Endoscopic surgery within the head is a common procedure in neurological surgery and otolaryngology. It avoids large cranial incisions and can reduce the need brain retraction and prolonged wound healing. Endoscopic surgery within the head also provides improved illumination and visualization of the target tissues because the camera of the endoscope is brought directly to the surgical site.
During this type of surgery, there may be local trauma to the tissues in the surgical pathway, resulting from pressure or abrasion caused by the surgical tools. Generally these tissues are the nasal mucosa, turbinates, nasal septum, and sphenoid/frontal/maxillary sinus. When transorbital approaches are used, orbital and periorbital tissue are subject to local trauma. Surgical pathway trauma can add to the trauma of the procedure and prolong the patient's recovery time. Liquids in the surgical pathway, such as mucous, blood, and soiled irrigation fluid, tend to obscure the view of the endoscope. This leads to the constant need for irrigation and suction of the obstructing liquids. In some cases the endoscope may also have to be removed, cleaned and replaced multiple times during a single procedure. This disadvantage tends to increase the complexity and time requirements of the operation. In addition, with each movement of a surgical tool into or out of the surgical pathway, the surrounding tissues are put at risk of additional trauma. Improved devices and methods are therefore needed.
An access sheath is provided to protect the nasal passageway during endoscopic trans nasal or intra ocular surgery. The access sheath protects the entrance of the naris and sinus from the placement and manipulation of surgical tools both during the initial placement and during manipulation and exchange of surgical tools. The access sheath may provide a guide port to help direct surgical tools into position. In some designs the access sheath may splint the sinus open, to help open and provide access past the turbinate. The access sheath may also help to keep surgical tools and especially an endoscope freer from obscuring matter and secretions
The access sheath may be flexible for placement in a folded or rolled up configuration, have a hoop or expansion capability to fill and splint the passage, be partially or totally fluid tight to reduce ingress of secretions, and be lubricious for the unobstructed motion of fine surgical tools during delicate micro surgery.
U.S. Pat. No. 8,986,201 B2 discloses an access sheath, which may be made of elastomer, and has many of the performance features described above. However, elastomer has an inherent draw back. Flexible elastomers are inherently tacky and hence create sliding friction on surgical tools. In some designs this has required additives or a coating on the surface of access sheath to reduce friction. Still generally additives and coatings cannot always provide the surgeon with the feel of a surgical tool sliding against a wet mucus layer.
A nasal access sheath made of a hard plastic material is manufactured in a way to make it flexible, in one embodiment, by using a braided tube. A braid can be made from multiple fibers of plastic monofilament. Monofilaments can be made of rigid and tough plastic such as PET (Polyethylene terephthalate) or Nylon. Monofilaments can even be made from stainless steel. The fibers remain flexible because they have a small diameter, such as 0.08 mm to 0.5 mm. The fibers may have a round cross section, a relatively flat cross section, or elliptical cross section. A plurality of fibers can be braided into a braided tube. As one example, 64 fibers are counter wound in a two over and two under braid. The angle (pics or pitch) of the braided fibers can select the circular profile of the resulting braided tube or sleeve. A braided tube additionally is flexible due to the loose association of the braided fibers and their ability to slide relative to each other but still maintain the intended braided pattern.
In the drawings, the same reference number indicates the same element in each of the views.
The braided tube 20, especially if made of plastic, can be placed around a mandrel 22 that causes the braided tube 20 to expand to a specific diameter or shape. The braided tube 20 can then be heat set in an oven. Upon cooling the braided tube 20 will be permanently formed into the shape of the mandrel 22. Heat setting mandrels can be made of hollow or solid stainless steel, Delrin (acetal homopolymer resin). Mandrels 22 may be made of Teflon (fluoropolymer), especially if intended to coat the braided tube in a plastic/rubber/silicone dispersion. Heat setting can be done at a variety of temperatures and time, depending on the braided tube material and the heat capacity of the mandrel 22. A useful heat set temperature for nylon or PET braids is 120° C. and 150° C. for half an hour, followed by cooling to room temperature in ambient conditions or a quench in water.
The braided tube 20 may be coated with a semipermeable or more preferably an impermeable membrane. A less tacky coating or a harder urethane material, of durometer 50 A or harder, may be used. It can be applied in thicknesses of 0.1 mm to 0.3 mm. Due to the thinness and the flexibility of the braided tube, even when coated the braided tube may still remain flexible, foldable, and be able to elongate.
