The invention generally relates to systems and methods for performing multiple precursor, neutral loss and product ion scans in a single ion trap.
The drive to miniaturize mass spectrometers has encouraged a wealth of unconventional methods of ionization, atmospheric pressure interfaces, vacuum systems, and mass analyzer combinations. Both continuous and discontinuous atmospheric pressure interfaces have been developed, allowing the coupling of ambient spray and plasma ionization methods with portable systems. The standard analyzer geometry has evolved from the 3D quadrupole ion trap (Paul trap), to cylindrical, rectilinear, linear, toroidal, and halo traps, as well as ion trap arrays, two-plate linear ion traps, wire ion traps, and other unusual devices.
However, the fundamental way in which mass analysis is performed in quadrupole ion traps has varied very little. This is surprising, especially given the clear constraints in terms of size, power, and complexity of miniature ion trap systems. Three unconventional mass scanning methodologies have appeared in the ion trap literature: digital ion trap frequency scanning, sinusoidal RF frequency scanning, and ac frequency scanning. Although digital technology is promising, especially with regards to high spectral resolution and scan speed, it requires a complete overhaul of existing instrumentation, and the power consumption of the technique is higher than for conventional methods. RF frequency scanning can similarly improve resolution and mass range but it can also reduce power consumption and hence instrument size.
AC frequency scanning techniques exist, but such methods are not particularly high performance. However, AC frequency scanning techniques offer increased instrument versatility while requiring virtually no instrument modifications. Among the unique capabilities made accessible by AC frequency scanning (also known as secular frequency scanning) are single analyzer precursor ion scans and neutral loss scans. These simple scans require simultaneous orthogonal excitation of precursor and product ions for fragmentation of a particular precursor ion in concert with the ejection and detection of a particular product ion. In the case of the precursor ion scan, the product ion m/z, and hence secular frequency under constant RF conditions, is fixed, whereas in the neutral loss scan the difference between precursor ion and product ion m/z is fixed, and with an added noise elimination scan, this requires a triple frequency scan.
The invention recognizes that both the precursor ion scan and the neutral loss scan suffer from low conversion of precursor ions to detected product ions using conventional scan rates (thousands of Dalton/charge per second). For example, in cases where each precursor ion is given ˜3 ms to fragment, typical estimated conversions are 5-10%, which implies that perhaps 90% of the precursor ions are left in the ion trap after a precursor ion scan. For the neutral loss scan, this is not the case because the precursor ions must be cleared from the ion trap during the scan to prevent artifact peaks.
The invention takes advantage of the inefficiency in fragmentation in the precursor scan, and utilizes that inefficiency in order to conduct multiple precursor ion scans on the same single ion population. That allows certain MS/MS permutations to be performed on a single ion population in a single ion trap. By performing multiple scans on the same ion population, the information obtained from those ions can be maximized, a particularly useful characteristic for resource-constrained ion traps with relatively low duty cycles and when sample size and/or access is highly limited. Exemplary combinations are multiple precursor ion scans, precursor ion scans followed by a neutral loss scan, precursor ion scans followed by product ion scans, and segmented neutral loss scans (i.e., different mass ranges being interrogated by different (or the same) neutral loss scans, which can be done at the same or different RF amplitudes), as well as simultaneous precursor and neutral loss scans.
In certain aspects, the invention provides systems including a mass spectrometer having a single ion trap, and a central processing unit (CPU), and storage coupled to the CPU for storing instructions that when executed by the CPU cause the system to apply at least one of the following ion scans to a single ion population in the single ion trap: multiple precursor ion scans, a plurality of segmented neutral loss scans, or multiple simultaneous neutral loss scans.
In other aspects, the invention provides methods for analyzing a single ion population that involve generating a single ion population that is transferred into a single ion trap of a mass spectrometer, and applying, via a CPU operably associated with the mass spectrometer, at least one of the following ion scans to the single ion population in the single ion trap: multiple precursor ion scans, a plurality of segmented neutral loss scans, or multiple simultaneous neutral loss scans.
In certain embodiments, the multiple precursor ion scans are applied sequentially to the single ion population. In other embodiments, the multiple precursor ion scans are applied simultaneously to the single ion population. In certain embodiments in which the ion scans are multiple precursor ion scans, the CPU causes the system to apply at least one additional scan to the single ion trap. The at least one additional scan may be a neutral loss scan. In such embodiments, the CPU may cause the system to apply the neutral loss scan simultaneously or sequentially with the multiple precursor ion scans. In other embodiments, the at least one additional scan is one or more product ion scans, applied after the precursor ion scans. In other embodiments, the at least one additional scan is a plurality of segmented neutral loss scans.
