Lower-limb orthoses and exoskeletons have been developed with different structures and control strategies to assist users during their locomotion. Rehabilitation orthoses and exoskeletons are tools that aim to relieve the repetitive and physically tasking duties of the clinicians and therapists as well as improving the patient's recovery efficacy [1]. Traditional control methodologies for rehabilitation exoskeletons are designed to replicate normal kinematics (joint angles/velocities) and thus fall into the category of kinematic control. This approach is especially useful for providing assistance to individuals with spinal cord injury, who cannot contribute to the kinematic patterns of their own legs. Many exoskeletons have adopted this control paradigm to generate missing function for the user's lower limbs, e.g., [2] [7]. Even though these devices have shown promising results, their controllers force patients to follow pre-defined walking patterns, which may not be desirable for patients with some control of their lower limbs, such as stroke patients [8].
An alternative control category, kinetic control, enforces kinetic goals (e.g., torques or energy) instead of kinematic trajectories, which could provide more flexible gait training paradigms. Instead of constraining a person's motion in a pre-defined manner, kinetic control could provide a supportive environment to allow the person to relearn their own personal, preferred gait. However, exoskeletons typically utilize kinematic strategies that compensate for chronic deficits instead of enabling recovery of patient's normative gait [9]. Related to kinetic control, although not designed for physical rehabilitation, the BLEEX enhances the ability of an able-bodied person to carry extra heavy loads, using force control to minimize the user's interaction forces with the exoskeleton so the person does not feel the weight of the backpack [10]. However, minimizing interaction forces with the exoskeleton does not offload the body weight of the human user as needed in rehabilitation. Kinetic control methods that could enable greater flexibility for powered exoskeletons need to be developed for gait rehabilitation systems.
Exemplary embodiments of the present disclosure comprise a nonlinear control method of potential energy shaping [11] that is well suited for kinetic control of exoskeletons. By altering the potential energy of the human dynamics in the closed loop, bodyweight support (BWS) can be provided virtually through the actuators of powered lower-limb exoskeletons, allowing persons to train their walking motions naturally as well as freeing up therapists to make corrections. However, the changing contact conditions and degrees of underactuation encountered during human walking present significant challenges to consistently matching a desired potential energy for the human in closed loop. Accordingly, contact-invariant ways of matching desired dynamics disclosed herein enable exoskeletal BWS. Beneficial effects of this control methodology are disclosed herein with simulations and experiments of a powered orthosis during human walking [12], [13]. This feedback control strategy is fundamentally task-invariant, and its parameterization allows systematic adjustments for patient-specific therapy.
Conventional techniques provide weight support for patients through a torso or hip harness attached to an overhead lift. However, conventional BWS training systems are stationary (i.e., mounted to a ceiling or above a treadmill) and labor-intensive for clinicians (who must move the full weight of the limbs), which constrains patients' therapies to clinical environments and greatly reduces the efficiency and frequency of training. Although current powered exoskeletons address the challenge of mobility, their control systems follow pre-defined joint patterns to offload body-weight. These conventional exoskeleton control systems discourage patients from actively participating in the training process and do not allow patients to relearn their own natural walking gait. The proposed devices and control systems virtually provide BWS to reduce the gravitational forces experienced by the patient's center of mass and lower extremities during locomotion without pre-defined patterns, which can allow patients to relearn their natural walking gaits while training outside of clinical environments.
Exemplary embodiments disclosed herein demonstrate validation of the potential energy shaping approach, which can be implemented in a highly backdrivable, torque-controlled powered ankle foot orthosis (PAFO). The orthosis dynamics are modeled with contact constraints corresponding to heel contact, flat foot, toe contact, and no contact (i.e., swing). In addition, energy shaping control laws are derived for the ankle actuator to provide virtual BWS to a human subject. Torque profiles from simulations provide a reference for the PAFO hardware design.
The mechanical and electronic design of the PAFO is demonstrated to validate its closed-loop torque control capabilities for implementing the potential energy shaping controller. Testing with an able-bodied person with this PAFO is also presented, demonstrating the feasibility of the potential energy shaping approach for both positive and negative virtual body-weight augmentation.
