This invention is in the field of microfluidics for the analysis of nucleic acids.
There is an unmet need for the development of instruments and technologies that would permit fully integrated (i.e., sample-in to results-out) focused nucleic acid analysis, defined as the rapid identification (by nucleic acid sequencing or fragment sizing) of a subset of a given human, animal, plant, or pathogen genome. Focused nucleic acid sequencing will enable end-users to make real-time clinical, forensic, or other decisions. For example, many common human diseases can be diagnosed based on less than 1000 base pairs of DNA sequence, orders of magnitude less than required to generate a complete human genome. Similarly, precise determination of the sizes of sets of less than 20 specific DNA fragments generated by short tandem repeat analysis is sufficient to identify a given individual. Depending on the application, focused nucleic analysis can be performed in a variety of settings, including hospital laboratories, physician's offices, the bedside, or, in the case of forensic or environmental applications, in the field.
There are several unmet needs for improved DNA sequencing and fragment sizing systems. First, there is an unmet need for DNA sequencing and fragment sizing instruments that are easy to use and do not require highly trained operators. Second, there is an unmet need for systems that eliminate all manual processing. As a result, only minimal operator training would be required and the system could be readily operated by individuals constrained by challenging environments such as would be encountered, for example, by a first responder wearing a haz-mat suit.
Third, there is an unmet need for ultrafast analysis that does not sacrifice the need for complete, accurate, and reliable data. For human identification applications, an appropriate time to result is 45 minutes or less, well under the days to weeks required using conventional technology. For clinical applications such as sequencing infectious agents to determine an appropriate treatment regimen, 90 minutes or less is a reasonable time to answer, allowing treatment with antibacterial and antiviral medications to be initiated shortly after a patient's arrival in an emergency room. Regardless of application, there is an unmet need to generate actionable data in real time. A short time to answer also allows a concomitant increase in sample throughput.
Fourth, there is an unmet need for miniaturization. Many DNA analysis systems require an entire laboratory and related support. For example, the high throughput Genome Sequencer FLX (Roche Diagnostics Corp, Indianapolis, Ind.) DNA sequencing system requires only a benchtop for installation but a large laboratory to perform the required library construction. Miniaturization is important both for laboratory and point-of-care use as well as field operation. It is also important for cost reduction per sample.
Fifth, there is an unmet need for ruggedization. For many applications, particularly those in forensics, the military, and homeland defense, the DNA analysis instrument must be operable in the field. Accordingly, the instrument must be capable of transport whether carried on a soldier's back, driven in a police vehicle, or dropped from a helicopter into a battlefield. Similarly, the instrument must be able to withstand and function under environmental extremes, including temperature, humidity and airborne particulates (e.g., sand).
Sixth, there is an unmet need for systems that can accept multiple sample types and perform highly multiplexed analyses in parallel. For most applications, capability of analysis of DNA from a single sample type in a singleplex reaction is not acceptable to perform meaningful DNA analysis.
Developers of microfluidics (also referred to as micro total analysis systems (μTAS) or lab-on-a-chip technologies, see, Manz et al., Sens. Actuators B 1990, 1, 244-248) who are seeking to condense complex series of laboratory manipulations onto biochips have recognized certain of these unmet needs, but to date, have been unable to design integrated biochips and instruments that perform all of the biochemical and physical processes necessary or desirable to allow microfluidic nucleic acid analysis to address these needs. As a result, focused nucleic acid analysis has not entered into widespread use in society today.
The development of microfluidic systems involves the integration of microfabricated components, such as microscale separations, reactions, microvalves and pumps and various detection schemes into fully functional devices (see, e.g., Pal et al., Lab Chip 2005, 5, 1024-1032). Since Manz et al. (supra), demonstrated capillary electrophoresis on a chip in the early 1990's, others have sought to improve upon it. Several groups have demonstrated integration of DNA processing functionality with biochip separation and detection. Integrated devices in a glass-PDMS (polydimethylsiloxane) hybrid structure have been reported (Blazej et al., Proc Natl Acad Sci USA 2006, 103, 7240-5; Easley et al., Proc. Natl. Acad. Sci. USA 2006, 103, 19272-7; and Liu et al., Anal. Chem. 2007, 79, 1881-9). Liu coupled multiplex polymerase chain reaction (PCR), separation and four dye detection for human identification by short tandem repeat (STR) sizing. Blazej coupled Sanger sequencing reaction, Sanger reaction cleanup, electrophoretic separation and four dye detection for DNA sequencing of pUC18 amplicon. Easley coupled solid phase extraction of DNA, PCR, electrophoretic separation and single color detection to identify the presence of bacterial infection in blood. An integrated silicon-glass device coupling PCR, electrophoretic separation and single color detection was demonstrated by Burns (Pal, 2005, Id.). A hybrid device coupling a glass-PDMS portion for PCR to a poly(methyl methacrylate) (PMMA) portion for electrophoretic separation and single color detection for identifying the presence of bacteria DNA was reported by Huang (Huang et al., Electrophoresis 2006, 27, 3297-305).
Koh et al., report a plastic device that coupled PCR to biochip electrophoretic separation and single color detection for identifying the presence of bacterial DNA (Koh et al., Anal. Chem. 2003, 75, 4591-8). A silicone based device that couples DNA extraction, PCR amplification, biochip electrophoretic separation and single color detection was reported by Asogawa (Asogowa M, Development of portable and rapid human DNA Analysis System Aiming on-site Screening, 18th International Symposium on Human Identification, Poster, Oct. 1-4, 2007, Hollywood, Calif., USA). U.S. Pat. No. 7,332,126 (Tooke et al.) describes the use of centrifugal force to effect microfluidic operations required for nucleic acid isolation and cycle sequencing. However, this approach is based on small sample volumes, (those of the order of one to a few μL). As a result, the device is not useful for the processing of large samples for the isolation and analysis of nucleic acids, especially in highly parallel fashion, because the fluid samples must be applied to the device while stationary, which is, the disc must be able to contain all the fluids required for operation prior to centrifugation (potentially up to 100 s of mL for a highly-parallel device). Secondly, the device is limited to sample preparation and cycle sequencing, of bacterial clones (e.g., plasmid DNA).
There are several deficiencies in those devices that attempt to integrate DNA processing with biochip electrophoretic separation. First, detection is limited by either information content per assay (most use single color detectors although some have up to four color detection systems) or throughput (single sample or two sample capability). Second, these devices do not represent complete sample-to-answer integration, e.g., Blazej's device requires off-board amplification of template DNA prior to cycle sequencing, while others use samples that require pre-processing of some sort (e.g., Easley and Tooke require lysis of the sample before addition). Third, some of the processing choices made for these devices negatively impact time-to-answer: for example, the hybridization-based method of Blazej requires more than 20 minutes for cleanup of the cycle sequencing product. Fourth, many of these devices are fabricated in part or in-whole glass or silicon. The use of these substrates and corresponding fabrication techniques make them inherently costly (Gardeniers et al., Lab-on-a-Chip (Oosterbroeck R E, van den Berg A, Eds.). Elsevier: London, pp 37-64 (2003)) and limit them to applications where reuse of the devices must be performed; for many applications (such as human ID) this leads to the risk of sample contamination. Finally, the demonstrated technology is inappropriate for two applications, human identification via STR analysis and sequencing. For example, the Easley and Pal devices both suffer from poor resolution—much worse than a single base. Fragment sizing applications (e.g., human identification by analysis of short tandem repeat profiles) and sequencing both require single base resolution.
In addition to the limitations of the prior art with respect to microfluidic integration, problems with respect to fluorescence detection also limit the widespread application of nucleic acid analysis beyond conventional laboratory work. The most widely used commercial sequencing kits (BigDye™ v3.1 [Applied Biosystems] and DYEnamic™ ET [GE Healthcare Biosciences Corp, Piscataway, N.J.]) are based on a twenty year old detection method for four color (see, e.g., U.S. Pat. Nos. 4,855,225; 5,332,666; 5,800,996; 5,847162; 5,847,162). This method is based on resolution of the emission signal of a dye-labeled nucleotide into four different colors, one representing each of the four bases. These four-color dye systems have several disadvantages, including inefficient excitation of the fluorescent dyes, significant spectral overlap, and inefficient collection of the emission signals. The four color dye systems are especially problematic because they limit the amount of information that can be gained from a given electrophoretic (or other) separation of sequenced products.
There is an unmet need for a system capable of achieving high information content assays in electrophoretic systems based on separation and detection of DNA fragments by both fragment size and by color (dye wavelengths). The maximum number of DNA fragments that can be distinguished by electrophoresis is determined by the readlength of the separation and the resolution of the device. The maximum number of colors that can be detected is determined in part by the availability of fluorescent dyes and the wavelength discrimination of the detection system. Current biochip detection systems are typically limited to single color, although up to 4 color detection has been reported.
STR analysis for human identification is an example of DNA fragment sizing based on color multiplexing and allows simultaneous analysis up to 16 loci (AmpFlSTR Identifiler kit; Applied Biosystems, Foster City, Calif.) and PowerPlex16 kit (Promega Corporation, Madison, Wis.). Using four or five fluorescent dyes, a single separation channel can discriminate among the sizes of the many allelic variants of each locus. Several fragment sizing applications would require more than 16 fragments to be separated and detected on a single lane. For example, the identification of pathogens by fingerprinting (i.e., the separation and detection of a large number of characteristic DNA fragments) and the diagnosis of aneuploidy by surveying the entire human genome can be accomplished by looking at several dozen or several hundred loci, respectively.
One approach to increasing the number of loci that can be detected in a single separation channel is to broaden the range of fragment sizes generated, in part by increasing the fragment sizes of additional loci. The use of longer fragments for additional loci is not ideal, however, as amplification of larger fragments is more susceptible to inhibitors and DNA degradation, leading to lower yields of longer fragments relative to shorter fragments. Furthermore, the generation of longer fragments also requires an increase in the extension time and hence, an increase in the total assay time. There is an unmet need to increasing the number of loci that can be detected in a given separation channel by increasing the number of dye colors that can be simultaneously detected.
