Transmission of blood-borne infections (e.g. Hep B, Hep C, & HIV) from health care exposures was previously believed to be uncommon. However, multiple outbreaks across the United States over the last several years have shed light on this problem. In a recent article (Ann Intern Med 2009; 150(1):33-39, which is incorporated herein by specific reference in its entirety) reviewing outbreaks of nonhospital health care-associated Hep B & Hep C virus transmission in the United States from 1998-2008. 33 known outbreaks were identified. These events have resulted in 448 cases of Hep B and Hep C and >60,000 individuals exposed to potential blood-borne pathogens. As recently as October 2009, Broward General Medical Center (Florida) sent letters to >1800 patients who were potentially exposed to contaminated fluids. In this instance, the nurse routinely used the same bag of saline on multiple patients.
One such large outbreak occurred in an outpatient oncology clinic in Nebraska from 2000-2001. In this instance, the nurse would reuse a syringe to perform a saline flush. Saline for multiple patients was acquired from a common bag. More than 600 patients were notified. 99 patients were diagnosed with Hep C. These were all breaches of standard practice and should have never occurred. In many instances, veteran health care personnel were responsible for such poor practices. These cases all have in common acquisition of saline or a drug from a common container via a contaminated needle and/or syringe. Despite the increasing use of single-dose vials, such outbreaks occur as providers still use these single-dose vials on multiple individuals to contain costs.
Presently, syringes tipped with needles or “needleless syringes” allow withdrawal of a liquid composition (e.g., saline or medication) from a reservoir. Use of a needleless device reduces the chance of an accidental stick by the operator. However, neither one of these systems prevents re-use of a contaminated syringe, needle or needleless device when acquiring more medication from the same or different reservoir. In addition, when a reservoir is accessed by a needle or a needleless device, the liquid composition is allowed to flow freely. Essentially, not only can sterile liquid flow from the reservoir to the syringe, but the contaminated syringe, needle or needless device can contaminate the fluid, and the contaminated fluid can move up the line and contaminate the reservoir and its liquid contents. Most of the outbreaks related to liquid reservoir contamination have occurred in the outpatient setting (e.g. clinics, surgery centers, hemodialysis units). As such, the present system and methods of use can be utilized in these settings.
While the present invention is usable with respect to syringes, needles, needless devices, or any other medical device configured to withdraw liquids from a sterile reservoir, the description herein will refer to syringes, however, such reference to syringes should be construed to include reference to any medical device configured to withdraw liquids from a sterile reservoir.
Accordingly, a system has been designed to overcome the foregoing problems of contamination. As such, the system is configured so that only one access event occurs for a sterile reservoir with a sterile syringe attached to a sterile syringe adapter. This adapter is configured to be received onto the syringe in a manner such that it must be removed before a needle or needleless system can be placed on the end of the syringe and the medication is injected into the patient. Additionally, the adapter was configured so that is can only be placed on a syringe one time. As such, once the adapter is removed from a syringe, it cannot be placed back on the syringe. Also, the system can include a back-up safety feature that is configured to prevent the adapter from allowing liquid to flow therethrough once it has been used one time. This prevents liquid from flowing from a sterile reservoir through the device once the device has already accessed a reservoir and had fluid flow therethrough.
When a syringe is placed to receive liquid from a reservoir, negative pressure or a “sucking” effect allows for the liquid to flow through the syringe. Accordingly, the adapter can include an access valve that only operates one way. After removal and use of the syringe, the access valve remains closed and does not allow for fluid to pass through the same adapter connection. The adapter and access valve is described in more detail below. At steady state and with positive pressure from the syringe, the access valve remains closed, not allowing any fluid to flow into or out of the reservoir, thus preventing accidental contamination of the reservoir.
The adapter as described herein can be removably couplable to a syringe. Also, the adapter can be removably couplable to a solution reservoir. Additionally, the adapter can be integrated to either the syringe or solution reservoir.
Development of a vacuum within a rigid solution reservoir, such as a glass bottle, occurs during liquid extraction. Once liquid (i.e. medication) is removed from a reservoir, a negative internal pressure results. This is not a problem with IV bags as the bag simply collapses under atmospheric pressure, normalizing pressures across the PVC membrane. In a rigid medication reservoir, such as a glass bottle, this problem is presently overcome by injecting air via a delivery assist device into the bottle through a valve prior to drug removal with a syringe. This issue of negative pressure can be addressed by placement of a millipore filter within a one-way valve and incorporating the filter and one-way valve into a delivery assist device such that air can enter into the rigid reservoir as fluid is extracted, thus preventing a vacuum effect. This one-way air valve and filter combination can be incorporated with delivery assist devices used only on rigid reservoirs, like a glass bottle.
The present invention relates to a safety-engineered, one-time use, syringe adaptor-connector system that restricts access to medical vials and intravenous bags (common reservoirs). Within the mechanism a delivery control valve (connector) is placed on the common reservoir, serving to limit the syringe type that can be used to access the liquid content within the reservoir. The syringe includes an adaptor, which is designed to limit access to a common reservoir to one-time use only. The system impedes the ability to access a common reservoir with a contaminated syringe.
The adoption of this safety-promoting device into medical practices provides an added measure of protection and safety that is not limited by human error, misunderstanding of safe syringe practice, or lack of education. This “lock and key” method is the most effective way in which to prevent the spread of infection through contaminated syringes.