Exemplary embodiments of the present disclosure include a method for controlling an orthosis device coupled to a leg of a person. In particular embodiments, the method comprises: measuring forces exerted on the orthosis device, where the orthosis device comprises an actuator, a plurality of support members and a plurality of sensors, and where the forces include a gravitational force exerted on the person; applying a torque to one of the plurality of support members by the actuator; and controlling the torque applied by the actuator, where the torque applied by the actuator is controlled to counteract the gravitational force exerted on the person; and torque applied by the actuator is not controlled to direct the orthosis device in a pre-determined pattern of motion.
In certain embodiments, the torque applied by the actuator is controlled by a closed-loop potential energy shaping control system. In particular embodiments, the torque applied by the actuator is kinetically controlled via a nonlinear control method. In some embodiments, controlling the torque to the actuator comprises varying the current to an electric motor. In specific embodiments, the electric motor is a permanent magnetic synchronous motor.
In certain embodiments, the orthosis device comprises a first support member placed under a foot of the person; the orthoses device comprises a second support member coupled to a shin of the person; and the actuator is configured to vary the angle between the first support member and the second support member. In particular embodiments, the orthoses device comprises a third support member coupled to the thigh of a user, and a second actuator configured to vary the angle between the third support member and the second support member. In some embodiments, measuring forces exerted on the orthosis device with a plurality of sensors comprises measuring a force in a ball area of the foot of the person via a first sensor in the first support member. In specific embodiments, measuring forces exerted on the orthosis device with a plurality of sensors comprises measuring a force in a heel area of the foot of the person via a second sensor in the first support member.
In certain embodiments, the actuator comprises an electric motor coupled to a gearbox and a first sprocket; the orthosis device comprises a second sprocket coupled to the first support member; and the orthosis device comprises a belt coupling the first sprocket and the second sprocket. Particular embodiments, further comprise measuring the torque applied to one of the plurality of support members by the actuator with a reaction torque sensor.
In some embodiments, the gearbox comprises a planetary gear transmission, and the reaction torque sensor measures the torque at an end of the planetary gear transmission. Specific embodiments further comprise measuring an angle of the second support member via an inertial measurement unit.
Certain embodiments include an orthosis device comprising: a first support member; a second support member; a plurality of sensors configured to measure forces exerted on the orthosis device, wherein the forces exerted on the orthosis device include a gravitational force; an actuator configured to apply a torque to the first or second support member; and a controller configured to control the torque applied by the actuator, wherein the torque is controlled to counteract the gravitational force exerted on the person and device and wherein the torque is not controlled to direct the orthosis device in a pre-determined pattern of motion. In particular embodiments, the controller comprises a closed-loop potential energy shaping control system. In some embodiments, the torque applied by the actuator is kinetically controlled via a nonlinear control method. In specific embodiments, the actuator comprises an electric motor, and wherein the controller is configured vary the current to the electric motor.
In certain embodiments, the first support member is configured to be placed under a foot of a person; the second support member is configured to be coupled to a shin of a person; and the actuator is configured to vary the angle between the first support member and the second support member. In particular embodiments, the plurality of sensors comprises a first sensor in the first support member configured to measure a force in a ball area of the foot of the person. Certain embodiments include a third support member configured to be coupled to the thigh of a person, and a second actuator configured to vary the angle betweent the third and second support members. In some embodiments, the plurality of sensors comprises a second sensor in the first support member configured to measure a force in a heel area of the foot of the person. Specific embodiments further comprise an inertial measurement unit configured to measure an angle of the second support member.
Certain embodiments further comprise an optical encoder configured to measure an angle between the first support member and the second support member. In particular embodiments, the actuator comprises an electric motor coupled to a gearbox and a first sprocket; the orthosis device comprises a second sprocket coupled to the first support member; and the orthosis device comprises a belt coupling the first sprocket and the second sprocket. Some embodiments further comprise a reaction torque sensor coupled to the gearbox. In specific embodiments, the gearbox comprises a planetary gear transmission, and wherein the reaction torque sensor is configured to measure torque at an end of the planetary gear transmission.
In the present disclosure, the term “coupled” is defined as connected, although not necessarily directly, and not necessarily mechanically.
The use of the word “a” or “an” when used in conjunction with the term “comprising” in the claims and/or the specification may mean “one,” but it is also consistent with the meaning of “one or more” or “at least one.” The term “about” means, in general, the stated value plus or minus 5%. The use of the term “or” in the claims is used to mean “and/or” unless explicitly indicated to refer to alternatives only or the alternative are mutually exclusive, although the disclosure supports a definition that refers to only alternatives and “and/or.”
