The present disclosure is generally related to methods of imaging tissues using near infra-red light sheet microscopy. The present disclosure is further generally related to an optical system for near infra-red light sheet microscopy.
Optical imaging of biological systems capable of high spatiotemporal resolution in vivo and ex vivo has revolutionized biology and medicine for visualization and understanding of structures, functions and dynamic processes at the cellular, and even molecular, scale (Jain, R. K. Cancer Cell 26: 605-622 (2014); Liu et al., Science 360: eaaq1392 (2018)). To circumvent light scattering by tissues, in vivo 3D imaging by non-linear two-photon fluorescence microscopy (670-1070 nm excitation) (Helmchen & Denk. Nat. Methods 2: 932-940 (2005); Lapadula et al., Chem. Mater. 26: 1062-1073 (2014); Alifu et al., Dyes Pigments 143: 76-85 (2017)) or three-photon microscopy (1300-1700 nm excitation) (Horton et al., Nat. Photonics 7: 205-209 (2013); Ouzounov et al., Nat. Methods 14: 388-390 (2017); Rowlands et al., Light-Sci. Appl. 6: e16255 (2017); Liu et al., Biomed. Opt. Express 6: 1857-1866 (2015)) has reached penetration depths of approximately 0.7-1.5 mm, benefiting from increased scattering mean free path of the near-infrared (NIR) excitation used (Horton et al., Nat. Photonics 7: 205-209 (2013)).
Light sheet microscopy (LSM) uses orthogonally arranged planar illumination and wide-field detection, capable of high speed 3D optical sectioning, low photo-damage (Huisken et al., Science 305: 1007-1009 (2004); Dodt et al., Nat. Methods 4: 331-336 (2007)) and volumetrically imaging/tracking with subcellular resolution (Liu et al., Science 360: eaaq1392 (2018)). Currently the excitation and emission of LSM are mostly in the visible range except for two photon excitation in the NIR at approximately 940 nm (Truong et al., Nat. Methods 8: 757-U102 (2011)) or three photon excitation at 1000 nm (Escobet-Montalbán et al., Opt. Lett. 43: 5484-5487 (2018)). Light scattering has limited use of LSM to imaging small transparent animals, organisms (zebrafish larvae, Drosophila larvae, Medaka embryo, C. elegans, etc.), mammalian tissue samples after chemical clearing (Dodt et al., Nat. Methods 4: 331-336 (2007); Chung et al., Nature 497: 332-337 (2013); Tomer et al., Nat. Protoc. 9: 1682-1697 (2014)), and mouse brain at a depth of approximately 200 μm after craniotomy (Bouchard et al., Nat. Photonics 9: 113 (2015)).
Several classes of fluorescence probes have been developed with emission in the NIR-II window (1000-1700 nm) including carbon nanotubes, quantum dots, organic conjugated polymers and molecular dyes, and rare-earth nanoparticles (Welsher et al., Nat. Nanotechnol. 4: 773-780 (2009); Welsher et al., Proc. Natl. Acad. Sci. USA 108: 8943-8948 (2011); Hong et al., Nat. Med. 18: 1841-1846 (2012); Hong et al., Nat. Photonics 8: 723-730 (2014); Hong et al., Nat. Commun. 5: 4206 (2014); Diao et al., Angew. Chem. Int. Edit. 54: 14758-14762 (2015); Diao et al., Nano Res. 8: 3027-3034 (2015); Antaris et al., Nat. Mater. 15: 235-242 (2016); Zhong et al., Nat. Commun. 8: 737 (2017); Won et al., Mol. Imaging. 11: 338-352 (2012); Naczynski et al., Nat. Commun. 4: 2199 (2013); Wan et al., Nat. Commun. 9: 1171 (2018); Zhang et al., Proc. Natl. Acad. Sci. USA 115: 6590-6595 (2018); Bruns et al., Nat. Biomed. Eng. 1: 0056 (2017)). With suppressed photon scattering and diminished autofluorescence in the long-wavelength region, these probes have facilitated one-photon wide-field (Welsher et al., Nat. Nanotechnol. 4: 773-780 (2009); Welsher et al., Proc. Natl. Acad. Sci. USA 108: 8943-8948 (2011); Hong et al., Nat. Med. 18: 1841-1846 (2012); Hong et al., Nat. Photonics 8: 723-730 (2014); Hong et al., Nat. Commun. 5: 4206 (2014); Diao et al., Angew. Chem. Int. Edit. 54: 14758-14762 (2015); Diao et al., Nano Res. 8: 3027-3034 (2015); Antaris et al., Nat. Mater. 15: 235-242 (2016); Zhong et al., Nat. Commun. 8: 737 (2017); Won et al., Mol. Imaging. 11: 338-352 (2012); Naczynski et al., Nat. Commun. 4: 2199 (2013)) or confocal (Wan et al., Nat. Commun. 9: 1171 (2018); Zhang et al., Proc. Natl. Acad. Sci. USA 115: 6590-6595 (2018)) fluorescence imaging in the NIR-II window for mouse models of cardiovascular and brain diseases and cancer (Hong et al., Nat. Photonics 8: 723-730 (2014); Wan et al., Nat. Commun. 9: 1171 (2018); Bruns et al., Nat. Biomed. Eng. 1: 0056 (2017)). Non-invasive imaging through the skin, skull and body tissues was achieved, with deep penetration depths and high signal-to-background ratio (SBR).
One aspect of the disclosure, therefore, encompasses embodiments of a method of optically sectioning a biological sample, the method comprising the steps: (a) contacting a biological sample with a fluorescent contrast agent having an excitation wavelength of between about 785 nm to about 2400 nm and a fluorescence emission having a wavelength of between about 800 nm to about 2400 nm; (b) irradiating the biological sample with at least one excitation light having a wavelength of between about 785 nm to about 2400 nm, wherein the excitation light is configured as a static light sheet and directed through a first plane of the biological sample; (c) orthogonally detecting an emitted fluorescence having a wavelength of between about 800 nm to about 2400 nm from the irradiated biological sample; and (d) generating a first digital image of the fluorescence from the irradiated first plane through the biological sample.
In embodiments of this aspect of the disclosure, the biological sample can be an isolated cell or population of isolated cells, a cultured cell or population of cultured cells, an isolated tissue or organ, or an animal or human subject.
In some embodiments of this aspect of the disclosure, the biological sample has raised features and the irradiated biological sample is imaged by transmitting the light sheet through the raised feature from a side thereof and detecting the emitted fluorescence at right angles to the plane of the illuminating light sheet.
In some embodiments of this aspect of the disclosure, the biological sample can have substantially planar regions and the irradiated biological sample can be illuminated by the light sheet at an angle with respect to the normal to the planar region and the emitted fluorescence is detected at right angles to the plane of the illuminating light sheet.
In some embodiments of this aspect of the disclosure, the fluorescent contrast agent can be administrated by vascular delivery to a tissue or organ of the animal or human subject.
In some embodiments of this aspect of the disclosure, the fluorescent contrast agent can be conjugated to a targeting ligand that can specifically bind to a molecular target.
