The invention relates to a method and apparatus for measuring a fluorescent sample on a substrate and preferably, though not exclusively, relates to a method and apparatus for measuring a fluorescent sample within a microfluidic channel of a microfluidic substrate.
Throughout this specification a reference to a fluorescent sample is to be taken as including a naturally fluorescent sample as well as a fluorescently labeled sample.
Throughout this specification a reference to a substrate is to be taken as a reference to a product, or component of a product, that is able to absorb an exciting radiation at a wavelength, and remit the radiation at a higher wavelength.
For many years, medical diagnostic tools have used classical biochemical techniques that involve bulky and expensive equipment such as spectrophotometry, gas chromatography (GC), mass spectrometry (MS), high-performance liquid chromatography (HPLC), paper and thin-layer chromatography (PC and TLC), and electrophoretic techniques coupled with fluorescence detection techniques. These standard analytical tools work effectively and efficiently. However, the tools are expensive, and they require costly consumables, sample throughput, and experienced and skilled operators. All these hinder rapid, inexpensive, and in-situ diagnosis of health-care requirements. Furthermore, such methods often require tedious and laborious processes. For these reasons, these tools are mostly used as confirmatory tools for the presumptive positive samples that are initially screened by some types of assay techniques.
Currently, the problems with quantitative immunoassay techniques are not significantly different from the classical biochemical techniques mentioned earlier. The performance of quantitative immunoassays is today largely restricted to centralized laboratories because of the need for long assay times, and relatively complex, bulky and expensive equipment, as well as highly trained operators. Most immunoassays remain within the walls of large centralized laboratories, far from the patients whose samples are collected and measured. If a wider range of the immunoassays are able to be run in a simpler way, less expensively and at the point of care or in the home health care environment, the health of many patients may be improved. To achieve this objective, a simple, compact, smart, robust, and inexpensive device providing high quality results is required.
Optical biosensors have some advantages, such as sensitivity, simplicity and immunity to electromagnetic wave interferences. Due to these advantages, optical biosensors are one type of biosensor exploited for immunoassay applications. There are many types of optical techniques which are commonly used for biosensing applications. Fluorescence-based sensors are highly developed due to their high sensitivity, versatility, accuracy and fairly good selectivity. A fluorescence method is also very suitable for miniaturization. The current technology to measure/detect fluorescent samples on a substrate such as, for example, inside a microfluidic channel, is performed by focusing the excitation light source onto the sample inside the microchannel and collecting the fluorescence emission of the sample using a set of complex lenses, mirrors, and optical filters. As a result, the fluorescence signal of microfluidic substrates may enter the detection system giving rise to a strong but unwanted fluorescence noise. The fluorescence from the sample of interest is usually very weak due to the low sample concentration. As a consequence, fluorescence noise due to the fluorescence of substrate may suppress the desired fluorescence signal from the sample of interest.
Currently, there are two approaches commonly used to avoid the noise due to the fluorescence of the substrate. The first approach is to incorporate a confocal fluorescence microscope to block the signals not from the thin layer within which the sample resides. This technique requires bulky, expensive and complicated optics. In the second approach, materials are selected with the substrate material having no, or a low, fluorescence property. Optical grade glass and silica are commonly used, since these materials do not fluoresce when they are excited by light within the visible wavelength range. However, these materials are relatively expensive and fabrication of microfluidic channels using these materials requires time-consuming photomask generation, photolithography and etching processes. As a consequence, a microfluidic chip made from optical grade glass or silica is relatively expensive.
Possible inexpensive materials suitable as substrate materials are polymer-based materials, such as polymethyl-methacrylate (PMMA), polycarbonate and Mylar. In addition, microfluidic channels using polymeric materials are easily fabricated by molding, embossing, casting or ablation processes. Complex models of microchannels in polymer sheets can be fabricated in less than an hour using a direct-write laser system. However, these materials exhibit relatively high fluorescence signals which in turn hinder their use for low fluorescence intensity detection. The intensity of the fluorescence background signal from the polymeric materials may be two orders of magnitude higher than the fluorescence signal of a sample within the microfluidic channel. Hence, there is a need to address the auto-fluorescence background noise of polymeric materials used in a polymeric microfluidic chip.
