The present invention relates to volume dimensioning devices.
Volume dimensioning devices, i.e. dimensioners, are devices that are used for estimating sizes of items (such as boxes) and the sizes of empty spaces (such as the volume left in a delivery truck). Dimensioners may be larger devices that are a static part of a larger logistical system in a distribution center or warehouse, or they may be smaller mobile devices designed for portable use. Mobile dimensioners that are certified for use in commerce can be used to charge customers for shipment based on the dimensions of an item. The National Conference on Weights and Measures (NCWM) issues a National Type Evaluation Program (NTEP) Certificate of Conformance to mobile dimensioners that have been evaluated and found to produce accurate measurements capable of meeting applicable requirements of the National Institute of Standards and Technology (NIST) Handbook 44, entitled “Specifications, Tolerances, and Other Technical Requirements for Weighing and Measuring Devices.”
Despite the certification process, mobile dimensioners can have variable tolerances in measurements as a result of the inherent variations that arise from different methods of measurement. The same item could be measured from two different locations, resulting in two different methods of measurement, each with a different angle relative to the item being measured as well as a different distance to the item being measured. Consequently, it is possible to have two different measurements for an item, both of which are certifiable and correct, all because of the variable tolerances in the methods of measurement. More specifically, because the accuracy dimension (referred to as “d” in the NIST and NTEP documentation) can change, two different yet valid measurements can be obtained simply by moving the mobile dimensioner around.
By way of a non-limiting example, assume that one dimension of an item to be shipped has been measured with a mobile dimensioner to be 9.5 with an accuracy dimension of 0.5 (i.e. d=0.5). Simply moving the mobile dimensioner side to side or further away might provide the same dimension with a measurement of 10 with an accuracy dimension of 1.0 (i.e. d=1.0). In yet other situations, it is possible to produce variable measurements for the same dimension with the same accuracy dimension. Again, simply by moving the mobile dimensioner in and out, it would be possible to go from a measurement of 10 with a d=1.0 to a measurement of 9 with a d=1.0.
One of the primary reasons behind government oversight of the measurement process is to ensure that vendors are not employing improper measurements in their business transactions with both customers and shipping companies. Since a mobile dimensioner has the inherent ability to produce different certifiable measurements, a disreputable vendor, could in practice, move the device back and forth within the useable range, for example closer and father away, always looking at the reported dimension and then picking the larger dimension for overcharging customers and the smaller dimension for cheating shippers. Therefore, over time, a disreputable vendor can employ a certified mobile dimensioner to determine a method of measurement designed to systematically defraud customers and shippers.
Therefore, a need exists for a mobile dimensioner designed to thwart activities intended to generate fraudulent measurements.
Accordingly one aspect of the present invention discloses a mobile dimensioner device, comprising: a display; one or more optical sensors; one or more measurement sensors; an input subsystem; a clock system; one or more processors; and memory containing instructions executable by the one or more processors whereby the device is operable to: receive a threshold time period; activate at least one of the one or more measurement sensors; derive a first set of dimensions for an object and an associated indication of the dimensional accuracy of each of the dimensions based on information received from the one or more measurement sensors; display, on the display, the first set of dimensions and the associated indication of the dimensional accuracy of each of the dimensions; determine the time interval since the first set of dimensions for the object was derived; if the time interval exceeds the threshold time period, receive a deactivation event.
In other exemplary embodiments, the threshold time period is defined by one of the group consisting of: defined by the manufacturer of the device, defined to comply with certification standards set by a certification organization, defined in response to input received via the input subsystem at the device, and defined in response to information received at the device from a server.
In additional exemplary embodiments, the deactivation event is selected from the group consisting of: a power off event for the device, an event that turns off the ability of the device to take measurements, an event that turns off the one or more measurement sensors of the device, an event that restricts the ability of the device to report results, an event that turns off one or more communication interfaces of the device, an event that deactivates the measurement sensors and displays the first set of dimensions, an event that deactivates the measurement sensors and places the device in a state requiring reset, and an event that deactivates the measurement sensors and deletes the first set of dimensions.
In further embodiments, the one or more optical sensors are selected from a group consisting of: a barcode sensor, a camera, and an image sensor.
In yet other embodiments, the one or more measurement sensors are selected from a group consisting of: point-cloud projection, structured light, and stereoscopic cameras and n-scopic cameras.
Another aspect of the present invention discloses a mobile dimensioner device, comprising: a display; one or more optical sensors; one or more measurement sensors; an input subsystem; one or more processors; and memory containing instructions executable by the one or more processors whereby the device is operable to: receive a threshold number of contrary events; activate at least one of the one or more measurement sensors; derive a first set of dimensions for an object and an associated indication of the dimensional accuracy of each of the dimensions based on information received from the one or more measurement sensors; display, on the display, the first set of dimensions and the associated indication of the dimensional accuracy of each of the dimensions; display, on the display, an indication to obtain a better measurement of the object; detect a number of contrary events; if the number of contrary events detected exceeds the threshold number of contrary events, receive a deactivation event.
In still other exemplary embodiments, the device is further operable to: derive a set of preliminary dimensions for an object based on information received from the one or more measurement sensors.
In more embodiments, the contrary event is an action that does not correspond to an indication to obtain a better measurement of the object.
In some embodiments, the threshold number of contrary events is defined by one of the group consisting of: defined by the manufacturer of the device, defined to comply with certification standards set by a certification organization, defined in response to input received via the input subsystem at the device, and defined in response to information received at the device from a server.
An additional aspect of the present invention discloses a mobile dimensioner device, comprising: a display; one or more optical sensors; one or more measurement sensors; an input subsystem; one or more processors; and memory containing instructions executable by the one or more processors whereby the device is operable to: activate at least one of the one or more measurement sensors; derive a set of first dimensions for an object and an associated indication of the first dimensional accuracy of each of the first dimensions based on information received from the one or more measurement sensors; display, on the display, the set of first dimensions and the associated indication of the first dimensional accuracy of each of the first dimensions; derive a set of second dimensions for an object and an associated indication of the second dimensional accuracy of each of the second dimensions based on information received from the one or more measurement sensors; display, on the display, the set of second dimensions and the associated indication of the first dimensional accuracy of each of the first dimensions; in response to an input to capture the set of second dimensions, determine if the second dimensional accuracy is greater than the first dimensional accuracy; if the second dimensional accuracy is greater than the first dimensional accuracy; then receive a deactivation event; and if the second dimensional accuracy is not greater than the first dimensional accuracy; then capture the second set of dimensions.
In yet other embodiments, the device further comprises: a communication interface.
In still more embodiments, the communication interface is selected from the group consisting of: Bluetooth, Ethernet, wireless Ethernet, USB, serial, and I2C.
In other embodiments, the device is further operable to: send the second set of dimensions to a server.
The foregoing illustrative summary, as well as other exemplary objectives and/or advantages of the invention, and the manner in which the same are accomplished, are further explained within the following detailed description and its accompanying drawings.
The present invention embraces the concept of restricting a mobile dimensioner from reporting systematically either the highest or lowest dimensions. Because the measurement results from a mobile dimensioner are not predictable, e.g. a mobile dimensioner used at its farthest range will not necessarily produce larger or smaller dimensions, a disreputable vendor must move the mobile dimensioner in and out and/or right or left looking for a specific measurement conducive to defrauding customers and shippers. This behavior must be repeated with each measurement because each item being measured will have a different size and will produce different results. In embodiments of the present invention, a mobile dimensioner device detects excessive measuring time and/or a repetitive range of motion and deactivates upon detection of this inappropriate behavior. In other embodiments of the present invention, a mobile dimensioner device detects when a measurement with a greater accuracy has been taken (i.e. a small accuracy dimension “d” value) and restricts the dimensioner from reporting measurements with less accuracy (i.e. a large accuracy dimension “d” value).
In general, device 100 may be implemented in any form of digital computer or mobile device. Digital computers may include, but are not limited to, laptops, desktops, workstations, fixed vehicle computers, vehicle mount computers, hazardous environment computers, rugged mobile computers, servers, blade servers, mainframes, other appropriate computers. Mobile devices may include, but are not limited to, cellular telephones, smart phones, personal digital assistants, tablets, pagers, two-way radios, netbooks, barcode scanners, radio frequency identification (RFID) readers, intelligent sensors, tracking devices, volume dimensioning devices, mobile dimensioners, and other similar computing devices.
In some embodiments of the present invention, the device 100 of
In other embodiments of the present invention, the device 100-1 may be connected to another device 100-2 via a wired communication channel 176, as shown in
In one embodiment, the device 100-1 is a mobile dimensioner device and the device 100-2 is a server than handles backend functions like invoicing customers for the packages being shipped. In this embodiment,
In general, as shown, the device 100 of
The processing system 110 also includes a peripherals interface 116 for communicating with other components of the device 100, including but not limited to, radio frequency (RF) circuitry 152, such as Wi-Fi and/or cellular communications circuitry such as wireless Ethernet, Bluetooth, and near field communication (NFC), audio circuitry 154 for the audio input component 153, such as a microphone, and audio output component 155, such as a speaker, one or more accelerometers 156, one or more other sensors 158, such as a location determination component such as a Global Positioning System (GPS) chip, and one or more external ports 160, which may be used for smart card readers or for wired connections such as wired Ethernet, USB, serial or I2C ports. The RF circuitry 152 and external ports 160 individually and collectively make up the communication interfaces for the device 100. The processing system 110 is also connected to a power system component 120 that is used to power the device 100, such as a battery or a power supply unit. The processing system 110 is also connected to a clock system component 130 that controls a timer for use by the disclosed embodiments.
The peripherals interface 116 may also communicate with an Input/Output (I/O) subsystem 140, which includes a display(s) controller 141 operative to control display(s) 142. In some embodiments the display(s) 142 is a touch-sensitive display system, and the display(s) controller 141 is further operative to process touch inputs on the touch sensitive display 142. The I/O subsystem 140 may also include a keypad(s) controller 143 operative to control keypad(s) 144 on the device 100. The I/O subsystem 140 also includes an optical sensor(s) controller 145 operative to control one or more optical sensor(s) 146. The optical sensor(s) may include, but is not limited to, a barcode sensor, a camera, and an image sensor. The I/O subsystem 140 also includes a measurement sensor(s) controller 147 operative to control one or more measurement sensor(s) 148. The measurement sensor(s) may include, but is not limited to, a point-cloud projection sensor, a structured light sensor, a stereoscopic camera, and an n-scopic camera. The components of device 100 may be interconnected using one or more buses, represented generically by the arrows of
In some embodiments, the activation event comprises a power on event or a power cycling event for the mobile dimensioner device 100. In other embodiments, the activation event comprises an event that turns on the ability of the mobile dimensioner device 100 to take measurements, such as turning on one or more measurement sensors 148. In other embodiments, the activation event comprises an event that turns on the communication interfaces 160 and/or 152 of mobile dimensioner device 100-1 to report measurement results to the server 100-2. In still other embodiments, an activation event may also comprise a reset of any existing measurements currently in memory 113 or storage 115 of the mobile dimensioner device 100, or a reset of the mobile dimensioner device itself. In additional embodiments, the activation event includes a preliminary scan of the object to derive some preliminary dimensions. Note that in some embodiments, the object can be empty space, e.g. the amount of dirt removed from a hole or the amount of space remaining in a delivery truck. However, in other embodiments, the object will be a package that is being shipped.
Next, the mobile dimensioner device 100 then resets a timer used in the detection of excessive measuring time. This timer is called the acc-meas-timer in
The mobile dimensioner device 100 then checks to see if the acc-meas-timer is greater than a specified threshold (Step 208). This threshold is called the acc-meas-time-threshold in
If the acc-meas-timer is less than or equal to the acc-meas-time-threshold (Path 207), i.e. the amount of time spent between obtaining and capturing an accurate measurement is not excessive, then the mobile dimensioner device 100 checks to see if an accurate measurement has been obtained and displayed (Step 210). If no accurate measurement has been obtained and displayed (Path 211), then the mobile dimensioner device 100 resets the acc-meas-timer (Step 222), and the process then continues to Step 208 as described above.
Returning to Step 210, if an accurate measurement has been obtained and displayed (Path 213), then the mobile dimensioner device checks to see if the measurement has been captured (Step 216). In some embodiments, the measurement may be captured in response to an input at the mobile dimensioner device 100, or may be set in some other manner. If the measurement is captured (Path 217), then the measurement results are reported (Step 218) and the process is complete (Step 220). If the measurement is not captured (Path 215), then the clock system 130 of the mobile dimensioner device 100 increments the acc-meas-timer with the passage of time (Step 206) and the process continues to Step 208 as described above.
