This invention relates to a gas inlet system for an isotope ratio spectrometer such as an isotope ratio optical spectrometer (IROS) or an isotope ratio mass spectrometer (IRMS). The invention also relates to a method of coupling an analyte gas to an isotope ratio spectrometer.
Isotope ratio analysis is used to measure the relative abundance of isotopes (isotope ratio) in a gaseous sample for example, the stable isotopic composition of oxygen and carbon (18O/16O and 13C/12C) is an important proxy indicator of paleoenvironmental changes recorded in carbonate minerals deposited for example, as marine sediments.
Isotope ratio mass spectrometry (IRMS) is a well-established technique for such analysis. IRMS offers relatively high throughput (several minutes of analysis time per sample) as well as high precision isotope measurements. However high precision is accomplished at the expense of flexibility; IRMS instruments accept analytes in the form of a relatively limited number of gases which must be isotopically representative of the original sample. One of the challenges in IRMS is to ensure that fractionation (i.e., a shift in the relative quantities of isotopes between the original sample analyte and the mass spectrometer) is minimized or prevented.
The two most common types of IRMS instruments are continuous flow and dual inlet. In dual inlet IRMS, purified gas obtained from a sample is alternated rapidly with a standard gas of known isotopic composition by means of a system of valves, so that a number of comparison measurements are made of both gases. In continuous flow IRMS, sample preparation occurs immediately before introduction to the IRMS, and the purified gas produced by the sample is measured just once. The standard gas may be measured before and after the sample, or after a series of sample measurements.
Whilst continuous flow IRMS instruments can achieve higher sample throughput and are more convenient to use than dual inlet instruments, the yielded data are of lower precision. A general review of IRMS and gas inlet systems for these may be found in Brenna et al, Mass Spectrometry Reviews, 1997, 16, p. 227-258.
IRMS is not, however, without disadvantages. It is not compatible with condensable gases or a sticky molecule such as water. If a mixture of gases is applied to the analyser, there is the danger of interferences by reactions within the ion source. Thus, generally, chemical preparation of the sample is necessary to transfer the isotope of interest to a molecule that is more easily analyzed, and to separate the sample from other gas molecules. Typically, though, the required steps of chemical conversion are time consuming, and may compromise overall accuracy and throughput. Moreover, in general terms, IRMS instruments tend to be expensive, voluminous, heavy, confined to a laboratory, and usually in need of a skilled operator.
Isotope ratio optical (usually infrared) spectrometry (IROS) is a more recently developed technique for isotope ratio analysis. Here, photo absorption by H2O molecules is measured and the isotopologies of H2O are calculated by spectroscopy. IROS has a number of benefits over IRMS, such as ease of use, cost and potential field portability. It also permits direct analysis of water, where IRMS requires initial conversion e.g. to H2 or CO2, or equilibration with CO2, followed by analysis in gaseous form.
Relative to IRMS however, IROS typically offers a smaller dynamic range, poorer linearity, a larger measurement cell volume and pressure, such that more sample is required, and a higher pressure in the analyte cell.
Nevertheless, the gas load of an IR spectrometer may be very high. Thus, dilution of the sample to a relatively low concentration is often not a disadvantage. Indeed, given the limited dynamic range of an IROS instrument, significant dilution of the sample might be mandatory. The use of dry air as a carrier gas in IROS presents a further challenge relative to IRMS (where, typically, Helium is used as a carrier gas instead). The diffusion coefficient of CO2 in air is 0.16 cm2/sec, compared with a diffusion coefficient of about 0.7 cm2/sec for CO2 in Helium. Thus CO2 mixes much more slowly (by a factor greater than 4) in an IROS instrument than in an IRMS device, such that, in an IROS instrument, mixing is a more challenging issue.
For the sample quantities required for infrared laser spectroscopy, the major part of the substance has to be transferred into the laser cell, for example by the use of a carrier gas. However, the transfer of the substance into the laser cell without fractionation (i.e. modification of the isotope signature) is demanding. In IRMS, fractionation can be avoided either by transferring only a small part of the substance, thus not disturbing the chemical equilibrium, or transferring (and measuring) substantially the whole sample. Different parts of the transferred sample may have different isotope signatures but the whole time dependent peak is used, and differences in isotope fractionation cancel one another out during integration. The same principle applies even if, a constant fraction of the sample is split away, along as that fraction is time constant, that is, the proportion of sample which is split away remains constant over the measurement period.
