical device may receive the first portion of the scattered light at the reflective surface from each of the respective spots on the sample and redirect the first portion of the scattered light to a collection area on the sample within the field of view of the spectrometer from which the coupled light is directed to the spectrometer. Here, at least the respective spots on the sample form an extended spot area on the sample. In some examples, the collection area forms a collection spot on the sample, and the extended spot area further includes the collection spot. In some examples, the collection area includes a diffuse reflective material configured to direct the coupled light into the spectrometer.
In some examples, the reflective surface is a corner cube mirror. In this example, the optical device may further illuminate a first spot on the sample with the input light via a corrugated window positioned between the illumination system and the sample. The input light may be scattered from the sample as the first portion of the scattered light and directed towards the corner cube mirror via a second spot on the sample. In addition, the optical device may further redirect the first portion of the scattered light back from the corner cube mirror back to a third spot on the sample via the corrugated window to produce the redirected scattered light that is coupled into the spectrometer via a fourth spot on the sample. Here, the first spot, the second spot, the third spot, and the fourth spot form an extended spot area on the sample. In some examples, the optical device further includes a corrugated mirror and a heat dissipating element adjacent to the corrugated mirror. In this example, the sample may be sandwiched between the corrugated window and the corrugated mirror.
In some examples, the reflective surface includes a first off-axis parabolic mirror and a second off-axis parabolic mirror. In this example, the optical device may further receive the first portion of the scattered light at the first off-axis parabolic mirror from a first spot on a first side of the sample, direct the first portion of the scattered light as a set of parallel rays from the first off-axis parabolic mirror to the second off-axis parabolic mirror, and direct reflected scattered light corresponding to the set of parallel rays to a second spot on a second side of the sample opposite the first side to produce the redirected scattered light. Here, the first spot and the second spot from an extended spot area on the sample.
The following provides an overview of examples of the present disclosure.
Example 1: An optical device, comprising: An optical device, comprising: a reflective surface positioned apart from a sample and configured to receive a first portion of scattered light from the sample and to redirect the first portion of the scattered light back to one or more discrete spots on the sample in a non-random manner to produce redirected scattered light from the sample; and a spectrometer configured to receive coupled light from the sample at an input thereof and to obtain a spectrum of the sample based on the coupled light, the coupled light comprising at least a portion of the redirected scattered light.
Example 2: The optical device of example 1, further comprising: an illumination system positioned between the reflective surface and the sample and configured to illuminate the sample with input light that is scattered from the sample as the scattered light.
Example 3: The optical device of example 2, wherein the coupled light further comprises a second portion of the scattered light directly coupled from the sample into the spectrometer.
Example 4: The optical device of example 3, wherein the one or more spots comprise a single spot within a field of view of the spectrometer, the single spot having an extended spot size based on the redirected scattered light, and wherein the reflective surface is further configured to receive another portion of the redirected scattered light and to redirect the other portion of the redirected scattered light back to the single spot on the sample to produce additional redirected scattered light, wherein the coupled light further comprises at least a portion of the additional redirected scattered light.
Example 5: The optical device of example 4, further comprising: a diffuse reflective material on each side of the sample to reflect at least a portion of the input light back to the reflective surface for redirection of the input light towards the single spot on the sample.
Example 6: The optical device of example 4, wherein the reflective surface comprises a half-sphere or a sphere.
Example 7: The optical device of any of examples 2 through 6, wherein the reflective surface comprises a first section adjacent the illumination system and a second section configured to redirect the first portion of the scattered light, wherein the first section has a first curvature and the second section has a second curvature different than the first curvature.
Example 8: The optical device of example 7, wherein the first section comprises two outside sections and the illumination system comprises two light sources, each positioned adjacent to one of the two outside sections.
Example 9: The optical device of example 7 or 8, wherein the reflective surface completely surrounds the sample.
Example 10: The optical device of example 9, further comprising: a sample holder configured to hold the sample, wherein the sample holder extends in one of two perpendicular directions.
Example 11: The optical device of example 2, wherein the reflective surface is configured to receive the first portion of the scattered light from a first spot on the sample having a spot area at least partially outside of a field of view of the spectrometer and to redirect the first portion of the scattered light to a second spot on the sample within the field of view of the spectrometer, wherein the first spot and the second spot form an extended spot area on the sample.
Example 12: The optical device of example 11, wherein the reflective surface is further configured to redirect the first portion of the scattered light to the second spot on the sample via a third spot on the sample that is at least partially outside of the field of view of the spectrometer, wherein the extended spot area further comprises the third spot.
Example 13: The optical device of example 2, wherein the illumination system comprises at least two light sources, each configured to direct a respective portion of the input light to a respective spot on the sample, and wherein the reflective surface is configured to receive the first portion of the scattered light from each of the respective spots on the sample and to redirect the first portion of the scattered light to a collection area within a field of view of the spectrometer from which the coupled light is directed to the spectrometer, wherein at least the respective spots on the sample form an extended spot area on the sample.