Returning to Step 208, if the acc-meas-timer is greater than the acc-meas-time-threshold (Path 209), i.e. the amount of time spent in obtaining and capturing an accurate measurement is excessive, then the process continues to Step 214 where a deactivation event is received by the mobile dimensioner device 100, and the mobile dimensioner device returns to a state where it waits for an activation event (Step 202). In some embodiments, the deactivation event comprises a power off event for the mobile dimensioner device 100 itself. In alternative embodiments, the deactivation event comprises placing the device in a state requiring a reset, such as a key sequence to reset or a simple power cycle reset. In other embodiments, the deactivation event comprises an event that turns off the ability of the mobile dimensioner device 100 to take measurements, such as an event that turns off or temporarily disables one or more measurement sensors 148. In some embodiments, any active measurements in the mobile dimensioner device at the time of the deactivation event may be cleared, i.e. deleted or erased. In other embodiments, any active measurements in the mobile dimensioner device at the time of the deactivation event may be displayed. In yet other embodiments, the deactivation event comprises events restricting the ability of the mobile dimensioner device 100-1 to report the results to the server 100-2, such as events that turn off the communication interfaces 152 and/or 160 of the mobile dimensioner device 100. In some embodiments, the deactivation events are initiated by the mobile dimensioner device 100 itself in response to the criteria met in accordance with
In this manner,
As described earlier, there are different embodiments for the activation event, including but not limited to: a power on event, a power cycling event, an event that turns on the ability to take measurements, an event that turns on the communication interfaces, an event that resets existing measurements, an event that resets the mobile dimensioner device, and an event that includes a preliminary scan of the object.
Next, the mobile dimensioner device 100 then resets a counter used in the detection of repetitive motion. This counter is called the contrary-event-counter in
A contrary event is an action by the mobile dimensioning device 100 that does not correspond to an indication for a better measurement of the object. As described earlier, an action may be a new measurement taken by the mobile dimensioning device 100, a movement of the mobile dimensioner device 100, or a combination of both.
An indication may either be text or graphics (or both) for a movement that the mobile dimensioner device 100 should take or a measurement that the mobile dimensioner device 100 should obtain (or both a movement and a measurement) in order to better measure the object being measured. By way of a non-limiting example, a movement indication may be a text instruction that provides directions, such as move left, move up, move in closer, etc., that allow the mobile dimensioner device 100 to be moved into a better position for measuring the subject being measured. In another non-limiting example, a movement indication may be an arrow that provides visual cues, such as move down, move right, move back further, etc., that allow the mobile dimensioner device 100 to be moved into a better position for measuring the subject being measured. A measurement indication may include, but is not limited to, a textual instruction, such as “measure the depth of the object”, that allows the mobile dimensioner device 100 to obtain a measurement of a particular dimension of the subject being measured. A measurement indication may also include, but is not limited to, a visual representation of the object being measured that highlights particular dimensions of the subject being measured for the mobile dimensioner device 100 to obtain, such as an icon of a box with the depth dimension blinking.
In this respect, a contrary event is more specifically defined as a measurement or movement (or both) by the mobile dimensioning device 100 that does not correspond to a text or graphic (or both) that provides information designed to help the mobile dimensioner device 100 obtain better measurements of the subject being measured. Accordingly, the contrary-event-counter is used to track the number of times that movements or measurements of the mobile dimensioner device 100 are not aligned with the goal of obtaining better measurements for the subject being measured.
In some embodiments, a contrary event also occurs whenever new accurate measurements are derived after an accurate measurement has already been derived but not captured.
Returning to
Next, the mobile dimensioner device 100 checks to see if the contrary-event-counter is greater than a specified threshold (Step 312). This threshold is called the contrary-event-threshold in
If the contrary-event-counter is less than or equal to the contrary-event-threshold (Path 307), i.e. the number of times that a contrary event has occurred is less than the allowed number, then the mobile dimensioner device 100 checks to see if an accurate measurement has been obtained and displayed (Step 314). If no accurate measurement has been obtained and displayed (Path 311), then the mobile dimensioner device 100 receives new actions (Step 316). In some embodiments, new actions may be new measurements taken by the mobile dimensioner device. In other embodiments, new actions may be movements of the mobile dimensioner device 100, including but not limited to movements in three dimensional space (up-and-down, side-to-side, front-to-back), or a repositioning of the viewing angle of the mobile dimensioner device 100 relative to the object being measured. In other embodiments, new actions include both new measurements and new movements. It should be noted that if the actions do not produce an accurate measurement that can be displayed, then the process set forth in
Next, the mobile dimensioner device 100 checks to see if there are any indications for obtaining a better measurement of the object being measured (Step 322). If not (Path 315), then the process continues as indicated by the connector A. If there are indications (Path 317), then the mobile dimensioner device checks to see if the new actions followed or corresponded to the indications (Step 326). If the new actions followed the indications (Path 325), then the process continues as indicated by connector A. If the new actions did not correspond to the indications (Path 323), then the indication-measurement-flag is set to TRUE (Step 332) and the process continues as indicated by connector A.
Returning to Step 314, if an accurate measurement has been obtained and displayed (Path 313), then the accurate-measurement-flag is set to TRUE (Step 320), and then the mobile dimensioner device checks to see if the measurement has been captured (Step 324). As described earlier, in some embodiments, the measurement may be captured in response to input at the mobile dimensioner device 100, or some other manner. If the measurement is captured (Path 321), then the measurement results are reported (Step 330) and the process is complete (Step 328). If the measurement is not captured (Path 319), then process then continues to Step 316 where new actions are received by the mobile dimensioner device 100, as already described.
Connector A from
Connector B from
As described earlier, there are different embodiments for the deactivation event, including but not limited to: a power off event, an event that turns off the ability to take measurements, an event that turns off sensors, an event that restricts the reporting of results, an event that turns off communication interfaces, an event that deactivates sensors and displays the last set of dimensions, an event that deactivates sensors and requires a device reset, and an event that deactivates and deletes the last set of dimensions.
In this manner,
The process begins in
As described earlier, there are different embodiments for the activation event, including but not limited to: a power on event, a power cycling event, an event that turns on the ability to take measurements, an event that turns on the communication interfaces, an event that resets existing measurements, an event that resets the mobile dimensioner device, and an event that includes a preliminary scan of the object.
Next, the mobile dimensioner device 100 then resets a counter used to track the number of times that a contrary event occurs, i.e. the contrary-event-counter (Step 406). The next steps in the process resets the accurate-measurement-flag (Step 408) which is used to track when an accurate measurement has been obtained. The indication-measurement-flag, which is used to track when indications for better measurements are not followed, is then reset (Step 410). Next, the mobile dimensioner device 100 resets the acc-meas-timer (Step 412), which is the timer used to track the amount of time that transpires before the mobile dimensioner device 100 derives an accurate measurement.
An additional timer, called the indication-timer, is then reset (Step 414). This timer is used to track the aggregate elapsed time that the mobile dimensioner device 100 spends in movements or measurements that are not aligned with the goal of obtaining better measurements for the subject being measured. The process then continues as indicated by connector C.
Connector C from
If the acc-meas-timer is less than or equal to the acc-meas-time-threshold (Path 435), i.e. the amount of time spent in obtaining and capturing an accurate measurement is not excessive, then the mobile dimensioner device 100 checks to see if the indication-timer is greater than a specified threshold (Step 458). This threshold is called the indication-time-threshold in
If the indication-timer is less than or equal to the indication-time-threshold (Path 439), the mobile dimensioner device 100 then checks to see if the contrary-event-counter is greater than a specified threshold (Step 460). This threshold is called the contrary-event-threshold in
If the contrary-event-counter is less than or equal to the contrary-event-threshold (Path 443), i.e. the number of times that a contrary event has occurred is less than or equal to the allowed number, then the process continues as indicated by connector F.
If the acc-meas-timer is greater than the acc-meas-time-threshold (Path 437), i.e. the amount of time spent in obtaining and capturing an accurate measurement is excessive, then the process continues to Step 462 where a deactivation event is received by the mobile dimensioner device 100. The process then continues as indicated by connector D.
If the indication-timer is greater than the indication-time threshold (Path 441), i.e. the aggregate elapsed time that the mobile dimensioner device 100 spends in movements or measurements that are not aligned with the goal of obtaining better measurements for the object being measured is greater than what is allowed, then the process continues to Step 462 where a deactivation event is received by the mobile dimensioner device 100. The process then continues as indicated by connector D.
If the contrary-event-counter is greater than the contrary-event-threshold (Path 445), i.e. the number of times that a contrary event has occurred is now greater than the allowed number, then the mobile dimensioner device 100 then the process continues to Step 462 where a deactivation event is received by the mobile dimensioner device 100. The process then continues as indicated by connector D.
As described earlier, there are different embodiments for the deactivation event, including but not limited to: a power off event, an event that turns off the ability to take measurements, an event that turns off sensors, an event that restricts the reporting of results, an event that turns off communication interfaces, an event that deactivates sensors and displays the last set of dimensions, an event that deactivates sensors and requires a device reset, and an event that deactivates and deletes the last set of dimensions.
Connector D of
Returning to Step 460, Connector F of
Next, the mobile dimensioner device 100 checks to see if there are any indications for obtaining a better measurement of the object being measured (Step 422). If not (Path 411), then the process continues as indicated by the connector E. If there are indications (Path 413), then the mobile dimensioner device checks to see if the new actions followed or corresponded to the indications (Step 426). If the new actions follow the indications (Path 421), then the incrementing of the indication-timer with the passage of time, if it has been running, is stopped or paused (Step 434) and the process continues as indicated by connector E. If the new actions do not correspond to the indications (Path 419), then the indication-measurement-flag is set to TRUE (Step 432), the indication-timer is incremented with the passage of time (Step 436), and the process continues as indicated by connector E.
Connector E of
The mobile dimensioner device 100 then checks to see if the number of contrary events is 0 (Step 442). If not (Path 431), then the process continues as indicated by connector C. If the number of contrary events is 0 (Path 433), then the acc-meas-time-threshold is augmented (Step 452), and the process continues as indicated by connector C. The augmentation of the acc-meas-time-threshold effectively rewards movements and measurements by the mobile dimensioner device 100 that are aligned with the goal of obtaining better measurements for the subject being measured by giving more time to derive an accurate measurement. Connector C from
Returning to Step 416, if an accurate measurement has been derived and displayed (Path 409), then the accurate-measurement-flag is set to TRUE (Step 420), and then the mobile dimensioner device checks to see if the measurement has been captured (Step 424). As described earlier, in some embodiments, the measurement may be captured in response to input at the mobile dimensioner device 100, or may be set in some other manner. If the measurement is captured (Path 417), then the measurement results are reported (Step 430) and the process is complete (Step 428). If the measurement is not captured (Path 415), then the clock system 130 of the mobile dimensioner device 100 increments the acc-meas-timer with the passage of time (Step 454), and the process then continues to Step 418 where new actions are received by the mobile dimensioner device 100. In alternative embodiments, once the clock system 130 of the mobile dimensioner device 100 increments the acc-meas-timer with the passage of time (Step 454), the process continues as indicated by connector C. In this embodiment, similar to
In this manner,
As described earlier, there are different embodiments for the activation event, including but not limited to: a power on event, a power cycling event, an event that turns on the ability to take measurements, an event that turns on the communication interfaces, an event that resets existing measurements, an event that resets the mobile dimensioner device, and an event that includes a preliminary scan of the object.
The mobile dimensioner device 100 first resets a variable used to store the previous measurements taken by the mobile dimensioner device 100 (Step 504). This is the prev-measurement variable in
Next, the mobile dimensioner device 100 checks to see if an accurate measurement has been derived and displayed (Step 510). If no accurate measurement has been derived and displayed yet (Path 507), then the mobile dimensioner device 100 receives new actions (Step 512). As described above, new actions are movements, measurements, or both. The process then continues as indicated by connector G.
Connector G in
Connector H of
As described earlier, deactivation events may be a power off event, an event requiring a reset, an event that turns off the ability to take measurements, an event that clears out any active measurements, an event restricting the ability to report the results, or any combination therein.
In this manner,
In this respect, the processes described in
To supplement the present disclosure, this application incorporates entirely by reference the following commonly assigned patents, patent application publications, and patent applications:
In the specification and/or figures, typical embodiments of the invention have been disclosed. The present invention is not limited to such exemplary embodiments. The use of the term “and/or” includes any and all combinations of one or more of the associated listed items. The figures are schematic representations and so are not necessarily drawn to scale. Unless otherwise noted, specific terms have been used in a generic and descriptive sense and not for purposes of limitation.