There is an unmet need to increase the capacity of Sanger sequencing separations (and therefore decrease the cost, labor, and space of the process) by increasing the number of DNA sequences that can be analyzed in a single separation channel. In addition, in some applications, multiple DNA fragments are sequenced that generate difficult to read “mixed sequence” data; there is a need to develop an approach in that mixed sequences can be interpreted correctly.
One approach to increasing the capacity of Sanger separation channels and developing the ability to interpret mixed sequences is to increase the number of dye colors utilized in the sequencing reactions. In both DNA sequencing and fragment sizing, multiple fragments labeled with different dyes can be detected at the same time. In general, the separation between peak emission wavelengths of adjacent dyes must be large enough relative to peak width of the dyes. Accordingly, the throughput of each separation channel can, for example, be doubled by utilizing two sets of 4 dyes in two independent sequencing reactions, and combining the products, and separating them on a single channel. This methodology requires the use of a total of 8 dye colors, with the first sequence reaction using a set of 4 dye colors applied to label the dideoxynucleuotide terminators, and the second reaction another set of 4 dye colors applied to the label the terminators; each set of dye colors is independent so that no overlap in interpretation of the two sequences is possible. Using this same approach, a set of 12 dyes can be utilized to allow simultaneous analysis of the sequence of three DNA fragments in a single channel, a set of 16 dyes allows the analysis of four sequences, and so on, dramatically increasing the information capacity of Sanger separations.
The novel instruments and biochips of this application satisfy many unmet needs, including those outlined above.
This invention provides fully integrated microfluidic systems to perform nucleic acid analysis. These processes include sample collection, DNA extraction and purification, amplification (that can be highly multiplexed), sequencing, and separation and detection of the DNA products.
The separation and detection modules of this invention are ruggedized and capable of better than single base resolution. They are capable of detecting six or more colors and as such are useful for generating high information content from sequencing and fragment sizing applications.
Highly multiplexed rapid PCR on biochips is the subject of an U.S. patent application, filed on Apr. 4, 2008, assigned U.S. Ser. No. 12/080,746, and entitled, “M
Accordingly, in a first aspect, the invention provides optical detectors comprising one or more light sources positioned for illuminating one or a plurality of detection positions on a substrate; one or a plurality of first optical elements positioned for collecting and directing light emanating from the detection positions on the substrate; and a light detector positioned to accept light from the first optical elements, wherein the light detector comprises a wavelength dispersive element to separate the light from the first optical elements according to light wavelength and positioned to provide a portion of the separated light to the detection elements, wherein each of the detection elements are in communication with a first control element for simultaneously collecting detection information from each of the detection elements and, wherein said light detector detects fluorescence from at least 6 dyes labeled to one or more biological molecules, each dye having a unique peak emission wavelength.
In a second aspect, the invention provides systems for separation and detection of biological molecules comprising, a component for simultaneously separating a plurality of biological molecules in one or a plurality of channels on a substrate, wherein each channel comprises a detection position; one or more light sources positioned to illuminate the detection positions on the substrate; one or a plurality of first optical elements positioned for collecting and directing light emanating from the detection positions; and a light detector positioned to accept light directed from the first optical elements, wherein the light detector comprises a wavelength dispersive element to separate the light from the first optical elements according to light wavelength and positioned to provide a portion of the separated light to the detection elements, wherein each of the detection elements are in communication with a first control element for simultaneously collecting detection information from each of the detection elements and, wherein said light detector detects fluorescence from at least 6 dyes labeled to one or more biological molecules, each dye having a unique peak wavelength.
In a third aspect, the invention provides methods for separating and detecting a plurality of biological molecules comprising, providing one or a plurality of analysis samples into one or a plurality of microfluidic channels on a substrate, wherein each microfluidic channel comprises a detection position, and each analysis sample independently comprises a plurality of biological molecules, each independently labeled with one of at least 6 dyes, each dye having a unique peak wavelength; simultaneously separating the plurality of labeled biological molecules in each microfluidic channel; and detecting the plurality of separated target analytes in each microfluidic channel by, illuminating each detection position with a light source; collecting the light emanating from each detection position; directing the collected light to a light detector; and (i) separating the collected light by light wavelength; and (ii) simultaneously detecting the fluorescence from at least 6 dyes labeled to one or more biological molecules, each dye having a unique peak wavelength.
In a fourth aspect, the invention provides integrated biochip systems comprising (a) a biochip comprising one or a plurality microfluidic systems, wherein each microfluidic system comprises a first reaction chamber in microfluidic communication with a separation chamber, wherein the first reaction chamber is adapted for nucleic acid extraction; nucleic acid purification; pre-nucleic acid amplification cleanup; nucleic acid amplification; post-nucleic acid amplification cleanup; pre-nucleic acid sequencing cleanup; nucleic acid sequencing; post-nucleic acid sequencing cleanup; reverse transcription; pre-reverse transcription cleanup; post-reverse transcription cleanup; nucleic acid ligation; nucleic acid hybridization; or quantification; and the separation chamber comprises a detection position; and (b) a separation and detection system comprising, (i) a separation element for simultaneously separating a plurality of target analytes in the separation chambers; (ii) one or more light sources positioned to illuminate the detection positions on the biochip; (iii) one or a plurality of first optical elements positioned for collecting and directing light emanating from the detection positions; and (iv) a light detector positioned to accept light directed from the first optical elements, wherein the light detector comprises a wavelength dispersive element to separate the light from the first optical elements according to light wavelength and positioned to provide a portion of the separated light to at least six detection elements, wherein each of the detection elements are in communication with a first control element for simultaneously collecting detection information from each of the detection elements; and wherein said light detector detects fluorescence from at least 6 dyes labeled to one or more biological molecules, each dye having a unique peak wavelength.
In a fifth aspect, the invention provides integrated biochip systems biochip system comprising (a) a biochip comprising one or a plurality microfluidic systems, wherein each microfluidic system comprises a first reaction chamber in microfluidic communication with a separation chamber, wherein the first reaction chamber is adapted for nucleic acid extraction; nucleic acid purification; pre-nucleic acid amplification cleanup; nucleic acid amplification; post-nucleic acid amplification cleanup; pre-nucleic acid sequencing cleanup; nucleic acid sequencing; post-nucleic acid sequencing cleanup; reverse transcription; pre-reverse transcription cleanup; post-reverse transcription cleanup; nucleic acid ligation; nucleic acid hybridization; or quantification; and the separation chamber comprises a detection position; and (b) a separation and detection system comprising, (i) a separation element for simultaneously separating a plurality of biological molecules comprising DNA sequences, in the separation chambers; (ii) one or more light sources positioned to illuminate the detection positions on the biochip; (iii) one or a plurality of first optical elements positioned for collecting and directing light emanating from the detection positions; and (iv) a light detector positioned to accept light directed from the first optical elements, wherein the light detector comprises a wavelength dispersive element to separate the light from the first optical elements according to light wavelength and positioned to provide a portion of the separated light to at least six detection elements, wherein each of the detection elements are in communication with a first control element for simultaneously collecting detection information from each of the detection elements; and wherein said light detector detects fluorescence from at least 8 dyes labeled to one or more DNA sequences, each dye having a unique peak wavelength, said dyes being members of at least two 4-dye containing subsets, such that said dye sets are capable of detecting at least two DNA sequences in a single channel, wherein the number of dyes is a multiple of four, and the number of DNA sequences to be detected is equal to that multiple, such that each of the different dyes is present in only one subset.
a is an illustration showing samples and PCR reagents have been loaded into a biochip of the invention.
b is an illustration showing sample delivery through channels to sample chambers (they are shown at different positions along the sample channels in order to illustrate the flow path.)
c is an illustration showing the samples in the sample chambers.
d is an illustration showing delivery of PCR reagents to reagent chambers.
e is an illustration showing withdrawal of excess PCR reagent.
f and 8g are illustrations showing the initial mixing step and retention of the liquids by the first set of capillary valves.
h through 8j are illustrations showing the mixed liquids delivered to the PCR chamber, at that point thermal cycling is initiated.
a through 9e are illustrations showing the delivery of cycle sequencing reagent to metering chambers in layer 1 and removal of excess reagent from the vicinity of the chambers.
f and 9g are illustrations showing the introduction of PCR product into a Sanger reaction chamber.
h-9k are illustrations showing mixing of Sanger reagent with PCR product by reciprocal motion.
a and 12b are illustrations showing delivery of a Sanger product to the UF input chambers.
c is an illustration showing the sequencing product delivered to the filtration chamber.
d is an illustration showing is nearly complete filtration of the sequencing product.
e through 12g are illustrations showing delivery of wash to the UF input chambers and subsequent removal of excess wash from the delivery channel.
h is an illustration showing the beginning of the first wash cycle; it is followed by filtration as in
i and 12j are illustrations showing elution liquid (the same liquid as the wash) being delivered to the UF input chambers.
k through 12m are illustrations showing a single cycle of pressurizing the UF input chambers with the output ports closed, and then releasing the pressure to cause reciprocal motion.
n is an illustration showing purified product ready for further processing or removal.
A. General Description of Integration
Using microfluidics allows fabrication of features to perform more than one function on a single biochip. Two or more of these functions can be connected microfluidically to enable sequential processing of a sample; this coupling is termed integration.
