The present invention is concerned with the separation of metal ions, in particular, radiometal ions, from other metal ions in aqueous solution. In particular, the present invention relates to methods of continuous separation of radiometal ions from target metal ions from which the radiometal ions have been generated, especially to such methods that may be used in the generation of radiometals and radiopharmaceuticals for medical and veterinary use, such as in positron emission tomography (PET). In addition, the present invention relates to continuous methods of production of radiometals, optionally including recycling of the separated target metal, and methods of production of radiolabeled compounds. Apparatus for carrying out such a separation is also provided, along with the use of such apparatus in separation of metal ions.
Over the past several decades, positron emission tomography (PET) has become one of the best available diagnostic options for cancer1. PET radiopharmaceuticals based on radiometals are gaining in popularity due to their ability to probe biological processes occurring on timescales from hours to days2.
Radiometals such as 68Ga, 89Zr, 64Cu, and 45Ti are finding increased use in peptide and antibody-based PET radiopharmaceuticals due to their widely ranging half-lives, which allow for matching with the circulation time of the biological vector of interest, high radiolabelling yields, little or no post-labelling purification requirement, and the possibility of carrying out late-stage radiolabelling2,3,4.
68Ga is experiencing a particularly high adoption rate in clinics5. For example, 68Ga-PSMA (prostate-specific membrane antigen) is emerging as the gold standard for prostate cancer diagnostics6, and other tracers are in development7. The synthesis of Ga-PSMA has been described8.
The convenient chelation chemistry and ready availability via the 68Ge generator contribute to this popularity9. However, this has also caused the so-called “Ga rush”: all medical 68Ga is supplied by gallium generators, which are based on 68Ge isotopes produced by large particle accelerators10,11. These generators can suffer from high prices, quality inconsistencies12, limited shelf life and inherently low yield9 of 68GaCl32. At medical conferences, the inventors have heard clinicians comment that they are not able to source enough 68Ga from the generators to support larger clinical trials. Accordingly, alternative sources of 68Ga are needed.
An alternative means of production of 68Ga is the irradiation of the stable isotope 68Zn using a cyclotron13. As many hospitals have their own cyclotrons, this is potentially a convenient means of production of 68Ga for radiotracers and the like. However, the cyclotron production of 68Ga from 68Zn and its separation requires a series of manual operations, entailing significant radiation exposure to the personnel carrying out those operations, and the process is not easily amenable to automation.
The production of 68Ga from a zinc salt solution target has recently been described, in which zinc chloride14 or zinc nitrate15 is used as the solution target in a cyclotron. However, the irradiated solution of 68Zn and 68Ga resulting from this step still requires a semi-manual separation on two solid-phase cartridges16. Although the procedure is capable of recovering the expensive 68Zn target material for re-use it is laborious and slow. Further, the eluted 68Ga needs to be re-formulated before it can be used in radiolabelling. Recently73, a cassette style apparatus for conducting ion exchange chromatographic separation of 68Zn and 68Ga has been described, in which 68Zn can be recovered in an acetone solution, and, it is said, can be re-used. However, it is necessary for 68Zn target solutions to be rigorously organic-free in order that they can be used as solution targets. Accordingly, there are improvements to be made in the production of 68Ga for medical and radiolabelling purposes.
45Ti shows promise as a PET radiometal due to its 85.7% positron branch, negligible secondary radiation, and facile production. Its three-hour half-life compares favourably with that of 68Ga (68 min) and can allow for longer transport distances. Furthermore, the sharper PET images of 45Ti due to its lower p endpoint energy (1.04 MeV for 45Ti versus 1.90 MeV for 68Ga) can be especially advantageous for small-animal PET. A number of small molecule 45Ti compounds have been synthesised and used for PET imaging and radiotracing.
The bombardment of naturally monoisotopic scandium with low energy protons from a medical cyclotron via the natSc(p,n)45Ti nuclear reaction is an attractive 45Ti production route17-19. Recovery of the radiometal is the first post-production step, which, for a highly hydrolysable metal such as Ti, becomes critical. Currently, the solid phase extraction from acidic solutions on to a cation or anion exchange resin is the predominant way to separate the 45Ti from its Sc matrix. The present inventors and others have previously used the PEG-functionalised diol20, cation exchange21-23 and hydroxamate24 resins. However, as for the separation of 68Ga, this method is capable of improvement for safety and efficiency, especially regarding re-use of the Sc target material.
Even longer transportation and post-injection imaging times are possible with 89Zr. The 89Zr radioisotope decays with a half-life of 3.27 days via electron capture (77%), and positron emission (23%) to 89Y.25 Since residence time of monoclonal antibodies (mAbs) in humans ranges from a few days to weeks, 89Zr appears to be an ideal radionuclide for use in immuno-PET. Conjugated via desferrioxamine (DFO)-derived bifunctional chelators, 89Zr-labelled Cetuximab, Trastuzumab, and J591 have been prepared and investigated pre-clinically and clinically.26 The synthesis of Zr-trastuzumab has been described27.
The proton bombardment of a thick target prepared from naturally monoisotopic 89Y (natY) at optimum energy of 14 MeV yields up to 58 MBq/μAh of the 89Zr radionuclide. The separation of zirconium from bulk yttrium typically involves the adsorption of the radionuclide onto a hydroxamate resin followed by elution with oxalic acid.28
89Zr obtained from a solution target of Y(NO3)3 has been described, along with column based separation methods to isolate the 89Zr as the hydrogen phosphate, instead of the conventional oxalate, avoidance of which is beneficial for reasons of its toxicity16.
64Cu is the most commonly used Cu radioisotope. It has a half life of 12.7 h, and so is well suited to PET studies conducted over a 48 h period. This half life also allows for transport distances longer than for 68Ga. 64Cu decays 17.4% by positron emission, and has a β+ maximum energy of 0.66 MeV with average energy of 0.28 MeV, allowing for very high quality PET images. In addition, 64Cu also decays by electron capture (43%) and by β− (43%), and so can be used as both a therapeutic and a diagnostic radionuclide. Radiopharmaceuticals based on Cu-thiosemicarbazones have been developed to measure blood flow, for example Cu-pyruvaldehyde-bis(N4-methylthiosemicarbazone) (Cu-PTSM), and, more recently, in the imaging of hypoxic tissues, for example Cu-diacetyl-bis(N4-methylthiosemicarbazone) (Cu-ASTM).75
Extraction of metal ions from an aqueous solution and separation of one metal ion from another both present in aqueous solution has been extensively studied. There are many known methods of carrying out liquid-liquid extraction to carry out such separations. For example:
Other methods of separation of various metal ions from one another are disclosed, which do not concern the metal ions of interest in the present invention37-45.
All of these methods are batch separations requiring manual input from personnel, and so are capable of improvement from the point of view of safety. The authors of the above separation procedures, except for Dejesus34, do not address the efficiency of the overall process. None of the above procedures consider the reuse of target material where radiometal ions are used.
The use of an automated separation system for the mixed liquid-liquid extraction/resin based separation of 99Tc from Mo ion comprised in aqueous solution has been described46, in which the use of the column in the separation is automated. This does not of course address the need for manual handling of radiometal solutions in the liquid-liquid extraction batch processes.
One of the important stages of liquid-liquid extraction is the phase separation stage, ie the stage at which the organic extractant and the aqueous phases are separated from one another following their mixing to allow partition of the solutes of the aqueous phase between the aqueous phase and the organic extractant. Traditionally, this has been carried out using such apparatus as a separatory funnel, in which a more dense phase and a less dense phase separate into individual layers and are allowed to flow out of the separatory funnel in turn. Recently, however, techniques for carrying out this stage using a microfiltration membrane have been described47,48. In these procedures, the basis of the phase separation is not density, as in the traditional methods, but interfacial tension between the phases. If the combination of the interfacial tension between the organic and aqueous phases and the membrane is appropriately selected, one of the phases will wet the membrane and the other will not, allowing one phase selectively to pass through the membrane. In this way, the membrane acts as a selective barrier between the phases. Pressure is applied to drive the liquid through the membrane, though again this must be carefully selected with reference to the interfacial tension of the phases and the membrane used in order that the phase separation is completely selective. This method of membrane based phase separation is distinct from extraction methods in which the membrane itself comprises the extractant phase, such as fibre-supported liquid extraction49, bulk liquid membranes50, emulsion liquid membranes, supported liquid membranes, polymer inclusion membranes and the like51.
It is an aim of the present invention to provide a method for the efficient on demand production of radioisotopes, particularly for medical and veterinary use.
It is an aim of the present invention to provide a means for the efficient separation of a radiometal ion from a target metal ion that minimises manual handling of the radioisotope and target material, to improve safety.
It is an aim of the present invention to provide a method for the production of radioisotopes that minimises manual handling of the radioisotopes to improve safety.
It is an aim of the present invention to provide a method of on demand production of radioisotopes, particularly for medical and veterinary use, in which the use of 68Ge isotopes produced by large particle accelerators is avoided.
It is an aim of the present invention to provide a method of on demand production of radioisotopes, particularly for medical and veterinary use, in which the entire process from generation of the radioisotope to obtaining a pure solution of the radioisotope suitable for synthesis of a radiotracer or radiopharmaceutical can be carried out as a continuous process with minimal manual intervention.
It is an aim of the present invention to provide a method of on demand production of radioisotopes, particularly for medical and veterinary use, in which the target material from which the radioisotope is generated can be reused to generate further radioisotope therefrom.
It is an aim of the present invention to provide a method of on demand production of radioisotopes, particularly for medical and veterinary use, in which a precise dose of a radiotracer or radiopharmaceutical can be supplied for a given patient, subject or purpose.
It is an aim of the present invention to provide apparatus suitable to carry out a method addressing one or more of the above aims.
Accordingly, in a first aspect, the present invention provides a method of separation of a radiometal ion from a target metal ion, comprising a first liquid-liquid extraction step in which an organic phase comprising an extractant and an interfacial tension modifier is mixed with an aqueous phase comprising the radiometal ion and the target metal ion in order that the radiometal ion is at least partially transferred to the organic phase, followed by a first phase separation step, wherein the phase separation is carried out in flow comprising the use of a microfiltration membrane to separate the phases based on the interfacial tension between the phases such that a permeate phase passes through the membrane and a retentate phase does not, wherein:
Preferably, in the method, a pressure ΔPmem is exerted across the microfiltration membrane by a pressure controller. Preferably, the pressure exerted across the microfiltration membrane, ΔPmem, is controlled to be less than the capillary pressure Pcap associated with the fluid passageways of the microfiltration membrane and the mixture of the aqueous phase and the organic phase, and is controlled to be greater than the pressure Pper required to cause the permeate phase to pass through the microfiltration membrane.
Preferably, the microfiltration membrane is hydrophobic, and the permeate phase is the organic phase. Suitably, however, a hydrophilic microfiltration membrane may be used, in which case the permeate phase will be the aqueous phase.
Preferably, the first liquid-liquid extraction step is conducted in flow. Preferably, the first liquid-liquid extraction step comprises mixing the aqueous phase and the organic phase such that stable liquid-liquid segmented flow of the mixture is established.
Preferably, the aqueous phase is an aqueous solution of a protic acid. Suitably, the aqueous phase comprises a concentration of aqueous hydrochloric acid or nitric acid of greater than or equal to 1 M, such as greater than or equal to 3M, preferably greater than or equal to 6M, and, in some embodiments, most preferably comprises a concentration of 12M aqueous hydrochloric acid or nitric acid. Suitably, the aqueous phase has a pH of less than or equal to 1. Suitably, the aqueous phase is an aqueous solution of nitric acid. Preferably, the aqueous phase is an aqueous solution of hydrochloric acid.