Number | Name | Date | Kind |
---|---|---|---|
3971065 | Bayer | Jul 1976 | A |
4026031 | Siddall et al. | May 1977 | A |
4279328 | Ahlbom | Jul 1981 | A |
4398811 | Nishioka et al. | Aug 1983 | A |
4495559 | Gelatt, Jr. | Jan 1985 | A |
4730190 | Win et al. | Mar 1988 | A |
4803639 | Steele et al. | Feb 1989 | A |
5184733 | Amarson et al. | Feb 1993 | A |
5198648 | Hibbard | Mar 1993 | A |
5220536 | Stringer et al. | Jun 1993 | A |
5331118 | Jensen | Jul 1994 | A |
5359185 | Hanson | Oct 1994 | A |
5384901 | Glassner et al. | Jan 1995 | A |
5548707 | LoNegro et al. | Aug 1996 | A |
5555090 | Schmutz | Sep 1996 | A |
5561526 | Huber et al. | Oct 1996 | A |
5590060 | Granville et al. | Dec 1996 | A |
5606534 | Stringer et al. | Feb 1997 | A |
5619245 | Kessler et al. | Apr 1997 | A |
5655095 | LoNegro et al. | Aug 1997 | A |
5661561 | Wurz et al. | Aug 1997 | A |
5699161 | Woodworth | Dec 1997 | A |
5729750 | Ishida | Mar 1998 | A |
5730252 | Herbinet | Mar 1998 | A |
5732147 | Tao | Mar 1998 | A |
5734476 | Dlugos | Mar 1998 | A |
5737074 | Haga et al. | Apr 1998 | A |
5767962 | Suzuki et al. | Jun 1998 | A |
5831737 | Stringer et al. | Nov 1998 | A |
5850370 | Stringer et al. | Dec 1998 | A |
5850490 | Johnson | Dec 1998 | A |
5869827 | Rando | Feb 1999 | A |
5870220 | Migdal et al. | Feb 1999 | A |
5900611 | Hecht | May 1999 | A |
5923428 | Woodworth | Jul 1999 | A |
5929856 | LoNegro et al. | Jul 1999 | A |
5938710 | Lanza et al. | Aug 1999 | A |
5959568 | Woolley | Sep 1999 | A |
5960098 | Tao | Sep 1999 | A |
5969823 | Wurz et al. | Oct 1999 | A |
5978512 | Kim et al. | Nov 1999 | A |
5979760 | Freyman et al. | Nov 1999 | A |
5988862 | Kacyra et al. | Nov 1999 | A |
5991041 | Woodworth | Nov 1999 | A |
6009189 | Schaack | Dec 1999 | A |
6025847 | Marks | Feb 2000 | A |
6035067 | Ponticos | Mar 2000 | A |
6049386 | Stringer et al. | Apr 2000 | A |
6053409 | Brobst et al. | Apr 2000 | A |
6064759 | Buckley et al. | May 2000 | A |
6067110 | Nonaka et al. | May 2000 | A |
6069696 | McQueen et al. | May 2000 | A |
6115114 | Berg et al. | Sep 2000 | A |
6137577 | Woodworth | Oct 2000 | A |
6177999 | Wurz et al. | Jan 2001 | B1 |
6189223 | Haug | Feb 2001 | B1 |
6232597 | Kley | May 2001 | B1 |
6236403 | Chaki | May 2001 | B1 |
6246468 | Dimsdale | Jun 2001 | B1 |
6333749 | Reinhardt et al. | Dec 2001 | B1 |
6336587 | He et al. | Jan 2002 | B1 |
6369401 | Lee | Apr 2002 | B1 |
6373579 | Ober et al. | Apr 2002 | B1 |
6429803 | Kumar | Aug 2002 | B1 |
6457642 | Good et al. | Oct 2002 | B1 |
6507406 | Yagi et al. | Jan 2003 | B1 |
6517004 | Good et al. | Feb 2003 | B2 |
6519550 | D'Hooge et al. | Feb 2003 | B1 |
6674904 | McQueen | Jan 2004 | B1 |
6705526 | Zhu et al. | Mar 2004 | B1 |
6781621 | Gobush et al. | Aug 2004 | B1 |
6824058 | Patel et al. | Nov 2004 | B2 |
6832725 | Gardiner et al. | Dec 2004 | B2 |
6858857 | Pease et al. | Feb 2005 | B2 |
6922632 | Foxlin | Jul 2005 | B2 |
6971580 | Zhu et al. | Dec 2005 | B2 |
6995762 | Pavlidis et al. | Feb 2006 | B1 |
7057632 | Yamawaki et al. | Jun 2006 | B2 |
7085409 | Sawhney et al. | Aug 2006 | B2 |
7086162 | Tyroler | Aug 2006 | B2 |
7104453 | Zhu et al. | Sep 2006 | B1 |
7128266 | Zhu et al. | Oct 2006 | B2 |
7137556 | Bonner et al. | Nov 2006 | B1 |
7159783 | Walczyk et al. | Jan 2007 | B2 |
7161688 | Bonner et al. | Jan 2007 | B1 |
7205529 | Andersen et al. | Apr 2007 | B2 |
7214954 | Schopp | May 2007 | B2 |
7277187 | Smith et al. | Oct 2007 | B2 |
7307653 | Dutta | Dec 2007 | B2 |
7310431 | Gokturk et al. | Dec 2007 | B2 |
7413127 | Ehrhart et al. | Aug 2008 | B2 |
7509529 | Colucci et al. | Mar 2009 | B2 |
7527205 | Zhu | May 2009 | B2 |
7586049 | Wurz | Sep 2009 | B2 |
7602404 | Reinhardt et al. | Oct 2009 | B1 |
7639722 | Paxton et al. | Dec 2009 | B1 |
7726575 | Wang et al. | Jun 2010 | B2 |
7780084 | Zhang et al. | Aug 2010 | B2 |
7788883 | Buckley et al. | Sep 2010 | B2 |
7974025 | Topliss | Jul 2011 | B2 |
8027096 | Feng et al. | Sep 2011 | B2 |
8028501 | Buckley et al. | Oct 2011 | B2 |
8050461 | Shpunt et al. | Nov 2011 | B2 |
8055061 | Katano | Nov 2011 | B2 |
8072581 | Breiholz | Dec 2011 | B1 |
8102395 | Kondo et al. | Jan 2012 | B2 |
8132728 | Dwinell et al. | Mar 2012 | B2 |
8134717 | Pangrazio et al. | Mar 2012 | B2 |
8149224 | Kuo et al. | Apr 2012 | B1 |
8194097 | Xiao et al. | Jun 2012 | B2 |
8201737 | Palacios Durazo et al. | Jun 2012 | B1 |
8212889 | Chanas et al. | Jul 2012 | B2 |
8228510 | Pangrazio et al. | Jul 2012 | B2 |
8230367 | Bell et al. | Jul 2012 | B2 |
8294969 | Plesko | Oct 2012 | B2 |
8305458 | Hara | Nov 2012 | B2 |
8310656 | Zalewski | Nov 2012 | B2 |
8313380 | Zalewski et al. | Nov 2012 | B2 |
8317105 | Kotlarsky et al. | Nov 2012 | B2 |
8322622 | Liu | Dec 2012 | B2 |
8339462 | Stec et al. | Dec 2012 | B2 |
8350959 | Topliss et al. | Jan 2013 | B2 |
8351670 | Ijiri et al. | Jan 2013 | B2 |
8366005 | Kotlarsky et al. | Feb 2013 | B2 |
8371507 | Haggerty et al. | Feb 2013 | B2 |
8376233 | Van Horn et al. | Feb 2013 | B2 |
8381976 | Mohideen et al. | Feb 2013 | B2 |
8381979 | Franz | Feb 2013 | B2 |
8390909 | Plesko | Mar 2013 | B2 |
8408464 | Zhu et al. | Apr 2013 | B2 |
8408468 | Horn et al. | Apr 2013 | B2 |
8408469 | Good | Apr 2013 | B2 |
8424768 | Rueblinger et al. | Apr 2013 | B2 |
8437539 | Komatsu et al. | May 2013 | B2 |
8441749 | Brown et al. | May 2013 | B2 |
8448863 | Xian et al. | May 2013 | B2 |
8457013 | Essinger et al. | Jun 2013 | B2 |
8459557 | Havens et al. | Jun 2013 | B2 |
8463079 | Ackley et al. | Jun 2013 | B2 |
8469272 | Kearney | Jun 2013 | B2 |
8474712 | Kearney et al. | Jul 2013 | B2 |
8479992 | Kotlarsky et al. | Jul 2013 | B2 |
8490877 | Kearney | Jul 2013 | B2 |
8517271 | Kotlarsky et al. | Aug 2013 | B2 |
8523076 | Good | Sep 2013 | B2 |
8528818 | Ehrhart et al. | Sep 2013 | B2 |
8544737 | Gomez et al. | Oct 2013 | B2 |
8548420 | Grunow et al. | Oct 2013 | B2 |
8550335 | Samek et al. | Oct 2013 | B2 |
8550354 | Gannon et al. | Oct 2013 | B2 |
8550357 | Kearney | Oct 2013 | B2 |
8556174 | Kosecki et al. | Oct 2013 | B2 |
8556176 | Van Horn et al. | Oct 2013 | B2 |
8556177 | Hussey et al. | Oct 2013 | B2 |
8559767 | Barber et al. | Oct 2013 | B2 |
8561895 | Gomez et al. | Oct 2013 | B2 |
8561903 | Sauerwein | Oct 2013 | B2 |
8561905 | Edmonds et al. | Oct 2013 | B2 |
8565107 | Pease et al. | Oct 2013 | B2 |
8570343 | Halstead | Oct 2013 | B2 |
8571307 | Li et al. | Oct 2013 | B2 |
8576390 | Nunnink | Nov 2013 | B1 |
8579200 | Samek et al. | Nov 2013 | B2 |
8583924 | Caballero et al. | Nov 2013 | B2 |
8584945 | Wang et al. | Nov 2013 | B2 |
8587595 | Wang | Nov 2013 | B2 |
8587697 | Hussey et al. | Nov 2013 | B2 |
8588869 | Sauerwein et al. | Nov 2013 | B2 |
8590789 | Nahill et al. | Nov 2013 | B2 |
8594425 | Gurman et al. | Nov 2013 | B2 |
8596539 | Havens et al. | Dec 2013 | B2 |
8596542 | Havens et al. | Dec 2013 | B2 |
8596543 | Havens et al. | Dec 2013 | B2 |
8599271 | Havens et al. | Dec 2013 | B2 |
8599957 | Peake et al. | Dec 2013 | B2 |
8600158 | Li et al. | Dec 2013 | B2 |
8600167 | Showering | Dec 2013 | B2 |
8602309 | Longacre et al. | Dec 2013 | B2 |
8608053 | Meier et al. | Dec 2013 | B2 |
8608071 | Liu et al. | Dec 2013 | B2 |
8611309 | Wang et al. | Dec 2013 | B2 |
8615487 | Gomez et al. | Dec 2013 | B2 |
8621123 | Caballero | Dec 2013 | B2 |
8622303 | Meier et al. | Jan 2014 | B2 |
8628013 | Ding | Jan 2014 | B2 |
8628015 | Wang et al. | Jan 2014 | B2 |
8628016 | Winegar | Jan 2014 | B2 |
8629926 | Wang | Jan 2014 | B2 |
8630491 | Longacre et al. | Jan 2014 | B2 |
8635309 | Berthiaume et al. | Jan 2014 | B2 |
8636200 | Kearney | Jan 2014 | B2 |
8636212 | Nahill et al. | Jan 2014 | B2 |
8636215 | Ding et al. | Jan 2014 | B2 |
8636224 | Wang | Jan 2014 | B2 |
8638806 | Wang et al. | Jan 2014 | B2 |
8640958 | Lu et al. | Feb 2014 | B2 |
8640960 | Wang et al. | Feb 2014 | B2 |
8643717 | Li et al. | Feb 2014 | B2 |
8646692 | Meier et al. | Feb 2014 | B2 |
8646694 | Wang et al. | Feb 2014 | B2 |
8657200 | Ren et al. | Feb 2014 | B2 |
8659397 | Vargo et al. | Feb 2014 | B2 |
8668149 | Good | Mar 2014 | B2 |
8678285 | Kearney | Mar 2014 | B2 |
8678286 | Smith et al. | Mar 2014 | B2 |
8682077 | Longacre | Mar 2014 | B1 |
D702237 | Oberpriller et al. | Apr 2014 | S |
8687282 | Feng et al. | Apr 2014 | B2 |
8692927 | Pease et al. | Apr 2014 | B2 |
8695880 | Bremer et al. | Apr 2014 | B2 |
8698949 | Grunow et al. | Apr 2014 | B2 |
8702000 | Barber et al. | Apr 2014 | B2 |
8717494 | Gannon | May 2014 | B2 |
8720783 | Biss et al. | May 2014 | B2 |
8723804 | Fletcher et al. | May 2014 | B2 |
8723904 | Marty et al. | May 2014 | B2 |
8727223 | Wang | May 2014 | B2 |
8740082 | Wilz | Jun 2014 | B2 |
8740085 | Furlong et al. | Jun 2014 | B2 |
8746563 | Hennick et al. | Jun 2014 | B2 |
8750445 | Peake et al. | Jun 2014 | B2 |
8752766 | Xian et al. | Jun 2014 | B2 |
8756059 | Braho et al. | Jun 2014 | B2 |
8757495 | Qu et al. | Jun 2014 | B2 |
8760563 | Koziol et al. | Jun 2014 | B2 |
8763909 | Reed et al. | Jul 2014 | B2 |
8777108 | Coyle | Jul 2014 | B2 |
8777109 | Oberpriller et al. | Jul 2014 | B2 |
8779898 | Havens et al. | Jul 2014 | B2 |
8781520 | Payne et al. | Jul 2014 | B2 |
8783573 | Havens et al. | Jul 2014 | B2 |
8789757 | Barten | Jul 2014 | B2 |
8789758 | Hawley et al. | Jul 2014 | B2 |
8789759 | Xian et al. | Jul 2014 | B2 |
8792688 | Unsworth | Jul 2014 | B2 |
8794520 | Wang et al. | Aug 2014 | B2 |
8794522 | Ehrhart | Aug 2014 | B2 |
8794525 | Amundsen et al. | Aug 2014 | B2 |
8794526 | Wang et al. | Aug 2014 | B2 |
8798367 | Ellis | Aug 2014 | B2 |
8807431 | Wang et al. | Aug 2014 | B2 |
8807432 | Van Horn et al. | Aug 2014 | B2 |
8810779 | Hilde | Aug 2014 | B1 |
8820630 | Qu et al. | Sep 2014 | B2 |
8822848 | Meagher | Sep 2014 | B2 |