In some embodiments of this aspect of the disclosure, the targeting ligand can have an affinity for a molecular target and is selected from an antibody, a peptide, an aptamer, or a nucleic acid.
In some embodiments of this aspect of the disclosure, the targeting ligand can be an antibody or fragment thereof selectively binding to an epitope of a polypeptide selected from the group consisting of: Programmed cell death protein 1 (PD-1), Programmed death-ligand 1 (PD-L1), and CD11b.
In some embodiments of this aspect of the disclosure, the fluorescent contrast agent can emit at a wavelength of between about 900 to about 2400 nm and can be an organic molecular dye, a conjugated polymeric dye, a polymer micelle-wrapped organic nanofluorophore, a carbon nanotube, a quantum dot, or a rare-earth down-conversion or up-conversion nanoparticle.
In some embodiments of this aspect of the disclosure, the method can further comprise the steps: (i) repeating steps (b)-(d), thereby irradiating a plurality of parallel planes perpendicular to the light sheet plane at various depths through the biological sample, and generating a plurality of digital images; and (ii) digitally combining the plurality of digital images to generate a three-dimensional image of the location of the fluorescence emitted by the contrast agent in the biological sample.
In some embodiments of this aspect of the disclosure, the excitation light can be delivered to biological sample by an illumination objective, wherein the numerical aperture (N.A) of the objective is configured to deliver the excitation light as a light sheet having a balanced waist thickness of between about 5 μm to about 20 μm and a Rayleigh length of between about 0.1 mm to about 6.0 mm.
In some embodiments of this aspect of the disclosure, the excitation light can be generated by a laser having a wavelength of between about 700 nm and about 2400 nm. Another aspect of the disclosure encompasses embodiments of a light sheet microscope comprising along an optical axis: an illumination objective positioned to direct an excitation light sheet through a plane of a biological sample; a plurality of achromatic lenses optimized for transmission of light between about 785 nm to about 2400 nm; a proximal first adjustable mechanical slit; a cylindrical lens; a distal second adjustable mechanical slit adjacent to the cylindrical lens and distal to the illumination objective, wherein the slit of said second slit is orientated at right-angles to the slit of the first slit; a pinhole; at least one excitation light source in the 600-2400 nm range; and at least one removable mirror disposed to direct an excitation light from the at least one light source along the optical axis of the illumination objective, the plurality of achromatic lenses, the proximal first adjustable mechanical slit, the cylindrical lens, the distal second adjustable mechanical slit, and the pinhole; a detection objective disposed to orthogonally receive fluorescent light emitted from a target irradiated by a light sheet from the illumination objective and to direct said fluorescent light to a detector operably connected to a computer system for generating a digital image of the fluorescent light; and at least one emission filter configured to only transmit fluorescent light having a wavelength of between about 785 nm to about 2400 nm.
In some embodiments of this aspect of the disclosure, the light sheet microscope can further comprise a right-angle prism disposed between the biological sample and the illuminating and receiving lenses.
In some embodiments of this aspect of the disclosure, when the illuminating light sheet is configured to illuminate the biological sample at an angle of less than 90° between a tangential plane of the biological sample and the illuminating light sheet, a solid, liquid or gas having a refractive index of about that of the biological sample can be disposed between the biological sample and the illuminating and receiving lenses.
In some embodiments of this aspect of the disclosure, the excitation light source can further comprise a shortpass filter to select the excitation wavelength.
In some embodiments of this aspect of the disclosure, the excitation light source can be a laser.
In some embodiments of this aspect of the disclosure, the excitation light source can emit an excitation light having a wavelength of between about 785 nm and 2400 nm.
In some embodiments of this aspect of the disclosure, the detector can be sensitive to light having a wavelength of between about 800 nm and about 2400 nm.
In some embodiments of this aspect of the disclosure, the detector can be an InGaAs camera sensitive to light having a wavelength of between about 800 nm and about 2400 nm.
In some embodiments of this aspect of the disclosure, the detector can be a small bandgap semiconductor-based camera sensitive to light having a wavelength of between about 800 nm to about 2400 nm.
Further aspects of the present disclosure will be more readily appreciated upon review of the detailed description of its various embodiments, described below, when taken in conjunction with the accompanying drawings.
This disclosure is not limited to particular embodiments described, and as such may, of course, vary. The terminology used herein serves the purpose of describing particular embodiments only, and is not intended to be limiting, since the scope of the present disclosure will be limited only by the appended claims.
Where a range of values is provided, each intervening value, to the tenth of the unit of the lower limit unless the context clearly dictates otherwise, between the upper and lower limit of that range and any other stated or intervening value in that stated range, is encompassed within the disclosure. The upper and lower limits of these smaller ranges may independently be included in the smaller ranges and are also encompassed within the disclosure, subject to any specifically excluded limit in the stated range. Where the stated range includes one or both of the limits, ranges excluding either or both of those included limits are also included in the disclosure.
Embodiments of the present disclosure will employ, unless otherwise indicated, techniques of medicine, organic chemistry, biochemistry, molecular biology, pharmacology, and the like, which are within the skill of the art. Such techniques are explained fully in the literature.
The following examples are put forth so as to provide those of ordinary skill in the art with a complete disclosure and description of how to perform the methods and use the compositions and compounds disclosed and claimed herein. Efforts have been made to ensure accuracy with respect to numbers (e.g., amounts, temperature, etc.), but some errors and deviations should be accounted for. Unless indicated otherwise, parts are parts by weight, temperature is in ° C., and pressure is at or near atmospheric. Standard temperature and pressure are defined as 20° C. and 1 atmosphere.
Before the embodiments of the present disclosure are described in detail, it is to be understood that, unless otherwise indicated, the present disclosure is not limited to particular materials, reagents, reaction materials, manufacturing processes, dimensions, frequency ranges, applications, or the like, as such can vary. It is also to be understood that the terminology used herein is for purposes of describing particular embodiments only, and is not intended to be limiting. It is also possible in the present disclosure that steps can be executed in different sequence, where this is logically possible. It is also possible that the embodiments of the present disclosure can be applied to additional embodiments involving measurements beyond the examples described herein, which are not intended to be limiting. It is furthermore possible that the embodiments of the present disclosure can be combined or integrated with other measurement techniques beyond the examples described herein, which are not intended to be limiting.
It should be noted that, as used in the specification and the appended claims, the singular forms “a,” “an,” and “the” include plural referents unless the context clearly dictates otherwise. Thus, for example, reference to “a support” includes a plurality of supports. In this specification and in the claims that follow, reference will be made to a number of terms that shall be defined to have the following meanings unless a contrary intention is apparent.
Each of the applications and patents cited in this text, as well as each document or reference cited in each of the applications and patents (including during the prosecution of each issued patent; “application cited documents”), and each of the PCT and foreign applications or patents corresponding to and/or claiming priority from any of these applications and patents, and each of the documents cited or referenced in each of the application cited documents, are hereby expressly incorporated herein by reference. Further, documents or references cited in this text, in a Reference List before the claims, or in the text itself; and each of these documents or references (“herein cited references”), as well as each document or reference cited in each of the herein-cited references (including any manufacturer's specifications, instructions, etc.) are hereby expressly incorporated herein by reference.
