This invention relates generally to the field of cardiovascular medicine, and in particular to the treatment of patients suffering from a spectrum of cardiac states, ranging from shock to pulseless electrical activity, in which the patient appears to be lifeless and in cardiac arrest yet retains some mechanical activity in the myocardial wall motion.
One common technique for treating persons suffering from cardiac arrest is the use of cardiopulmonary resuscitation (CPR). In this procedure, the patient's chest is repeatedly compressed, often in combination with periodic ventilations. Administration of electrical countershock and drugs intended to assist in restoration of cardiopulmonary function to chest compression and ventilation, constitutes advanced life support. For a variety of reasons, the effectiveness of CPR has been limited. Hence, devices or techniques which can improve the effectiveness of CPR are greatly needed.
In additional to sudden cardiac arrest, refractory-shock (which is referred to herein as “shock”) is often fatal. For example, if not properly stabilized, a person suffering from shock can progress into cardiac arrest, which, because it is not sudden in nature, is usually fatal. Emergency medicine and critical care practitioners approach the treatment of shock principally by attempting to alleviate the cause because there are no non-invasive techniques that may beneficial in assisting circulation. Hence, devices and techniques are also needed to treat those suffering from refractory shock and shock that is progressing toward cardiac arrest.
There is no general consensus as to when it is the appropriate to start administering CPR as the patient's blood pressure progressively decreases. This relates to a lack of demonstrated efficacy and concern that chest compression may interfere with residual cardiac function, even though CPR may at some point be beneficial in shock patients as they progress to cardiac arrest. Hence there a need for a device or technique to prevent CPR from interfering with residual cardiac function.
Unlike cardiac arrest caused by ventricular fibrillation, pulseless electrical activity (PEA) is a heterogeneous entity with respect to cardiac function and hemodynamics. PEA is a clinical condition characterized by unresponsiveness and lack of palpable pulse in the presence of organized cardiac electrical activity. Pulseless electrical activity has previously been referred to as electromechanical dissociation (EMD). During PEA, electrical activity of the heart may or may not be indicative of cardiac mechanical movements and particularly cardiac output.
Pulseless electrical activity is not necessarily a condition of complete mechanical quiescence in the heart. In PEA, the heart may have a regular organized electrical rhythm such as supraventricular or ventricular rhythms. These cardiac rhythms may not be associated with mechanical activity of the heart in PEA.
As an example of cardiac mechanical patterns during PEA, patients may have weak ventricular contractions and detectable aortic pressure—which is a condition referred as pseudo-PEA. Various studies have documented that between 40-88% of patients with PEA had residual cardiac mechanical activity (pseudo-PEA). In pseudo-PEA, the patient may appear lifeless and without a pulse, despite some degree of residual left ventricle function and hemodynamics. The outcome of patients suffering PEA has tended to be worse than those in ventricular fibrillation, possibly reflecting the potential of CPR chest compressions and residual myocardial mechanical activity to interfere with each other's efficacy. Hence there is a need for a device or technique to enhance the efficacy of CPR in PEA.
Disclosed herein are techniques and systems for treating those suffering from a variety of myocardial pathophysiologic states ranging in hemodynamics from awake patients in refractory shock to those who appear to be lifeless, yet who still retain some degree of residual myocardial mechanical function. It has been observed when performing open chest cardiac massage, that coordinating compression and relaxation with the heart's residual mechanical activity often improves recovery of cardiac function. Extrapolating from this, if mechanical myocardial function is present but inadequate, as in PEA, external chest compressions should likely be directed toward assisting cardiac ejection—that is compressing the chest during its intrinsic contractions—and then releasing the chest so as not to interfere with ventricular filling. CPR that is not synchronized with the heart's residual mechanical function may result in application of the compression phase when the left ventricle is trying to fill, resulting in significantly decreased cardiac output on the next ejection secondary to the Frank-Starling Law. Interference with ventricular filling by compression of the chest can be so deleterious that it can, in and of itself, cause complete loss of residual myocardial function resulting in true cardiac arrest.
A system is disclosed here that detects residual myocardial activity in an apparently lifeless patient and outputs signals to trigger chest compressions by mechanical chest compression devices; to audibly indicate when to initiate such chest compressions, or to other interventions that benefit from synchronization with residual myocardial activity. These other interventions may include but are not limited to: abdominal counter-pulsation, ventilation, phasic limb-compression, myocardial electrical stimulation, intravascular fluid shifting, intravascular balloon inflation-deflation, intra-esophageal or intra-pericardial balloon inflation, application of transthoracic electromagnetic irradiation, and the like.
A method is disclosed here for improving the cardiac output of patients suffering from a range of pathophysiologic states such as pulseless electrical activity or shock, which having some residual myocardial wall mechanical activity. According to the method, residual myocardial activity is sensed to determine the presence of residual ventricular phasic movement, with or without residual left or right ventricular pump function, but having an apparent ejection phase and a relaxation phase. A compressive force is repeatedly applied based on the sensed myocardial activity such that, for example, the compressive force is applied during at least some of the ejection phases and is ceased during at least some of the relaxation phases to permit cardiac filling, thereby creating or enhancing cardiac output and organ perfusion. The synchronization with the sensed myocardial activity may also be used when the patient's chest is actively lifted during decompression. In this way, the chances for improving the outcome of patients suffering from shock or cardiac arrest are improved.
The compressive force may be applied over a variable range of time intervals. For example, the compressive force may be applied for only a certain portion of the contraction or ejection phase, such as at the beginning, middle or end. As another example, the compressive force may be applied during each and every sensed contraction or ejection phase, or only during certain contraction or ejection phases.
The start of the chest compression and the duration of the compression can be adjusted to improve patient outcomes. For example, the adjustments to the start time or duration may be adjusted to optimize the chest compressions or other phasic therapy, where the adjustment is based on feedback of a patient condition or physiologic parameter during one or more prior chest compressions. The feedback signal may, for example, indicate a rate or amount of cardiac ejection or filling, cardiac output or other indicator of mechanical activity of the heart or arterial blood flow. The feedback signal is coupled to the therapy by logic circuits so as to vary the synchronized phasic therapies, e.g., chest compressions, and vary the application of the therapies. By varying the therapies and their application and subsequently re-measuring the feedback signals, the logic circuits can determine which synchronized therapy, or therapies, and pattern of synchronized therapy is optimal and most effective to improve cardiac ejection, cardiac output or otherwise improve the condition of the patient. For example, the logic circuit may vary each of the synchronized therapies and combinations of therapies to determine which pattern of therapy or therapies when synchronized with residual myocardial synchronization results in the greatest measured cardiac output or results in some other measurable condition that indicates that the phasic therapy(ies) are being applied optimally.
Electrical stimulation of the heart may be applied in conjunction with or in addition to chest compressions. The electrical stimulations may be synchronized with electrical signals (ECG/EKG) of the inherent heartbeat, which may be slow and weak, or if there are no regular electrical heart signals, with pulsatile flow or myocardial movements. For example, the electrical stimulation may be synchronized with arterial pulses, such as aortic pressure (AoP), based on detected pulsatile pressure, blood flow, or myocardial movements.