8824692 | Sheerin et al. | Sep 2014 | B2 |
8824696 | Braho | Sep 2014 | B2 |
8842849 | Wahl et al. | Sep 2014 | B2 |
8844822 | Kotlarsky et al. | Sep 2014 | B2 |
8844823 | Fritz et al. | Sep 2014 | B2 |
8849019 | Li et al. | Sep 2014 | B2 |
D716285 | Chaney et al. | Oct 2014 | S |
8851383 | Yeakley et al. | Oct 2014 | B2 |
8854633 | Laffargue | Oct 2014 | B2 |
8866963 | Grunow et al. | Oct 2014 | B2 |
8868421 | Braho et al. | Oct 2014 | B2 |
8868519 | Maloy et al. | Oct 2014 | B2 |
8868802 | Barten | Oct 2014 | B2 |
8868803 | Caballero | Oct 2014 | B2 |
8870074 | Gannon | Oct 2014 | B1 |
8879639 | Sauerwein | Nov 2014 | B2 |
8880426 | Smith | Nov 2014 | B2 |
8881983 | Havens et al. | Nov 2014 | B2 |
8881987 | Wang | Nov 2014 | B2 |
8897596 | Passmore et al. | Nov 2014 | B1 |
8903172 | Smith | Dec 2014 | B2 |
8908995 | Benos et al. | Dec 2014 | B2 |
8910870 | Li et al. | Dec 2014 | B2 |
8910875 | Ren et al. | Dec 2014 | B2 |
8914290 | Hendrickson et al. | Dec 2014 | B2 |
8914788 | Pettinelli et al. | Dec 2014 | B2 |
8915439 | Feng et al. | Dec 2014 | B2 |
8915444 | Havens et al. | Dec 2014 | B2 |
8916789 | Woodburn | Dec 2014 | B2 |
8918250 | Hollifield | Dec 2014 | B2 |
8918564 | Caballero | Dec 2014 | B2 |
8925818 | Kosecki et al. | Jan 2015 | B2 |
8928896 | Kennington et al. | Jan 2015 | B2 |
8939374 | Jovanovski et al. | Jan 2015 | B2 |
8942480 | Ellis | Jan 2015 | B2 |
8944313 | Williams et al. | Feb 2015 | B2 |
8944327 | Meier et al. | Feb 2015 | B2 |
8944332 | Harding et al. | Feb 2015 | B2 |
8950678 | Germaine et al. | Feb 2015 | B2 |
D723560 | Zhou et al. | Mar 2015 | S |
8967468 | Gomez et al. | Mar 2015 | B2 |
8971346 | Sevier | Mar 2015 | B2 |
8976030 | Cunningham et al. | Mar 2015 | B2 |
8976368 | Akel et al. | Mar 2015 | B2 |
8978981 | Guan | Mar 2015 | B2 |
8978983 | Bremer et al. | Mar 2015 | B2 |
8978984 | Hennick et al. | Mar 2015 | B2 |
8985456 | Zhu et al. | Mar 2015 | B2 |
8985457 | Soule et al. | Mar 2015 | B2 |
8985459 | Kearney et al. | Mar 2015 | B2 |
8985461 | Gelay et al. | Mar 2015 | B2 |
8988578 | Showering | Mar 2015 | B2 |
8988590 | Gillet et al. | Mar 2015 | B2 |
8991704 | Hopper et al. | Mar 2015 | B2 |
8996194 | Davis et al. | Mar 2015 | B2 |
8996384 | Funyak et al. | Mar 2015 | B2 |
8998091 | Edmonds et al. | Apr 2015 | B2 |
9002641 | Showering | Apr 2015 | B2 |
9007368 | Laffargue et al. | Apr 2015 | B2 |
9010641 | Qu et al. | Apr 2015 | B2 |
9014441 | Truyen et al. | Apr 2015 | B2 |
9015513 | Murawski et al. | Apr 2015 | B2 |
9016576 | Brady et al. | Apr 2015 | B2 |
D730357 | Fitch et al. | May 2015 | S |
9022288 | Nahill et al. | May 2015 | B2 |
9030964 | Essinger et al. | May 2015 | B2 |
9033240 | Smith et al. | May 2015 | B2 |
9033242 | Gillet et al. | May 2015 | B2 |
9036054 | Koziol et al. | May 2015 | B2 |
9037344 | Chamberlin | May 2015 | B2 |
9038911 | Xian et al. | May 2015 | B2 |
9038915 | Smith | May 2015 | B2 |
D730901 | Oberpriller et al. | Jun 2015 | S |
D730902 | Fitch et al. | Jun 2015 | S |
D733112 | Chaney et al. | Jun 2015 | S |
9047098 | Barten | Jun 2015 | B2 |
9047359 | Caballero et al. | Jun 2015 | B2 |
9047420 | Caballero | Jun 2015 | B2 |
9047525 | Barber et al. | Jun 2015 | B2 |
9047531 | Showering et al. | Jun 2015 | B2 |
9049640 | Wang et al. | Jun 2015 | B2 |
9053055 | Caballero | Jun 2015 | B2 |
9053378 | Hou et al. | Jun 2015 | B1 |
9053380 | Xian et al. | Jun 2015 | B2 |
9057641 | Amundsen et al. | Jun 2015 | B2 |
9058526 | Powilleit | Jun 2015 | B2 |
9064165 | Havens et al. | Jun 2015 | B2 |
9064167 | Xian et al. | Jun 2015 | B2 |
9064168 | Todeschini et al. | Jun 2015 | B2 |
9064254 | Todeschini et al. | Jun 2015 | B2 |
9066032 | Wang | Jun 2015 | B2 |
9070032 | Corcoran | Jun 2015 | B2 |
D734339 | Zhou et al. | Jul 2015 | S |
D734751 | Oberpriller et al. | Jul 2015 | S |
9082023 | Feng et al. | Jul 2015 | B2 |
9082195 | Holeva et al. | Jul 2015 | B2 |
9142035 | Rotman | Sep 2015 | B1 |
9171278 | Kong et al. | Oct 2015 | B1 |
9233470 | Bradski et al. | Jan 2016 | B1 |
9235899 | Kirmani et al. | Jan 2016 | B1 |
9299013 | Curlander et al. | Mar 2016 | B1 |
9424749 | Reed et al. | Aug 2016 | B1 |
9486921 | Straszheim et al. | Nov 2016 | B1 |
20010027995 | Patel et al. | Oct 2001 | A1 |
20010032879 | He et al. | Oct 2001 | A1 |
20020054289 | Thibault et al. | May 2002 | A1 |
20020067855 | Chiu et al. | Jun 2002 | A1 |
20020109835 | Goetz | Aug 2002 | A1 |
20020118874 | Chung et al. | Aug 2002 | A1 |
20020158873 | Williamson | Oct 2002 | A1 |
20020167677 | Okada et al. | Nov 2002 | A1 |
20020179708 | Zhu et al. | Dec 2002 | A1 |
20020196534 | Lizotte et al. | Dec 2002 | A1 |
20030038179 | Tsikos et al. | Feb 2003 | A1 |
20030053513 | Vatan et al. | Mar 2003 | A1 |
20030063086 | Baumberg | Apr 2003 | A1 |
20030078755 | Leutz et al. | Apr 2003 | A1 |
20030091227 | Chang et al. | May 2003 | A1 |
20030156756 | Gokturk et al. | Aug 2003 | A1 |
20030197138 | Pease et al. | Oct 2003 | A1 |
20030225712 | Cooper | Dec 2003 | A1 |
20030235331 | Kawaike et al. | Dec 2003 | A1 |
20040008259 | Gokturk et al. | Jan 2004 | A1 |
20040019274 | Galloway et al. | Jan 2004 | A1 |
20040024754 | Mane et al. | Feb 2004 | A1 |
20040066329 | Zeitfuss et al. | Apr 2004 | A1 |
20040073359 | Ichijo et al. | Apr 2004 | A1 |
20040083025 | Yamanouchi et al. | Apr 2004 | A1 |
20040089482 | Ramsden | May 2004 | A1 |
20040098146 | Katae et al. | May 2004 | A1 |
20040105580 | Hager et al. | Jun 2004 | A1 |
20040118928 | Patel et al. | Jun 2004 | A1 |
20040122779 | Stickler et al. | Jun 2004 | A1 |
20040155975 | Hart et al. | Aug 2004 | A1 |
20040165090 | Ning | Aug 2004 | A1 |
20040184041 | Schopp | Sep 2004 | A1 |
20040211836 | Patel et al. | Oct 2004 | A1 |
20040214623 | Takahashi et al. | Oct 2004 | A1 |
20040233461 | Armstrong et al. | Nov 2004 | A1 |
20040258353 | Gluckstad et al. | Dec 2004 | A1 |
20050006477 | Patel | Jan 2005 | A1 |
20050117215 | Lange | Jun 2005 | A1 |
20050128196 | Popescu et al. | Jun 2005 | A1 |
20050168488 | Montague | Aug 2005 | A1 |
20050211782 | Martin | Sep 2005 | A1 |
20050264867 | Cho et al. | Dec 2005 | A1 |
20060047704 | Gopalakrishnan | Mar 2006 | A1 |
20060078226 | Zhou | Apr 2006 | A1 |
20060108266 | Bowers | May 2006 | A1 |
20060112023 | Horhann | May 2006 | A1 |
20060151604 | Zhu et al. | Jul 2006 | A1 |
20060159307 | Anderson et al. | Jul 2006 | A1 |
20060159344 | Shao et al. | Jul 2006 | A1 |
20060213999 | Wang et al. | Sep 2006 | A1 |
20060232681 | Okada | Oct 2006 | A1 |
20060255150 | Longacre | Nov 2006 | A1 |
20060269165 | Viswanathan | Nov 2006 | A1 |
20060291719 | Ikeda et al. | Dec 2006 | A1 |
20070003154 | Sun et al. | Jan 2007 | A1 |
20070025612 | Iwasaki et al. | Feb 2007 | A1 |
20070031064 | Zhao et al. | Feb 2007 | A1 |
20070063048 | Havens et al. | Mar 2007 | A1 |
20070116357 | Dewaele | May 2007 | A1 |
20070127022 | Cohen et al. | Jun 2007 | A1 |
20070143082 | Degnan | Jun 2007 | A1 |
20070153293 | Gruhlke et al. | Jul 2007 | A1 |
20070171220 | Kriveshko | Jul 2007 | A1 |
20070177011 | Lewin et al. | Aug 2007 | A1 |
20070181685 | Zhu et al. | Aug 2007 | A1 |
20070237356 | Dwinell et al. | Oct 2007 | A1 |
20070291031 | Konev et al. | Dec 2007 | A1 |
20070299338 | Stevick et al. | Dec 2007 | A1 |
20080013793 | Hillis et al. | Jan 2008 | A1 |
20080035390 | Wurz | Feb 2008 | A1 |
20080056536 | Hildreth et al. | Mar 2008 | A1 |
20080062164 | Bassi et al. | Mar 2008 | A1 |
20080077265 | Boyden | Mar 2008 | A1 |
20080164074 | Wurz | Jul 2008 | A1 |
20080204476 | Montague | Aug 2008 | A1 |
20080212168 | Olmstead et al. | Sep 2008 | A1 |
20080247635 | Davis et al. | Oct 2008 | A1 |
20080273191 | Kim et al. | Nov 2008 | A1 |
20080273210 | Hilde | Nov 2008 | A1 |
20080278790 | Boesser et al. | Nov 2008 | A1 |
20090059004 | Bochicchio | Mar 2009 | A1 |
20090081008 | Somin et al. | Mar 2009 | A1 |
20090095047 | Patel et al. | Apr 2009 | A1 |
20090134221 | Zhu et al. | May 2009 | A1 |
20090195790 | Zhu et al. | Aug 2009 | A1 |
20090225333 | Bendall et al. | Sep 2009 | A1 |
20090237411 | Gossweiler et al. | Sep 2009 | A1 |
20090268023 | Hsieh | Oct 2009 | A1 |
20090272724 | Gubler | Nov 2009 | A1 |
20090273770 | Bauhahn et al. | Nov 2009 | A1 |
20090313948 | Buckley et al. | Dec 2009 | A1 |
20090318815 | Barnes et al. | Dec 2009 | A1 |
20090323084 | Dunn et al. | Dec 2009 | A1 |
20090323121 | Valkenburg | Dec 2009 | A1 |
20100035637 | Varanasi et al. | Feb 2010 | A1 |
20100060604 | Zwart et al. | Mar 2010 | A1 |
20100091104 | Sprigle | Apr 2010 | A1 |
20100118200 | Gelman et al. | May 2010 | A1 |
20100128109 | Banks | May 2010 | A1 |
20100161170 | Siris | Jun 2010 | A1 |
20100171740 | Andersen et al. | Jul 2010 | A1 |
20100172567 | Prokoski | Jul 2010 | A1 |
20100177076 | Essinger et al. | Jul 2010 | A1 |
20100177080 | Essinger et al. | Jul 2010 | A1 |
20100177707 | Essinger et al. | Jul 2010 | A1 |
20100177749 | Essinger et al. | Jul 2010 | A1 |
20100202702 | Benos et al. | Aug 2010 | A1 |
20100208039 | Stettner | Aug 2010 | A1 |
20100211355 | Horst et al. | Aug 2010 | A1 |
20100217678 | Goncalves | Aug 2010 | A1 |
20100220849 | Colbert et al. | Sep 2010 | A1 |
20100220894 | Ackley et al. | Sep 2010 | A1 |
20100223276 | Al-Shameri et al. | Sep 2010 | A1 |
20100245850 | Lee et al. | Sep 2010 | A1 |
20100254611 | Arnz | Oct 2010 | A1 |
20100303336 | Abraham | Dec 2010 | A1 |
20100315413 | Izadi et al. | Dec 2010 | A1 |
20100321482 | Cleveland | Dec 2010 | A1 |
20110019155 | Daniel et al. | Jan 2011 | A1 |
20110040192 | Brenner et al. | Feb 2011 | A1 |
20110043609 | Choi et al. | Feb 2011 | A1 |
20110099474 | Grossman et al. | Apr 2011 | A1 |
20110169999 | Grunow et al. | Jul 2011 | A1 |
20110188054 | Petronius et al. | Aug 2011 | A1 |
20110188741 | Sones et al. | Aug 2011 | A1 |
20110202554 | Powilleit et al. | Aug 2011 | A1 |
