The invention generally relates to systems and methods for mass spectrometry analysis of samples.
Biofluids (e.g., complex mixtures such as blood, saliva, or urine) are routinely separated using chromatography before the MS measurement in order to minimize suppression effects on analyte ionization and to pre-concentrate the analytes. Recently, systems and methods have been developed that allow for sample preparation and pre-treatment to be combined with the ionization process (See Ouyang et al., WO 2010/127059).
These systems and methods use wetted porous material, named paper spray ionization, for direct, qualitative and quantitative analysis of complex biofluids (“paper spray”). Analyte transport is achieved by wicking in a porous material with a macroscopically sharp point and a high electric field is used to perform ionization and chemical analysis of compounds present in biological samples. Pneumatic assistance is not required to transport the analyte: a voltage is simply applied to the wet paper, held in front of a mass spectrometer.
The inventions herein generally relate to various modifications to the wetted porous material used in the paper spray process described above. Such modified substrates lend themselves to more efficient analysis of certain samples as well as improved analytical results for those samples.
In certain aspects, the invention generally relates to systems and methods for ion generation using a wetted porous substrate that substantially prevents diffusion of sample into the substrate. Paper spray has been applied for analysis of biofluids, such as dried blood spots. Generally, the whole blood is deposited onto hydrophilic chromatography paper and dried to form a dried blood spot. The liquid blood is hydrophilic, and when it is dropped onto a hydrophilic porous sample substrate, such as chromatography paper, the fresh blood sample diffuses into the substrate. When the whole blood is dried on the paper substrate, the microstructure of the paper substrate is modified. During the paper spray, the efficiency of the extraction of chemical from the dried blood spot is limited.
In these embodiments, systems and methods of the invention limit or prevent interaction of the sample with a porous substrate, thus allowing efficient transfer of the sample through the substrate for analysis. In these embodiments, the invention provides a mass spectrometry probe including at least one porous substrate connected to a high voltage source, in which the porous substrate includes a material that substantially prevents diffusion of a sample into the substrate. In certain embodiments, the porous substrate is discrete from a flow of solvent. The substrate may have any shape, although shapes having sharp points are preferred. Exemplary shapes include triangles or cones. The material for the porous substrate will depend on the properties of the sample to be analyzed. For example, when the sample is hydrophilic, the substrate is less hydrophilic than the sample. Alternatively, when the sample is hydrophobic, the substrate is less hydrophobic than the sample. An exemplary substrate for analyzing a biological fluid is silanized paper.
In other embodiments, the probe further includes a discrete amount of a solvent, e.g., a droplet or droplets, applied to the porous substrate. The solvent is capable of diffusing into the substrate. The solvent is applied as a droplet or droplets, and in an amount sufficient to wet the porous substrate. Once applied to the porous substrate, the solvent can assist transport of the sample through the porous material. The solvent can contain an internal standard. The solvent/substrate combination can allow for differential retention of sample components with different chemical properties. In certain embodiments, the solvent minimizes salt and matrix effects. In other embodiments, the solvent includes chemical reagents that allow for on-line chemical derivatization of selected analytes.
In another embodiment, the invention provides a system for analyzing a sample material including at least one porous substrate connected to a high voltage source, in which the porous substrate includes a material that substantially prevents diffusion of a sample into the substrate, and a mass analyzer. In certain embodiments, the porous substrate is discrete from a flow of solvent. The mass analyzer is for a mass spectrometer or a handheld mass spectrometer. Exemplary mass analyzers include a quadrupole ion trap, a rectalinear ion trap, a cylindrical ion trap, a ion cyclotron resonance trap, an orbitrap, a time of flight, a Fourier Transform ion cyclotron resonance, and sectors.
Another embodiment of the invention provides a method for analyzing a sample that involves contacting a sample to a porous substrate, in which the porous substrate includes a material that substantially prevents diffusion of the sample into the substrate, applying a solvent that is capable of diffusing into the substrate to the sample, resulting in diffusion of components of the sample into the substrate, applying a high voltage to the substrate to generate ions of the components that are expelled from the porous substrate, and analyzing the expelled ions. In certain embodiments, the porous substrate is kept separate from a flow of solvent. Exemplary samples include chemical species or biological species. In certain embodiments, the sample is a biological fluid, e.g., complex mixtures such as blood, saliva, or urine.
In certain embodiments, the solvent is capable of mixing with the sample. In other embodiments, the solvent is not capable of mixing with the sample but is capable of extracting the components from the sample. In certain embodiments, prior to applying the high voltage, the method further involves drying the substrate. In certain embodiments, analyzing involves providing a mass analyzer to generate a mass spectrum of analytes in the sample.
Another embodiment of the invention provides a method for analyzing blood that involves contacting a blood sample to a porous substrate, in which the porous substrate includes a material that substantially prevents diffusion of the blood sample into the substrate, applying to the blood sample a solvent that is capable of diffusing into the substrate, resulting in diffusion of components of the blood sample into the substrate, applying a high voltage to the substrate to generate ions of the components that are expelled from the porous substrate, and analyzing the expelled ions. In certain embodiments, the porous substrate is kept separate from a flow of solvent. In certain embodiments, the solvent is capable of mixing with the blood sample. In other embodiments, the solvent is not capable of mixing with the blood sample but is capable of extracting the components (e.g., proteins) from the blood sample. In certain embodiments, prior to applying the high voltage, the method further involves drying the substrate to produce a dried blood spot.
Another embodiment of the invention provides a method for analyzing a protein from a blood sample that involves contacting a blood sample comprising at least one protein to a porous substrate, in which the porous substrate includes a material that substantially prevents diffusion of the blood sample into the substrate, applying to the blood sample a solvent that is capable of diffusing into the substrate, resulting in diffusion of the protein into the substrate, applying a high voltage to the substrate to generate ions of the protein that are expelled from the porous substrate, and analyzing the expelled ions. In certain embodiments, the porous substrate is kept separate from a flow of solvent.
Other aspects of the invention generally relate to ion generation using a wetted porous material and a drying agent. As previously mentioned, paper spray has been applied for analysis of biofluids, such as dried blood spots. One current method for analyzing a dried blood spot involves spotting 15 microliters of drug-spiked blood onto Whatman 31ETF card paper, and allowing it to air dry for at least 2 hours. This drying time is important, otherwise the blood can run through the paper during analysis, thus reducing the efficiency of the paper spray.
These aspects of the invention combine a drying agent with the porous substrate used in paper spray in order to rapidly dry a sample that is applied to the substrate. Rapid drying decreases waiting time and allows for sample analysis within approximately two to five minutes of applying a sample to the substrate. With respect to analysis of biological fluids, probes, systems, and methods of the invention provide a point-of-care device for rapid and convenient use. In a particular embodiment, probes, systems, and methods of the invention use anhydrous salts to dry and thus restrict flow of liquid blood during paper spray analysis.
In certain embodiments, the invention provides a mass spectrometry probe including at least one porous substrate connected to a high voltage source, in which the porous substrate includes a drying agent. Any drying agent that is compatible with the sample and does not interfere with analysis by mass spectrometry may be used. An exemplary drying agent is an anhydrous salt. In certain embodiments, the porous substrate also includes an internal standard.
In certain embodiments, the porous substrate is discrete (i.e., separate or disconnected from) from a flow of solvent. Instead, a sample is either spotted onto the porous substrate or the porous substrate is wetted and used to swab a surface containing the sample. The porous substrate with spotted or swabbed sample is then wetted and connected to a high voltage source to produce ions of the sample which are subsequently analyzed. The sample is transported through the porous substrate without the need of a separate solvent flow.
Probes, systems, and methods of the invention combine sample preparation and pre-treatment with the ionization process needed for mass analysis of samples. Probes, systems, and methods of the invention allow for rapid and direct analysis of chemicals in raw biological samples of complex matrices, such as biofluids and tissues, without sample preparation. In particular embodiments, probes, systems, and methods of the invention allow for the analysis of a dried spots of blood or urine.
Exemplary porous materials include paper, e.g., filter paper, or PVDF membrane. The porous material can be of any shape. In certain embodiments, the porous material is provided as a triangular piece or cone.
In certain embodiments, the probe further includes a discrete amount of a solvent, e.g., a droplet or droplets, applied to the porous material. The solvent is applied as a droplet or droplets, and in an amount sufficient to wet the porous material. Once applied to the porous material, the solvent can assist transport of the sample through the porous material. The solvent can contain an internal standard. The solvent/substrate combination can allow for differential retention of sample components with different chemical properties. In certain embodiments, the solvent minimizes salt and matrix effects. In other embodiments, the solvent includes chemical reagents that allow for on-line chemical derivatization of selected analytes.
Another embodiment of the invention provides a system for analyzing a sample material including at least one porous substrate connected to a high voltage source, in which the porous substrate includes a drying agent, and a mass analyzer. In certain embodiments, the porous substrate is discrete from a flow of solvent. The mass analyzer is for a mass spectrometer or a handheld mass spectrometer. Exemplary mass analyzers include a quadrupole ion trap, a rectalinear ion trap, a cylindrical ion trap, a ion cyclotron resonance trap, an orbitrap, a time of flight, a Fourier Transform ion cyclotron resonance, and sectors.
Another embodiment of the invention provides a method for analyzing a sample that involves contacting a sample to a porous substrate, in which the porous substrate includes a drying agent, applying a high voltage to the porous material to generate ions of an analyte in the sample that are expelled from the porous material, and analyzing the expelled ions. In certain embodiments, the porous substrate is kept separate from a flow of solvent. In certain embodiments, methods of the invention further involve, prior to applying the high voltage, drying the substrate. In certain embodiments, methods of the invention further involve applying a solvent to the substrate. The solvent is applied as a droplet or droplets, and in an amount sufficient to wet the porous material. Once applied to the porous material, the solvent can assist transport of the sample through the porous material. In certain embodiments, analyzing involves providing a mass analyzer to generate a mass spectrum of analytes in the sample. Exemplary samples include chemical species or biological species. In certain embodiments, the sample is a biological fluid, e.g., complex mixtures such as blood, saliva, or urine.
Another embodiment of the invention involves contacting a blood sample to a porous substrate, in which the porous substrate includes a drying agent, drying the blood sample on the substrate, applying a solvent to the substrate, applying a high voltage to the substrate to generate ions of the components that are expelled from the porous substrate, and analyzing the expelled ions. In certain embodiments, applying the solvent and the high voltage occur simultaneously. In certain embodiments, applying the solvent and the high voltage occur sequentially.
Other aspects of the invention generally relate to ion generation using a modified wetted porous substrate. These aspects of the invention recognize that the porous material used in paper spray plays an important role in determining the resolution of target compounds from complex samples and also in determining the transfer efficiency of analyte ions. Accordingly, embodiments of the invention provide a mass spectrometry probe that includes at least one porous substrate connected to a high voltage source, in which at least a portion of the porous substrate includes a material that modifies an interaction between a sample and the substrate. In certain embodiments, the porous substrate is discrete from a flow of solvent. Exemplary porous substrates include paper, e.g., filter paper, or PVDF membrane. The porous material can be of any shape. In certain embodiments, the porous material is provided as a triangular piece or cone.
In these embodiments, the material may be any material that modifies the interaction between the sample and the substrate. In certain embodiments, the material modifies the interaction between the sample and substrate during sample deposition. In other embodiments, the material modifies the interaction between the sample and substrate during sample elution. The material may coat at least a portion of the substrate. Alternatively, the material may impregnate at least a portion of the substrate. In certain embodiments, the material is silica.
In particular embodiments, the silica coats a surface of the substrate.
In certain embodiments, the probe further includes a discrete amount of a solvent, e.g., a droplet or droplets, applied to the porous material. The solvent is applied as a droplet or droplets, and in an amount sufficient to wet the porous material. Once applied to the porous material, the solvent can assist transport of the sample through the porous material. The solvent can contain an internal standard. The solvent/substrate combination can allow for differential retention of sample components with different chemical properties. In certain embodiments, the solvent minimizes salt and matrix effects. In other embodiments, the solvent includes chemical reagents that allow for on-line chemical derivatization of selected analytes.