Prior to describing the various embodiments, the following definitions are provided and should be used unless otherwise indicated.
LSM, Light Sheet Microscope (Microscopy); LSFM, Light Sheet Fluorescence Microscope (Microscopy); Full Width at Half Maximum, FWHM; NIR, Near infra-red; NA, numerical aperture; InGaAs, Indium Gallium Arsenide; PEG, polyethylene glycol; SBR, signal:background ratio; CSQD, core-shell quantum dot; 3D, three-dimensional; PD-L1, programmed-death ligand 1; PD-1, Programmed cell death protein 1; BST, blood supply time; FOV, field of view; NaErF4, Erbium rare-earth nanoparticles; TBI, traumatic brain injury
The term “achromatic lens” or “achromat” as used herein refers to a lens that is designed to limit the effects of chromatic and spherical aberration. Achromatic lenses are corrected to bring two wavelengths (typically red and blue) into focus on the same plane.
The term “biological sample” as used herein refers to an isolated or cultured cell or population of cells, an isolated tissue or organ removed from an animal or human subject, or an animal or human subject living or dead. The term biological sample can refer to a sample of plant or animal origin.
The term “biomarker” as used herein refers to an antigen such as, but not limited to, a peptide, polypeptide, protein (monomeric or multimeric) that may be found on the surface of a cell, an intracellular component of a cell, or a component or constituent of a biofluid such as a soluble protein in a serum sample and which is a characteristic that is objectively measured and evaluated as an indicator of a tumor or tumor cell. The presence of such a biomarker in a biofluid or a biosample isolated from a subject human or animal can indicate that the subject is a bearer of a cancer. A change in the expression of such a biomarker may correlate with an increased risk of disease or progression, or predictive of a response of a disease to a given treatment. Exemplary biomarkers useful in the systems and methods of the disclosure can be, but are not limited to, such as activin A; IL-18 BPa, adiponectin/acrp30, IL-18 receptor α/IL-1 R5, AgRP, IL-18 receptor β/AcPL, ALCAM, IL-2 receptor α, angiogenin, IL-2 receptor α, AR (amphiregulin), IL-3, Axl, IL-4, B7-1/CD80, I-TAC/CXCL11, BCMA/TNFRSF17, leptin (OB), BDNF, LIF, β-NGF, LIGHT/TNFSF14, BLC/BCA-1/CXCL13, LIGHT/TNFSF14, BMP-5, MCP-2, BTC, MCP-3, cardiotrophin-1/CT-1, MCP-4/CCL13, CTLA-4/CD152, M-CSF, CXCL16, MMP-10, Dtk, MMP-13, EGF, MMP-9, EGF receptor/ErbB1, MSP α-chain, endoglin/CD105, MSP β-chain, eotaxin/CCL11, NAP-2, eotaxin-2/MPIF-2, NGF R, eotaxin-3/CCL26, NT-4, ErbB3, OSM, Fas/TNFRSF6, osteoprotegerin, Fas Ligand, PDGF receptor β, FGF Basic, PDGF-AA, FGF-4, PDGF-AB, FGF-6, PDGF-BB, FGF-7/KGF, PIGF, FGF-9, P-selectin, follistatin, RAGE, GITR/TNFRF18, RANTES, HB-EGF, SCF, HCC-4/CCL16, SCF receptor/CD117, HGF, sgp130, 1-309, Siglec-9, IGFBP-1, siglec-5/CD170, IGFBP-2, Tarc, IGFBP-3, TGFα, IGF-I, TNF RI/TNFRSF1A, IGF-I, TNF RII/TNFRSF1B, IGF-I S receptor, TNFβ, IGF-II, TRAIL R1/DR4/TNFRSF 10/, IGF-II, TRAIL R3/TNFRSF 10C, IL-1α, TRAIL R4/TNFRSF 10D, IL-1β, TRANCE, IL-1 R4/ST2, TREM-1, IL-1 sRI, TROP/TNFRSF19, IL-1 sRI, uPAR, IL-10, VCAM-1 (CD106), IL-10 receptor β, VE-cadherin, IL-13 receptor α1, VEGF, IL-13 receptor α2, VEGF R2 (KDR), IL-17, VEGF R3, and the like, or any combination thereof. It is considered within the scope of the disclosure for a cancer or cancer cell to be characterized by at least one biomarker and more typically by a plurality (a panel) of such markers.
The term “contacting” as used herein refers to any method that places a biological sample to be examined by the apparatus and methods of the disclosure in physical contact with a probe as herein disclosed. Isolated cells and tissues, for example, may be immersed in a solution or suspension of the probe or nanoparticle probe whereupon the probe may attach to or be absorbed by the sample. Organs isolated from an animal or human subject may, for example, be immersed in a probe solution or the solution can be delivered to throughout all or part of the organ through the vasculature of the organ. Similarly, the probe composition may be administered to an animal or human subject by intravascularly, subcutaneously, or by any other method know to one of skill in the art.
The term “dye” as used herein refers to any reporter group whose presence can be detected by its light absorbing or light emitting properties. For example, Cy5 is a reactive water-soluble fluorescent dye of the cyanine dye family. Cy5 is fluorescent in the red region (about 650 to about 670 nm). It may be synthesized with reactive groups on either one or both of the nitrogen side chains so that they can be chemically linked to either nucleic acids or protein molecules. Labeling is done for visualization and quantification purposes. Cy5 is excited maximally at about 649 nm and emits maximally at about 670 nm, in the far red part of the spectrum; quantum yield is 0.28. FW=792. Suitable fluorophores (chromes) for the probes of the disclosure may be selected from, but not intended to be limited to, fluorescein isothiocyanate (FITC, green), cyanine dyes Cy2, Cy3, Cy3.5, Cy5, Cy5.5 Cy7, Cy7.5 (ranging from green to near-infrared), Texas Red, and the like. Derivatives of these dyes for use in the embodiments of the disclosure may be, but are not limited to, Cy dyes (Amersham Bioscience), Alexa Fluors (Molecular Probes, Inc.), HILYTE™ Fluors (AnaSpec), and DYLITE™ Fluors (Pierce, Inc.).
The term “fluorescence” as used herein refers to a luminescence that is mostly found as an optical phenomenon in cold bodies, in which the molecular absorption of a photon triggers the emission of a photon with a longer (less energetic) wavelength. The energy difference between the absorbed and emitted photons ends up as molecular rotations, vibrations or heat. Sometimes the absorbed photon is in the ultraviolet range, and the emitted light is in the visible range, but this depends on the absorbance curve and Stokes shift of the particular fluorophore.
Spectroscopic, photochemical, biochemical, immunochemical, electrical, optical or chemical means can be used to detect such labels. The detection device and method may include, but is not limited to, optical imaging, electronic imaging, imaging with a CCD camera, integrated optical imaging, and mass spectrometry. Further, the amount of labeled or unlabeled probe bound to the target may be quantified. Such quantification may include statistical analysis.
