This invention relates to ion mass analytical technology using ion trap, and more particularly relates to a linear ion trap analyzer with electric field optimization.
The technology for traditional quadrupole ion trap was greatly developed after its invention in the 1950's, and was applied in a wide variety of mass spectrometer instrument system. Many articles and patents related to this field were collected in the book “Practical Aspects of Ion Trap Mass Spectrometry” written by R. E. March and J. F. J. Todd. Usually, the three-dimensional ion trap (3D-IT) with rotary symmetry comprises a trapping volume for mass analyzing surrounded by a ring electrode and an opposing pair of end-cap electrodes. RF voltage is applied on the ring electrode to form a substantial quadrupole field to confine the ions, and a dipole AC voltage is applied between the opposing pair of end-caps to excite the ions motion and ejected out mass-selectively, to achieve mass scan of the ion trap.
The two-dimensional linear ion trap (2D-LIT) mass spectrometry instruments have been widely used because of their high sensitivity and storage capacity after commercialization. There are many designs for 2D-LIT. Commonly, as shown in
Over years, many scientists made effort on improving the performance of ion trap in mass scan through optimizing the trapping field. For example, to overcome the effects of negative fringe field around the ejection hole during resonant ejection in 3D-IT, Kawato et al. introduced embossment flanges on the round edge of the ejection hole in U.S. Pat. No. 6,087,658. For the same problem, in U.S. Pat. No. 6,911,651, Senko et al. stretched the distance between the end-caps and made concentric recess around the outlet hole.
Above all, field improvement in ion trap by amendment on electrodes highly depends on the mechanical accuracy. Once the modified electrode is formed, the amendment of field is fixed and optimized for certain analytical condition. If the working cycle of ion trap contains more than one stage and needs different field optimization conditions, these methods may not be useful.
The designer produces a kind of 3D-IT with more than one circular electrodes in U.S. Pat. No. 5,468,958. These electrodes are applied with RF voltage of different ratios. The electric field can be adjusted by changing the ratios. An amended electrode is embedded in the end-cap electrode to introduce a field component which can be adjusted by voltages to optimize trapping field in a small range (in U.S. Pat. No. 7,279,681, L I Gangqiang et al). While in U.S. Pat. No. 6,608,303 by Amy et al., a thin metal electrode on which a RF potential with particular phase was applied, was embedded in the ejection hole to optimize field around.
The design and accuracy are simplified. The field inside can be adjusted through outside. and these technologies are used on linear ion trap gradually. In CN1585081, Chuanfan Ding designed a kind of linear ion trap surrounded by printed circuit boards. As using a lot of individual adjustable electrodes, flexible field adjustment, as well as larger ion capacity and lower cost are achieved.
But in all the above technologies, all the electrodes invoked to correct electric field depend on high frequency power supply which can accurately control voltages that are applied on these electrodes. This high frequency power supply could be a usual RF-resonant high frequency power supply or alternatively a high frequency switch power supply used by digital ion trap. Anyway, the instruments become complicated with the additional power supply.
A field adjusting electrode is placed behind the injection hole of one end cap in 3D ion trap and is driven by a DC voltage to affect respectively ion motions during injection and ejection in U.S. Pat. No. 7,285,773 by Dingli. Although this kind of local corresponding correction hasn't fully improved high frequency field components, yet as for ions which are excited, motion characteristics have been greatly improved. Since field adjusting electrode only needs to apply with a DC voltage rather than a high frequency voltage, instruments could be simplified and adjustment could be easy. But this patent is not for linear ion trap shown in
Besides, ions could eject from the two through slots on X electrodes after resonant excitation in linear ion trap, so two detectors need to place behind X electrodes to obtain maximum signal which may increase cost.
One of the purposes of this invention is to design a proper field adjusting electrode and its corresponding power supply, optimize the electric field inside the linear ion trap and ions motion characteristics as well as ions ejection from one through slot as many as possible.
An aspect of the invention provides a linear ion trap analyzer, comprising a ion trapping volume multiple surrounded by columnar electrodes, whereas, the generatrix of said columnar electrodes are parallel to the central axis of the trapping volume, at least a part of said columnar electrodes is applied with high frequency voltage to form in said trapping volume the trapping electric field which is dominated by two dimensional quadrupole field. At least one through slot for ion ejection orientated in one direction perpendicular to the said axis, wherein AC electric field superposition is applied to invoke dipole excitation in said one direction; In this invention, an elongated field adjusting electrode is set inside the slot on one columnar electrode opposite to the through slot or between the two columnar electrodes, wherein the potential on the field adjusting electrode is set as the sum of a portion of the high frequency voltage applied to one adjacent columnar electrode and a DC voltage offset, which is adjustable. Through adjusting the geometry or the location of the elongated electrode or the potential on it, one or more objectives, including field optimization inside the ion trap as well as ion motion characteristics of resonant ejection, can be realized.