There is a range of possible functions or component processes that must be integrated to achieve any given application, though not all processes must be implemented for any given application. As a result, the chosen integration methods must be appropriate for effectively connecting a number of different component processes in different sequences. Processes that can be integrated include, but are not limited to, the following:
B. General Description of Integration Components and their Uses
1. Sample Collection and Initial Processing For many applications, the following discrete components are advantageously integrated into the biochip: sample insertion; removal of foreign matter; removal of interfering nucleic acids; and concentration of cells of interest. Generally, a pre-processing component of the biochip accepts samples, performs initial removal of particulates and foreign nucleic acid containing cells, and concentrates the cells of interest into small volumes. One approach is to use a sample tube that can accept a swab (e.g., resembling a “Q-tip”) and that is filled with lysis solution to perform the lysis and extraction step. The swab can be placed in contact with a number of cell-containing sites, including a bloodstain, a fingerprint, water, an air filter, or a clinical site (e.g., buccal swab, wound swab, nasal swab). The interface of these tubes with other components of the biochip may include a filter for removal of foreign matter. Another approach is to use a large-volume blood or environmental sample acquisition cartridge, which processes a 1-100 mL of sample. In the case of blood, a leukocyte reduction medium can remove human white blood cells and interfering DNA while passing microbes containing nucleic acids of interest. For environmental samples, large-mesh filters can be used to remove dust and dirt, while small-mesh filters (e.g., filters of <20 μm, <10 μm, <5 μm, <2.5 μm, <1 μm, <0.5 μm, <0.2 μm, <0.1 μm) can be used to trap microbes, concentrating them in a small volume. These pre-processing components can be separate consumables or can be attached to the integrated biochip at time of manufacture. Alternatively, the biochip can be designed to perform differential lysis to separate cells by type (e.g., sperm from vaginal epithelial cells or red blood cells from bacteria).
2. Lysis and Extraction
A variety of lysis and extraction methods can be employed. For example, a typical procedure involves the application of heat after mixing of the sample with a small quantity of a degradative enzyme such as proteinase-K, which breaks down cell walls and releases nucleic acids. Other useful methods are sonication and ultrasonication, either or both performed sometimes in the presence of beads.
For example, lysis and extraction can be performed on a sample containing 106 cells or less. Depending on the application, a smaller number of starting cells can be utilized in the biochips and methods of the invention, less than 105, less, than 104, less than 103, less than, 102, less than 10, and, in cases when multi-copy sequences are to be analyzed, less than 1.
3. Purification of Nucleic Acids
One form of nucleic acid purification can be achieved by inserting a purification medium between an input and output channel. This purification medium can be silica fiber based and use chaotropic-salt reagents to lyse the biological sample, expose the DNA (and RNA) and bind the DNA (and RNA) to the purification media. The lysate is then transported via the input channel through the purification medium to bind the nucleic acids. Bound nucleic acid is washed by an ethanol based buffer to remove contaminants. This can be accomplished by flowing wash reagents via the input channel through the purification membrane. Bound nucleic acid is then eluted from the membrane by flowing an appropriate low salt buffer (e.g., Boom U.S. Pat. No. 5,234,809). A variation of this method involves the use of a differently-configured solid phase. For example, silica gel can be employed to bind nucleic acid. Paramagnetic silica beads can be used, and their magnetic properties employed to immobilize them against a channel or chamber wall during binding, wash, and elution steps. Non-magnetized silica beads may also be employed, either packed within a tight ‘column’ where they are retained by frits (typically manufactured into the plastic of the device, but these may also be inserted during assembly) or “free” during certain phases of their operation: Free beads can be mixed with nucleic acids and then flowed against a frit or a weir in the device to trap them so that they do not interfere with downstream processes. Other formats include sol-gels with silica particles distributed in the gel medium and polymer monoliths with silica particle inclusions, in which the carrier is cross-linked for greater mechanical stability. Essentially, any nucleic acid purification method that is functional in a conventional setting can be adapted to the integrated biochips of this invention.
4. Nucleic Acid Amplification
A variety of nucleic acid amplification methods can be employed, such as PCR and reverse-transcription PCR, which required thermal cycling between at least two, and more typically, three temperatures. Isothermal methods such as strand displacement amplification can be used, and multiple displacement amplification can be used for whole genome amplification. The teachings of the U.S. patent application entitled “M
5. Nucleic Acid Quantification
One approach to quantification in a microfluidic format is based upon real-time PCR. In this method of quantification, a reaction chamber is fabricated between an input and output channel. The reaction chamber is coupled to a thermal cycler and an optical excitation and detection system is coupled to the reaction chamber to allow fluorescence from the reaction solution to be measured. The amount of DNA in the sample is correlated to the intensity of the fluorescence from the reaction chamber per cycle. See, e.g., Heid et al., Genome Research 1996, 6, 986-994. Other quantitation methods include the use of intercalating dyes such as picoGreen, SYBR, or ethidium bromide, either prior to or after amplification, which may then be detected using either fluorescence or absorbance.
6. Secondary Purifications
For STR analysis, multiplex-amplified and labeled PCR product can be used directly for analysis. However, electrophoretic separation performance can be greatly improved by purification of the product to remove ions necessary for PCR that interfere with the separation or other subsequent steps. Similarly, purification following cycle sequencing or other nucleic acid manipulations can be useful. Collectively, any purification step following the initial extraction or purification of nucleic acid can be considered a secondary purification. A variety of methods can be employed, including ultrafiltration, in that small ions/primers/unincorporated dye labels are driven through a filter, leaving the desired product on the filter that then can be eluted and applied directly to the separation or subsequent module. Ultrafiltration media include polyethersulfone and regenerated cellulose “woven” filters, as well as track-etch membranes, in which pores of highly-uniform size are formed in an extremely thin (1-10 μm) membrane. The latter have the advantage of collecting product of size larger than the pore size on the surface of the filter, rather than capturing the product at some depth below the surface. The amplified nucleic acids may also be purified using the same methods outlined above (i.e., classic solid phase purification on silica). Still further methods include hydrogels, cross-linked polymers that have the property of pore size variability, that is, the pore size changes in response to environmental variables such as heat and pH. In one state, the pores are tight and PCR product cannot pass through. As the pores dilate, hydrodynamic or electrophoretic flow of product through the pores is possible. Another method is the use of hybridization, either non-specific hybridization of product to random DNA immobilized on a surface (such as the surface of beads) or specific hybridization, in that a complement to a sequence tag on the product is on the solid surface. In this approach, the product of interest is immobilized through hybridization and unwanted material removed by washing; subsequent heating melts the duplex and releases the purified product.
7. Cycle Sequencing Reaction
Classic cycle sequencing requires thermal cycling, much as PCR. The preferred methods are those employing dye-labeled terminators, such that each extension product bears a single fluorescent label corresponding to the final base of the extension reaction.
8. Injection, Separation, and Detection
Injection, separation and detection of labeled nucleic acid fragments into the electrophoresis channel can be performed in a variety of ways, which have been described in the U.S. patent application entitled “P
C. Integration Methods
The biochip also contains several different means for integrating the functional modules. These means involve the transport of liquids from point to point on the biochip, the control of flow rates for processes that are flow-rate dependent, (e.g., some washing steps, particle separation, and elution), the gating of fluid motion in time and space on the biochip (e.g., through the use of some form of valve), and the mixing of fluids.
A variety of methods can be used for fluid transport and controlled fluid flow. One method is positive-displacement pumping, in that a plunger in contact with either the fluid or an interposing gas or fluid drives the fluid a precise distance based on the volume displaced by the plunger during the motion. An example of such a method is a syringe pump. Another method is the use of integrated elastomeric membranes that are pneumatically, magnetically, or otherwise actuated. Singly, these membranes can be used as valves to contain fluids in a defined space and/or prevent premature mixing or delivery of fluids. When used in series, however, these membranes can form a pump analogous to a peristaltic pump. By synchronized, sequential actuation of membranes fluid can be “pushed” from its trailing side as membranes on the leading side are opened to receive the moving fluid (and to evacuate any displaced air in the channels of the device). A preferred method for actuation of these membranes is pneumatic actuation. In such devices, the biochip is comprised of fluidic layers, at least one of that has membranes, one side of that is exposed within the fluid channels and chambers of the device. The other side of the membrane is exposed to a pneumatic manifold layer that is plumbed to a pressure source. The membranes are opened or closed by the application of pressure or vacuum. Valves that are normally open or normally closed can be used, changing state under the application of pressure or vacuum. Note that any gas can be used for actuation, as the gas does not contact the fluids under analysis.
Yet another method for driving fluids and controlling flow rates is to apply vacuum or pressure directly on the fluids themselves, by altering the pressure at the leading, trailing, or both menisci of the fluid. Appropriate pressures (typically in the range of 0.05-3 psig) are applied. Flow rates also can be controlled by properly sizing the fluidic channels, as the flow rate is proportional to the pressure differential across the fluid and the hydraulic diameter to the fourth power and inversely proportional to the length of the channel or the liquid plug and the viscosity.
Fluid gating can be achieved using a variety of active valves. The former can include piezoelectric valves or solenoid valves that can be directly incorporated into the chip, or applied to the biochip such that ports on the main chip body communicate with the valves, directing fluid into the valves and then back into the chip. One drawback to these types of valves is that for many applications, they are likely to be difficult to manufacture and too expensive to incorporate into disposable integrated devices. A preferable approach is to use of membranes as valves, as discussed above. For example, membranes actuated by 10 psig can be used to successfully contain fluids undergoing PCR.
In some applications, capillary microvalves, which are passive valves, can be preferable. Essentially, microvalves are constrictions in the flow path. In microvalves, surface energy and/or geometric features such as sharp edges can be used to impede flow when the pressure applied to the fluid is below a critical valve, termed the burst pressure, which is generally given by the relation:
P
valveα(γ/dH)*sin(θc)
where γ is the surface tension of the liquid, dH is the hydraulic diameter of the valve (defined as 4 * (cross-sectional area)/cross-sectional perimeter), and θc is the contact angle of the liquid with the valve surface.
Properties that make passive valves preferable for certain applications include: extremely low dead-volume (typically in the picoliter range), and small physical extent (each being only slightly larger than the channels leading to and from the valve). Small physical extent allows for a high density of valves on a given surface of the biochip. Additionally, certain capillary valves are very simple to manufacture, consisting essentially of a small hole in a sheet of plastic, with or without a surface treatment. Judicious use of capillary valves can reduce the total number of membrane valves required, simplify the overall manufacture and create a robust system.