Another embodiment of the invention provides a system for analyzing a sample material including at least one porous substrate connected to a high voltage source, in which at least a portion of the porous substrate includes a material that modifies an interaction between a sample and the substrate, and a mass analyzer. In certain embodiments, the porous substrate is discrete from a flow of solvent. The mass analyzer is for a mass spectrometer or a handheld mass spectrometer. Exemplary mass analyzers include a quadrupole ion trap, a rectalinear ion trap, a cylindrical ion trap, a ion cyclotron resonance trap, an orbitrap, a time of flight, a Fourier Transform ion cyclotron resonance, and sectors.
Another embodiment of the invention provides a method for analyzing a sample that involves contacting a sample to a porous substrate, in which at least a portion of the porous substrate includes a material that modifies an interaction between a sample and the substrate, applying a high voltage to the porous material to generate ions of an analyte in the sample that are expelled from the porous material, and analyzing the expelled ions. In certain embodiments, the porous substrate is kept separate from a flow of solvent. Exemplary samples include chemical species or biological species. In certain embodiments, the sample is a biological fluid, e.g., complex mixtures such as blood, saliva, or urine.
Another embodiment of the invention provides a method for analyzing blood that involves contacting a blood sample to a porous substrate, in which at least a portion of the porous substrate comprises a silica coating that modifies an interaction between the blood sample and the substrate, applying a high voltage to the porous substrate to generate ions of an analyte in the blood sample that are expelled from the porous substrate, and analyzing the expelled ions.
The invention generally relates to ion generation using modified wetted porous materials. In certain aspects, the invention generally relates to systems and methods for ion generation using a wetted porous substrate that substantially prevents diffusion of sample into the substrate. In other aspects, the invention generally relate to ion generation using a wetted porous material and a drying agent. In other aspects, the invention generally relates to ion generation using a modified wetted porous substrate in which at least a portion of the porous substrate includes a material that modifies an interaction between a sample and the substrate.
A method of generating ions from fluids and solids for mass spectrometry analysis is described. Porous materials, such as paper (e.g. filter paper or chromatographic paper) or other similar materials are used to hold and transfer liquids and solids, and ions are generated directly from the edges of the material when a high electric voltage is applied to the material (
In certain embodiments, the porous material is any cellulose-based material. In other embodiments, the porous material is a non-metallic porous material, such as cotton, linen wool, synthetic textiles, or plant tissue. In still other embodiments, the porous material is paper. Advantages of paper include: cost (paper is inexpensive); it is fully commercialized and its physical and chemical properties can be adjusted; it can filter particulates (cells and dusts) from liquid samples; it is easily shaped (e.g., easy to cut, tear, or fold); liquids flow in it under capillary action (e.g., without external pumping and/or a power supply); and it is disposable.
In particular embodiments, the porous material is filter paper. Exemplary filter papers include cellulose filter paper, ashless filter paper, nitrocellulose paper, glass microfiber filter paper, and polyethylene paper. Filter paper having any pore size may be used. Exemplary pore sizes include Grade 1 (11 μm), Grade 2 (8 μm), Grade 595 (4-7 μm), and Grade 6 (3 μm), Pore size will not only influence the transport of liquid inside the spray materials, but could also affect the formation of the Taylor cone at the tip. The optimum pore size will generate a stable Taylor cone and reduce liquid evaporation. The pore size of the filter paper is also an important parameter in filtration, i.e., the paper acts as an online pretreatment device. Commercially available ultrafiltration membranes of regenerated cellulose, with pore sizes in the low nm range, are designed to retain particles as small as 1000 Da. Ultra filtration membranes can be commercially obtained with molecular weight cutoffs ranging from 1000 Da to 100,000 Da.
In certain embodiments, the substrate includes a material that substantially prevents diffusion of a sample into the substrate. The material for the porous substrate will depend on the properties of the sample to be analyzed. For example, when the sample is hydrophilic, the substrate is less hydrophilic than the sample. Alternatively, when the sample is hydrophobic, the substrate is less hydrophobic than the sample. Thus the sample remains substantially on top of the substrate until an appropriate solvent is applied to the substrate. The solvent is capable of diffusing into the substrate. The solvent interacts with the sample, causing the sample or components of the sample to diffuse into the substrate. In certain embodiments, the solvent is capable of mixing with the sample and both the solvent and the sample diffuse into the substrate. In other embodiments, the solvent is not capable of mixing with the sample but is capable of extracting the components from the sample that diffuse into the substrate along with the solvent.
In other embodiments, the porous material includes a drying agent in order to rapidly dry a sample that is applied to the substrate. Any drying agent that is compatible with the sample and does not interfere with analysis by mass spectrometry may be used. An exemplary drying agent is an anhydrous salt, such as magnesium sulfate, sodium sulfate, sodium carbonate, or calcium chloride. Other exemplary drying agents include blood coagulants, such as alum powder. In certain embodiments, the porous substrate also includes an internal standard. Rapid drying decreases waiting time and allows for sample analysis within approximately two to five minutes of applying a sample to the substrate. With respect to analysis of biological fluids, such a substrate allows for a point-of-care device for rapid and convenient use.
In other embodiments, the substrate includes a material that modifies an interaction between the sample and the substrate. The material may be any material that modifies the interaction between the sample and the substrate. In certain embodiments, the material modifies the interaction between the sample and substrate during sample deposition. In other embodiments, the material modifies the interaction between the sample and substrate during sample elution. The material may coat at least a portion of the substrate. Alternatively, the material may impregnate at least a portion of the substrate. In certain embodiments, the material is silica. In particular embodiments, the silica coats a surface of the substrate.
The substrate is then connected to a high voltage source to produce ions of the sample which are subsequently mass analyzed. The sample is transported through the porous material without the need of a separate solvent flow. Pneumatic assistance is not required to transport the analyte; rather, a voltage is simply applied to the porous material that is held in front of a mass spectrometer.
In certain embodiments, the porous material is integrated with a solid tip having a macroscopic angle that is optimized for spray. In these embodiments, the porous material is used for filtration, pre-concentration, and wicking of the solvent containing the analytes for spray at the solid type.
Probes of the invention work well for the generation of micron scale droplets simply based on using the high electric field generated at an edge of the porous material. In particular embodiments, the porous material is shaped to have a macroscopically sharp point, such as a point of a triangle, for ion generation. Probes of the invention may have different tip widths. In certain embodiments, the probe tip width is at least about 5 μm or wider, at least about 10 m or wider, at least about 50 μm or wider, at least about 150 μm or wider, at least about 250 μm or wider, at least about 350 μm or wider, at least about 400 or wider, at least about 450 μm or wider, etc. In particular embodiments, the tip width is at least 350 μm or wider. In other embodiments, the probe tip width is about 400 μm. In other embodiments, probes of the invention have a three dimensional shape, such as a conical shape.
As mentioned above, no pneumatic assistance is required to transport the droplets. Ambient ionization of analytes is realized on the basis of these charged droplets, offering a simple and convenient approach for mass analysis of solution-phase samples.
Sample solution is directly applied on the porous material held in front of an inlet of a mass spectrometer without any pretreatment. Then the ambient ionization is performed by applying a high potential on the wetted porous material. In certain embodiments, the porous material is paper, which is a type of porous material that contains numerical pores and microchannels for liquid transport. The pores and microchannels also allow the paper to act as a filter device, which is beneficial for analyzing physically dirty or contaminated samples.
In other embodiments, the porous material is treated to produce microchannels in the porous material or to enhance the properties of the material for use as a probe of the invention. For example, paper may undergo a patterned silanization process to produce microchannels or structures on the paper. Such processes involve, for example, exposing the surface of the paper to tridecafluoro-1,1,2,2-tetrahydrooctyl-1-trichlorosilane to result in silanization of the paper. In other embodiments, a soft lithography process is used to produce microchannels in the porous material or to enhance the properties of the material for use as a probe of the invention. In other embodiments, hydrophobic trapping regions are created in the paper to pre-concentrate less hydrophilic compounds.
Hydrophobic regions may be patterned onto paper by using photolithography, printing methods or plasma treatment to define hydrophilic channels with lateral features of 200˜1000 μm. See Martinez et al. (Angew. Chem. Int. Ed. 2007, 46, 1318-1320); Martinez et al. (Proc. Natl Acad. Sci. USA 2008, 105, 19606-19611); Abe et al. (Anal. Chem. 2008, 80, 6928-6934); Bruzewicz et al. (Anal. Chem. 2008, 80, 3387-3392); Martinez et al. (Lab Chip 2008, 8, 2146-2150); and Li et al. (Anal. Chem. 2008, 80, 9131-9134), the content of each of which is incorporated by reference herein in its entirety. Liquid samples loaded onto such a paper-based device can travel along the hydrophilic channels driven by capillary action.
Another application of the modified surface is to separate or concentrate compounds according to their different affinities with the surface and with the solution. Some compounds are preferably absorbed on the surface while other chemicals in the matrix prefer to stay within the aqueous phase. Through washing, sample matrix can be removed while compounds of interest remain on the surface. The compounds of interest can be removed from the surface at a later point in time by other high-affinity solvents. Repeating the process helps desalt and also concentrate the original sample.
In certain embodiments, methods and systems of the invention use a porous material, e.g., paper, to hold and transport analytes for mass spectral analysis. Analytes in samples are pre-concentrated, enriched and purified in the porous material in an integrated fashion for generation of ions with application of a high voltage to the porous material. In certain embodiments, a discrete amount of transport solution (e.g., a droplet or a few droplets) is applied to assist movement of the analytes through the porous material. In certain embodiments, the analyte is already in a solution that is applied to the porous material. In such embodiments, no additional solvent need be added to the porous material. In other embodiments, the analyte is in a powdered sample that can be easily collected by swabbing a surface. Systems and methods of the invention allow for analysis of plant or animal tissues, or tissues in living organisms.
Methods and systems of the invention can be used for analysis of a wide variety of small molecules, including epinephrine, serine, atrazine, methadone, roxithromycin, cocaine and angiotensin I. All display high quality mass and MS/MS product ion spectra (see Examples below) from a variety of porous surfaces. Methods and systems of the invention allow for use of small volumes of solution, typically a few μL, with analyte concentrations on the order of 0.1 to 10 μg/mL (total amount analyte 50 pg to 5 ng) and give signals that last from one to several minutes.
Methods and systems of the invention can be used also for analysis of a wide variety of biomolecules, including proteins and peptides. Methods of the invention can also be used to analyze oligonucleotides from gels. After electrophoretic separation of oligonucleotides in the gel, the band or bands of interest are blotted with porous material using methods known in the art. The blotting results in transfer of at least some of the oligonucleotides in the band in the gel to the porous material. The porous material is then connected to a high voltage source and the oligonucleotides are ionized and sprayed into a mass spectrometer for mass spectral analysis.
Methods and systems of the invention can be used for analysis of complex mixtures, such as whole blood or urine. The typical procedure for the analysis of pharmaceuticals or other compounds in blood is a multistep process designed to remove as many interferences as possible prior to analysis. First, the blood cells are separated from the liquid portion of blood via centrifugation at approximately 1000×g for 15 minutes (Mustard, J. F.; Kinlough-Rathbone, R. L.; Packham, M. A. Methods in Enzymology; Academic Press, 1989). Next, the internal standard is spiked into the resulting plasma and a liquid-liquid or solid-phase extraction is performed with the purpose of removing as many matrix chemicals as possible while recovering nearly all of the analyte (Buhrman, D. L.; Price, P. I.; Rudewicz, P. J. Journal of the American Society for Mass Spectrometry 1996, 7, 1099-1105). The extracted phase is typically dried by evaporating the solvent and then resuspended in the a solvent used as the high performance liquid chromatography (HPLC) mobile phase (Matuszewski, B. K.; Constanzer, M. L.; Chavez-Eng, C. M., Ithaca, New York, Jul. 23-25, 1997; 882-889). Finally, the sample is separated in the course of an HPLC run for approximately 5-10 minutes, and the eluent is analyzed by electrospray ionization-tandem mass spectrometry (Hopfgartner, G.; Bourgogne, E. Mass Spectrometry Reviews 2003, 22, 195-214).
