SEPARATION OF AN ORGANIC PHASE FROM A MIXTURE COMPRISING ORGANIC AND AQUEOUS PHASES BY SOLID PHASE SYSTEMS

Abstract
The present invention relates to a method for separation of a mixture comprising an organic and an aqueous phase making use of a solid phase system. Still further the invention refers to a specific solid phase system, namely a spin column which can be used in order to carry out said method in an easy way. According to preferred embodiments the organic phase is phenol or phenol/chloroform.
Description

BRIEF DESCRIPTION OF THE DRAWINGS


FIG. 1 shows a longitudinal section of a spin column, as it can be used in the present invention, inserted in a collection microtube (centrifugation tube).



FIG. 2 shows a longitudinal section of the spin column of FIG. 1 before and after centrifugation.



FIGS. 3-5 show longitudinal sections of the spin column of FIG. 1 filled with different solid phase systems according to Example 1 (see below).



FIG. 6 shows a longitudinal section of an alternative embodiment of a spin column.





The spin column of FIG. 1 comprises a reservoir 1 for the absorption/filtration medium and a filter 2, which is a barrier for that medium in order to keep it in the column during centrifugation, but which can be passed by the aqueous phase. The two elements are building the spin column, which is removable incorporated in a centrifugation tube 3. After centrifugation of an emulsion 4 of aqueous and organic phase using the spin column filled with a suitable adsorption medium 5 the separated aqueous phase 6 can be collected at the bottom of the centrifugation tube by removing the spin column (see FIG. 2).


In FIGS. 3-5 the reservoir 1 is filled with Sephadex (7) (FIG. 3), two layers of Sephadex (7) and PVPP (8), which are not mixed (FIG. 4), and a mixture 9 of Sephadex and PVPP (FIG. 5), respectively. Only in the case of the embodiment shown in FIG. 3, there were still two phases in the bottom of the centrifugation tube after centrifugation meaning that Sephadex is not suitable to adsorb the organic phase, while in the cases of the embodiments of FIGS. 4 and 5 there was only the aqueous phase in the bottom of the centrifugation tube (see also Example 1).


In the alternative embodiment shown in FIG. 6 the spin column, consisting of reservoir 1 and filter 2, is not inserted in but placed on the centrifugation tube.


Materials and Methods

The following examples do not limit the scope of the invention. The effectiveness of the method according to the present invention is clearly demonstrated:


Preparation of Spin Columns

A suspension of insoluble polymers in AE buffer (0.5 mM EDTA; 10 mM Tris; pH 9.0) is set up and swollen over night at 4° C.


For the preparation of the spin columns the suspension was homogenized by vortexing and an aliquot was pipetted into the column. The gel bed was formed by a short centrifugation. This pipetting/centrifugation procedure was repeated until the height of the bed made about ¾ of the volume of the inner reservoir of the column.


EXAMPLE 1
Phenol Binding Property of Poly(Vinylpolypyrrolidone)

Three spin columns were filled with different solid phase systems:

  • column A: Sephadex G-10; GE Healthcare; Cat No. 17-0010-01
  • column B: mixture of Sephadex G-10 and poly(vinylpolypyrrolidone); Sigma-Aldrich Co. Cat No. P6755
  • column C: two layers of solids: poly(vinylpolypyrrolidone) (below); Sephadex G-10 (above)


On each of the columns A, B and C 200 μL of a mixture of 100 μL water and 100 μL phenol/chloroform was placed. The columns were centrifuged for 30 seconds with 6000 g.


Only in the case of column A with no poly(vinylpolypyrrolidone) being present there was still a two-phase system left after the centrifugation step.


This result shows, that poly(vinylpolypyrrolidone) polymer is suited for adsorption of organic phase like phenol/chloroform and Sephadex G-10 does not lead to a separation.


EXAMPLE 2
Extraction of Plasmid DNA and Subsequent Linearization by Restriction
Extraction:

5 μL pUC21-solution (5 μg pUC21), 10 μL 1M sodium chloride solution and 35 μL TE buffer (1 mM EDTA; 10 mM Tris; pH 7.5) were mixed with 50 μL phenol and 50 μL phenol/chloroform. The mixture was homogenized by vortexing and added to a spin column filled with adsorption medium, which was prepared by mixing the following components:


4 g Sephadex G-25; GE Healthcare; Cat No. 17-0032-01


1 g poly(vinylpolypyrrolidone)


1 g poly(ethylene-co-acrylic acid) sodium salt acrylic acid 5 wt.-%; Sigma-Aldrich Co. Cat No. 426733


in AE buffer (0.5 mM EDTA; 10 mM Tris; pH 9.0)


The column was centrifuged for 30 seconds with 6000 g.


Restriction:

A mixture of 25 μL of the column flow through, 4 μL 10×reaction buffer, 10 μL BiDest and 1 μL enzyme (1. BamHI; 2. HindIII; 3. Sau3A) was heated for 60 minutes at 37° C. in a water bath.