Ventilations are another phasic therapy that may be applied to the patient based on the sensed myocardial activity or hemodynamics. The patient may be ventilated manually or by a mechanical ventilator. The ventilation may be synchronized with chest compressions or other resuscitative therapies in conditions such as shock or pseudo-PEA.
The compressive force may be applied using a variety of devices or equipment. Some examples include mechanical chest compression devices, inflatable vests, nerve stimulators, abdominal compression devices, chest or abdominal active decompression devices, limb phasic compression devices, and the like. Further, the compressive force may be applied at different locations on the chest, abdomen, limbs, or back, such as the left lateral chest, cardiac point of maximal impulse and the like.
The myocardial activity may be sensed using a variety of sensing systems. Such systems may include electrocardiography, Doppler ultrasonography, plethysmography, phonocardiography, echocardiography, transthoracic impedance and the like. These may be incorporated into a probe that is coupled to the chest, abdomen, back, extremities, or a combination of these, or placed within the body, such as within the esophagus, trachea, or stomach. These various types of sensors may detect myocardial activity by detecting, for example, cardiac electrical activity, contractions, other movements of the heart, and palpable pulses of arteries. These measurements may be made from standard locations such as the precordium, but also from the esophagus, trachea, or abdomen. Variations in the skin indicative of pulsating blood flow, and the rhythm and chemical content of the breath, may also be utilized.
The sensors and algorithms optimally suited to sense myocardial activity may depend on characteristics of a particular patient. Further, the sensors optimally suited to sense myocardial activity may change during the course of treating. To determine the optimal sensor or sensors that best indicate myocardial activity, the system may include algorithms to validate the sensor(s) and to correlate sensor output data to a desired patient response, such as improved cardiac output. To validate the sensors, the system may apply or prompt the application of therapies such as chest compressions at a predetermined rate, force or vector and compare the outputs of the sensors to expected sensor outputs or otherwise determine which sensor(s) generate signals that most accurately indicate the response of the patient to the predetermined chest compressions. The validation of sensors results in an identification of sensors and arrangements of sensors that generate signals that most accurately measure or predict the response of the patient. The sensors may be validated at the initiation of therapy and may be revalidated periodically during treatment of the patient, such as at regular intervals or when a substantial change, e.g., beyond a threshold amount, occurs in the response of the patient to treatment.
The validated sensors or validated arrangement of sensors are those sensors that have been determined to most accurately measure or predict a predetermined response(s) of the patient. Once the sensors have been validated, signals generated only by the sensors, or pattern of sensors, identified in the validation process are used to provide feedback to the algorithms that determine the application of phasic therapies such a chest compressions and ventilation. Using these signals, the algorithms may generate and adjust a regimen for chest compressions and ventilations of the patients. The regimen may dictate the force to be applied by the chest compressions, the frequency of the chest compressions, the shape and duration of the force applied by the chest compressions, the synchronization and phasing of the chest compressions with sensed myocardial activity, the location on the chest or other body location, e.g., legs, of compressions, and a vector of the chest or other compressions. The algorithms may vary the regimen to optimize a condition of the patient, such as to increase sensed cardiac output.
In some cases, chest compressions may be performed manually, such as using traditional CPR techniques. In such cases, an audio or visual signal may be produced to indicate when the ejection phase is sensed. The generated signals may indicate to a rescuer when to apply the chest compressions, whether to apply more or less force during the compressions, or whether to apply the compressions to a different location on the chest. In this way, the rescuer will be prompted as to when, how and where to apply the compressive force to the patient. The tone, volume, or other parameter, of the synchronizing prompt may be varied so as to assist the rescuer in providing optimal CPR. In some cases, the chest, abdomen, or extremities may also be actively or passively compressed or decompressed in an alternating manner with chest compressions, and in synchronization with either cardiac ejection or filling.
A system is disclosed here for improving the cardiac output and prognosis of a patient who is suffering from impaired myocardial mechanical states such as pulseless electrical activity or shock but having residual myocardial wall motion. The system comprises a myocardial activity sensor that is adapted to sense movement of the myocardial wall and or myocardial valvular motion to determine the presence of residual ventricular contract and relaxation, and/or pump function having an ejection phase and a filling phase. The system may also include a compression device that is configured to repeatedly apply a compressive force to the heart, either through the chest wall, intrathoracically through the pericardium, or directly to the myocardium through an endoscope and pericardial window. Further, a controller is employed to receive signals from the myocardial activity sensor and to control operation of the compression device such that the compression device repeatedly applies a compressive force to the heart such that the compressive force is applied during at least some of the ejection phases and is ceased during at least some of the relaxation phases to permit residual cardiac filling, thereby enhancing cardiac output and organ perfusion.
As an option to using a mechanical compression device or as an initial treatment applied before the compression device is setup on a patient, chest compressions may be performed manually. In such cases, the system may include a cadence device that is configured to produce audio and/or visual signals indicative of when compressive forces are to be applied and ceased. This same cadence system may be utilized to synchronization other therapies phasic therapies such as ventilation or abdominal counterpulsation.
The myocardial activity sensors that may be used include electrocardiography sensors, Doppler ultrasonography sensors, plethysmography sensors, phonocardiography sensors, echocardiography sensors, transthoracic impedance sensors, magnetic resonance imaging, and radiographic fluoroscopy. These sensors may be placed on the patient's chest, abdomen, back or extremities, within body cavities such as the esophagus, or some distance from the patient in the case of technologies like radiography or magnetic resonance imaging. If the patient has an arterial pressure catheter in place, the controller may also utilize that signal for synchronization. Further, the controller may be configured to apply the compressive force during each sensed ejection phase or during only at certain ejection phases. As another option, the controller may be configured to apply the compressive force for only a certain duration of the ejection phase.
The system may further include a ventilator device that is configured to provide ventilation to the patient based on the sensed residual myocardial mechanical activity. The controller may also vary the pattern of individual ventilations so as to optimize synchronization.
A sensor may detect the expansion and relaxation of the chest due to ventilation or chest compressions. The sensor may be a plastic adhesive strip applied to the chest that stretches and contracts with the movement of the chest due. The stretch and contraction of the adhesive strip may be detected as a change in an electrical property, e.g., resistance, of the strip, optically due to a change in transmissivity or reflection of the strip or by other means. The stretch and contraction of the adhesive strip causes the adhesive strip sensor to generate signals indicative of the expansion and relaxation of the chest. These signals may be used by the algorithms to predict when blood is being drawn into the heart as the chest relaxes (expands) or when blood is being forced from the heart as the chest is compressed.
The phasic device may be a mechanical compression device, an inflatable vest, a nerve stimulator, or the like. Further, the system may include a lifting device that is configured to actively decompress the chest during the relaxation phase, or compress the abdomen during chest decompression.
In another embodiment, a logic circuit may be used to vary the phasic therapeutic device or devices such that the optimal pattern and combination can be determined and applied. This pattern may be variable over time and the invention will monitor for the possibility by occasionally varying the pattern of therapies and adjusting according to an indicator of hemodynamics or predictor of outcome.
During resuscitation of patients suffering cardiac arrest, the presence and degree of residual left ventricular mechanical (physical) activity may vary over time. The system may be configured to detect transient periods of left ventricular mechanical activity and to synchronize therapies only during these periods to assist residual cardiac mechanical activity and achieve a greater cardiac output.