The terms “comprise” (and any form of comprise, such as “comprises” and “comprising”), “have” (and any form of have, such as “has” and “having”), “include” (and any form of include, such as “includes” and “including”) and “contain” (and any form of contain, such as “contains” and “containing”) are open-ended linking verbs. As a result, a method or device that “comprises,” “has,” “includes” or “contains” one or more steps or elements, possesses those one or more steps or elements, but is not limited to possessing only those one or more elements. Likewise, a step of a method or an element of a device that “comprises,” “has,” “includes” or “contains” one or more features, possesses those one or more features, but is not limited to possessing only those one or more features. Furthermore, a device or structure that is configured in a certain way is configured in at least that way, but may also be configured in ways that are not listed.
Other objects, features and advantages of the present invention will become apparent from the following detailed description. It should be understood, however, that the detailed description and the specific examples, while indicating specific embodiments of the invention, are given by way of illustration only, since various changes and modifications within the spirit and scope of the invention will be apparent to those skilled in the art from this detailed description.
The patent or application file contains at least one drawing executed in color. Copies of this patent or patent application publication with color drawing(s) will be provided by the Office upon request and payment of the necessary fee.
Referring now to
The embodiment shown in
In the embodiment shown in
In addition, motor 110 is coupled a torque sensor 114 and torque sensor amplifier 116. Device 100 further comprises a motor controller 111 and an inertial measurement unit 119. In the embodiment of
In a particular embodiment, motor 110 may be configured as a permanent magnetic synchronous motor (PMSM) and gearbox 112 may be configured as a two stage planetary gear transmission (e.g. TPM 004X-031x-1x01-053B-W1-999, Wittenstein, Inc.). In specific embodiments, belt 140 may also be configured as a poly chain GT Carbon timing belt (e.g. 8MGT 720, Gates Industry, Inc.). Such a configuration can provide the desired torque outputs and can be used to move the actuator's weight closer to the user's center of mass, which can minimize the metabolic burden of the added weight [19].
In certain embodiments of device 100 with an overall transmission ratio of 43.71:1, an efficiency of 0.9, and a motor peak torque of 1.29 Nm, the maximum output torque that actuator 190 can achieve is 50 Nm. In addition, actuator 190 can provide 288 W peak power, which should be sufficient for normal human walking speed under the control algorithm disclosed herein. The combination of a high torque PMSM with a low ratio transmission can minimize backdrive torque and provide comfort to the user.
To achieve accurate torque control performance, motor 110 can be configured as a PMSM with distributed winding, which has sinusoidal back-EMF [20]. Such a configuration can be utilized to reduce the torque ripple and to produce smoother torque output. In a particular embodiment, controller 111 may be configured as a field oriented motor controller (e.g. GSol-Gut-35/100, Elmo Motion Control, Inc.). Such a configuration has a lower response time and torque ripple compared to trapezoidal motor control [21]. In certain embodiments, Hall sensors and a resolver can be coupled to motor 110 to obtain accurate position feedback for controller 111.
Given the requirements of the proposed algorithm, the user's gait phase, ankle angle, and absolute shank angle can be measured by force sensors 182 and 184 (e.g. FlexiForce A301, Tekscan, Inc.), encoder 135, and inertial measurement unit 119. Sensors 182 and 184 can be embedded into lower support member 180 which can be configured as an insole which is placed beneath the user's foot for detecting the phase of gait, e.g., stance vs. swing. In exemplary embodiments, force sensors 182 and 184 can be placed within the normal COP trajectory to provide precise readings, where force sensor 182 can be placed under the ball of the foot, while force sensor 184 can be placed under the heel.
In certain embodiments, lower support member 180 can be produced on a Connex 350 3D printer and made from a rubber-like polyjet photopolymer. In a particular embodiment, encoder 135 is configured as an optical incremental encoder (e.g., 2048 CPR, E6-2048-250-IE-S-H-D-3, US Digital, Inc.), which can be used to obtain the ankle angle. In addition, IMU 119 may be configured as model number 3DM-GX4-25-RS232-SK1, LORD MicroStrain, Inc. and installed on intermediate support member 150 to obtain the absolute shank angle (e.g. the absolute angle of intermediate support member 150. Particular embodiments may include a safety button that must be held continuously by the user to power device 100, i.e., an enable signal. The user could release the button to disable the device 100 at any point during use, e.g., if balance was the lost.