The term “fluorophore” as used herein refers to any reporter group whose presence can be detected by its light emitting properties. The term “contrast agent” as used herein refers to an agent that when delivered to a cell, a tissue, or an animal or human subject can improve the image.
The term “light sheet microscopy” as used herein refers to illumination done perpendicularly to the direction of observation. The expanded beam of a light, most advantageously, but not limited to, a laser light is focused in only one direction by a cylindrical lens, or by a combination of a cylindrical lens and a microscope objective as the latter is available in better optical quality and with higher numerical aperture than the first. This way a thin sheet of light or “light sheet” is created in the focal region that can be used to excite fluorescence only in a thin slice (usually a few micrometers thin) of the sample.
In light sheet fluorescence microscopy (LSFM) the sheet of light that illuminates the specimen is orthogonal to the detection path and only fluorophores close to the focal plane of the detection system are detectable. By reducing photobleaching effects while imaging, the specimen is exposed to 5000 times less energy than in a confocal microscope. The light sheet stage allows for 3-dimensional movement and rotation around a vertical axis, such that imaging from all angles can be possible. Additionally, data acquisition in LSFM technology allows the detector to collect all pixels in one image as opposed to one pixel at a time, which has a great impact on the rate of image acquisition (100 frames per second) in contrast to the confocal at 1-5 frames per second. As a result, LSFM permits lengthier imaging time course experiments.
The term “nanoparticle” as used herein refers to a particle having a diameter of between about 1 nm and about 1000 nm, preferably between about 100 nm and 1000 nm, and most preferably between about 50 nm and 700 nm. Similarly, by the term “nanoparticles” is meant a plurality of particles having an average diameter of between about 50 nm and about 1000 nm.
The term “orthogonal” as used herein refers to two objects having axes, wherein the axes are at right angles to each other.
The term “molecular target” as used herein refers to a molecule that can be specifically bound by a ligand such as, but not limited to, a peptide, protein, nucleic acid, or a small molecule ligand. The target may be specific to or concentrated in or on a cell, a tissue, or an animal tissue, organ, or cell type compared to other tissues, organs, or cell types of the animal subject. A molecular target may be a biomarker.
The term “quantum dot” (quantum dots) as used herein refers to semiconductor nanocrystals or artificial atoms, which are semiconductor crystals that contain anywhere between 100 to 1,000 electrons and range from about 2 to about 10 nm. Some quantum dots can be between about 10 to about 20 nm in diameter. Quantum dots have high quantum yields, which makes them particularly useful for optical applications. Quantum dots are fluorophores that fluoresce by forming excitons, which can be thought of as the excited state of traditional fluorophores, but which have much longer lifetimes of up to 200 nanoseconds. This property provides quantum dots with low photobleaching. The energy level of quantum dots can be controlled by changing the size and shape of the quantum dot, and the depth of the quantum dots' potential. One of the optical features of small excitonic quantum dots is coloration, which is determined by the size of the dot. The larger the dot, the redder, or more towards the red end of the spectrum the fluorescence. The smaller the dot, the bluer or more towards the blue end it is. The bandgap energy that determines the energy and hence the color of the fluoresced light is inversely proportional to the square of the size of the quantum dot. Larger quantum dots have more energy levels which are more closely spaced, thus allowing the quantum dot to absorb photons containing less energy, i.e. those closer to the red end of the spectrum. Because the emission frequency of a dot is dependent on the bandgap, it is therefore possible to control the output wavelength of a dot with extreme precision. Colloidally prepared quantum dots are free floating and can be attached to a variety of molecules via metal coordinating functional groups. These groups include but are not limited to thiol, amine, nitrile, phosphine, phosphine oxide, phosphonic acid, carboxylic acids or other ligands. By bonding appropriate molecules to the surface, the quantum dots can be dispersed or dissolved in nearly any solvent or incorporated into a variety of inorganic and organic films.
The terms “subject”, “individual”, or “patient” as used herein are used interchangeably and refer to a cell, a tissue, or an animal preferably a warm-blooded animal such as a mammal. Mammal includes without limitation any members of the Mammalia. A mammal, as a subject or patient in the present disclosure, can be from the family of Primates, Carnivora, Proboscidea, Perissodactyla, Artiodactyla, Rodentia, and Lagomorpha. In a particular embodiment, the mammal is a human. In other embodiments, animals can be treated; the animals can be vertebrates, including both birds and mammals. In aspects of the disclosure, the terms include domestic animals bred for food or as pets, including equines, bovines, sheep, poultry, fish, porcines, canines, felines, and zoo animals, goats, apes (e.g. gorilla or chimpanzee), and rodents such as rats and mice.
The term “upconverting nanoparticle (UCNP)” as used herein refers to nanoscale particles that exhibit photon upconversion. In photon upconversion, two or more incident photons of relatively low energy are absorbed and converted into one emitted photon with higher energy. Generally, absorption occurs in the infrared, while emission occurs in the visible or ultraviolet regions of the electromagnetic spectrum. UCNPs are usually composed of lanthanide- or actinide-doped transition metals and are of particular interest for their applications in bio-imaging and bio-sensing at the deep tissue level. Upconversion can take place in both organic and inorganic materials, through a number of different mechanisms. Organic molecules that can achieve photon upconversion through triplet-triplet annihilation are typically polycyclicaromatic hydrocarbons (PAHs). Inorganic materials capable of photon upconversion often contain ions such as Ln3+, Ti2+, Ni2+, Mo3+, Re4+, Os4+, and the like.
Near-infrared-II Light Sheet Microscopy through Scattering Tissues: The present disclosure encompasses a method of three-dimensional (3D) fluorescence imaging in the visible by light sheet microscopy (LSM), a powerful tool for biological imaging with high spatiotemporal resolution. Optical excitation and emission wavelengths were extended up to approximately 1320 nm and to at least 1700 nm respectively into the NIR-II (800-2400 nm) region for near-infrared (NIR) light sheet microscopy through scattering tissues. Suppressed scattering of both excitation and emission photons allowed optical sectioning at greater than 2 mm depth in non-cleared brain tissues. NIR-II LSM enabled non-invasive in vivo imaging of live mice, revealing never-before-seen dynamic processes such as highly abnormal tumor microcirculation, and 3D molecular imaging of an important immune checkpoint protein, programmed-death ligand 1 (PD-L1) receptors at the single cell scale in tumors. In vivo two-color near-infrared light sheet sectioning enabled simultaneous volumetric imaging of tumor vasculatures and PD-L1 proteins in live mammals.
Near-infrared optical sectioning of live mammals in three dimensions: Mammals show high genetic similarity to humans and have been widely used for disease model studies. However, light scattering has limited 3D light sheet microscopy to imaging small transparent animals or cleared tissues ex vivo. Non-invasive in vivo 3D optical sectioning of live mammals with high spatiotemporal resolution has been challenging. Near-infrared light sheet microscopy was developed with optical excitation and emission wavelengths in the 1000-1700 nm, greatly suppressing light scattering by tissues for in vivo approximately 10 mm3 volumetric imaging of live mice with single cell resolution. Highly abnormal tumor microcirculation dynamics and molecular distribution of programmed-death ligand 1 (PD-L1) receptors were revealed in live tumors.