Capillary valves implemented in devices of the invention are of two types: In-plane valves, in that the small channels and sharp corners of the valves are formed by creating “troughs” in one layer and bonding this layer to a featureless lid (typically another layer of the device); and through-hole valves, in that small (typically 250 μm or less) holes are made in an intermediate layer between two fluidics-carrying layers of the device. In both cases, treatment with fluoropolymer can be used to increase the contact angle of fluids in contact with the valves.
In machined prototypes, through-hole valves are both easier to implement and exhibit greater valving pressures, as illustrated in
Mixing can be accomplished in a variety of ways. First, diffusion can be used to mix fluids by co-injecting the two fluids into a single channel, usually of small lateral dimension and of sufficient length such that the diffusion time
t
D=(width)2/(2*Diffusion constant)
is satisfied at the given flow rates. Unfortunately, this type of mixing is typically inadequate for mixing large volumes quickly, because the diffusion or mixing time scales with the channel width squared Mixing can be enhanced in a variety of ways, such as lamination, in that the fluid stream is divided and recombined. (Campbell and Grzybowski Phil. Trans. R. Soc. Lond. A 2004, 362, 1069-1086); or through the use of fine microstructure to create chaotic advection within the flow channel (Stroock et al., Anal. Chem. 2002. 74, 5306-4312). In systems using active pumps and valves, mixing can be accomplished by cycling fluid between two points on the device multiple times. Finally, the latter also can be accomplished in systems using capillary valves. A capillary valve disposed between two channels or chambers acts as a pivot for fluid flow; as fluid flows from one channel into the other through the capillary, the trailing meniscus is trapped if sufficiently low pressure is used to pump the fluid. Reversal of the pressure drives the fluid back into the first channel, and it is again pinned at the capillary. Multiple cycles can be used to efficiently mix components.
Approaches to separation and detection in microfluidic formats are described in the U.S. patent application entitled “P
The upper portion of
D. Fabrication Methods
The devices of the invention can be primarily composed of plastics. Useful types of plastics include, but are not limited to: cyclic olefin polymer (COP); cyclic olefin copolymer (COC); (both of that have excellent optical quality, low hygroscopicity, and high operating temperatures when of sufficient molecular weight); poly(methyl methacrylate) (PMMA) (readily machinable and can be obtained with excellent optical properties); and polycarbonate (PC) (highly-moldable with good impact resistance and a high operating temperature). More information about materials and fabrication methods are contained in the U.S. patent application entitled “M
A variety of methods can be used to fabricate the individual parts of the biochip and to assemble them into a final device. Because the biochip can be composed of one or more types of plastic, with the possible inclusion of inserted components, the methods of interest pertain to creation of individual parts followed by post-processing of parts and assembly.
Plastic components can be fabricated in several ways, including injection molding, hot embossing and machining Injection molded parts can be comprised of both gross features (such as fluid reservoirs) as well as fine features (such as capillary valves). In some cases, it can be preferable to create fine features on one set of parts and larger features on another set, because the approaches to injection molding of these differently-sized features can vary. For large reservoirs (measuring several (about 1-50 mm) mm on a side and with depths of several mm (about 1-10 mm) and capable of accommodating 100 s of μL), conventional molding can be employed using machined injection molding tools, or tools created by burning into a steel or other metal using a graphite electrode that has been machined to be a negative of the tool.
For fine features, both tool creation and molding process can be varied. Tools are typically created using a lithographic process on a substrate of interest (for example, isotropic etch of glass, or deep reactive ion etching or other processes on silicon). The substrate can then be electroplated with nickel (usually after deposition of a chromium layer to promote adhesion) and the substrate removed, for example, by etching in an acid. This nickel “daughter” plate is the injection molding tool. The molding process can be somewhat different than above, as well: For fine, shallow features, compression-injection molding, in which the mold is physically compressed slightly after plastic has been injected into the cavity, has been found to be superior to standard injection molding in terms of fidelity, precision, and reproducibility.
For hot embossing, similar issues regarding gross and fine features as discussed above hold, and tools can be created as above. In hot embossing, plastic resin in the form of pellets, or as a pre-formed blank of material created through molding or embossing, can be applied to the tool surface or a flat substrate. A second tool may then brought into contact at precisely controlled temperature and pressure in order to raise the plastic above its glass transition temperature and to cause material flow to fill the cavities of the tool(s). Embossing in a vacuum can avoid the problem of air becoming trapped between tool and plastic.
Machining also can be employed to create parts. High-speed computer numerical controlled (CNC) machines can be used to create many individual parts per day from either molded, extruded, or solvent-cast plastic. Proper choice of milling machine, operating parameters, and cutting tools can achieve high surface quality (surface roughnesses of 50 nm are achievable in high-speed milling of COC (Bundgaard et al., Proceedings of IMechE Part C: J. Mech. Eng. Sci. 2006, 220, 1625-1632). Milling can also be used to create geometries that can be difficult to achieve in molding or embossing and to readily mix feature sizes on a single part (for example, large reservoirs and fine capillary valves can be machined into the same substrate). Another advantage of milling over molding or embossing is that no mold-release agents are needed to release the fabricated part from a molding tool.
Post-processing of individual parts includes optical inspection (that can be automated), cleaning operations to remove defects such as burrs or hanging plastic, and surface treatment. If optical-quality surfaces are required in machined plastic, polishing with a vapor of a solvent for the plastic can be used. For example, for PMMA, dichloromethane can be used, while for COC and COP, cyclohexane or toluene can be used.
Prior to assembly, surface treatments can be applied. Surface treatment can be performed to promote or reduce wetting (i.e., to change the hydrophilicty/hydrophobicity of the part); to inhibit the formation of bubbles within microfluidic structures; to increase the valving pressure of capillary valves; and/or to inhibit protein adsorption to surfaces. Coatings that reduce wettability include fluoropolymers and/or molecules with fluorine moieties that are exposed to the fluid when the molecules are adsorbed or bonded to the surfaces of the device. Coatings can be adsorbed or otherwise deposited, or they can be covalently linked to the surface. The methods that can be used to make such coatings include dip coating, passing coating reagent through the channels of the assembled device, inking, chemical vapor deposition, and inkjet deposition. Covalent bonds between coating molecules and the surface can be formed by treatment with oxygen or other plasma or UV-ozone to create an activated surface, with either subsequent deposition or co-deposition of the surface treatment molecule on the surface (see, Lee et al. Electrophoresis 2005, 26, 1800-1806; and Cheng et al., Sensors and Actuators B 2004, 99, 186-196.)
Assembly of component parts into the final device can be performed in a variety of ways. Inserted devices, such as filters, can be die-cut and then placed with a pick-and-place machine.
Thermal diffusion bonding can be used, for example for the bonding of two or more layers of the same material, each of that is of uniform thickness. Generally, the parts can be stacked and the stack placed into a hot press, where the temperature can be raised to the vicinity of the glass transition temperature of the material comprising the parts, to cause fusion at the interfaces between the parts. An advantage of this method is that the bonding is “general”, i.e., any two stacks of layers of roughly the same dimensions can be bonded, regardless of the internal structure of the layers, because heat and pressure are applied uniformly across the layers.
Thermal diffusion bonding may also be used to bond more complex parts, such as those that are not planar on their bonding or opposing surfaces, by using specially-created bonding cradles. Such cradles conform to the outer surface of the layers to be bonded.
Other bonding variations include solvent-assisted thermal bonding, in that a solvent such as methanol partially solubilizes the plastic surface, enhancing bond strength at a lower bonding temperature. A further variation is the use of spin-coated layers of lower-molecular weight material. For example, a polymer of the same chemical structure but of a lower molecular weight than the substrate components can be spun onto at least one layer to be bonded, the components assembled, and the resulting stack bonded, by diffusion bonding. During thermal diffusion bonding, the low-molecular weight components can pass through their glass transition temperature at a lower temperature than the components and diffuse into the substrate plastic.
Adhesives and epoxies can be used to bond dissimilar materials and are likely to be used when bonding components fabricated in different ways. Adhesive films can be die cut and placed on components. Liquid adhesive may also be applied through spin-coating Inking of adhesive onto structured parts (such as in nanocontact printing) can be successfully used to apply adhesive to structured surfaces without a need to “direct” the adhesive onto particular areas.
In one example, a biochip of the invention can be assembled as shown in
An integrated biochip for DNA extraction and amplification by PCR is shown in
The biochip was constructed of 4 layers of thermoplastic as shown in
The channels of the biochip were of cross-sectional dimensions ranging from 127 μm×127 μm to 400 μm×400 μm, while reservoirs ranged from 400 μm×400 μm in cross-section to 1.9×1.6 mm; both channels and reservoirs extend for distances as short as 0.5 mm to several 10 s of mm. The capillary valves used in the biochip were of 127 μm×127 μm size for “in-plane” valves and 100 μm in diameter for through-hole capillary valves.
Certain channels, reservoirs, and capillary valves of the four machined layers were treated with a hydrophobic/olephobic material, PFC 502A (Cytonix, Beltsville, Md.). Surface treatment was performed by coating with a wetted Q-tip followed by air-drying at room temperature. The dried fluoropolymer layer was less than 10 μm thick as determined by optical microscopy. Surface treatment serves two purposes: to prevent the formation of bubbles within liquids, especially within low-surface-tension liquids, such as cycle sequencing reagent, which can occur as the liquid rapidly wets the walls of channels or chambers (and “closes off” a bubble before the air can be displaced), and to enhance the capillary burst pressure at that capillary valves resist liquid flow. The regions left untreated were the thermal cycling chambers for PCR and cycle sequencing.