The sensor functions may be utilized to determine the vector of left ventricular ejection and to optimize the force vector of chest compression spatially. This might be done utilizing an array of Doppler probes placed over the chest to detect the velocity of residual myocardial motion from multiple locations and calculate the vector of that motion.
The vector of left ventricular blood flow ejection is generally from the point of maximal impulse in the left lateral chest between 4th and 6th intercostal spaces near the lateral clavicular line toward the medial cephalad direction. The system disclosed here can determine the vector and align the force of chest compression with the vector to assist ejection of blood and minimize interference with ventricular filling.
Utilizing an indicator of cardiac output, such as exhaled end-tidal carbon dioxide or vital organ oxymetry, the controller circuit could apply synchronized therapies during progressive shock and determine if they benefit the patient through increased blood flow.
A system is disclosed here to treat a patient having a heart and a chest, the system comprising: a least one sensor monitoring cardiac activity in the patient by detecting at least one of myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion; a logic controller receiving signals from the at least one sensor and generating control commands for controlling one or more phasic therapies and synchronizing the one or more phasic therapies with the monitored cardiac activity in the patient; and wherein the logic controller executes an algorithm stored in memory associated with the logic controller, wherein the algorithm causes the logic controller to generate commands to vary patterns of the application of the one or more phasic therapies, and thereafter detect changes in at least one of the sensed myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion due to variations in the patterns, and determine one of the patterns of phasic therapies corresponding to a desired level of at least one of sensed myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion hemodynamics and organ perfusion.
A method is disclosed here to treat a patient in shock comprising: sensing myocardial motion or pulsatile blood flow in the patient; repeatedly applying a phasic therapy to the patient synchronized to the sensed actual myocardial motion or pulsatile blood flow, wherein the phasic therapy includes repeatedly applying a compressive force to the chest or an electrical shock to the heart of the patient, and adjusting the compressive force or electrical shock depending on whether the force or shock coincides with a heart beat as indicated by sensed myocardial motion or pulsatile blood flow.
A system is disclosed here to treat a patient having a heart and a chest, the system comprising: a least one sensor monitoring cardiac activity in the patient by detecting at least one of myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion; a logic controller receiving signals from the at least one sensor and generating control commands for controlling one or more phasic therapies and synchronizing the one or more phasic therapies with the monitored cardiac activity in the patient; and wherein the logic controller executes an algorithm stored in memory associated with the logic controller, wherein the algorithm causes the logic controller to generate commands to vary patterns of one or more phasic therapies, and thereafter detect changes in at least one of the sensed parameters due to variation in the pattern of phasic therapies. The logic circuit would then determine which one of the patterns of phasic therapies corresponding to a desired level of at least one of sensed myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion hemodynamics and organ perfusion.
A method is disclosed here to treat a patient comprising: sensing a natural rate of myocardial activity of the heart of the patient, and repeatedly applying a phasic therapy to the patient synchronized to the sensed myocardial activity, wherein the phasic therapy includes repeated myocardial electrical stimulation applied at a rate faster than the sensed natural rate of myocardial activity. The method may further comprise a sensing system comparing at least one of the sensed myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion hemodynamics and organ perfusion with and without application of the phasic therapies to determine which of the phasic therapies optimally augments hemodynamics or perfusion.
A method is disclosed here to treat a patient having a heart and a chest, the system comprising: monitoring cardiac activity in the patient by detecting with at least one sensor at least one of myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion; receiving the signals from the at least one sensor and, based on the signals, synchronizing one or more phasic therapies applied to the patient to the monitored cardiac activity in the patient; varying the one or more phasic therapies; detect changes in at least one of the sensed myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion due to the variations in the one or more phasic therapies; determine one of the variations of the phasic therapies corresponding to a desired level of at least one of sensed myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion hemodynamics and organ perfusion.
The method may further comprise comparing at least one of the sensed myocardial pump activity, myocardial mechanical activity, hemodynamics and organ perfusion with and without application of the phasic therapies to determine which of the phasic therapies optimally augments hemodynamics or perfusion.
The invention relates to techniques and devices that may be used to increase cardiac output for patients suffering from a wide variety of ailments ranging from shock to pulseless electrical activity (PEA) where the patient appears to be lifeless yet has some residual mechanical heart activity. One exemplary technique of the invention is to sense when the heart is beating and then synchronize chest compressions, or other resuscitative techniques, with movement of the myocardial wall. In this way, various techniques may be used to optimally synchronize chest compressions (or other elements of CPR) with residual left ventricular function to improve the outcome of such patients. Hence, the invention may be used to synchronize the compression force of external devices, on or around the chest, with the ejection phase of the residual left ventricular function, and the relaxation phase with residual cardiac filling. In another aspect, the system and method disclosed herein provides various techniques and devices for sensing residual mechanical function, and then turn this information into a useful data stream that may be used to operate the various components of resuscitative technology, including adjuncts to blood flow, ventilation, and cardiac stimulating technology.
Such techniques may be used with patients suffering from a wide range of ailments. One exemplary use is for patients who are believed to be in cardiac arrest with pulseless electrical activity (PEA) and non-detectable blood pressures, but who still have residual left ventricular function to some degree. However, it will be appreciated that the invention is not intended to be limited to only such a use, but to a wide range of conditions where there is some organized electrical (but impaired) mechanical cardiac activity.
For example, at one end of such a spectrum is normal spontaneous circulation, where the cardiac outputs are normal and left ventricular mechanical and pumping function are normal. Below that level is hypotension then compensated shock. In such cases, the blood pressure and the patient's pulse are still palpable and there may be good cardiac output. However, for various reasons, the cardiac output is unable to meet the metabolic demands of the body and homeostasis is at risk. This is evident by parameters such as decreasing urine output and increasing serum lactate, which are markers of inadequate organ function.
Below compensated shock is the state of uncompensated shock. This is a state in which the myocardium and the cardiovascular system are no longer able to provide adequate amounts of blood flow, oxygen and nutrients to meet the needs of vital organs, and the function of those organs is affected to the extent that they are beginning to become damaged. Blood pressures in this state might be, for example, 70/30 mm Hg. systolic over diastolic. Also, urine output may cease, and the patient may become confused because of inadequate cerebral function. Importantly, as shock progresses, the pathways to multi-organ system failure are initiated.
Below classical uncompensated shock is what might be called “extreme shock” which borders on cardiac arrest. In this case, the patient exhibits some residual myocardial function including some left ventricular ejection, but cardiac output is wholly inadequate to meet the needs of vital organs. For example, cardiac output might be less than 1 liter per minute, blood pressure might be 50/20, urine output may be minimal or absent, and the patient may be stuporous or comatose. Further, the patient may appear to be near death with significantly impaired cerebral function and stupor bordering on coma. If untreated, extreme shock will result in true cardiac arrest in a timeframe of minutes. Generally, it is not possible to palpate arterial pulses manually in this range, and such patients may be classified as PEA by clinical personnel even thought their heart continues to beat.