Preferably, the radiometal ion and the target metal ion are defined as follows:
In some embodiments, the radiometal ion is a Ti ion and the target metal ion is a Sc ion. In those embodiments, preferably the aqueous phase is a solution in 12M HCl. Preferably, the extractant is a solvent having the ability to function as a bidentate ligand for Ti via two oxygen atoms, preferably thus forming a five membered ring, as well as having a suitable interfacial tension with 12 M (37%) HCl. Suitable extractants may be maltol, vanillin, eugenol, and guaiacol (o-methoxyphenol). Suitable interfacial tension modifiers are solvents having similar properties to the extractant, but that are not able to function as a bidentate ligand for the Ti ion, such that it does not interfere with the ability of the extractant to interact with the Ti ions, such as fluorobenzene, trifluorotoluene, thiophene and anisole. Preferably, the extractant is guaiacol and the interfacial tension modifier is anisole. More preferably, the anisole is present in an amount of at least 10% v/v. Preferably, the flow ratio of the aqueous phase to the organic phase is 1 to greater than or equal to 3. Preferably, the microfiltration membrane is a PTFE membrane. Preferably, a pressure controller is present in the form of a PFA diaphragm. Most preferably, the microfiltration membrane is a PTFE membrane having a pore size of 0.2 μm, the PFA diaphragm has a thickness of 0.002″ (0.0508 mm). Suitably, the combined flow rate of the organic phase and aqueous phase may be selected in the range of 0.01 mL/min to 12 mL/min, such as 0.1 mL/min to 10 mL/min, or 0.2 mL/min to 8 mL/min, or 0.2 mL/min to 5 mL/min, or 0.2 mL/min to 2 mL/min, such as 0.5 mL/min or 1.00 mL/min.
In some embodiments, the radiometal ion is a Ga ion and the target metal ion is a Zn ion. In those embodiments, preferably the extractant is selected from the group consisting of diethylether, butylmethyl ether, diisopropyl ether, tetrahydropyran, methyl hexyl ether, dibutyl ether and diamyl ether. In some embodiments, the extractant is selected from the group consisting of diethylether, butylmethyl ether, tetrahydropyran, methyl hexyl ether, and dibutyl ether. In some embodiments, the extractant is selected from the group consisting of butylmethyl ether, tetrahydropyran, methyl hexyl ether, and dibutyl ether. More preferably the extractant is selected from butyl methyl ether, diisopropyl ether, dibutyl ether and diethyl ether, yet more preferably the extractant is selected from butyl methyl ether and diisopropyl ether, and most preferably the extractant is diisopropyl ether. In some embodiments, the extractant is not diisopropyl ether, and/or is not diethyl ether. Preferably, the interfacial tension modifier is selected from the group consisting of: a fluorinated aromatic hydrocarbon; an aromatic hydrocarbon; an alkoxybenzene; a halogenated alkane, for example selected from the group consisting of 1,2-dichloroethane, 1,1,2-trichloroethane, 1,1,1-trichloroethane, hexachloroethane and bromoethane; and an alkane; more preferably, the interfacial tension modifier is selected from the group consisting of toluene, anisole, 1,2-dichloroethane, trifluorotoluene and heptane. Yet more preferably, the interfacial tension modifier is selected from the group consisting of toluene and trifluorotoluene, and most preferably the interfacial tension modifier is trifluorotoluene. Preferably, the ratio of the extractant to the interfacial tension modifier is 1:2 by volume. Preferably, the aqueous phase is a solution in 6M HCl. Where this is so, the extractant is preferably selected from diethyl ether, diisopropyl ether, dibutyl ether, butyl methyl ether and hexyl methyl ether, more preferably from diethyl ether, diisopropyl ether and hexyl methyl ether. Equally preferably, the aqueous phase is a solution in 3M HCl. Where this is so, the extractant is preferably selected from diethyl ether and diisopropyl ether, more preferably diisopropyl ether. The conditions under which the separation is carried out may include a concentration of zinc salt, such as ZnCl2, of more than 5 m, such as 7 m, where m indicates molality (moles of solute per kg solvent).
Where the radiometal ion is a Ga ion and the target metal ion is a Zn ion, it is useful for the method to further comprise a back extraction procedure, in order that the Ga ion can be brought back into the aqueous phase for use in radiolabeling reactions. This back extraction procedure comprises, following the first phase separation step, a first back-extraction step in which an organic phase comprising the radiometal ion is mixed with an aqueous solution of a protic acid in order that the radiometal ion is at least partially transferred to the aqueous solution, followed by a back-extraction phase separation step, in which the phase separation is carried out in flow comprising the use of a microfiltration membrane to separate the phases based on the interfacial tension between the phases such that a permeate phase passes through the membrane and a retentate phase does not, in order to obtain an aqueous solution comprising the radiometal ion. Preferably, the aqueous solution of a protic acid is an aqueous solution of less than 6 M HCl, such as less than 3 M HCl, more preferably 0.001 to 1 M HCl, most preferably 0.1 M HCl.
Where the radiometal ion is a Ga ion and the target metal ion is a Zn ion, it is useful for the method to further comprise a scrubbing procedure, in order to reduce the quantity of Zn in the organic phase and thus improve the purity with which the Ga is obtained. This scrubbing procedure comprises, following the first phase separation step, a first scrubbing step in which an organic phase comprising the radiometal ion and the target metal ion is mixed with an aqueous solution of a protic acid in order that the target metal ion is at least partially transferred to the aqueous solution, followed by a scrubbing phase separation step, in which the phase separation is carried out in flow comprising the use of a microfiltration membrane to separate the phases based on the interfacial tension between the phases such that a permeate phase passes through the membrane and a retentate phase does not, in order to obtain an aqueous solution comprising the target metal ion, and an organic phase comprising the radiometal ion and a decreased quantity of the target metal ion. Preferably, the aqueous solution of a protic acid is an aqueous solution of at least 8 M HCl.
In order to obtain a high purity of Ga, it is preferable that the method further comprises, following the first liquid-liquid extraction step and the first phase separation step, and in this order: the scrubbing procedure described above; and then a first back extraction procedure as described above. Yet more preferably, the method can further comprise, following the first back extraction procedure: a second liquid-liquid extraction step and a second phase separation step as described above; and then a second back extraction procedure as described above. Preferably, the aqueous solution comprising the radiometal ion obtained from the first back extraction procedure is acidified prior to its introduction into the second liquid-liquid extraction step as the aqueous phase, preferably to a 6N acid concentration.
Preferably, the microfiltration membrane is selected from a PTFE membrane with PP support and a PTFE membrane. Preferably, a pressure controller is present in the form of a PFA diaphragm. Most preferably, the microfiltration membrane is selected from a PTFE membrane with PP support and a PTFE membrane, and has a pore size of 0.2 μm, the PFA diaphragm has a thickness of 0.002″ (0.0508 mm). Suitably, the combined flow rate of the organic phase and aqueous phase may be selected in the range of 0.01 mL/min to 12 mL/min, such as 0.1 mL/min to 10 mL/min, or 0.2 mL/min to 8 mL/min, or 0.2 mL/min to 5 mL/min, or 0.2 mL/min to 2 mL/min, such as 0.5 mL/min or 1.00 mL/min.
In some embodiments, the radiometal ion is a Zr ion and the target metal ion is a Y ion. Preferably, the extractant is a solvent having the ability to function as a bidentate ligand for Zr via two oxygen atoms, preferably thus forming a five membered ring, as well as having a suitable interfacial tension with 12 M (37%) HCl. Suitable extractants may be maltol, vanillin, eugenol, and guaiacol (o-methoxyphenol). Suitable interfacial tension modifiers are solvents having similar properties to the extractant, but that are not able to function as a bidentate ligand for the Zr ion, such that it does not interfere with the ability of the extractant to interact with the Zr ions, such as fluorobenzene, trifluorotoluene, thiophene and anisole. Preferably, the extractant is guaiacol (o-methoxyphenol), and the interfacial tension modifier is anisole. Preferably, the anisole is present in an amount of at least 10% v/v. Preferably, the aqueous phase is a solution in 12 M HCl. Preferably, the flow ratio of the aqueous phase to the organic phase is 1 to greater than or equal to 3, and more preferably is 1:5. Alternatively, where the radiometal ion is a Zr ion and the target metal ion is a Y ion, it is preferable that the extractant is 0.1 M trioctylphosphine oxide (TOPO), the interfacial tension modifier is hexane, and the aqueous phase is a solution in 6 M HCl.
Preferably, the microfiltration membrane is a PTFE membrane. Preferably, a pressure controller is present in the form of a PFA diaphragm. Most preferably, the microfiltration membrane is a PTFE membrane having a pore size of 0.2 μm, the PFA diaphragm has a thickness of 0.002″ (0.0508 mm). Suitably, the combined flow rate of the organic phase and aqueous phase may be selected in the range of 0.01 mL/min to 12 mL/min, such as 0.1 mL/min to 10 mL/min, or 0.2 mL/min to 8 mL/min, or 0.2 mL/min to 5 mL/min, or 0.2 mL/min to 2 mL/min, such as 0.5 mL/min or 1.00 mL/min.
In some embodiments, the radiometal ion is a Cu ion and the target metal ion is a Ni ion. Preferably, the extractant is a species having the ability to act as a monodentate or bidentate ligand for the Cu ion, as well as having a suitable interfacial tension with 6 M HCl.
Suitable extractants may be: one or more trialkyl phosphine oxides; one or more alkylphosphoric acid monoalkyl esters; one or more diketones having the structure R3—C(═O)CH2C(═O)—R4, in which R3 and R4 are each independently an alkyl or an aryl group; and one or more aldoximes or ketoximes in which the substituent(s) of the oxime group are aromatic groups.
Suitable interfacial tension modifiers are solvents, such as branched or unbranched cyclic or acyclic aliphatic hydrocarbons having from five to sixteen carbon atoms, or aromatic hydrocarbons. Preferably, the interfacial tension modifier is selected from n-pentane, n-hexane, n-heptane, n-octane, n-nonane, n-decane, n-undecane, i-hexane, neo-hexane, i-heptane, neo-heptane, cyclohexane, cycloheptane, cyclooctane, kerosene, light petroleum, benzene, naphthalene, toluene, ethylbenzene, dimethylbenzene, iso-octane and mixtures thereof, more preferably selected from toluene, hexane, heptane and mixtures thereof.
Preferably, the extractant is selected from:
More preferably, the extractant is trioctyl phosphine oxide.
Preferably, the extractant, preferably trioctylphosphine oxide, is present in a concentration of at least 0.1 M in the interfacial tension modifier. Preferably, the extractant, preferably trioctyl phosphine oxide, is present in a concentration of from 0.1 M to 0.4 M in the interfacial tension modifier. Preferably, the extractant is 0.4 M trioctyl phosphine oxide, and the interfacial tension modifier is toluene; alternatively, the extractant is 0.1 M trioctylphosphine oxide and the interfacial tension modifier is hexane or heptane. Preferably, the aqueous phase is a solution in 6 M HCl.
Preferably, the flow ratio of the aqueous phase to the organic phase is 1 to greater than or equal to 1, more preferably, 1 to greater than or equal to 3, and most preferably in the range of from 1 to greater than or equal to 3 to 1 to less than or equal to 5.