After surface treatment, the layers were bonded as shown in
Pneumatic instrumentation was developed for driving fluids within the biochips of the invention. Two small peristaltic pumps provided pressure and vacuum. Positive pressure output was divided among three regulators that have the range of approximately 0.05-3 psig. The vacuum was ported to a regulator with an output vacuum of approximately (−0.1)-(−3) psig. A fourth, higher pressure was taken from a cylinder of N2 to a further regulator or alternatively from a higher-capacity pump. The positive and negative pressures were applied to a series of 8 pressure-selector modules. Each module was equipped with solenoid valves that could choose an output pressure to be transmitted to the biochip from among the 5 inputs. The output pressure lines terminated on at least one pneumatic interface. This interface clamped to the chip with O-rings positioned over the chip ports on the input side of the chip (the ports along the top of the figures).
Immediately above the biochip ports were additional solenoid valves (i.e., gate valves; 8 per interface) that accept the output pressure lines from the pressure-selector modules. These valves, in close proximity, to the chip provide a low dead-volume interface (approximately 13 μL) between the pressure line and the chip. A low dead-volume interface can prevent unintentional motion of certain liquids on the biochip when pressure is applied to move other liquids (the small gas volume between the liquid plug and the closed valve determines the maximum amount the plug can move, for example, due to compression of the gas as pressure is applied). All pressure-selector valves and gate valves were operated under computer control using a script-based LabView™ program. An important feature of this system is that short pressure cycles times are possible. Some fluidic control events could be performed that required pulses of pressure as short as 30 msec and/or complex pressure profiles could be utilized where pressure could be switched from one value to another (i.e., one regulator to another) rapidly (that is, with time lags of no more than 10-20 msec).
The samples consisted of a bacterial suspension of approximately 106 cells/mL of E. coli DH5 transformed with pGEM sequencing plasmid insert (pUC18 sequencing target). PCR reagent consisted of dNTPs KOD Taq Polymerase (Novagen, Madison, Wis.) at concentration 0.1 μm
A 1.23 μL sample of the bacterial suspension was added to each of the four ports 104, each comprising through holes 202 and 336 in layers 1 and 2, respectively. The sample then resided in sample channels 303 in layer 2. Next, 10 μL of PCR reagent was added to port 105, comprising of through holes 217 and 306 in layers 1 and 2. The PCR reagent then resided in chamber 307 in layer 2 (see,
In operation, air displaced by samples and downstream processes (such as metering of reagents, mixing of fluids) was evacuated through ports on the output end of the chip, 108, comprised of through-holes 227. The final volume of the PCR reaction can be increased or decreased as desired.
The biochip was placed in the pneumatic manifold described above. The following automated pressure profile was performed with no delays between steps. Unless otherwise noted, the pneumatic interface valves, corresponding to ports along the input side of the chip, were closed during all steps.
A pressure of 0.12 psig was applied to ports 104 for 15 sec to drive the samples down channels 303 to through-hole 304. The samples passed through through-hole 304 and emerged on the other side of layer 2 in sample chamber 204 of layer 1 and were driven to the first mixing junction 205. At the first mixing junction the samples were retained by capillary valves 210 (see,
A pressure of 0.12 psig was applied to port 105 for 10 sec to drive the PCR reagent through through-hole 320. The PCR reagent emerged on the other side of layer 2 in distribution channel 208, and moved into the metering chambers 209, which define a volume of reagent equal to the sample volume, where they were retained by capillary valves 211 at mixing junction 205. (see,
A pressure of 0.12 psig was applied to port 107 (comprised of through-holes 203 and 305) with port 105 open to atmosphere for 3 sec to empty channel 208 (see,
A pressure of 0.8 psig was applied to ports 107 and 105 for 0.03 sec and a pressure of 0.7 psig was simultaneously applied to ports 104 for 0.03 sec to initiate mixing of the samples and PCR reagents by bursting liquids past the capillary valves 210 and 211 (see,
A pressure of 0.12 psig was applied to ports 104 and 107 for 10 sec to pump the samples and PCR reagents into mixing channels 214, with retention at capillary valves 210 and 211. Passage through the mixing bulbs 212 into the constrictions 213 created added hydraulic resistance to flow, decreasing the high velocity imparted by the previous high pressure pulse.
A pressure of 0.7 psig was applied to ports 104 and 107 for 0.03 sec to detach the liquid from capillary valves 210 and 211 (see
A pressure of 0.12 psig was applied for 3 sec to ports 104 and 107 to pump liquid through mixing channel 214 to capillary valves 219, where they were retained (see,
A pressure of 0.7 was psig applied for 0.1 sec to ports 104 and 107 to drive the mixture of the samples and PCR reagents through through-holes 315 and 402 and through the body of layers 2 and 3, and into PCR chamber 502 (see,
A pressure of 0.12 psig was applied for 3 sec to ports 104 and 107 to complete pumping of the mixture of the samples and PCR reagents into chamber 502. The leading edge of the mixture of the samples and PCR reagents then passed through through-holes 403 and 316, emerged into layer 1, and was pinned at capillary valve 220 (see,
The biochip was then pressurized to 30 psig N2 and thermally cycled for PCR amplification via a Peltier using a gas bladder compression mechanism as described the U.S. patent application entitled, “M
Sample, reagent volumes, and PCR chamber sizes were chosen such that the liquid filled the region between valves 219 and valves 220. As a result there liquid/vapor interfaces of small cross sectional area (typically 127 μm×127 μm) were located approximately 3 mm from the thermally cycled bottom surface of layer 4. The application of pressure during thermal cycling inhibited outgassing by dissolved oxygen in the sample. The small cross-sectional area of the liquid/vapor interface and distance from the Peltier surface both inhibited evaporation.
The observed temperature at the top of the biochip during cycling never exceeded 60° C., and, as a result, the vapor pressure at the liquid/vapor interfaces was significantly lower than it would have been for such interfaces if they were within the PCR chamber. For a 2 μL sample, 1.4 μL of which is within chamber 502 and the remainder is within the through-holes and capillary valves, the observed evaporation was less than 0.2 μL over 40 cycles of PCR. The volume of non—cycled fluid-0.6 μL in this case—can be reduced by the choice of smaller diameters for the through-holes.
PCR was performed using the following temperature profile:
The biochips described in Example 1 were used. PCR product generated in tubes using the protocol outlined in Example 1 was added to both sample and PCR reagent ports of the biochip as described above. 50 μL of a cycle sequencing reagent (BigDye™ 3.1/BDX64, MCLab, San Francisco) was added to port 106 (comprised of through-holes 215 and 308) and chamber 309. After installation of two pneumatic interfaces (one for the input and one for the output end of the chip), the PCR product was processed as described in Example 1 through to the PCR chamber, but without the PCR thermal cycling step. The disposition of the fluids in the chip was as shown in
The following pressure profile was carried out using the pneumatic system software; all solenoid valves corresponding to chip ports were closed unless otherwise noted:
1. A pressure of 0.1 psig was applied to port 106 with ports 109 open to atmosphere for 10 sec to pump cycle sequencing reagent into channel 310 (see,
2. A pressure of 0.7 psig was applied for 0.2 sec on ports 106 and 108 (comprised of through-holes 216 and 314) to drive cycle the sequencing reagent from channel 304 through through-holes 311, through the body of layer 2, and into in the cycle sequencing reagent metering chambers 218 on layer 1 (see,
3. A pressure of 0.1 psig was applied to port 106 with ports 109 open to atmosphere, driving the cycle sequencing reagent to the capillary valves 221, where it was retained (see,
4. A pressure of 0.1 psig was applied to port 108 with port 106 open to atmosphere for 1 sec to drive excess cycle sequencing reagent backwards into chamber 101, leaving channel 310 empty (see,
5. A pressure of 0.7 psig was applied to ports 104 and 107 for 0.1 sec with ports 109 open to atmosphere to drive PCR product past capillary valve 220 and into through hole 317, passing through the body of layer 2 and through-hole 404 in layer 3, and into the cycle sequencing chamber 503 of layer 4 (see,
6. A pressure of 0.1 psig was applied to ports 109 with ports 104 and 107 open to atmosphere for 5 sec to drive PCR sample back to the through-holes. Capillary action retained the liquids at the entrance of the through-hole, preventing a trapped air bubble from appearing between the PCR product and chamber 503 (see,
7. A pressure of 0.7 psig was applied to port 108 for 0.2 sec with ports 109 open to atmosphere to drive cycle sequencing reagent into chamber 503, while simultaneously applying 0.1 psig to ports 104 and 107, to contact the PCR product with the sequencing reagent (see,
8. A pressure of 0.1 psig was applied for 10 sec to ports 104, 107 and 108 with ports 109 open to atmosphere to drive PCR product and Sanger reagent into the chamber. The trailing meniscus of the PCR product and that of the sequencing reagent were pinned at the capillary valves 220 and 221 (see,
9. 5 vacuum pulses of 0.25 psig vacuum and duration 0.1 sec were applied to port 108 with ports 109 open to atmosphere to draw both liquids partially backwards into reagent metering chamber 218 (see,
Steps 9-10 were repeated an additional two times to effect mixing of the sequencing reagent and PCR product.
The biochip was then pressurized to 30 psig N2 and thermally cycled using the following temperature profile:
A 4-sample biochip for the performance of sequencing product purification was constructed of four layers, as discussed in Example 1, and is shown in
1. Four 10 μL, samples of cycle sequencing product generated in tube reactions using pUC18 template and KOD enzyme were added to ports 1104 in the first layer and driven through channel 1105 in the second layer to the chamber 1106 in the second layer. 200 μL, of deionized water was added to port 1120 (a through-hole in the first layer) to reservoir 1121 in the second layer. The biochip was then installed in two pneumatic interfaces.
The following pressure profile was carried out using the pneumatic system software. All solenoid valves corresponding to biochip ports were closed unless otherwise noted.
2. A pressure of 0.09 psig was applied to ports 1104 with ports 1119 open to atmosphere for 5 sec to drive the sequencing product to capillary valves 1108 in layer 1, where they were retained.