The mass spectrometer may be any type of mass spectrometer, such as a bench-top or miniature (portable) mass spectrometer. In certain embodiments, the mass spectrometer is a miniature mass spectrometer. In certain embodiments, the system further includes an ionization source, which may be any ionization source known in the art.
The single ion population may be generated from any type of sample. Exemplary samples include biological samples (e.g., human tissue or body fluids, such as oral fluids), agricultural samples, industrial samples, environmental samples, or combinations thereof.
In other aspects, systems and methods of the invention can be applied to multiple reaction monitoring (MRM). MRM may be performed in which multiple precursor scans are applied to a single ion population in a single ion trap under conditions in which the experiment is performed in a frequency (mass) range in which other ions do not occur and that subsequent experiments would then be possible.
In other aspects, the invention provides methods for analyzing a sample that involve generating a single ion population from a sample that is transferred into a single ion trap of a mass spectrometer; and applying, via a CPU operably associated with the mass spectrometer, at least one of the following ion scans to the single ion population in the single ion trap: multiple precursor ion scans, a plurality of segmented neutral loss scans, or multiple simultaneous neutral loss scans, thereby analyzing the sample. Exemplary samples include biological samples, agricultural samples, industrial samples, environmental samples, or combinations thereof.
The invention generally relates to systems and methods for performing for performing multiple precursor, neutral loss and product ion scans in a single ion trap. Methods of performing precursor ion scans as well as neutral loss scans in a single linear quadrupole ion trap have recently been described and demonstrated. The invention generally relates to methodology for performing permutations of MS/MS scan modes, that is, ordered combinations of precursor, product, and neutral loss scans, following a single ion injection event. Exemplary permutations include 1) multiple precursor ion scans, 2) precursor ion scans followed by a single neutral loss scan, 3) precursor ion scans followed by product ion scans, and 4) segmented neutral loss scans. In addition, the common product ion scan can be performed earlier in the sequence under certain conditions. Multiple precursor ion scans can be performed simultaneously as can precursor ion scans with a neutral loss scan.
The systems and methods can be used to analyze any type of sample or chemical, as described in more detail here. Certain exemplary compounds were used to illustrate the systems and methods of the invention. For such exemplary demonstrations, amphetamine, methamphetamine, 3,4-methylenedioxyamphetamine, 3,4-methylenedioxymethamphetamine, 3,4-methylenedioxyethylamphetamine, cocaine, noroxycodone, oxycodone, buphedrone HCl, and N-ethylcathinone were purchased from Cerilliant (Round Rock, Tex., USA). HPLC grade methanol was purchased from Fisher Scientific (Hampton, N.H., USA). Oral fluid samples were spiked with amphetamine standards and subsequently diluted ten-fold in 95:4.9:0.1 acetonitrile:water:formic acid.
The skilled artisan will appreciate that any ionization source and technique can be used with the systems and methods of the invention, as described in more detail herein. In exemplary embodiments, nanoelectrospray ionization using a 1.5 kV potential was utilized. Borosilicate glass capillaries (1.5 mm O.D., 0.86 mm I.D.) from Sutter Instrument Co. (Novato, Calif., USA) were pulled to 2 μm tip diameters using a Flaming/Brown micropipette puller (model P-97, Sutter Instrument Co.). The nanospray electrode holder (glass size 1.5 mm) was purchased from Warner Instruments (Hamden, Conn., USA) and was fitted with 0.127 mm diameter silver wire, part number 00303 (Alfa Aesar, Ward Hill, Mass.).
All scans were performed using a Finnigan LTQ linear ion trap mass spectrometer (San Jose, Calif., USA) modified previously to perform orthogonal excitation (Snyder et al., J. Am. Soc. Mass Spectrom. 10.1007/s13361-017-1707-y; and Snyder et al., Anal. Chem. 2017, the content of each of which is incorporated by reference herein in its entirety). The ion trap has dimensions x0=4.75 mm, y0=4 mm, and three axial sections of lengths 12, 37, and 12 mm. The RF frequency was tuned to 1.166 MHz. The RF amplitude was held constant throughout ionization, ion cooling, and mass scan segments by substituting the RF modulation signal between the RF detector board and RF amplifier with a low voltage DC pulse from an external function generator. All AC waveforms were generated by using two Keysight 33612A (Chicago, Ill., USA) arbitrary waveform generators. Inverse Mathieu q scans were generated in Matlab (Snyder et al., Rapid Commun. Mass Spectrom. 2016, 30, 2369-2378, the content of which is incorporated by reference herein in its entirety) exported as .csv files, and imported to the waveform generators.
Precursor ion scans were performed by applying to the y electrodes of the linear ion trap a low voltage (˜200 mVpp) swept frequency to generate an inverse Mathieu q scan for precursor ion excitation while simultaneously applying a higher voltage (˜600 mVpp) fixed frequency to the x electrodes at a particular product ion's secular frequency.