Methods and systems of the invention avoid the above sample work-up steps. Methods and systems of the invention analyze a dried blood spots in a similar fashion, with a slight modification to the extraction procedure. First, a specialized device is used to punch out identically sized discs from each dried blood spot. The material on these discs is then extracted in an organic solvent containing the internal standard (Chace, D. H.; Kalas, T. A.; Naylor, E. W. Clinical Chemistry 2003, 49, 1797-1817). The extracted sample is dried on the paper substrate, and the analysis proceeds as described herein.
Examples below show that methods and systems of the invention can directly detect individual components of complex mixtures, such as caffeine in urine, 50 μg of cocaine on a human finger, 100 μg of heroin on a desktop surface, and hormones and phospholipids in intact adrenal tissue, without the need for sample preparation prior to analysis (See Examples below). Methods and systems of the invention allow for simple imaging experiments to be performed by examining, in rapid succession, needle biopsy tissue sections transferred directly to paper.
Analytes from a solution are applied to the porous material for examination and the solvent component of the solution can serve as the electrospray solvent. In certain embodiments, analytes (e.g., solid or solution) are pre-spotted onto the porous material, e.g., paper, and a solvent is applied to the material to dissolve and transport the analyte into a spray for mass spectral analysis.
In certain embodiments, a solvent is applied to the porous material to assist in separation/extraction and ionization. Any solvents may be used that are compatible with mass spectrometry analysis. In particular embodiments, favorable solvents will be those that are also used for electrospray ionization. Exemplary solvents include combinations of water, methanol, acetonitrile, and THF. The organic content (proportion of methanol, acetonitrile, etc. to water), the pH, and volatile salt (e.g. ammonium acetate) may be varied depending on the sample to be analyzed. For example, basic molecules like the drug imatinib are extracted and ionized more efficiently at a lower pH. Molecules without an ionizable group but with a number of carbonyl groups, like sirolimus, ionize better with an ammonium salt in the solvent due to adduct formation.
In certain embodiments, a multi-dimensional approach is undertaken. For example, the sample is separated along one dimension, followed by ionization in another dimension. In these embodiments, separation and ionization can be individually optimized, and different solvents can be used for each phase.
In other embodiments, transporting the analytes on the paper is accomplished by a solvent in combination with an electric field. When a high electric potential is applied, the direction of the movement of the analytes on paper is found to be related to the polarity of their charged forms in solution. Pre-concentration of the analyte before the spray can also be achieved on paper by placing an electrode at a point on the wetted paper. By placing a ground electrode near the paper tip, a strong electric field is produced through the wetted porous material when a DC voltage is applied, and charged analytes are driven forward under this electric field. Particular analytes may also be concentrated at certain parts of the paper before the spray is initiated.
In certain embodiments, chemicals are applied to the porous material to modify the chemical properties of the porous material. For example, chemicals can be applied that allow differential retention of sample components with different chemical properties. Additionally, chemicals can be applied that minimize salt and matrix effects. In other embodiments, acidic or basic compounds are added to the porous material to adjust the pH of the sample upon spotting. Adjusting the pH may be particularly useful for improved analysis of biological fluids, such as blood. Additionally, chemicals can be applied that allow for on-line chemical derivatization of selected analytes, for example to convert a non-polar compound to a salt for efficient electrospray ionization.
In certain embodiments, the chemical applied to modify the porous material is an internal standard. The internal standard can be incorporated into the material and released at known rates during solvent flow in order to provide an internal standard for quantitative analysis. In other embodiments, the porous material is modified with a chemical that allows for pre-separation and pre-concentration of analytes of interest prior to mass spectrum analysis.
The spray droplets can be visualized under strong illumination in the positive ion mode and are comparable in size to the droplets emitted from a nano-electrospray ion sources (nESI). In the negative ion mode, electrons are emitted and can be captured using vapor phase electron capture agents like benzoquinone. Without being limited by any particular theory or mechanism of action, it is believed that the high electric field at a tip of the porous material, not the fields in the individual fluid channels, is responsible for ionization.
The methodology described here has desirable features for clinical applications, including neotal screening, therapeutic drug monitoring and tissue biopsy analysis. The procedures are simple and rapid. The porous material serves a secondary role as a filter, e.g., retaining blood cells during analysis of whole blood. Significantly, samples can be stored on the porous material and then analyzed directly from the stored porous material at a later date without the need transfer from the porous material before analysis. Systems of the invention allow for laboratory experiments to be performed in an open laboratory environment.
References and citations to other documents, such as patents, patent applications, patent publications, journals, books, papers, web contents, have been made throughout this disclosure. All such documents are hereby incorporated herein by reference in their entirety for all purposes.
Various modifications of the invention and many further embodiments thereof, in addition to those shown and described herein, will become apparent to those skilled in the art from the full contents of this document, including references to the scientific and patent literature cited herein. The subject matter herein contains important information, exemplification and guidance that can be adapted to the practice of this invention in its various embodiments and equivalents thereof.
The following examples are intended to further illustrate certain embodiments of the invention, and are not to be construed to limit the scope of the invention. Examples herein show that mass spectrometry probes of the invention can ionize chemical and biological samples, allowing for subsequent mass analysis and detection. An exemplary probe was constructed as a paper triangle, which was used to generate micron scale droplets by applying a high potential on the paper. The analytes were ionized from these electrically charged droplets and transported into a conventional mass spectrometer.
Examples below show that a wide range of samples could be directly analyzed in the ambient environment by probes of the invention in both of pure state and complex mixtures. The results showed that paper-based spray has the following benefits: it operated without sheath gas, i.e., few accessories were required for in situ analysis; biological samples (dried blood, urine) could be stored on the precut filter papers for months before analysis; filter paper minimized matrix effects seen with electrospray or nano electrospray in many samples (blood cells, salt and proteins) and enhanced the MS signal of chemicals in complex samples; powdered samples were easily collected by swabbing surfaces using paper pieces and then directly analyzed; the paper could be pretreated to contain internal standards that were released at known rates during solvent flow in quantitative analysis; and the paper could be pretreated to contain matrix suppression or absorption sites or to perform ion exchange or to allow on-line chemical derivatization of selected analytes.
Detection of most analytes was achieved as low as ppb levels (when examined as solutions) or in the low ng to pg range (when solids were examined) and the detection time was less than one minute. Certain Examples below provide a protocol for analyzing a dried blood spot, which can also be used for in situ analysis of whole blood samples. The dried blood spot method is also demonstrated to be compatible with the storage and transport of blood sample for blood screening and other clinical tests.
Devices of the invention integrated the capabilities of sampling, pre-separation, pre-concentration and ionization. Methods and systems of the invention simplify the problem of sample introduction in mass analyzers.
Filter paper was cut into triangular pieces with dimensions of 10 mm long and 5 mm wide and used as a sprayer (
Samples were directly applied to the paper surface that served as a sample purification and pre-concentration device. Filter paper allowed liquid samples to move through the hydrophilic network driven by capillary action and electric effects and to transport them to the tip of the paper. Separation could take place during this transport process. Sample solution was sprayed from the tip and resulted in ionization and MS detection when a high voltage (˜4.5 kV) was applied to the paper surface.
All experiments were carried out with a Finnigan LTQ mass spectrometer (Thermo Electron, San Jose, CA). The typical temperature of the capillary inlet was set at 150° C. while 30° C. for heroin detection. The lens voltage was set at 65 V for sample analysis and 240 V for survival yield experiment. Tandem mass spectra were collected using collision-induced dissociation (CID) to identify analytes in tested samples, especially for complex mixtures and blood samples.
Spray was produced by applying a high potential on the wetted paper triangle. One paper triangle was placed in front of the inlet of LTQ with its sharp tip facing to the inlet, separated by 3 mm or more. Typically, 10 uL sample solution was applied to wet the paper triangle. The solution can wet or saturate the paper or form a thin layer of liquid film on the surface of there paper. A high potential (3-5 kV) was applied between the paper triangle and mass inlet to generate an electric field, which induced a charge accumulation on the liquid at the tip of paper triangle. The increasing coulombic force breaks the liquid to form charged droplets and then the solvent evaporated during the flight of droplets from the paper tip to the mass analyzer. Paper spray required no sheath gas, heating or any other assistance to remove the solvent.
When liquid accumulated on the paper triangle, a Taylor cone was observed at the tip when examined with a microscope. The droplets formed were clearly visible under strong illumination. The Taylor cone and visible spray disappeared after a short time of evaporation and spray. However, the mass signal lasted for a much longer period (several minutes). This revealed that the paper triangle could work in two modes for mass analysis. In a first mode, the liquid was transported inside the paper at a rate faster than the liquid could be consumed as spray at the paper tip, resulting in a large cone being formed at the paper tip and droplets being generated. In a second mode, the liquid transport inside the paper was not able to move at a rate fast enough to keep up with the spray consumption, and droplets were not visible. However, it was observed that ionization of analytes did take place. The first mode provided ESI like mass spectra and the second mode provided spectra with some of the features APCI spectra. In the latter case, the paper triangle played a role analogous to a conductive needle to generate a high electric field to ionize the molecules in the atmosphere. It was observed that the mass signal in the first mode was stronger than the mass signal in the second mode by approximately two orders of magnitude under the conditions and for the samples tested.
a. Probe Materials
A number of porous materials were tested to generate charged droplets for mass spectrometry. The materials were shaped into triangles having sharp tips and sample solution was then applied to the constructed probes. Data herein show that any hydrophilic and porous substrate could be used successfully, including cotton swab, textile, plant tissues as well as different papers. The porous network or microchannels of these materials offered enough space to hold liquid and the hydrophilic environment made it possible for liquid transport by capillary action. Hydrophobic and porous substrates could also be used successfully with properly selected hydrophobic solvents.
For further investigation, six kinds of commercialized papers were selected and qualitatively tested to evaluate their capabilities in analyte detection. Filter papers and chromatography paper were made from cellulose, while glass microfiber filter paper was made from glass microfiber.
It was hypothesized that the glass fiber paper was working on mode II and prohibiting efficient droplet generation, due to the relative large thickness (˜ 2 mm). This hypothesis was proved by using a thin layer peeled from glass fiber paper for cocaine detection. In that case, the intensity of the background increased and a cocaine peak was observed. All filter papers worked well for cocaine detection, (
b. Probe Shape and Tip Angle
Many different probe shapes were investigated with respect to generating droplets. A preferred shape of the porous material included at least one tip. It was observed that the tip allowed ready formation of a Taylor cone. A probe shape of a triangle was used most often. As shown in
A paper triangle was mounted on a 2D moving stage to determine how the mass signal was affected by the relative positions of the paper triangle and the mass spectrometer inlet. The paper triangle was moved 8 cm in the y-direction in a continuous manner and 3 cm in the x-direction with a 2 mm increment for each step (
Spray duration was also tested (
In another test, the paper was sealed with a PTFE membrane to prevent evaporation of solution, which prolonged the spray time by about three times. These results indicate that paper spray offers long enough time of spray for data acquisition even using 5 uL solution, and the intensity of signal is stable during the entire spray period.
Probes of the invention include a porous material, such as paper, that can function to both separate chemicals in biological fluids before in situ ionization by mass spectrometry. In this Example, the porous material for the probe was chromatography paper. As shown in
The chromatography paper thus allowed for sample collection, analyte separation and analyte ionization. This represents a significant simplification of coupling chromatography with MS analysis. Chromatography paper is a good material for probes of the invention because such material has the advantage that solvent movement is driven by capillary action and there is no need for a syringe pump. Another advantage is that clogging, a serious problem for conventional nanoelectrospray sources, is unlikely due to its multi-porous characteristics. Therefore, chromatography paper, a multi-porous material, can be used as a microporous electrospray ionization source.