Unfortunately, neither principle is possible for IROS. It is not possible to transfer only a small part of the substance for IROS because most or all of the total substance is required for analysis. Likewise transferring the whole sample and relying upon integration of the whole time dependent peak is not possible for IROS either. The signal in the laser cell has to remain as constant as possible. A conventional “peak shaped” transient signal typically has a relatively small start and end part and these cannot be evaluated. These initial and final parts of the transient signal often differ in isotope distribution from the rest and it is therefore necessary to have these integrated into the whole sample result.
When the majority of the sample is transferred out of the closed volume by a carrier gas, diffusion occurs at the boundary carrier gas/analyte, which is what leads to fractionation.
Against this background, the present invention seeks to provide a gas inlet system for an isotope ratio spectrometer that allows an improved supply of analyte gas to the spectrometric analyzer. Although the present invention seeks in particular to address at least some of the challenges presented by an IROS instrument, it is nevertheless also concerned with improving the manner of supply of analyte gas to an IRMS device as well.
According to a first aspect of the present invention, there is provided a gas inlet supply for an isotope ratio spectrometer.
The invention also extends to the combination of such a gas inlet supply with an optical spectrometer such as an infrared spectrometer, or with a mass spectrometer.
According to a further aspect of the present invention, there is provided a method of coupling an analyte gas to an Isotope Ratio Spectrometer.
The gas inlet supply and method of the invention employ an intermediate reservoir of variable volume, i.e. located intermediate between the supply of analyte gas and the spectrometer. A constant decrease of the reservoir volume can be used to generate a constant flow of gas to the spectrometer. This permits a wide variety of analyte gas sources (e.g. vial, bag, syringe, sampling tube, gas chromatograph, TOC analyzer, laser desorption, combustion or ablation cells, and so forth) to be coupled to the spectrometer via the variable volume reservoir, so that, for example, a constant flow and/or pressure of gas to the spectrometer can be achieved, mixing of the analyte gas with a carrier can be improved, dilution of the analyte by a carrier can be carried out in a controlled and quantifiable manner, analyte concentration can be determined, and so forth. By locating the variable volume reservoir between the analyte source and the spectrometer, it is also possible, in preferred embodiments, to buffer a pulsed analyte supply in the reservoir and then deliver a constant or quasi-constant flow of analyte from the reservoir to the spectrometer.
As described in more detail below, the gas inlet supply and method of the invention preferably employ an open split. In this way, the gas inlet system preferably operates at or close to atmospheric pressure. The reservoir of variable volume is preferably fluidly coupled to the open split. The variable volume reservoir and open split can be used together to adjust the gas flow, so that a signal intensity from the analyte is in the optimum measurement range of the spectrometer and that it can be matched with optional reference gas pulses.
These advantages are especially useful in the case of an isotope ratio optical spectrometer. In another aspect of the present invention, there is provided a method of cleaning a gas inlet system for an isotope ratio mass spectrometer comprising the steps of: filling a reservoir in the gas inlet system with a carrier gas, and expelling the carrier gas from the reservoir to the isotope ratio mass spectrometer. Preferably, the carrier gas is expelled to the isotope ratio mass spectrometer via one or more gas supply lines. Preferably the reservoir is a variable volume reservoir and the carrier gas fills it by expanding the volume thereof so as to draw the carrier gas into the reservoir.
Preferably the carrier gas is expelled by compressing the variable volume reservoir.
Further advantages and preferred arrangements will become apparent upon review of the following description and drawings, and from the accompanying claims.
The invention may be put into practice in a number of ways, some of which will now be described by way of example only and with reference to the accompanying drawings in which:
The isotope ratio analyzer 1 is also supplied with carrier gas from a carrier gas supply 2, along a carrier gas supply line. At an arbitrary point along the carrier gas supply line is located an open split 3. The first supply valve 7 is also selectively connectable to a second open split 8.
Open splits are a well-known solution for gas flow management. In general terms, open splits comprise a region, e.g. for mixing gases, open to the atmosphere. Gases to be analyzed emerge from a line (in the case of
An example of an open split design for isotope ratio mass spectrometry is shown in U.S. Pat. No. 5,424,539. A similar design is described in U.S. Pat. No. 5,661,038. Other examples may be seen in U.S. Pat. No. 7,928,369 and WO-A-2007/112,876. Gas inlet systems configured for auto dilution of samples using an open split are also known in the form of the Thermo Scientific Gasbench™ and Thermo Scientific ConFlo™ interfaces for isotope ratio mass spectrometry (www.thermoscientific.com).