Preferably, the microfiltration membrane is a PTFE membrane. Preferably, a pressure controller is present in the form of a PFA diaphragm. Most preferably, the microfiltration membrane is a PTFE membrane having a pore size of 0.2 μm, the PFA diaphragm has a thickness of 0.002″ (0.0508 mm). Suitably, the combined flow rate of the organic phase and aqueous phase may be selected in the range of 0.01 mL/min to 12 mL/min, such as 0.1 mL/min to 10 mL/min, or 0.2 mL/min to 8 mL/min, or 0.2 mL/min to 5 mL/min, or 0.2 mL/min to 2 mL/min, such as 0.5 mL/min or 1.00 mL/min.
In a second aspect, the present invention provides a method of generation of radiometal ions from a target metal, comprising:
Preferably, the method further comprises, following step c, recycling the aqueous solution of the target metal ions for use in a subsequent irradiation step b. Suitably, the recycling of the aqueous solution of the target metal ions comprises the step of treating the aqueous solution of the target metal ions to remove any organic solvents from the solution. Preferably, the treatment step comprises passage of the aqueous solution of the target metal ions through a reverse phase chromatography column having a stationary phase suitable for adsorption of any trace organic solvents. Preferably, the reverse phase chromatography column is a C18 column, ie an octadecyl carbon chain bonded silica stationary phase column.
Suitably, the irradiation is conducted using a cyclotron. Suitably, when irradiating Zn to produce 68Ga, or Sc to produce 45Ti, the irradiation may comprise bombardment of the target metal ion solution with protons having an energy of 12-13 MeV, preferably 12.5 MeV, for example at a current of 5 μA for 5 to 20 min (Zn) or at a current of 10-20 μA for 5-15 min (Sc). For irradiation of Y to produce 89Zr, the irradiation may comprise bombardment of the target metal ion solution with protons having an energy of 12.5-15 MeV, for example at a current of 25 μA for around 5 min. For irradiation of 64Ni to produce 64Cu, the irradiation may comprise irradiation with 11 MeV protons using 20 μA current for 360 min. The mixture of radiometal ions and target metal ions in aqueous solution may comprise a concentration of target metal ion salt, such as chloride salt, of at least 0.1 M, preferably 1 M, or preferably more than 5 m, such as 7 m.
In a third aspect, the present invention provides a method of generation of radiometal ions from a target metal, comprising:
Suitably, the irradiation is conducted using a cyclotron. Suitably, when irradiating Zn to produce 68Ga, or Sc to produce 45Ti, the irradiation may comprise bombardment of the target metal with protons having an energy of 12-13 MeV, preferably 12.8 MeV, for example at a current of 10 μA for 160 min (Zn) or at a current of 10-20 μA for 5-15 min (Sc). For irradiation of Y to produce 89Zr, the irradiation may comprise bombardment of the target metal with protons having an energy of 13.1 MeV, for example at a current of 25 μA for around 5 min. For irradiation of 64Ni to produce 64Cu, the irradiation may comprise irradiation with 11 MeV protons using 20 μA current for 360 min.
Suitably, Ni is irradiated in the form of an electroplated layer. Suitably, the irradiated Ni (the solid mixture of radiometal and target metal) is dissolved in a 30% HCl for 30 min at 60° C., and additional 5 minutes at 80° C., then diluted to 6M HCl. Suitably, Sc, Y and Zn are each irradiated in the form of a metal foil. Suitably, the irradiated Sc and Y foils (the solid mixture of radiometal and target metal) are dissolved in 30-37%12M HCl at ambient temperature for a sufficient time to dissolve the foil, usually a few minutes, then diluted to 12M HCl. Suitably, the irradiated Zn foils (the solid mixture of radiometal and target metal) are dissolved in 3 M or 6 M HCl at ambient temperature for a sufficient time to dissolve the foil, usually a few minutes.
These solutions can be directly used for the liquid-liquid extraction.
In this aspect, the aqueous solution containing target metal ions resulting from the liquid-liquid extraction cannot be directly recycled for further irradiation as it is not a solid metal foil, though it can be recycled by use in a process according to the second aspect of the invention. Preferably, if the said solution is to be used in a process according to the second aspect of the invention, it is first subjected to a step of treatment to remove any organic solvents from the solution. Preferably, the treatment step comprises passage of the aqueous solution of the target metal ions through a reverse phase chromatography column having a stationary phase suitable for adsorption of any trace organic solvents. Preferably, the reverse phase chromatography column is a C18 column, ie an octadecyl carbon chain bonded silica stationary phase column.
In a fourth aspect, the present invention provides a method of production of a radiolabelled pharmaceutical, wherein the radiometal used in the radiolabeling is selected from 45Ti and 89Zr, comprising the method of the second or the third aspect of the invention, followed by the step of reaction of the solution of separated radiometal ions resulting from step c with a reactive precursor of the radiolabelled pharmaceutical. Reaction protocols for the production of radiolabeled pharmaceuticals are well known to the skilled person. For example, the synthesis of an 89Zr containing radiotracer has been described27, in which 89Zr in a 1 M HEPES buffer is pH adjusted to within the range 6.8-7.2 with 2M sodium hydroxide or 2M hydrochloric acid. Trastuzumab-DFO (10 mg/mL) is added to the solution to create the reaction solution, which is pumped through a single channel reactor at a total flow rate of 20 μL/min, optionally followed by incubation at 37° C. for 1 h by halting the flow. The product, 89Zr-Trastuzumab, is collected in a microcentrifuge tube and the radiochemical yield confirmed by instant TLC. A method of producing a 45Ti containing radiotracer is described18 and is reproduced in Example 4. Copper radionuclide containing radiopharmaceuticals, such as Cu-ASTM used in imaging hypoxic tissues, have been described75.
In a fifth aspect, the present invention provides a method of production of a radiolabelled pharmaceutical, wherein the radiometal used in the radiolabeling is 68Ga, comprising the method of the second or the third aspect of the invention, and further comprising:
a back extraction procedure as described above in the first aspect of the invention, in which the organic phase is that resulting from the separation step of the second or third aspect of the invention; followed by
reaction of the aqueous solution resulting from the back extraction procedure with a reactive precursor of the radiolabelled pharmaceutical.
Reaction protocols for the production of radiolabeled pharmaceuticals are well known to the skilled person. For example, the synthesis of a 68Ga-containing radiotracer has been described,8 in which 40 μl of an aqueous solution of 68Ga is added to a PSMA conjugate (0.1-1 nmol in 0.1 M HEPES buffer, pH 7.5, 100 μl) and 10 μl HEPES buffer (2.1 M in H2O). The pH of the solution is adjusted using NaOH. The reaction mixture is incubated at room temperature or 95° C., depending on the conjugate used. A 68Ga-PSMA radiotracer is produced.
In a sixth aspect, the present invention provides the use of phase separation in flow, comprising the use of a microfiltration membrane to separate an organic phase from an aqueous phase based on the interfacial tension between the phases such that a permeate phase passes through the membrane and a retentate phase does not, in the liquid-liquid extraction of a radiometal ion from a target metal ion. Features mentioned in connection with the first aspect of the invention are also relevant to the sixth aspect of the invention.
In a seventh aspect, the present invention provides apparatus for conducting separation of a radiometal ion from a target metal ion by means of a liquid-liquid extraction and phase separation carried out in continuous flow, the phase separation being preferably according to the first aspect of the invention, comprising:
a first inlet for an aqueous phase comprising the radiometal ion and the target metal ion;
a second inlet for an organic phase comprising an extractant and an interfacial tension modifier;
one or more mixers for mixing the organic phase and the aqueous phase;
tubing to convey the mixture of the organic phase and the aqueous phase;
a phase separation apparatus comprising a microfiltration membrane to separate the organic phase from the aqueous phase based on the interfacial tension between the phases such that a permeate phase passes through the membrane and a retentate phase does not;
a first outlet for the aqueous phase exiting the phase separation apparatus;
a second outlet for the organic phase exiting the phase separation apparatus.
Preferably, the phase separation apparatus further comprises a pressure controller to control the pressure ΔPmem exerted across the microfiltration membrane.
Preferably, the pressure controller is in the form of a diaphragm. Suitably, the diaphragm is made of a polymer selected from the group consisting of perfluoroalkoxyalkane (PFA), latex, polytetrafluoroethylene (PTFE), fluorinated ethylene propylene (FEP), fluoroelastomers (FMK), perfluoroelastomers (FFKM), tetrafluoro ethylene/polypropylene rubbers (FEPM), neoprene, nitrile rubber, and polyethylene. Preferably, the diaphragm is made of perfluoroalkoxyalkane (PFA). Preferably, the diaphragm thickness is 0.002″ (0.0508 mm).
Suitably, the microfiltration membrane is made from a polymer selected from the group consisting of polytetrafluoroethylene (PTFE), polyvinylidene fluoride (PVDF), cellulose acetate, polysulfane, polysulfone, polyethersulfone, polypropylene, polyethylene, and polyvinyl chloride (PVC). Preferably, the microfiltration membrane is made from polytetrafluoroethylene (PTFE). Suitably, the microfiltration membrane has a pore size selected in the range 0.1 to 1.0 μm, such as 0.1 to 0.5 μm, or 0.2 to 0.5 μm. Preferably, the microfiltration membrane has a pore size of 0.2 μm.
Suitably, the one or more mixers comprise mixers selected from the group consisting of: Y-junction mixing tees; T-junction mixing tees; static mixers; packed beds containing sand, stainless steel beads or glass beads; or combinations thereof. Preferably, the one or more mixers are one T-junction mixing tee and two static mixers. Preferably, the T-junction mixing tee is made of polyethyletherketone (PEEK) and the static mixers are made of polytetrafluoroethylene (PTFE).
Preferably, the apparatus further comprises a mixing loop between the one or more mixers and the phase separation apparatus. Preferably, the mixing loop is made of PFA tubing. Preferably, the mixing loop is 108 cm long.
Preferably, the tubing used in the apparatus is PFA tubing.
Preferably, the apparatus further comprises a pump, such as a syringe pump, upstream of each of the first inlet and the second inlet to drive the aqueous phase and the organic phase, respectively, therethrough.
In an eighth aspect, the present invention provides apparatus for conducting separation of a radiometal ion from a target metal ion, particularly the separation of 68Ga from Zn, by means of a liquid-liquid extraction carried out in continuous flow, followed by back-extraction of the radiometal ion, comprising:
a first apparatus according to the seventh aspect of the invention for conducting the liquid liquid extraction;
a second apparatus according to the seventh aspect of the invention for conducting back extraction of the radiometal ion, in which the first inlet is for an aqueous phase for back-extraction of the radiometal ion, and the second inlet is for the organic phase containing the radiometal ion obtained from the first apparatus. Suitably, the second outlet of the first apparatus is connected directly or indirectly to the second inlet of the second apparatus.
Preferably, the apparatus further comprises a third apparatus according to the seventh aspect of the invention for conducting a second liquid liquid extraction; and a fourth apparatus according to the seventh aspect of the invention for conducting a second back extraction of the radiometal ion, in which the first inlet is for an aqueous phase for the second back-extraction of the radiometal ion, and the second inlet is for the organic phase containing the radiometal ion obtained from the fourth apparatus. Suitably, the first outlet of the second apparatus is connected directly or indirectly to the first inlet of the third apparatus. Suitably, the second outlet of the third apparatus is connected directly or indirectly to the second inlet of the fourth apparatus.