In an alternate embodiment a PET (Polyethylene Terephthalate) braided tube can be used with a uniform coat of a silicone dispersion (Nusil MED16-6606). Despite being silicone this combination provides a slick surface relative to surgical tools. This can be attributed to the mechanical nature of the structure. The braid surface provides a non-continuous, undulating surface where a full surface contact is replaced by a series of discrete contact points. Discrete contact points reduce the surface area of contact and hence reduce the friction between access sheath 40 and the surgical tool.
A single layer braided tube coated in an impermeable silicone, such as NUSIL silicone dispersion 6061, was found to be a good coating as it applies in a thin layer and despite being silicone (that has inherent tackiness) has little tack. The resulting friction as tested showed 50 grams of friction. This is about the same friction as hydrophilic coated devices when tested new. The lubricious nature of the material also does not degrade over time, unlike hydrophilic coatings.
Examples of lubricious coatings are: ceramic coatings. Slick-Sil coating (by Surface Solutions Group), Parylene coating, and hydrophilic coatings. Similar coatings can provide better friction reduction but may not feel as lubricious as mucous membrane.
Hydrophilic coating provide lubricity similar to mucous membranes, however they require wetting with water or saline to activate, and need periodic or continuous re-wetting to stay slick. Hydrophilic coatings also wear away after multiple abrasions with surgical tools and may not withstand the long procedure time of skull base neurosurgery. Examples of lubricious additives are: barium sulfate, powdered Teflon (fluoropolymer) glass fillers, and ceramic fillers. These can reduce the surface tack but also tend to provide a surgical tool feel that is different from mucous membranes.
The elastomeric coating on the internal surface of the braided tube 20 may be reduced. If the mandrel 22 is created from a semi flexible rubber or jacketed in rubber it results in a flexible surface. If the braided tube 20 is loaded over this surface and stretched to tightly engage the mandrel 22, the internal contact points of the braided tube embed slightly into the semi-flexible mandrel surface. This effectively masks the internal surface of the braid. A coating step fully coats the external braided tube while leaving the highest contact points of the internal braided tube uncoated and fully retaining the inherent lubricity of the hard plastic monofilament of the braided tube. A semi flexible surface may also be achieved by jacketing the mandrel in polyolefin shrink tubing. When the braid/jacket mandrel are heated for shaping, the shrink tubing softens and the braided tube will slightly embed. This similarly creates a braid/mandrel assembly that has a partially masked inner braid surface for a follow up coating step.
Referring now to
As is apparent from the description of
A monofilament material having round or flat fibers may be used as the braided tube material. A sheet or strip of braid material may also be used in place of a tube, with the sheet or strip formed into a tube during the manufacturing process. For example, a strip of braid material may be wrapped around the mandrel and formed into a tube via the heat setting. A coating may be applied to at least part of the heat set braided tube.
Internal contact points of the braided tube material may optionally be embedded into the mandrel surface, and a coating applied onto at least part of the heat set braided material. One or both ends of the heat set braided material may be folded or rolled to form an atraumatic end.
The access sheath 40 may be compressed or folded for low profile placement and high profile working position. In a simple case the access sheath can be folded by hand and slid into the sinus manually. Loading tools or kits may also be used.
The loading tool 52 has a flared proximal end 54 to allow the access sheath 40 to be easily inserted into the loading tool 52 and compacted into a small volume. A tubular body 62 is joined to the flared proximal end 54 of the loading tool 52. The tubular body 62, which may be straight or have a slight taper towards the distal end, is designed to fit into the nasal opening. The slot 56 along the side of the loading tool 52 provides a conduit for the access sheath 40 to be deployed and released.
The kit in
Turning to
As shown in
In use the instrument 80 is opened, as shown in
The designs of
Referring to
The elastomer may extend proximally 1-50 mm on the external surface of the access sheath 40. The external extended elastomeric surface 124 provides a user selectable section that may be cut to a desired length. When cut, a portion of the external extended elastomeric surface 124 remains on the access sheath 40 and provides an atraumatic distal rim. An external rim of elastomer may similarly be used on the proximal end of the access sheath. This provides a section at the proximal end that maintains the integrity of the braided tube and avoid fraying. The external rim of elastomer on the proximal rim, if used, may only be on the external surface so as not to create friction on surgical tools passing through the internal channel 45.