However, a most preferred arrangement of open split is described and explained in our co-pending, as yet unpublished application nos. GB 1306806.9, GB 1306807.7 and GB 1306808.5, the contents of which are explicitly incorporated herein by reference in their entirety. In essence, the open split may be in the form of any suitable opening to atmosphere, e.g. a tube or capillary that is open to atmosphere. The open split 3 in the preferred arrangement of
A variety of different analyte gas supply arrangements may be employed, and some examples will be described in connection with
Linking all of the different preferred embodiments, however, is the variable volume reservoir 5. It is the variable volume reservoir 5 that permits the buffering of various analyte supplies, both continuous and pulsed, so as to permit the controlled delivery of analyte gas to the analyzer. The variable volume reservoir likewise permits auto dilution, cleaning of the gas lines, and various other advantageous procedures which will be detailed subsequently.
In general terms, the variable volume container 5 should ideally meet one or more of the following criteria:
For the purposes of obtaining an accurate measurement, it is preferable to adjust the reservoir volume continuously during a measurement. This provides a constant pressure at, and/or a flow into, the analyzer. Typically a flow through the analyzer will be held constant by controlling the variable reservoir such that the pressure is constant.
Three general approaches exist. Firstly, a sensor may be connected to the reservoir to allow a feedback control of the pressure. Secondly, after calibration or calculation, the volume may be simply continuously adjusted to give a constant flow. Thirdly, feedback from the analyzer may be employed to improve the flow rate and/or pressure. Such feedback may likewise be used to update a calibrated flow control.
Various methods and devices that enable such controlled delivery as a continuous operation are shown and discussed below.
Syringe Type Variable Volume Reservoir
In the first type of variable volume reservoir 5 (
In the preferred embodiment of a syringe type variable volume reservoir, a shaft seal (43) is provided between the inner and outer cylinders. The volume between both cylinders represents the reservoir of variable volume. The cylinders may be formed of various materials such as glass, stainless steel, other metals, polymer materials and so forth. The shaft seal may be formed form PTFE, PFA, Viton, or other sealing materials. Should both cylinders be machined and fitted to a suitable tolerance, the shaft seal might be omitted. Alternatively, one or both of the cylinders might be partially deformable (elastic) so as to form the sealing function.
The difference of the pressure inside the variable volume reservoir 5, and atmospheric pressure, generates a force. The force may be measured, and the pressure inside the variable volume reservoir 5 can therefore be determined, within the accuracy of the friction of the sealing.
In a most preferred embodiment (
The advantage of the arrangement described here is relatively low cost and that the volume within the variable volume reservoir 5 is, at any time, strictly proportional to the distance the inner cylinder is moving relative to the outer cylinder, so that it can be controlled very accurately (since, in this case, the volume is determined by the diameter of the outer cylinder—which is fixed—and the separation between the front face of the plunger and the base of the variable volume reservoir 5).
As still a further alternative (
Bellows Type Variable Volume Reservoir
(
It is preferable that decrease and increase of the volume in the variable volume reservoir 5 takes place through the application of a unidirectional force, again preferably by way of a linear actuator. As shown in
Again in a most preferred embodiment of a bellow type variable volume reservoir, a constant flow of gas into the analyzer may be achieved by application of a constant driving speed to the linear actuator. Moreover, welded metal may be employed to provide the bellows. Bellows cut from polymeric material might equally be employed. The advantage of the bellows type variable volume reservoir 5 is the potential extremely high leak tightness.
Bag Type Variable Volume Reservoir
As a variation of the bellows type variable volume reservoir, and as shown in
Although the bag type variable volume reservoir more particularly lends itself to pressure control, nevertheless a constant gas flow rate may be achieved, for example by control using a mass flow controller (MFC).
In any of the different embodiments described above (syringe or cylinder type, bellows type or bag type variable volume reservoir 5), some general parameters and characteristics are envisaged.