20110234389 | Mellin | Sep 2011 | A1 |
20110235854 | Berger et al. | Sep 2011 | A1 |
20110249864 | Venkatesan et al. | Oct 2011 | A1 |
20110254840 | Halstead | Oct 2011 | A1 |
20110279916 | Brown et al. | Nov 2011 | A1 |
20110286007 | Pangrazio et al. | Nov 2011 | A1 |
20110286628 | Goncalves et al. | Nov 2011 | A1 |
20110288818 | Thierman | Nov 2011 | A1 |
20110301994 | Tieman | Dec 2011 | A1 |
20110303748 | Lemma et al. | Dec 2011 | A1 |
20110310227 | Konertz et al. | Dec 2011 | A1 |
20120024952 | Chen | Feb 2012 | A1 |
20120056982 | Katz et al. | Mar 2012 | A1 |
20120057345 | Kuchibhotla | Mar 2012 | A1 |
20120067955 | Rowe | Mar 2012 | A1 |
20120074227 | Ferren et al. | Mar 2012 | A1 |
20120081714 | Pangrazio et al. | Apr 2012 | A1 |
20120111946 | Golant | May 2012 | A1 |
20120113223 | Hilliges et al. | May 2012 | A1 |
20120113250 | Farlotti et al. | May 2012 | A1 |
20120126000 | Kunzig et al. | May 2012 | A1 |
20120140300 | Freeman | Jun 2012 | A1 |
20120168512 | Kotlarsky et al. | Jul 2012 | A1 |
20120179665 | Baarman et al. | Jul 2012 | A1 |
20120185094 | Rosenstein et al. | Jul 2012 | A1 |
20120190386 | Anderson | Jul 2012 | A1 |
20120193423 | Samek | Aug 2012 | A1 |
20120197464 | Wang et al. | Aug 2012 | A1 |
20120203647 | Smith | Aug 2012 | A1 |
20120218436 | Rodriguez et al. | Aug 2012 | A1 |
20120223141 | Good et al. | Sep 2012 | A1 |
20120224026 | Bayer et al. | Sep 2012 | A1 |
20120236288 | Stanley | Sep 2012 | A1 |
20120242852 | Hayward et al. | Sep 2012 | A1 |
20120256901 | Bendall | Oct 2012 | A1 |
20120261474 | Kawashime et al. | Oct 2012 | A1 |
20120262558 | Boger et al. | Oct 2012 | A1 |
20120280908 | Rhoads et al. | Nov 2012 | A1 |
20120282905 | Owen | Nov 2012 | A1 |
20120282911 | Davis et al. | Nov 2012 | A1 |
20120284012 | Rodriguez et al. | Nov 2012 | A1 |
20120284122 | Brandis | Nov 2012 | A1 |
20120284339 | Rodriguez | Nov 2012 | A1 |
20120284593 | Rodriguez | Nov 2012 | A1 |
20120293610 | Doepke et al. | Nov 2012 | A1 |
20120294549 | Doepke | Nov 2012 | A1 |
20120299961 | Ramkumar et al. | Nov 2012 | A1 |
20120300991 | Free | Nov 2012 | A1 |
20120313848 | Galor et al. | Dec 2012 | A1 |
20120314030 | Datta | Dec 2012 | A1 |
20120314058 | Bendall et al. | Dec 2012 | A1 |
20120316820 | Nakazato et al. | Dec 2012 | A1 |
20130019278 | Sun et al. | Jan 2013 | A1 |
20130038881 | Pesach et al. | Feb 2013 | A1 |
20130038941 | Pesach et al. | Feb 2013 | A1 |
20130043312 | Van Horn | Feb 2013 | A1 |
20130050426 | Sarmast et al. | Feb 2013 | A1 |
20130075168 | Amundsen et al. | Mar 2013 | A1 |
20130094069 | Lee et al. | Apr 2013 | A1 |
20130101158 | Lloyd et al. | Apr 2013 | A1 |
20130156267 | Muraoka et al. | Jun 2013 | A1 |
20130175341 | Kearney et al. | Jul 2013 | A1 |
20130175343 | Good | Jul 2013 | A1 |
20130200150 | Reynolds et al. | Aug 2013 | A1 |
20130201288 | Billerbaeck et al. | Aug 2013 | A1 |
20130208164 | Cazier et al. | Aug 2013 | A1 |
20130211790 | Loveland et al. | Aug 2013 | A1 |
20130223673 | Davis et al. | Aug 2013 | A1 |
20130257744 | Daghigh et al. | Oct 2013 | A1 |
20130257759 | Daghigh | Oct 2013 | A1 |
20130270346 | Xian et al. | Oct 2013 | A1 |
20130287258 | Kearney | Oct 2013 | A1 |
20130291998 | Konnerth | Nov 2013 | A1 |
20130292475 | Kotlarsky et al. | Nov 2013 | A1 |
20130292477 | Hennick et al. | Nov 2013 | A1 |
20130293539 | Hunt et al. | Nov 2013 | A1 |
20130293540 | Laffargue et al. | Nov 2013 | A1 |
20130306728 | Thuries et al. | Nov 2013 | A1 |
20130306731 | Pedraro | Nov 2013 | A1 |
20130307964 | Bremer et al. | Nov 2013 | A1 |
20130308013 | Li et al. | Nov 2013 | A1 |
20130308625 | Park et al. | Nov 2013 | A1 |
20130313324 | Koziol et al. | Nov 2013 | A1 |
20130313325 | Wilz et al. | Nov 2013 | A1 |
20130329012 | Bartos | Dec 2013 | A1 |
20130329013 | Metois et al. | Dec 2013 | A1 |
20130342342 | Sabre et al. | Dec 2013 | A1 |
20130342343 | Harring et al. | Dec 2013 | A1 |
20130342717 | Havens et al. | Dec 2013 | A1 |
20140001267 | Giordano et al. | Jan 2014 | A1 |
20140002828 | Laffargue et al. | Jan 2014 | A1 |
20140008439 | Wang | Jan 2014 | A1 |
20140009586 | McNamer et al. | Jan 2014 | A1 |
20140021259 | Moed et al. | Jan 2014 | A1 |
20140025584 | Liu et al. | Jan 2014 | A1 |
20140031665 | Pinto et al. | Jan 2014 | A1 |
20140034731 | Gao et al. | Feb 2014 | A1 |
20140034734 | Sauerwein | Feb 2014 | A1 |
20140036848 | Pease et al. | Feb 2014 | A1 |
20140039693 | Havens et al. | Feb 2014 | A1 |
20140042814 | Kather et al. | Feb 2014 | A1 |
20140049120 | Kohtz et al. | Feb 2014 | A1 |
20140049635 | Laffargue | Feb 2014 | A1 |
20140058612 | Wong et al. | Feb 2014 | A1 |
20140061306 | Wu et al. | Mar 2014 | A1 |
20140062709 | Hyer et al. | Mar 2014 | A1 |
20140063289 | Hussey et al. | Mar 2014 | A1 |
20140064624 | Kim et al. | Mar 2014 | A1 |
20140066136 | Sauerwein et al. | Mar 2014 | A1 |
20140067104 | Osterhout | Mar 2014 | A1 |
20140067692 | Ye et al. | Mar 2014 | A1 |
20140070005 | Nahill et al. | Mar 2014 | A1 |
20140071430 | Hansen et al. | Mar 2014 | A1 |
20140071840 | Venancio | Mar 2014 | A1 |
20140074746 | Wang | Mar 2014 | A1 |
20140076974 | Havens et al. | Mar 2014 | A1 |
20140078341 | Havens et al. | Mar 2014 | A1 |
20140078342 | Li et al. | Mar 2014 | A1 |
20140078345 | Showering | Mar 2014 | A1 |
20140091147 | Evans et al. | Apr 2014 | A1 |
20140097238 | Ghazizadeh | Apr 2014 | A1 |
20140098091 | Hori | Apr 2014 | A1 |
20140098792 | Wang et al. | Apr 2014 | A1 |
20140100774 | Showering | Apr 2014 | A1 |
20140100813 | Showering | Apr 2014 | A1 |
20140103115 | Meier et al. | Apr 2014 | A1 |
20140104413 | McCloskey et al. | Apr 2014 | A1 |
20140104414 | McCloskey et al. | Apr 2014 | A1 |
20140104416 | Giordano et al. | Apr 2014 | A1 |
20140104451 | Todeschini et al. | Apr 2014 | A1 |
20140104664 | Lee | Apr 2014 | A1 |
20140106594 | Skvoretz | Apr 2014 | A1 |
20140106725 | Sauerwein | Apr 2014 | A1 |
20140108010 | Maltseff et al. | Apr 2014 | A1 |
20140108402 | Gomez et al. | Apr 2014 | A1 |
20140108682 | Caballero | Apr 2014 | A1 |
20140110485 | Toa et al. | Apr 2014 | A1 |
20140114530 | Fitch et al. | Apr 2014 | A1 |
20140124577 | Wang et al. | May 2014 | A1 |
20140124579 | Ding | May 2014 | A1 |
20140125842 | Winegar | May 2014 | A1 |
20140125853 | Wang | May 2014 | A1 |
20140125999 | Longacre et al. | May 2014 | A1 |
20140129378 | Richardson | May 2014 | A1 |
20140131438 | Kearney | May 2014 | A1 |
20140131441 | Nahill et al. | May 2014 | A1 |
20140131443 | Smith | May 2014 | A1 |
20140131444 | Wang | May 2014 | A1 |
20140131445 | Ding et al. | May 2014 | A1 |
20140131448 | Xian et al. | May 2014 | A1 |
20140133379 | Wang et al. | May 2014 | A1 |
20140135984 | Hirata | May 2014 | A1 |
20140136208 | Maltseff et al. | May 2014 | A1 |
20140139654 | Taskahashi | May 2014 | A1 |
20140140585 | Wang | May 2014 | A1 |
20140151453 | Meier et al. | Jun 2014 | A1 |
20140152882 | Samek et al. | Jun 2014 | A1 |
20140152975 | Ko | Jun 2014 | A1 |
20140158468 | Adami | Jun 2014 | A1 |
20140158770 | Sevier et al. | Jun 2014 | A1 |
20140159869 | Zumsteg et al. | Jun 2014 | A1 |
20140166755 | Liu et al. | Jun 2014 | A1 |
20140166757 | Smith | Jun 2014 | A1 |
20140166759 | Liu et al. | Jun 2014 | A1 |
20140168380 | Heidemann et al. | Jun 2014 | A1 |
20140168787 | Wang et al. | Jun 2014 | A1 |
20140175165 | Havens et al. | Jun 2014 | A1 |
20140175172 | Jovanovski et al. | Jun 2014 | A1 |
20140191644 | Chaney | Jul 2014 | A1 |
20140191913 | Ge et al. | Jul 2014 | A1 |
20140192187 | Atwell et al. | Jul 2014 | A1 |
20140192551 | Masaki | Jul 2014 | A1 |
20140197238 | Liu et al. | Jul 2014 | A1 |
20140197239 | Havens et al. | Jul 2014 | A1 |
20140197304 | Feng et al. | Jul 2014 | A1 |
20140203087 | Smith et al. | Jul 2014 | A1 |
20140204268 | Grunow et al. | Jul 2014 | A1 |
20140205150 | Ogawa | Jul 2014 | A1 |
20140214631 | Hansen | Jul 2014 | A1 |
20140217166 | Berthiaume et al. | Aug 2014 | A1 |
20140217180 | Liu | Aug 2014 | A1 |
20140225918 | Mittal et al. | Aug 2014 | A1 |
20140225985 | Klusza et al. | Aug 2014 | A1 |
20140231500 | Ehrhart et al. | Aug 2014 | A1 |
20140232930 | Anderson | Aug 2014 | A1 |
20140240454 | Hirata et al. | Aug 2014 | A1 |
20140247279 | Nicholas et al. | Sep 2014 | A1 |
20140247280 | Nicholas et al. | Sep 2014 | A1 |
20140247315 | Marty et al. | Sep 2014 | A1 |
20140263493 | Amurgis et al. | Sep 2014 | A1 |
20140263645 | Smith et al. | Sep 2014 | A1 |
20140267609 | Laffargue | Sep 2014 | A1 |
20140268093 | Tohme et al. | Sep 2014 | A1 |
20140270196 | Braho et al. | Sep 2014 | A1 |
20140270229 | Braho | Sep 2014 | A1 |
20140270361 | Amma et al. | Sep 2014 | A1 |
20140278387 | DiGregorio | Sep 2014 | A1 |
20140282210 | Bianconi | Sep 2014 | A1 |
20140284384 | Lu et al. | Sep 2014 | A1 |
20140288933 | Braho et al. | Sep 2014 | A1 |
20140297058 | Barker et al. | Oct 2014 | A1 |
20140299665 | Barber et al. | Oct 2014 | A1 |
20140306833 | Ricci | Oct 2014 | A1 |
20140307855 | Withagen et al. | Oct 2014 | A1 |
20140312121 | Lu et al. | Oct 2014 | A1 |
20140313527 | Askan | Oct 2014 | A1 |
20140319219 | Liu et al. | Oct 2014 | A1 |
20140319220 | Coyle | Oct 2014 | A1 |
20140319221 | Oberpriller et al. | Oct 2014 | A1 |
20140320408 | Zagorsek et al. | Oct 2014 | A1 |
20140326787 | Barten | Nov 2014 | A1 |
20140332590 | Wang et al. | Nov 2014 | A1 |
20140333775 | Naikal et al. | Nov 2014 | A1 |
20140344943 | Todeschini et al. | Nov 2014 | A1 |
20140346233 | Liu et al. | Nov 2014 | A1 |
20140347553 | Ovsiannikov et al. | Nov 2014 | A1 |
20140350710 | Gopalkrishnan et al. | Nov 2014 | A1 |
20140351317 | Smith et al. | Nov 2014 | A1 |
20140353373 | Van Horn et al. | Dec 2014 | A1 |
20140361073 | Qu et al. | Dec 2014 | A1 |
20140361082 | Xian et al. | Dec 2014 | A1 |
20140362184 | Jovanovski et al. | Dec 2014 | A1 |