In the embodiment shown, torque sensor 114 is configured as a reaction torque sensor (e.g., TPM 004+, Wittenstein, Inc.) located between a casing for actuator 190 and intermediate support member 150 to measure the real torque output from actuator 190. Information from torque sensor 114 can be used to reduce the actuator torque error caused by the nonlinear transmission efficiency, the variable motor torque constant, and the backdrive torque.
As disclosed herein, exemplary embodiments of device 100 comprise a control system that provides bodyweight support (BWS) for a user. The use of device 100 can prove beneficial for people who have retained partial control of their motor skills, including for example those who have sustained a stroke and other neurological injury. Persons using device 100 are provided with BWS during conventional gait retraining to help them produce the coordinated muscle activities needed for walking. Device 100 can also be used to augment the performance of able-bodied users, e.g., offloading the weight of the user's body or backpack to minimize metabolic burden in soldiers.
Installing torque sensor 114 at the end of gearbox 112, instead of at the end of the belt 140, avoids additional mass at the ankle joint. By measuring the torque on the case of actuator 190, instead of the output shaft, a non-contact torque sensor can be avoided. This is beneficial since noncontact torque sensors are usually more expensive, larger in size, and heavier than the adopted reaction torque sensor. The system schematic is shown in
In order to provide accurate torque tracking, a torque control system was built with two closed loops as shown in the schematic of
T
e=(3P/2)·λm·iq, (23)
where P is the number of motor poles, λm an is the motor flux linkage, and iq is the active current in the d-q rotating reference frame [22].
One common methodology to realize torque control is by estimating the active current feedback, transmission ratio, and efficiency of the actuator. However, due to the nonlinear relationship between the motor winding current and the actuator output torque, an outer closed torque control loop was designed to eliminate the torque error. This torque controller tracked the reference torque commanded by the high-level control algorithm, i.e., the stance or swing controller, depending on the contact condition determined by the force sensors. In certain embodiments, controller 111 can be configured as a real time micro-controller (e.g. myRIO-1900, National Instrument, Inc. with a dual-core ARM microprocessor and a Xilinx FPGA) to implement the control algorithms.
Modeling Dynamics
Exemplary embodiments of device 100 can be controlled using only feedback local to the leg of a person using device 100. As shown in
Stance Leg Dynamics
The stance leg is modeled as a kinematic chain with respect to an inertial reference frame (IRF) defined at either the heel or toe, depending on the phase of the stance period (to be discussed later). The configuration of this leg is given by qst=(px, py, ϕ, θa, θk)T, where px and py are the Cartesian coordinates of the heel, Φ is the angle of the heel defined with respect to the vertical axis, and θa and θk are the angles of the ankle and knee, respectively. The Lagrangian dynamics can be derived in the form
M
st(qst){umlaut over (q)}st+Cst(qst,{dot over (q)}st){dot over (q)}st+Nst(qst)+Al(qst)Tλ=Bstust+Bhvst+Jst(qst)TF, (1)
where Mst is the inertia/mass matrix, Cst is the Coriolis/centrifugal matrix, Nst is the gravitational forces vector. Al∈c×5 is the constraint matrix defined as the gradient of the constraint functions, c is the number of contact constraints that may change during different contact conditions, and l∈{heel, flat, toe} indicates the contact configuration. The Lagrange multiplier λ is calculated using the method in [14]. Assuming the orthosis has actuation at the ankle joint, i.e., ust, the matrix Bst=(01×3, 1,0)T maps orthosis torque into the coordinate system. The interaction forces F=(Fx, Fy, Mz)T∈
3×1 between the hip of stance model and the swing thigh are composed of 3 parts: two linear forces and a moment in the sagittal plane [14]. Force vector F is mapped into the system's dynamics by the body Jacobian matrix Jst(qst)∈
3×5. The human input term vst=[vha, vhk]T∈
2×1 provides torques at the ankle and knee joints, i.e., vha and Vhk, which are mapped into the dynamical system through Bh=(02×3, I2×2)T
5×2. While designing the energy shaping controller, we make no assumptions about the human inputs or interaction forces.