The mixing of the gas within the variable volume reservoir 5 should be optimized. Optimization is achieved by considering the notional connection of two points within the variable volume reservoir 5 by a curved line. The line is constrained to be inside the container at each point, as short as possible, and continuously either rising or falling, or having no slope with respect to the gravitational centre of the earth. The length of the line for any two points within the volume of the variable volume reservoir 5 should not exceed 150 mm. The line should also be smaller than the cube root of the volume of the variable volume reservoir 5, multiplied by a factor of 5 (or 10). The volume v within the variable volume reservoir 5 is defined by (Vmin+(Vmax−Vmin))*0,1<V<Vmax
With Vmin=minimum Volume, Vmax=maximum volume
In this configuration, the gas supply lines/capillaries and so forth are not taken into account.
Likewise, in common with each of the various embodiments of variable volume reservoirs 5, the exit of the reservoir 5 should preferably be at approximately half the height or at least not in the top or bottom 20% of the height. This arrangement minimises mixing effects caused by diffusion of the gases as diffusion is at a minimum rate in this middle section.
Another consideration to minimize mixing effects is the position of the gas exit from the variable volume reservoir 5. If different gases are filled into the volume subsequently, the different gases typically have a different specific weight. Thus, they will be “stacked” horizontally. In the following, the gases will mix by diffusion effects.
The degree of mixing of course increases with time. But it is also observed that mixing effects are less pronounced in the “middle” section. that is, neither at the top nor at the bottom of the variable volume reservoir 5 (here, the terms “top” and “bottom” are meant in the gravitational sense). Therefore, a preferred embodiment of the invention locates the gas entry/exit towards the “middle” of the height—and, at least, not in the top or bottom 20% of the height of the variable volume reservoir.
Also, it is preferred if the variable volume is “flat”. In the case of a syringe this means that it is preferred if the syringe is positioned horizontally rather than vertically.
Optimising the mixing of gas inside the variable volume reservoir 5 as detailed above, and locating the exit of the container neither at the top nor at the bottom of the variable volume reservoir 5 ensures that mixing inside the volume by diffusion is facilitated.
The currently most preferred arrangement for a variable volume reservoir is a glass syringe having a maximum volume of around or in excess of 100 ml, controlled by a stepper motor.
Having described a number of preferred embodiments for the variable volume reservoir 5, various different configurations of the gas inlet system embodying the present invention will now be described, referring once again, initially, to
I First Configuration: Use of Carrier Gas to Supply Analyte to Analyzer
In the first implementation of the gas inlet system of
(i) Analyte Collection
Analyte gas is supplied from the first source of analyte gas 9. The analyte gas supply from the first source of analyte gas 9 is entrained with carrier gas. The flow of carrier gas and entrained analyte is continuous in the embodiment of this section of the description.
Prior to filling the variable volume reservoir 5, the first source of analyte gas 9 is connected via the first supply valve 7 to the second open split 8. In this case, analyzer valve 4 is closed, and so is second supply valve 6, so that, with the variable valve reservoir 5 in an initial minimal volume configuration, there is no gas flow through the connecting gas lines between the analyzer valve 4, the first supply valve 7, the second supply valve 6 and the inlet/outlet from the variable volume reservoir 5.
Analyte collection may be achieved, in the gas inlet system embodying the present invention, in two different modes of operation, either by sampling the whole analyte for a certain time period, or by splitting away part of the analyte entrained in the continuous carrier gas supply from the first source of analyte gas 9 with a part flowing directly into the analyzer 1.
(a) Sampling the Whole Analyte
In the following description, a series of steps are described. The skilled reader will readily appreciate that, unless the context so requires, the order of the steps is not critical to the resultant configuration and effect. The skilled person will also recognise that, whatever the order of the steps, a delay, or no delay, may be present between one, some or all of the steps.
Referring again to
The resultant configuration of the gas inlet system of
The carrier gas supply 2 supplies a carrier gas flow which is larger than the intake of the analyzer 1 by an amount x. This means that, in the case where the analyzer valve 4 is closed, the entire amount of x, representing the difference between the analyzer gas intake and the carrier gas supply, will flow through the first open split 3 to atmosphere or to another device having a constant pressure. In the arrangement of
In consequence of this particular configuration, variable volume reservoir 5 advantageously is filled with the whole of the flow from the first source of analyte gas 9, plus also an amount of flow from the carrier gas supply 2, via the first open split 3. In this way, the whole sample is transferred into the variable volume container, along with carrier gas from the carrier gas supply 2.