According to one embodiment, a linear ion trap analyzer mentioned above may further include an electric circuit for applying voltages on the said field adjusting electrode comprising a capacitor for coupling the high frequency voltage to the said field adjusting electrode from the said adjacent columnar electrode, and a resistor and/or an inductor for applying a DC voltage superposition on the said high frequency voltage, and the DC voltage is controlled by a DC voltage source.
According to one embodiment, a linear ion trap analyzer mentioned above may further include an electric circuit for applying voltages to the said field adjusting electrode comprising a capacitor for coupling the high frequency voltage to the said field adjusting electrode from the said adjacent columnar electrode, and the diodes for applying a DC voltage superposition on said high frequency voltage, wherein the DC voltage controlled by a DC voltage power supply and the DC amplitude of the said power supply substantially equals to the sum of the required DC voltage offset of field adjusting electrode and the positive or negative peak value of the said high frequency voltage.
According to one embodiment, at least part of the columnar surface of the said columnar electrodes is hyperbolic columnar surface.
According to one embodiment, at least part of the columnar surface of the said columnar electrodes is planar columnar surface.
According to one embodiment, at least part of the columnar surface of the said columnar electrodes is step shaped columnar surface.
According to one embodiment, at least part of the columnar surface of the said columnar electrodes includes the planar patterns of printed circuits on the surface.
According to one embodiment, the said field adjusting electrodes comprise of single or multiple sections of segmented electrodes
According to one embodiment, the said high frequency voltages are generated by digital switches and in rectangular waveforms.
According to one embodiment, the strength or the frequency of the trapping electric field is scanned while the said AC electric field superposition is applied to invoke dipole excitation in the direction perpendicular to the said axis and to invoke the ions trapped inside the linear ion trap to eject out resonantly according to their mass-to-charge ratios; and also includes controlling means to alter the DC voltage applied on the said field adjusting electrode when the said scan is reversed or the scan speed is changed.
According to one embodiment, the said value of DC voltage applied on the said field adjusting electrode is adjusted to improve the ejection efficiency through the outlet slot opposite to the field adjusting electrode during the scan where ions eject out resonantly according to their mass-to-charge ratios.
According to one embodiment, the value of DC voltage applied on the said field adjusting electrode is set to generate a DC high order field during the said AC electric field is applied to invoke dipole excitation of at least one ion, wherein the DC high order field alters the secular frequency of the said at least one ion from the frequency of the said AC electric field and break their resonance when the amplitude of the ion motion is close to the field radius of the linear ion trap, so that the ion avoid being further excited.
The electrode field can be adjusted according to the need of real working mode through field adjusting electrode in linear ion trap. It has a great influence on ion kinetic character while resonance ejection. And part of positive ions which could eject from left side could be reflected by field adjusting electrode as long as the DC voltage is high enough applied on field adjusting electrode. Thus more ions eject from the outlet slot of right X electrode to increase the outlet efficiency of single side.
To make the purposes, characteristics and advantages mentioned above in this invention more obvious and easier to understand, the following is the embodiments of this invention combined with figures demonstrated in detail, including:
The linear ion trap related should be demonstrated before further description to this invention.
A linear ion trap was usually described as space either surrounded by a set of poles or defined by several electrodes extended along axial direction. In order to involve the substance of linear ion trap, columnar electrodes are used in stead of poles or electrodes extended along axis. The so called columnar surface is defined as such curved surface formed by straight lines parallel to a fixed line (here defined as Z axis) and moving along a directrix. These moving straight lines are called generatrix of columnar surface. Multiple columnar electrodes, the linear ion trap formation, are not necessarily columns but have columnar surface and their generatrix are parallel to each other as well as a central axis (z axis), which is coupled clearly with the statement of electrodes extending along the axial direction. Also, the columnar surfaces are not necessarily very long, so the linear ion trap is not necessarily elongated. Moreover, planar is also involved as a special case (That is, directrix is a straight line or a polyline.). In other words, for several planar electrode surfaces, as long as they are placed parallel to z axis, and space surrounded by those surfaces can be formed to trap ions under proper situation, are also involved in the discussion about the electrode geometry in this invention.