Similarly, neutral loss scans required three identical inverse Mathieu q scans with appropriate trigger delays. Trigger delays are described for example in Snyder et al. (Anal. Chem. 2017), the content of which is incorporated by reference herein in its entirety. A first frequency scan (˜200 mVpp) was used for precursor ion excitation, a second frequency scan (˜600 mVpp) with trigger delay was applied to reject remaining precursor ions subsequent to their excitation, and finally a third frequency sweep (˜600 mVpp) with a trigger delay larger than the artifact delay was used for product ion ejection. The fixed neutral loss selected was directly proportional to the time delay between the excitation and ejection sweeps. Scan rates for precursor ion scans as well as product ion scans ranged from 200 Da/s to 800 Da/s. Each scan was calibrated separately using a linear fit of m/z vs. time.
Permutation scans were performed by applying the appropriate waveforms back-to-back (or simultaneously in the case of multiple precursor and precursor plus neutral loss scans). Only one ion injection was used for each permutation and automatic gain control was turned off. Injection time was varied from 5 ms to 25 ms, depending on sample concentration (generally 1-10 ppm, viz. g/L). Each mass spectrum shown here is the average of 10 scans.
Precursor ion scans and neutral loss scans are possible in single quadrupole ion traps using double resonance excitation, that is, by simultaneously exciting a precursor ion and ejecting a particular product ion so that the detection of that product ion occurs during the unique time during which its precursor fragments. Unlike CID in beam-type instruments (e.g. sectors and triple quadrupoles), CID in ion traps requires a relatively long time to increase internal energies because 1) helium is used as the collision partner and 2) collision energies are quite small. Hence many collisions, and thus more time, are required for fragmentation in ion traps. For the precursor scans and neutral loss scans, the low fragmentation efficiency translates into relatively low sensitivity for conventional scan rates. However, precursor ion scans, if performed under low AC amplitude conditions, do not clear the ion trap and thus if only 10% of the precursor ions are converted to product ions, then the other 90% of the ions are left in the trap for reexamination. This characteristic makes available permutations of MS/MS scan modes.
Exemplary MS/MS permutations is shown in Table 1.
a Product ion scans can also be performed earlier in the sequence provided the masses of the fragments do not fall into ranges of interest in the other scan types
b Multiple reaction monitoring (MRM) experiments can also be done at the end of any permutation but they are not considered further as they are not scans.
Multiple precursor ion scans can be performed on the same ion population so long as the precursor ions are not ejected and fully fragmented. Precursor ion scans can also be followed by a single neutral loss scan. Because the neutral loss scan clears the ion trap with an ‘artifact rejection’ frequency sweep, no subsequent scans are possible using a single ion injection event. Any number of product ion scans can succeed precursor ion scans as well. Finally, although two neutral loss scans cannot interrogate the same mass range (for a single ion injection event), one can be used for a segmented neutral loss scan wherein different neutral loss scans are performed over different mass ranges. In each of the scans the AC amplitude is optimized while constant RF amplitude is used. presumably also at different RF amplitudes.
Multiple precursor ion scans are allowed because each precursor ion scan (at scan rates of hundreds of Daltons/charge per second) converts <10% of precursor ions to product ions. Permutations of precursor ion scans could be useful for scanning an analyte population for different molecular functionalities and for monitoring more than one class of molecules. An example of a double precursor ion scan is shown in
Precursor ion scans can be followed by a single neutral loss scan. Because the neutral loss scan clears the trap of ions, no other scans are subsequently possible. Nonetheless, like permutations of precursor ion scans, precursor ion scans followed by neutral loss scans may be useful for examining an ion population for different functional groups.
Product ion scans may follow precursor ion scans as well. A useful example of the utility of this scan mode is shown in
Because neutral loss scans clear the precursor ions from the ion trap via the ‘artifact rejection’ frequency sweep, no other scan modes may follow them. So although neutral loss scans may not be repeated in the same mass range, segmented neutral loss scans are allowed. These are similar to segmented full mass scans wherein different mass ranges are interrogated at dissimilar RF amplitudes to improve resolution and mass accuracy. Neutral loss scans can also be ‘segmented’ so that different mass ranges can be analyzed for differing neutral losses. Segmenting the scan allows better mass spectral resolution to be obtained as well as better fragmentation efficiency for higher mass ions. Moreover, often different classes of molecules will occupy different mass ranges so that multiple classes of molecules could be monitored with a single ion injection event (e.g. fatty acids and complex phospholipids in tissue).