(b) Splitting Away Part of the Analyte
As in the sampling of the whole analyte in (a) above, again with the gas inlet system in the initial configuration (variable volume reservoir at the minimum volume, valves 4, 6 and 7 closed), the analyzer valve 4 is opened. The first source of analyte gas 9 is connected to the variable volume reservoir 5. As previously, valve 6 is closed. Thus in this configuration, the gas inlet system is functionally as illustrated in
The variable volume reservoir 5 is then expanded at a rate which is this time lower than the gas flow from the first source of analyte gas 9. Any flow of gas from the first source of analyte gas 9 in excess of the uptake capacity of the variable volume reservoir 5 (on account of its relatively slow rate of expansion) then flows through the valve 4 and is sucked into the analyzer 1 along with carrier gas from the carrier gas supply 2.
If the difference between the flow into the variable volume reservoir 5 and the sample flow from the first source of analyte gas 9 is larger than the uptake of the analyzer 1, then any excess analyte gas is passed along the carrier gas supply line to the first open split 3. Although this particular implementation is not preferred, nonetheless it can be advantageous when the sample amount is known to be very high, since, in that case, even if the flow from the first source of analyte gas 9 is not known, the gas composition flowing into the analyzer 1 will be certain to be identical to the gas composition both from the first source of analyte gas 9 and at the variable volume reservoir 5.
More generally, the advantages of splitting away part of the analyte are:
For both applications, measuring the sample concentration during the sampling process can be useful. To permit this, however, it is desirable that both the volume of the variable volume reservoir 5, and the time derivative of that, and the flow of analyte gas from the first source of analyte gas 9 can be controlled sufficiently accurately. For this reason, the syringe type variable volume reservoir 5, as described above, provides particular advantages.
Moreover, it is desirable either to control the flow from the first source of analyte gas, for example by using a mass flow controller (mfc), a proportional valve, a volume flow controller or a similar device, or by measuring the flow from the first source of analyte gas 9 using a mass flow meter or the like.
Collection of analyte within the variable volume reservoir 5 can be terminated at any point between the minimum and maximum volumes of the variable volume reservoir. To do this, the following steps are taken. Firstly, the variable volume expansion of the reservoir 5 is stopped. Next, the first supply valve 7 is switched so that the first source of analyte gas 9 is once again connected to the second open split 8. The analyzer valve 4 is also then closed. The resultant configuration is shown in
(ii) Preparation for Measurement
Various optional steps may be taken in preparation for measurement of the analyte.
(a) Determination of Analyte Concentration
If the variation in the amount of sample is small and falls within the linearity range of the analyzer 1, then no further action is necessary and measurement may commence directly.
If the user knows the amount of analyte at least some extent, then the concentration within the variable volume reservoir 5 can be calculated from the known values of the relevant flows and volumes.
If the procedure for splitting away part of the analyte (I(i)(b) above) has been employed, then, knowing the concentration of the split away gas, and the flow rates, concentration within the variable volume reservoir can be calculated.
Otherwise, part of the gas that has been stored in the variable volume container (I(i) above) can be used to determine the concentration of analyte within the volume of the variable volume reservoir. To do this, starting with the valves in the configuration shown in
It is feasible to carry out the determination of concentration of the analyte within the variable volume reservoir without first closing the analyzer valve 4 This is the preferred mode of operation as no pressure peak is generated downstream of the analyser valve 4 after opening it. This way, it is ensured that no gas is lost through the open split 3. As a final alternative for determination of the concentration of analyte, it may be carried out during measurement of the analyte itself, rather than as a preliminary step.
(b) Auto Dilution: Split Away Part of the Sample
It may be that the concentration of analyte within the variable volume reservoir 5 once it has been collected is too high for the subsequent measurement. In this case, it is desirable to dilute the sample within the variable volume reservoir 5.
To achieve this, the first supply valve 7 is configured to connect the first source of analyte gas 9 to the second open split 8. The second supply valve 6 is closed so as to isolate the second source of analyte gas 11. The analyzer valve 4 is open. Thus the configuration of the gas inlet system 20 is as shown in
Next, the volume of the variable volume reservoir 5 is increased again to a desired amount. The speed of volume increase should be smaller than the difference between the uptake of the analyzer 1, and the carrier gas flow from the carrier gas supply 2. In that case, the variable volume reservoir 5 sucks in carrier gas from the carrier gas supply 2. Mixing takes place so that the resultant concentration of analyte within the variable volume reservoir 5 is lower at the end of the procedure than at the beginning.