As already described, probes and methods of the invention offer a simple and convenient ionization method for mass spectrometry. Paper triangles were spotted with different compounds and connected to a high voltage source to produce ions. All experiments were carried out with a Finnigan LTQ mass spectrometer (Thermo Electron, San Jose, CA). Data herein show that a variety of chemicals could be ionized in solution phase, including amino acid, therapeutic drugs, illegal drugs and peptides.
Data herein show that probes of the invention work well over the mass/charge range from 50 to over 1000 for detection of pure compounds. Data further shows that detection was achieved down to as low as 1 ng/mL for most chemicals, including illegal drugs, such as heroin, cocaine and methadone.
Complex mixtures such as urine, blood, and cola drink were examined using methods, devices, and systems of the invention. All experiments were carried out with a Finnigan LTQ mass spectrometer (Thermo Electron, San Jose, CA).
Body fluids, such as plasma, lymph, tears, saliva, and urine, are complex mixtures containing molecules with a wide range of molecular weights, polarities, chemical properties, and concentrations. Monitoring particular chemical components of body fluids is important in a number of different areas, including clinical diagnosis, drug development, forensic toxicology, drugs of abuse detection, and therapeutic drug monitoring. Tests of blood, including the derived fluids plasma and serum, as well as on urine are particularly important in clinical monitoring.
A wide variety of chemicals from blood are routinely monitored in a clinical setting. Common examples include a basic metabolic panel measuring electrolytes like sodium and potassium along with urea, glucose, and creatine and a lipid panel for identifying individuals at risk for cardiovascular disease that includes measurements of total cholesterol, high density lipoprotein (HDL), low density lipoprotein (LDL), and triglycerides. Most laboratory tests for chemicals in blood are actually carried out on serum, which is the liquid component of blood separated from blood cells using centrifugation. This step is necessary because many medical diagnostic tests rely on colorimetric assays and therefore require optically clear fluids. After centrifugation, detection of the molecule of interest is carried in a number of ways, most commonly by an immunoassay, such as an enzyme-linked immunosorbent assay (ELISA) or radioimmunoassay (RIA), or an enzyme assay in which the oxidation of the molecule of interest by a selective enzyme is coupled to a reaction with a color change, such as the tests for cholesterol (oxidation by cholesterol oxidase) or glucose (oxidation by glucose oxidase).
There is considerable interest in the pharmaceutical sciences in the storage and transportation of samples of whole blood as dried blood spots on paper (N. Spooner et al. Anal Chem., 2009, 81, 1557). Most tests for chemicals found in blood are carried out on a liquid sample, typically serum or plasma isolated from the liquid whole blood. The required storage, transportation, and handling of liquid blood or blood components present some challenges. While blood in liquid form is essential for some tests, others can be performed on blood or other body fluids that have been spotted onto a surface (typically paper) and allowed to dry.
Probes and methods of the invention can analyze whole blood without the need for any sample preparation. The sample was prepared as follows. 0.4 uL blood was directly applied on the center of paper triangle and left to dry for about 1 min. to form a dried blood spot (
Different drugs were spiked into whole blood and the blood was applied to probes of the invention as described above. Detection of different drugs is described below.
Imatinib (GLEEVEC), a 2-phenylaminopyrimidine derivative, approved by the FDA for treatment of chronic myelogenous leukemia, is efficacious over a rather narrow range of concentrations. Whole human blood, spiked with imatinib at concentrations including the therapeutic range, was deposited on a small paper triangle for analysis (
Atenolol, a β-blocker drug used in cardiovascular diseases, was tested using the dried blood spot method to evaluate paper spray for whole blood analysis. Atenolol was directly spiked into whole blood at desired concentrations and the blood sample was used as described above for paper spray. The protonated atenolol of 400 pg (1 ug/mL atenolol in 0.4 uL whole blood) in dried blood spot was shown in mass spectra, and the MS/MS spectra indicated that even 20 pg of atenolol (50 ug/mL atenolol in 0.4 uL whole blood) could be identified in the dried blood spot (
It was also observed that the paper medium served a secondary role as a filter, retaining blood cells. Significantly, samples were analyzed directly on the storage medium rather than requiring transfer from the paper before analysis. All experiments were done in the open lab environment. Two additional features indicated that the methodology had the potential to contribute to increasing the use of mass spectrometry in primary care facilities: blood samples for analysis were drawn by means of a pinprick rather than a cannula; and the experiment was readily performed using a handheld mass spectrometer (
Systems and methods of the invention were compatible with a handheld mass spectrometer. Paper spray was performed using a handheld mass spectrometer (Mini 10, custom made at Purdue University). Analysis of whole blood spiked with 10 μg/mL atenolol. Methanol/water (1:1; 10 μL) was applied to the paper after the blood (0.4 uL) had dried (˜ 1 min) to generate spray for mass detection (
Sample collection by paper wiping followed by analysis using probes of the invention was used for fast analysis of agrochemicals on fruit. Chromatography paper (3×3 cm) wetted with methanol was used to wipe a 10 cm2 area on the peel of a lemon purchased from a grocery store. After the methanol had dried, a triangle was cut from the center of the paper and used for paper spray by applying 10 μL methanol/water solution. The spectra recorded (
Systems and methods of the invention were used to analyze human prostate tumor tissue and normal tissue. Tumor and adjacent normal tissue sections were 15 μm thick and fixed onto a glass slide for an imaging study using desorption electrospray ionization (DESI). A metal needle was used to remove a 1 mm2×15 μm volume of tissue from the glass slide from the tumor region and then from the normal region and place them onto the surface of the paper triangle for paper spray analysis.
A droplet of methanol/water (1:1 v:v; 10 μl) was added to the paper as solvent and then 4.5 kV positive DC voltage applied to produce the spray. Phospholipids such as phosphatidylcholine (PC) and sphingomyelin (SM) were identified in the spectrum (
The administration of a drug depends on managing the appropriate dosing guidelines for achievement of a safe and effective outcome. This guideline is established during clinical trials where the pharmacokinetics (PK) and pharmacodynamics (PD) of the drug are studied. Clinical trials use PK-PD studies to establish a standard dose, which may be fixed or adjusted according formulas using variables like body mass, body surface area, etc. However, the drug exposure, i.e. the amount of drug circulating over time, is influenced by a number of factors that vary from patient to patient. For example, an individuals' metabolic rate, the type and level of plasma proteins, and pre-existing conditions such as renal and/or hepatic impairment all play a role in affecting the exposure of the drug in vivo. Further, administration of a drug in combination with other medications may also affect exposure. As a result, it is often difficult to predict and prescribe an optimum regimen of drug administration.
Over- or underexposure to a drug can lead to toxic effects or decreased efficacy, respectively. To address these concerns, therapeutic drug monitoring (TDM) can be employed. TDM is the measurement of active drug levels in the body followed by adjustment of drug dosing or schedules to increase efficacy and/or decrease toxicity. TDM is indicated when the variability in the pharmacokinetics of a drug is large relative to the therapeutic window, and there is an established relationship between drug exposure and efficacy and/or toxicity. Another requirement for TDM is that a sufficiently precise and accurate assay for the active ingredient must be available. Immunoassays and liquid chromatography mass spectrometry (LC-MS) are commonly used methods for TDM. In comparison with immunoassay, LC-MS has advantages which include wide applicability, high sensitivity, good quantitation, high specificity and high throughput. Probes of the invention may be coupled with standard mass spectrometers for providing point-of-care therapeutic drug monitoring.
The drug Imatinib (GLEEVEC in USA and GLIVEC in Europe/Australia, for the treatment of chronic myelogenous leukemia) in a dried blood spot was analyzed using paper spray and a lab-scale LTQ mass spectrometer. Quantitation of Imatinib in whole blood was achieved using the MS/MS spectra with a known concentration of Imatinib-d8 being used as the internal standard (
Multiple-tip devices were fabricated and applied for high throughput analysis (
To prevent the contaminant during spray, blanks were inserted between two sample tips.
Direct analysis of chemicals in animal tissue using probes of the invention was performed as shown in
Lipid profiles were obtained for human prostate tissues (1 mm2×15 μm, Panel (C) & (D)) removed from the tumor and adjacent normal regions. Phospholipids such as phosphatidylcholine (PC) and sphingomyelin (SM) were identified in the spectra. The peak of [PC(34:1)+K]+ at m/z 798 was significantly more intense in tumor tissue (panel C) and peaks [SM(34:1)+Na]+ at m/z 725, [SM(36:0)+Na]+ at m/z 756, and [SM(36:4)+Na]+ at m/z 804 were significantly lower compared with normal tissue (panel D).
For analysis of target analytes which have relatively low ionization efficiencies and relatively low concentrations in mixtures, derivatization is often necessary to provide adequate sensitivity. On-line derivatization can be implemented by adding reagents into the spray solution, such as methanol/water solutions containing reagents appropriate for targeted analytes. If the reagents to be used are stable on paper, they can also be added onto the porous material when the probes are fabricated.
As a demonstration, 5 μL methanol containing 500 ng betaine aldehyde chloride was added onto a paper triangle and allowed to dry to fabricate a sample substrate preloaded with a derivatization reagent for the analysis of cholesterol in serum. On-line charge labeling with betaine aldehyde (BA) through its reaction with hydroxyl groups has been demonstrated previously to be very effective for the identification of cholesterol in tissue (Wu et al., Anal Chem. 2009, 81:7618-7624). When the paper triangle was used for analysis, 2 μL human serum was spotted onto the paper to form a dried spot and then analyzed by using paper spray ionization. A 10 μL ACN/CHCl3 (1:1 v:v) solution, instead of methanol/water, was used for paper spray to avoid reaction between the betaine aldehyde and methanol.
The comparison between analysis using a blank and a reagent-preloaded paper triangle is shown in
Pre-concentration of chemicals on the paper surface using photoresist treatment. Chromatography paper was rendered hydrophobic by treatment with SU-8 photoresist as described previously (Martinez et al., Angew Chem Int. Ed., 2007, 46:1318-1320). Then 5 μl bradykinin 2-9 solution (100 ppm in pure H2O) was applied on the paper surface. When the solution was dry, the paper was put into water and washed for 10 s. After washing, the paper triangle was held in front of the MS inlet, 10 μl pure MeOH was applied as solvent and the voltage was set at 4.5 kV for paper spray. The same experiment was done with untreated paper substrate for comparison.
The polarity of the voltage applied to the probe need not match that used in the mass analyzer. In particular, it is possible to operate the probes of the invention with a negative potential but to record the mass spectrum of the resulting positively changed ions. In negative ion mode, a large current of electrons (or solvated electrons) is produced in paper spray. These electrons, if of suitable energy, can be captured by molecules with appropriate electron affinities to generate radical anions.
Alternatively, these electrons might be responsible for electron ionization of the analyte to generate the radical cation or alternatively ESI might involve a solvent molecule which might then undergo charge exchange with the analyte to again generate the radical cation. If this process occurs with sufficient energy, characteristic fragment ions might be produced provided the radical cation is not collisionally deactivated before fragmentation can occur.
An experiment was done on a benchtop LTP using toluene vapor, with a probe of the invention conducted at −4.5 kV with methanol:water as solvent applied to the paper. The spectrum shown in
An interesting note is that the “E1” fragment ions were most easily produced when the source of toluene vapor was placed close to the MS inlet; i.e., in the cathodic region of the discharge between the paper tip and MS inlet. This suggests that direct electron ionization by energetic electrons in the “fall” region might be at least partly responsible for this behavior.
a. Solvent Applied to Liquid Blood on a Porous Substrate and Dried Blood on a Porous Substrate
Systems and methods described herein use anhydrous salts (drying agents) to dry and thus restrict the flow of liquid blood during paper spray analysis.
b. Drying Agent for Rapid Analysis of Whole Blood
A drying agent, such as magnesium sulfate, sodium sulfate, sodium carbonate, or calcium chloride is mixed and pre-spotted along with the internal standard, thus saving a sample preparation step. Another potential agent, is a blood coagulant such as alum powder. As is seen in
Further improvements include (i) drying/crystallizing the drying agent on the paper; (ii) drying agent choice; (iii) Compatibility with paper type, e.g. thickness & pore size; (iv) combination of a drying agent and oven drying.