Below the state of extreme shock is pulseless electrical activity (PEA) cardiac arrest, which importantly also has a spectrum of conditions and a range of hemodynamics. For example, at its upper end, PEA has both left ventricular mechanical function and cardiac output, but not sufficient enough to be detected as a peripheral radial or femoral pulse. If an intra-arterial catheter is placed into the patient, the blood pressure might be only 45/25, with blood pressure measurable only in major arteries of the chest, neck or groin. A Doppler probe placed over the neck or groin may detect forward blood flow. Blood flow is so profoundly inadequate that the patient will generally appear lifeless and their pupils may dilate and become fixed. Further, they appear to be in cardiac arrest despite the presence of residual pump function and forward flow. The high end of PEA dynamics overlaps the low end of “extreme shock.” In such cases, the clinical personnel may not be able to distinguish the differences. The electrocardiogram, while showing organized electrical activity, is variable in its pathology and may be relatively normal in its QRS configuration. The inventor has termed electromechanical dissociation (EMD) with residual myocardial mechanical activity “pseudo-EMD.”
Below the “high end” stage of PEA is electromechanical dissociation with almost absent left ventricular function. The blood pressure measured by intravascular catheters just above the aortic valve will show aortic pulsations but the blood pressures measured are on the order of 25/15 mm HG, and there will be almost no associated forward blood flow. Without application of CPR, oxygen delivered to the vital organs will be essentially absent and irreparable injury to organs such as the brain occurs within minutes. The electrocardiogram rarely has a normal appearing QRS configuration, and the overall pattern of the ECG may be slurred out and irregular.
The final stage of PEA is an organized electrical rhythm but no left ventricular mechanical function. This is true cardiac arrest. A catheter measuring pressures above the aortic valve will detect no pressure pulse and echocardiography will show no cardiac movement. Further, the cardiac output is 0 and the patient is in complete global ischemia and cardiac arrest. Without application of CPR, oxygen delivered to the vital organs will be 0, and irreparable injury to organs such as the brain occurs within minutes. The overall pattern of the ECG is invariably slurred out and irregular.
Along the spectrum described above, the invention may be used in all cases where there is some myocardial mechanical activity and synchronized resuscitative therapies may improve cardiac output. In such cases, the invention may be used to detect residual mechanical activity and to synchronize such activity of the heart with resuscitation techniques, such as those used in CPR (including chest compressions/decompressions and/or ventilation). Hence, the invention may be utilized in any pathophysiologic state from true cardiac arrest, to pseudo-EMD PEA, through the various stages of shock, or in any hemodynamics state in which residual myocardial mechanical function with and without cardiac output exists. By synchronizing chest compressions and/or decompressions, among other potentially cyclical therapies, both ejection and filling phases of the cardiac cycle may be augmented. In so doing, cardiac output and organ profusion may be increased, thereby improving the outcome of patients with impaired hemodynamics.
As one particularly important example, one clinical situation that often occurs and is challenging for physicians, is when patients progress from shock to apparent PEA cardiac arrest. In the earlier stages of this process, physicians tend to treat such patients with intravenous medications and possibly controlled ventilation. While drugs such as antibiotics may be administered to patients in states such as septic shock, pressor drugs such as dopamine continue to be a mainstay of treatment. Pressors, however, have generally not been shown to improve the outcome of such patients despite raising the blood pressure. This may be because they improve blood pressure but also raise vital organ oxygen utilization, such that the overall balance between oxygen supply and demand is not improved. Pressor drugs also have significant direct vital organ toxicity.
If, however, these parenteral therapies do not stabilize the patient, their shock may progress inexorably towards more and more extreme states and eventually become cardiac arrest. Many practitioners in emergency medicine and critical care continue to be unsure—and the medical literature remains unclear—as to which point a patient whose blood pressure is dropping should begin to receive chest compressions. Indeed, physicians generally do not apply techniques such as external chest compress before subjective loss of vital signs. This is because CPR, and in particular chest compressions, can interfere with cardiac function and in particular cardiac filling if applied in an unsynchronized manner. For instance, a patient whose blood pressure is 60/40 who begins to receive chest compressions out of synchronization with heart function could rapidly progress into full cardiac arrest. More specifically, in performing CPR without synchronization, application of the compression phase when the left ventricle is trying to fill may significantly decrease cardiac output on the next ejection secondary to the Frank Starling Law of the heart. Hence, by detecting myocardial mechanical function, chest compressions can be synchronized with the ejection phase so that patients in shock may be treated without exacerbating their condition and possibly moving them toward cardiac arrest.
Hence, the issue as to when chest compressions should begin when a patient is progressing through the stages of shock may be addressed by synchronizing chest compressions, and possibly other mechanical adjuncts, with the ejection and relaxation phases, so that the clinician may be more confident that chest compressions are assisting and not interfering with residual circulatory function. In this way, the clinician does not need to be as concerned with the question as to when to begin chest compressions. In this manner, the invention may act to allow use of external mechanical adjuncts in the treatment of any form of shock in a manner similar to the methods by which intra-aortic balloon counterpulsation has been applied in cardiogenic shock. The invention may thus allow application of such adjuncts in the pre-hospital, and Emergency Department environments.
Another advantage of using synchronization is that it may be performed as an adjunct to therapies directed at the cause of the shock, such as antibiotics or thrombolysis, enhancing vital organ perfusion while these therapies are being administered. Indeed, improved hemodynamics may not only stave off organ injury, it may improve the efficacy of parenteral therapies. Further, synchronized chest compressions are unlikely to have significant organ toxicity, unlike pressor drugs.
As described above, one particular application of the invention is in connection with those suffering from pulseless electrical activity (PEA). PEA is one of the three broad-types of cardiac arrest, the other two being ventricular fibrillation and asystole. PEA is also referred to as electromechanical disassociation (EMD). PEA has been described as “the presence of organized electrical activity on the electrocardiogram but without palpable pulses.” Rosen P, Baker F J, Barkin R M, Braen G R, Dailey R H, Levy R C. Emergency Medicine Concepts and Clinical Practice. 2nd ed. St Louis: C V Mosby, 1988. Unlike ventricular fibrillation, which can be specifically reversed with electrical countershock, PEA does not have a specific countermeasure. This may explain the traditionally worse outcome of patients in PEA compared to ventricular fibrillation. Unfortunately, the incidence of PEA is increasing, possible because early risk modification is changing the natural history of cardiovascular disease. It is now reported by some authorities that the majority of patients in cardiac arrest are in PEA at the time of EMS arrival. Additionally, a significant fraction of patients that are shocked out of ventricular fibrillation, or resuscitated from asystole, will experience PEA at some point during their resuscitation. The combination of these circumstances mean that a large majority of patients receiving advanced life support for treatment of cardiac arrest will have PEA at some time during resuscitation. Hence, now or in the near future PEA may supersede classical ventricular fibrillation in importance. It may already have done so.
Many patients with PEA have residual cardiac mechanical activity, and many have detectable blood pressures. This condition may be referred to as pseudo-EMD PEA. In such cases, the patient may appear lifeless and without a pulse. However, there often remains some degree of residual left ventricular function. Hence, one important feature of the invention is to sense when the patient still has some myocardial function and then to synchronize phasic resuscitation therapies, especially compression of the chest, with the heart's residual mechanical function. In this way, the compression phase of CPR may occur during the ejection phase, and the relaxation phase can allow elastic recoil of the chest—with associated decreases in intrathoracic pressure when the left ventricle is trying to fill. In this way, synchronizing phasic resuscitative therapies with residual ventricular ejection and filling, may improve hemodynamics, the rate of a return to spontaneous circulation (ROSC), and long term survival.