(c) Pre-Adjust the Pressure Inside the Variable Volume Reservoir
In the idealized situation that there is no flow restriction between the variable volume reservoir 5 and the analyzer valve 4, the flow of gas at the analyzer valve 4 would always be equal to the time derivative of the volume of the variable volume reservoir 5. It is thus desirable that the restriction between the variable valve reservoir 5 and the analyzer valve 4 is as small as possible. Where the restriction is considered non-negligible, it is necessary to take into account the fact that the gas flow through the analyzer valve 4 depends upon the pressure inside the variable volume reservoir 5 and also upon the restriction. The pressure must first be built up. In mathematical terms:
Where p is the pressure within the variable volume reservoir 5, V is the volume within the variable volume reservoir 5, R is the restriction, jout represents the flow rate of gas through the analyzer valve 4; jres is the rate of change of volume, dV(t)/dt of the variable volume reservoir 5.
There is a steady state, in which p=jres*R. This steady state is characterized by a steady flow of gas through the analyzer valve 4 (jout). In order to reach this steady state as fast as possible, analyzer valve 4 and second supply valve 6 are closed, and first supply valve 7 is arranged so as to connect the first source of analyte gas 9 to the second open split 8. This is the configuration shown in
Next, the volume V within the variable volume reservoir 5 is decreased until p is reached. This can be calculated from the (known) volume of the variable volume container, and the known start pressure, which will be atmospheric pressure, because the variable volume reservoir has been connected to the first open split 3.
Finally, the analyzer valve 4 is opened so as to provide the configuration shown in
(d) Determining the Optimum Equilibrium Flow into the Analyzer
Prior to measurement, the optimum equilibrium flow into the analyzer is determined according to the following equation:
jres=janalyzer*(canalyzer/cres) Equation 2
Where jres represents dV(t)/dt of the variable volume container reservoir; janalyzer represents the flow of gas into the analyzer; canalyzer represents the desired concentration of analyte within the analyzer 1, and cres represents the determined or estimated concentration of analyte in the variable volume reservoir 5.
In this manner, using the variable volume reservoir, the dynamic range of the setup is significantly increased, because the speed of the volume change may be adjusted across a wide range. It has been measured and shown that, in this manner, a factor of at least 25 in the dynamic range can be achieved. This has to be multiplied by the dynamic range of the analyzer 1.
(iii) Measurement
Measurement of the analyte in the variable volume reservoir is commenced (again, the order or steps is arbitrary) with the gas inlet system 20 configured as shown in
As has been explained above, it is advantageous to have a steady, constant flow of analyte gas into the analyzer 1. Thus, the decrease of the volume of the variable volume reservoir should preferably be constant with time. Tests have shown that it is possible to generate a constant analyte flow into the analyzer 1 using the configuration of
At the commencement of measurement of the analyte, it is possible to adjust the analyte flow into the analyzer 1. A plot of analyte concentration in the analyzer 1 as a function of arbitrary time (horizontal axis) is shown in
(iv) Cleaning of the Gas Lines and the Variable Volume Reservoir
Typically, the minimum volume of the variable volume reservoir 5 is as close as possible to zero. Likewise, the volume of the various gas transport lines within the gas inlet system is preferably as close to zero as possible. On that basis, memory of remaining gas inside the lines and the variable volume reservoir can often be neglected.
If not, however, then following measurement (in particular, prior to filling the variable volume reservoir with fresh analyte for subsequent measurements), the gas supply lines and the variable volume reservoir 5 can be flushed with zero air (i.e., air free of CO2) using the following procedure.
Firstly, the second supply valve 6 is closed and the first supply valve 7 is configured to connect the first source of analyte gas 9 to the second open split 8. Analyzer valve 4 is open. The configuration of the gas inlet system is thus as shown in
The volume of the variable volume reservoir 5 is decreased to a minimum volume. Any gas remaining within the variable volume reservoir 5 thus leaves the system through the analyzer valve 4 and the first open split 3.