In a ninth aspect of the invention, the present invention provides apparatus for conducting separation of a radiometal ion from a target metal ion, particularly the separation of 68Ga from Zn, by means of a liquid-liquid extraction carried out in continuous flow, followed by scrubbing of target metal ion from the organic phase, and then back-extraction of the radiometal ion, comprising:
a first apparatus according to the seventh aspect of the invention for conducting the liquid liquid extraction;
a second apparatus according to the seventh aspect of the invention for conducting scrubbing of the organic phase exiting the first apparatus, in which the first inlet is for an aqueous phase for scrubbing the organic phase, and the second inlet is for the organic phase containing the radiometal ion obtained from the first apparatus;
a third apparatus according to the seventh aspect of the invention for conducting back extraction of the radiometal ion, in which the first inlet is for an aqueous phase for back-extraction of the radiometal ion, and the second inlet is for the organic phase containing the radiometal ion obtained from the second apparatus. Suitably, the second outlet of the first apparatus is connected directly or indirectly to the second inlet of the second apparatus. Suitably, the second outlet of the second apparatus is connected directly or indirectly to the second inlet of the third apparatus.
Preferably, the apparatus further comprises a fourth apparatus according to the seventh aspect of the invention for conducting a second liquid liquid extraction; a fifth apparatus according to the seventh aspect of the invention for conducting a second back extraction of the radiometal ion, in which the first inlet is for an aqueous phase for the second back-extraction of the radiometal ion, and the second inlet is for the organic phase containing the radiometal ion obtained from the fourth apparatus. Suitably, the first outlet of the third apparatus is connected directly or indirectly to the first inlet of the fourth apparatus. Suitably, the second outlet of the fourth apparatus is connected directly or indirectly to the second inlet of the fifth apparatus. Preferably, the apparatus further comprises means for acidification of the aqueous phase between the first outlet of the third apparatus and the first inlet of the fourth apparatus.
In each of the eighth and ninth aspects of the invention, the preferred features of each apparatus are as recited for the seventh aspect of the invention. The features and preferred features for each apparatus may be the same or different; preferably, each apparatus is the same.
In a tenth aspect, the present invention provides apparatus for on-demand production of a radiometal from a target metal, comprising:
apparatus for irradiation of a target metal;
apparatus for separation of the radiometal from the target metal according to any one of the seventh to ninth aspects of the invention.
Suitably, the apparatus for irradiation of a target metal comprises a cyclotron, such as a GE PETTrace PT800 cyclotron. Suitably, the apparatus for irradiation of a target metal further comprises means for cooling the target metal, such as direct water cooling. Where the target metal is provided in solution, the apparatus comprises a liquid target chamber, preferably made of niobium, such as a GE PETTrace liquid target chamber.
In an eleventh aspect, the present invention provides apparatus for on-demand production of a radiolabeled compound, comprising:
apparatus for irradiation of a target metal;
apparatus for separation of the radiometal from the target metal according to any one of the seventh to ninth aspects of the invention;
apparatus for reaction of the radiometal solution obtained from the separation step with a reactive precursor of the radiolabeled compound.
Suitably, the apparatus for irradiation of a target metal comprises a cyclotron, such as a GE PETTrace PT800 cyclotron. Suitably, the apparatus for irradiation of a target metal further comprises means for cooling the target metal, such as direct water cooling. Where the target metal is provided in solution, the apparatus comprises a liquid target chamber, preferably made of niobium, such as a GE PETTrace liquid target chamber.
Suitably, the apparatus for reaction of the radiometal solution comprises continuous flow reaction apparatus, such as has been widely described in the literature47. For example, the apparatus may comprise, in suitable combinations for the reaction to be carried out: pumps, such as syringe pumps; mixers, such as Y-junction mixing tees, T-junction mixing tees, static mixers, packed beds containing sand, stainless steel beads or glass beads; mixing and/or reaction loops of tubing of suitable length for the reaction process; heating and/or cooling apparatus such as water, ice or oil baths through which the reaction tubing passes.
Apparatus for Separation of Radiometal Ions
As noted above, liquid-liquid extraction (LLE) is a widely used means of separation of components of a solution by partitioning the components between two different solvents. Traditionally, this has been conducted between immiscible solvents which separate under the influence of gravity in such apparatus as a separatory funnel, or which are forced to separate by use of apparatus such as a centrifuge. The extraction is based on the relative solubilities of the components of the solution in the chosen immiscible liquids, usually an aqueous phase and an organic phase. In some cases, the system of organic and aqueous phases plus one or more components may form an emulsion or “third phase” which can prevent or make less effective the partitioning of the components between easily separable organic and aqueous phases.
“Extraction” is used to describe the transfer of a component from the aqueous phase to the organic phase, whereas “stripping” or “back-extraction” describes the transfer of a component of interest from the organic phase to the aqueous phase. Removal of an unwanted component from the organic phase is described as “scrubbing”.
In order that an extraction or back-extraction takes place efficiently, it is necessary for the organic and aqueous phases containing the components to be extracted to be thoroughly mixed, in order to permit partitioning of the components between the phases according to their relative solubilities in the phases, followed by a means of separating the two phases from one another. Traditionally, in a separatory funnel, this would be carried out by shaking the separatory funnel containing the phases and components, followed by allowing the phases to separate under gravity, and running off the phases in turn from the bottom of the funnel. More than one extraction step can be carried out to ensure the maximum extraction of the desired components.
It has been recognized by the present inventors that it would be desirable to avoid the handling of solutions containing radioactive materials by an operator, as would be required by these traditional manual methods of liquid-liquid extraction. Nonetheless, such separations are attractive as they are simple and inexpensive to carry out, compared with other methods such as chromatographic methods that may require expensive media, and may result in the need to dispose of radioactively-contaminated media following use. In addition, liquid-liquid extraction can be conducted on a wide range of scales.
Recently, new methods of, and apparatus for, conducting mixing of fluids and separation of fluids have been developed, which methods may be conducted in a continuous manner. The present inventors have recognized that such apparatus and methods could potentially have applicability in the separation of metal ions, in particular the separation of radiometal ions from target metal ions. As far as the present inventors are aware, no such use has been made of these new methods and apparatus in the field of the present invention.
Continuous flow synthesis apparatus has been developed recently that allows reactions to be carried out in a continuous manner. A review52 of transformations that have been carried out in continuous flow systems lists transformations such as hydrogenations and reductions, oxidation, acid or base catalyzed bond-forming reactions, transition metal catalyzed bond-forming reactions, esterification reactions, protection and deprotection reactions, photocatalysis and enzymatic reactions. Particular attention has been paid to the use of reactive and/or toxic, particularly gaseous, reagents, as they can be generated and used in a closed system: for example, cyanogen bromide53, chlorine azide54, ethylene gas55, (meth)acryloyl chloride56, chlorine gas57 and diazomethane58. The use of such systems in the automated production of drugs has been suggested59. Some continuous purifications involving metals have been reported, such as removal of excess ligand from nanoparticles60 and extraction of leached copper from a target compound61. However, so far as the present inventors are aware, such technologies have not been applied to the separation of metal ions from one another.
WO2004/087283 describes systems which may be used for liquid-liquid separations, amongst other uses, and in which the separation is carried out by means of differential wetting of arrays of capillary tubes. For example, a hydrophilic and a hydrophobic liquid may be mixed, and the mixture brought in contact with one or more capillary tubes having a hydrophobic coating. The hydrophobic liquid thus wets the capillary tube and rises up it, whereas the hydrophilic liquid does not wet the capillary tube and does not enter it. In this way, the hydrophobic liquid passes through the array of capillary tubes and is separated from the hydrophilic liquid.
WO2014/026098 describes a membrane separation apparatus suitable for the separation of a first fluid (permeate) from a second fluid (retentate) based on the interfacial tension of the two fluids. In particular, it relates to a system in which a pressure controller is included in the apparatus to apply pressure across the microfiltration membrane that is independent of the pressure downstream of the device, and which can control the selectivity of the membrane for the passage of the fluids, such that one fluid can be allowed to pass selectively through the membrane thus separating it from the other fluid. Such a membrane separation unit is available from Zaiput Flow Technologies. A schematic of such a separator 10 is depicted on the right hand side of
Without wishing to be bound by theory48, it is believed that, in order that the separation is complete, the capillary pressure Pcap associated with the fluid passageways in the membrane and the mixture of fluids to which the membrane is exposed, must not be exceeded, or both fluids will be forced through the membrane. Thus, ΔPmem, the pressure difference across the membrane, may not exceed Pcap. Pcap is quantified as:
where θ is the contact angle formed between the solid material of the membrane, the first fluid to be separated and the second fluid to be separated, r is the radius of the membrane pores, and y is the interfacial tension with respect to the first fluid to be separated and the second fluid to be separated.
Further, in order to ensure that the whole of the first fluid passes through the membrane, ΔPmem must exceed the pressure Pper needed to cause the permeate liquid to pass through the membrane. Pper is quantified as:
where μ is the viscosity of the permeate phase, Q is the entering permeate fluid volumetric flow rate, L is the membrane thickness, n is the number of pores, and R is the pore radius; this assumes that the membrane acts as an array of cylindrical pores.
In addition, the separator must be operated at a flow rate which is suited to the available membrane area; if the flow rate is excessive, both phases may exit both outlets.
Where the pressure drop along the length of the membrane is negligible compared to Pcap−Pper, then ΔPmem can be assumed to be constant along the length of the membrane, and the conditions for successful separation are
P
cap
>ΔP
mem
>P
per
The first inequality is satisfied by selection of the microfiltration membrane material and pore size in a range appropriate for the separation, and the second by ensuring that the pressure on the retentate side of the membrane is greater than that on the permeate side of the membrane; this additional pressure is provided by the pressure controller. In practice, the actual operating range of pressures is often narrower than the theoretical range given above.
While the apparatus for mixing of the two phases is described here as a mixing tee followed by static mixers, it will be appreciated that other combinations of mixing apparatus (either passive or active) can be used depending on the degree of mixing required, the nature of the fluids to be mixed, and the volume of fluid to be mixed. For example: fewer or more static mixers may be employed; the mixing tee may be a Y-junction mixer or a T-junction mixer; a packed bed reactor housing sand, stainless steel or glass beads may replace one or more of the mixers depicted in
The variable length mixing loop, along with the other tubing used in the apparatus, is made from a material chosen with reference to its chemical compatibility with the solvent system, and the pressures that it will need to withstand in operation; the present inventors have found that PFA tubing is suitable for use with the solvent systems used herein. The length of the mixing loop, and the other mixers used, are selected in order to ensure an adequate degree of mixing of the phases, and a residence time in the apparatus sufficient to ensure efficient partitioning of the metal ions between the phases, for the chosen solvent system and the metal ions to be separated.
For the separation of radiometals, the total production time is a critical parameter in the system since the radiometal is continuously undergoing decay back to the target metal. Therefore, the shortest possible residence time is desirable, and the mixing of the phases must be optimized to ensure efficient extraction in as short a time as possible.
It has been found by the present inventors that achieving liquid-liquid segmented flow (sometimes referred to as “slug flow”, though this term more usually refers to gas-liquid mixtures) in the mixed fluid stream passing through the tubing 80, mixing loop 130 and on to the membrane separator 10 is of importance in the present invention. Liquid-liquid segmented flow describes a flow pattern through a tube or pipe in which a first fluid is dispersed in a second fluid in the form of segments of varying length. During stable liquid-liquid segmented flow, the first fluid is shed from the back of the segment at the same rate as it is picked up at the front of the segment, and so the segment length remains constant as it travels along the tube. The present inventors have found that the high mass transfer in liquid-liquid segmented flow systems is particularly beneficial in allowing the efficient partition of components between two phases for the purposes of liquid-liquid extraction. Accordingly, the mixers provided in the apparatus are selected such that liquid-liquid segmented flow is provided in the mixing loop 130 for the combination and volume of fluids used. The present inventors have found that, in the solvent systems used herein, liquid-liquid segmented flow is achieved by mixing of the phases through mixing tee 90. When static mixers are used also, the performance of the extraction was further improved.