The invention may incorporate various non-invasive sensing technologies (represented by sensor 12 in
A wide variety of equipment and device may be used to provide chest compressions. For example, various types of automated compression systems may be use to compress the chest. These include systems, such as the AutoPulse Resuscitation System, by ZOLL Circulation, Inc. of Sunnyvale, Calif., the Thumper manufactured by Michigan Instruments or the LUCAS device, and the like. Further, the invention is not limited to automated compression systems, but may be used with manual techniques as well. For example, the invention may be used to provide an audio and/or visual signal to indicate to a rescuer as to when to manually apply chest compressions. Further, in some cases a suction device may be adhered to the chest so that the chest may be actively lifted intermittently with chest compressions.
Using either manual or automated equipment, the invention may be configured to synchronize external chest compressions with any residual mechanical activity of the myocardium such that when the myocardium enters pumping or systole phase, CPR is in the chest compression phase. Further, when the heart enters its refilling or diastole phase, chest compressions enter the relaxation phase. Sensory data may be passed through a logic circuit and outputs of that circuit used to control when during cardiac ejection or filling of synchronization occurs. These relationships may be varied over time to optimize the efficacy.
In addition to synchronizing chest compressions with residual heart function, the invention may also be use to synchronize ventilations with residual heart function. For example, inspiration and expiration may be synchronized with residual myocardial function so as to increase cardiac output. For instance, inspiration may be synchronized to systole and expiration with diastole. To apply ventilations, the invention may use a traditional ventilator or ventilations may be provided manually, such as by using a ventilatory bag. In the latter case, an audio and/or visual signal may be provided to the rescuer as to when to apply proper ventilations.
With both chest compressions and ventilations, the timing, frequency and/or duration may be varied depending on the particular treatment. For example, chest compressions may occur during the entire systole phase, or only during a portion of it. Further, chest compressions may occur every systole phase or during only certain systole phases. A similar scenario may occur with ventilations. The controller may use one or more sensory inputs, and a logic circuit utilizing and indicator or indicators of efficacy, to optimize the effect of synchronization on hemodynamics.
The system disclosed herein may be utilized with any therapy that may benefit from synchronization with residual myocardial mechanical function in apparently lifeless patients. Chest compression and decompression, abdominal counter-pulsation, ventilation, phasic limb-compression, myocardial electrical stimulation, intravascular fluid shifting, intravascular or intra-pericardial balloon inflation-deflation, application of transthoracic electromagnetic irradiation, among others. The controller logic circuit may vary the pattern of synchronization among multiple therapies so as to determine the optimal pattern with respect to increasing hemodynamics.
Myocardial electrical stimulation is, for example, external electrical shocks delivered through metal paddles or electrodes applied to the chest, or electrical signals applied directly to the heart from an internal pacemaker modified to synchronize myocardial electrical stimulations to, for example, myocardial wall function or detected pulsatile blood flow.
To sense myocardial wall function, a variety of noninvasive devices and technologies may be used. For example, one technology that may be used is electrocardiography (ECG). ECG may be an attractive detection method because it is already used in most clinical situations during resuscitation. However, because myocardial activity is not always present with ECG during PEA, it may need to be used in combination with other sensing techniques as described below. Another example of a sensing technique that may be used is Doppler ultrasonography (DOP). Doppler ultrasound uses the Doppler shift of ultrasonic wave to quantify the blood flow in peripheral vessels. This may be applied with a transducer on the neck for carotid flow, the groin for femoral flow, or a transthoracic or intraesophageal transducer for aortic flow. A Doppler probe may also be placed at the cardiac point of maximum impulse to detect movement of blood within the myocardium. An array of Doppler probes may be used to determine the vector of residual myocardial mechanical function and align chest compression and relation with that vector.
A dynamic pressure sensor detects pulsatile flow by sensing the oxygen content in a peripheral vein. The oxygen content sensed by the ROSS sensors as blood pulses through the peripheral vein. Similarly a pulse oximetry sensor may also be used to detect the oxygen content in a blood vessel in, for example, the toes, fingers or ear lobes. The oxygen content of the blood may be used to determine when to initiate and terminate CPR and mechanical or electrical cardiac stimulation. For example, if the ROSS sensor or pulse oximetry sensor detects pulsatile flow and an oxygen content above a threshold, the system may reduce the force of chest compressions or terminate chest compressions. Similarly, if the ROSS or oximetry sensor detects no pulsatile flow or an oxygen level falling below a threshold, the system may initiate manual chest compressor or electrical cardiac stimulation. The system may adjust various parameters of phasic therapies based on trends in the sensed oxygen status.
The data regarding pulses in peripheral blood vessels may be utilized to estimate residual myocardial mechanical function, such as the cardiac ejection phase, based on stored information regarding the delay between the myocardial mechanical function and pulse pressure or pulsatile flow in the peripheral blood vessel.
A further sensing technique that may be used is plethysmography (PLETH). Plethysmography may be applied by measuring changes in the transthoracic AC electrical impedance with heart motion. A further technique that may be used is phonocardiography (PHONO). Phonocardiography records the acoustical energy detected by a stethoscope over the heart. Still a further technique that may be used is echocardiography (ECHO). With echocardiography or ultrasound imaging of the heart, left ventricular ejection can be quantified. In some cases, echocardiograph detection of heart function may be combined with ECG. Also, sensitivity may be improved through the use of intravenously injected microbubbles or other ultrasound enhancing technologies.
It may be optimal to combine a number of these detection systems so as to increase the sensitivity and specificity of detecting residual myocardial mechanical function. Additionally, it may be optimal to incorporate a logic circuit which compares combinations of sensing technologies to an indicator of actual cardiac output, such as end-tidal carbon dioxide or aortic flow. In this manner, the invention could determine which combination of sensing technologies are most predictive of improvements derived from synchronization.
Additionally, the logic circuit of the invention might be capable of varying the synchronized therapeutics against indicators of actual cardiac output so as to determine which pattern of synchronized therapy is most effective. It may vary synchronization within one therapeutic device or multiple therapeutic devices so as to identify the optimal pattern.
Referring now to
The sensor 12 may be a sensor or sensors for a variety of sensing systems such as electrocardiography, Doppler ultrasonography, plethysmography, phonocardiography, echocardiography, transthoracic impedance and the like. The sensor 12 may be incorporated into a probe that is coupled to the chest, abdomen, back, extremities, or a combination of these, or placed within the body, such as within the esophagus, trachea, or stomach. These various types of sensors detect myocardial activity by detecting, for example, cardiac electrical activity; physical contractions and other movements of the heart, palpable pulses of arteries in, for example, the esophagus, trachea, or stomach; variations in the skin indicative of pulsating blood flow and the rhythm and chemical content of the breath.
The data collected by sensor 12 is transmitted to a controller 14 having signal processing and logic capabilities. A further description of controller 14 will be described hereinafter with reference to
As an alternative to applying automated chest compressions, the invention may also be used with manual techniques. In such cases, controller 14 may include a speaker 22 and/or a light 24 that provide information to a rescuer as to when to apply chest compressions and/or decompressions. For example, speaker 22 may be configured as a metronome to apply a repeating signal, or could give instructions in a human understandable voice. Light 24 may be configured to repeatedly flash to indicate when to apply chest compressions and/or decompressions. It will also be appreciated that a force transducer may be placed between the hands of the person providing manual chest compression and the patients such that the force, timing and vector of chest compression can be sensed so that the accuracy of synchronization is evaluated.