In accordance with a first preferred aspect there is provided a method for measuring a fluorescent sample on a substrate. The method comprises exciting the fluorescent sample with an exciting light source for the generation of a sample fluorescent optical signal and a substrate fluorescent optical signal. The substrate fluorescent optical signal is substantially eliminated leaving the sample fluorescent optical signal. The sample fluorescence optical signal is then processed. The substrate may be a microfluidic channel of a microfluidic substrate.
The substantially eliminating the substrate fluorescence signal may be by:
Alternatively, the substantially eliminating the substrate fluorescence signal may be by:
Further alternatively, the substantially eliminating the substrate fluorescence signal may be by:
According to a second preferred aspect there is provided a method for measuring a fluorescent sample on a substrate, the method comprising:
According to a third preferred aspect there is provided a method for measuring a fluorescent sample on a substrate, the method comprising:
According to a fourth preferred aspect there is provided a method for measuring a fluorescent sample on a substrate, the method comprising:
The filter may be low-pass filter; and the acquired signal may be acquired by an optical receiver. The unity-gain sine wave signal may be generated by a phase-locked loop (“PLL”) circuit. A quadrature phase shifter may generate the two sine wave signals, which may be 90° out of phase with each other. The unity-gain square wave signal may be in phase with the fluorescence signal of the sample.
The fluorophore may be a label, and the label may be fluorescein. The light source may be a 470 nm blue LED with 3460 mcd in intensity.
The PLL circuit may comprise a phase detector to generate an average DC voltage proportional to the phase difference, a low-pass filter (LPF1) to suppress high frequency components generated by the phase detector, and a voltage-controlled oscillator (VCO) to control the VCO frequency to oscillate at a frequency identical to the input frequency with a finite phase difference.
The substrate may be a microfluidic substrate having at least one microfluidic channel. The microfluidic substrate may be made of a polycarbonate material.
The optical receiver may be any one from the group consisting of: photodiode, avalanche photodiode, photomultiplier tube and CCD detector. An optical filter may filter off signals with a wavelength outside the range of interest.
According to a fifth aspect there is provided an apparatus for measuring a fluorescent sample on a substrate, the method comprising:
The filtering circuit may be an electrical filtering circuit and may comprise:
Alternatively, the electrical filtering circuit may comprise:
Further alternatively, the electrical filtering circuit may comprise:
According to a sixth preferred aspect there is provided an apparatus for measuring a fluorescent sample on a substrate, the apparatus comprising:
According to a seventh aspect there is provided an apparatus for measuring a fluorescent sample on a substrate, the apparatus comprising:
According to an eighth preferred aspect there is provide an apparatus for measuring a fluorescent sample on a substrate, the apparatus comprising:
The present invention provides a method to eliminate the noise from the fluorescence of the substrate (background noise) by filtering of the fluorescence signals. The method may use combination of one or more of: a phase-locked loop, quadrature demodulator and annihilator circuits. Preferably, the method is able to completely eliminate the fluorescence noise from substrate materials so that very weak fluorescence signal arising from the low concentration sample of interest is detectable.
Advantageously, the present invention is able to discriminate a very weak fluorescence signal of sample of interest from a large unwanted fluorescence signal caused by the fluorescence background of the microfluidic substrate. Hence, the present invention may enable detection of very low concentrations of a sample of interest on a substrate made of materials with high fluorescence properties, such as polymeric sheet materials.
In order that the present invention may be fully understood and readily put into practical effect, there shall now be described by way of non-limitative example only preferred embodiments of the present invention, the description being with reference to the accompanying illustrative drawings.
In the drawings:
Due to the difference in fluorescent emission lifetime of the fluorescein and the substrate, a phase difference in two output signals can be created.
When a fluorescent sample is excited by modulated light source of appropriate wavelength, the sample will fluoresce with the same frequency but will be phase shifted with respect to the exciting source by an amount proportional to the fluorescence lifetime of the sample. The fluorescent intensity of the sample, in this case, can be simply extracted by optical filtering of the wavelength of the exciting light source. However, if there are two fluorescent samples with similar fluorescence wavelengths but with slightly different lifetimes, optical filtering will only remove the signal from the excitation light source but will not be able to separate the information encoded in the two fluorescence samples. Due to the difference in their fluorescence lifetimes, the fluorescence signal from the two samples will be phase shifted by different amounts. The method proposed here aims to substantially eliminate the fluorescence signal from one of the samples, which we refer to as the background or substrate signal, so as to enhance the signal from the sample of interest.