Liquid-liquid segmented flow may be determined by visual inspection of the mixture as it flows through the tubing, or may be detected for example by a phototransistor device which clips on to the outside of the tubing and detects a phase interface by alteration in current flow depending on the amount of light received. These devices can detect phase interfaces even in mixtures of colourless liquids. Such devices are available from Optek Technology (OPB350 and OCB350 series). These devices can also be used at the outlets of the separator to detect whether retention or breakthrough of a phase has occurred.
As discussed above, the membrane separator 10 comprises two main components: a polymer microfiltration membrane 30 and a thin diaphragm 60 (
The diaphragm 60 acts to modulate the pressure between the aqueous and organic sides of the membrane 30. The aqueous phase is retained by the membrane 30, while the organic phase permeates through the membrane 30. The physical properties and geometry of the membrane 30 as well as the chemical nature of the aqueous and organic phases and their interactions with the membrane surface determine the capillary and permeation pressures. The interactions between the aqueous and the organic phases determine the interfacial tension. The interplay between these parameters determines whether the conditions are within the operating range of the system. If they are not, incomplete phase separation will occur.
Pressure control may be provided by controlling the pressure at each of the outlets of the separation apparatus; however, to do so makes it difficult to integrate the apparatus with other downstream components. Accordingly, it is preferable to use a pressure controller, as described in WO2014/026098 and shown in
The diaphragm may be made from a polymer selected from the group consisting of perfluoroalkoxyalkane (PFA), latex, polytetrafluoroethylene (PTFE), fluorinated ethylene propylene (FEP), fluoroelastomers (FMK), perfluoroelastomers (FFKM), tetrafluoro ethylene/polypropylene rubbers (FEPM), neoprene, nitrile rubber, and polyethylene. The diaphragm material should be selected primarily with regard to its resistance to the solvent system to be used in the separation: for example, its acid resistance and/or resistance to organic solvents. The elasticity of the diaphragm following long periods of deformation or physical degradation will also be affected by the choice of diaphragm material. As perfluoroalkoxyalkane (PFA) is a very robust material and is mechanically strong, it is the preferred choice of diaphragm material.
The choice of diaphragm thickness is important as this directly affects the pressure exerted on across the membrane; this must be selected in combination with the membrane properties and solvent system to arrive at a functional apparatus for a given separation. Preferably, the diaphragm thickness is 0.002″ (0.0508 mm).
The microfiltration membrane may be made from a polymer selected from the group consisting of polytetrafluoroethylene (PTFE), polyvinylidene fluoride (PVDF), cellulose acetate, polysulfane, polysulfone, polyethersulfone, polypropylene, polyethylene, and polyvinyl chloride (PVC). As well as chemical compatibility with the solvent system to be used in the separation, the membrane material should be selected having regard to the wettability of the material by the organic phase to be used (for hydrophobic materials such as listed above). It has been found by the present inventors that polytetrafluoroethylene (PTFE) fulfils these requirements for the solvent systems used herein.
The choice of membrane pore size affects the upper and lower boundaries of the pressure of the system, as explained in detail above. Each individual membrane will have a range of pore sizes, and so the pore size specified herein is the manufacturer's specification of pore size, which will represent an average value. The pore size must be selected in combination with the membrane material, the diaphragm thickness and the solvent system to arrive at a functional apparatus for a given separation. Suitably, the microfiltration membrane has a pore size selected in the range 0.1 to 1.0 μm, such as 0.1 to 0.5 μm, or 0.2 to 0.5 μm. Preferably, the microfiltration membrane has a pore size of 0.2 μm.
It will be appreciated from the foregoing discussion that it may be possible for a number of different combinations of membrane material, pore size and diaphragm thickness to provide suitable separation conditions for a given solvent system. Similarly, a number of different solvent systems may be separable in an apparatus having a given combination of membrane material, pore size and diaphragm thickness.
Turning now to
Turning now to
In the scrubbing stage 272 shown in
Turning now to
Separation Methods
The selection of an appropriate extractant to conduct a liquid-liquid extraction in which the phase separation is conducted in flow comprising the use of a microfiltration membrane to separate the phases based on interfacial tension is crucial: the system must provide selective extraction of the metal ion of interest, as little extraction as possible of the target metal ion, must be stable in the presence of the strongly acidic solutions often used in the generation of radiometals (to dissolve irradiated metal foils or irradiated electroplated layers, and/or to avoid the hydrolysis of susceptible metals such as Ti or Zr), and must have a sufficiently high interfacial tension with the aqueous phase that complete separation can be achieved using the microfiltration membrane. This is a much more demanding set of criteria than need be applied to standard batch liquid-liquid separations carried out on the basis of density.
While the present inventors have made reference to literature reports of extraction of particular individual metal ions in batch processes, and in some cases to separation of metal ions from one another in batch processes, it was not expected that these literature conditions would be directly applicable in the processes of the present invention. Indeed, a number of the literature conditions simply did not work at all under the necessary conditions. For example, it had been reported that liquid-liquid batch separation of 68Ga ions from 68Zn ions in an aqueous solution of protic acid could be carried out using isopropyl ether28. However, under the conditions required for 68Ga production (in particular, the presence of around 7 m ZnCl2 concentration), the phase equilibrium simply did not allow efficient separation; the present inventors found that 65% of the ZnCl2 present migrated into the organic phase, thus heavily contaminating the 68Ga solution with Zn. Other conditions were found to be too inefficient for application in the process of the present invention, for example the use of 1-octanol in the separation of 45Ti from Sc, which allowed only 50% extraction efficiency. Yet other extractants attempted by the present inventors did not provide clean phase separation when the microfiltration membrane was applied, but instead led to breakthrough, retention or the formation of emulsions.
It was surprisingly discovered by the present inventors that it is possible to adjust the properties of an extractant with respect to the aqueous phase such that an extractant that did not provide clean phase separation when used alone could do so with the addition of a carefully selected second solvent, here referred to as an interfacial tension modifier. The interfacial tension modifier must not interfere with the interactions between the radiometal ion and the extractant, must not extract the target metal ion to any significant degree, must not dissolve water to any significant extent, and must be able to adjust the properties of the interfacial tension of the overall solvent system (extractant, aqueous phase and interfacial tension modifier (if present)) with respect to the microfiltration membrane such that complete separation of the phases by the microfiltration membrane was possible. The amount of the interfacial tension modifier must be selected carefully to provide optimum separation, as must the relative flow rates of the organic phase (extractant and interfacial tension modifier) and the aqueous phase. The interfacial migration (ie the tendency of one phase to contaminate the other) is a critical parameter which must be minimized to prevent the contamination of aqueous phase with the organic phase, which would make the process incompatible with recycling the target metal solution for a further irradiation step, due to stringent organic-free requirements for aqueous cyclotron solution targets, particularly ZnCl2.
Thus, for the separation of 45Ti from Sc, and for the separation of 89Zr from Y, the extractant chosen is a solvent having the ability to function as a bidentate ligand for the radiometal via two oxygen atoms, preferably thus forming a five membered ring, as well as having a suitable interfacial tension with 12 M (37%) HCl. Suitable extractants may be maltol, vanillin, eugenol, and guaiacol (o-methoxyphenol), with guaiacol being particularly preferred. The interfacial tension modifier is a solvent having similar properties to the extractant, though not having the ability to function as a bidentate ligand for the radiometal ion, such that it does not interfere with the ability of the extractant to interact with the radiometal ions, as well as the ability to modify the interfacial tension of the overall system to allow complete separation. Suitable interfacial tension modifiers may be fluorobenzene, trifluorotoluene, thiophene and anisole, with anisole being particularly preferred. For the preferred system of guaiacol and anisole, an amount of anisole of at least 10% v/v is found to perform particularly well, and the optimum flow ratio for the organic phase to the aqueous phase to be greater than 3 to 1, and in some cases 5 to 1. Alternatively, where the radiometal ion is a Zr ion and the target metal ion is a Y ion, it is preferable that the extractant is 0.1 M trioctylphosphine oxide (TOPO), the interfacial tension modifier is hexane, and the aqueous phase is a solution in 6 M HCl.
For the separation of 68Ga from Zn, the use of ether extractants was found to work well on combination with an interfacial tension modifier selected from the group consisting of: a fluorinated aromatic hydrocarbon; an aromatic hydrocarbon; an alkoxybenzene; a halogenated alkane, for example selected from the group consisting of 1,2-dichloroethane, 1,1,2-trichloroethane, 1,1,1-trichloroethane, hexachloroethane and bromoethane; and an alkane; particularly, selected from the group consisting of toluene, anisole, 1,2-dichloroethane, trifluorotoluene and heptane, with trifluorotoluene being the most preferred. Preferably, the ratio of the extractant to the interfacial tension modifier is 1:2 by volume, and the optimum flow ratio for the organic phase to the aqueous phase is greater than 3 to 1.
For the separation of 64Cu from 64Ni, the use of a phosphine oxide extractant was found to work well with an interfacial modifier selected from an aromatic hydrocarbon and an aliphatic alkane which may be cyclic or acyclic; for example selected from the group consisting of n-pentane, n-hexane, n-heptane, n-octane, n-nonane, n-decane, n-undecane, i-hexane, neo-hexane, i-heptane, neo-heptane, cyclohexane, cycloheptane, cyclooctane, kerosene, light petroleum, benzene, naphthalene, toluene, ethylbenzene, dimethylbenzene and iso-octane and mixtures thereof; particularly selected from the group consisting of toluene, heptane and hexane. Preferably, the concentration of the extractant in the interfacial tension modifier is at least 0.1 M, such as from 0.1 M to 0.4 M, and the optimum flow ratio for the organic phase to the aqueous phase is from 5:1 to 3:1.
Separation of Ti and Sc
Materials
Guaiacol (99%), anisole (99%), 1-octanol (99%), titanium (IV) chloride (neat), titanium (IV) chloride solution (0.09 M in 20% HCl), hydrochloric acid (37%), sulfuric acid (95.0-98.0%) and pyridine-2,6-dicarboxylic acid (dipic) (98%) were purchased from Sigma Aldrich and used without further purification. TLC plates (Silica gel on TLC Al foil) were purchased from Sigma Aldrich. Scandium (III) chloride (anhydrous, 99.9%) and scandium foil (250 μm, 99.9% pure, rare earth analysis) were purchased from Alfa Aesar. Custom Ti and Sc ICP standards were purchased from Inorganic Ventures (100 ppm of each metal in a 5% HCl solution). Salan62 and (salan)Ti(dipic)3 were synthesized according to the literature procedures.