During stance phase, the locomotion of the stance leg can be separated into three sub-phases: heel contact, flat foot, and toe contact, as depicted in
(1) Heel Contact: The heel is fixed to the ground as the only contact point, about which the stance leg rotates. The IRF is defined at the heel, yielding the constraint aheel(q)=0 and the constraint matrix Aheel=Δqaheel, where
a
heel:=(px,py)T⇒Aheel=(I2×2,02×3). (2)
(2) Flat Foot: At this configuration, the foot is flat on the ground slope, where ϕ is equal to the slope angle. The IRF is still defined at the heel, which yields the constraint aflat(q)=0 and the constraint matrix Aflat=Δqaflat, where
a
flat:=(px,py,ϕ−γ)T⇒Aflat=(I3×3,03×2) (3)
(3) Toe Contact: The toe contact condition begins when the Center of Pressure (COP), the point along the foot where the ground reaction force is imparted, reaches the toe. During this phase the toe is the only contact point, about which the stance leg rotates. The IRF is defined at this contact point to simplify the contact constraints. The coordinates of the heel are then defined with respect to the toe, yielding the constraint atoe(q)=0 and the constraint matrix Atoe=Δqatoe:
Swing Leg Dynamics
We choose the hip as a floating base for the swing leg's kinematic chain in
By deriving the equations of motion, we obtain
M
sw(qsw){umlaut over (q)}sw+Csw(qsw,{dot over (q)}sw){dot over (q)}sw+Nsw(qsw)=Bswusw+Bhusw−Jsw(qsw)TF, (5)
where Msw is the inertia/mass matrix, Csw is the Coriolis/centrifugal matrix, Nsw is the gravitational forces vector. The matrix Bsw=(01×4, 1)T maps the orthosis torque usw into the system. The vector F contains the interaction forces between the swing leg and hip (including human hip torques), and Jsw(qsw)∈3×5 maps F into the dynamics. The human input vector vsw=[vsk; vsa]T∈
2×1 contains human knee and ankle torques vsk and vsa, respectively, which are mapped into the coordinate system through Bh. There are no contact constraints during swing, i.e., Asw=0.
Equivalent Constrained Dynamics
In this section the equations of motion will be expressed as equivalent constrained dynamics in order to derive an underactuated control law that achieves the desired potential energy for a given contact condition [12], [13]. For the sake of generality the subscripts associated with specific contact conditions will be dropped. To begin, one can calculate the Lagrange multiplier λ, based on the results in [14], [15] as
λ={circumflex over (λ)}+{tilde over (λ)}u+
{circumflex over (λ)}=W({dot over (A)}{dot over (q)}−AM−1(C{dot over (q)}+N−Bhv)),
{tilde over (λ)}=WAM−1B,
W=(AM−1AT)−1. (6)
Plugging in λ and A, dynamics (1) become:
M
λ
q+C
λ
q+N
λ
=B
λ
u+B
hλ
v+J
λ
T
F, (7)
where
M
λ
=M,
C
λ
=[I−A
T
WAM
−1
]C+A
T
W{dot over (A)},
N
λ
=[I−A
T
WAM
−1
]N,
B
λ
=[I−A
T
WAM
−1
]B,
B
hλ
=[I−A
T
WAM
−1
]B
h,
J
λ
=J[I−A
T
WAM
−1]T. (8)
Given (7), the desired form of the equivalent constrained dynamics is given as
M
λ
{umlaut over (q)}−C
λ
{dot over (q)}+Ñ
λ
=B
hλ
v+J
λ
T
F, (9)
where
Ñ
λ
=[I−A
T
WAM]
−1
Ñ, (10)
given the desired gravitational forces vector Ñ that will be introduced later for each configuration. Based on the results in [12] and [13], the desired dynamics (9) can be achieved in closed loop if the following thatching condition is satisfied:
B
λ
⊥(Nλ−Ñλ)=0. (11)
The underactuated potential shaping control law is then
u=(BλTBλ)−1BλT(Nλ−Ñλ). (12)
During swing, we have Asw=0, hence (11) and 12) reduce to the classical matching condition and control law in [11].
We choose Ñst in (10) by replacing the gravitational constant in Nst with {tilde over (g)}<g for BWS and {tilde over (g)}>g for reverse BWS. The upper body segments are lumped into a single point mass at the hip in the stance dynamics. Assuming the stance knee is rigid enough to provide a lever arm from the ankle to the hip, ankle torques will directly map to forces along the stance leg, which can be used to shape the weight of that leg. We approximate a rigid stance knee by setting its angle to zero, i.e., θk=0, in the potential energy before deriving, the gravitational forces vector Nst that is used to evaluate the matching condition (11) and calculate the control law (12). As a consequence, the row corresponding to this DOF in Nst vanishes. We now prove for each contact condition that the weights of the stance shank, thigh, and hip links can be shaped by the orthosis ankle actuator.