One of the disadvantages of performing multiple discrete MS/MS scans in sequence is that insufficient ions may remain after the first scan for several reasons. It is possible that most of the precursor ions fragment in the first scan, or if enough collision energy is imparted to the precursors then they may collide with the orthogonal electrodes (y direction, in our case) and hence be lost before any other scans take place. In this case it is possible to perform simultaneous MS/MS scans. That is, one may perform multiple simultaneous precursor ion scans, or simultaneous precursor and neutral loss scans.
Finally, an example of a simultaneous neutral loss scan performed on a mixture of amphetamine, methamphetamine, 3,4-methylenedioxyamphetamine, and 3,4-methylenedioxymethamphetamine is shown in
The final experiment performed in this work was translating the MS/MS scan modes to a complex mixture. Oral fluid was chosen as an appropriate sample, as it has previously been examined for illicit drugs by swab touch spray tandem mass spectrometry.58 In this work amphetamine standards were spiked into the oral fluid at concentrations ranging from 1 ppm to 100 ppm and subsequently it was diluted ten-fold in 50:4.9:0.1 acetonitrile:water:formic acid to improve nanospray performance. A full scan of the nanosprayed solution of 100 ppb (final concentration, 1 ppm in oral fluid) is shown in
An inverse Mathieu q scan is described in U.S. application Ser. No. 15/789,688, the content of which is incorporated by reference herein in its entirety. An inverse Mathieu q scan operates using a method of secular frequency scanning in which mass-to-charge is linear with time. This approach contrasts with linear frequency sweeping that requires a complex nonlinear mass calibration procedure. In the current approach, mass scans are forced to be linear with time by scanning the frequency of a supplementary alternating current (supplementary AC) so that there is an inverse relationship between an ejected ion's Mathieu q parameter and time. Excellent mass spectral linearity is observed using the inverse Mathieu q scan. The rf amplitude is shown to control both the scan range and the scan rate, whereas the AC amplitude and scan rate influence the mass resolution. The scan rate depends linearly on the rf amplitude, a unique feature of this scan. Although changes in either rf or AC amplitude affect the positions of peaks in time, they do not change the mass calibration procedure since this only requires a simple linear fit of m/z vs time. The inverse Mathieu q scan offers a significant increase in mass range and power savings while maintaining access to linearity, paving the way for a mass spectrometer based completely on AC waveforms for ion isolation, ion activation, and ion ejection.
Methods of scanning ions out of quadrupole ion traps for external detection are generally derived from the Mathieu parameters au and qu, which describe the stability of ions in quadrupolar fields with dimensions u. For the linear ion trap with quadrupole potentials in x and y,
q
x
=−q
y=8zeV0-p/Ω2(x02+y02)m (1)
a
x
=−a
y=16zeU/Ω2(x02+y02)m (2)
where z is the integer charge of the ion, e is the elementary charge, U is the DC potential between the rods, V0-p is the zero-to-peak amplitude of the quadrupolar radiofrequency (rf) trapping potential, Ω is the angular rf frequency, x0 and y0 are the half distances between the rods in those respective dimensions, and m is the mass of the ion. When the dimensions in x and y are identical (x0=y0), 2r02 can be substituted for (x02+y02). Solving for m/z, the following is obtained:
m/z=4V0-p/qxΩ2r02 (3)
m/z=8U/axΩ2r02 (4)
Ion traps are generally operated without DC potentials (au=U=0) so that all ions occupy the q axis of the Mathieu stability diagram. In the boundary ejection method, first demonstrated in the 3D trap and in the linear ion trap, the rf amplitude is increased so that ions are ejected when their trajectories become unstable at q=0.908, giving a mass spectrum, i.e. a plot of intensity vs m/z since m/z and rf amplitude (i.e. time) are linearly related.
The basis for an inverse Mathieu q scan is derived from the nature of the Mathieu parameter qu (eq. 3). In order to scan linearly with m/z at constant rf frequency and amplitude, the qu value of the m/z value being excited should be scanned inversely with time t so that
q
u
=k/(t−j) (5)
where k and j are constants determined from the scan parameters. In the mode of operation demonstrated here, the maximum and minimum qu values (qmax and qmin), which determine the m/z range in the scan, are specified by the user. Because the inverse function does not intersect the q axis (e.g. qu=1/t), the parameter j is used for translation so that the first q value is qmax. This assumes a scan from high q to low q, which will tend to give better resolution and sensitivity due to the ion frequency shifts mentioned above.
The parameters j and k are calculated from the scan parameters,
j=q
min
Δt/(qmin−qmax) (6)
k=−q
max
j (7)
where Δt is the scan time. Operation in Mathieu q space gives advantages: 1) the waveform frequencies depend only on the rf frequency, not on the rf amplitude or the size or geometry of the device, which implies that the waveform only has to be recalculated if the rf frequency changes (alternatively, the rf amplitude can compensate for any drift in rf frequency), and 2) the mass range and scan rate are controlled by the rf amplitude, mitigating the need for recalculating the waveform in order to change either parameter. It is important to note that we purposely begin with an array of qu values instead of m/z values for these very reasons.