The membrane separator module was similar to those manufactured by Zaiput Flow Technologies. The aqueous and the organic phases were combined through a tee and mixed with two static mixers and mixing tubing. The aqueous phase was retained by the membrane, while the organic phase permeated through the membrane. Under the conditions of direct extraction, the radionuclide 45Ti was selectively extracted from scandium into the organic phase. Pall PTFE membranes were used for all experiments (47 mm diameter, 0.1/0.2/0.5 μm pore size, polypropylene support for the 0.1/0.2 μm pore sizes). Perfluoroalkoxy alkane (PFA) diaphragms (0.001″/0.002″/0.005″ (0.0254/0.0508/0.1270 mm)) were purchased from McMaster Carr. All PFA tubing ( 1/16″ (1.5875 mm) OD, 0.03″ (0.762 mm) ID) was purchased from Idex Health and Science. PTFE static mixers were purchased from Stamixco. The 15 mL plastic centrifuge tubes with screw caps were purchased from VWR.
Radionuclide Production and Separation
For all experiments, the cyclotron target material, scandium, was used at its natural abundance level.
45Ti was produced by 10-20 μA proton irradiation of 30-60 mg scandium foil, for 5-15 min using a GE PETtrace cyclotron. To minimize coproduction of the 44Ti (half-life=60.0 years), a 500 μm thick aluminium foil was used to degrade the incidental 16 MeV beam to approximately 13 MeV. The irradiated foil was digested in 30-37% M HCl. The mixture was filtered and centrifuged if necessary. The solution was diluted with concentrated hydrochloric acid to make the final dilution ca. 12 M in HCl. These dilutions were used as the aqueous phase for the LLEF.
Instrumentation and Methods
The solutions for the continuous membrane-based separation were pumped using either the KDS 100 Legacy Syringe (radioactive experiments) or the Harvard Apparatus PHD 2000 Programmable and Infusion syringe pumps (non-radioactive experiments). The NMR spectra were taken on Agilent 400 MR operating at 400.445 MHz (1H). Radio-TLC was performed with a Raytest MiniGita TLC scanner using chloroform/ethyl acetate (1/1, (v/v)) as a mobile phase. The HPLC and radio-HPLC analyses were performed on a Hitachi Chromaster equipped with a Carrol&Ramsey 105-S radio-detector and a Hitachi 5430 double diode array detector. Column: Phenomenex Luna 3μ C18(2) (100 Å, 100 mm×2.00 mm). Flow: 0.5 mL/min. Eluents: (A) 0.1% (v/v) CF3COOH in Milli-Q water, (B) 0.1% (v/v) CF3COOH in CH3CN. The radiochemical identity of [45Ti](salan)Ti(dipic) was established by comparing its retention time with that of its natural abundance isotopomer. The radiochemical conversion (RCC) was determined by radio-HPLC or radio-TLC and calculated as:
RCP=(Areaproduct/Total Area)*100%.
The extent of extraction (extraction %) was determined from the radioactivity measurements and using inductively coupled plasma atomic emission spectroscopy (ICP-AES, Agilent 5100 Dual View) of the aqueous phase. Samples of the aqueous phase were collected before the LLE and after 5, 15, 30, and 45 minutes of LLE. 0.35 mL of each sample was digested in 5 mL with 10% (v/v) H2SO4 for 6 hours at 160° C. 2.7 mL of the digested sample was diluted up to 10 mL with Milli-Q water to reach a total acid concentration of 5% (v/v). Calibration standards (Inorganic Ventures) were prepared to match the sample matrix (Sc) with concentrations of 22.2, 18, 15, 10, and 5 ppm Ti and Sc and run prior to every set of samples. Samples were analyzed in radial view at a viewing height of 8 mm. The extraction to the organic phase was calculated from the concentration of Ti and Sc in the aqueous phase before and after the LLE by E=((cbefore LLE−cafter LLE)/cbefore LLE)−100%.
Batch LLE and Separation
For the non-radioactive work, 0.5 mL 0.01 M TiCl4 and 0.01 M ScCl3 in 37% HCl was mixed with extractant(s) and shaken for 2 min in a centrifuge tube. The phases were allowed to separate by gravity. The concentration of Ti and Sc in the aqueous phase before and after the LLE was measured by ICP-AES.
For the batch LLE of 45Ti, a centrifuge tube was charged with 2 mL of the solution of 45Ti (10-50 MBq) in 37% HCl and 2 mL of the organic phase. The mixture was shaken vigorously, spun for 15 minutes, and centrifuged at 4000 rpm to separate the phases.
An Eppendorf 5702 centrifuge was used to assist in phase separation. For batch experiments, the phase mixing was performed using a IKA ROCKER 3D digital shaker.
Continuous Membrane-Based LLE and Separation
The continuous liquid-liquid extraction and phase separation in flow was performed using a membrane-based separator with a PFA diaphragm for integrated pressure control. A flow schematic of the apparatus is depicted in
The preliminary screening experiments were performed in batch using gravity separation. It had been reported64 that 45Ti can be extracted from aqueous HCl into 1-octanol, presumably as [45Ti]Ti n-octyloxide 1, Structure 1.
a20% HCl;
btrifluorotoluene/hexafluoropropanol (1/1, (v/v))
We observed little extraction when a solution of 45Ti in 20% HCl was used (Table 1, entry 1). Using 37% HCl (12 M) significantly improved the extraction. 1,2-Decanediol used as co-extractant as a 0.1 M solution in 1-octanol gave only a slight improvement over neat 1-octanol. 2,3-naphthalene diol, reported, in the context of the extractive spectrophotometric determination of Ti content in rocks, to extract Ti at pH=4,65 showed similarly modest performance. (Table 1, entries 3 and 4). During these batch extractions, we noticed a significant increase in the volume of the 1-octanol phase suggestive that conc. HCl was migrating into the organic phase. In an attempt to improve the phase separation we turned to perfluorinated extractants.66 Disappointingly, the fluorous analog of 1-octanol, CF3(CF2)5CH2CH2OH, formed an emulsion. A 0.05 M solution of C10F21CH2CH(OH)CH2OH in trifluorotoluene/hexafluoropropanol (1/1, (v/v)) failed to extract any 45Ti. (Table 1, entry 5). Previously, it was reported, in the context of investigating catalytic complexes of Ti, that guaiacol (o-methoxyphenol) and maltol were able easily to form moisture-sensitive but isolable complexes with titanium tetrachloride.67 Gratifyingly, using neat guaiacol as an extractant, we were able to extract 75% of activity into the organic phase, presumably as Structure 2.
The LLE and phase separation in flow were performed using a membrane-based separator with a PFA diaphragm for integrated pressure control (
A flow schematic of the experimental setup is analogous to what has been depicted in
To find the optimum extraction conditions, the diaphragm thickness, membrane pore size, organic phase composition, flow rate ratios, and residence times were varied. Once the optimal materials and conditions were determined, a study was conducted to determine the shortest residence time, thereby minimizing the dead volume and overall processing time, while still maintaining high extraction. The residence time was varied by varying the length of the mixing tubing after the static mixers while maintaining a constant flow rate. All systems were operated for 60 min each and samples were collected every 15 min.
Investigation of Membrane Pore Size, Diaphragm Thickness and Organic Phase Composition
Due to the harsh nature of both the aqueous and organic solvents used for this extraction, both the membrane and diaphragm had to be extremely stable. Therefore, PTFE membranes were used for all experiments described herein.
Complete separation requires both the diaphragm thickness and membrane pore size be chosen so that Pdia lies between the Pcap and Pperm pressures. In general, low interfacial tension mixtures often require smaller pore size membranes and thinner diaphragms. In these experiments, polytetrafluoroethylene (PTFE) membranes were tested using the following pore sizes: 0.1, 0.2, and 0.5 μm. Three different diaphragm film thicknesses were also tested: 0.001″, 0.002″, and 0.005″ (0.025 mm, 0.051 mm, and 0.127 mm).
Since guaiacol had shown the most selective and highest extraction efficiency for Ti over Sc in batch, it was chosen as a candidate for translation into flow, with 1-octanol also being investigated.
It can be seen from Table 2 that no satisfactory conditions could be found for the use of 1-octanol as extractant. As the extraction efficiency was only around 50%, it was decided not to investigate this extractant further.
A solution of 0.01 M TiCl4 and 0.01 M ScCl3 in 37% HCl was extracted into guaiacol using the membrane separator.
Occasional retention and/or breakthrough of the aqueous into the organic phase was observed with a 0.2 μm PTFE/PP membrane, 0.002″ (0.051 mm) diaphragm thickness, and 0.2 mL/min total flow rate. The situation was remedied by adding various amounts of anisole as an interfacial tension modifier, which is structurally similar to guaiacol but acted to increase the interfacial tension, as shown in Table 3. Although the guaiacol/anisole mixture performed much better than 1-octanol, its extraction efficiency was not high enough. Therefore, the ratio of guaiacol to anisole was varied as well, as summarized in Table 5.
Investigation of Flow Rate Ratios and Organic Phase Compositions Since a 0.2 μm membrane and a 0.002″ (0.051 mm) diaphragm was the only combination that led to complete phase separation, it was used for all of the optimization experiments.
The organic mixtures of guaiacol and anisole were used to extract Ti, (0.01 M) from 37% HCl at a total flow rate of 0.20 mL/min and aqueous to organic ratios of 1/5, 1/3, 1/1, 3/1, and 5/1 (v/v). Corresponding flow rates are shown in Table 4.
The composition of the organic phase needed to both selectively extract only Ti and have a high enough interfacial tension with the HCl phase that complete separation could be achieved. It was determined that extraction was directly correlated with guaiacol concentration, that is a higher guaiacol concentration led to higher extraction up to a maximum Ti extraction of 90% with 90% guaiacol. Guaiacol concentrations above 90% led to incomplete phase separation. A summary of the phase separation performance using various organic phase compositions is shown in Table 5.
In addition to the composition of the organic phase, the relative ratios of aqueous to organic flow rates were also varied. When comparing relative flow rate ratios of 1/1, 1/3, and 1/5 (v/v) (aq. to org.) it was determined that 1/1 gave the lowest extraction. A ratio of 1/3 gave a higher extraction, but 1/5 did not yield a further increase in performance. All ratios where the aqueous flow rate was higher led to lower extraction efficiency. Therefore, a flow rate ratio of 1/3 was chosen as to avoid using excess solvent (Table 5).
Investigation of Scalability and Stability of the Extraction
In order to determine the scalability and stability of the extraction, the total flow rates were scaled five-fold to a total flow rate of 1.00 mL/min, while maintaining the same flow rate ratios and residence times. The extraction performance was identical to the original scale, and the maximum extraction of 90% was still achieved at 90% guaiacol and a flow rate ratio of 1/3 (
Investigation of Residence Time
After an optimal system was developed, the residence time of mixing was varied to minimize the dead volume and decrease the total amount of time spent in the system. This was achieved by increasing or decreasing the length of the PFA tubing used for mixing. The following lengths were tested with their corresponding residence times at 0.20 mL/min: 10 cm (13.7 s), 25 cm (34.2 s), 54 cm (73.9 s), 108 cm (147.8 s), 216 cm (295.6 s).
The total production time is a critical parameter in this system since the radioactive Ti is continuously undergoing decay back to Sc (t½=3 hours). Therefore, the shortest possible residence time is desirable. Residence times of the mixing tubing were varied from 13.7 s up to −5 min. The extraction efficiency was the same for all residence times. Therefore the shortest residence time was the most optimal, and resulted in a total system residence time of less than 1 min (
Conclusion—Optimal Conditions
Overall, an organic phase consisting of 90% guaiacol and 10% anisole, total flow rates of 0.20 or 1.00 mL/min, an aqueous to organic flow rate ratio of 1/3, and a residence time of 13.7 s led to highest and most efficient extraction resulting in 90.3±1.1% extraction of Ti and 0% extraction of Sc.