Again,
In this embodiment, an elongated field adjusting electrode 5 is placed in the middle of X electrode 1a oppositely faced to the ejection slot. Voltage on this electrode is set to the sum of at least a portion of the high frequency voltage V1a applied to on nearby X electrode 1a and a DC voltage VDC, that is:
Vfae=cV1aVDC0<c≦1
in which, the high frequency voltage V1a includes the original high frequency quadrupole driving voltage and the dipole excitation AC voltage. Before resonant ejection, the amplitude of ions motion gradually becomes larger and larger and negative high order field will reduce secular frequency when ions move close to the ejection slot. For example, positive ion will oscillate near to the field adjusting electrode 5 when the high frequency quadrupole voltages on X electrode 1a and 1b turn positive. If VDC is made positive, positive ions will obtain extra reversing force, so that secular frequency reduction can be avoided. This helps ion ejection quickly.
Moreover, when VDC is properly adjusted, the positive voltage can make more positive ions eject from columnar electrode 1b, increasing single-side ejection efficiency of ions. This will save a detector, comparing with both-side ejection.
As mentioned above, planar electrodes, as a special case of columnar electrodes, can also be used to comprise linear ion trap.
Vfae=cV1a+VDC0<c≦1
It should be pointed out that the back shape of the field adjusting electrode 15 (apart from trapped ions) was designed just to make the mechanical assembling easy. This embodiment does not limit its specific shape.
Every columnar electrode contains only one planar surface parallel to axis in this embodiment and the electric field is quite different from two-dimensional quadrupole electric field, which may not be ideal enough to influence ions motion characteristics only through the field adjusting electrode adjustment. If multiple planar surfaces are used to form step shaped columnar surfaces or ones whose generatrix is polyline, a more similar electric field will be formed as that formed by hyperbolic columnar surfaces. This kind of design has been opened in CN1925102A. A field adjusting electrode can also be set in the middle of the electrode opposite to an outlet slot in this ion trap and be applied with voltage equivalent to composition of at least part of high frequency voltage and a DC voltage.
In this embodiment, in order to obtain a good quadrupole electric field inside the rectangular linear ion trap built by planar electrodes, each electrode surface can be composed of several sub-electrodes, on which high frequency voltage with certain proportion is applied separately to form a similar electric field with that formed by hyperbolic columnar electrodes. The details of these ion traps can be found in Chinese publication No. CN1585081.
Vfae=cV1a+VDC0<c≦1
Using the said field adjusting electrode 25, harmful effects caused by outlet slot on ion motion can be further overcome, increasing single-side ion ejection efficiency.
There are lots of methods/devices/circuits used to superimpose high frequency voltage and DC voltage applied on different kinds of field adjusting electrodes mentioned above. Two examples are shown as follows.
Generally, the ratio of peak values of VDC and V1a should be 0 to 5% if field adjusting electrode is basically even with the adjacent columnar electrode on one side of trapping volume (shown as
The disadvantage of this option is that the resistance must be large enough, generally several mega or several tens of mega ohms Otherwise, the RF power supply will be affected and the RF voltage applied on field adjusting electrode 5, 15, 25 will be insufficient. However, DC voltage component applied on the field adjusting electrode could not be set up or adjusted quickly if the coupled resistance is much too large.
In order to solve this conflict, the option is brought forward in another embodiment of this invention, which superimposes high frequency voltage component of the adjacent columnar electrodes obtained by coupling capacitor and DC voltage component through a diode.
According to
Using this method mentioned above, a DC voltage could be superposed with a high frequency voltage. By changing the value of V1, adjustment of amplitude of DC voltage superimposed will be realized. Resistor 34B of several kilo-ohms to several hundreds of kilo-ohms plays a role of current limit, which would satisfy with the need of DC voltage set-ups on the field adjusting electrode.
When providing positive DC for field adjusting electrode, the output voltage V1A of positive DC supply 32A is higher than V1a(0−p)+VDC (that is, V1+2V1a(0−p)), thus diode 35A is reversed and out of work. When providing the needed negative DC component to field adjusting electrode, positive electric supply 32A is connected to the field adjusting electrode through resistor 34A and diode 35A. Diode 35A will be forward for a while if the output V1A of DC power supply 32A is lower than the positive peak value V1a(0−p) of high frequency voltage V1a. Capacitor 33 will be charged or discharged by power supply 32A through resistor 34A and diode 35A. After several cycles, the maximum peak value of output Vfae will be decreased to the level of V1, which equals to a DC level superposition on high frequency voltage VDC=V1−V1a(0−p). Diode 35B will be reversed and negative power supply is out of work as long as the output V1 of negative power supply 32B is lower than VDC−V1a(0−p) (that is, V1−2V1a(0−p)).