3. A pressure of 0.6 psig was applied to ports 1104 with ports 1119 open to atmosphere for 0.1 sec to burst the samples through capillary valves 1108 in layer 1 and deliver them through through-holes 1111 in layer 2 into UF input chambers 1112 in layer 2.
4. A pressure of 0.09 psig was applied for to ports 1104 with ports 1119 open to atmosphere for 10-30 sec (different times were used in different experiments) to complete delivery of sequencing product to chambers 1112. Sequencing product was retained by capillary valves 1113 in layer 2 (see
5. A pressure of 0.8 psig was applied to port 1124 with ports 1119 and 1104 open to atmosphere for 0.5 sec to drive sequencing product through valves 113 into filtration chambers 1115. This also cleared input capillary valves 1108 of retained liquid.
6. A pressure of 0.09 psig was applied to port 1124 with ports 1119 open to atmosphere for 10-30 sec to complete delivery of the sequencing product to chamber 1115. Sequencing product was retained at valve 1113 (see,
7. A pressure of 7.5 psig was slowly applied to all ports of the chip for ultrafiltration. During ultrafiltration, the sequencing product meniscus remains pinned at 1113 while the leading edge of the liquid “retracts” as liquid was driven through the filter 1116. 10 μL of sequencing product required ˜120 sec for filtration. The pressure was released after filtration (see,
8. A pressure of 0.09 psig was applied to port 1120 with port 1124 open to atmosphere for 3 sec to drive water into channel 1122 (in layer 4) and partially-fill overflow chamber 1123 (see,
9. A pressure of 0.8 psig was applied to ports 1120 and 1124 with ports 1119 open to atmosphere to drive water through through-hole capillary valves 1110 in channel 1122 into chambers 1112.
10. A pressure of 0.09 psig was applied to port 1120 with ports 1119 open to atmosphere for 10-30 sec to complete delivery of liquid to chambers 1112, where it was retained by valves 1113. (see,
11. A pressure of 0.09 psig was applied to port 1124 with port 1120 open to drive water in chamber 1123 and channel 1122 back into chamber 1121 (see,
12. A pressure of 0.8 psig was applied to port 1124 with ports 1119 and 1104 open to atmosphere for 0.5 sec to drive water through valves 113 into filtration chambers 1115. This also cleared input capillary valves 1108 of retained liquid.
13. A pressure of 0.09 psig was applied to port 1124 with ports 1119 open to atmosphere for 10-30 sec to complete delivery of water to chamber 1115. Sequencing product was retained at valve 1113 (see,
The water was driven through the UF filter as in step 6 above, completing the first wash Steps 8-13 were repeated one additional time.
Steps 8-12 were repeated to partially-fill chambers 1115 with a final volume of water used for elution (see,
Vacuum of 1.6 psig was applied to ports 1104 with all other ports closed for 1 sec, drawing some water from chambers 1115 into chambers 1112 (the maximum motion being dictated by the creation of a vacuum of equal magnitude in the dead-space between the meniscus of the liquid and the solenoid valves corresponding to ports 1119), (see,
Ports 1104 were opened to atmosphere for 1 sec, allowing the liquid to move back into chamber 1115 due to the partial vacuum generated between the liquid and the valves corresponding to ports 1119 (see,
16-17 was repeated 50× to create 50 elution cycles.
A pressure of 0.09 psig was applied to port 1124 for 10 sec with ports 1119 open to atmosphere to drive liquids such that its trailing meniscus was pinned at 1113.
A pressure of 0.7 psig/0.05 sec was applied to port 1124 with ports 1119 open to atmosphere to detach the eluent (see,
The samples were retrieved and run directly on Genebench™ as described, yielding up to 479 QV20 bases.
Injection is performed electrokinetically with a pre-concentration step using counterelectrodes. The input well 1305, illustrated in
Separation occurs in separation channels 1306, and detection occurs via laser-induced fluorescence in the detection region 1307. In this biochip, a recess 1308 is provided to allow, for example, a Peltier block (not shown) to mate with the lower surface of 1301 to provide thermal cycling for PCR and cycle sequencing. Pneumatic interfaces (not shown) within the instrument clamp to the ends of the chip to provide microfluidic control.
1. Separation Instrument
DNA separation is carried out on a biochip and instrumentation as described in U.S. Patent Application Publication No. US2006-0260941-A1. Separation chips can be glass (see, U.S. Patent Application Publication No. US2006-0260941-A1) or plastic (the U.S. patent application entitled, “PLASTIC MICROFLUIDIC SEPARATION AND DETECTION PLATFORMS”, U.S. Ser. No. 12/080,745, filed on Apr. 4, 2008), each of which are hereby incorporated by reference in their entirety.
2. Excitation and Detection Instrumentation
The instrument comprises excitation and detection subsystems for interacting with and interrogating a sample. Samples typically include one or more biological molecules (including but not limited to DNA, RNA, and proteins) that are labeled with dyes (e.g., fluorescent dyes). The excitation subsystem comprises an excitation source or sources and an excitation beam path with optical elements including lenses, pinholes, mirrors and objectives, to condition and focus the excitation source in an excitation/detection window. Optical excitation of a sample can be accomplished by a series of laser types, with emission wavelengths in the visible region, between 400 to 650 nm. Solid state lasers can provide emission wavelengths of approximately 460 nm, 488 nm, and 532 nm. These lasers include, for example, the Compass, Sapphire and Verdi products from Coherent (Santa Clara, Calif.). Gas lasers include argon-ion and helium neon with emission in the visible wavelengths at approximately 488 nm, 514 nm, 543 nm, 595 nm, and 632 nm. Other lasers with emission wavelengths in the visible region are available from CrystaLaser (Reno, Nev.). In one embodiment, a 488 nm solid state laser Sapphire 488-200 (Coherent, Santa Clara, Calif.) can be utilized. In another embodiment, a light source with wavelength beyond the visible range can be used for exciting dyes having absorption and/or emission spectra beyond the visible range (e.g., infrared or ultra-violet emitting dyes). Alternatively optical excitation can be achieved by the use of non-laser light sources with emission wavelengths appropriate for dye excitation, including light emitting diodes, and lamps.
The detection subsystem comprises one or more optical detectors, a wavelength dispersion device (which performs wavelength separation), and one or a series of optical elements including, but not limited to, lenses, pinholes, mirrors and objectives to collect emitted fluorescence from fluorophore-labeled DNA fragments that are present at the excitation/detection window. The fluorescence emitted can be from a single dye or a combination of dyes. In order to discriminate the signal to determine its contribution from the emitting dye, wavelength separation of the fluorescence can be utilized. This can be achieved by the use of dichroic mirrors and bandpass filter elements (available from numerous vendors including Chroma, Rockingham, Vt.; and Omega Optical, Brattleboro, Vt.). In this configuration, the emitted fluorescence passes through a series of dichroic mirrors where one portion of the wavelength will be reflected by the mirror to continue traveling down the path, and the other portion will pass through. A series of discrete photodetectors, each one positioned at the end of the dichroic mirror will detect light over a specific range of wavelengths. A bandpass filter can be positioned between the dichroic mirror and photodetector to further narrow the wavelength range prior to detection. Optical detectors that can be utilized to detect the wavelength-separated signals include photodiodes, avalanche photodiodes, photomultiplier modules, and CCD cameras. These optical detectors are available from suppliers such as Hamamatsu (Bridgewater, N.J.).
In one embodiment, wavelength components are separated by the use of dichroic mirrors and bandpass filters and these wavelength components are detected with Photomultiplier Tube (PMT) detectors (H7732-10, Hamamatsu). The dichroic mirror and bandpass components can be selected such that incident light on each of the PMTs consists of a narrow wavelength band corresponding to the emission wavelength of the fluorescent dye. The band pass is typically selected to be centered about the fluorescent emission peak with a band pass of wavelength range of between 1 and 50 nm. The system is capable of eight color detection and can be designed with eight PMTs and a corresponding set of dichroic mirrors and bandpass filters to divide the emitted fluorescence into eight distinct colors. More than eight dyes can be detected by applying additional dichroic mirrors, bandpass filters and PMT.
Another method of discriminating the dyes that make up the fluorescence signal involves the use of wavelength dispersive elements and systems such as prisms, diffraction gratings, transmission gratings (available from numerous vendors including ThorLabs, Newton, N.J.; and Newport, Irvine, Calif.; and spectrographs (available from numerous vendors including Horiba Jobin-Yvon, Edison, N.J.). In this mode of operation, the wavelength components of the fluorescence are dispersed over a physical space. Detector elements placed along this physical space detect light and allow the correlation of the physical location of the detector element with the wavelength. Detectors suitable for this function are array-based and include multi-element photodiodes, CCD cameras, back-side thinned CCD cameras, multi-anode PMT. One skilled in the art will be able to apply a combination of wavelength dispersion elements and optical detector elements to yield a system that is capable of discriminating wavelengths from the dyes used in the system.
In another embodiment, a spectrograph is used in place of the dichroic and bandpass filters to separate the wavelength components from the excited fluorescence. Details on spectrograph design is available in John James, Spectrograph Design Fundamental, Cambridge, UK: Cambridge University Press, 2007. The spectrograph P/N MF-34 with a concave holographic grating with a spectral range of 505-670 nm (P/N 532.00.570) (HORIBA Jobin Yvon Inc, Edison, N.J.) is used in this application. Detection can be accomplished with a linear 32-element PMT detector array (H7260-20, Hamamatsu, Bridgewater, N.J.). Collected fluorescence is imaged on the pinhole, reflected, dispersed, and imaged by the concave holographic grating onto the linear PMT detector that is mounted at the output port of the spectrograph. The use of a PMT-based detector takes advantage of the low dark noise, high sensitivity, high dynamic range, and rapid response characteristic of PMT detectors. The use of a spectrograph and multi-element PMT detector for detection of excited fluorescence allows for flexibility in the number of dyes and the emission wavelength of dyes that can be applied within the systems and within the lane, without the need for physically reconfiguring the detection system (dichroic, bandpass and detectors) of the instrument. The data collected from this configuration is a wavelength dependent spectra across the visible wavelength range for each scan for each lane. Generating a full spectrum per scan provides dye flexibility both in terms of dye emission wavelength and number of dyes that can be present within a sample. In addition, the use of the spectrometer and linear multi-element PMT detector also allows for extremely fast read-out rates as all the PMT elements in the array are read-out in parallel.