Referring to
Eight embodiments of implementing the method may be used. The first embodiment uses a combination of phase-locked loop and quadrature demodulation techniques. This embodiment is suitable if the phase difference between the desired signal and the unwanted signal is close to 90°. The second embodiment uses a phase-locked loop technique only, which is more general, and used to minimize the background noise for signals with phase difference that is not close to 90°. The third embodiment uses a combination of phase-locked loop and quadrature annihilator techniques. This embodiment is suitable provided there is a phase difference between the desired signal and the unwanted signal. The fourth uses quadrature demodulator and phase locked loop systems, and the fifth uses only the quadrature demodulator. The sixth uses a quadrature annihilator and a phase-locked loop; whereas the seventh and eighth both use a quadrature annihilator.
Referring to
1.1 Light-Source [1]
Different light sources may be used as the excitation light source 10. Examples are lasers of appropriate wavelength including laser diodes, LEDs or a broadband light source. The wavelength of the light source 10 must be able to excite the fluorophore of the sample. For an example, if the fluorophore is fluorescein label, a light source 10 with a wavelength between 450 nm and 500 nm is required. Ideally, the wavelength of the light source 10 should only have a single wavelength, for example, light emitted by a laser or laser diode. However, a laser source is usually relatively bulky and expensive. In addition, lasers and laser diodes are only available at certain wavelengths. These may not be suitable for certain types of fluorophores and chromophores. If the light source 10 is an LED that covers the excitation wavelength region and also the fluorescence wavelength region, a suitable low-pass optical filter or band-pass optical filter is installed between the LED and the microfluidic card. The light beam, particularly from an LED, diverges, so that a focusing lens system is required to focus the light into the sample 20 inside the microfluidic channel.
The intensity of fluorescence emissions from the sample and the substrate is proportional to the intensity of excitation light source 10. However, if the intensity of light source 10 is too high, it may cause quick photodestruction or photobleaching of the fluorophore. Experiments have shown that a 470 nm blue LED with 3460 mcd in intensity is sufficient to excite fluorescein.
For the purpose of the noise elimination method proposed in this first embodiment, the light source 10 must be modulated at a frequency such that the phase difference between the fluorescence desired signal and unwanted signal is 90°. The modulation frequency to achieve this condition depends on the difference between the fluorescence lifetime of fluorophore of the sample and substrate. The modulation frequency is calculated using equation (1) as follows:
where Δt is fluorescence lifetime difference between the labeled sample and the substrate.
As an example, if Δt=1 ns, the fmod=250 MHz. From tests, in general, an LED can only be modulated up to 50 MHz. Therefore, if a high frequency modulation is required, a laser source or laser diode may have to be used.
1.2 Phase Delay Generator
The phase delay generator 50 is used to shift the phase of the sine wave signal generated by the sine wave generator 30. The sine wave signal is used to modulate the light source 10. The phase delay may be manually tuned. In this embodiment, the phase delay is carefully tuned to φs, where φs is the phase of the desired fluorescence signal, so that the output of phase delay generator 50 is sin (ωt+φs). This output of the phase delay generator 50 is input into the phase-locked loop circuitry to ensure the phase is locked.
1.3 Phase-Locked Loop (“PLL”) System
A simple PLL system 60 comprises a phase detector 61, low-pass filter (LPF1) 62 and voltage-controlled oscillator (“VCO”) 63. The modulating signal, sin (ωt), generated by the sine wave generator 30 is firstly delayed, either manually or automatically, by a certain phase φs. If the locked condition is achieved, the frequency and the phase of the PLL system output are similar to the frequency and the phase of the desired signal. The phase detector 61 serves as an “error amplifier” in the feedback loop, which minimizes the phase difference, Δφ=φs−φo≈0, between the delayed modulating signal and the output signal of the VCO 63. The loop is considered locked if Δφ≈0 is invariant with time, and the input and output frequencies are equal.
Under the phase-locked condition, all the signals in the loop reach a steady state. The phase detector 61 generates an average DC voltage that is proportional to Δφ. Referring to
The output signal of the VCO 63 may be a unity-gain sine wave signal, sin (ωt+φs), which has the same frequency and phase as the signal to be detected, ys(t).