The present disclosure encompasses NIR-II LSM developed using organic dyes and PbS/CdS core/shell quantum dot (CSQD) probes, extending excitation and emission to the approximately 785-1320 nm and approximately 1000-1700 nm regimes respectively. However, the methods herein described can be readily adapted for other fluorescent labels that can be targeted to specific cells, tissues or organs by conjugating the label(s) to ligands having specific binding affinity for the desired target.
Suppressed light scattering of both excitation and emission allowed up to 10 mm3 volumetric imaging of mouse brain with a penetration depth of at least 2 mm. NIR LSM readily afforded in vivo imaging of mouse tumor models non-invasively, enabling real-time observation of unusual tumor microcirculation, and 3D molecular imaging of immune checkpoint proteins at cellular scale in live mammals.
The LSM methods of the disclosure can use multiple switchable lasers with Gaussian beams (658 nm, 785 nm and 1319 nm) cylindrically focused into static light sheets for optical sectioning and an InGaAs camera for orthogonally detecting 900-1700 nm fluorescence (
With a 785 nm light sheet, the glycerol-cleared brain tissue imaging depth limit increased (
The lateral full width at half maximum (FWHM) values of the smallest cerebral vasculatures imaged in the three emission regions (850-1000 nm, 1100-1200 nm and 1500-1700 nm,
As a light sheet propagated in a scatting medium such as an intralipid phantom (van Staveren et al., Appl. Opt. 30: 4507-4514 (1991); Johns et al. Opt. Express 13: 4828-4842 (2005)) mimicking the glycerol-cleared brain tissue, Monte Carlo simulations (Johns et al. Opt. Express 13: 4828-4842 (2005)) and experiments showed light sheet decaying in intensity in the X-Y plane and spreading in Z due to tissue scattering (
Three-dimensional light sheet microscopy of mouse brain (
Under the 785 nm LS excitation, the maximum 1500-1700 nm fluorescence signal detected in the cerebral vasculatures of mouse brain cortex layer as a function of depth Z showed two attenuation regions (
The capability of NIR-II LSM performing volumetric imaging through scattering tissues at the 1-10 mm3 scale enabled non-invasive in vivo 3D imaging of protruding features on live mice related to disease models (
To further exploit NIR-II LSM, in vivo two-color molecular imaging and vasculature imaging of PD-L1 expressing MC38 tumor (Tang et al., Cancer Cell 29: 285 (2016)) was performed using a renal excretable organic dye (Zhang et al., Adv. Mater. 28; 6872 (2016); Weizhi et al., Adv. Mater. 0, 2400106 (2018)) (TT dye: excitation approximately 785 nm/emission approximately 1000-1200 nm) conjugated to anti-PD-L1 antibodies and unconjugated PEGylated PbS/CdS CSQD (excitation approximately 1319 nm/emission 1500-1700 nm) intravenously injected into tumor bearing mice. PD-L1 is an important immune checkpoint protein expressed by tumors as a powerful way of T cell immunity evasion. PD-L1 blocking by antibody immunotherapy is a novel approach to treat various cancers. In vivo PD-L1 imaging is important to fundamental understanding of cancer immunity, and to immunotherapy prognosis since treatment efficacy correlated with PD-L1 levels in the tumor (Balar & Weber, Cancer Immunol. Immun. 66: 551 (2017); Song et al., Plos One 8, e65821 (2013)). 24 h post injection of anti-PD-L1-TT dye, wide-field imaging was first performed resulting in much brighter anti-PD-L1-TT dye signals in MC38 tumors (36) (
PD-L1 is an important immune checkpoint protein expressed by certain tumors as a powerful way of cancer evasion of a body's immune surveillance through PD-L1 (on tumor) and PD-1 (on T cells) binding (Balar & Weber Cancer Immunol. Immun. 66: 551-564 (2017)). PD-L1 blocking by antibody immunotherapy is a breakthrough approach to activate the immune system and treat cancer. In vivo PD-L1 imaging/mapping is important to fundamental understanding of cancer immunity and immunotherapy mechanism (Balar & Weber Cancer Immunol. Immun. 66: 551-564 (2017); Song et al., Plos One 8: e65821 (2013)). To this end, a 45° oblique NIR-II LSM was developed (
An erbium-based rare-earth nanoparticles (ErNPs) (Zhong. et al., Nat. Commun. 8: 737 (2017)) (excitation 975 nm/emission 1500-1700 nm, see Method) conjugated to PD-L1 antibodies were injected intravenously and 24 h later, PEGylated PbS/CdS CSQD conjugated to anti-PD-1 (excitation about 1319 nm/emission 1500-1700 nm) were injected. At 2 h post injection, dynamic imaging into the tumor at a fixed illumination plane by oblique NIR-II LSM (
Strong scattering of light by scalp/skull typically required scalp removal (Wang et al., Nat. Methods 15: 789-792 (2018); Kawakami et al., Biomed. Opt. Express 6: 891-901 (2015)). To demonstrate non-invasive oblique NIR-II LSM into non-protruding tissues, an intact mouse head was imaged through the layers of scalp, skull, meninges and brain cortex, 2 h post injection of PEGylated PbS/CdS CSQD probes (
Further, in a mouse model of TBI (Zhang et al., Adv. Mater. 28: 6872-6879 (2016)) (
The disclosure encompasses 3D near-infrared light sheet microscopy for in vivo and ex vivo deep tissue volumetric imaging through highly scattering biological tissues. Light sheet microscopy with both excitation and emission in the NIR-II window avoided shadows or stripes along the illumination direction by suppressing tissue scattering and adsorption effects encountered by visible LSM (Russo et al., Science 353: 783-785 (2016)). Non-invasive NIR-II LSM enabled in vivo observation of dynamic processes and molecular imaging at cellular scale. NIR LSM imaging can be further advanced by developing brighter fluorophores with added colors and applying new configurations of LSM (Itoh et al., Opt. Lett. 41: 5015-5018 (2016)). Recent developments such as non-coherent structured illumination (Tomer et al., Cell 163: 1796-1806 (2015)) and optical lattices illumination (Liu et al. Science 360, 1392 (2018)) could be introduced to improve the resolution and contrast of NIR-II LSM. Real time molecular imaging of multiple targets by rapid sectioning 3D tissues of live mammals should become possible.
Accordingly, one aspect of the disclosure, therefore, encompasses embodiments of a method of optically sectioning a biological sample, the method comprising the steps: (a) contacting a biological sample with a fluorescent contrast agent having an excitation wavelength of between about 785 nm to about 2400 nm and a fluorescence emission having a wavelength of between about 800 nm to about 2400 nm; (b) irradiating the biological sample with at least one excitation light having a wavelength of between about 785 nm to about 2400 nm, wherein the excitation light is configured as a static light sheet and directed through a first plane of the biological sample; (c) orthogonally detecting an emitted fluorescence having a wavelength of between about 800 nm to about 2400 nm from the irradiated biological sample; and (d) generating a first digital image of the fluorescence from the irradiated first plane through the biological sample.