1) Heel Contact: We decompose Mst into the submatrices M1∈2×2, M2∈
2×3, M3∈
3×2, and M4∈
3×3, to simplify the multiplication between Aheel and Mst−1 with blockwise inversion. Following the derivation in [12], [13]:
where V1=[V11, V12]=M2M4−1, V11∈2×1, V12∈
2×2 and P=[1,0]T. The subscript (k, z) indicates rows k through z of a matrix.
Let Ñλ1 be the desired (constrained) gravitational forces vector defined by (10). We choose the annihilator of Bλ1 as
where P⊥=[0, 1] is used as an annihilator for P. Plugging terms into (11), the matching condition holds if Ñst(3,3)=Nst(3,3), not shaping the heel orientation DOF. Therefore the control law uheel is defined by (12) after satisfying the matching condition with this assumption.
2) Flat Foot: At this configuration let M1∈3×3, M2∈
3×2, M3∈
2×3, M4∈
2×2. The same procedure yields
where V2=M2M4−1∈3×2. The exact flat-foot control law uflat is given by (12) after satisfying (11) with the annihilator
3) Toe Contact: At the toe contact configuration we decompose Mst as in the Flat Foot case to simplify the derivation. With the same procedure we obtain
where V3 and V4 are defined as
We choose the annihilator of Bλ3 as
Plugging in (17) and (18), the left-hand side of (11) is
B
λ3
⊥(Nλ3−Ñλ3)=V3(Nst(1,3)−Ñst(1,3)). (19)
The matching condition is not satisfied unless we assume Ñst(1,3)=Nst(1,3), i.e., not shaping the unactuated DOF ϕ (recall that px and py are constrained). The toe-contact controller utoe is then defined by (12) under this assumption.
For the swing leg, there ale no contact constraints defined in the dynamics so the matching condition simplifies. We replace g with {tilde over (g)} in Nsw to define the desired gravitational forces vector Nsw. Letting Bsw⊥=[I4×4, 04×1], we have Bsw⊥Bsw=0 rank (Bsw⊥)=4. The loft-band side of the matching condition (11) with A=0 is
B
sw
⊥(Nsw−Ñsw)=(Nsw(1,4)−Ñsw(1,4)),
The matching condition can be satisfied if first four rows of Nsw, which correspond to unactuated DOFs, are unshaped: Ñsw(1,4)=Nsw(1,4). Only links distal to the swing actuator can be shaped, i.e., the foot mass. This could assist individuals with weakened dorsiflexors (i.e., drop foot), Given Asw=0, the swing controller reduces from (12) to
u
sw=(BswTBsw)−1BswT(Nsw−Ñsw), (20)
where Ñsw=[Nsw(1,4)T,Ñsw(5)T]T.
In order to understand the torques required for the potential energy shaping strategy, we simulate it on humanlike biped model, i.e., combining the stance and swing legs together in
T=(BswT,01×3)Tul+(01×3,BswT)Tusw+vh,
v
h=[01×3,vha,vhk,vhh,vhsk,vhsa]T∈8×1, (21)
where l∈{heel, flat, toe} indicates the stance controller based on the contact condition, and vh is the vector of human inputs. The human torque for a single joint in vh is given by
v
hj
=−K
pj(θj−θjeq)−Kdj{dot over (θ)}j, (22)
where Kpj, Kdj, θjeq respectively correspond to the stiffness, viscosity, and equilibrium angle of joint j∈{a, k, h, sk, sa}.
Biped locomotion is modeled as a hybrid dynamical system which includes continuous and discrete dynamics. Impacts happen when the swing heel contacts the ground and when contact constraints change between the heel contact and flat foot configurations. Note that no impact occurs when switching between the flat foot and toe contact configurations, but the location of the IRF does change from heel to toe. Based on [16], the hybrid dynamics and impact maps during one step are computed in the following sequence:
Simulation Results
Average values from adult males [18] were chosen for the model parameters as in [12], [13] with trunk masses grouped at the hip. Following the same procedure presented in [12], the investigators first tuned the human impedance controller's gains to find a stable gait and then implemented the energy shaping control laws disclosed herein.