1.4 Quadrature Phase Shifter
If the modulation frequency for the light source 10 is carefully tuned, the unwanted noise signal yn is 90° out of phase to the desired signal ys. Hence, the output of the optical receiver 40 becomes:
The output of the PLL 60, sin (ωt+φs), is then input to the quadrature phase shifter 70, to generate two sine wave signals which are 90° out of phase with each other, that is, i(t)=sin (ωt+φs) and q(t)=cos (ωt+φs). These two signals are input into a respective mixer and both are mixed with the signal generated by the optical receiver 40 y(t)total. The output from the mixers comprises DC components and AC components. To obtain the DC components only, the AC components of the mixer output must be filtered off. After filtering by two low-pass filters, two respective DC signals, the desired DC signal
and unwanted DC signal due to the fluorescence of the substrate
are generated. The mathematical basis for this technique is described later.
These results show the generation of a pure DC component of the desired/signal
is generated. It also demonstrates that this method is able to filter off the noise to generate only DC components of the desired signal. The DC components of the desired signal are proportional to the peak intensity of the modulated fluorescence desired signal. This technique simplifies the signal measurement and data processing. A volt-meter is able to measure the output of the fluorescence detection system 100. Hence, the value of the DC level of Vi indicates the concentration of the sample.
1.5 Sample and Microfluidic Card
Microfluidic channels are constructed in a microfluidic card. Microfluidic channels based on polycarbonate materials, such as PMMA and Mylar, are relatively cheap and easily fabricated. Thus, polycarbonate materials are an important commercial material for microfluidic applications. Microfluidic cards are often considered a consumable item. However, the fluorescence background of polycarbonate materials is considerably high, particularly compared to the fluorescence intensity of the sample of interest immobilized inside a microfluidic channel.
Referring to
1.6 Optical Receiver
The working region of the optical receiver 40 must cover the fluorescence wavelength of the sample of interest. For example, if the fluorophore used is fluorescein having a fluorescence wavelength range between 500 nm and 600 nm, the optical receiver 40 must be sensitive to this wavelength range. An optical filter, either a long-pass filter or a band-pass filter, may need to be installed before the optical receiver 40, to filter off signals with a wavelength outside the range of interest. As the light source 10 in this embodiment is modulated at a certain modulation frequency, the optical receiver 40 must also be able to respond to the modulated signals at that frequency. There are many types of optical receivers 40 suitable for this embodiment such as photodiodes, avalanche photodiodes, and photomultiplier tubes. A photomultiplier tube is a very sensitive light detector, but the size is relatively bulky, and requires high biasing voltage (1000V). A photodiode is cheap, compact, and simple, but its sensitivity is not as good as a photomultiplier tube and may not be suitable for very weak fluorescence signals. An avalanche photodiode may be used as they are relatively inexpensive, compact, and have good sensitivity. For a multichannel detection system, an array of CCD detectors may be used.
1.7 Step by Step Analysis
The frequency of the light source modulation is carefully tuned so that the phase difference between desired signal ys and unwanted signal yn is 90°. That is, they are 90° out of phase.
φn=φs+90°
After the optical receiver 40, the signal generated by the sample and microfluidic system 100 is:
Referring to
i(t)=sin (ωt+φs) (4)
and
q(t)=cos (ωt+φs) (5)
Mixing of (3) and (4) results in:
Mixing of (3) and (5) results in:
There are DC components and AC components in (6) and (7). The AC components are filtered off using low pass filters 80, leaving only respective DC components, that is:
Recall from
is the desired component, and
is the unwanted signal. Equation (8) shows that this technique is able to help eliminate the noise due to auto-fluorescence of the substrate. Equation (8) also shows that
depends only on amplitudes As which is proportional to the concentration of the labeled sample, so that the magnitude of DC level of Vi is proportional to the concentration of the sample.
Therefore, after the elimination of the noise due to the auto fluorescence of the substrate, the fluorescence signal of the labeled sample remains and can be subsequently processed to provide the necessary or desired output.
If the fluorescence lifetime of a sample of interest is almost similar to the fluorescence lifetime of the substrate, a high frequency system, including a light source 10 and detector, is required in order to apply the technique proposed in the first embodiment. The availability of high frequency optical sources and detectors remains problematic due to limitations of the existing technologies. Where a high frequency system in order of GHz is not viable due to technical or cost constraints, the technique proposed in the first embodiment described above can be modified.
