This invention relates to methods of calibrating a mass spectrometer which involve experimentally determining the mass to charge ratios of a plurality of chemical compounds in a reference standard using a mass spectrometer at first and second scan speeds; interpolating mass to charge ratios for each of the chemical compounds at a scan speed different from the first and second scan speeds; and constructing a calibration curve using the interpolated mass to charge ratios. The interpolating step may be carried out using probabilistic methods.
Mass spectrometers require regular mass calibration and this is almost always done by constructing a calibration curve based on the mass spectrum of a known reference standard which typically includes a plurality of chemical compounds having different masses. Mass calibration is required for many reasons, including, but not limited to, changes in performance caused by subtle changes in the electronics of a spectrometer or conditions—such as the temperature, humidity or cleanliness—in the room in which the spectrometer is located. Therefore constant monitoring is needed to ensure consistent high mass accuracy.
GB2405991 describes a known method of calibrating a mass spectrometer which involves experimentally determining the mass to charge ratios of a plurality of chemical compounds in a reference standard. FIG. 1 of GB2405991 depicts a calibration curve which shows the experimentally determined mass to charge ratio (Measured y) as a function of the expected mass to charge ratio (True x) for four chemical compounds in a reference standard. FIG. 2 of GB2405991 depicts the same data but this time shows the offset between the measured value and the true value as a function of the true value. Using either of these calibration curves it is possible to determine the true value for any measured value obtained using the mass spectrometer on which the calibration was performed. The calibration curve may be constructed using probabilistic methods, for example as described in GB2405991, or by using more conventional methods, such as least squares regression. In experiments, true values can then be obtained from the calibration curve for any given measured value.
It has been found that the mass calibration of a mass spectrometer is heavily dependent on the scan speed, i.e. the time the analyser spends scanning for ions. Scan speeds are typically quoted in Da/e/s, where: “Da” represents a unit of mass; “e” represents the elementary charge; and “s” represents the time in seconds.
Dependence on scan speed is particularly pertinent in cases where the mass spectrometer has a quadrupole or ion trap mass analyser.
In order to illustrate the dependence on scan speed reference is made to the following figures, in which:
The calibration curves show the offset between the measured mass to charge ratio and the true mass to charge ratio (y-axis) as a function of the true mass to charge ratio (x-axis). This format is the same as illustrated by FIG. 2 of GB2405991.
It is clear from
A problem with known mass calibration methods is that they do not account for scan speed. As highlighted, ignoring scan speed can cause errors in obtaining true values from a calibration curve where an experiment is performed using a scan speed that differs from the scan speed that was used to perform the calibration.
It is, therefore, a non-exclusive aim of the invention to overcome, or at least substantially reduce, the aforementioned problems.
According to a first aspect of the invention there is provided a method of calibrating a mass spectrometer comprising:
The method may comprise:
The interpolating step may be performed using probabilistic methods.
The calibration curve may be constructed using probabilistic methods.
The method may comprise experimentally determining the mass to charge ratios of at least three (e.g. three) chemical compounds in the reference standard at each scan speed.
According to a second aspect of the invention there is provided a mass spectrometer having associated therewith a computer for performing data analysis functions of data produced by the mass spectrometer, the computer performing the method of the first aspect of the invention.
The invention is particularly advantageous in quadrupole mass spectrometers. However, the present invention may also be useful in other types of mass spectrometer.
The mass spectrometer may further comprise:
The mass spectrometer may also comprise a chromatography or other separation device upstream of an ion source.
In some embodiments, the chromatography separation device may comprise a liquid chromatography or gas chromatography device.
In some embodiments, the separation device may comprise: (i) a Capillary Electrophoresis (“CE”) separation device; (ii) a Capillary Electrochromatography (“CEC”) separation device; (iii) a substantially rigid ceramic-based multilayer microfluidic substrate (“ceramic tile”) separation device; or (iv) a supercritical fluid chromatography separation device.
Embodiments of the invention will now be described, by way of example only, with reference to the following figures, in which:
Each set of data was interpolated to estimate a mass to charge ratio for a scan speed other than one of the five scan speeds that were used to create the sets of data. The chosen scan speed could be any scan speed (e.g. 1500 Da/e/s) from which the user wishes to obtain a calibration curve. Interpolating the sets of data provides estimated mass to charge ratios for the three chemical compounds at the chosen scan speed. A calibration curve (not shown) utilising the estimated mass to charge ratios at the chosen scan speed was constructed.
Each set of data was interpolated to estimate a mass to charge ratio for a scan speed other than one of the ten scan speeds that were used to create the sets of data. The chosen scan speed could be any scan speed (e.g. 1500 Da/e/s) from which the user wishes to obtain a calibration curve. Interpolating the sets of data provides estimated mass to charge ratios for the three chemical compounds at the chosen scan speed. A calibration curve (not shown) utilising the estimated mass to charge ratios at the chosen scan speed was constructed.
Data corresponding to
In some embodiments, the interpolating step may be performed using probabilistic methods, such as those described in GB2405991 and Calibration and Interpolation by John Skilling; AIP Conference Proceedings 872, 321 (2006), the contents of which are herein incorporated by reference.
Methods according to the invention enable a user to construct a calibration curve for any scan speed without having to perform a calibration experiment at that scan speed. Advantageously, this can reduce the time it takes to calibrate a mass spectrometer, leaving more time available in the day to acquire experimental data.
To summarise, methods of the invention are an improvement over existing methods because calibration curves can be constructed for any given scan speed from data that has been obtained using a discrete number of scan speeds, e.g. five or ten scan speeds as per the described embodiments.
When used herein the term “low scan speed” is intended to mean scan speeds in the range 5 Da/e/s to 5000 Da/e/s. The term “high scan speed” is intended to mean any scan speed greater than 5000 Da/e/s.
When used in this specification and claims, the terms “comprises” and “comprising” and variations thereof mean that the specified features, steps or integers are included. The terms are not to be interpreted to exclude the presence of other features, steps or components.
The features disclosed in the foregoing description, or the following claims, or the accompanying drawings, expressed in their specific forms or in terms of a means for performing the disclosed function, or a method or process for attaining the disclosed result, as appropriate, may, separately, or in any combination of such features, be utilised for realising the invention in diverse forms thereof.
Although certain example embodiments of the invention have been described, the scope of the appended claims is not intended to be limited solely to these embodiments. The claims are to be construed literally, purposively, and/or to encompass equivalents.
Number | Date | Country | Kind |
---|---|---|---|
2013007.6 | Aug 2020 | GB | national |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/EP2021/073185 | 8/20/2021 | WO |