For notational purposes, 35% BWS corresponds to {tilde over (g)}=0.65·g, whereas 35% reverse BWS corresponds to {tilde over (g)}=1.35·g. The torque profiles for these conditions are shown in
Because the simulations in [13] show that the flat-foot controller uflat is equivalent to the other stance phase controllers (uheel and utoe), the investigators also performed a simulation using uflat during all of stance. Given the similar behavior in
Actuator Control System Testing
Due to the fact that the magnitude of the backdrive torque could be greater than the reference torque during swing, the user would feel resistance at the ankle during swing. Therefore, the backdrive torque should be compensated by the closed torque control loop to make the PAFO more transparent to the user. The investigators conducted an experiment to verify the effects of backdrive torque compensation, where they put the PAFO on a fixed frame and moved the ankle joint manually to mimic ankle motion. The experimental results are shown in
An experiment was also conducted to test the performance of torque tracking, where two reference torques, i.e., a 20 Nm step signal and a 35 Nm step signal, were given to approximate the situation when 20% or higher percentages of BWS are applied. Based on the results shown in
The investigators experimentally tested the control algorithm on an adult male subject walking with the PAFO on a treadmill, where the experiment setup is shown in
A safety harness was attached to the subject's torso to minimize the risk of falling. The subject was initially given time to find a natural gait with the unpowered exoskeleton on the treadmill. The subject was told not to use the handrails of the treadmill unless balance was lost. Once the subject started walking naturally with the orthosis, the investigators started the experiments and recorded data.
For this initial validation study the investigators conducted two experiments with limited weight augmentation: 20% BWS and 10% reverse BWS. The investigators stopped at 10% reverse BWS because the subject was already struggling to walk with that amount of weight addition. At the beginning of each experiment, the investigators asked the subject to stand straight to initialize the feedback of the PAFO. Then, the subject started walking on the treadmill at a constant speed of 0.536 m/s while holding the safety button to keep the PAFO system powered. The investigators recorded data for 15 steps for each condition once steady walking was observed. After the data was collected, the BWS condition was changed and the investigators ran the experiment again.
The experimental results for 20% BWS and 10% reverse BWS are shown in
From
The estimated mass parameters in Table I did not need to be accurate, since the BWS percentage could be adjusted easily in the program based on the preference of the subject. The controller did not require velocity feedback, and precise contact measurement was not needed. These experiments therefore demonstrate that the potential energy shaping control strategy can be implemented with relative ease. For both experiments, the subject was able to walk safely and comfortably with both positive and negative weight augmentation, motivating future studies with additional human subjects, including patients, to understand the effects of this weight augmentation.
Referring now to
The embodiment shown in
Referring now to
In summary, the devices and methods disclosed herein present an implementation of potential energy shaping for torque control on a powered ankle-foot orthosis. Exemplary embodiments include a potential energy shaping controller for the ankle actuator derived and simulated on a biped model. Based on the simulation results, a highly backdrivable powered ankle-foot orthosis can be used to implement and test this torque control strategy for a clinically relevant orthosis for persons including, for example, stroke patients. Initial experiments demonstrate that the disclosed PAFO control system can track the reference torque generated by the high-level control.
All of the devices, systems and/or methods disclosed and claimed herein can be made and executed without undue experimentation in light of the present disclosure. While the devices, systems and methods of this invention have been described in terms of particular embodiments, it will be apparent to those of skill in the art that variations may be applied to the devices, systems and/or methods in the steps or in the sequence of steps of the method described herein without departing from the concept, spirit and scope of the invention. All such similar substitutes and modifications apparent to those skilled in the art are deemed to be within the spirit, scope and concept of the invention as defined by the appended claims.
The contents of the following references are incorporated by reference herein:
This application claims priority to U.S. Provisional Patent Application Ser. No. 62/266,959 filed Dec. 14, 2015 and entitled “Torque Control Methods for Powered Orthosis”, the contents of which are incorporated herein by reference.
This invention was made with government support under HD080349 by the National Institute of Child Health and Human Development of the NIH. The government has certain rights in the invention.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US16/65558 | 12/8/2016 | WO | 00 |
Number | Date | Country | |
---|---|---|---|
62266959 | Dec 2015 | US |