Chest compressions may be applied at a variety of locations. Examples include the sternal area, parasternal areas, circumferentially, the back, and the like. The abdomen may be compressed or counterpulsed broadly or with specific emphasis on the areas of the abdominal aorta or inferior vena cava. The extremities may be compressed rhythmically. The pattern of ventilation may be varied.
Controller 14 configured to receive data from sensor 12 and then process the signals in order to operate chest compression device 16, speaker 22 or light 24. More specifically, controller 14 is configured to synchronize external chest compressions and/or decompressions with any residual mechanical activity of the myocardium sensed by sensor 12. In this way, when the myocardium enters the pumping or systole phase, chest compression device 16 is configured to force interface member 18 against the chest to apply a chest compression. When the heart enters its refilling or diastole phase, controller 14 is configured to lift interface member 18 so that no compressive forces are being applied to the chest. It is understood that the therapeutic impulses may be restricted to a portion of each phase.
The vest 21 may include separately inflatable chambers 23 wherein each chamber is coupled via a conduit 25 to an air pump 27. Valves 29 in the conduits are actuated by the controller 14 to cause the inflation and deflation of the chamber associated with the valve. By selectively inflating and deflating the chambers, specific locations on the chest and back of the patient that are compressed can be to enhance the chest compressions and increase the cardiac output. As an alternative to a vest with chambers, the chest compression device may include a force interface member 18 that is segmented into a plurality of separately actuated pads 18a, 18b applied to compress the chest. Depending on which of the pads 18a, 18b are actuated to compress the chest and the sequence in with one or more of the pads are actuated the position on the chest of the chest compressions and the vector of the force applied by the chest compression can be varied to enhance, for example, cardiac output.
System 10 further includes a ventilation system 26 that is coupled to controller 14. For example, ventilation system 26 may comprise a ventilator that is in fluid communication with a mask 28. Controller 14 may be configured to synchronize inspiration and expiration to residual myocardial function as detected by sensor 12. For example, ventilation system 26 may be configured to provide positive pressure ventilations during systole and allow for expiration during diastole, or vice versa. Controller 14 may also be configured to coordinate operation of ventilation system 26 with chest compression device 16. As an option to using a mechanical ventilator, the invention may also utilize other techniques, such as a ventilatory bag that may be mechanically squeezed by the patient. In such cases, speaker 22 or light 24 may be actuated to indicate to the rescuer as to when to apply proper ventilations.
Referring now to
The controller may combine inputs from a number of sensing systems so as to increase the sensitivity and specificity of detecting residual myocardial mechanical function. Additionally, it may be optimal to incorporate a logic circuit, possibly within a microprocessor, which compares combinations of sensing technologies to an indicator of actual cardiac output, such as end-tidal carbon dioxide or aortic flow. In this manner, the invention could determine which combination of sensing technologies are most predictive of improvements derived from synchronization. Additionally, the logic circuit of the invention might be capable of varying the synchronized therapeutics and comparing the combinations to amount of residual myocardial synchronization and measured cardiac output so as to determine which pattern of synchronized therapy is most effective.
As previously mentioned, chest or abdominal compressions and/or ventilations may be applied during different times of the cardiac cycle and may be varied over the cycles themselves. The arterial blood pressure shown in
In summary, by utilizing a sensor, or combination of sensors, an apparently lifeless patient's residual myocardial wall function may be detected and the application of phasic resuscitative therapies, including chest compressions and/or decompressions as well as abdominal counterpulsations, and ventilations may be precisely controlled so that the application of CPR components enhances, and does not interfere with the existing mechanical activity of the heart. The device may also potentially be used in patients suffering severe shock with residual signs of life.
Referring to
As illustrated in step 50, the patient may periodically be provided with ventilations. The phases of ventilations may also be synchronized with the sensed ejection phases and refilling phases as measured in step 44. Further, the ventilations may be coordinated with the application of the chest compressions.
In some cases, the patient's chest may be actively lifted in an alternating manner with chest compressions as illustrated in step 52. In such cases, the chest may be lifted during the filling phase as measured in step 44.
As another optional step 54, the patient may periodically be provided with medications as part of the treatment. Examples of medications that may be applied include epinephrine, vasopressin, amiodarone, and the like. Alternative phasic therapies for use in step 46 may also be synchronized with residual myocardial activity. These other phasic therapies may include, among others: abdominal counterpulsation, ventilation, phasic limb-compression, myocardial electrical stimulation, intravascular fluid shifting, intravascular balloon inflation-deflation, application of transthoracic electromagnetic irradiation.
Throughout the procedure, the patient's heart may be continually monitored to determine myocardial activity as well as other physiological conditions. For example, after each of the treatments in steps 46, 48, 50, 52 and 54, the condition and response of the patient is monitored. Depending on the sensed response and condition, the selection and application of these treatments may be adjusted to achieve a desired response and condition of the patient. Depending on the patient's condition, any of the items discussed in steps 44-54 may be varied or stopped over time. At step 56, the process ends.
The controller analyzes 504 the output signals to determine which of the output signals or group of signals best indicate the condition of the patient, such as cardiac output. The algorithm 500 may compare the actual output signals to expected sensor output signals 506 stored in the memory of the controller. Based on the comparison, the controller identifies 508 the sensor(s) generating signal(s) that accurately and clearly report the condition of the patient in response to the regimen of chest compression(s).
The sensors identified in step 510 are deemed to be best suited to sense myocardial activity may depend on the particular patient and the circumstances of the PEA condition. The sensor validation procedure 500 may be performed at the initiation of chest compressions and periodically thereafter, especially if myocardial output does not improve in an expected manner.
Once the sensors have been validated, signals generated by the sensors identified in the validation process are used to provide feedback to the algorithms, such as shown in
The sensor signals from step 602 provide the controller and the health care provider with information from which to determine whether to initiate chest compressions. For example, if the ECG signals indicate a steady, regular heart beat, the controller may determine that chest compressions are not needed, in steps 604 and 606. The signals from step 602 may be analyzed in step 604 to determine whether, for example, the ECG signals do not indicate a regular or sufficiently frequent heart beat. If the ECG signals indicate an irregular or infrequent heart beat, the controller may determine (604) that chest compressions are needed (606) to augment the remaining natural cardiac activity.
In addition, signals from the sensors detecting pulsatile blood flow and actual myocardial motion may be compared to the signals of cardiac electrical activity to confirm that the cardiac electrical activity is synchronized with actual cardiac output. If the there is no detectable cardiac electrical activity or if the cardiac electrical activity is disassociated with pulsatile or actual ejection of blood from the heart, the controller may rely on sensors detecting actual myocardial movement or pulsatile blood flow to monitor cardiac movement and output. The controller may perform a sensor validation algorithm (
After chest compressions have been initiated (step 606), the controller executes algorithms (
In step 610, the controller executes algorithms (see
Examples of parameters of the chest compression that may be varied and optimized include: the depth of the compressions made into the chest; the during of each compression, the velocity of each compression, the force applied to the chest during each compression, the rate of the chest compression, the shape of the compression (such as the duration at the depth of the compression), the location of the compression on the chest, and the phase of synchronization between the chest compression and the sensed cardiac activity. While varying one or more of these parameters, the response of the patient to the chest compression is monitor and a determination is made at to which combination of parameter setting yields the most advantageous response, such as strongest arterial pulse flow.