Once an array of Mathieu qu values is chosen, they are converted to secular frequencies, which proceeds first through the calculation of the Mathieu βu parameter,
a conversion that can be done by using the algorithm described in Snyder et al. (Rapid Commun. Mass Spectrom. 2016, 30, 1190), the content of which is incorporated by reference herein in its entirety. The final step is to convert Mathieu βu values to secular frequencies (eqns. 9, 10) to give applied AC frequency vs time. Each ion has a set of secular frequencies,
ωu,n=|2n+βu|Ω/2 −∞<n<∞ (9)
where n is an integer, amongst which is the primary resonance frequency, the fundamental secular frequency,
ωu,0=βuΩ/2 (10)
This conversion gives an array of frequencies for implementation into a custom waveform calculated in a mathematics suite (e.g. Matlab).
Prior work used a logarithmic sweep of the AC frequency for secular frequency scanning, but, as described here, the relationship between secular frequency and m/z is not logarithmic, resulting in very high mass errors during mass calibration.
In theory, once the Mathieu qu parameters are converted to secular frequencies, a waveform is obtained. However, this waveform should not be used for secular frequency scanning due to the jagged edges observed throughout the waveform (i.e. phase discontinuities). In the mass spectra, this is observed as periodic spikes in the baseline intensities. Instead, in order to perform a smooth frequency scan, a new parameter Φ is introduced. This corresponds to the phase of the sinusoid at every time step (e.g. the ith phase in the waveform array, where i is an integer from 0 to v*Δt−1). Instead of scanning the frequency of the waveform, the phase of the sinusoid is instead scanned in order to maintain a continuous phase relationship. The relationship between ordinary (i.e. not angular) frequency f and phase Φ is:
f(t)=(½π)(dΦ/dt)(t) (11)
so that
Φ(t)=Φ(0)+2π∫0f(τ)dτ (12)
where variable τ has been substituted for time tin order to prevent confusion between the integration limit t and the time variable in the integrand. Thus, the phase of the sine wave at a given time t can be obtained by integrating the function that describes the frequency of the waveform as a function of time, which was previously calculated.
We begin with the phase of the waveform set equal to zero:
Φ(0)=0(t=0) (13)
The phase is then incremented according to eqns. 14 and 15, which accumulates (integrates) the frequency of the sinusoid, so that
Δ=ωu,0/v (14)
Φ(i+1)=Φ(i)+Δ (15)
where v is the sampling rate of the waveform generator. Note that ωu,0 is the angular secular frequency (2*π*fu,0, where fu,0 is the ordinary secular frequency in Hz) in units of radians/sec. Thus, sweeping through phase Φ (
Because the relationship between secular frequency and time is approximately an inverse function, the phase will be swept according to the integral of an inverse function, which is a logarithmic function. However, because the relationship between secular frequency and m/z is only approximately an inverse relationship, the phase Φ will deviate from the log function and thus cannot be described analytically (due to eq. 8).
Any ion trap known in the art can be used in systems of the invention. Exemplary ion traps include a hyperbolic ion trap (e.g., U.S. Pat. No. 5,644,131, the content of which is incorporated by reference herein in its entirety), a cylindrical ion trap (e.g., Bonner et al., International Journal of Mass Spectrometry and Ion Physics, 24(3):255-269, 1977, the content of which is incorporated by reference herein in its entirety), a linear ion trap (Hagar, Rapid Communications in Mass Spectrometry, 16(6):512-526, 2002, the content of which is incorporated by reference herein in its entirety), and a rectilinear ion trap (U.S. Pat. No. 6,838,666, the content of which is incorporated by reference herein in its entirety).
Any mass spectrometer (e.g., bench-top mass spectrometer of miniature mass spectrometer) may be used in systems of the invention and in certain embodiments the mass spectrometer is a miniature mass spectrometer. An exemplary miniature mass spectrometer is described, for example in Gao et al. (Anal. Chem. 2008, 80, 7198-7205.), the content of which is incorporated by reference herein in its entirety. In comparison with the pumping system used for lab-scale instruments with thousands of watts of power, miniature mass spectrometers generally have smaller pumping systems, such as a 18 W pumping system with only a 5 L/min (0.3 m3/hr) diaphragm pump and a 11 L/s turbo pump for the system described in Gao et al. Other exemplary miniature mass spectrometers are described for example in Gao et al. (Anal. Chem., 2008, 80, 7198-7205.), Hou et al. (Anal. Chem., 2011, 83, 1857-1861.), and Sokol et al. (Int. J. Mass Spectrom., 2011, 306, 187-195), the content of each of which is incorporated herein by reference in its entirety.