20140363015 | Braho | Dec 2014 | A1 |
20140369511 | Sheerin et al. | Dec 2014 | A1 |
20140374483 | Lu | Dec 2014 | A1 |
20140374485 | Xian et al. | Dec 2014 | A1 |
20140379613 | Nishitani et al. | Dec 2014 | A1 |
20150001301 | Ouyang | Jan 2015 | A1 |
20150001304 | Todeschini | Jan 2015 | A1 |
20150003673 | Fletcher | Jan 2015 | A1 |
20150009100 | Haneda et al. | Jan 2015 | A1 |
20150009301 | Ribnick et al. | Jan 2015 | A1 |
20150009338 | Laffargue et al. | Jan 2015 | A1 |
20150009610 | London et al. | Jan 2015 | A1 |
20150014416 | Kotlarsky et al. | Jan 2015 | A1 |
20150021397 | Rueblinger et al. | Jan 2015 | A1 |
20150028102 | Ren et al. | Jan 2015 | A1 |
20150028103 | Jiang | Jan 2015 | A1 |
20150028104 | Ma et al. | Jan 2015 | A1 |
20150029002 | Yeakley et al. | Jan 2015 | A1 |
20150032709 | Maloy et al. | Jan 2015 | A1 |
20150036876 | Marrion et al. | Feb 2015 | A1 |
20150039309 | Braho et al. | Feb 2015 | A1 |
20150040378 | Saber et al. | Feb 2015 | A1 |
20150042791 | Metois et al. | Feb 2015 | A1 |
20150048168 | Fritz et al. | Feb 2015 | A1 |
20150049347 | Laffargue et al. | Feb 2015 | A1 |
20150051992 | Smith | Feb 2015 | A1 |
20150053766 | Havens et al. | Feb 2015 | A1 |
20150053768 | Wang et al. | Feb 2015 | A1 |
20150053769 | Thuries et al. | Feb 2015 | A1 |
20150062366 | Liu et al. | Mar 2015 | A1 |
20150062369 | Gehring et al. | Mar 2015 | A1 |
20150063215 | Wang | Mar 2015 | A1 |
20150063676 | Lloyd et al. | Mar 2015 | A1 |
20150069130 | Gannon | Mar 2015 | A1 |
20150071819 | Todeschini | Mar 2015 | A1 |
20150083800 | Li et al. | Mar 2015 | A1 |
20150086114 | Todeschini | Mar 2015 | A1 |
20150088522 | Hendrickson et al. | Mar 2015 | A1 |
20150096872 | Woodburn | Apr 2015 | A1 |
20150099557 | Pettinelli et al. | Apr 2015 | A1 |
20150100196 | Hollifield | Apr 2015 | A1 |
20150102109 | Huck | Apr 2015 | A1 |
20150115035 | Meier et al. | Apr 2015 | A1 |
20150116498 | Vartiainen et al. | Apr 2015 | A1 |
20150127791 | Kosecki et al. | May 2015 | A1 |
20150128116 | Chen et al. | May 2015 | A1 |
20150129659 | Feng et al. | May 2015 | A1 |
20150133047 | Smith et al. | May 2015 | A1 |
20150134470 | Hejl et al. | May 2015 | A1 |
20150136851 | Harding et al. | May 2015 | A1 |
20150136854 | Lu et al. | May 2015 | A1 |
20150142492 | Kumar | May 2015 | A1 |
20150144692 | Hejl | May 2015 | A1 |
20150144698 | Teng et al. | May 2015 | A1 |
20150144701 | Xian et al. | May 2015 | A1 |
20150149946 | Benos et al. | May 2015 | A1 |
20150161429 | Xian | Jun 2015 | A1 |
20150163474 | You | Jun 2015 | A1 |
20150169925 | Chen et al. | Jun 2015 | A1 |
20150169929 | Williams et al. | Jun 2015 | A1 |
20150186703 | Chen et al. | Jul 2015 | A1 |
20150193644 | Kearney et al. | Jul 2015 | A1 |
20150193645 | Colavito et al. | Jul 2015 | A1 |
20150199957 | Funyak et al. | Jul 2015 | A1 |
20150204662 | Kobayashi et al. | Jul 2015 | A1 |
20150204671 | Showering | Jul 2015 | A1 |
20150213647 | Laffargue et al. | Jul 2015 | A1 |
20150229838 | Hakim et al. | Aug 2015 | A1 |
20150269403 | Lei et al. | Sep 2015 | A1 |
20150276379 | Ni et al. | Oct 2015 | A1 |
20150301181 | Herschbach | Oct 2015 | A1 |
20150308816 | Laffargue et al. | Oct 2015 | A1 |
20150325036 | Lee | Nov 2015 | A1 |
20150355470 | Herschbach | Dec 2015 | A1 |
20160040982 | Li et al. | Feb 2016 | A1 |
20160048725 | Holz et al. | Feb 2016 | A1 |
20160063429 | Varley et al. | Mar 2016 | A1 |
20160088287 | Sadi et al. | Mar 2016 | A1 |
20160090283 | Svensson et al. | Mar 2016 | A1 |
20160090284 | Svensson et al. | Mar 2016 | A1 |
20160101936 | Chamberlin | Apr 2016 | A1 |
20160102975 | McCloskey et al. | Apr 2016 | A1 |
20160104019 | Todeschini et al. | Apr 2016 | A1 |
20160104274 | Jovanovski et al. | Apr 2016 | A1 |
20160109219 | Ackley et al. | Apr 2016 | A1 |
20160109220 | Laffargue et al. | Apr 2016 | A1 |
20160109224 | Thuries et al. | Apr 2016 | A1 |
20160112631 | Ackley et al. | Apr 2016 | A1 |
20160112643 | Laffargue et al. | Apr 2016 | A1 |
20160138247 | Conway et al. | May 2016 | A1 |
20160138248 | Conway et al. | May 2016 | A1 |
20160138249 | Svensson et al. | May 2016 | A1 |
20160169665 | Deschenes et al. | Jun 2016 | A1 |
20160187186 | Coleman et al. | Jun 2016 | A1 |
20160187210 | Coleman et al. | Jun 2016 | A1 |
20160191801 | Sivan | Jun 2016 | A1 |
20160202478 | Masson et al. | Jul 2016 | A1 |
20160343176 | Ackley | Nov 2016 | A1 |
20170115490 | Hsieh et al. | Apr 2017 | A1 |
20170182942 | Hardy et al. | Jun 2017 | A1 |
Number | Date | Country |
---|---|---|
2004212587 | Apr 2005 | AU |
3335760 | Apr 1985 | DE |
10210813 | Oct 2003 | DE |
102007037282 | Mar 2008 | DE |
1111435 | Jun 2001 | EP |
1443312 | Aug 2004 | EP |
2286932 | Feb 2011 | EP |
2381421 | Oct 2011 | EP |
2533009 | Dec 2012 | EP |
2722656 | Apr 2014 | EP |
2779027 | Sep 2014 | EP |
2833323 | Feb 2015 | EP |
2843590 | Mar 2015 | EP |
2845170 | Mar 2015 | EP |
2966595 | Jan 2016 | EP |
3006893 | Mar 2016 | EP |
3012601 | Mar 2016 | EP |
3007096 | Apr 2016 | EP |
2503978 | Jan 2014 | GB |
2525053 | Oct 2015 | GB |
2531928 | May 2016 | GB |
H04129902 | Apr 1992 | JP |
200696457 | Apr 2006 | JP |
2007084162 | Apr 2007 | JP |
2008210276 | Sep 2008 | JP |
2014210646 | Nov 2014 | JP |
2015174705 | Oct 2015 | JP |
20110013200 | Feb 2011 | KR |
20110117020 | Oct 2011 | KR |
20120028109 | Mar 2012 | KR |
9640452 | Dec 1996 | WO |
0077726 | Dec 2000 | WO |
0114836 | Mar 2001 | WO |
2006095110 | Sep 2006 | WO |
2007015059 | Feb 2007 | WO |
2011017241 | Feb 2011 | WO |
2012175731 | Dec 2012 | WO |
2013021157 | Feb 2013 | WO |
2013033442 | Mar 2013 | WO |
2013163789 | Nov 2013 | WO |
2013166368 | Nov 2013 | WO |
2013173985 | Nov 2013 | WO |
2013184340 | Dec 2013 | WO |
2014019130 | Feb 2014 | WO |
2014023697 | Feb 2014 | WO |
2014102341 | Jul 2014 | WO |
2014110495 | Jul 2014 | WO |
2014149702 | Sep 2014 | WO |
2014151746 | Sep 2014 | WO |
2015006865 | Jan 2015 | WO |
2016020038 | Feb 2016 | WO |
2016061699 | Apr 2016 | WO |
Entry |
---|
Office Action in counterpart European Application No. 13186043.9 dated Sep. 30, 2015, pp. 1-7. |
Lloyd et al., “System for Monitoring the Condition of Packages Throughout Transit”, U.S. Appl. No. 14/865,575, filed Sep. 25, 2015, 59 pages, not yet published. |
McCloskey et al., “Image Transformation for Indicia Reading,” U.S. Appl. No. 14/928,032, filed Oct. 30, 2015, 48 pages, not yet published. |
Great Britain Combined Search and Examination Report in related Application GB1517842.9, dated Apr. 8, 2016, 8 pages. |
Search Report in counterpart European Application No. 15182675.7, dated Dec. 4, 2015, 10 pages. |
Wikipedia, “3D projection” Downloaded on Nov. 25, 2015 from www.wikipedia.com, 4 pages. |
M.Zahid Gurbuz, Selim Akyokus, Ibrahim Emiroglu, Aysun Guran, An Efficient Algorithm for 3D Rectangular Box Packing, 2009, Applied Automatic Systems: Proceedings of Selected AAS 2009 Papers, pp. 131-134. |
European Extended Search Report in Related EP Application No. 16172995.9, dated Aug. 22, 2016, 11 pages. |
European Extended search report in related EP Application No. 15190306.9, dated Sep. 9, 2016, 15 pages. |
Collings et al., “The Applications and Technology of Phase-Only Liquid Crystal on Silicon Devices”, Journal of Display Technology, IEEE Service Center, New, York, NY, US, vol. 7, No. 3, Mar. 1, 2011 (Mar. 1, 2011), pp. 112-119. |
European extended Search report in related EP Application 13785171.3, dated Sep. 19, 2016, 8 pages. |
El-Hakim et al., “Multicamera vision-based approach to flexible feature measurement for inspection and reverse engineering”, published in Optical Engineering, Society of Photo-Optical Instrumentation Engineers, vol. 32, No. 9, Sep. 1, 1993, 15 pages. |
El-Hakim et al., “A Knowledge-based Edge/Object Measurement Technique”, Retrieved from the Internet: URL: https://www.researchgate.net/profile/Sabry—E1 -Hakim/publication/44075058—A—Knowledge—Based—EdgeObject—Measurement—Technique/links/00b4953b5faa7d3304000000.pdf [retrieved on Jul. 15, 2016] dated Jan. 1, 1993, 9 pages. |
H. Sprague Ackley, “Automatic Mode Switching in a Volume Dimensioner”, U.S. Appl. No. 15/182,636, filed Jun. 15, 2016, 53 pages, Not yet published. |
Bosch Tool Corporation, “Operating/Safety Instruction for DLR 130”, Dated Feb. 2, 2009, 36 pages. |
European Search Report for related EP Application No. 16152477.2, dated May 24, 2016, 8 pages. |
Mike Stensvold, “get the Most Out of Variable Aperture Lenses”, published on www.OutdoorPhotogrpaher.com; dated Dec. 7, 2010; 4 pages, [As noted on search report retrieved from URL: http;//www.outdoorphotographer.com/gear/lenses/get-the-most-out-ofvariable-aperture-lenses.html on Feb. 9, 2016]. |
Houle et al., “Vehical Positioning and Object Avoidance”, U.S. Appl. No. 15/007,522 [not yet published], filed Jan. 27, 2016, 59 pages. |
United Kingdom combined Search and Examination Report in related GB Application No. 1607394.2, dated Oct. 19, 2016, 7 pages. |
European Search Report from related EP Application No. 16168216.6, dated Oct. 20, 2016, 8 pages. |
Peter Clarke, Actuator Developer Claims Anti-Shake Breakthrough for Smartphone Cams, Electronic Engineering Times, p. 24, May 16, 2011. |
Spiller, Jonathan; Object Localization Using Deformable Templates, Master's Dissertation, University of the Witwatersrand, Johannesburg, South Africa, 2007; 74 pages. |
Leotta, Matthew J.; Joseph L. Mundy; Predicting High Resolution Image Edges with a Generic, Adaptive, 3-D Vehicle Model; IEEE Conference on Computer Vision and Pattern Recognition, 2009; 8 pages. |
European Search Report for application No. EP13186043 dated Feb. 26, 2014 (now EP2722656 (Apr. 23, 2014)): Total pp. 7. |
International Search Report for PCT/US2013/039438 (WO2013166368), dated Oct. 1, 2013, 7 pages. |