In one embodiment, the present invention include a two-part system that has: 1) a syringe adapter that is attached to the functional tip of a syringe, where this component is attached at the time the syringe is manufactured or can be an independent safety device that can be coupled to the syringe prior to use; 2) a delivery assist device which is incorporated into the medication reservoir (e.g. IV bag, glass vial, etc.). The syringe adapter can be integrated with a syringe, or it can be removably couplable to the syringe. Once removed it cannot be reused. The adapter can be prepackaged to be coupled to a syringe, and then the adapter can be removed from the syringe so that the syringe can receive a needle or the like. The delivery assist device can be integrated with a liquid reservoir (e.g., container), or it can be removably coupled to the liquid reservoir.
The components of the system are configured such that in order to access liquid from a bottle or bag that has the delivery assist device in place, a syringe adaptor must be utilized. This system would not allow an individual with a standard syringe & needle to access the medication. The syringe adapter can be placed on the syringe at the time of manufacturing and subsequently sterilized. With this adapter, the liquid can be removed from its reservoir when the reservoir has the delivery assist device. Once the liquid is in the syringe, the syringe adaptor will not allow access of an IV or other tubes contaminated with body fluids. The syringe adapter must be removed from the syringe before a needle can be placed on the syringe such that patient IV access can occur. Once the adapter is removed, the design will not allow it to be placed back on a syringe. Before additional liquid can be accessed from the reservoir again, a new sterile syringe and syringe adapter must be used. This two part “lock and key” design (e.g., 1 component built into the liquid reservoir and 1 component attached to the syringe) prevents access of a liquid from a common reservoir via a contaminated syringe. A functional “lock” (delivery assist device) and a “key” (syringe adapter) that does not allow someone to access that medication reservoir with a non-sterile syringe can be important to stop the reservoir from being contaminated. Once used, the syringe adapter is rendered useless with the delivery assist device.
In one embodiment, as an added safety feature the delivery assist device, located on the liquid reservoir, can be designed such that any positive pressure from a syringe will not allow flow of contaminated material back into the common reservoir. An example can include a one-way valve in the delivery assist device, but other safety features can be used. This design can help prevent potential “work-arounds”, i.e. trying to find a way to use a standard syringe to access the liquid through this delivery device.
In one embodiment, the delivery assist device and/or syringe adapter can employ electronics or a mechanical system that only allows the delivery assist device to open only for a new and sterile syringe adapter. For example, the electronics or a mechanical system can be configured to only open a valve or otherwise allow access to the reservoir with a sterile syringe adapter (e.g., unused).
For example, there may be a small chip on the delivery assist device that recognizes if a given adapter has already been allowed access to this delivery assist device or to another delivery assist device. If this is the case, the electronics or mechanical system won't allow flow of the fluid from the reservoir through the deliver assist device (or valve therein). Alternatively, the electronics or mechanical system do not allow full coupling of the adapter to the delivery assist device, thus not allowing one to create the negative pressure needed to extract the fluid from the reservoir. As an example, in the case of the 2-way valve mechanism (e.g.,
The connector includes a one-way valve, to allow for uni-directional flow of the fluid, and an internal pressure relief device, used to relieve a negative pressure within the medication vial. This allows the fluid to transfer from the vial to the syringe. Below are the possible choices for one-way valves and pressure relief devices:
To limit the syringe to one-time use only, the adaptor itself can be locked out or otherwise become inoperable for connecting or otherwise being usable as an adapter to transfer fluid from a liquid reservoir. This can occur via several different mechanisms. In one embodiment, the adaptor can be deformed or mechanically altered when separated from the reservoir, or alternatively can be physically broken apart.
The adaptor can have a unibody design, which can include engineered perforations; enabling the adaptor to either separate into individual parts or plastically deform upon removal from the connector. Once the adaptor portion of the syringe has been dismembered it cannot be reassembled to build a functioning adaptor again, preventing re-use of the syringe.
While many of the drawings represent as a cylindrical shape, the shape of the components is not limited to a cylinder, and could be produced in many forms including a square, rectangle, oval, star, hexagon, or pentagon etc.
Examples of suitable materials that can be used to make the syringe adapter include, but are not limited to, silicone elastomers, which can be used for flexible components such as membranes and flexible body. Stainless steel can be used in fluid pathway components. Polyester can be used in external housing of adaptor or connector. Polyethylene can be used for fluid pathways within the device. Polycarbonate can be used in external housing where clarity, strength, and tight connections are important. Polypropylene can be used in internal or external housing of adaptor and connector. Alternate Design Specifications:
In the first design, one way flow is required. This provides a secondary safety mechanism, preventing backflow in addition to preventing syringe reuse. However, one could use only one of the safety mechanisms in the design. Therefore an alternate design is one that prevents syringe reuse but allows for flow in both directions. Such a design would not require a pressure equalization method.
In this embodiment, the system can allow flow in both directions (into and out of the syringe) but will be open only during connection of the connector to the adaptor. The system cannot be open with a needle, instead the system will contain an intrinsic key in which to open the system. In this embodiment, the system should be made inoperable upon disconnect to prevent reuse. The system should not function when a needle is used to tamper with the system.
Another method of using a “lock and key” mechanism would be to place a duckbill (or other) valve in the connector which is manually open the valve upon insertion of the adaptor, allowing flow in either direction. This would occur with a “key” located on the adaptor that would manually push the gates of the valve open. The key would be of a shape other than a needle in order to prevent tampering of the device. This will eliminate the need for a pressure relief device, allowing one to physically push air in and pull fluid out once the syringe adaptor has been connected to the connector.
This application claims the benefit of and priority to U.S. Provisional Application No. 61/406,343, filed Oct. 25, 2010, which is incorporated herein by reference in its entirety.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/US11/57768 | 10/25/2011 | WO | 00 | 4/25/2013 |
Number | Date | Country | |
---|---|---|---|
61406343 | Oct 2010 | US |