In certain embodiments, a desiccator may be used to store the pre-loaded paper in a moisture-free atmosphere. Such a desiccator could be any container that contains the commercially-available Drierite. This scheme is shown in
Paper spray mass spectrometry is applied to oncology drugs in fresh whole blood samples supported on filter paper substrates instead of dry blood as done previously. Addition of the coagulant alum clotted the blood and allowed for immediate sample analysis. The coagulant did not interfere with the function of the paper spray nor did it add features to the mass spectra. Quantitative analysis of therapeutic drugs in the blood was achieved utilizing internal standards which were pre-spotted onto the filter paper. Eight oncology drugs were examined, with lower limits of detection ranging between 0.5 and 17 ng/mL and linear dynamics ranges greater than two orders of magnitude. Inter-day accuracies of quality controls for pazopanib ranged from 102 to 118%, with imprecisions of 9 to 13%. This one-step method was conducted using 10 μL of blood, a drop of solvent, and took about 45 seconds per trial. These results indicate applicability to point-of-care therapeutic drug monitoring in a clinical setting.
a. Introduction
Every drug has an optimally effective therapeutic range, outside of which the drug is harmful or ineffective. Concentrations of active pharmaceutical compounds in a patient's circulatory system vary with the individual (A. Y H. Lu, Drug Metab. Dispos., 1998, 26, 1217-1222). One procedure for determining the upper limit is to increase drug dosage until dose-limiting toxicity occurs. Alternative approaches include pharmacogenomics (W. E. Evans and M. V. Relling, Science, 1999, 286, 487-491; W. E. Evans and H. L. McLeod, N. Engl. J. Med., 2003, 348, 538-549; and W. E. Evans and M. V Relling, Nature, 2004, 429, 464-468) and pharmacokinetic modeling (N. H. G. Holford and L. B. Sheiner, Clin Pharmacokinet, 1981, 6, 429-453; and L. E. Gerlowski and R. K. Jain, J. Pharm. Sci., 1983, 72, 1103-1127) which use genetics and statistics to account for the variation in uptake and metabolism of drugs. Another possibility is therapeutic drug monitoring (TDM), in which drug exposure levels within the body are measured directly by a quantitative method. Historically, immunoassays and high performance liquid chromatography with optical or mass spectrometric detection have been the methods-of-choice for therapeutic drug monitoring. Immunoassays have been used for decades for point-of-care monitoring of blood, and recent research has provided faster results for point-of-care applications (X. Yang, J. Janatova, J. M. Juenke, G. A. McMillin and J. D. Andrade, Anal. Biochem., 2007, 365, 222-229; and T. Tachi, T. Hase, Y Okamoto, N. Kaji, T. Arima, H. Matsumoto, M. Kondo, M. Tokeshi, Y Hasegawa and Y Baba, Anal. Bioanal. Chem., 2011, 401, 2301-2305) by implementing microfluidic methods (B. Zheng and R. F. Ismagilov, Angew. Chem. Int. Ed., 2005, 44, 2520-2523). The ultimate goal for TDM is individualized dosing during treatment using a platform technology for fast, accurate, and inexpensive assays directly from untreated biofluid.
This Example shows the use of coagulants for rapid quantitative analysis of oncology drugs from fresh whole blood by paper spray mass spectrometry. Oncology drug analysis has been chosen as a model problem for point-of-care analysis due to the need for fast results during chemotherapy. The quantitative results herein show that paper spray has sufficient sensitivity and precision, with regard to clinical regulations, to provide the first one-step analysis of fresh whole blood for applications in point-of-care mass spectrometry.
b. Materials
Most drug standards were purchased from Sigma-Aldrich (St. Louis, MO, USA) and used without any further purification. All organic solvents (HPLC grade) were purchased from VWR Scientific (Chicago, IL, USA). Blood card paper (grade 31 ETF) was obtained as a sample from Whatman (Piscataway, NJ, USA). Bovine whole blood with potassium ethylene diamine tetraacetic acid (K2EDTA) was purchased from Innovative Research (Novi, MI, USA) and stored at 4° C.
c. Preparation of Samples
All stock solutions were prepared in methanol, and working solutions were prepared in 50:50 methanol/water. Spiked blood samples were prepared by adding the necessary quantity of working solution not to exceed 5% of the volume of blood (typically, addition of 50 □L 20× concentration solution to 950 □L of bovine whole blood). Before sample preparation, the blood was incubated to a temperature of 37° C. The blood standards were vortexed for 10 seconds and analyzed within 1 hour of preparation.
d. Mass Spectrometry and Paper Spray Ionization
For blood sample analysis, the back edge of the paper triangle was held by a toothless copper alligator clip which was connected to the high voltage power supply (
Dried blood spot analysis was performed in the same fashion as fresh blood, but without the coagulant and with an extra two hours for blood drying, in accordance to the procedure of Manicke et al. (J. Am. Soc. Mass. Spectrom., 2011, 22, 1501-1507). All experiments were performed using a TSQ Quantum Access MAX (Thermo Scientific, San Jose, CA, USA). Data collection was executed in selected reaction monitoring (SRM) mode (R. W. Kondrat and R. G. Cooks, Anal. Chem., 1978, 50, A81-A92), observing the desired fragment intensity for the drugs and internal standards by collision-induced dissociation (CID). Scan times were 250 ms, alternating between the drug and internal standard, and data were collected until the spray voltage was manually turned off at 15 seconds, totaling 30 scans for each fragment ion.
e. Data Analysis
The two sets of 30 scans were plotted as individual ion chronograms, showing ion current versus time for each fragment ion. The area under the curve corresponding to the drug was compared to that of the internal standard and used for quantitation. Limits of detection were determined using a calibration curve consisting of four trials each of 0, 50, 100, 500, and 1000 ng/mL drug standard solutions. Limits of detection were determined by multiplying the t-value (95% confidence) by the standard deviation of the blank, divided by the slope of the calibration curve determined by 1/x2 weighted linear least squares (G. L. Long and J. D. Winefordner, Anal. Chem., 1983, 55, A712-A724). Reported precision values are the relative standard deviation of concentration determined from multiple blood spots taken from a single blood sample. Accuracy is defined to be the measured concentration divided by the actual concentration multiplied by 100%.
f. Results
One of the primary challenges faced by rapid blood analysis with paper spray is the contamination of the spray tip by blood components. In the case of freshly-spotted blood that was not dried, oxygenated hemoglobin—the red component of blood—was seen to run to the tip (see
One approach for the rapid paper spray analysis of fresh blood samples was the addition of coagulants to the paper. A pre-spotted coagulant, potash alum, quickly clotted the blood (within seconds) and allowed the spray solvent to wick through the blood spot and to the tip of the paper, dissolving analytes in the process (
g. Therapeutic Drug Monitoring
Two important parameters characterizing the method are precision and accuracy. The imprecision should not exceed a relative standard deviation of 15% (20% at the lower limit of quantitation). The inaccuracy of quality control samples likewise should not exceed ±15% (±20% at the lower limit of quantitation). Table 2 is a representative quality control table for analysis of pazopanib in blood, portraying the inter- and intra-day accuracy and reproducibility of the oncology drug assay (n=5 for each concentration, each day). Pazopanib is an anti-cancer drug which has been approved by the U.S. FDA.
The quality control results from Table 2 suggest that this method is a viable bioanalytical method for point-of-care therapeutic drug monitoring. Most of the quality control samples yielded a positive bias, which was most likely due to error in the sample preparation process. All of the inter-day percent accuracies and standard deviations for this data set produced errors under 15%, on average 10.8%. This is comparable to previous work on dried blood spots by paper spray with pre-spotted internal standards, which had about 10% relative standard deviations (N. E. Manicke, P. Abu-Rabie, N. Spooner, Z. Ouyang and R. G. Cooks, J. Am. Soc. Mass. Spectrom., 2011, 22, 1501-1507). The most unusual result, from 750 ng/mL on Day 1, seems especially erroneous and was most likely caused by experimental errors during the paper spray process.
It is conceivable that in the course of point-of-care work, reanalysis of a sample will be needed due to instrument malfunction or required immediate result confirmation. Therefore, reinjection reproducibility was analyzed in which the solvent was reapplied and paper spray was performed on the same blood spot for a second time. In general, paper spray does not have high recovery of the analyte. That is, the amount of analyte extracted and sprayed from the paper toward the mass spectrometer is only a portion of the available analyte (Z. Zhang, W. Xu, N. E. Manicke, R. G. Cooks, Z. Ouyang, Anal. Chem., 2012, DOI: 10.1021/ac202058w). Although the relatively lower recovery contributed to poorer sensitivity, it did allow for reanalysis of the same blood spot. Table 3 shows how analyzing the same blood sample a second time did not affect the accuracy of the results, with the exception of the 10 ng/mL samples, the concentrations of which were less than three times the lower limit of quantitation.
It should be noted that the parameters for paper spray were chosen carefully for optimal performance. To achieve precision of better than 15%, it was necessary to use an isotopically-labeled internal standard due to the various affinities of the analytes for the substrate, as well as inefficiencies in ionization and ion transfer. In addition, although volumes of blood less than 5 μL clotted more efficiently, 10 μL of blood was typically used in order to improve the standard deviation of the method. It was difficult, for example, to control the blood volume, spot size, and spot shape for smaller volumes of micro-pipetted blood. An automated technique would allow for smaller blood volumes to be analyzed more reproducibly. The spray solvent, methanol, was chosen to achieve proper wicking through the paper, good analyte extraction, and low background noise. The background noise was reduced by using the lowest spray voltage possible, typically 3.0 kV
A wide variety of oncology drugs was explored by the same method. These drugs varied in molecular weight from 261 g/mol (cyclophosphamide) to 854 g/mol (paclitaxel). There are many factors that affected the lower limits of detection for these drugs, including solubility in the spray solvent, affinity for the substrate and blood components, proton affinity (or Na+ affinity in the case of the taxanes which form sodium ion adducts), background noise level, and selectivity of the SRM transition. A previous study demonstrated that matrix effects cause negligible analyte suppression within the therapeutic range for drugs by paper spray, and that paper spray MS measures the total drug concentration in blood (free plus bound drug) (N. E. Manicke, Q. A. Yang, H. Wang, S. Oradu, Z. Ouyang and R. G. Cooks, Int. J. Mass spectrom., 2011, 300, 123-129).
Oncology drugs pazopanib, tamoxifen, imatinib, cyclophosphamide, paclitaxel, docetaxel, topotecan, and irinotecan were each analyzed individually to determine lower limits of detections (LLOD's). As Table 4 shows, the method described herein for rapid drug analysis from fresh blood gave LLOD's ranging from 0.5 ng/mL for pazopanib, to 17 ng/mL for topotecan; all these values are in the same order of magnitude as the dried blood spot experiments by paper spray (N. E. Manicke, P. Abu-Rabie, N. Spooner, Z. Ouyang and R. G. Cooks, J. Am. Soc. Mass. Spectrom., 2011, 22, 1501-1507). The LLOD's were within one order of magnitude compared to previously-reported HPLC-MS/MS methods of therapeutic drugs.
2H313C-
13C215N-
aTherapeutic Ranges for plasma concentrations (R. Regenthal, M. Krueger, C. Koeppel and R. Preiss, J. Clin. Monitor Comp., 1999, 15, 529-544.)
When analyzing fresh blood, it was hypothesized that any biomolecules that are soluble in the spray solvent and ionizable by paper spray will be detected by the mass spectrometer.
h. Conclusion
Quantitative analysis of oncology drugs by paper spray with alum, a coagulant, shows that this method has capabilities for point-of-care therapeutic drug monitoring. Utilizing a bench-top mass spectrometer, inter- and intra-day percent accuracies and imprecisions of quality controls and standards were within regulated guidelines (<15% error) using pre-spotted isotopically-labeled internal standards. The time span from blood spotting to results took under 60 seconds, due to the online extraction of paper spray and the coagulating ability of pre-spotted alum. The excellent limits of detection suggest that this method should be easily transferrable to a portable mass spectrometer where the larger blood volumes that can be handled quickly with the drying agent will be necessary for adequate analytical performance. The use of inexpensive expendables means that disposable cartridges could be used which would reduce sample contact and lower the potential risk for cross-contamination.