In embodiments of this aspect of the disclosure, the biological sample can be an isolated cell or population of isolated cells, a cultured cell or population of cultured cells, an isolated tissue or organ, or an animal or human subject.
In some embodiments of this aspect of the disclosure, the biological sample has raised features and the irradiated biological sample is imaged by transmitting the light sheet through the raised feature from a side thereof and detecting the emitted fluorescence at right angles to the plane of the illuminating light sheet.
In some embodiments of this aspect of the disclosure, the biological sample can have substantially planar regions and the irradiated biological sample can be illuminated by the light sheet at an angle with respect to the normal to the planar region and the emitted fluorescence is detected at right angles to the plane of the illuminating light sheet.
In some embodiments of this aspect of the disclosure, the fluorescent contrast agent can be administrated by vascular delivery to a tissue or organ of the animal or human subject.
In some embodiments of this aspect of the disclosure, the fluorescent contrast agent can be conjugated to a targeting ligand that can specifically bind to a molecular target.
In some embodiments of this aspect of the disclosure, the targeting ligand can have an affinity for a molecular target and is selected from an antibody, a peptide, an aptamer, or a nucleic acid.
In some embodiments of this aspect of the disclosure, the targeting ligand can be an antibody or fragment thereof selectively binding to an epitope of a polypeptide selected from the group consisting of: Programmed cell death protein 1 (PD-1), Programmed death-ligand 1 (PD-L1), and CD11b.
In some embodiments of this aspect of the disclosure, the fluorescent contrast agent can emit at a wavelength of between about 900 to about 2400 nm and can be an organic molecular dye, a conjugated polymeric dye, a polymer micelle-wrapped organic nanofluorophore, a carbon nanotube, a quantum dot, or a rare-earth down-conversion or up-conversion nanoparticle.
In some embodiments of this aspect of the disclosure, the method can further comprise the steps: (i) repeating steps (b)-(d), thereby irradiating a plurality of parallel planes perpendicular to the light sheet plane at various depths through the biological sample, and generating a plurality of digital images; and (ii) digitally combining the plurality of digital images to generate a three-dimensional image of the location of the fluorescence emitted by the contrast agent in the biological sample.
In some embodiments of this aspect of the disclosure, the excitation light can be delivered to biological sample by an illumination objective, wherein the numerical aperture (N.A) of the objective is configured to deliver the excitation light as a light sheet having a balanced waist thickness of between about 5 μm to about 20 μm and a Rayleigh length of between about 0.1 mm to about 6.0 mm.
In some embodiments of this aspect of the disclosure, the excitation light can be generated by a laser having a wavelength of between about 700 nm and about 2400 nm.
Another aspect of the disclosure encompasses embodiments of a light sheet microscope comprising along an optical axis: an illumination objective positioned to direct an excitation light sheet through a plane of a biological sample; a plurality of achromatic lenses optimized for transmission of light between about 785 nm to about 2400 nm; a proximal first adjustable mechanical slit; a cylindrical lens; a distal second adjustable mechanical slit adjacent to the cylindrical lens and distal to the illumination objective, wherein the slit of said second slit is orientated at right-angles to the slit of the first slit; a pinhole; at least one excitation light source in the 600-2400 nm range; and at least one removable mirror disposed to direct an excitation light from the at least one light source along the optical axis of the illumination objective, the plurality of achromatic lenses, the proximal first adjustable mechanical slit, the cylindrical lens, the distal second adjustable mechanical slit, and the pinhole; a detection objective disposed to orthogonally receive fluorescent light emitted from a target irradiated by a light sheet from the illumination objective and to direct said fluorescent light to a detector operably connected to a computer system for generating a digital image of the fluorescent light; and at least one emission filter configured to only transmit fluorescent light having a wavelength of between about 785 nm to about 2400 nm.
In some embodiments of this aspect of the disclosure, the light sheet microscope can further comprise a right-angle prism disposed between the biological sample and the illuminating and receiving lenses.
In some embodiments of this aspect of the disclosure, when the illuminating light sheet is configured to illuminate the biological sample at an angle of less than 90° between a tangential plane of the biological sample and the illuminating light sheet, a solid, liquid or gas having a refractive index of about that of the biological sample can be disposed between the biological sample and the illuminating and receiving lenses.
In some embodiments of this aspect of the disclosure, the excitation light source can further comprise a shortpass filter to select the excitation wavelength.
In some embodiments of this aspect of the disclosure, the excitation light source can be a laser.
In some embodiments of this aspect of the disclosure, the excitation light source can emit an excitation light having a wavelength of between about 785 nm and 2400 nm. In some embodiments of this aspect of the disclosure, the detector can be sensitive to light having a wavelength of between about 800 nm and about 2400 nm.
In some embodiments of this aspect of the disclosure, the detector can be an InGaAs camera sensitive to light having a wavelength of between about 800 nm and about 2400 nm.
In some embodiments of this aspect of the disclosure, the detector can be a small bandgap semiconductor-based camera sensitive to light having a wavelength of between about 800 nm to about 2400 nm.
As mentioned above, compounds of the present disclosure and pharmaceutical compositions can be used in combination of one or more other therapeutic agents for treating viral infection and other diseases. For example, compounds of the present disclosure and pharmaceutical compositions provided herein can be used in combination with other anti-viral agents to treat viral infection.
While embodiments of the present disclosure are described in connection with the Examples and the corresponding text and figures, there is no intent to limit the disclosure to the embodiments in these descriptions. On the contrary, the intent is to cover all alternatives, modifications, and equivalents included within the spirit and scope of embodiments of the present disclosure.
NIR-II fluorescence probes: Organic nanofluorophore p-FE dye PEGylated core-shell quantum dots PbS/CdS CSQD (
150 μg anti-mouse PD-1 (BioXCell; cat. #: BE0146; clone RMP1-14) was dissolved in 300 μL PBS buffer. 6 μL EZ-LINK® Sulfo-NHS-LC-Biotin (1.7 mg/mL in DMSO) was added to the solution, and the solution was stirred at room temperature for 1.5 hr. Unreacted biotin was removed by a centrifugal filter (cutoff=100 k). The final product (referred to anti-mouse-PD-1-biotin) was suspended in 100 μL PBS buffer.
Prepared PbS/CdS-streptavidin and anti-mouse-PD-1-biotin was mixed, and stirred at room temperature for 2 hr. Excess antibody was removed by a centrifugal filter (cutoff=300 k). Similar method was used to conjugate anti-mouse/human-CD11 b (BioXCell; cat. #: BE0007; clone M1/70) to PbS/CdS CSQD. The TT dye (Zhang et al., Adv. Mater. 28: 6872-6879 (2016) exhibits similar spectroscopic properties as the p-FE dye with an excitation at approximately 785 nm and emission of approximately 1000-1200 nm. The TT dye contains an azide group that can be used for conjugation to PD-L1 antibodies by click chemistry for molecular imaging. Purified anti-mouse-PD-L1 antibody was purchased from Selleckchem. Conjugation was through copper-free click chemistry using linker DBCO-PEG4-NHS (Click Chemistry Tools).