In a word, whether positive or negative DC voltage is superimposed, the DC component supplied by DC power supply is equal to the sum of the DC voltage needed and the peak value of high frequency voltage (positive or negative phase).
When the driving voltage of ion trap is digital square waveform, the diode coupling option can be described by
The output of DC high voltage supply 42 is +V and the out put of DC high voltage supply 41 is −V. The high frequency square waveform is generated by switch 44 and 45. The switch 44 and 45 can be on and off in turn controlled by an outside controller so that square waveform with peak value of V can be generated.
When switch 44 is on and 45 is off, diode 35 is forward and capacitor 33 is charged by DC power supply 32 through resistor 34 and diode 35. The output equals to V1. When switch 44 is off and 45 is on, diode 35 is reversed and the amplitude level of output equals to ((+V) +V1−(−V)).
The method mentioned above can realize to superimpose DC voltage to the high frequency square waveform, wherein the amplitude of the DC voltage superimposed equals to V1− (−V) and the amplitude can be adjusted by changing the value of V1.
The diodes 35 or 35A and 35B used in the circuit mentioned above should have high reverse breakdown voltage, low junction capacity, large positive peak current and quick reverse recovery capability. The diode in the embodiment can be replaced by using serial multiple diodes.
With the help of the field adjusting electrode, field components in the linear ion trap can be adjusted according to the need of real working mode, which can help improve ions motion characteristics obviously during the resonant ejection.
If DC voltage applied on field adjusting electrode is set higher, such as 80 V (dotted line c) shown in the figure, the resonant frequency will increase rather than decrease when ions amplitude reaches around 3 mm. Ions may get fully resonant with the dipole electric field when they move around 3.5 mm under forward scan and they are excited fast and eject from outlet slot, which would cause spectrum with high resolution.
For the DC voltage on field adjusting electrode is adjusted to a higher value (for example a proper one higher than 0V), part of positive ions which may eject from the field-adjusting-electrode side can be reflected back by the said field adjusting electrode and thus more positive ions can eject from the opposite side through the outlet slot on X electrode. In other words, ions prefer to eject from the outlet slot which increase ions single-side ejection efficiency. The said proper DC voltage can be obtained by practical measurement although the value of said DC voltage may be different in specific applications.
On the contrary, since dipole excitation frequency is lower than ions motion frequency, when reverse scan is carried out, lower DC voltage on the field adjusting electrode (for instance, dotted line b in the figure) can help ions eject and obtain higher resolution. With the help of field adjusting electrode, proper voltages can be chosen according to different scan modes and scan speeds so that optimization under proper situation could be realized. Since combination of forward and reverse scans can be used in precursor ions selectivity, precursor spectrum with high resolution can also be realized through DC voltage optimization on field adjusting electrode under proper situation.
Using field adjusting electrode, it could be obtained not only to optimize process of ions scans and ejection as well as mass-selectively isolation, but also to improve effects of excited precursor collision induced dissociation. For example, the DC voltage of 0V or 120V in
It is only part of the functions that influence ions motion using field adjusting electrode. In fact, it can be developed by anyone who is familiar with ion trap working principles. Besides, in the embodiment only one field adjusting electrode is placed along the field axis, which could be replaced by multiple field adjusting electrode segments to adjust fringe field components separately. The location of field adjusting electrode can be either in the slit on the electrode opposite to the outlet slot or aperture or between the pair of electrode in ejection direction. The top of electrode can be even with surrounding X electrode, or put deeply inside the slit, only that the electric field generated can infiltrate and influence the field inside the ion trap. The field adjusting electrode is not necessarily completely straight. It could have gurgitations, gradient, being curved to correct field ununiformity along the axial direction of the ion trap. All these changes can be easily achieved by people with skill in the same field using knowledge from this invention, which should be covered by this invention.
Number | Name | Date | Kind |
---|---|---|---|
5468958 | Franzen et al. | Nov 1995 | A |
6087658 | Kawato | Jul 2000 | A |
6608303 | Amy et al. | Aug 2003 | B2 |
6909089 | Londry et al. | Jun 2005 | B2 |
6911651 | Senko et al. | Jun 2005 | B2 |
7279681 | Li et al. | Oct 2007 | B2 |
7285773 | Ding et al. | Oct 2007 | B2 |
20090127456 | Makarov et al. | May 2009 | A1 |
Number | Date | Country |
---|---|---|
1585081 | Feb 2005 | CN |
1925102 | Mar 2007 | CN |
Number | Date | Country | |
---|---|---|---|
20120248307 A1 | Oct 2012 | US |