This method could also find a place in home-use applications, for example in an assisted living facility to ensure day-to-day medication levels. The small blood consumption by paper spray, several microliters, would also allow co-monitoring of blood by separate aliquots and analyzing electrolytes, blood gases, and glucose levels. In addition to therapeutic oncology drugs and metabolites, paper spray may be applied for analysis of biocompounds such as fatty acids, lipids, and proteins, as well as other applications including drugs of abuse, steroids, and human growth hormones.
a. Analysis of Whole Blood
When fresh (liquid) blood is dropped onto sample substrates of less hydrophobicity, such as print paper or Silanized paper, the blood drop stays on the top of the substrate without diffusing into the porous substrate (
The solvent can extract the chemicals from the blood and also facilitate the spray to generate ions from the substrate (
b. Direct Detection of Proteins from Whole Fresh Blood
Even though there is a large amount of proteins in whole blood, it is difficult to detect any protein spectra from a dried blood spot using chromatography paper. Without being limited by any particular theory or mechanism of action, it is believed that when blood is dried on paper, the proteins in the blood could be denatured and combine strongly with the substrate especially if the substrate is polar.
10 μL of bovine whole blood was dropped onto a paper surface. Then 50 μL methanol with 1% acetic acid was applied as solvent. 4 kV DC voltage was applied on the copper clip to produce spray.
c. Extraction of Chemicals from Fresh Whole Blood Using Organic Solvents that are not Miscible with Blood
Most of the blood sample stayed on top of the paper substrate forming a film of liquid blood instead of being absorbed into the paper. When the solvent was applied, it went through the blood sample to the tip. When high voltage is applied, a strong spray plum was produced from the tip of the paper triangle (
d. Extraction of Chemicals from Fresh Whole Blood Using Organic Solvents which are Miscible with Blood
When high voltage was applied, a strong spray plum was produced from the tip of the paper triangle (
e. Applying Sample Extraction to Make Dried Sample Spots for Later Analysis
10 μL whole fresh blood was spotted on to the print paper. Then 10 μL acetonitrile was applied onto the blood sample before it dried to extract nicotine from the blood sample. After the sample and solvent were dried on the paper, another 20 μL acetonitrile was applied as spray solvent and 4 kV DC was applied to generate paper spray. LOQ of 1 ppb could be achieved. The spectra were recorded using TSQ mass spectrometer (
The determination of tobacco nicotine alkaloids from biofluids is of great importance for a smoking test, to tobacco cessation treatment, and to the study of exposure to secondhand smoke and effect of tobacco use on individual health. Paper spray mass spectrometry has been developed for direct, quantitative analysis of tobacco nicotine alkaloids from dried and liquid biofluids such as blood, urine and saliva. Limit of quantitation as low as several ng/mL were obtained for nicotine, cotinine, trans-3′-hydroxycotinine and anabasine. Due to the fast and convenient characteristics of this method, it shows potential for significantly improving the analytical efficiency in clinical labs and also for point-of-care tobacco use assessment.
a. Introduction
Tobacco use is the chief preventable cause of disease and death in the US (Centers for Disease Control and Prevention. Annual Smoking-Attributable Mortality, Years of Potential Life Lost, and Economic Costs-United States, 1995-1999. Morbidity and Mortality Weekly Report 2002; 51(14):300-3). However, due to it addiction nature, to quit smoking is extremely difficult. Based on the data from CDC, in 2010, about 19.3% of all adults are smokers in the US (Centers for Disease Control and Prevention. Vital Signs: Current Cigarette Smoking Among Adults Aged ≥18 Years-United States, 2005-2010. Morbidity and Mortality Weekly Report 2011; 60(33):1207-12) and only less than 10 percent who quit smoking for a day remain abstinent one year later (M. C. Fiore, Med. Clin. N. Am. 1992, 76, 289-303). Smokers have much higher health risks for a series of serious diseases, such as coronary heart disease, stroke, and lung cancer. The adverse health effects of smoking accounts for about one fifth deaths each year (Centers for Disease Control and Prevention. Annual Smoking-Attributable Mortality, Years of Potential Life Lost, and Economic Costs-United States, 1995-1999. Morbidity and Mortality Weekly Report 2002; 51(14):300-3) and $193 billion were spent annually due to cigarette smoking caused health-related economic losses in the US (Centers for Disease Control and Prevention. Smoking-Attributable Mortality, Years of Potential Life Lost, and Productivity Losses-United States, 2000-2004. Morbidity and Mortality Weekly Report 2008; 57(45):1226-8).
Because of the significant adverse health effects of tobacco use, more and more insurance companies request life insurance buyers to do smoking tests. Different rating categories will be given based on the smoking status. Many employers also use smoking testing to evaluate prospective employees for tobacco use.
For tobacco use assessment and other tobacco use related clinical diagnosis, nicotine and its metabolites are most frequently analyzed chemicals (I. Kim, M. A. Huestis, J. Mass Spectrom. 2006, 41, 815-821; and J. Hukkanen, P. Jacob, N. L. Benowitz, Pharmacol. Rev. 2005, 57, 79-115). Nicotine is the major addictive chemical in tobacco which makes people difficult to quit smoking once started (N. L. Benowitz, Annu. Rev. Pharmacol. Toxicol. 2009, 49, 57-71; and N. L. Benowitz, N. Engl. J. Med. 2010, 362, 2295-2303). Nicotine is also the major component in the pharmacotherapy for smoking cessation (J. E. Henningfield, N. Engl. J. Med. 1995, 333, 1196-1203). To reduce the withdrawal symptoms during smoking cessation, nicotine medications such as nicotine gum are often utilized which provide lower, and more stable blood nicotine concentration.
Nicotine is mainly metabolized in the liver by an enzyme CYP2A6 (J. Hukkanen, P. Jacob, N. L. Benowitz, Pharmacol. Rev. 2005, 57, 79-115). The half-life of nicotine is relatively short (about 2 hours), thus the assessment of nicotine in biofluids cannot reflect the status of tobacco use accurately. Cotinine is the major metabolite of nicotine with a much longer half-life (about 16 hours). Because of the long half-life, cotinine is being used as the biomarker for cigarette smoking and environmental tobacco smoke exposure. Cotinine is further metabolized by the same enzyme CYP2A6 to trans-3′-hydroxycotinine (3HC). Recent study showed that the ratio of 3HC to cotinine could provide a convenient measure to phenotype individuals for CYP2A6 activity which is a biomarker of nicotine metabolism guiding dosage in nicotine medication (N. L. Benowitz, O. F. Pomerleau, C. S. Pomerleau, P. Jacob, Nicotine & Tobacco Research 2003, 5, 621-624; P. Jacob, L. S. Yu, M. J. Duan, L. Ramos, O. Yturralde, N. L. Benowitz, Journal of Chromatography B-Analytical Technologies in the Biomedical and Life Sciences 2011, 879, 267-276; and D. Dempsey, P. Tutka, P. Jacob, F. Allen, K. Schoedel, R. F. Tyndale, N. L. Benowitz, Clin. Pharmacol. Ther. 2004, 76, 64-72). Although nicotine and its metabolites are widely used for tobacco use assessment, they cannot be used to distinguish smokers from persons taking nicotine medications. In such a condition, another tobacco alkaloid anabasine can be used which only exists in tobacco but not in nicotine-medication products (P. Jacob, L. Yu, A. T. Shulgin, N. L. Benowitz, Am. J. Public Health 1999, 89, 731-736; and P. Jacob, D. Hatsukami, H. Severson, S. Hall, L. Yu, N. L. Benowitz, Cancer Epidemiol. Biomark. Prev. 2002, 11, 1668-1673).
This Example shows the power of this technique for direct quantitation of biomarkers of tobacco use from biofluids, including blood, urine and saliva. Besides the testing them in dried spots, the Example also demonstrates the direct analysis of fresh liquid samples.
b. Materials
Bovine whole blood was purchased from Innovative Research (Novi, MI). Synthetic urine sample was purchase from CST Technologies (Great Neck, NY). Saliva sample was donated by Dr. Ouyang who is the PI of this study.
Nicotine, cotinine, and anabasine were purchased from Sigma-Aldrich (St. Louis, MO). Trans-3′-hydroxycotinine and trans-3′-hydroxycotinine-d3 were purchased from Santa Cruz Biotechnology (Santa Cruz, CA). Nicotine-d3 and cotinine-d3 were purchased from CDN Isotopes (Quebec, Canada). Anabasine-d4 was purchased from United States Biological (Swampscott, MA).
All experiments were performed using a TSQ Quantum Access Max mass spectrometer (Thermo Scientific, San Jose, CA). All analytes were monitored in the selected reaction monitoring (SRM) mode. The SRM transitions are shown in Table 5 below.
To get the calibration curves, bovine whole blood was spiked with each chemical and corresponding isotope labeled internal standard. For dried blood spot analysis, the paper was loaded with 5 μL whole blood, placed in the open air overnight, and then stored in a sealed plastic bag containing desiccant at room temperature.
Grade 31ET chromatography paper was purchased from Whatman (Piscataway, NJ). Printer paper (letter size, 75 g/m2) was purchased from Xerox (Norwalk, CT).
For this Example, the paper was cut into a triangle (base 7 mm, height 12 mm). A copper clip was used to hold the paper triangle in front of the mass spectrometer. 4 kV DC high voltage was applied through the copper clip to induce paper spray ionization from tip of the paper triangle. Unless otherwise noted, 90% acetonitrile with 10% water was used as the spray solvent.
For the calibration curves, 3 samples were repeated at each concentration. The lower limit of quantitation (LLOQ) was calculated as 10 times of standard deviation of blank samples only with internal standards divided by the slope of the calibration curve.
c. Results
Nicotine, cotinine, 3HC, and anabasine (
Different paper substrates were tested trying to minimize the matrix interference from the paper substrates and improve the LOQ. Chromatography paper with different thickness (grade 1 0.18 mm and 31 ET 0.5 mm), silica coated paper, and printer paper were tested. Interestingly, it was found that printer paper was one of the best paper substrates for these four chemicals. For 3HC and cotinine, similar results were obtained using chromatography paper and printer paper. However, for nicotine and anabasine, printer paper provided much lower background signal than chromatography paper thus significantly improved the LOQ. It was found that when organic solvent was continuously applied on printer paper, the background was low for the first several minutes, and then the background will gradually increase. It was observed that organic solvent are more difficult to penetrate through the paper substrate to the reverse side of the paper. These observations suggest that the lower background from printer paper may be because matrix chemicals were more slowly extracted from the printer paper substrate. The slow extraction of matrix chemicals may be due to the smaller pore size of the printer paper.
For point-of-care test, people may want to get the results as fast as possible. Even though the analysis of dried blood spot using paper spray can be done within one minute, the time to allow the blood dry on the paper can take hours. Addition of the coagulant alum has been shown to allow paper spray analysis before the blood is dried. See Examples above. Here, paper spray was also tested for direct analysis of liquid blood samples without any additive. As shown in
Besides blood, cotinine was also tested directly from fresh liquid saliva and urine. 5 μL saliva sample was added onto Grade 1 chromatography paper, and then 30 μL acetonitrile was applied as solvent. The calibration curve is shown in
d. Conclusion
Data herein show direct quantitation of tobacco nicotine alkaloids from biological fluids in the form of both dried spots and fresh liquid. This method shows potential in quantitating human exposure to tobacco and in studying the nicotine metabolism.
In this Example, paper spray was coupled with a commercial TSQ and a home-made MINI 11 for analysis of different target drugs in dried blood spots, respectively. Different from previous reports, silica coated paper was used for paper spray, rather than chromatography paper or filter paper, in order to improve the analysis of therapeutic drugs in dried blood spots (DBS).