With the extraction conditions optimized for natTi, we turned to the radioactive isotopomer, 45Ti. While the concentration of Sc was comparable in both non-radioactive and radioactive cases (0.01 M vs. 0.03-0.13 M correspondingly) the concentration of the radiometal in the Sc-containing matrix solution was lower than that of its natural abundance isotopomer by 10 orders of magnitude, ranging from 1 to 10 picomoles. At these concentrations, even trace levels of impurities or water could potentially lead to side-reaction or hydrolysis and, as a consequence, change the extraction efficiencies of 45Ti. To our delight, however, the LLE of 45Ti in flow using a guaiacol-anisole 9/1 (v/v) mixture and a flow rate ratio of 1/3 (aq. to org.), with a residence time of 13.7 s showed that the extraction efficiency of 45Ti was consistent with that of natTi (84.8±2.4% and 90.3±1.1% correspondingly), (
Finally, to examine if the extracted solution of 45Ti can be directly used for radiolabelling, we attempted a synthesis of [45Ti](salan)Ti(dipic) 3, a Ti-antineoplastic, previously used for animal 45Ti-PET and ex vivo radiotracing (Scheme 1).18
To that end, the organic phase after the continuous LLE of 45Ti was collected and reacted with an equimolar solution of salan and 2,6-pyridine dicarboxylic acid (dipic) in pyridine at 60° C. An essentially complete (98.7%) conversion to the desired product 3 was observed within 15 min as evidenced by radio-TLC (red peak for the product 3 and only traces of unreacted 2, green peak), proving the high quality and reactivity of extracted 45Ti (
Separation of Zr and Y
Materials
Guaiacol (99%, natural), anisole (99%, ReagentPlus), hydrochloric acid (37%, ACS reagent), sulfuric acid (95.0-98.0%), and trioctylphosphine oxide (TOPO, ≥98.5%) were purchased from Sigma Aldrich. High purity hydrochloric acid (37%, “Ultrapur”) was purchased from Merck. All purchased chemicals were used without further purification. Yttrium foil (99.9%) was purchased from Alfa Aesar. ZrCl4 and YCl3 were purchased from Sigma Aldrich.
Pall PTFE membranes were used for all experiments (47 mm diameter, 0.1/0.2/0.5 μm pore size, polypropylene support). Perfluoroalkoxy alkane (PFA) diaphragms (0.001″/0.002″/0.005″ (0.0254/0.0508/0.1270 mm)) were purchased from McMaster Carr. All PFA tubing ( 1/16″ (1.5875 mm) OD, 0.03″ (0.762 mm) ID) was purchased from Idex Health and Science. PTFE static mixers were purchased from Stamixco. Two syringe pumps (KDS 100 Legacy Syringe Pump) and a dose calibrator (CRC-55tR, CII Capintec, Inc.) were used for the experiments.
Radionuclide Production and Separation
For all experiments, the cyclotron target material (yttrium) was used at its natural abundance level.
89Zr was produced by proton bombardment of yttrium foils on a PETTrace PT800 cyclotron. The 640 μm thick, 5 mm×5 mm foils were cut and sandwiched between a silver disc and a 500 μm Al degrader and placed in the target holder, providing direct water cooling on the rear face of the silver. Based on SRIM calculations the Al foil degrades the incident proton energy from the nominal 16.5 to approx. 13.1 MeV, bringing the energy below the threshold for co-production (<100 μb) of both 88Y and 88Zr. The irradiated foil was digested in 30-37% HCl. The mixture was filtered and centrifuged if necessary. If needed, the solution was diluted with water to make the final dilution ca. 6 M in HCl. These dilutions were used as the aqueous phase for the LLEF.
Instrumentation and Methods
For 89Zr work, the extent of extraction (extraction %) was determined from the radioactivity measurements and using inductively coupled plasma atomic emission spectroscopy (ICP-AES, Agilent 5100 Dual View) of the aqueous phase. Samples of the aqueous phase were collected before the LLE and after 5, 15, 30, and 45 minutes of LLE. 0.35 mL of each sample was digested in 5 mL with 10% (v/v) H2SO4 for 6 hours at 160° C. 2.7 mL of the digested sample was diluted up to 10 mL with Milli-Q water to reach a total acid concentration of 5% (v/v).
An Eppendorf 5702 centrifuge was used to assist in phase separation. The membrane separator module was similar to those manufactured by Zaiput Flow Technologies. The solutions for the continuous membrane-based separation were pumped using either the KDS 100 Legacy Syringe (radioactive experiments) or the Harvard Apparatus PHD 2000 Programmable and Infusion syringe pumps (non-radioactive experiments). For batch experiments, the phase mixing was performed using an IKA ROCKER 3D digital shaker.
The continuous liquid-liquid extraction in flow (LLEF) was performed using the apparatus depicted in
Earlier reports indicated that zirconium can be extracted from its acidic solutions into an organic phase containing trioctylphosphine oxide (TOPO).35 However, the preliminary batch experiments containing the equimolar solutions of 0.01 M ZrCl4 and YCl3 in 37% HCl produced a 3-phase mixture.
We began by testing whether the phase separation in this extraction system can be improved under the LLEF conditions.
Starting with the conditions optimized for Ti/Sc extraction (0.2 μm membrane, 0.002″ (0.051 mm) diaphragm, and 0.05/0.15 mL/min aq/org flow rate) we found that extensive breakthrough of the third phase occurred at both 0.2 μm and 0.1 μm membrane pore size (Table 6, entry 1). Lowering the concentrations of ZrCl4 and YCl3 to 0.001 M and then to 0.0005 M did not result in any improvement in the phase separation either (Table 6, entries 3-4). Only by lowering the concentration of HCl from 12 M to 6 M can a complete phase separation be achieved (Table 6, entry 4).
Unable to use the literature conditions for LLEF of Zr, we turned to 9/1, (v/v) guaiacol/anisole mixture, which performed extremely well for the phase separation and LLEF of Ti. Complete phase separation occurred in 37% HCl at both 0.001 M and 0.01 M ZrCl4 and YCl3 in a wide range of flow rates (Table 6, entries 5-6).
The extraction efficiencies of Zr from 0.01 M ZrCl4 solution in 37% HCl, also containing 0.01 M YCl3 were investigated at different flow rates using the guaiacol/anisole, 9/1 (v/v) mixtures (
The extraction of 89Zr from its solution in 37% HCl, also containing 0.01 M YCl3 was explored at low flow rates (0.033/0.166 mL/min, aq/org) using the guaiacol/anisole, 9/1 (v/v) mixtures. (
Separation of Gallium and Zinc
Materials
Anisole (99%, ReagentPlus), hydrochloric acid (37%, ACS reagent), zinc chloride (≥98%), sulfuric acid (95.0-98.0%), dibutyl ether, butyl methyl ether, tetrahydropyran, hexyl methyl ether, α,α,α-trifluorotoluene, and toluene, were purchased from Sigma Aldrich. Diethyl ether, diisopropyl ether (≥99%), and high purity hydrochloric acid (37%, “Ultrapur”) were purchased from Merck. Heptane (99.7%) and 1,2-dichloroethane were purchased from VWR Chemicals. All purchased chemicals were used without further purification. Zinc foil (99.9%) was purchased from Alfa Aesar.
Pall PTFE membranes were used for all experiments (47 mm diameter, 0.1/0.2/0.5 μm pore size, polypropylene support). Perfluoroalkoxy alkane (PFA) diaphragms (0.001″/0.002″/0.005″ (0.0254/0.0508/0.1270 mm)) were purchased from McMaster Carr. All PFA tubing ( 1/16″ (1.5875 mm) OD, 0.03″ (0.762 mm) ID) was purchased from Idex Health and Science. PTFE static mixers were purchased from Stamixco. The 15 mL plastic centrifuge tubes with screw caps were purchased from VWR.
Combined Radiogallium (66Ga, 67Ga, 68Ga) and 65Zn Production and Purification
For all experiments, the cyclotron target material (zinc) was used at its natural abundance level.
Production: These radionuclides were produced simultaneously, by proton bombardment of stacked Zn and Cu foils. The incident 16.5 MeV proton beam would first encounter a 250 μm thick, 831 mg Zn foil before entering a 500 μm thick, 327 mg Cu foil. Incident energy on the Cu foil was calculated to appx. 12.8 MeV, making the 500 μm foil a thick target (range in Cu only 370 μm). The foils were irradiated for 160 minutes at 10 μA resulting in an integrated current of 26.2 μAh. The irradiated Zn foil, containing gallium radioisotopes was dissolved in a small amount of 3 M or 6 M hydrochloric acid and then added to either the 7 molal (m) or 1 M stock solution of ZnCl2 also prepared in 3 M or 6 M hydrochloric acid.
65Zn Purification: The irradiated Cu foil (327 mg, containing 5.6 MBq of 65Zn) was dissolved in 1.7 mL of concentrated HNO3 at 60° C. The deep blue solution was evaporated to dryness at 150° C. using vigorous Ar flow. The blue solid was re-dissolved in 2.5 mL 1 M HCl, and loaded onto TK200 resin (3 g). The resin was first eluted with 1 M HCl, which removed all the copper (a total of 14 mL), and then with water, which eluted the zinc (a total of 25 mL). The fractions containing the highest amount of 65Zn were collected, the solution was evaporated to dryness, and added to either the 7 molal (m) or 1 M stock solution of ZnCl2 prepared as described above. The resulting solution, containing 100-300 kBq of 65Zn and radiogallium (present mostly as 67Ga) and simulating a cyclotron-irradiated liquid target mixture was used as the aqueous phase for the LLE.
Instrumentation and Methods
Gallium and zinc were quantified by measuring radioactivities from 67Ga, 68Ga, and 65Zn radioisotopes using the CRC-55tR, CII (Capintec, Inc) dose calibrator and Princeton Gammatech LGC 5 and Ortec GMX 35195-P gamma spectrometers.
An Eppendorf 5702 centrifuge was used to assist in phase separation. The membrane separator module was similar to those manufactured by Zaiput Flow Technologies. The solutions for the continuous membrane-based separation were pumped using either the KDS 100 Legacy Syringe (radioactive experiments) or the Harvard Apparatus PHD 2000 Programmable and Infusion syringe pumps (non-radioactive experiments). For batch experiments, the phase mixing was performed using an IKA ROCKER 3D digital shaker.
For the phase separation studies, a centrifuge tube was charged with 1.3 mL of the 7 m ZnCl2-3 M HCl, or 7 m ZnCl2-6 M HCl solution and various amounts of organic phase were added. The mixture was shaken for 30 minutes and centrifuged at 4000 rpm to separate the phases.
For the batch LLE of gallium, a centrifuge tube was charged with 1.3 mL of the 7 m ZnCl2-3 M HCl, or 7 m ZnCl2-6 M HCl solution, also containing 67Ga, 68Ga, and 65Zn radioisotopes, and various amounts of organic phase were added. The mixture was shaken for 30 minutes and centrifuged at 4000 rpm to separate the phases.