In certain embodiments, the systems of the invention include an ionizing source, which can be any type of ionizing source known in the art. Exemplary mass spectrometry techniques that utilize ionization sources at atmospheric pressure for mass spectrometry include paper spray ionization (ionization using wetted porous material, Ouyang et al., U.S. patent application publication number 2012/0119079), electrospray ionization (ESI; Fenn et al., Science, 1989, 246, 64-71; and Yamashita et al., J. Phys. Chem., 1984, 88, 4451-4459.); atmospheric pressure ionization (APCI; Carroll et al., Anal. Chem. 1975, 47, 2369-2373); and atmospheric pressure matrix assisted laser desorption ionization (AP-MALDI; Laiko et al. Anal. Chem., 2000, 72, 652-657; and Tanaka et al. Rapid Commun. Mass Spectrom., 1988, 2, 151-153,). The content of each of these references is incorporated by reference herein in its entirety.
Exemplary mass spectrometry techniques that utilize direct ambient ionization/sampling methods include desorption electrospray ionization (DESI; Takats et al., Science, 2004, 306, 471-473, and U.S. Pat. No. 7,335,897); direct analysis in real time (DART; Cody et al., Anal. Chem., 2005, 77, 2297-2302); atmospheric pressure dielectric barrier discharge Ionization (DBDI; Kogelschatz, Plasma Chemistry and Plasma Processing, 2003, 23, 1-46, and PCT international publication number WO 2009/102766), and electrospray-assisted laser desorption/ionization (ELDI; Shiea et al., J. Rapid Communications in Mass Spectrometry, 2005, 19, 3701-3704). The content of each of these references in incorporated by reference herein its entirety.
Processor 1086 which in one embodiment may be capable of real-time calculations (and in an alternative embodiment configured to perform calculations on a non-real-time basis and store the results of calculations for use later) can implement processes of various aspects described herein. Processor 1086 can be or include one or more device(s) for automatically operating on data, e.g., a central processing unit (CPU), microcontroller (MCU), desktop computer, laptop computer, mainframe computer, personal digital assistant, digital camera, cellular phone, smartphone, or any other device for processing data, managing data, or handling data, whether implemented with electrical, magnetic, optical, biological components, or otherwise. The phrase “communicatively connected” includes any type of connection, wired or wireless, for communicating data between devices or processors. These devices or processors can be located in physical proximity or not. For example, subsystems such as peripheral system 1020, user interface system 1030, and data storage system 1040 are shown separately from the data processing system 1086 but can be stored completely or partially within the data processing system 1086.
The peripheral system 1020 can include one or more devices configured to provide digital content records to the processor 1086. For example, the peripheral system 1020 can include digital still cameras, digital video cameras, cellular phones, or other data processors. The processor 1086, upon receipt of digital content records from a device in the peripheral system 1020, can store such digital content records in the data storage system 1040.
The user interface system 1030 can include a mouse, a keyboard, another computer (e.g., a tablet) connected, e.g., via a network or a null-modem cable, or any device or combination of devices from which data is input to the processor 1086. The user interface system 1030 also can include a display device, a processor-accessible memory, or any device or combination of devices to which data is output by the processor 1086. The user interface system 1030 and the data storage system 1040 can share a processor-accessible memory.
In various aspects, processor 1086 includes or is connected to communication interface 1015 that is coupled via network link 1016 (shown in phantom) to network 1050. For example, communication interface 1015 can include an integrated services digital network (ISDN) terminal adapter or a modem to communicate data via a telephone line; a network interface to communicate data via a local-area network (LAN), e.g., an Ethernet LAN, or wide-area network (WAN); or a radio to communicate data via a wireless link, e.g., WiFi or GSM. Communication interface 1015 sends and receives electrical, electromagnetic or optical signals that carry digital or analog data streams representing various types of information across network link 1016 to network 1050. Network link 1016 can be connected to network 1050 via a switch, gateway, hub, router, or other networking device.
Processor 1086 can send messages and receive data, including program code, through network 1050, network link 1016 and communication interface 1015. For example, a server can store requested code for an application program (e.g., a JAVA applet) on a tangible non-volatile computer-readable storage medium to which it is connected. The server can retrieve the code from the medium and transmit it through network 1050 to communication interface 1015. The received code can be executed by processor 1086 as it is received, or stored in data storage system 1040 for later execution.