Referring to
As shown in
Turning to
The angle section 154 may allow the proximal end of the sheath 40 to be more easily stretched and/or deflected. This allows for more versatile movement of surgical tools extending through the sheath during surgery. As shown in
One method for placing a surgical access sheath includes loading a surgical access sheath into a loading tool, with the surgical access sheath comprising a braid material, and with the loading tool having a conical proximal end, and a tubular distal end, and a slot extending from the conical proximal end to the tubular distal end; inserting the loading tool into a body orifice; inserting a surgical tool into an internal channel of the surgical access sheath; moving the surgical tool to move the surgical access sheath out of the loading tool through the slot; and withdrawing the loading tool from the body orifice.
Another method for placing a surgical access sheath includes placing a surgical access sheath into a low profile delivery position, with the surgical access sheath comprising a braid material; loading the surgical access sheath into a tube; inserting the loading tool into a body orifice; moving a plunger into the tube to eject the surgical access sheath out of the tube and into the body orifice; and withdrawing the tube from the body orifice.
Another method for placing a surgical access sheath includes providing a scissor-like loading tool having a first jaw pivotally attached to a second jaw, with the first jaw having a channel and the second jaw movable at least partially into the channel when the scissor-like loading tool is in a closed position; providing a surgical access sheath comprising a braid material, with the access sheath having an internal channel; placing the access sheath around one of the first jaw and the second jaw, while the jaws are in an open position; moving the jaws into the closed position, to fold and grasp the surgical access sheath; inserting the jaws into a body orifice; opening the jaws to release the surgical access sheath; and removing the jaws from the body orifice.
A surgical kit includes an access sheath comprising a braid material; and a loading tool having a conical proximal end, and a tubular distal end, and a slot extending from the conical proximal end to the tubular distal end. The loading tool may comprise a flexible material. Typically the slot extends along an entire length of the loading tool. The slot may have a width equal to 25% to 45% of a minimum diameter of the tubular distal end. The access sheath may have a conical section, an angle section joined to the conical section, with the conical section having a central axis AF not parallel to a central axis AN of the angle section, a body section joined to the angle section, with the body section having a length at least twice the length of the angle section. Alternatively, the loading tool may have a tube and a plunger slidable into the tube, with the access sheath foldable or compressible to fit into the tube, and with the access sheath expandable when ejected from the tube by the plunger. If used, the tube may have an outside diameter of 5 to 20 mm, and it may be transparent or translucent. A scissor-like loading tool may also be used, with the scissor-like loading tool having a first jaw pivotally attached to a second jaw, with the first jaw having a channel and the second jaw movable at least partially into the channel when the scissor-like loading tool is in a closed position, to fold the access sheath.
Surgical instruments must be able to pass across the surface of the braided structure with minimal friction and without catching on the braid or the interstitial space between the braid elements. For the structure to be useful it must be comprised of a tight, dense, braid pattern with minimal interstitial space.
The shapes of the interstitial spaces may change during use as the braided material stretches or compresses. As used here, an interstitial space is defined as an equilateral quadrilateral with a dimension represented by the length of one side, in the as manufactured shape without any forces acting on the sheath.
The braid material of the access sheath may have interstitial spaces with a dimension of between 0.50 mm (0.020 in.) to 0.75 mm (0.030 in.), or 0.25 mm (0.010 in.) to 1.50 mm (0.060 in.)
Minimizing the interstitial space is significant because a small interstitial space between the braid elements creates a weave which is difficult for tools to snag on or poke through. A small interstitial space between the braid fibers also allows a coating of a filling or water proofing material, such as silicone (polysiloxane) to span the interstitial space. The larger the interstitial spaces, the more difficult it is for the coating to maintain coverage across the spaces. Spaces not fully covered, or filled in, with a coating create voids or pinholes which may tend to cause the tip of a tool to catch or snag in the braid material. Materials such as silicone may be used for the filling or water proofing material, including polymers, natural and synthetic rubbers, and elastomers generally. These may be applied to the braided material by dipping, spraying, rolling, etc. The filling material fills in the interstitial spaces, and may or may not also form a layer thickness above the braided material, on the inside and/or the outside surface of the sheath.
If the interstitial space is too small, it may create issues in manufacture and use of the sheath, as the braid sheath material in effect approaches becoming a solid material. Additionally, some interstitial space allows a coating to access and encapsulate the braid fibers. Without interstitial space the coating would only be able to form a surface film. A surface film may be prone to delamination with compression, stretch, or tool passage and create a higher friction coefficient. Additionally, there is a frictional benefit from having a surface that in not continuous. The discontinuous surface formed by the fibers of the mesh material reduce the surface area in contact with tools and directly reduces static and dynamic friction acting on the tools. For these reasons it is preferred to maintain an interstitial space of at least 0.25 mm (0.010 in).