Referring to
Referring to
2.1 Step by Step Analysis
After mixing by the mixer, the desired fluorescence signal ys(t) and the unwanted background fluorescence noise yn(t) are modulated by the switching signal from the PLL 60, as follows.
Hence, the output of the mixer is:
xtotal(t)=xs(t)+xn(t)
The phase difference between signal and noise is Δφ=φn−φs.
Frequency, fmod=ω/2π; and Period, T=1/f
The DC voltage for the desired signal after integration, Vs, is calculated as follows
Hence, the DC fluorescence desired signal is
And the DC voltage of the unwanted signal Vn, noise due to background fluorescence of the substrate, is calculated as follows:
The total DC signal of the fluorescence desired signal, and unwanted noise at the output of integration system 90 illustrated in
2.2 The Ideal Case: yn(t) is completely eliminated and ys(t) is Maximized
To completely eliminate yn(t), the following must be made:
To maximize the desired signal ys(t), make the
cos (φx−φs)=±1φx−φs=k2π (k2=0, ±1, ±2 . . . ) (16)
Thus, from equations (15) and (16), to completely eliminate the noise and maximize the desired signal, the switching signal, u(t), and modulation frequency, fmod, are generated and adjusted such that u(t) is in phase with the desired signal ys(t), and ys(t) has a 90° phase difference from the noise signal yn(t).
Suppose, ideal requirements are selected as follows:
However, the ideal case may not be achievable due to current technical limitations or other constraints. For an example, due to very short lifetime difference between the fluorophore fluorescence lifetime and the microfluidic substrate auto-fluorescence lifetime which may be in order of nm, a high frequency modulation, in order of GHz, is required to achieve the ideal case. If an LED is used as the excitation source, which cannot be modulated at such a high frequency, the ideal condition above is not possible. Although the ideal condition may not be achievable, the technique is still useful to minimize the effect of the unwanted background signal from the substrate. Hence, the technique remains applicable for detecting very low concentrations of fluorescently labeled samples in a high fluorescence background microfluidic substrate both for ideal and non-ideal cases. Below is an analysis based on the general case, the ideal and non-ideal cases.
2.3 General Case
Suppose for simplification, that φs=0, so the phase difference between signal and noise is Δφ=φn. For general cases, φx and φn are arbitrary, so that the DC voltage for signal and noise are:
Equation (20) shows that in order to completely eliminate the noise, (φx−φn)=90°=π/2 (ideal condition). However, sometimes this condition cannot be achieved due to reasons explained previously. Usually, there is a small offset angle (β), so that φx=φn+(π/2+β). Hence, u(t) has following phase relationship with φn, that is, (β is small offset angle):
Thus, the DC voltages for signal and noise become,
Therefore, the output for general cases:
And the signal to noise ratio is:
Equations (23 and 24) show that the smaller the β, the higher the signal to noise ratio
Hence, to minimize the noise level, β has to be made as small as possible. Since φn, β, ω, and An are constant parameters in a measurement, equation (23) shows us Vout is proportional to As which is proportional to the concentration of sample. Hence, Vout is proportional to the concentration of sample of interest.
Referring to
The difference in the third embodiment compared to the first embodiment is that the phase delay is tuned to φn, where φn is the phase of the unwanted fluorescence signal, so that the output of the phase delay generator 50 is sin (ωt+φn), which is input into the phase-locked loop circuitry in order to ensure the phase is locked. The tuning can be carried out by inserting a same microfluidic card, without the sample, and tuning the phase delay generator 50 so that the output of the low pass filter (LPF) 80 is zero. A calibrator 95 can be incorporated to tune the phase delay generator 50 automatically using the output of the low pass filter 80 as the error signal to the calibrator 95.
The frequency of light source excitation is tuned so that there is a phase difference between desired signal ys generated by the sample and unwanted signal yn generated by the microfluidic system 102. Since the phase difference is not required to be 90°, the frequency of excitation is much lower than required for the first embodiment.
