The present invention relates to electrical measurements and, in particular, to impedance meters.
Referring to
A basic requirement for measurements is that a meter be calibrated so the measurements actually correspond to the desired units of measure. It is important to not only initially calibrate a meter, but also, to keep it calibrated.
A method for verifying the adjustment for the purpose of calibration of an impedance meter having at least a first and a second measurement range includes measuring within the first range a first measured value of a test impedance; measuring within the second range a second measured value of the test impedance; and comparing the first and second measured values to verify the calibration of the impedance meter.
Problems with the calibration of the meter 10 may be identified by measuring a measured impedance value of a test impedance 12 using at least two different measurement ranges. If the first measured impedance value and the second measured impedance value do not correspond (even if the true values for the test impedance are unknown), the calibration of the meter 10 becomes suspect. It is at least out of calibration and perhaps malfunctioning as well.
Referring to
If a resistive test impedance falls within an overlapping area between ranges, the test impedance can be used to check the calibration of those ranges. If it is not in an overlapping area, it may be necessary to have another resistive test impedance, both with a known value so that the ranges may be independently verified.
In the case of a capacitive (or inductive) test impedance, it will likely be possible to use a single test impedance for both ranges. This is accomplished by measuring the test impedance at two different frequencies. The frequencies are chosen so that the magnitude of the respective measured impedance value is within the meter measurement range desired.
Usually, the best measurement results occur when the test impedance is near the value of the current-sensing resistor. Because of this, results can be further improved if the frequencies are further chosen to have the desired impedance value near the respective current-sensing resistor value.
It is not necessarily required to know the actual capacitive value of the test impedance because the same impedance is used for both measurements; it is just scaled by the chosen frequencies. If the scaled values do not correspond the meters results are suspect.
Referring again to
The different test impedances may be external to the meter 10, or they may be conveniently provided within the meter itself to provide a self-test function.
It should be evident that this disclosure is by way of example and that various changes may be made by adding, modifying or eliminating details without departing from the fair scope of the teaching contained in this disclosure. The invention is therefore not limited to particular details of this disclosure except to the extent that the following claims are necessarily so limited.
Number | Name | Date | Kind |
---|---|---|---|
3824461 | Preikschat | Jul 1974 | A |
4105967 | Macemon | Aug 1978 | A |
4963830 | Roth et al. | Oct 1990 | A |
5012181 | Eccleston | Apr 1991 | A |
5345182 | Wakamatsu | Sep 1994 | A |
5463323 | Wakamatsu | Oct 1995 | A |
5793214 | Wakamatsu | Aug 1998 | A |
5818243 | Wakamatsu | Oct 1998 | A |
5886529 | Wakamatsu | Mar 1999 | A |
6510392 | Doi et al. | Jan 2003 | B2 |
6737875 | Davis et al. | May 2004 | B2 |
7614274 | Goeke | Nov 2009 | B1 |
7834641 | Goeke | Nov 2010 | B1 |
7839195 | Feng et al. | Nov 2010 | B1 |
20020087277 | Doi et al. | Jul 2002 | A1 |
20070255992 | Sasagawa | Nov 2007 | A1 |
20080071487 | Zeng et al. | Mar 2008 | A1 |
Number | Date | Country |
---|---|---|
2246639 | Feb 1992 | GB |
62195576 | Aug 1987 | JP |
Entry |
---|
U.S. Appl. No. 11/759,487, filed Jun. 7, 2007, Inventor: Wayne C. Goeke, Title: Phase-Gain Calibration of Impedance/Admittance Meter, 8 pages of specification/claims/abstract and 1 page of drawings. |