Instruments may employ a staring mode of operation, to detect multiple lanes simultaneously and multiple wavelengths simultaneously. In one configuration, the excitation beam is simultaneously impinged on all lanes at the same time. The fluorescence from this is collected by a two dimensional detector such as a CCD camera or array. In this staring mode of collection, a wavelength dispersive element is used. One dimension of the detector represents the physical wavelength separation, while the other dimension represents the spatial or lane-lane separation.
For simultaneous excitation and detection of multiple samples, a scanning mirror system (62) (P/N 6240HA, 67124-H-0 and 6M2420X40S100S1, Cambridge technology, Cambridge Mass.) is utilized to steer both the excitation and detection beam paths in order to image each of the lanes of the biochip. In this mode of operation, the scanning mirror steers the beam paths, scanning sequentially from lane to lane from the first lane to the last lane, and the repeating the process again from the first lane to the last lane again. A lane-finding algorithm such as that described in U.S. Patent Application Publication No. US2006-0260941-A1 is used to identify location of lane.
An embodiment of an optical detection system for simultaneous multiple lane and multi dye detection is shown in
In one embodiment, the fluorescence excitation and detection assembly 40 includes a laser 60, a scanner 62, one or more light detectors 64, and various mirrors 68, spectrograph, and lenses 72 for transmitting a laser beam emitted from the laser 60 through opening 42 to the test module 55 and back to the light detectors 64. The scanner 62 moves the incoming laser beam to various scanning positions relative to the test module 55. Specifically, the scanner 62 moves the laser beam to a pertinent portion of each micro channel within the test module 55 to detect respective separate components. The multi element PMT 64 collects data (e.g. the fluorescent signals from DNA fragments of varying length) from the test module 55 and provides the data electronically through a cable attached to a port to a data acquisition and storage system located outside the protective cover. In one embodiment, the data acquisition and storage system can include a ruggedized computer available from Option Industrial Computers (l3 audreuil-Dorion, Quebec, Canada).
In another embodiment (a “staring mode”), the excitation source is incident on all the detection spots simultaneously, and fluorescence from all detection spots is collected simultaneously. Simultaneous spectral dispersion (wavelength spectra of detected fluorescence) and spatial dispersion (detection spots) can be performed with a two dimensional detector array. In this configuration, the 2-dimensional detector array is positioned in the system such that spectral components are imaged and detected across one dimension of the array detector (row), while spatial components are imaged and detected across the other dimension of the array detector.
A preferred instrument utilizes a scanning mode of operation, rather than a “staring” mode. In scanning mode, signal for each channel is required to be collected, integrated, and read-out while the scanner is coincident with the lane being interrogated and before it is incident on the next channel. A detector with fast readout allows for optimal light collection and integration, translating into higher signal to noise performance. Ideally, the read-out time of the detector should be significantly less than the total time which the scanner is coincident with the channel. The multi-element PMT can be read-out in less than 0.7 ms and this read-out time is far less than the integration time for detection for each individual channel.
Fluorescence incident on the pinhole can be dispersed by the grating according to its wavelength composition and focused onto the linear multi-anode PMT detector array. The detector provides 32 current outputs, one for each of the elements in the array that correspond to the number of photons incident on the element. During multiple samples (or lanes) detection, when the laser is in position exciting the selected lane, the integrator circuitry will integrate the PMT output current to generate an output voltage proportional to the integrated PMT current. At the same time, the single ended output voltage is converted to differential mode using the Analog Devices (Norwood, Mass.) differential driver IC (P/N SSM2142). At the end of the integration time (defined by scan rate and number of lane), the data acquisition system will read the differential signal and save the data in its buffer. After the data have been saved, the data acquisition system will move the scanner to shift the laser beam to the next selected lane, at the same time resetting the integrator circuitry.
Each single element PMT module has its own integrator circuitry. For an 8 color detection system, there are 8 PMT modules and 8 integrator circuitries. Additional colors can be added using corresponding numbers of PMT modules and integrator circuitry.
Since each of the PMT elements (H77260-20, Hamamatsu, Japan) has a similar or more rapid signal response as a single PMT tube (H7732-10, Hamamatsu, Japan), and the readout is in parallel, this detector is able to operate very rapidly. When coupled with the spectrometer, this spectrometer and multi-anode detector system is able to provide full spectral scans across the visible spectrum (450 nm to 650 nm) with readout-times of less than 0.1 ms.
The ability to provide fast refresh rates allows this spectrometer/detector system to be applied to scanning mode implementation of detection of multiple lanes sequentially within a single run. The use of PMT based detectors provides for low noise, high sensitivity and high dynamic range, and fast response. The 140 mm spectrometer with a concave holographic grating (Horiba Jobin-Yvon) and multianode PMT detector is the H7260-20 detector (Hamamatsu, Japan). Other spectrometers configurations and multi-anode PMT detectors can also be used for this application.
Determination of nucleotide bases from the electrophoregrams was achieved using signal processing algorithms to correct, filter, and analyze the data. This process consisted of locating a callable signal, correcting the signal baseline, filtering out noise, removing color cross-talk, identifying signal peaks, and determining associated bases. Locating the callable signal was performed to remove extraneous data from the beginning and end of the signal and accomplished by employing a threshold. Next, the background was removed from the signal, so that the signal had a common baseline for all detected colors. Finally, a lowpass filter was applied to remove high frequency noise from the signal.
To disambiguate the detected colors, a weighted matrix was calculated and applied to the signal to amplify the color-space of the nucleotide-dye spectrum. Calculation of this color separation matrix was accomplished using the methods of Li et al. Electrophoresis 1999, 20, 1433-1442. In this adaptation, a “m×n” color separation matrix is calculated from correlating the “m” number of dyes utilized in the assay with the “n” number of detector elements. The conversion of the signal from the detector space (PMT elements), to the dye space is performed by matrix manipulation as follows: D=CSM×PMT, where D is the signal in dye space for each of the m dyes, CSM is the color separation matrix, and PMT is a matrix with the signal from each of the n elements of the detector.
Next, the peaks in the color separated signal were identified using a combination of zero-crossing filters and frequency analysis. Finally, for fragment sizing applications, the corrected traces were allele-called to identify each fragment and to assign a fragment size based on a sizing standard. For DNA sequencing applications, the corrected traces were base-called to associate one of the four nucleotides with each peak in the trace. A detailed description of base calling can be found in Ewing et al. Genome Research, 1998, 8, 175-185, and Ewing et al., Genome Research, 1998, 8, 186-194, the disclosures of that are hereby incorporated by reference in their entirety.
3. Dye Labels
Dye labels attached to oligonucleotides and modified oligonucleotides can be synthesized or obtained commercially (e.g. Operon Biotechnologies, Huntsville, Ala.). A large number of dyes (greater than 50) are available for application in fluorescence excitation applications. These dyes include those from the fluorescein, rhodamine AlexaFluor, Biodipy, Coumarin, and Cyanine dye families. Furthermore, quenchers are also available for labeling oligo sequences to minimize background fluorescence. Dyes with emission maxima from 410 nm (Cascade Blue) to 775 nm (Alexa Fluor 750) are available and can be used. Dyes ranging between 500 nm to 700 nm have the advantage of being in the visible spectrum and can be detected using conventional photomultiplier tubes. The broad range of available dyes allows selection of dye sets that have emission wavelengths that are spread across the detection range. Detection systems capable of distinguishing many dyes have been reported for flow cytometry applications (see, Perfetto et al., Nat. Rev. Immunol. 2004, 4, 648-55; and Robinson et al., Proc of SPIE 2005, 5692, 359-365).
Fluorescent dyes have peak excitation wavelengths that are typically 20 to 50 nm blue-shifted from their peak emission wavelength. As a result, use of dyes over a wide range of emission wavelengths may require the use of multiple excitation sources, with excitation wavelengths to achieve efficient excitation of the dyes over the emission wavelength range. Alternatively, energy transfer dyes can be utilized to enable a single laser, with a single emission wavelength, to be used for exciting all dyes of interest. This is achieved by attaching an energy transfer moiety to the dye label. This moiety is typically another fluorescent dye with an absorption wavelength that is compatible with the excitation wavelength of the light source (e.g. laser). Placement of this absorber in close proximity with an emitter allows the absorbed energy to be transferred from the absorber to the emitter, allowing for more efficient excitation of the long wavelength dyes (Ju et al., Proc Natl Acad Sci USA 1995, 92, 4347-51).
Dye labeled dideoxynucleuotides are available from Perkin Elmer, (Waltham, Mass.).
The following example illustrates the separation and detection of nucleic acid fragments labeled with 6 fluorescent dyes, and demonstrates the color resolution capability of the spectrometer/multi element excitation/detection system. DNA fragments were labeled with 6-FAM, VIC, NED, PET dyes by applying fluorescently labeled primers in a multiplexed PCR amplification reaction. In this reaction, 1 ng of human genomic DNA (9947A) was amplified in a 25 μL reaction in according to the manufacturers recommended conditions (AmpFlSTR Identifiler, Applied Biosystems). 2.7 μL of the PCR product was removed and mixed with 0.3 μL of GS500-LIZ sizing standard (Applied Biosystems) and 0.3 μL of HD400-ROX sizing standard. HiDi (Applied Biosystems) was added to a total of 13 μL and the sample was inserted into the sample well of the separation biochip and subjected to electrophoresis.