Next, the volume of the variable volume reservoir 5 is increased again. The speed of the volume increase of the reservoir 5 should be smaller than the difference between the analyzer uptake capacity and the carrier gas flow from the carrier gas supply 2. In this case, the variable volume reservoir 5 is then filled up with carrier gas from the carrier gas supply 2.
Once the variable volume reservoir has been filled with carrier gas, that carrier gas can be expelled again by cyclically repeating the above steps one or more times (i.e. flushing the carrier gas in the variable volume reservoir back out through the supply lines to the first open split 3).
(II) Sampling of a Limited Sampling Amount by Vacuum
For this type of sampling the first supply valve 7 is effectively closed relative to the variable volume reservoir 5 i.e. the first source of analyte gas 9 is connected to the open split 8 and therefore not to the variable volume reservoir 5.
In the alternative implementation of embodiments of the present invention, a second source of analyte gas 11 is supplied to the analyzer using the variable volume reservoir 5. In each of the following sections, the second source of analyte gas 11 is connected into the gas inlet system through a port 10. It is preferable that connection and disconnection between the second source of analyte gas 11 and the gas inlet system 20 is straightforward. The removable coupling described has a screw thread and is preferably a finger tight connector, as used in liquid chromatography, with a preferred thread style of either 10-32 coned, or ¼-28 coned, or 10-32 flat bottom or ¼-28 flat bottom. A Cajon port may be employed instead.
(i) Analyte Collection
Referring once again to
Functionally, the configuration is as shown in
The volume of the variable volume reservoir 5 is then increased. The second supply valve 6 is then closed again so that, functionally, the arrangement of
Depending upon the type of sampling container that represents the second source of analyte gas 11 (some examples of which will be described in further detail below), there will now be a (partial) vacuum inside the variable volume reservoir 5. Thus, the volume of the variable volume reservoir 5 is decreased again until the pressure inside it reaches atmospheric pressure. The amount of decrease necessary can either be calculated from the known volume of the reservoir 5 according to the following expression:
Where Vsc,min is the minimum volume of the sampling container (i.e. variable volume reservoir 5)—which may be zero—and V is the volume of the variable volume reservoir 5 following introduction of analyte. Alternatively, the pressure in the variable volume reservoir 5 can be measured and set to atmospheric pressure. Techniques for pressure measurement are set out above as part of the description of some preferred embodiments of variable volume reservoir 5.
(ii) Preparation for Measurement
(a) Determination of Analyte Concentration
As with the continuous carrier gas supply embodiments described above in Section I(ii), various optional steps may be taken in the preparation for measurement. Determination of analyte concentration is essentially as described in Section I(ii)(a) above. Either the variations of the amount of sample are small and fall within the linearity range of the analyzer 1, so that no actions are necessary, or the user may know the amount to some extent so that the concentration within the variable volume reservoir 5 can be calculated from the known flows and volumes, or part of the gas within the variable volume reservoir 5 may be used to determine the concentration of analyte within that reservoir 5. The technique is as described above in I(ii)(a): the analyzer valve 4 and the second supply valve 6 are closed and the first supply valve 7 is set to connect the first source of analyte gas 9 to the second open split 8. The volume within the variable volume reservoir 5 is then decreased to a known amount. The analyzer valve 4 is then opened and the concentration of the analyte is determined and integrated using the analyzer 1. Again, by analogy, it is possible to carry out this procedure with the analyzer valve 4 left open but that is not the preferred technique.
(b) Auto-Dilution: Splitting Away of Part of the Sample
The procedure for diluting the concentration of analyte within the variable volume reservoir 5 by splitting away part of the sample is identical to that described in Section I(ii)(b) above.
(c) Pre-Adjustment of the Pressure Inside the Variable Volume Reservoir
Likewise, the procedure here is exactly the same as in Section I(ii)(c) above.
(iii) Measurement
Measurement of the analyte within the analyzer 1 employs the same technique as described in I(iii) above.