The braided material used to form the sheath advantageously has one or more of the following properties:
A braided tube that can expand to approximately 4 times its initial low profile diameter. This means when braided, it is made in its lowest profile diameter but with the application of compression (by pushing the ends towards each other), the diameter of the braided tube can expand to 4 times the diameter.
The braided tube has an outside diameter of approximately 10 mm (⅜ in.) and can expand to approximately 40 mm (1½ in.).
The braided tube has as a full load braid pattern, meaning that any individual strand passes over two strands and then under two strands and then back over two strands, with this path continuous throughout the braided tube and for each individual fiber.
The braided tube is made from 0.25 mm (0.010 in.) diameter, round PET (polyethylene terephthalate) monofilament.
The braided tube is made from 96 individual fibers.
The braided tube, when placed over a 25 mm diameter (1 in.) rod has approximately one pic per millimeter or 24-26 pics per one inch, i.e., the count of crossing pairs of fibers along a straight longitudinal line on the braid surface per unit length. It is a measure of the braid density at that particular diameter.
An access sheath that has a high coefficient of friction may impede free movement of surgical instruments. The sheath should also advantageously maintain lubricity for the entire duration of the surgical procedure, which may be several hours, with surgical instruments extended and withdrawn through the sheath multiple times. Lubricity refers to the friction occurring when an object (such as surgical tool) slide across the material of the sheath, i.e., how slippery the material is.
An access sheath without inherent lubricity requires use of an external surgical lubricant, or using water or saline solution as a lubricant. However, this requires repeated or continual application of the surgical lubricant or flushing of water or saline, to maintain consistent lubricity. Hence, using an external lubricant introduces an additional step to the surgical procedure, which slows the surgical procedure, while also potentially obscuring visualization of the surgical field.
A hydrophilic coating may optionally be applied to the surface of the access sheath. However, this requires wetting the device to activate the coating. Hydrophilic coatings may also decrease in lubricity over the course of a long procedure with large numbers of instrument passes.
The present sheath may be provided as a composite structure of a braided material and water proofing material. The water proofing material may be silicone. The present composite structure sheath can provide long lasting lubricity, in contrast a simple molded rubber access sheath. The present braided material composite sheath does not require the application of lubricants and maintains its performance throughout the length of the procedure.
The silicone (or equivalent material) may be applied to the braided material by dip coating. The silicone is provided to make the braided material water proof, i.e., in the sense of preventing water from passing through the braided material. The silicone is not needed to provide lubricity because the braided material itself provides lubricity. The silicone fills in the interstitial spaces between the fibers of the braided material. Advantageously, the silicone has a maximum layer thickness of 0.13 mm (0.0005 inches). Maximum layer thickness refers to the thickness of the silicone, not including the thickness of the braided material. Stated differently, advantageously the thickness of the silicone does not project above or below the fibers by more than 0.13 mm.
For example, if the braided material is made of fibers having a 0.25 mm (0.01 inch) diameter, the total thickness of the silicone is advantageously 0.25 mm to 0.51 mm. Excessive thickness of silicone (or equivalent material) decreases lubricity.
The composite sheath achieves its performance because it has a non-continuous contact surface. Rather, the surface is a series of distinct contact points due to the undulating surface texture created by the spaced apart fibers.
Instead of instruments contacting a flat surface of an elastomer, instruments have minimal contact area with the bumps formed by the crossing of round monofilament fibers. The sheath allows a high durometer and less tacky material (such as PET monofilaments) to still have the foldability, stretchability, and conformability, that would otherwise require an elastomer with a lubricant as described above.
A method of insertion, placement and deployment of a braided access sheath 166 is described below. This method uses a pulling force rather than a pushing force as the unique braided structure of the sheath will tend to resist pushing forces as the diameter of the sheath increases with pushing forces.
Once folded, the sheath 166 can be grasped and held in position with any standard medical tool 168 such as a pituitary forceps, a curved hemostat, a standard hemostat, or any medical instrument that can hold the folded sheath as shown in
An access sheath may protect the anatomy from multiple passes of instruments, provide a hood to keep the scope clean, and minimize the run-in of blood to the operating site.