Data storage system 1040 can include or be communicatively connected with one or more processor-accessible memories configured to store information. The memories can be, e.g., within a chassis or as parts of a distributed system. The phrase “processor-accessible memory” is intended to include any data storage device to or from which processor 1086 can transfer data (using appropriate components of peripheral system 1020), whether volatile or nonvolatile; removable or fixed; electronic, magnetic, optical, chemical, mechanical, or otherwise. Exemplary processor-accessible memories include but are not limited to: registers, floppy disks, hard disks, tapes, bar codes, Compact Discs, DVDs, read-only memories (ROM), Universal Serial Bus (USB) interface memory device, erasable programmable read-only memories (EPROM, EEPROM, or Flash), remotely accessible hard drives, and random-access memories (RAMs). One of the processor-accessible memories in the data storage system 1040 can be a tangible non-transitory computer-readable storage medium, i.e., a non-transitory device or article of manufacture that participates in storing instructions that can be provided to processor 1086 for execution.
In an example, data storage system 1040 includes code memory 1041, e.g., a RAM, and disk 1043, e.g., a tangible computer-readable rotational storage device such as a hard drive. Computer program instructions are read into code memory 1041 from disk 1043. Processor 1086 then executes one or more sequences of the computer program instructions loaded into code memory 1041, as a result performing process steps described herein. In this way, processor 1086 carries out a computer implemented process. For example, steps of methods described herein, blocks of the flowchart illustrations or block diagrams herein, and combinations of those, can be implemented by computer program instructions. Code memory 1041 can also store data, or can store only code.
Various aspects described herein may be embodied as systems or methods. Accordingly, various aspects herein may take the form of an entirely hardware aspect, an entirely software aspect (including firmware, resident software, micro-code, etc.), or an aspect combining software and hardware aspects. These aspects can all generally be referred to herein as a “service,” “circuit,” “circuitry,” “module,” or “system.”
Furthermore, various aspects herein may be embodied as computer program products including computer readable program code stored on a tangible non-transitory computer readable medium. Such a medium can be manufactured as is conventional for such articles, e.g., by pressing a CD-ROM. The program code includes computer program instructions that can be loaded into processor 1086 (and possibly also other processors) to cause functions, acts, or operational steps of various aspects herein to be performed by the processor 1086 (or other processor). Computer program code for carrying out operations for various aspects described herein may be written in any combination of one or more programming language(s), and can be loaded from disk 1043 into code memory 1041 for execution. The program code may execute, e.g., entirely on processor 1086, partly on processor 1086 and partly on a remote computer connected to network 1050, or entirely on the remote computer.
In certain embodiments, the systems of the invention can be operated with a Discontinuous Atmospheric Pressure Interface (DAPI). A DAPI is particularly useful when coupled to a miniature mass spectrometer, but can also be used with a standard bench-top mass spectrometer. Discontinuous atmospheric interfaces are described in Ouyang et al. (U.S. Pat. No. 8,304,718 and PCT application number PCT/US2008/065245), the content of each of which is incorporated by reference herein in its entirety.
A wide range of heterogeneous samples can be analyzed, such as biological samples, environmental samples (including, e.g., industrial samples and agricultural samples), and food/beverage product samples, etc.
Exemplary environmental samples include, but are not limited to, groundwater, surface water, saturated soil water, unsaturated soil water; industrialized processes such as waste water, cooling water; chemicals used in a process, chemical reactions in an industrial processes, and other systems that would involve leachate from waste sites; waste and water injection processes; liquids in or leak detection around storage tanks; discharge water from industrial facilities, water treatment plants or facilities; drainage and leachates from agricultural lands, drainage from urban land uses such as surface, subsurface, and sewer systems; waters from waste treatment technologies; and drainage from mineral extraction or other processes that extract natural resources such as oil production and in situ energy production.
Additionally exemplary environmental samples include, but certainly are not limited to, agricultural samples such as crop samples, such as grain and forage products, such as soybeans, wheat, and corn. Often, data on the constituents of the products, such as moisture, protein, oil, starch, amino acids, extractable starch, density, test weight, digestibility, cell wall content, and any other constituents or properties that are of commercial value is desired.
Exemplary biological samples include a human tissue or bodily fluid and may be collected in any clinically acceptable manner. A tissue is a mass of connected cells and/or extracellular matrix material, e.g. skin tissue, hair, nails, nasal passage tissue, CNS tissue, neural tissue, eye tissue, liver tissue, kidney tissue, placental tissue, mammary gland tissue, placental tissue, mammary gland tissue, gastrointestinal tissue, musculoskeletal tissue, genitourinary tissue, bone marrow, and the like, derived from, for example, a human or other mammal and includes the connecting material and the liquid material in association with the cells and/or tissues. A body fluid is a liquid material derived from, for example, a human or other mammal. Such body fluids include, but are not limited to, mucous, blood, plasma, serum, serum derivatives, bile, blood, maternal blood, phlegm, saliva, sputum, sweat, amniotic fluid, menstrual fluid, mammary fluid, peritoneal fluid, urine, semen, and cerebrospinal fluid (CSF), such as lumbar or ventricular CSF. A sample may also be a fine needle aspirate or biopsied tissue. A sample also may be media containing cells or biological material. A sample may also be a blood clot, for example, a blood clot that has been obtained from whole blood after the serum has been removed.