In step 612, the controller synchronizes ventilation of the patient and electrical stimulation of the heart to the chest compressions or to contraction/ejection of the heart. The electrical stimulation may be repeated and coordinated with the chest compressions. In step 614, the sensors, e.g., validated sensors, detect or measure the response of the patient to one or more of the compressions, ventilation and electrical stimulations. In step 616, a determination is made that the detected or measured response is achieving a desired result or outcome in the patient. If a desired result or outcome is not being achieved, the controller may adjust the compressions, ventilation or electrical stimulations until the desired result or outcome is achieved.
The audible commands may include computer generated voice commands such as, during the chest compressions, “press softer”, “press harder”, “press deeper”, “press shallower”, “press faster (or slower)” and “press lower (or higher) on the abdomen”. Similarly, visible commands may be computer generated display images corresponding to these voice commands. The audible and visible commands may be the result of computer analysis of feedback signals generated from sensors monitoring the pulsatile flow, myocardial activity, breathing or other condition of the patient.
The force of the chest compressions to be applied is indicated by the length of the dotted line 706, 708 shown in
The chest compressions are synchronized with the heart beat. In this example, every third chest compression 810 coincides with the heart beat. It is desirable that the chest compression be synchronized with the ejection phase of the heartbeat. For example, the start of the chest compression should coincide with the QRS electrical signals 812 that precedes the ejection phase of the cardiac cycle.
A timer in the system that controls or triggers the chest compressions generates a timing signal 806 that triggers the start 814 and end 816 of each chest compression. The timing signal 806 triggers chest compressions at regular intervals, such as about every second. The regular intervals of the chest compressions may, over time, become unsynchronized with the heart beat 802.
To maintain synchronization between the chest compressions and the heart beat, a timer or counter generates an error signal 808. The timer is in the system that controls or triggers the chest compressions. The error signal is used to measure the period between the QRS signal 812 and the chest compression 810, e.g., the initiation of the chest compression, nearest the QRS signal. If the QRS signal 812 and the chest compression are synchronized, an error period 818 may be instantaneous or brief as shown by the error signal 808. A longer error signal 820 results if the initiation of the chest compression does not occur at about the same time as the QRS signal. As an alternative to the QRS signal, the error period 818, 820 may be determined based on sensed pulsatile flow or sensed mechanical myocardial activity.
The error period is triggered by the chest compression timing signal 806 and particularly by the signal 814 initiating the chest compression. The error period 820 may be a positive period if the chest compression starts 814 before the QRS period. A positive error period is applied by the system to delay the next chest compression by the duration of the period. The delay should cause the chest compression that coincides with the next heart beat to by synchronized with that heart beat. Similarly, the error period 820 may be a negative period if the QRS signal precedes the chest compression signal. A negative error period 820 may be applied to advance the occurrence of the next chest compression by the duration of the negative error period. The advance should cause the chest compression that coincides with the next heart beat to be synchronized with the heart beat.
The determination of the error period is similar to a phase lock loop control technique conventionally used in control systems. The delay or advance due to an error period 820 to the chest compression signal 814 may occur entirely in a period between two chest compressions, or may be distributed evenly between two or more periods depending on the length of the delay or advance. Similarly, the error period 820 may not result in a delay or advance if the period is shorter than a threshold duration, e.g., 10 milliseconds.
The chest compression is terminated 906 before a safety period 908 before the T wave. The period may be a short duration such as 10 to 200 milliseconds. The safety period 908 is applied to ensure that the chest compression does not continue to the T wave particularly during the portion 910 of the T wave during which the heart is vulnerable to commotio cordis, which is a disruption of the heart rhythm due to a blow to the heart during the T wave.
The electrical stimulation pulses 1008 may be applied at a frequency greater than the natural heat beat, such as in the manner shown and described in connection with
The electrical pulse signals 1008 may be applied to the chest of the patient or directly to the heart for each heart beat. The electrical signal is applied for each heart beat to assist the heart in restoring natural electrical stimulation, to resynchronize the natural electrical stimulation to the myocardial mechanical activity or to supplement the natural electrical stimulation to increase the ejection force from the myocardial mechanical activity.
The electrical pulse signals may be a “pacer pulse” such as that delivered by a conventional pacemaker and having a value of less than 500 milliamps (mA). Alternatively, the electrical pulse signals may shock the heart by delivering a pulse of between 500 mA to 5 A, which is similar to a low energy defibrillator pulse.
The application of the electrical pulse signals for each heart beat is in contrast to a conventional pacemaker device that does not issue an electrical stimulation for each heart beat. A conventional pacemaker issues an electrical stimulation one if and when a timer expires without the occurrence of a natural heart beat. A conventional pacemaker issues an electrical signal when a natural heartbeat does not occur in a prescribed period and does not issue an electrical signal that is synchronized with a naturally heartbeat.
The invention has now been described in detail for purposes of clarity and understanding. However, it will be appreciated that certain changes and modifications may be practiced within the scope of the appended claims.
This application is a divisional of U.S. patent application Ser. No. 13/182,800 (U.S. Pat. No. 9,259,543) filed Jul. 14, 2011, which is a continuation-in-part (CIP) of U.S. patent application Ser. No. 12/685,289, (U.S. Pat. No. 8,870,797) filed Jan. 11, 2010, which is a divisional application of U.S. patent application Ser. No. 10/973,775, (U.S. Pat. No. 7,645,247) filed Oct. 25, 2004, and claims the benefit of U.S. Provisional Patent Application No. 61/419,525 filed Dec. 3, 2010, the entirety of all of these applications are incorporated by reference.