The run-in of blood into the operating site comes primarily from the anatomical dissection performed in the proximal sinus cavity prior to the placement of the access sheath. This may include the creation of a nasal flap, removal of turbinates or bleeding from instrument passage. Hence, the ability of an access sheath to stop or minimize blood run-in is advantageous.
Relative to blood run-in, an access sheath having a thin wall is provided, so that it can be folded into a minimum compact profile during placement but then have sufficient memory, expansion range, and expansion strength to expand against the anatomy when deployed. Expanding outwardly or splinting against the sinus cavity maximizes the open visual surgical field allowing better exposure for the surgery as well as providing a tamponading force against any bleeding in the proximal sinus cavity.
The access sheath may be made of a composite coated braid structure having a wall thickness of 0.38 mm to 0.88 mm (0.015 to 0.035 in.), or 0.52 to 0.72 mm (0.020 to 0.029 in.).
Correspondingly the access sheath has an expansion strength of 100-200 or 170 to 200 grams force as measured with a 3 mm (0.38 in.) rod pressing into the sheath at the longitudinal and vertical center of the sheath for 6 mm (0.25 in.) displacement. Expansion strength refers to the amount of outward force the sheath can exert on surrounding tissue. For sheaths having a generally oval cross section, for example as shown in
A metric on the splinting efficiency of the access sheath is expansion strength per unit of wall thickness, expressed as:
expansion strength/wall thickness=T ratio in gm/mm.
For example, a sheath having an expansion strength of 190 gm and a 0.62 mm (0.025 in.) wall thickness would have a T ratio of 190 gm/0.62 mm=306 gm/mm]. Generally sheaths having a T ratio of about 200 to 400 may be used.
The sheath shown in
An alternative embodiment is made of a single layer of braid material, such that the wall thickness of the sheath is equal to the diameter of the braid fibers plus the thickness of the layer or coating (e.g., silicone) on the inside and on the outside of the braid fibers. However, sheaths may optionally be made with two or more layers of braid material, over entire or substantially the entire sheath, or over selected portions of the sheath. For example, a central area of the sheath may have a narrow reinforcement band with a second layer of braid material, to locally increase expansion strength. Similarly, a reinforcement band of a second layer of material may be provided at the distal and/or proximal end of the sheath.
Along with splinting the anatomy to provide improved surgical access, the expansion strength of the access sheath provides a tamponade against the sinus nasal cavity anatomy wall. This helps reduce seeping of blood flow that could result from minor abrasions during placement of the access sheath and/or existing bleeding injuries. Tamponade of blood flow then results in a cleaner surgical field as migrating blood flow along the surface and to the distal end of the access sheath is minimized. Post surgery the sheath may be left in place to continue to tamponading, and as an aid to healing and recovery while providing an open breathing passageway to better allow the patient to breath more freely.
Additionally, the expansion strength and associated anatomical fill, helps to maintain the access sheath in position. If the access sheath conforms to and slightly impinges on the anatomy, it then forms an interlocking shape with the anatomy which tends to resist migration of the sheath.
Thus, a novel surgical sheath and methods have been shown and described. Each of the features described in association with a specific embodiment may of course also be used in other embodiments described. Various changes and substitutions may of course be made without departing from the spirit and scope of the invention. The invention, therefore, should not be limited except by the following claims and their equivalents.
This application is a continuation-in-part of U.S. patent application Ser. No. 15/680,947 filed Aug. 18, 2017 and now pending, which is a continuation-in-part of U.S. patent application Ser. No. 15/340,718 filed Nov. 1, 2016, now U.S. Pat. No. 9,494,621, which is a continuation of U.S. patent application Ser. No. 14/626,184 filed Feb. 19, 2015 and now abandoned, which is a continuation of U.S. patent application Ser. No. 13/798,990 filed Mar. 13, 2013, now U.S. Pat. No. 8,986,201. U.S. patent application Ser. No. 15/680,947 claims priority to U.S. Provisional Patent Application No. 62/396,746 filed Sep. 19, 2016 and U.S. Provisional Patent Application No. 62/377,400 filed Aug. 19, 2016. Each of the applications listed above is incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
62396746 | Sep 2016 | US | |
62377400 | Aug 2016 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14626184 | Feb 2015 | US |
Child | 15340718 | US | |
Parent | 13798990 | Mar 2013 | US |
Child | 14626184 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 15680947 | Aug 2017 | US |
Child | 16211776 | US | |
Parent | 15340718 | Nov 2016 | US |
Child | 15680947 | US |