The continuous liquid-liquid extraction and phase separation in flow was performed using a membrane-based separator with a PFA diaphragm for integrated pressure control. A flow schematic of the apparatus is depicted in
The earlier work established that dialkyl ethers, and in particular diethyl ether, efficiently and selectively extracted gallium from 5-6 M hydrochloric acid solutions in batch.68-71 Since dialkyl ethers are generally non-toxic, readily available low boiling point liquids, we decided to evaluate this class of compounds for further development in LLE and membrane-based separation of gallium from zinc. Tetrahydropyran (THP) diethyl (Et2O), diisopropyl (iPr2O), dibutyl (Bu2O), butyl methyl (BuOMe), and hexyl methyl (HexOMe) ethers were chosen as the extractants. The preliminary experiments showed that the presence of concentrated (7 m) ZnCl2 dramatically influenced the phase equilibrium. A single phase was observed by mixing equal volumes of diethyl ether, and a concentrated solution of ZnCl2 prepared in 6 M HCl. Lowering the concentration of HCl to 5, and then to 4 M still produced a single phase. At 3.5 M HCl two phases finally separated but extensive migration of aqueous into the organic phase was observed (Table 7, entry 8, Et2O/aq=6.22, (v/v)). Lowering the concentration of HCl further led to a gradual decrease in the ratio Et2O/aq, (v/v), ie a decrease in the migration of aqueous into the organic phase (Table 7, entries 2-4). Unexpectedly, this trend was opposite to what one observed when ZnCl2 was not present.69 The extraction efficiency of Ga and Ga/Zn selectivity were also disappointing.
68Zn
68Ga
The interfacial migration is a critical parameter which had to be minimized to prevent the contamination of aqueous phase with the organic phase, which would make the process incompatible with the implementation of continuous LLEF due to stringent organic-free requirements for the ZnCl2-based aqueous cyclotron solution targets. The significant interfacial migration would also lead to low interfacial tension, which might cause a phase breakthrough during membrane separation. Our strategy was to find a suitable interfacial tension modifier which provided for reliable phase separation with no or little interfacial migration while keeping good Ga extraction efficiency and high Ga/Zn selectivity. Given its low capacity to dissolve water72, toluene was initially chosen as an interfacial tension modifier for screening the phase separation in the series R1OR2/ZnCl2—HCl.
Having established a preferred interfacial tension modifier for phase separation, we investigated its performance for Ga/Zn extraction selectivity in the series of extractants R1OR2/ZnCl2—HCl at 6 M and 3 M HCl. Table 8 shows that TFT in combination with any of the six ethers in the screening set allowed for excellent gallium extraction efficiencies (entries 1-6). At 6 M HCl, the iPr2O and BuOMe were found to be the best performers extracting up to 95% Ga in batch. As expected, the Ga extraction efficiency decreased substantially in 3 M HCl. Nevertheless, a 2/1 (v/v) mixture of TFT and iPr2O was able to extract 77% of Ga and only 1% of Zn (entry 3). THP co-extracted the highest amount of Zn from 6 M and 3 M HCl.
iPr2O
Next, we translated the batch experiments into fully continuous flow experiments using the apparatus depicted in
The aqueous phase was formed by a 7 m ZnCl2/3 M HCl mixture and the organic phase consisted of a 2/1, (v/v) mixture of TFT used as an interfacial tension modifier and the series of ethers were used as the extractant. The aqueous and the organic phases were combined through a tee and mixed with two static mixers and mixing tubing. The aqueous phase was retained by the membrane, while the organic phase permeated through the membrane. For the membrane-based separator, we used optimized conditions established in the previously described work on 45Ti separation: flowrate org/aq, (mL/h)=45/15; 0.2 μm membrane pore size, 0.002″ (0.051 mm) diaphragm, and 108 cm mixing tube. The 68Ga was selectively extracted into the organic phase. The organic phase can then be either collected or directly re-routed into the second separation module, where 0.1 M HCl was used as the aqueous phase. After the second stage LLEF, 68Ga was selectively back-extracted into the aqueous phase together with the residual Zn.
Table 9 shows that gallium stripping was uniformly high (>90%) across the series. On the other hand, little selectivity was observed for Zn stripping, so that a single-stage LLEF/back-extraction sequence delivered the desired gallium solution in 0.1 M HCl containing as much as 10 mg/mL of zinc (the presence of greater than 10 mg/mL of Zn is indicated in
To decrease the amount of co-extracted Zn, two consecutive (two-stage) liquid-liquid extractions/back-extractions in flow were performed. In this experiment, two extraction/back-extraction modules can be combined. After the first stage of extraction/back-extraction, the mixture Ga containing residual zinc was acidified to 6 M HCl and subjected to the second stage of extraction/back-extraction under analogous conditions.
After the second stage, the 71% of original gallium was recovered in the final solution for radiolabeling, which also contained 100 μg/mL of Zn.
Even higher purity gallium solution can be obtained, if an extra step of liquid-liquid extraction of Zn using 8 M HCl is included (
Radioqallium production: Approximately 3.5 ml of target solution (5 M ZnCl2 in 3 M aq. HCl) was loaded in a GE PETtrace liquid target. The target chamber was made of niobium to limit corrosion. The target front foil was 250 μm niobium foil, bringing the proton energy down to 12.5 MeV from the nominal 16.5 MeV. The target was not pressurized, but left open to ensure no pressure buildup in the chamber. Bombardment was performed at a current of 5 μA for 6 minutes. After irradiation the produced Ga-68 was quantified by gamma spectroscopy on a 10% GeLi detector, calibrated using certified Eu-152 and Ba-133 sources. The produced activity at saturation was calculated to 204 MBq/μA.
Liquid-liquid extraction of radioqallium followed by irradiated target solution purification: 2.5 mL of the irradiated target solution was used as the aqueous phase for LLE. iPr2O/TFT (1/2) was used as the organic phase. The phases were separated using the membrane separator with a 0.2 μm PTFE/PP membrane and a 2 mil (0.0508 mm) diaphragm. The aqueous flow rate was 0.25 mL/min and the organic flow rate was 0.75 mL/min. Samples of the aqueous and organic phase after the LLE were collected and the activity of 66,67,68Ga (radiogallium) was measured with gamma spectroscopy. 57% of radiogallium was extracted into the organic phase. The aqueous phase was then passed through a C18 cartridge after the LLE, in order to remove trace organic solvent, and used directly for a second irradiation.
Radioqallium production from a re-used cyclotron target solution: Approximately 3.5 ml of a 1:1 target solution (5 M ZnCl2 in 3 M aq. HCl) and LLE-purified target solution from the first bombardment (Example 13) was loaded in a GE PETtrace liquid target. The target chamber was made of niobium to limit corrosion. The target front foil was 250 μm niobium foil, bringing the proton energy down to 12.5 MeV from the nominal 16.5 MeV. The target was not pressurized, but left open to ensure no pressure buildup in the chamber. Bombardment was performed at a current of 5 μA for 5 minutes. After ended irradiation the produced Ga-68 was quantified by gamma spectroscopy on a 10% GeLi detector, calibrated using certified Eu-152 and Ba-133 sources. The produced activity at saturation was calculated to 258 MBq/μA.
Liquid-liquid extraction of radiogallium from the re-used irradiated target solution: The LLE procedure described in Example 13 was used to extract Ga from 2.5 ml of the target solution from the second bombardment, which led to a radiogallium extraction of 62%.
Separation of Cu and Ni
Materials
Hydrochloric acid (37%, ACS reagent), zinc chloride (≥98%), trioctylphosphine oxide (≥98.5%), cobalt chloride, iron chloride, silver chloride, copper (II) chloride and toluene were purchased from Sigma Aldrich. High purity hydrochloric acid (37%, “Ultrapur”) was purchased from Merck. Heptane (99.7%) and hexane were purchased from VWR Chemicals. Nickel-64 (99.6% isotope-enriched) was purchased from Campro Scientific. All purchased chemicals were used without further purification.
Pall PTFE membranes were used for all experiments (47 mm diameter, 0.1/0.2/0.5 μm pore size, polypropylene support). Perfluoroalkoxy alkane (PFA) diaphragms (0.001″/0.002″/0.005″ (0.0254/0.0508/0.1270 mm)) were purchased from McMaster Carr. All PFA tubing ( 1/16″ (1.5875 mm) OD, 0.03″ (0.762 mm) ID) was purchased from Idex Health and Science. PTFE static mixers were purchased from Stamixco. The 15 mL plastic centrifuge tubes with screw caps were purchased from VWR.
Radionuclide Production and Separation
For all experiments, the cyclotron target material was Nickel-64 (99.6% isotope-enriched).
Copper-64 Production and Purification
Production: 64Cu was produced via the 64Ni(p,n)64Cu reaction using a water-cooled solid target mounted on the beam line of a PETtrace (GE Healthcare) cyclotron. The target consisted of approximately 80 mg of 64Ni metal (enriched to 99%) electroplated on a silver disk backing. The target was irradiated with a proton beam with an incident energy of 16.5 MeV and a beam current of 20 μA. After irradiation, the silver disk backing was transferred into a hot cell where it was treated with 30% HCl for 30 min. at 60° C., and then for 5 min. at 80° C., resulting in a clear green solution containing a mixture of 64CuCl2 and 64NiCl2.
Purification: The solution was decanted, diluted to 6M HCl and loaded onto a Dowex 1×8 (chloride form 200-400 mesh) column. The column was washed with 21 mL of 6 M HCl, and then with 33 mL of 5 M HCl. Finally, the column was eluted with 10 mL of 1 M HCl, which elutes 64Cu. Final evaporation from aqueous HCl yielded 2-6 GBq of 64Cu as 64CuCl2 with specific activity, 300-3000 TBq/mmol and radionuclidic purity of 99%.
Instrumentation and Methods
64Cu was quantified by measuring radioactivities from 64Cu radioisotopes using the CRC-55tR, CII (Capintec, Inc) dose calibrator and Princeton Gammatech LGC 5 and Ortec GMX 35195-P gamma spectrometers. Cu, Ni, Ag, Fe, Co and Zn were quantified by ICP.
An Eppendorf 5702 centrifuge was used to assist in phase separation. The membrane separator module was similar to those manufactured by Zaiput Flow Technologies. The solutions for the continuous membrane-based separation were pumped using either the KDS 100 Legacy Syringe (radioactive experiments) or the Harvard Apparatus PHD 2000 Programmable and Infusion syringe pumps (non-radioactive experiments). For batch experiments, the phase mixing was performed using an IKA ROCKER 3D digital shaker.
The continuous liquid-liquid extraction and phase separation in flow was performed using a membrane-based separator with a PFA diaphragm for integrated pressure control. A flow schematic of the apparatus is depicted in
A single stage liquid-liquid extraction in flow of copper and copper-64 was performed using the setup depicted in
64Cu in 6M
With 0.4 M TOPO in toluene up to 93% copper extraction was achieved with ratios of 1:3 and 1:5, while the extraction was 85% with a 1:1 ratio (
Extraction with 0.4 M TOPO is not sensitive to a change in interfacial tension modifier (toluene vs. hexane) and remains efficient (93%) when the concentration of TOPO is decreased to 0.1 M. No nickel is extracted to the organic phase.
Whilst the invention has been described with reference to preferred embodiments, it will be appreciated that various modifications are possible within the scope of the invention.
In this specification, unless expressly otherwise indicated, the word ‘or’ is used in the sense of an operator that returns a true value when either or both of the stated conditions is met, as opposed to the operator ‘exclusive or’ which requires that only one of the conditions is met. The word ‘comprising’ is used in the sense of ‘including’ rather than in to mean ‘consisting of’. All prior teachings acknowledged herein are hereby incorporated by reference. No acknowledgement of any prior published document herein should be taken to be an admission or representation that the teaching thereof was common general knowledge in Australia or elsewhere at the date hereof.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/EP2019/071156 | 8/6/2019 | WO | 00 |
Number | Date | Country | |
---|---|---|---|
62715471 | Aug 2018 | US | |
62772803 | Nov 2018 | US |