Electrophoretic separation of DNA using Genebench consists of a series of four operations: pre-electrophoresis, loading, injection and separation. These operations are carried out on a microfluidic biochip, which is heated to a uniform temperature of 50° C. The biochip contains 16 channel systems for separation and detection multiple, each consisting of an injector channel and a separation channel. DNA for analysis is separated by electrophoretic transport of the DNA through a sieving matrix along the separation channel. The separation length of the biochips is ranges from 160 to 180 mm.
The first step is pre-electrophoresis, which is accomplished by applying a 160 V/cm field along the length of the channel for six (6) minutes. Separation buffer (TTE1X) is pipetted into the anode, cathode and waste wells. Samples for analysis are pipetted into the sample wells and a 175 V applied from the sample well to the waste well for 18 seconds, followed by the application of 175 V across the sample and waste well and 390 V at the cathode for 72 seconds, to load the sample into the separation channel. Injection of the sample is accomplished by applying a 160 V/cm field along the length of the separation channel while fields of 50 V/cm and 40 V/cm are applied across the sample and waste wells respectively. Separation is continued with the injection voltage parameters for 30 min during that an optical system detects the separating bands of DNA. The data collection rate is 5 Hz and PMT gains are set to −800 V.
Sixteen samples containing amplified DNA were loaded for simultaneous separation and detection. The signals from each of the 32-elements of the PMT were collected as a function of time to generate an electrophoregram. The resulting electrophoregram (
In this example, an 8 dye separation and detection of acids labeled with fluorescent dyes is shown. Forward and reverse primer pair sequences for 8 loci are selected from the published sequences (Butler et al., J Forensic Sci 2003, 48, 1054-64).
The loci selected are CSF1P0, FGA, THO1, TPDX, vWA, D3S1358, D5S818 and D7S820, although any of the loci and hence primer pairs described in the paper can also be used in this example. Each of the forward primers for the primer pairs is labeled with a separate fluorescent dye (Operon Biotechnologies, Huntsville, Ala.). Dyes selected for attachement to the primers include Alexa Fluor Dyes 488, 430, 555, 568, 594, 633, 647, and Tamra. Numerous other dyes are available and can also be used as labels. Each locus is amplified separately following the PCR reaction protocols of (Butler, 2003, Id.) to yield a reaction solution with fragments labeled with their respective dyes. Template for PCR reaction is 1 ng of human gemonic DNA (type 9947A from Promega, Madison Wis.).
Each PCR reaction was purified by cleaning up through a PCR cleanup column, where primers (labeled and dye-labeled primers) and enzymes are removed, and the PCR buffer is exchanged by the DI eluant. The resulting product of clean is a solution of labeled DNA fragments in DI water. Cleanup of dye labeled products follows the protocol of Smith using MinElute™ columns (Qiagen, Valencia, Calif.). A total of eight reactions are performed. Eight cleaned up PCR reactions were mixed together in a ratio to generate peaks of equivalent signal strengths, generating a mixture containing fragments labeled with 8 different dyes. Alternatively, primers for 8 loci can be mixed together to form a master primer mix for multiplexed amplification.
This solution is separated and detected with the instrument and protocol as described in Example 1. The grating of the spectrograph is adjusted such that the emission of the 8 dyes falls across the 32 pixels of the detector elements. The amount of sample loaded for analysis is to be adjusted such that detected signals fall within the dynamic range of the detection system.
The following example illustrates separation/detection of labeled DNA fragments with the spectrometer/multi-element PMT system of
In this example, separation and detection of fragments from cycle sequencing of two DNA templates are carried out simultaneously in a single separation channel. The cycle sequencing reactions can be prepared by either dye labeled terminator reactions or dye labeled primer reactions as follows:
For Dye Labeled Terminator Reactions:
Cycle sequencing reaction for each template fragment consisting of a sequencing primer appropriate for the template sequence of interest, and reagents for conducting DNA sequencing including cycle sequencing buffer, polymerase, oligonucleotides, dideoxynucleotides and labeled dideoxynucleotides is prepared. Eight different dyes are utilized for the labeling. In the first cycle sequencing reaction, one set of 4 dye labeled dideoxy nucleotides is used. In the second cycle sequencing reaction, another set of 4 dye labeled dideoxy nucleotides (with emission wavelengths different than those of the four used in the first cycle sequencing reaction) is used. Each cycle sequencing reaction is carried out separately following a protocol that thermally cycles each reaction multiple times. Each thermal cycle includes a denature, anneal and extension step with temperatures and times following the protocols of Sanger (see, Sanger et al., Proc Natl Acad Sci USA 1977, 74, 5463-7). The cycle sequencing product from the two reactions are combined to form a sample that consist of labeled DNA fragments, with a total of eight unique dyes, from each of the two DNA templates.
For Dye Labeled Primer Reactions:
Alternatively, the sample for separation and detection can be fabricated by using primer labeled cycle sequencing. Four cycle sequencing reactions are carried out for each DNA template. Each reaction is a cycle sequencing reaction consisting of a labeled sequencing primer, and reagents for conducting DNA sequencing including cycle sequencing buffer, polymerase, oligonucleuotides. In addition, each reaction will include one of the dideoxynucleuotides (ddATP, ddTTP, ddCTP, or ddGTP) and one labeled primer. Each dye associated with the primer is unique in emission wavelength and is correlated with the type of dideoxy nucleotide in the cycle sequencing solution (ddATP, ddTTP, ddCTP, or ddGTP). Each cycle sequencing reaction is carried out separately following a protocol that thermally cycles each reaction multiple times. Each thermal cycle includes a denature, anneal and extension step with temperatures and times following the protocols of Sanger (see, Sanger, 1977, Id.). For cycle sequencing the second DNA template, another set of 4 dyes (with emission wavelength different to that of the four used in the first cycle sequencing reaction) is applied. The product of all eight reactions (each with a different dye) are mixed together to form a sample that consist of DNA fragments from each of the two DNA templates.
Sample for Separation and Detection:
Each of the sequencing reactions is cleaned up by ethanol precipitation. Separation and detection of the sample follows the protocol of Example 8. The result of the separation and detection is the generation of two distinct DNA sequences, corresponding to each of the two template DNA fragments.
The methods of this example can be modified to allow the use of dyes in multiples of four to allow detection of that multiple of DNA sequences in a single separation channel (e.g. 12 dyes for the detection of 3 sequences simultaneously, 16 dyes for the detection of 4 sequences simultaneously, 20 dyes for the detection of five sequences simultaneously, and so on). Finally, separation of the labeled fragments need not be limited to electrophoresis.
There are several applications of nucleic acid analysis that can be applied to clinical diagnostics, including DNA and RNA sequencing and fragment size determination. In this example, the use of simultaneous detection of 10 colors allows the interrogation of up to 500 loci. Analysis of the size of large number of fragments can be utilized to identify pathogens or to characterize many loci within an individual's genome, for example. In the setting of prenatal and pre-implantation genetic diagnosis, aneuploidy is currently diagnosed by karyotyping and by fluorescent in situ hybridization (FISH). In FISH studies, the presence of two signals per cell indicates that two copies of a given locus are present within that cell, one signal indicates monosomy or partial monosomy, and three signals indicates trisomy or partial trisomy. FISH typically utilizes approximately 10 probes to assess whether or not a cell contains a normal chromosomal complement. This approach does not allow a detailed view of the entire genome, however, and cells that appear normal by FISH may well have major abnormalities that are not detected by the technique.
The teachings of the present invention make use of multicolor separation and detection to allow approximately 500 chromosomal loci widely dispersed about across all chromosomes to be assessed to allow a much more detailed analysis of chromosomal structure. In this example, primer pair sequences for approximately 500 loci are selected from published sequences, with each locus present as a single copy per haploid genome. In addition, 10 sets of 50 primer pairs are selected such that each set defines a corresponding set of DNA fragments such that none of the fragments are of the identical size. For each set, the forward primers for the primer pairs are labeled with one fluorescent dye, and no two sets share the same dye. Dyes selected for attachment to the primers are Alexa Fluor Dyes 488, 430, 555, 568, 594, 633, 647, 680, 700, and Tamra. Numerous other dyes are available and can also be used as labels. The loci can be amplified in one or several parallel PCR reactions as described in “M
The number of loci, dyes, and separation channels can be varied based on the desired application. Smaller numbers of fragments can be detected if desired by utilizing a smaller number of dye labels or generating fewer DNA fragments per label; in this way, less than 500, less than 400, less than 300, less than 200, less than 100, less than 75, less than 50, less than 40, less than 30, or less than 20 fragments can be detected as desired. The maximum number of loci that can be identified per lane is based on the read length and resolution of the separation system (e.g., single base pair resolution of DNA fragments ranging from 20 to 1500 base pairs results in hundreds of fragments) multiplied by the number of distinct dyes that can be detected (as noted supra, dozens are available). Accordingly, thousands of loci can be identified in a single separation channel, and the number will increase as additional dyes are developed.
This application is a continuation of U.S. Ser. No. 12/080,751, filed on 4 Apr. 2008 and entitled “Integrated Nucleic Acid Analysis”, which claims the benefit of the filing date, under 35 U.S.C. §119(e), of U.S. Provisional Application Ser. No. 60/921,802 filed 4 Apr. 2007; U.S. Provisional Application Ser. No. 60/964,502 filed 13 Aug. 2007; and U.S. Provisional Application Ser. No. 61/028,073 filed 12 Feb. 2008, each of that is hereby incorporated by reference in its entirety. This application also incorporates by reference, in their entireties, two U.S. Patent applications filed on Apr. 4, 2008; the first entitled “METHODS FOR RAPID MULTIPLEXED AMPLIFICATION OF TARGET NUCLEIC ACIDS”, U.S. Ser. No. 12/080,746; and the second entitled “PLASTIC MICROFLUIDIC SEPARATION AND DETECTION PLATFORMS”, U.S. Ser. No. 12/080,745.
Number | Date | Country | |
---|---|---|---|
61028073 | Feb 2008 | US | |
60964502 | Aug 2007 | US | |
60921802 | Apr 2007 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 12080751 | Apr 2008 | US |
Child | 13191952 | US |