3.1 Step by Step Analysis
The signal ytotal generated by the optical receiver 40 is composed of the signal ys generated by the sample and unwanted signal yn generated by the substrate material as follows:
The output q(t) of the quadrature phase shifter 70 is given by:
q(t)=cos (ωt+φn) (26)
Mixing of (25) and (26) results in:
There is a DC component and also AC components in equation (27). Suppose the AC components are filtered off using a low pass filter, only the DC component, which is the desired component Vs, is left given by:
For a fixed frequency of light source modulation, the phase difference (φs−φn) is a constant. Hence the desired component Vs depends only on amplitude As which is proportional to the concentration of the labeled sample. As a result, the magnitude of DC level of Vs is proportional to the concentration of sample. Furthermore, the desired component Vs is zero when an identical microfluidic card without a sample is excited by the modulated light source 10. This is used to tune the phase delay generator 50 automatically by incorporating a calibrator 95 to tune the phase delay generator 50 to lock onto the phase of the unwanted signal yn using Vs as the error signal to the calibrator 95. The phase of the phase delay generator 50 is then fixed by disconnecting the calibrator 95 from the phase delay generator 50 after successful calibration. The testing of an identical microfluidic card with a sample then proceeds with the phase-locked loop locked to the phase of the unwanted signal yn.
Equation (28) shows that this technique is able to eliminate the noise due to auto-fluorescence of microfluidic substrate. The technique is also applicable to eliminate noise yg given by:
The coefficients αk and the phases φk for k=2, . . . , ∞ of the unwanted signal yg can be unknown because mixing (26) with components of yg in equation (29) results in
The resultant signal in equation (30) and A0 cos (ωt+φn) is filtered out using the same low pass filter. The tuning of the phase delay generator 50 is still required in order to lock the fundamental phase of the unwanted signal yg so that the phase-locked loop and the quadrature phase shifter 70 generate the required annihilator signal cos (ωt+φn). The same calibrator 95 may be used to tune the phase delay generator 50 for the unwanted signal yg.
This embodiment is almost the same as that of the first embodiment. The output of phase lock loop 60 is 90 degree out of phase with the input signal. In order to get the same output, as proposed in the first embodiment, another 90 degree phase shift should be added in quadrature demodulator 70, as shown in
4.1 Step by Step Analysis
This is substantially the same as for the first embodiment. After the optical receiver 70, the signal generated by the sample and microfluidic card 20 is:
The output of the PLL 60 is:
VPLL=sin (ωt+φs−90°) (11)
It is fed into the quadrature phase shifter 70, resulting in signals (4) and (5), respectively. Equations (6) and (7) are applied for mixing the output of optical receiver 40 and the two outputs of quadrature phase shifter 70 respectively. The fluorescence signal (8) and unwanted noise signal (9) can be obtained after passing through respective low pass filters.
Phase shift of the signals received from the substrate and the sample depend on the fluorescence life time and modulation frequency. The phase locked loop 60 can be removed in this embodiment. Assuming that the signal from the sine wave generator 30 is stable and controlled in relation to, amongst others, phase and frequency, The phase delay can be turned to φs, where φs is the phase of the wanted fluorescence signal, so that the output of phase delay generator so is sin (ωt+φs), which will be fed into the quadrature phase shifter 70 and results in two quadrature signals. The two signals are then mixed with output of the optical receiver 40, to give the desired DC signal after the low pass filter 80, as shown in
This embodiment is similar to the third embodiment that consists of 6 main blocks, namely the tunable modulated light source system [I], the phase delay generator [11], the phase-locked loop circuitry [III], the inverter [IV], the sample (sample and microfluidic card) [V], and the optical receiver [VI]. It can also be considered as a special case of the third embodiment where its PLL output is 90 degree out of phase with its input. The system architecture is shown in
6.1 Step by Step Analysis
The input and output of the phase lock loop 60 are 90 degree out of phase.