Lloyd, Ryan and Scott McCloskey, “Recognition of 3D Package Shapes for Singe Camera Metrology” IEEE Winter Conference on Applications of computer Visiona, IEEE, Mar. 24, 2014, pp. 99-106, {retrieved on Jun. 16, 2014}, Authors are employees of common Applicant. |
European Office Action for application EP 13186043, dated Jun. 12, 2014(now EP2722656 (Apr. 23, 2014)), Total of 6 pages. |
Zhang, Zhaoxiang; Tieniu Tan, Kaiqi Huang, Yunhong Wang; Three-Dimensional Deformable-Model-based Localization and Recognition of Road Vehicles; IEEE Transactions on Image Processing, vol. 21, No. 1, Jan. 2012, 13 pages. |
U.S. Appl. No. 14/801,023, Tyler Doomenbal et al., filed Jul. 16, 2015, not published yet, Adjusting Dimensioning Results Using Augmented Reality, 39 pages. |
Wikipedia, YUV description and definition, downloaded from http://www.wikipeida.org/wiki/YUV on Jun. 29, 2012, 10 pages. |
YUV Pixel Format, downloaded from http://www.fource.org/yuv.php on Jun. 29, 2012; 13 pages. |
YUV to RGB Conversion, downloaded from http://www.fource.org/fccyvrgb.php on Jun. 29, 2012; 5 pages. |
Benos et al., “Semi-Automatic Dimensioning with Imager of a Portable Device,” U.S. Appl. No. 61/149,912, filed Feb. 4, 2009 (now expired), 56 pages. |
Dimensional Weight—Wikipedia, the Free Encyclopedia, URL=http://en.wikipedia.org/wiki/Dimensional—weight, download date Aug. 1, 2008, 2 pages. |
Dimensioning—Wikipedia, the Free Encyclopedia, URL=http://en.wikipedia.org/wiki/Dimensioning, download date Aug. 1, 2008, 1 page. |
European Patent Office Action for Application No. 14157971.4-1906, dated Jul. 16, 2014, 5 pages. |
European Patent Search Report for Application No. 14157971.4-1906, dated Jun. 30, 2014, 6 pages. |
Caulier, Yannick et al., “A New Type of Color-Coded Light Structures for an Adapted and Rapid Determination of Point Correspondences for 3D Reconstruction.” Proc. of SPIE, vol. 8082 808232-3; 2011; 8 pages. |
Kazantsev, Aleksei et al. “Robust Pseudo-Random Coded Colored STructured Light Techniques for 3D Object Model Recovery”; ROSE 2008 IEEE International Workshop on Robotic and Sensors Environments (Oct. 17-18, 2008) , 6 pages. |
Mouaddib E. et al. “Recent Progress in Structured Light in order to Solve the Correspondence Problem in Stereo Vision” Proceedings of the 1997 IEEE International Conference on Robotics and Automation, Apr. 1997; 7 pages. |
Proesmans, Marc et al. “Active Acquisition of 3D Shape for Moving Objects” 0-7803-3258-X/96 1996 IEEE; 4 pages. |
Salvi, Joaquim et al. “Pattern Codification Strategies in Structured Light Systems” published in Pattern Recognition; The Journal of the Pattern Recognition Society, Received Mar. 6, 2003; Accepted Oct. 2, 2003; 23 pages. |
EP Search and Written Opinion Report in related matter EP Application No. 14181437.6, dated Mar. 26, 2015, 7 pages. |
Hetzel, Gunter et al.; “3D Object Recognition from Range Images using Local Feature Histograms,”, Proceedings 2001 IEEE Conference on Computer Vision and Pattern Recognition. CVPR 2001. Kauai, Hawaii, Dec. 8-14, 2001; pp. 394-399, XP010584149, ISBN: 978-0-7695-1272-3. |
Second Chinese Office Action in related CN Application No. 201520810685.6, dated Mar. 22, 2016, 5 pages, no references. |
European Search Report in related EP Application No. 15190315.0, dated Apr. 1, 2016, 7 pages. |
Second Chinese Office Action in related CN Application No. 2015220810562.2, dated Mar. 22, 2016, 5 pages. English Translation provided [No references]. |
European Search Report for related Application EP 15190249.1, dated Mar. 22, 2016, 7 pages. |
Second Chinese Office Action in related CN Application No. 201520810313.3, dated Mar. 22, 2016, 5 pages. English Translation provided [No references]. |
U.S. Appl. No. 14/800,757 , Eric Todeschini, filed Jul. 16, 2015, not published yet, Dimensioning and Imaging Items, 80 pages. |
U.S. Appl. No. 14/747,197, Serge Thuries et al., filed Jun. 23, 2015, not published yet, Optical Pattern Projector; 33 pages. |
U.S. Appl. No. 14/747,490, Brian L. Jovanovski et al., filed Jun. 23, 2015, not published yet, Dual-Projector Three-Dimensional Scanner; 40 pages. |
Search Report and Opinion in related GB Application No. 1517112.7, dated Feb. 19, 2016, 6 Pages. |
U.S. Appl. No. 14/793,149, H. Sprague Ackley, filed Jul. 7, 2015, not published yet, Mobile Dimensioner Apparatus for Use in Commerce; 57 pages. |
U.S. Appl. No. 14/740,373, H. Sprague Ackley et al., filed Jun. 16, 2015, not published yet, Calibrating a Volume Dimensioner; 63 pages. |
Intention to Grant in counterpart European Application No. 14157971.4 dated Apr. 14, 2015, pp. 1-8. |
Decision to Grant in counterpart European Application No. 14157971.4 dated Aug. 6, 2015, pp. 1-2. |
Leotta, Matthew, Generic, Deformable Models for 3-D Vehicle Surveillance, May 2010, Doctoral Dissertation, Brown University, Providence RI, 248 pages. |
Ward, Benjamin, Interactive 3D Reconstruction from Video, Aug. 2012, Doctoral Thesis, Univesity of Adelaide, Adelaide, South Australia, 157 pages. |
Hood, Frederick W.; William A. Hoff, Robert King, Evaluation of an Interactive Technique for Creating Site Models from Range Data, Apr. 27-May 1, 1997 Proceedings of the ANS 7th Topical Meeting on Robotics & Remote Systems, Augusta GA, 9 pages. |
Gupta, Alok; Range Image Segmentation for 3-D Objects Recognition, May 1988, Technical Reports (CIS), Paper 736, University of Pennsylvania Department of Computer and Information Science, retrieved from Http://repository.upenn.edu/cis—reports/736, Accessed May 31, 2015, 157 pages. |
Reisner-Kollmann,Irene; Anton L. Fuhrmann, Werner Purgathofer, Interactive Reconstruction of Industrial Sites Using Parametric Models, May 2010, Proceedings of the 26th Spring Conference of Computer Graphics SCCG 10, 8 pages. |
Drummond, Tom; Roberto Cipolla, Real-Time Visual Tracking of Complex Structures, Jul. 2002, IEEE Transactions on Pattern Analysis and Machine Intelligence, vol. 24, No. 7; 15 pages. |
European Search Report for Related EP Application No. 15189214.8, dated Mar. 3, 2016, 9 pages. |
Santolaria et al. “A one-step intrinsic and extrinsic calibration method for laster line scanner operation in coordinate measuring machines”, dated Apr. 1, 2009, Measurement Science and Technology, IOP, Bristol, GB, vol. 20, No. 4; 12 pages. |
Search Report and Opinion in Related EP Application 15176943.7, dated Jan. 8, 2016, 8 pages. |
European Search Report for related EP Application No. 15188440.0, dated Mar. 8, 2016, 8 pages. |
United Kingdom Search Report in related application GB1517842.9, dated Apr. 8, 2016, 8 pages. |
Great Britain Search Report for related Application on. GB1517843.7, dated Feb. 23, 2016; 8 pages. |
Padzensky, Ron; “Augmera; Gesture Control”, Dated Apr. 18, 2015, 15 pages. |
Grabowski, Ralph; “New Commands in AutoCADS 2010: Part 11 Smoothing 3D Mesh Objects” Dated 2011, 6 pages. |
Theodoropoulos, Gabriel; “Using Gesture Recognizers to Handle Pinch, Rotate, Pan, Swipe, and Tap Gestures” dated Aug. 25, 2014, 34 pages. |
European Extended Search Report in related EP Application No. 16190017.0, dated Jan. 4, 2017, 6 pages. |
European Extended Search Report in related EP Application No. 16173429.8, dated Dec. 1, 2016, 8 pages [Only new references cited: US 2013/0038881 was previously cited]. |
Extended European Search Report in related EP Application No. 16175410.0, dated Dec. 13, 2016, 5 pages. |
Wikipedia, “Microlens”, Downloaded from https://en.wikipedia.org/wiki/Microlens, pp. 3. {Feb. 9, 2017 Final Office Action in related matter}. |
Fukaya et al., “Characteristics of Speckle Random Pattern and Its Applications”, pp. 317-327, Nouv. Rev. Optique, t.6, n.6. (1975) {Feb. 9, 2017 Final Office Action in related matter: downloaded Mar. 2, 2017 from http://iopscience.iop.org}. |
European Examination report in related EP Application No. 14181437.6, dated Feb. 8, 2017, 5 pages. [References have been previously cited]. |
European extended search report in related EP Application 16190833.0, dated Mar. 9, 2017, 8 pages. [only new art has been cited; US Publication 2014/0034731 was previously cited]. |
United Kingdom Combined Search and Examination Report in related Application No. GB1620676.5, dated Mar. 8, 2017, 6 pages [References have been previously cited; WO2014/151746, WO2012/175731, US 2014/0313527, GB2503978]. |
European Exam Report in related , EP Application No. 16168216.6, dated Feb. 27, 2017, 5 pages, [References have been previously cited; WO2011/017241 and US 2014/0104413]. |
Thorlabs, Examiner Cited NPL in Advisory Action dated Apr. 12, 2017 in related commonly owned application, downloaded from https://www.thorlabs.com/newgrouppage9.cfm?objectgroup—id=6430, 4 pages. |
Eksma Optics, Examiner Cited NPL in Advisory Action dated Apr. 12, 2017 in related commonly owned application, downloaded from http://eksmaoptics.com/optical-systems/f-theta-lenses/f-theta-lens-for-1064-nm/, 2 pages. |
Sill Optics, Examiner Cited NPL in Advisory Action dated Apr. 12, 2017 in related commonly owned application, http://www.silloptics.de/1/products/sill-encyclopedia/laser-optics/f-theta-lenses/, 4 pages. |
Chinese Notice of Reexamination in related Chinese Application 201520810313.3, dated Mar. 14, 2017, English computer Translation provided, 7 pages. [No new art cited]. |
Extended European search report in related EP Application 16199707.7, dated Apr. 10, 2017, 15 pages. |
Ulusoy et al., One-Shot Scanning using De Bruijn Spaced Grids, 2009 IEEE 12th International Conference on Computer Vision Workshops, ICCV Workshops, 7 pages [Cited in EP Extended search report dated Apr. 10, 2017]. |
U.S. Appl. No. 14/519,179 for Dimensioning System With Multipath Interference Mitigation filed Oct. 21, 2014 (Thuries et al.); 30 pages. |
U.S. Appl. No. 14/264,173 for Autofocus Lens System for Indicia Readers filed Apr. 29, 2014, (Ackley et al.); 39 pages. |