In one embodiment, the biological sample can be a blood sample, from which plasma or serum can be extracted. The blood can be obtained by standard phlebotomy procedures and then separated. Typical separation methods for preparing a plasma sample include centrifugation of the blood sample. For example, immediately following blood draw, protease inhibitors and/or anticoagulants can be added to the blood sample. The tube is then cooled and centrifuged, and can subsequently be placed on ice. The resultant sample is separated into the following components: a clear solution of blood plasma in the upper phase; the buffy coat, which is a thin layer of leukocytes mixed with platelets; and erythrocytes (red blood cells). Typically, 8.5 mL of whole blood will yield about 2.5-3.0 mL of plasma.
Blood serum is prepared in a very similar fashion. Venous blood is collected, followed by mixing of protease inhibitors and coagulant with the blood by inversion. The blood is allowed to clot by standing tubes vertically at room temperature. The blood is then centrifuged, wherein the resultant supernatant is the designated serum. The serum sample should subsequently be placed on ice.
Prior to analyzing a sample, the sample may be purified, for example, using filtration or centrifugation. These techniques can be used, for example, to remove particulates and chemical interference. Various filtration media for removal of particles includes filer paper, such as cellulose and membrane filters, such as regenerated cellulose, cellulose acetate, nylon, PTFE, polypropylene, polyester, polyethersulfone, polycarbonate, and polyvinylpyrolidone. Various filtration media for removal of particulates and matrix interferences includes functionalized membranes, such as ion exchange membranes and affinity membranes; SPE cartridges such as silica- and polymer-based cartridges; and SPE (solid phase extraction) disks, such as PTFE- and fiberglass-based. Some of these filters can be provided in a disk format for loosely placing in filter holdings/housings, others are provided within a disposable tip that can be placed on, for example, standard blood collection tubes, and still others are provided in the form of an array with wells for receiving pipetted samples. Another type of filter includes spin filters. Spin filters consist of polypropylene centrifuge tubes with cellulose acetate filter membranes and are used in conjunction with centrifugation to remove particulates from samples, such as serum and plasma samples, typically diluted in aqueous buffers.
Filtration is affected in part, by porosity values, such that larger porosities filter out only the larger particulates and smaller porosities filtering out both smaller and larger porosities. Typical porosity values for sample filtration are the 0.20 and 0.45 μm porosities. Samples containing colloidal material or a large amount of fine particulates, considerable pressure may be required to force the liquid sample through the filter. Accordingly, for samples such as soil extracts or wastewater, a pre-filter or depth filter bed (e.g. “2-in-1” filter) can be used and which is placed on top of the membrane to prevent plugging with samples containing these types of particulates.
In some cases, centrifugation without filters can be used to remove particulates, as is often done with urine samples. For example, the samples are centrifuged. The resultant supernatant is then removed and frozen.
After a sample has been obtained and purified, the sample can be analyzed to determine the concentration of one or more target analytes, such as elements within a blood plasma sample. With respect to the analysis of a blood plasma sample, there are many elements present in the plasma, such as proteins (e.g., Albumin), ions and metals (e.g., iron), vitamins, hormones, and other elements (e.g., bilirubin and uric acid). Any of these elements may be detected using methods of the invention. More particularly, methods of the invention can be used to detect molecules in a biological sample that are indicative of a disease state.
References and citations to other documents, such as patents, patent applications, patent publications, journals, books, papers, web contents, have been made throughout this disclosure. All such documents are hereby incorporated herein by reference in their entirety for all purposes.
Various modifications of the invention and many further embodiments thereof, in addition to those shown and described herein, will become apparent to those skilled in the art from the full contents of this document, including references to the scientific and patent literature cited herein. The subject matter herein contains important information, exemplification and guidance that can be adapted to the practice of this invention in its various embodiments and equivalents thereof.
The present application claims the benefit of and priority to U.S. provisional application Ser. No. 62/537,676, filed Jul. 27, 2017, the content of which is incorporated by reference herein in its entirety.
This invention was made with government support under NNX16AJ25G awarded by the National Aeronautics and Space Administration (NASA). The government has certain rights in the invention.
Number | Date | Country | |
---|---|---|---|
62537676 | Jul 2017 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 16047267 | Jul 2018 | US |
Child | 17159407 | US |