The results showed that after separation of the plasmid DNA the same could be linearized by restriction endonucleases and that hence a potential rest amount of phenol has got no influence on the enzymatic reaction of the restriction. In addition, it was shown that the salt was efficiently removed during the separation by gel filtration.


EXAMPLE 3
Extraction of Linearized Plasmid DNA and Subsequent Ligation
Extraction:

50 μL linearized pUC21-solution (Example 2) were mixed with 50 μL phenol/chloroform. The mixture was homogenized by vortexing and added to a spin column filled with adsorption medium, which was prepared by mixing the following components:


6 g Sephadex G-10


2 g poly(vinylpolypyrrolidone)


in AE buffer


The column was centrifuged for 30 seconds with 6000 g.


Ligation:

A mixture of 30 μL of the column flow through, 4 μL 10×reaction buffer, 5 μL BiDest and 1 μL T4 DNA ligase was kept for 150 minutes at room temperature.


The result showed, that after the separation the linearized plasmide DNA could be modified by ligation and a potential rest amount of phenol has got no influence on the enzymatic reaction of a ligation.


EXAMPLE 4
Extraction of Genomic DNA and Subsequent PCR Amplification
Extraction:

100 μL purified human gDNA (160 ng/μL) and 10 μL 5 M sodium chloride solution were mixed with 100 μL phenol/chloroform. The mixture was homogenized by vortexing and added to a spin column filled with adsorption medium, which was prepared by mixing the following components:


6 g Sephadex G-10


2 g poly(vinylpolypyrrolidone)


in AE buffer


The column was centrifuged for 30 seconds with 6000 g.


Amplification:

After the separation the gDNA could be amplified by PCR. This shows that a potential rest amount of phenol has got no influence on the enzymatic reaction of a PCR.


EXAMPLE 5
Comparison of the Duration of Different Separation Methods

Classical phenol extraction: 100 μL sample+100 μL phenol/chloroform; vortexing (10 sec); centrifugation (30 sec); transferring supernatant in fresh tube; adding salt and ethanol; vortexing and centrifugation (10 min); discarding supernatant; adding 70% ethanol; centrifugation (3 min); discarding supernatant; drying pellet (10 min); resuspending in buffer.


Total duration: 25 minutes


Separation with Phase Lock Gel™: prespinning Phase Lock Gel™ tube (20-30 sec; 12000-16000×g); 100 μL sample+100 μL phenol/chloroform; mixing (vortexing not recommended) (10 sec); applying all to the column; centrifugation (5 min; 12000-16000×g); transferring (decanting) supernatant in fresh tube.


Total duration: 6 minutes


Separation according to the invention: 100 μL sample+100 μL phenol/chloroform; vortexing (10 sec); applying all to the column; centrifugation (30 sec; 6000×g).


Total duration: 1 minute


The duration of the separation process is clearly diminished when the method according to the present invention is used.


REFERENCE NUMBERS




  • 1 reservoir


  • 2 filter


  • 3 collection microtube (centrifugation tube)


  • 4 emulsion of aqueous and organic phase


  • 5 adsorption medium


  • 6 aqueous phase


  • 7 Sephadex


  • 8 PVPP


  • 9 mixture of Sephadex and PVPP


Claims
  • 1. A method for the separation of a mixture of an organic and an aqueous phase, wherein the organic phase is adsorbed by a solid phase system.
  • 2. The method according to claim 1, wherein the organic phase comprises phenol or phenol/chloroform.
  • 3. The method according to claim 1, wherein the solid phase system comprises a water-insoluble polymer.
  • 4. The method according to claim 3, wherein the water-insoluble polymer is capable of adsorbing organic solvent.
  • 5. The method according to claim 4, wherein the organic solvent is selected from phenol or phenol/chloroform.
  • 6. The method according to claim 1, wherein the solid phase system comprises one ore more polymers out of the group consisting of poly(vinylpolypyrrolidone), poly(ethylene-co-acrylic acid) sodium salt acrylic acid 5 wt.-%, poly(vinylidene fluoride), poly(tetrafluorothylene), poly(4-vinylphenol) and poly(styrene-co-maleic anhydride).
  • 7. The method according to claim 1, wherein the solid phase system comprises additionally a gel filtration medium.
  • 8. The method according to claim 7, wherein the gel filtration medium is capable of adsorbing salts from the aqueous phase.
  • 9. The method according to claim 1, wherein the solid phase system is incorporated into a spin column.
  • 10. The method according to claim 9, wherein a mixture of an organic and an aqueous phase is added and the separation step is carried out by centrifugation.
  • 11. A spin column comprising a water-insoluble polymer capable of adsorbing organic solvent selected from phenol or phenol/chloroform.
  • 12. The spin column according to claim 11, wherein the water-insoluble polymer is one or more selected out of the group consisting of poly(vinylpolypyrrolidone), poly(ethylene-co-acrylic acid) sodium salt acrylic acid 5 wt.-%, poly(vinylidene fluoride), poly(tetrafluorothylene), poly(4-vinylphenol) and poly(styrene-co-maleic anhydride).
  • 13. The spin column according to claim 11 comprising additionally a gel filtration medium.