Dichloromethane/isopropanol solvent was identified as an optimal spray solvent for the analysis. The comparison was made with paper spray using chromatography paper as substrate with methanol/water as solvent for the analysis of verapamil, citalopram, amitriptyline, lidocaine and sunitinib in dried blood spots. It was demonstrated that the efficiency of recovery of the analytes was notably improved with the silica coated paper and the limit of quantitation (LOQ) for the drug analysis was 0.1 ng mL−1 using a commercial triple quadrupole mass spectrometer. The use of silica paper substrate also resulted in a sensitivity improvement of 5-50 fold in comparison with chromatography papers, including the Whatmann ET31 paper used for blood card. Analysis using a handheld miniature mass spectrometer Mini 11 gave LOQs of 10˜20 ng mL-1 for the tested drugs, which is sufficient to cover the therapeutic ranges of these drugs.
a. Introduction
Accurate measurement of therapeutic drugs and their metabolites in blood plays an important role in drug discovery and disease therapy. Storing whole blood samples as the dried blood spots (DBS) on paper is being adopted for the analysis of drugs in blood. In comparison with conventional way of collecting blood with test tubes, DBS has some special advantages including small sample volume (typically less than 50 μL), improved chemical stability for many analytes in blood and easy sample storage and transfer at ambient temperatures (Rao, R. N.; Maurya, P. K.; Ramesh, M.; Srinivas, R.; Agwane, S. B. Biomed. Chromat. 2010, 24, 1356-1364; and Li, W. K.; Zhang, J.; Tse, F. L. S. Biomed. Chromat. 2011, 25, 258-277). DBS analysis can be used for a wide range of applications, including toxicology (Barfield, M.; Spooner, N.; Lad, R.; Parry, S.; Fowles, S. J. Chromatogr. B: Analyt. Technol. Biomed. Life Sci. 2008, 870, 32-37) and pharmacokinetics studies (Lawson, G.; Tanna, S.; Mulla, H.; Pandya, H. J. Pharm. Pharmacol. 2009, 61, A33; Suyagh, M. F.; Laxman, K. P.; Millership, J.; Collier, P.; Halliday, H.; McElnay, J. C. J. Chromatogr. B: Analyt. Technol. Biomed. Life Sci. 2010, 878, 769-776; and Spooner, N.; Lad, R.; Barfield, M. Anal. Chem. 2009, 81, 1557-1563) for drug discovery as well as therapeutic drug monitoring (TDM) to assist in dosage optimization during therapy (Ntale, M.; Mahindi, M.; Ogwal-Okeng, J. W.; Gustafsson, L. L.; Beck, O. J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2007, 859, 137-140; Lejeune, D.; Souletie, I.; Houze, S.; Le Bricon, T.; Le Bras, J.; Gourmel, B.; Houze, P. J. Pharm. Biomed. Anal. 2007, 43, 1106-1115; Ronn, A. M.; Lemnge, M. M.; Angelo, H. R.; Bygbjerg, I. C. Therap. Drug Monitor. 1995, 17, 79-83; Tawa, R.; Hirose, S.; Fujimoto, T. J. Chromatogr. B: Biomed. Appl. 1989, 490, 125-132; Croes, K.; McCarthy, P. T.; Flanagan, R. J. J. Anal. Toxicol. 1994, 18, 255-260; Heine, R.; Rosing, H.; van Gorp, E. C. M.; Mulder, J. W.; van der Steeg, W. A.; Beijnen, J. H.; Huitema, A. D. R. J. Chromatogr. B: Analyt. Technol. Biomed. Life Sci. 2008, 867, 205-212; Koal, T.; Burhenne, H.; Romling, R.; Svoboda, M.; Resch, K.; Kaever, V. Rapid Commun. Mass Spectrom. 2005, 19, 2995-3001; Cheung, C. Y.; van derHeijden, J.; Hoogtanders, K.; Christiaans, M.; Liu, Y. L.; Chan, Y. H.; Choi, K. S.; van de Plas, A.; Shek, C. C.; Chau, K. F.; Li, C. S.; van Hooff, J.; Stolk, L. Transplant Int. 2008, 21, 140-145; Coombes, E. J.; Gamlen, T. R.; Batstone, G. F.; Leigh, P. N. Ann. Clin. Biochem. 1984, 21, 519-522; Fujimoto, T.; Tsuda, Y.; Tawa, R.; Hirose, S. Clin. Chem. 1989, 35, 867-869; and Li, P. K.; Lee, J. T.; Conboy, K. A.; Ellis, E. F. Clin. Chem. 1986, 32, 552-555)
Regardless of how blood samples are collected or stored, similar procedures have been applied for the chemical analysis of the therapeutic drugs in blood. The analytes are extracted from the blood sample using organic solvents, then separated using chromatography and analyzed using mass spectrometry (MS), ultraviolet (UV), fluorescence (FL), or immunoassay. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) has been the mainstream method for the quantitation of drugs in blood (Taylor, P. J.; Tai, C.-H.; Franklin, M. E.; Pillans, P. I. Clin. Biochem. 2011, 44, 14-20; Saint-Marcoux, F.; Sauvage, F.-L.; Marquet, P. Anal. Bioanal. Chem. 2007, 388, 1327-1349; Korecka, M.; Shaw, L. M. Ann. Transplant. 2009, 14, 61-72; and Checa, A.; Oliver, R.; Hernandez-Cassou, S.; Saurina, J. Anal. Chim. Acta 2009, 647, 1-13). High sensitivity and selectivity are obtained with LC-MS/MS for quantitative analysis of therapeutic drugs in blood. The standard procedure involves complex sample preparation and analyte separation prior to the MS analysis. These steps are essential for minimizing matrix effects and improving the detection limits (Breadmore, M. C.; Theurillat, R.; Thormann, W. Electrophoresis 2004, 25, 1615-1622; and Lee, E. D.; Henion, J. D. Rapid Commun. Mass Spectrom. 1992, 6, 727-733), but could take about 30 minutes to several hours. This time frame might not be rate determining for in-lab analysis of many samples where parallel high throughput approaches can be used, but could be so when small numbers of samples are analyzed and rapid decision making is required. Used in this way, LC-MS/MS can be applied to a wide range of drug compounds; however, this method must be performed in analytical laboratories and only by experienced chemists.
For analysis of therapeutic drugs in DBS using paper spray, the drug compounds are extracted from the blood matrix by the wetting solution. The extent of the extraction, relative to other compounds eluting from the DBS, has a direct impact on the sensitivity of the analysis. The affinity of the drug compounds to a paper substrate also affects the extraction efficiency and causes loss of compound during transport to the paper tip by the spray solvent and inefficiency in ionization. The chemical properties of the solvent need to be optimized for the extraction and transfer process, but they also need to be suitable for the spray ionization process, which is important to the MS analysis sensitivity. The selection of the combination of a paper substrate and a solvent is important to the overall performance of the paper spray MS analysis.
Generally, chromatography paper has mainly been used for paper spray substrates with methanol/water as the wetting and spray solvent. This Example investigates the use of silica-coated paper for the analysis of therapeutic drugs in dried blood spots. Silica is by far the most widely used matrix for chromatographic separation owing to its chemical and mechanical stability, variable pore size, well documented chemistry for surface modification, and excellent demonstrated performance for separations. Silica-coated paper is an excellent substrate for resolving a wide variety of compounds (Marinetti, G. V. In Lipid Chromatographic Analysis; Wuthier, R. E., Ed.; Marcel Dekker: New York, 1976; Vol. 1, pp 59-109; Valadon, L. R. G.; Mummery, R. S. Phytochemistry 1972, 11, 413-414; and Egan, R. W. Anal. Biochem. 1975, 68, 654-657).
In this Example, paper coated with silica gel was adopted as the paper spray substrate and relatively less polar solvents were used. The selection of the substrate/solvent systems was first optimized for drug analysis using a commercial triple quadrupole mass spectrometer and the performance using a handheld ion trap mass spectrometer (Mini 11; Gao, L.; Sugiarto, A.; Harper, J. D.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2008, 80, 7198-7205; Hou, K. Y.; Xu, W.; Xu, J. A.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2011, 83, 1857-1861; and Sokol, E.; Noll, R. J.; Cooks, R. G.; Beegle, L. W.; Kim, H. I.; Kanik, I. Int. J. Mass Spectrom. 2011, In Press, Corrected Proof) was then characterized for a set of therapeutic drugs from DBSs. The combination of the paper spray with a miniature mass spectrometer is of a great interest for point-of-care applications where small-size equipment and simple operational procedures are highly desirable. In the Mini 11 handheld mass spectrometer, a single stage discontinuous atmospheric pressure interface (DAPI; Gao, L.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2008, 80, 4026-4032) was used with an ion trap. A TSQ (Thermo Scientific, Inc., San Jose, CA) with multiple differential pumping stages and triple quadrupole mass analyzers can provide 50 times better LOQ for the analysis of therapeutic drugs in blood using paper spray.
b. Materials
The procedure of paper spray used in this Example was similar to that previously reported (Wang, H.; Liu, J. J.; Cooks, R. G.; Ouyang, Z. Angew. Chem., Int. Ed. 2010, 49, 877-880; Wang, H.; Manicke, N. E.; Yang, Q.; Zheng, L.; Shi, R.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2011, 83, 1197-1201; Manicke, N. E.; Yang, Q. A.; Wang, H.; Oradu, S.; Ouyang, Z.; Cooks, R. G. Int. J. Mass Spectrom. 2011, 300, 123-129; and Liu, J.; Wang, H.; Manicke, N. E.; Lin, J.-M.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2010, 82, 2463-2471). Briefly, the standards used herein, including verapamil, sunitinib, citalopram, amitriptyline, and lidocaine, were prepared as follows: drug solutions at 100× concentration were prepared by dilution of the stock solutions into 1:1 methanol/water. The 100× standards were then spiked into bovine whole blood (with sodium citrate as anticoagulant, Innovative Research, Novi, MI) by pipetting 5 μL of the standard into 495 μL of blood. The blood samples with lower concentrations of drugs were prepared serially by diluting 40 μL of higher drug content blood samples with 360 μL of blood. The concentrations of the drugs in the final blood samples were 0.01, 0.1, 1, 10, 1000, and 10,000 ng mL−1, respectively. DBS samples were prepared by spotting a fixed volume (5 μL) of blood onto the paper substrate and drying for at least 4 h at room temperature. Samples were stored at room temperature in a sealed plastic bag containing desiccant.
For paper spray, the DBS substrate was cut into a triangle (10 mm height and 5 mm base width). A copper clip was used to hold the paper triangle and to apply the high voltage needed for the spray. The distance between the tip of the paper triangle and the inlet to the mass spectrometer was about 5 mm. The silica-coated ion-exchange paper Grade SG81 (0.27 mm thick) and the chromatography paper Grade 4 (0.21 mm thick) and Grade ET31 (0.50 mm thick) were purchased from Whatman International Ltd. (Maidstone, England) and used without further chemical treatment. The Grade 4 chromatography paper is of similar thickness to the silica-coated paper while the thicker Grade ET31 is used for making the commercial blood cards (Whatman FTA DMPK-C card). The 9:1 methanol/water (v/v) solvent has been found to be generally optimal for analysis of drugs in DBS on the cellulose filter papers. The MS analysis was carried out in positive ion mode with a spray voltage at 3.5 kV. The LOQ value for each drug was defined as the lowest concentration within the set of linear responses. Three replicate measurements were made for each sample.
Images of surfaces of the chromatography paper (Whatman Grade 4, 0.21 mm thick) and the silica-coated paper (Whatman Grade SG81, 0.27 mm thick)) were recorded using a FEI NOVA nanoSEM field emission scanning electron microscope (SEM, FEI Company, Hillsboro, Oregon). The substrates were sputter-coated with platinum for 1.0 min before the analysis and the accelerating voltage for the Everhart-Thornley detector (ETD, routine imaging) or through-the-lens detector (TLD, high magnification/resolution imaging) was 5 kV with a working distance of about 5.0 mm.