U.S. Appl. No. 14/453,019 for Dimensioning System With Guided Alignment, filed Aug. 6, 2014 (Li et al.); 31 pages. |
U.S. Appl. No. 14/452,697 for Interactive Indicia Reader , filed Aug. 6, 2014, (Todeschini); 32 pages |
U.S. Appl. No. 14/231,898 for Hand-Mounted Indicia-Reading Device with Finger Motion Triggering filed Apr. 1, 2014 (Van Horn et al.); 36 pages. |
U.S. Appl. No. 14/715,916 for Evaluating Image Values filed May 19, 2015 (Ackley); 60 pages. |
U.S. Appl. No. 14/513,808 for Identifying Inventory Items in a Storage Facility filed Oct. 14, 2014 (Singel et al.); 51 pages. |
U.S. Appl. No. 29/458,405 for an Electronic Device, filed Jun. 19, 2013 (Fitch et al.); 22 pages. |
U.S. Appl. No. 29/459,620 for an Electronic Device Enclosure, filed Jul. 2, 2013 (London et al.); 21 pages. |
U.S. Appl. No. 14/483,056 for Variable Depth of Field Barcode Scanner filed Sep. 10, 2014 (McCloskey et al.); 29 pages. |
U.S. Appl. No. 14/531,154 for Directing an Inspector Through an Inspection filed Nov. 3, 2014 (Miller et al.); 53 pages. |
U.S. Appl. No. 29/525,068 for Tablet Computer With Removable Scanning Device filed Apr. 27, 2015 (Schulte et al.); 19 pages. |
U.S. Appl. No. 29/468,118 for an Electronic Device Case, filed Sep. 26, 2013 (Oberpriller et al.); 44 pages. |
U.S. Appl. No. 14/340,627 for an Axially Reinforced Flexible Scan Element, filed Jul. 25, 2014 (Reublinger et al.); 41 pages. |
U.S. Appl. No. 14/676,327 for Device Management Proxy for Secure Devices filed Apr. 1, 2015 (Yeakley et al.); 50 pages. |
U.S. Appl. No. 14/257,364 for Docking System and Method Using Near Field Communication filed Apr. 21, 2014 (Showering); 31 pages. |
U.S. Appl. No. 14/327,827 for a Mobile-Phone Adapter for Electronic Transactions, filed Jul. 10, 2014 (Hejl); 25 pages. |
U.S. Appl. No. 14/334,934 for a System and Method for Indicia Verification, filed Jul. 18, 2014 (Hejl); 38 pages. |
U.S. Appl. No. 29/530,600 for Cyclone filed Jun. 18, 2015 (Vargo et al); 16 pages. |
U.S. Appl. No. 14/707,123 for Application Independent DEX/UCS Interface filed May 8, 2015 (Pape); 47 pages. |
U.S. Appl. No. 14/283,282 for Terminal Having Illumination and Focus Control filed May 21, 2014 (Liu et al.); 31 pages. |
U.S. Appl. No. 14/619,093 for Methods for Training a Speech Recognition System filed Feb. 11, 2015 (Pecorari); 35 pages. |
U.S. Appl. No. 29/524,186 for Scanner filed Apr. 17, 2015 (Zhou et al.); 17 pages. |
U.S. Appl. No. 14/705,407 for Method and System to Protect Software-Based Network-Connected Devices From Advanced Persistent Threat filed May 6, 2015 (Hussey et al.); 42 pages. |
U.S. Appl. No. 14/614,706 for Device for Supporting an Electronic Tool on a User's Hand filed Feb. 5, 2015 (Oberpriller et al.); 33 pages. |
U.S. Appl. No. 14/628,708 for Device, System, and Method for Determining the Status of Checkout Lanes filed Feb. 23, 2015 (Todeschini); 37 pages. |
U.S. Appl. No. 14/704,050 for Intermediate Linear Positioning filed May 5, 2015 (Charpentier et. al.); 60 pages. |
U.S. Appl. No. 14/529,563 for Adaptable Interface for a Mobile Computing Device filed Oct. 31, 2014 (Schoon et al.); 36 pages. |
U.S. Appl. No. 14/705,012 for Hands-Free Human Machine Interface Responsive to a Driver of a Vehicle filed May 6, 2015 (Fitch et al.); 44 pages. |
U.S. Appl. No. 14/715,672 for Augumented Reality Enabled Hazard Display filed May 19, 2015 (Venkatesha et al.); 35 pages. |
U.S. Appl. No. 14/695,364 for Medication Management System filed Apr. 24, 2015 (Sewell et al.); 44 pages. |
U.S. Appl. No. 14/664,063 for Method and Application for Scanning a Barcode With a Smart Device While Continuously Running and Displaying an Application on the Smart Device Display filed Mar. 20, 2015 (Todeschini); 37 pages. |
U.S. Appl. No. 14/735,717 for Indicia-Reading Systems Having an Interface With a User's Nervous System filed Jun. 10, 2015 (Todeschini); 39 pages. |
U.S. Appl. No. 14/527,191 for Method and System for Recognizing Speech Using Wildcards in an Expected Response filed Oct. 29, 2014 (Braho et al.); 45 pages. |
U.S. Appl. No. 14/702,110 for System and Method for Regulating Barcode Data Injection Into a Running Application on a Smart Device filed May 1, 2015 (Todeschini et al.); 38 pages. |
U.S. Appl. No. 14/535,764 for Concatenated Expected Responses for Speech Recognition filed Nov. 7, 2014 (Braho et al.); 51 pages. |
U.S. Appl. No. 14/687,289 for System for Communication via a Peripheral Hub filed Apr. 15, 2015 (Kohtz et al.); 37 pages. |
U.S. Appl. No. 14/747,197 for Optical Pattern Projector filed Jun. 23, 2015 (Thuries et al.); 33 pages. |
U.S. Appl. No. 14/674,329 for Aimer for Barcode Scanning filed Mar. 31, 2015 (Bidwell); 36 pages. |
U.S. Appl. No. 14/702,979 for Tracking Battery Conditions filed May 4, 2015 (Young et al.); 70 pages. |
U.S. Appl. No. 29/529,441 for Indicia Reading Device filed Jun. 8, 2015 (Zhou et al.); 14 pages. |
U.S. Appl. No. 14/747,490 for Dual-Projector Three-Dimensional Scanner filed Jun. 23, 2015 (Jovanovski et al.); 40 pages. |
U.S. Appl. No. 14/740,320 for Tactile Switch for a Mobile Electronic Device filed Jun. 16, 2015 (Barndringa); 38 pages. |
U.S. Appl. No. 14/695,923 for Secure Unattended Network Authentication filed Apr. 24, 2015 (Kubler et al.); 52 pages. |
U.S. Appl. No. 14/740,373 for Calibrating a Volume Dimensioner filed Jun. 16, 2015 (Ackley et al.); 63 pages. |
U.S. Appl. No. 13/367,978, filed Feb. 7, 2012, (Feng et al.); now abandoned. |
U.S. Appl. No. 14/462,801 for Mobile Computing Device With Data Cognition Software, filed on Aug. 19, 2014 (Todeschini et al.); 38 pages. |
U.S. Appl. No. 14/596,757 for System and Method for Detecting Barcode Printing Errors filed Jan. 14, 2015 (Ackley); 41 pages. |
U.S. Appl. No. 14/277,337 for Multipurpose Optical Reader, filed May 14, 2014 (Jovanovski et al.); 59 pages. |
U.S. Appl. No. 14/200,405 for Indicia Reader for Size-Limited Applications filed Mar. 7, 2014 (Feng et al.); 42 pages. |
U.S. Appl. No. 14/662,922 for Multifunction Point of Sale System filed Mar. 19, 2015 (Van Horn et al.); 41 pages. |
U.S. Appl. No. 14/446,391 for Multifunction Point of Sale Apparatus With Optical Signature Capture filed Jul. 30, 2014 (Good et al.); 37 pages. |
U.S. Appl. No. 29/528,165 for In-Counter Barcode Scanner filed May 27, 2015 (Oberpriller et al.); 13 pages. |
U.S. Appl. No. 29/528,890 for Mobile Computer Housing filed Jun. 2, 2015 (Fitch et al.); 61 pages. |
U.S. Appl. No. 14/614,796 for Cargo Apportionment Techniques filed Feb. 5, 2015 (Morton et al.); 56 pages. |
U.S. Appl. No. 29/516,892 for Table Computer filed Feb. 6, 2015 (Bidwell et al.); 13 pages. |
U.S. Appl. No. 29/523,098 for Handle for a Tablet Computer filed Apr. 7, 2015 (Bidwell et al.); 17 pages. |
U.S. Appl. No. 14/578,627 for Safety System and Method filed Dec. 22, 2014 (Ackley et al.); 32 pages. |
U.S. Appl. No. 14/573,022 for Dynamic Diagnostic Indicator Generation filed Dec. 17, 2014 (Goldsmith); 43 pages. |
U.S. Appl. No. 14/529,857 for Barcode Reader With Security Features filed Oct. 31, 2014 (Todeschini et al.); 32 pages. |
U.S. Appl. No. 14/519,195 for Handheld Dimensioning System With Feedback filed Oct. 21, 2014 (Laffargue et al.); 39 pages. |
U.S. Appl. No. 14/519,211 for System and Method for Dimensioning filed Oct. 21, 2014 (Ackley et al.); 33 pages. |
U.S. Appl. No. 14/519,233 for Handheld Dimensioner With Data-Quality Indication filed Oct. 21, 2014 (Laffargue et al.); 36 pages. |
U.S. Appl. No. 14/533,319 for Barcode Scanning System Using Wearable Device With Embedded Camera filed Nov. 5, 2014 (Todeschini); 29 pages. |
U.S. Appl. No. 14/748,446 for Cordless Indicia Reader With a Multifunction Coil for Wireless Charging and EAS Deactivation, filed Jun. 24, 2015 (Xie et al.); 34 pages. |
U.S. Appl. No. 29/528,590 for Electronic Device filed May 29, 2015 (Fitch et al.); 9 pages. |
U.S. Appl. No. 14/519,249 for Handheld Dimensioning System With Measurement-Conformance Feedback filed Oct. 21, 2014 (Ackley et al.); 36 pages. |
U.S. Appl. No. 29/519,017 for Scanner filed Mar. 2, 2015 (Zhou et al.); 11 pages. |
U.S. Appl. No. 14/398,542 for Portable Electronic Devices Having a Separate Location Trigger Unit for Use in Controlling an Application Unit filed Nov. 3, 2014 (Sian et al.); 22 pages. |
U.S. Appl. No. 14/405,278 for Design Pattern for Secure Store filed Mar. 9, 2015 (Zhu et. al.) 23 pages. |
U.S. Appl. No. 14/590,024 for Shelving and Package Locating Systems for Delivery Vehicles filed Jan. 6, 2015 (Payne); 31 pages. |
U.S. Appl. No. 14/568,305 for Auto-Contrast Viewfinder for an Indicia Reader filed Dec. 12, 2014 (Todeschini); 29 pages. |
U.S. Appl. No. 29/526,918 for Charging Base filed May 14, 2015 (Fitch et al.); 10 pages. |
U.S. Appl. No. 14/580,262 for Media Gate for Thermal Transfer Printers filed Dec. 23, 2014 (Bowles); 36 pages. |
Ralph Grabowski, “Smothing 3D Mesh Objects,” New Commands in AutoCAD 2010: Part 11, Examiner Cited art in related matter Non Final Office Action dated May 19, 2017; 6 pages. |
European Exam Report in related EP Application No. 15176943.7, dated Apr. 12, 2017, 6 pages. |
European Exam Report in related EP Application No. 15188440.0, dated Apr. 21, 2017, 4 pages. |
European Exam Report in related EP Application No. 16152477.2, dated Jun. 20, 2017, 4 pages [No art to be cited]. |
European Exam Report in related EP Application 16172995.9, dated Jul. 6, 2017, 9 pages [No new art to be cited]. |
United Kingdom Search Report in related Application No. GB1700338.5, dated Jun. 30, 2017, 5 pages. |
European Search Report in related EP Application No. 17175357.7, dated Aug. 17, 2017, pp. 1-7 [No new art to be cited]. |
Number | Date | Country | |
---|---|---|---|
20170010141 A1 | Jan 2017 | US |