After the inverter (96), the signal becomes
yinverter=cos (ωt+φn);
this is the same as that in (25). Thus, the output of the low pass filter 10 is given by
6.2 Results
The conditions are:
Modulation Frequency: f=100 kHz;
Fluorescence signal: ys=0.5 sin (2π*105t+45.83°);
Noise signal: yn=sin (2π*105t+1.83°);
ytotal=0.5 sin (2π*105t+45.83°)+sin (2π*105t+1.83°); (35)
Before any sample is input into the microfluidic card, calibration takes place in order to tune the phase at phase delay (II), as shown in
The input signal at output of optical receiver 40:
ytotal=yn=sin (2π*105t+1.83°);
The output of PLL 60 (before calibration is done):
yPLL=sin (ωt+φx−90)
where φx is the phase that tuned;
After inverter (96), the signal becomes:
y90=cos (2π*105t+φx)
After the multiplication, the resultant signal is:
So the output signal is:
At the end of calibration, φx=1.18 and yresult=0. The switch as shown in
The resultant signal at the output of the multiplexer 97 is:
And after the LPF 80, the output signal is:
yresult=0.25 sin (44°)=0.1737
As shown in the
Offset appears at the output of optical receiver 40. It is also needed for biasing the electronic components. Offset is added without changing any other parameters:
Modulation Frequency: f=100 kHz;
Fluorescence signal: ys=0.5 sin (2π*105t+45.83°)+1.2;
Noise signal: yn=sin (2π*105t+1.83°)+1.2;
Similarly, calibration takes place before pumping any sample. The input signal is:
ytotal=yn=sin (2π*105t+1.83°)+1.2;
The output of PLL 60 (before calibration is done):
yPLL=sin (ωt+φx−90)+1.2
where φx is the phase that is tuned.
After the multiplication, the resultant signal is:
After LPF 80, the high frequency aspect is filtered, and results in:
At the end of calibration, φx=1.18 and yresult=1.44.
After pumping the samples, the signal becomes the mixture of the two signals, as shown in (36). Thus,
And after the LPF 80, the output signal is:
yresult=0.25 sin (44°)+2.88=3.054
As shown in
This is another special case of the third embodiment, and is relevant when the signal after the sine wave generator 30 is stable and well controlled in so far as, for example, phase and frequency. The phase shift is tuned to be (φn+90°) by using the calibrator 95. Thus, the output of the phase delay can be directly fed into the mixer 97, and after the LPF 80, the desired DC value can be obtained. The system architecture is shown in
7.1 Step by Step Analysis
The signal at the output of the sine wave generator 30 is tuned by (φn+90°), thus,
After calibration, samples are pumped into the microfluidic card 20, and two signals' mixture is produced at the output of the optical receiver 40. The output of the phase delay 50 is directly fed into the mixer 97, and passed through LPF 80 to filter out the high frequency components. The desired DC signal can be obtained. Detail mathematical derivation is the same as that in the third embodiment.
This scheme is almost the same as that in the seventh embodiment. However, in the seventh embodiment tuning (φn+90°) to achieve signal cos (ωt+φn) directly is much more difficult than tuning at φn followed by a 90 degree phase shift. This is an improvement of the seventh embodiment, though one more block (quadrature phase shifter III) 98 is added into the system. The system architecture is shown in
Because this technique enables the elimination of the noise signal due to the fluorescence background of a substrate by using the difference between the fluorescent emission lifetimes of the fluorescein and the substrate, polymeric materials having high background fluorescence may be used as the substrate to reduce cost, simplify fabrication and increase robustness. Polymeric sheets are relatively cheaper than optical grade glasses and silica wafers, which are presently commonly used as substrate. The substrate may be a microfluidic channel of a microfluidic substrate. In that case the cost of the microfluidic chips can be reduced, and they may be able to considered as consumable items by being disposable cartridges at the point-of-care immunoassay system. As a polymeric microfluidic chip is also relatively robust, it may be suitable for less trained users who are the target of point-of-care diagnosis in distributed-diagnostic and home-health-care system.
Whilst there has been described in the foregoing description preferred embodiments of the present invention, it will be understood by those skilled in the technology concerned that many variations or modifications in details of design or construction may be made without departing from the present invention.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/SG2005/000343 | 10/6/2005 | WO | 00 | 10/29/2008 |
Publishing Document | Publishing Date | Country | Kind |
---|---|---|---|
WO2007/040459 | 4/12/2007 | WO | A |
Number | Name | Date | Kind |
---|---|---|---|
5270548 | Steinkamp | Dec 1993 | A |
5340715 | Slovacek et al. | Aug 1994 | A |
20020092973 | Nagle et al. | Jul 2002 | A1 |
Number | Date | Country |
---|---|---|
WO 0109605 | Feb 2001 | WO |
Number | Date | Country | |
---|---|---|---|
20090218515 A1 | Sep 2009 | US |