Gamma camera gantry systems have been developed that utilize robotic, multi-axes, rectilinear or polar coordinate actuators to move the imaging detectors. Programmable devices control these actuators along many different paths of detector motion. These gantries often have one or more vertical supports to which some of the actuators are attached, and moving parts that allow one or more types of detector motion.
Because gamma rays cannot be reflected or refracted, gamma cameras use collimator devices to limit the direction and area from which gamma rays can enter the camera's imaging detector. Collimators are typically made from high Z-materials such as lead having thousands of small diameter holes through which the gamma rays pass. The diameter and length of the holes define the imaging characteristics of the collimator. The amount of lead between the holes determines the gamma ray energy that can be imaged.
Most gamma cameras have more than one collimator type, with many having more than five types.
Modern gamma cameras have collimators that often weigh several hundred pounds. This weight makes it very difficult for a person to manually remove one type of collimator and replace it with another. In addition, storage of these collimators takes up floor space in the imaging suite.
For some time, gamma cameras have used collimator exchangers to eliminate the need to manually lift the collimators. These exchangers typically use a collimator storage cart that an operator rolls up to the imaging detector. The operator then pulls or pushes the existing collimator from the head to the storage cart. The cart is then rolled away. The new collimator (on another cart) is then rolled to the detector and pushed or pulled onto the detector.
Another way of doing this used an automated system to drive the cart. In time, the manual positioning of the collimator storage cart and the manual removal/installation of the collimator was automated. Regardless, the cart took up valuable floor space; thereby increasing the size of the room required to house the camera system.
Philips' SKYLight system was one of the first large field of view, robotic, multi-axis, programmable gantry gamma cameras. This introduced a new approach to automatic collimator exchange. The detector heads were robotically moved to a location where a wall mounted collimator holder would swing/deploy from a position parallel to the wall, to a position perpendicular to the wall. The robotic gantry then places the currently installed collimator onto the holder and the holder swings back parallel to the wall. The gantry would then travel to a new location where it would pick up a new collimator from a similarly wall mounted, swinging collimator holder.
For dual imaging head systems, each side of the deployed holder could hold a collimator. This approach had the advantage of placing the collimators against the wall thereby reducing the floor space required to store the collimators. However, it has the costly disadvantage of requiring an actuator on each wall mounted, swinging collimator storage holder in order to orient the collimator to be parallel to the imaging detector surface for mounting/de-mounting of the collimator. In addition, placing a collimator on each side of the holder extends the depth of the holder. This extended depth as well as the need to house the actuators could extend the size of the room required to house the camera and collimator exchanger.
In an effort to provide more space to meet the required detector-to-isotopic point source separation distance for calibration, camera systems are sometimes located so that the detector heads can be aligned to face towards a door on the wall opposite the detectors. This allows the isotopic point source to be placed in a hallway gaining the required separation distance.
Imaging a patient in a hospital bed with robotic, multi-axis, programmable gantry gamma camera systems that are floor mounted, has required that the hospital bed be placed across the track that runs along the floor supporting the vertical gantry tower.
The present application describes additional movement capability in a gantry camera system, that allows special functions not possible with prior art devices.
More specifically, embodiments disclose a new movement of a 180 degree rotation on a gantry tower, that rotation of the gantry tower, and in a way that allows the camera and heads to be moved away from the patient being imaged.
The inventor found that many of the existing gantry systems have a limited amount of motion.
The inventor recognized that a significant limitation, and source of many different problems, would be solved or reduced if these gantry systems enabled their vertical support to rotate about its vertical axis. This new motion correspondingly allows a decrease in the size of the room housing the camera, and housing its associated optional collimators. The recognition of this problem allowed the inventor to conceive of new features allowed by new movements that are disclosed herein.
The imaging detectors on a gamma camera often require periodic calibration. Some of the intrinsic calibrations require that an isotopic point source be positioned at a distance of five times (5×) the maximum dimension of the detector's field of view. For a 21″×16″ field of view (which is a size of many, but not all, detectors), this distance is 8.75 feet from the face of the detector. Proper calibration requires that there be no obstructions between the isotopic source and the detector.
Robotic, multi-axis, programmable gantry gamma camera systems are often located in a place that requires that the imaging detectors be spaced from the wall. This inability to get close to a wall, combined with the need for the 5× detector size distance from the detector to a location where an isotopic point source can be placed, even further increases the necessary room size. An embodiment addresses this problem by allowing this movement that allows placing the imaging detector against a wall, facing away from the wall.
On occasion, it is necessary to perform imaging procedures on patients that cannot be easily removed from their hospital bed. Often these imaging procedures are whole body scans where it is desirable to have both detectors above the patient and able to scan down the body. This may even further increase the room size.
The inventor realized that a means of enabling a multi-axis, programmable gantry gamma camera system to be housed in a small room while still allowing for an automatic collimator exchanger, the ability to perform intrinsic calibrations with isotopic point sources and the ability to easily perform imaging procedures with a patient in a hospital bed that has been moved to the gamma camera would be useful.