NIR-II light sheet microscope system: As show in
This spatial filter improved the circularity and quality of the illumination beam to generate a uniform light sheet across the field of view. The excitations could be selected by removable mirrors (M2-M4).
Excitation power was measured by a laser power meter (3A-SH, NOVA II, OPHIR). Before the laser entering a cylindrical lens (CL), a vertically arranged adjustable mechanical slit parallel to the CL was used to adjust the span range of light sheet along Y-axis direction (
For in vivo imaging, the actual excitation intensity illuminated on the target mouse was approximately 0.8 W cm−2, approximately 1.1 W cm−2 and approximately 3.4 W cm−2 the for 658 nm, 785 nm and 1319 nm laser, respectively, which is below the safety limit for laser exposure shorter than 10 s (658 nm: 1.1 W cm−2; 785 nm: 1.6 W cm−2, 1319 nm: 5.5 W cm−2).
For the oblique LSM (
The light sheet was positioned to pass through a tissue sample or a tumor protrusion on a mouse or flat mouse head, fluorescence imaging was done by focusing on and normal to the light sheet plane through a detection objective (02) and a 200-mm tube lens, using a liquid-nitrogen-cooled InGaAs camera (2D-OMA V, Princeton Instruments) or a water-cooled InGaAs camera (Ninox 640, Raptor Photonics) after filtered by selected emission filters. The focal lengths of L1, L2, L3, L4 were 60 mm, 100 mm, 30 mm and 60 mm, respectively. All the optical components were made by Thorlabs. For illumination objective, a 5× objective (N.A=0.15, Nikon LU Plan) or a 10× objective (N.A=0.25, Bausch & Lomb Optical Co.) was used. For imaging, a 4× objective (N.A=0.1, Bausch & Lomb Optical Co.), a 10× objective (N.A=0.25, Bausch & Lomb Optical Co.), or a 50× objective (NA=0.6, Nikon CF Plan) was used.
Light sheet shape, resolution and field of view considerations: Two orthogonal slits were mounted in the illumination arm to adjust the light sheet shape by changing the effective numerical aperture NA and the span range along Y-axis direction (
The effective NA was estimated using NA=n sin α=n sin(arctan(D/2f)), where n is the refractive index, α is the half of aperture angle, D is the illumination width of light sheet (adjusted by slit S1 for LSM imaging) as it exiting the illumination objective, and f is the focal length of illumination objective. At a given width of slit S1, D was measured experimentally by putting a scattering paper close to the aperture of the illumination objective. D was adjusted by S2 when the cylindrical lens was rotated by 90° for imaging the side view of the light sheet (
The resolution of NIR-II LSM is limited by the diffraction limit and scattering that still existed in NIR-II (
Using the 0.5-NA 50× detection objective, the diffraction limited resolution along Z was 2.6 μm (850-1000 nm), 3.2 μm (1100-1200 nm), and 4.4 μm (1500-1700 nm). These analysis suggested that the Z resolution of the current NIR-II LSM is down to about 10 μm, suitable for single-cell resolution along Z.
X-Y diffraction limited resolution was higher, about 2.3 μm (850-1000 nm), 2.8 μm (1100-1200 nm) and 3.9 μm (1500-1700 nm) using the 10×, 0.25-NA detection objective; and 0.9 μm (850-1000 nm), 1.2 μm (1100-1200 nm) and 1.6 μm (1500-1700 nm) using the 50×, 0.6-NA detection objective (estimated by Rayleigh criteria, 0.61λ/NA).
The LS waist and Rayleigh length are contradicting factors, optimizing one means degrading performance in the other. The selecting actual NA for each experiment were tradeoffs of these two parameters to obtain uniform LS that are as thin as possible across a large enough FOV. The actual NA and corresponding waist and confocal length for each data set were summarized in Table 2.
Light sheet microscopy 3D volumetric imaging/scanning: For 3D normal NIR-II LSM imaging (
Different detection windows were selected by using corresponding long-pass and short-pass filters. Synchronous control of 3D translation stage movement and image record was realized using LabView software through a data acquisition card (NI USB-6008, National Instruments).
For oblique LSM (
Maximum intensity projections (
Mouse handling and tumor xenograft: C57BL/6 and BALB/c female mice were from Charles River. 4-6-week-old C57BL/6 (or BALB/c) mice were shaved using hair removal lotion and inoculated with about 1 million MC38 (or CT 26) cancer cells on the right flank of back near the hind limb or on the ear for tumor growth. The sample sizes of mice were selected based on previously reported studies. Mice were randomly selected from cages for all experiments. During in vivo imaging, all mice were anaesthetized by a rodent anesthesia machine with 2 l min−1 O2 gas mixed with 3% isoflurane.
Ex vivo NIR-II light sheet microscopy of mouse brains: For ex vivo LSM of mouse brains in various NIR-1 and NIR-II regions (
For NIR-II imaging mouse brains (
For a 5× illumination objective and the 4× (
In vivo wide-field NIR-II fluorescence imaging: The NIR-II wide-field fluorescence images in
For three-color wide-field imaging of PD-L1, PD-1 expressing cells and vessels in CT26 tumor bearing BALB/c mouse, 200 μL anti-PD-L1-Er (10 mg/mL) were injected intravenously. After 24 h, 200 μL anti-PD-1-PEGylated PbS/CdS CSQD (O.D.=0.5 at 808 nm) were injected and in another 29 h, 250 μL p-FE (O.D.=5 at 808 nm) was injected into the tail vein.
Wide-field imaging (
In vivo NIR-II light sheet microscopy of tumors: For three-plex light sheet imaging of PD-L1, PD-1 expressing cells and vessels in CT26 tumor on BALB/c mouse, 200 μL anti-PD-L1-Er (10 mg/mL) was first injected intravenously. After 24 h, 200 μL anti-PD-1-PEGylated PbS/CdS CSQD (O.D.=0.5 at 808 nm) was injected. In another 29 h, 250 μL p-FE (O.D.=5 at 808 nm) was injected into the tail vein to label the vessels. 2 min post injection of p-FE, non-invasive oblique NIR-II LSM (
The 1500-1700-nm fluorescence signal emitted from PD-1 labeled by PbS/CdS CSQD was recorded under CW 1319-nm light sheet excitation and 100-ms exposure time, during which no fluorescence emitted from ErNPs due to the lack of absorption at 1319 nm. Then illumination was switched to 14 ms long 975 nm pulse illumination and then switched off. After a wait time of 1 ms, optical sectioning/image recording of 1500-1700-nm fluorescence only detected the long life-time ErNPs for PD-L1 imaging using an exposure time of 100 ms. Fluorescence images recording at every optical sectioning was realized by synchronously controlling a 1319-nm laser, a 975-nm laser, a motorized stage and the camera using Labview software through a data acquisition card (NI USB-6210). Then another scan was performed to record p-FE signal in 1100-1300 nm in vessels with 100-ms exposure. The scanning step was 4 μm along the X direction. The one-step movement required 100 ms and was confined by the stage (M-VP-25XL, Newport).