A TSQ Quantum Access Max (Thermo Scientific, San Jose, CA), operated in the selected reaction monitoring (SRM) mode was used and specific product ions produced by collision-induced dissociation (CID) were monitored. The XCALIBUR software (MS control software) was used for control of the TSQ Quantum Access Max MS system and data acquisition. Argon gas (99.995% purity) was used as collision gas. The temperature MS inlet capillary was 300° C. The SRM and instrumental parameters used for the drug compounds were as follows: verapamil: m/z 455→303, tube lens 97 V, Q2 offset (collision energy) 28 V; sunitinib: m/z 399→283, tube lens 116 V, Q2 offset 28 V; citalopram: m/z 325→109, tube lens 121 V, Q2 offset 28 V; amitriptyline: m/z 278→233, tube lens 108 V, Q2 offset 17 V; lidocaine: m/z 235→86, tube lens 94 V, Q2 offset 18 V.
The Mini 11 (Gao, L.; Sugiarto, A.; Harper, J. D.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2008, 80, 7198-7205; Hou, K. Y.; Xu, W.; Xu, J. A.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2011, 83, 1857-1861; and Sokol, E.; Noll, R. J.; Cooks, R. G.; Beegle, L. W.; Kim, H. I.; Kanik, I. Int. J. Mass Spectrom. 2011, In Press, Corrected Proof) a handheld rectilinear ion trap mass spectrometer, has a discontinuous atmospheric pressure interface (DAPI; Gao, L.; Sugiarto, A.; Harper, J. D.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2008, 80, 7198-7205; and Hou, K. Y.; Xu, W.; Xu, J. A.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2011, 83, 1857-1861) and a pumping system with a 10 L/s trubomolecular pump (Pfeiffer HiPace 10, Pfeiffer Vacuum Inc., Nashua, NH) and a 5 L/min diaphragm pump (1091-N84.0-8.99, KNF Neuberger Inc., Trenton, NJ). The flow restricting capillary in the DAPI is of 5 cm length and 250 μm I.D. The DAPI was opened for 12 ms to introduce the ions generated by paper spray and the mass analysis was performed 600 ms after the DAPI was closed and the vacuum had reached an approximate level. The precursor ions were trapped at an RF voltage of 275 Vp-p at 1.0 MHz and the product ions were analyzed with a RF scan from 206 Vp-p to 4,500 Vp-p and a resonance ejection at q=0.70. The frequency and the amplitude of the excitation signal for the CID were as follows: amitriptyline, 99.40 kHz and 0.39 Vp-p; citalopram: 84.60 kHz and 0.63 Vp-p; lidocaine: 78.76 kHz and 1.22 Vp-p; sunitinib: 84.89 kHz and 0.73 Vp-p; verapamil: 80.58 kHz and 0.37 Vp-p.
c. Results
SEM images of the substrate surfaces are shown in
As discussed herein, the paper spray process involves analyte elution as well as spray ionization. The solvent properties affect not only the extraction of the analytes from the dried blood spots, but also their transfer across the paper as well as the ion formation during the spray. For the silica-coated paper, pure solvents with a range of polarities were first investigated, including water, hexane, dichloromethane, methanol, ethanol, isopropanol, butyl alcohol.
It is well-known that the electrospray process is highly dependent on the polarity and the volatility of the solvent. Typically solvent mixtures are used to optimize the spray process for analysis of target analytes (Tian, Z. X.; Kass, S. R. J. Am. Chem. Soc. 2008, 130, 10842-1084; Lavarone, A. T.; Jurchen, J. C.; Williams, E. R. J. Am. Soc. Mass Spectrom. 2000, 11, 976-985; Venter, A. R.; Kamali, A.; Jain, S.; Bairu, S. Anal. Chem. 2010, 82, 1674-1679; and Li, J. W.; Dewald, H. D.; Chen, H. Anal. Chem. 2009, 81, 9716-9722). Addition of a non- or less-polar solvent component of low boiling point helps to enhance the generation of droplets of suitable size during the spray and facilitates their subsequent desolvation (Kebarle, P.; Tang, L. Anal. Chem. 1993, 65, A972-A986). Paper spray is expected to share similar spray characteristics with electrospray but the efficiency of the analyte extraction step needs also to be considered in the selection of solvent composition. After finding isopropanol as the best pure solvent for paper spray with silica-coated substrates, dichloromethane was mixed with isopropanol to produce solvents with increased volatility. The direct sampling ionization of the verapamil in the dried blood spots was performed using a set of isopropanol/dichloromethane solvents in different ratios, as shown in
The elution efficiency was also characterized for the three solvent/substrate systems, 9:1 dichloromethane/isopropanol (v/v) for silica-coated paper and 9:1 methanol/water (v/v) for Grade 4 and Grade ET31 chromatography papers, using both pure verapamil sample spots and dried blood spots containing verapamil. The pure verapamil sample spots were prepared by dropping 5 μL water solution containing verapamil (500 ng mL−1) onto the paper and drying the substrate completely. The voltage was applied and solvent added multiple times to produce many paper spray events using the same substrate bearing a single sample spot while the signal of fragment ion m/z 303 was monitored. Spray solvent of 25 μL was used each time and the solvent was not added until the monitored ion signal decreased to a minimum, when the spray solvent was also exhausted. Solvent consumption took 5-8 s for dichloromethane/isopropanol on silica-coated paper but 40-70 s for methanol/water on chromatography paper substrates. For both pure analytes and blood samples, much higher peak intensities were was observed for the silica-coated paper substrate (
Similar procedure was applied to dried blood spots containing verapamil. The blood spots were each prepared with 5 μL of blood sample containing 500 ng mL−1 verapamil. The matrix in the dried blood spots is much more complex than that in the pure verapamil sample spots. For silica-coated paper with 9:1 dichloromethane/isopropanol, analyte release during the first paper spray event was still dominant, although the peak intensity for verapamil decreased about 3 times, presumably due to the matrix effect. For chromatography papers with methanol/water, no significant difference in elution pattern was observed except that the maxima of the verapamil detected appeared at later elutions (
Without be limited by any particular theory or mechanism of action, several factors could account for the difference between these two substrate/solvent systems and the improvement with the silica-coated paper substrate. The blocking of the pores in the cellulosic framework by the silica resulted in a more concentrated sample on the top surface of the substrate and also a less binding interaction between the analyte and the cellulose. This helps to improve analyte elution during the paper spray process. The polarity of the solvent systems might also play an important role. The polarity of methanol/water solvent (9:1, v/v) is higher than that of dichloromethane/isopropanol solvent (9:1, v/v) (polarity index: water 10.2, methanol 5.1, isopropanol 3.9 and dichloromethane 3.1). Low polarity organic compounds, such as verapamil, dissolve better in dichloromethane/isopropanol solvent, which makes the extraction of these chemicals from dried blood more efficient.
The performance of the improved substrate/solvent system was characterized for quantitative analysis of therapeutic drugs in whole blood samples using paper spray. Dried blood spots on paper spray substrates were prepared by depositing 5 μL whole blood containing verapamil at a concentration from 0.01 ng mL−1 to 10,000 ng mL−1 and drying the sample on substrate completely. MS analysis was performed using the triple quadrupole mass spectrometer and the peak intensity of m/z 303 from verapamil ((M+H)+, m/z 455) was recorded. As shown in
In this Example, paper spray using the silica paper substrates was also characterized using a miniature rectilinear ion trap mass spectrometer, Mini 11. This was done by quantitative analysis of verapamil, sunitinib, citalopram, amitriptyline, and lidocaine in dried blood spots. MS/MS analysis was performed using the procedure previously described (Gao, L.; Sugiarto, A.; Harper, J. D.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2008, 80, 7198-7205; Hou, K. Y.; Xu, W.; Xu, J. A.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2011, 83, 1857-1861; and Sokol, E.; Noll, R. J.; Cooks, R. G.; Beegle, L. W.; Kim, H. I.; Kanik, I. Int. J. Mass Spectrom. 2011, In Press, Corrected Proof) to acquire the intensities of the characteristic fragment ions from the drug compounds.
A comparison study was done for silica-coated paper and chromatography paper (Whatman Grade ET31, 0.50 mm thick) using paper spray and the Mini 11 for analysis of therapeutic drugs in dried blood spots. The LOQs obtained and the therapeutic windows for lidocaine, amitriptyline, sunitinib, verapamil, and citalopram are listed in Table 6.
aLOQ for each drug is defined as the lowest concentration within the set of linear responses in the sensitivity plot as shown in FIGS. 58 and 59A-59B. Three replicates for each sample.
Using the chromatography paper with 9:1 methanol/water, the performance of the paper spray/Mini 11 system was only good enough to cover the therapeutic range for lidocaine, but not for the other 5 drugs in DBSs. With the silica paper substrate and 9:1 dicholoramethane/isopropanol, improvements of 5-50 fold were obtained in LOQs for these drugs, which made the performance of the miniature systems adequate for their quantitative analysis from dried blood spots.
d. Conclusions
Paper spray with a silica coated paper substrate was characterized using a commercial triple quadrupole and a home-built miniature ion trap mass spectrometer for analysis of therapeutic drugs in dried blood spots. The overall analysis efficiency can be greatly improved by using low boiling and low polarity solvent, 9:1 dichloromethane/isopropanol (v/v). The LOQ of analysis of a set of therapeutic drugs in dried blood spots, including verapamil, citalopram, amitriptyline, lidocaine, and sunitinib, was obtained as low as 0.1 ng mL−1 with the commercial triple quadrupole and 10-20 ng mL−1 with a miniature ion trap mass spectrometer. The triple quadrupole is inherently better suited to SRM analysis than is an ion trap instrument so this factor as well as the reduced performance of the Mini MS is responsible for the difference. Compared to chromatography paper, in each case there is a 5-50 fold improvement with the silica paper substrate.
The present application is a continuation of U.S. nonprovisional application Ser. No. 17/395,909, filed Aug. 6, 2021, which is a continuation of U.S. nonprovisional application Ser. No. 16/905,193, filed Jun. 18, 2020, which is a continuation of U.S. nonprovisional application Ser. No. 16/107,296, filed Aug. 21, 2018, which is a continuation of U.S. nonprovisional application Ser. No. 15/710,072, filed Sep. 20, 2017, which is a continuation of U.S. nonprovisional application Ser. No. 15/359,031, filed Nov. 22, 2016, which is a is a continuation of U.S. nonprovisional application Ser. No. 14/987,154, filed Jan. 4, 2016, which is a is a continuation of U.S. nonprovisional application Ser. No. 14/512,579, filed Oct. 13, 2014, which is a continuation of U.S. nonprovisional application Ser. No. 14/119,548, filed Nov. 22, 2013, which is a 35 U.S.C. § 371 national phase application of PCT application number PCT/US12/40521, filed Jun. 1, 2012, which claims the benefit of and priority to U.S. provisional patent application Ser. No. 61/492,933, filed Jun. 3, 2011, U.S. provisional patent application Ser. No. 61/492,937, filed Jun. 3, 2011, and U.S. provisional patent application Ser. No. 61/492,947, filed Jun. 3, 2011, the content of each of which is incorporated by reference herein in its entirety.
This invention was made with government support under RR031246 and EB009459 awarded by the National Institutes of Health. The government has certain rights in the invention.
Number | Date | Country | |
---|---|---|---|
20230341360 A1 | Oct 2023 | US |
Number | Date | Country | |
---|---|---|---|
61492933 | Jun 2011 | US | |
61492947 | Jun 2011 | US | |
61492937 | Jun 2011 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 17395909 | Aug 2021 | US |
Child | 18128584 | US | |
Parent | 16905193 | Jun 2020 | US |
Child | 17395909 | US | |
Parent | 16107296 | Aug 2018 | US |
Child | 16905193 | US | |
Parent | 15710072 | Sep 2017 | US |
Child | 16107296 | US | |
Parent | 15359031 | Nov 2016 | US |
Child | 15710072 | US | |
Parent | 14987154 | Jan 2016 | US |
Child | 15359031 | US | |
Parent | 14512579 | Oct 2014 | US |
Child | 14987154 | US | |
Parent | 14119548 | US | |
Child | 14512579 | US |