Number | Name | Date | Kind |
---|---|---|---|
4198963 | Barkalow et al. | Apr 1980 | A |
4273114 | Barkalow et al. | Jun 1981 | A |
4928674 | Halperin et al. | May 1990 | A |
5020516 | Biondi et al. | Jun 1991 | A |
5490820 | Schock et al. | Feb 1996 | A |
5496257 | Kelly | Mar 1996 | A |
5496351 | Plicchi et al. | Mar 1996 | A |
6053869 | Kawagishi et al. | Apr 2000 | A |
6234985 | Lurie et al. | May 2001 | B1 |
6298267 | Rosborough et al. | Oct 2001 | B1 |
6351671 | Myklebust et al. | Feb 2002 | B1 |
6480733 | Turcott | Nov 2002 | B1 |
6620116 | Lewis | Sep 2003 | B2 |
6752771 | Rothman et al. | Jun 2004 | B2 |
7220235 | Geheb et al. | May 2007 | B2 |
7645247 | Paradis | Jan 2010 | B2 |
8870797 | Paradis | Oct 2014 | B2 |
20010027279 | Rothman et al. | Oct 2001 | A1 |
20040039313 | Sherman et al. | Feb 2004 | A1 |
20040230140 | Steen | Nov 2004 | A1 |
20050080363 | Jensen et al. | Apr 2005 | A1 |
20050256415 | Tan et al. | Nov 2005 | A1 |
20060100556 | Hargens et al. | May 2006 | A1 |
20060190045 | Marcus et al. | Aug 2006 | A1 |
20070060785 | Freeman et al. | Mar 2007 | A1 |
20080082135 | Arcot-Krislinamurthy et al. | Apr 2008 | A1 |
20080161894 | Ben-David et al. | Jul 2008 | A1 |
Entry |
---|
Astlion et al., “Effect of rescuer fatigue on performance of continuous external chest compressions over 3 min”, Resuscitation, Nov. 2002;55(2):151-5. |
Bayly et al., “Synchronous versus nonsyrichronous high-frequency jet ventilation: effects on cardiorespiratory vaiiables and airway pressures in postoperative patients”. Crit Care Med. Oct. 1987; 15(10):915-7. |
Bendielid et al., Correlation between measured inferior vena cava diameter and right atrial pressure depends on the echocardiographic method used in patients who are mechanically ventilated, J Am Soc Echocardiogr, Sep. 2002; 15(9):944-9. |
Bocka et al., Electromechanical Dissciation in Human Beings: An Echocardiographic Evaluation, Annals of Emergency Medicine, May 1988, page No. 450-452. |
Boese et al,, “Optomizlng temporal resolution in CT with retrospective ECG gating”, Radiologe, Feb. 2000;40(2): 123-9. |
Bottiger et al, Emergenc.y medicine—new concepts and therapies improve outcome from cardiac arrest, Anasthesiol Intensivnied Notfallmed Schmerztver. Jan. 2003; 38(1):63-7. |
Bradley et al.. “Blood flow: magnetic resonance imaging”, Radiology. Feb. 1985; 154(2)443-50. |
Brindley et al. “Predictors of survival following in-hospital adult carchopulmonary resuscitation”. Division of Critical Care Medicine, University of Alberta, Edmonton, no date. |
Brindley “Predictors of survival following in-hospital adult cardiopulmonary resuscitation” CMAJ •Aug. 20, 2002; 167 (4) (2002). |
Cobb et al, “Changing incidence of out-of-hospital •rentricular fibrillation, 1980-2000”. JAMA. Dec. 18, 2002(288(23):3008-13. |
Counselman et al., “The status of bedside ultrasongraphy training in emergency medicine residency programs”, Acad Emerg Med. Jan. 2003;10(1 ):37-42. |
Etiology, Electrophysiology, and Myocardial Mechanics of Pulseless Electrical Activity, Ture and Psedo-Pulseless Electrical Activity, p. 331-332 1989, 1992. |
Gebber GL, “Basis for phase relations between baroreceptor and sympathetic nervous discharge”, Am J Physiol, Feb. 1976:, 230(2): 263-70. |
Hayes et al., “Monitoring during cardiac arrest: are we there yet?”, Curr Opin Crit Care. Jun. 2003; 9(3): 211-7. |
International Search Report dated Mar. 5, 2012. |
Lanzer et al., “Cardiac imaging using gated magnetic resonance”, Radiology. Jan. 1984; 1501(1)0;121-7. |
Lanzer et al., “ECG-synchronized cardiac MR imaging: method and evaluation”, Radiology. Jun. 1985; 155(3):;681-6. |
Linder et at. “New mechanical methods for cardipulmonary resuscitation (CPR). Literature study and analysis of effectiveness”, Anaesthesist, Mar. 1997;46(3)220-30. |
Liu et al, “An efficient MR phosphorus spectroscopic localization technique for studying ischemic heart”, J Magn Resort imaging. Nov. 1999:, 10(5):892-8. |
Lorusso et al.. Hemodynamic effects In acute cardiomyoplasty of different wrapped muscle activation times as measured by pressure-vol. relation, J Card Surg, May-Jun. 1996;11 (3):217-25 |
Marek et at, “Correction of morphology and indicators of atrial function using long-term physiologic cardiac pacing in patients with prior chronic at dissociation or WI pacing”, Vnltr Lek. Sep. 1999 ;p. 45(9):513-7. |
Modersohn et al., “Diastolic heart function—pathophysiology, tharacterization, and therapeutic approaches”. Clin. Cardiol. Dec. 1993; 16(12):850-8. |
Morris-Thurgood et at, “Pacing in heart failure: improved ventricular interaction in diastole rather than systolic re-synchronization”, Europace, Oct. 2000;(4)2:271-5; duscussion 276. |
Ordonez et al., “Assessment of the effectiveness of descending aortomyop!asty for nonischemic cardiac failure by means of the subentiocaidial viability index”, Ann Thorac Cardiovasc Surg. Feb. 2001;7(1 ):17-22. |
Parish et al. “Aortic Pressure during Htiman Cardia Arrest: identification of Psedo-Electromechanical Dissociation”, Chest, Jan. 1992, pp. 123-128. |
Parish et al., “Success chzmges the probiem:why ventricular fibrillation is decling, why pulseless electricil activity is emerging, and what to do about it”, Resuscitation, Jul. 2003; 58(1 ): 31. |
Rudíger, A et al., “Frequency and outcome of in-hospital resuscitation outside the ICU-setting”, Swiss Med Wkly., Jan. 24, 2004: 134(3-4):59-62. |
Thornton et al., “Dynamic stiffness and implications for assisting the operation of the left ventricle”. IMA J Math Appl Med Bioi. Dec. 1996; 134 (3-4):275-95. |
Tortoli et al., Toward a better quantitative Measurement of Aortic Flow, Elsevier, Ultrasound in Med & Biol vol, 28, No. 2 pp. 249-257, 2002. |
Veenstra et al, “Electrotonic interactions between aggregates of chick embryo cardiac pacemaker cells” Am J Physio. Mar. 1968; 250(3 Pt 2):H453-63. |
Vinet et al, “Analysis of an iterative difference equation model of the cardiac cell membrane”, J Theor Bioi. Sep. 21, 1994; 170(2):201-14. |
Vural et al., “Optimization of synchronization delay in latissimus dorsi dynamic canhomyoplasty”, Ann Thorac Surg. May 1998;65(5)1231-4. |
Wenzel “Arginine vasopressin during cardiopulmonary resucitation: Laboratoiy evidence, clinical experience and recommendations, and a view to the future” Crit Care Med 2002. vol. 30. No. 4 (Suppl.) (2002). |
Wenzel, “Arginine vasopressin during cardiopulmonary resuscitation: laboratory evidence, clinical experience and recommendations, and a view to the future”, Department of Anesthesiologh and Critical Care Medicine, Leopold-Franzens-University, Innsbruck, Austria. |
Wenzel, Drug theraphy in cardiopulmonary resucitation, Wien K!in Wochenschr, Dec. 17, 2001;113(23-24):915-26. |
Number | Date | Country | |
---|---|---|---|
20160158098 A1 | Jun 2016 | US |
Number | Date | Country | |
---|---|---|---|
61419525 | Dec 2010 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 13182800 | Jul 2011 | US |
Child | 15040682 | US | |
Parent | 10973775 | Oct 2004 | US |
Child | 12685289 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 12685289 | Jan 2010 | US |
Child | 13182800 | US |