Example 14: The optical device of example 13, wherein the collection area forms a collection spot on the sample, and the extended spot area further comprises the collection spot.
Example 15: The optical device of example 13, wherein the collection area comprises a diffuse reflective material configured to direct the coupled light into the spectrometer.
Example 16: The optical device of any of examples 13 through 15, wherein the reflective surface comprises respective outside sections, each having one of the two light sources positioned adjacent thereto, an inside section, and respective additional sections between the respective outside sections and the inside section, wherein the inside section and the respective additional sections are configured to redirect the first portion of the scattered light to the collection area, wherein the outside sections, the additional sections, and the inside section each comprise a different respective curvature.
Example 17: The optical device of example 2, wherein the reflective surface comprises a corner cube mirror, and further comprising: a corrugated window positioned between the illumination system and the sample positioned, wherein the illumination system is configured to illuminate the sample with the input light via the corrugated window at a first spot on the sample, the input light being scattered from the sample as the first portion of the scattered light and directed towards the corner cube mirror via a second spot on the sample, wherein the corner cube mirror is configured to redirect the first portion of the scattered light back to a third spot on the sample via the corrugated window to produce the redirected scattered light that is coupled into the spectrometer via a fourth spot on the sample, wherein the first spot, the second spot, the third spot, and the fourth spot form an extended spot area on the sample.
Example 18: The optical device of example 17, further comprising: a corrugated mirror, the sample being sandwiched between the corrugated window and the corrugated mirror; and a heat dissipating element adjacent to the corrugated mirror.
Example 19: The optical device of example 2, wherein the reflective surface comprises a first off-axis parabolic mirror and a second off-axis parabolic mirror, the first off-axis parabolic mirror being configured to receive the first portion of the scattered light from a first spot on a first side of the sample and to direct the first portion of the scattered light as a set of parallel rays to the second off-axis parabolic mirror, the second off-axis parabolic mirror being configured to collect the set of parallel rays and to direct reflected scattered light corresponding to the set of parallel rays to a second spot on a second side of the sample opposite the first side to produce the redirected scattered light, wherein the first spot and the second spot from an extended spot area on the sample.
Example 20: A method for increasing collection of a spectrometer using the optical device of any one of examples 1 through 19.
Within the present disclosure, the word “exemplary” is used to mean “serving as an example, instance, or illustration.” Any implementation or aspect described herein as “exemplary” is not necessarily to be construed as preferred or advantageous over other aspects of the disclosure. Likewise, the term “aspects” does not require that all aspects of the disclosure include the discussed feature, advantage or mode of operation. The term “coupled” is used herein to refer to the direct or indirect coupling between two objects. For example, if object A physically touches object B, and object B touches object C, then objects A and C may still be considered coupled to one another—even if they do not directly physically touch each other. For instance, a first object may be coupled to a second object even though the first object is never directly physically in contact with the second object. The terms “circuit” and “circuitry” are used broadly, and intended to include both hardware implementations of electrical devices and conductors that, when connected and configured, enable the performance of the functions described in the present disclosure, without limitation as to the type of electronic circuits, as well as software implementations of information and instructions that, when executed by a processor, enable the performance of the functions described in the present disclosure.
One or more of the components, steps, features and/or functions illustrated in
It is to be understood that the specific order or hierarchy of steps in the methods disclosed is an illustration of exemplary processes. Based upon design preferences, it is understood that the specific order or hierarchy of steps in the methods may be rearranged. The accompanying method claims present elements of the various steps in a sample order, and are not meant to be limited to the specific order or hierarchy presented unless specifically recited therein.
The previous description is provided to enable any person skilled in the art to practice the various aspects described herein. Various modifications to these aspects will be readily apparent to those skilled in the art, and the generic principles defined herein may be applied to other aspects. Thus, the claims are not intended to be limited to the aspects shown herein, but are to be accorded the full scope consistent with the language of the claims, wherein reference to an element in the singular is not intended to mean “one and only one” unless specifically so stated, but rather “one or more.” Unless specifically stated otherwise, the term “some” refers to one or more. A phrase referring to “at least one of” a list of items refers to any combination of those items, including single members. As an example, “at least one of: a, b, or c” is intended to cover: a; b; c; a and b; a and c; b and c; and a, b and c. All structural and functional equivalents to the elements of the various aspects described throughout this disclosure that are known or later come to be known to those of ordinary skill in the art are expressly incorporated herein by reference and are intended to be encompassed by the claims. Moreover, nothing disclosed herein is intended to be dedicated to the public regardless of whether such disclosure is explicitly recited in the claims. No claim element is to be construed under the provisions of 35 U.S.C. § 112(1) unless the element is expressly recited using the phrase “means for” or, in the case of a method claim, the element is recited using the phrase “step for.”
This application claims priority to and the benefit of Provisional Application No. 63/403,160, filed in the U.S. Patent and Trademark Office on Sep. 1, 2022, the entire content of which is incorporated herein by reference as if fully set forth below in its entirety and for all applicable purposes.
Number | Date | Country | |
---|---|---|---|
63403160 | Sep 2022 | US |