A robotic, multi-axis, gantry gamma camera system 100 is shown in
In an embodiment, a collimator 120 is placed on the face 121 of the imaging detector. Spare collimators 125 are stored on a wall-mounted rack. However, unlike the prior art, the collimator holders are mounted on the wall, so that the collimator surfaces are flat against the wall (with the collimator holes perpendicular to the wall). These collimator holders are maintained as flat on the wall and do not swing. One embodiment provides only one collimator on each holder. An embodiment shows two collimator holders 125, 126, one vertically above the other. Preferably, there are 6 or more collimator holders as shown.
A rotation stage 140 is actuated to rotate the vertical gantry tower 110, along with everything that is connected to the tower, e.g., its associated detector arm(s) 130, 132 and imaging detectors 131, 133.
When the device is rotated by 90° around the vertical axis, it allows both imaging detectors to be positioned in tandem along the axis of the imaging table, e.g. adjacent a patient hospital bed (above or below the bed) . Doing so enables more of the axial extent of the patient's body to be imaged simultaneously. In addition, orienting the detector heads in this manner enables a patient's hospital bed to be in the same position as was the patient imaging table, parallel to the gantry tower track.
Additional axes of motion can be used with the rotation stage 140 to allow advantages that are not reasonably predictable from the prior art.
An axis-of-motion shown as 122 allows moving the imaging detector surface 121. The surface can be placed parallel to the collimator holder(s) 125 for mounting/de-mounting. This embodiment allows 180° rotation about the system's vertical axis shown by the arrow 142, enabling the imaging detectors to be re-oriented from facing perpendicular to the imaging table axis, to facing perpendicularly away form the table axis and parallel towards the wall mounted collimator storage holder(s) 125.
The addition of this 180° rotation about the system's vertical axis also unexpectedly enables the detectors to be positioned close to one wall, to allow the maximum distance from the imaging surface to the opposite wall.
The gantry arms 130 can also be moved in the direction 134 to allow the motion and functions described herein.
This system allows a number of advantages that provide significant advantages, and advantages which are not reasonably expectable from the prior art.
The addition of the 180° rotation about the robotic, multi-axis, programmable gantry camera system's vertical axis enables each collimator holder to be fixed in place. No actuator is required to swing the collimator(s) into position. This significantly reduces system complexity by eliminating actuators on each collimator holder. Since accurate positioning need not be maintained on a swinging apparatus, complexity and cost is further reduced.
Additionally, because each collimator holder holds only one collimator, the depth of collimator holder system is minimized and the need to enlarge the room is reduced or eliminated.
The addition of this 180° rotation about the robotic, multi-axis, programmable gantry camera system's vertical axis also enables the detectors to be positioned close to one wall so that there is the maximum distance from the imaging surface to the opposite wall. This minimizes the size of the room into which the robotic, multi-axis, programmable gantry gamma camera can be installed while still providing space to perform system calibrations with an isotopic point source.
The ability to rotate the vertical gantry by 90° allows whole body hospital bed imaging to be performed without the need to place the hospital bed across the vertical tower gantry track. This avoids the need to lift the bed over the tracks and reduces the risk of accidental collision of the camera system with the patient's hospital bed. The operation is explained with relation to
The entire movement and structure can be carried out by a controller 150 which can be a computer that automatically determines movement of the various structures and their speeds of movement according to a preprogrammed sequence. A user interface 152 may allow these devices to be controlled remotely for example.
The general structure and techniques, and more specific embodiments which can be used to effect different ways of carrying out the more general goals are described herein.
Although only a few embodiments have been disclosed in detail above, other embodiments are possible and the inventors intend these to be encompassed within this specification. The specification describes specific examples to accomplish a more general goal that may be accomplished in another way. This disclosure is intended to be exemplary, and the claims are intended to cover any modification or alternative which might be predictable to a person having ordinary skill in the art. For example, other kinds of medical imaging systems could be used with this embodiment. Other differences between the scans could be used. Other compensations can be applied to the scans.
Also, the inventor intend that only those claims which use the words “means for” are intended to be interpreted under 35 USC 112, sixth paragraph. Moreover, no limitations from the specification are intended to be read into any claims, unless those limitations are expressly included in the claims. The computers described herein may be any kind of computer, either general purpose, or some specific purpose computer such as a workstation. The computer may be an Intel (e.g., Pentium or Core 2 duo) or AMD based computer, running Windows XP or Linux, or may be a Macintosh computer. The computer may also be a handheld computer, such as a PDA, cellphone, or laptop.
The programs may be written in C or Python, or Java, Brew or any other programming language. The programs may be resident on a storage medium, e.g., magnetic or optical, e.g. the computer hard drive, a removable disk or media such as a memory stick or SD media, wired or wireless network based or Bluetooth based Network Attached Storage (NAS), or other removable medium or other removable medium. The programs may also be run over a network, for example, with a server or other machine sending signals to the local machine, which allows the local machine to carry out the operations described herein.
Where a specific numerical value is mentioned herein, it should be considered that the value may be increased or decreased by 20%, while still staying within the teachings of the present application, unless some different range is specifically mentioned. Where a specified logical sense is used, the opposite logical sense is also intended to be encompassed.
This application claims priority from U.S. Provisional application No. 60/894,856, filed Mar. 14, 2007, the disclosure of which is herewith incorporated by reference.
Number | Date | Country | |
---|---|---|---|
60894856 | Mar 2007 | US |