For results shown in
In vivo NIR-II light sheet microscopy of mouse head: To image (
Traumatic brain injury (TBI): The TBI was performed based on reported procedures with modifications (Zhang et al., Adv. Mater. 28: 6872-6879 (2016)). Briefly, the benchmark stereotaxic impactor was mounted on a stereotaxic frame at 45° (David Kopf Instruments, Tujunga, Calif., USA). Anesthetized mice were placed on a customized foam mold in a prone position. To induce TBI, the tip was driven towards the mouse head at a speed of 4.0-4.5 mm/s, a dwell time of 0.2 s set by the electronic control box, and an impact depth of 3 mm adjusted by the stereotaxic device. After 2 h recovery, the 4-6-week-old C57BL/6 mouse was intravenously injected with 200 μL anti-CD11b-PEGylated PbS/CdS CSQD (O.D.=0.5 at 808 nm) and monitored by wide-field system and oblique NIR-II LSM mounted with a 5× illumination objective and a 10× imaging objective 24 h post injection. The Z scanning increment, exposure time and excitation and emission wavelengths are summarized in Table 2.
Statistics and data analysis: Data analysis was performed in MATLAB 2017 or Origin 9.0. The standard deviation and mean shown in
Study of light sheet propagation in different media: Light sheet propagation in glycerol solutions was studied using light sheets with different NA and excitation wavelengths. Experiments were performed in glycerol containing PEGylated PbS/CdS CSQD uniformly dispersed in glycerol. The emission was collected in 1500-1700-nm window excited by 658 nm, 785 nm and 1319 nm LS illuminations. To directly observe light transmission in glycerol, the cylindrical lens was rotated by 90° and mechanical slits were used to control the NA and the spanning range along Y (
To study the propagation of the light sheet with wavelength of 658 nm, 785 nm and 1319 nm in a scattering medium, imaging of LS propagation was performed in intralipid solutions of different concentrations (
μs(λ)=0.016λ−2.4 (1)
μs′=10.094×conc.+0.433 (2)
g(λ)=1.1−0.58λ (3)
where A is the wavelength, μs′=μs(1−g) is the reduced scattering coefficient and conc. is the concentration of intralipid.
Equations 1 and 3 were only used for 10.00% intralipid solution. The available spectral range of Equation 2 is between 750 nm to 830 nm. The μs of 10% intralipid solution was first calculated at different wavelengths using Equation 1. Then the μs of 1.25%, 2.50% and 5.00% intralipid solutions was obtained based on the linear relationship between μs and concentration of Equation 2 and μs′=μs(1−g). These parameters are summarized in Table 1.
The illumination waist measured in water at NA=0.039 was used as initial FWHM of incident light in Monte Carlo simulations. The simulated results were consistent with the experimental observations (
The simulation results were compared with the experimental observations in glycerol-cleared mouse brain in
The LS excitation intensity along incident direction is another important parameter for imaging in scattering tissue, as it affects the transmission distance of excitation in the tissue and determines the illumination field. As the intralipid concentration increased, the intensity along propagation direction attenuated faster but the 1319 nm excitation decayed the slowest compared to the 658 nm and 785 nm excitations (
For high-quality optical sectioning in LSM, both uniform light sheet waist and available illumination field should be ensured across the field of view.
The μs, g of intralipid at different concentrations and wavelengths were calculated by Equations 1-3. The scattering coefficient of brain for excitation wavelength was mimicked by 2.5% intralipid and the anisotropy factor of brain was from the value measured by Shi et al., (J. Biophotonics 9: 38-43 (2016)). The μs in the various emission wavelengths were extracted from
850-1,000
850-1,000
850-1,000
850-1,000
An animation was constructed with images taken at various depth Z in mouse brain tissue by NIR light sheet microscopy under the same 785 nm excitation while detecting emissions in NIR-1 (850-1000 nm, p-FE), NIR-IIa (1100-1200 nm, p-FE) and NIR-IIb (1500-1700 nm, PEGylated PbS/CdS CSQD probes) respectively (as shown in
NIR light sheet images in the X-Y plane taken at various depth Z in mouse brain tissue by detecting 1500-1700 nm emission of PbS/CdS core-shell quantum dots in vessels under various excitations using 658 nm, 785 nm and 1319 nm light sheets respectively. Such imaging utilized an important property of quantum dots, i.e., ultra-wide excitation ranges (ultraviolet to greater than 1300 nm,
NIR light sheet images in the X-Z plane (Yin the range of 0-640 μm with 5 μm increasement) reconstructed from X-Y images at various depth Z in mouse brain tissue by detecting 1500-1700 nm emission (PbS/CdS core-shell quantum dots) under various excitations using 658 nm, 785 nm and 1319 nm light sheets respectively. Suppressed scattering of longer wavelength LS was gleaned from X-Z cross sectional images, with reduced FWHM of feature sizes along the depth Z direction, corresponding to higher vertical resolution and better sectioning capability along Z.
Ex vivo NIR-II Light sheet microscopy images of vasculatures in mouse brain as shown in
In vivo time-course LSM imaging of blood perfusion into tumor vasculatures by recording the p-FE nanofluorophore (200 μL, O.D.=4 at 808 nm, 785 nm excitation, 1000-1200 nm detection) signals immediately following intravenous injection into the mouse tail-vein. For imaging, light sheet sections were made through the tumor (at a fixed plane and below the top of the xenograft MC38 tumor of approximately 8 mm in diameter) at a depth of Z of approximately 300 μm. Imaging was recorded through the same plane at approximately 1 fps. The LS illumination position Z was controlled by a 3D motorized translation stage. A 4× detection objective and a 5× illumination objective were used. Blood flows in tumor vasculatures were found irregular and intermittent with turning-on and shutting-off behavior, oscillatory/fluctuating flowing patterns and even flow direction reversal in the same vasculature in a tumor.
In vivo light sheet microscopy of anti-PD-L1-TT and vasculatures in a MC38 xenograft tumor on a mouse for the data in
This application claims priority to and the benefit of U.S. Provisional Application 62/682,322 titled “NEAR INFRA-RED LIGHT SHEET MICROSCOPY THROUGH SCATTERING TISSUES” filed Jun. 8, 2018 and to U.S. Provisional Application 62/828,728 titled “LIGHT SHEET MICROSCOPY IN THE NEAR-INFRARED II WINDOW” filed Apr. 3, 2019, the entire disclosures of which are incorporated herein by reference in their entireties.
This invention was made with Government support under contracts NS105737 awarded by the National Institutes of Health. The Government has certain rights in the invention.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US19/35346 | 6/4/2019 | WO | 00 |
Number | Date | Country | |
---|---|---|---|
62682322 | Jun 2018 | US | |
62828728 | Apr 2019 | US |