(iv) Cleaning of the Gas Lines and the Volume
Again, similar considerations apply as with the use of a continuous analyte gas supply as described in Section I(iv) above. The variable volume reservoir 5 and gas supply lines can be flushed with zero air prior to commencement of analysis. The second supply valve 6 is closed and the first supply valve 7 is configured to connect the first source of analyte gas 9 with the second open split 8. The analyzer valve 4 is opened. The configuration of the gas inlet system configuration is thus as shown in
Next, the analyser valve 4 is closed and the second supply valve 6 is opened, so that the volume of the variable volume reservoir 5 is increased. This creates a vacuum between the first supply valve 7, second source of analyte gas 11, variable volume reservoir 5 and the analyser valve 4. The gas between the second supply valve 6 and the second source of analyte gas 11 is now distributed over the whole volume between the first supply valve 7, second source of analyte gas 11, variable volume reservoir 5 and the analyser valve 4. Hence the amount of gas between the second supply valve 6 and the second source of analyte gas 11 is significantly decreased. Next, the second supply valve 6 is closed so that the volume of the variable volume reservoir 5 is decreased to the minimum volume. The analyser valve 4 is opened again so that any gas between the first supply valve 7, second source of analyte gas 11, variable volume reservoir 5 and the analyser valve 4 is released through the open split 3.
The volume of the variable volume reservoir 5 is then increased again at a rate sufficiently slow enough to ensure that only carrier gas, rather than air, enters the volume between the first supply valve 7, second source of analyte gas 11, variable volume reservoir 5 and the analyser valve 4 via the open split 3. The rate at which the volume increases should be less than the difference between the rate of uptake by the analyser 1 and the rate of carrier gas flow from the carrier gas supply 2. At this stage the region between the first supply valve 7, second supply valve 6, variable volume reservoir 5 and the analyser valve 4 is filled with carrier gas and there is a vacuum between the second supply valve 6 and second source of analyte gas 11. To fill this region with carrier gas, the analyser valve 4 is closed and the volume of the variable volume reservoir 5 is decreased to generate an overpressure. This overpressure needs to be large enough so that once the second supply valve 6 is opened the pressure is still higher than atmospheric pressure. This procedure may be repeated several times.
III Optional Further Features of the Gas Inlet System
Further additional or alternative components may be employed. For example, chemical and/or cryo traps may be employed along the first and/or second analyte gas supply lines from the first and second sources of analyte gas 9,11, in order to remove unwanted gases from the analyte stream.
Although the arrangement of
In general terms, in addition to the variable volume reservoir 5 embodiments of the present invention may include one or more of the following components: at least one open split, a plurality of valves to allow isolation of the variable volume reservoir 5 from the open split or splits, at least one inlet port and, if there are several inlet ports, suitable valves to switch between them, a connection to the analyzer, and a connection to the carrier gas supply.
Analyte Gas Supply
Various methods for the delivery of analyte gas to the gas inlet system embodying the present invention are envisaged. This may be separated, generally, into specific arrangements for the delivery of a continuous flow of analyte gas (that is, suitable as a gas supply for the first source of analyte gas 9), and alternative arrangements for providing a peak or pulsed supply of analyte gas, representing the second source or analyte gas 11. Analyte gas supplies of the first type are shown in
For the provision of a continuous flow of gas from the first source of analyte gas 9, a vial 18 may be used. There a number of techniques by which analyte can be generated inside the gas phase of the vial 18. For example, a gaseous sample may be collected inside the closed volume of a vial 18, either because the analyte molecules are already in gaseous form, or by the reaction of the first gaseous molecule with another gaseous molecule either in the closed volume of the vial 18 or in a separate (subsequent) reaction cell. For example, hydrocarbon maybe decomposed into CO2 in the reaction cell.
Alternatively, the gaseous analyte may be provided through equilibration between a solid or liquid phase and a gaseous head space in the vial 18. This equilibration can be supported, for example, through heat, radiation, mechanical movement and so forth. Still further, a gaseous analyte gas supply may be generated through a chemical reaction of a solid or liquid with an added reagent and/or heat radiation, catalysis and so forth.
A syringe, (again not shown in
A vacuum is generated by opening the variable volume reservoir 5 so as to expand the volume thereof. This causes the gaseous sample within the vial 18 to be sucked along the gas supply line, into the first supply valve 7.
Because the pressure difference generates a significant flow rate, particularly upon commencement, diffusion processes are negligible. Two options then present. Firstly, if the quantity is small, it may be advantageous to transfer most or all of it into the analyzer 1. In that case it is important that the volume of the variable volume reservoir 5 is substantially larger than the volume of the vial 18. If the volume of the vial 18 is, for example, 12 ml and the volume of the variable volume reservoir 5 is 100 ml, following transfer, 89% of the sample would be in the variable volume reservoir 5 and the resultant pressure would be 107 mbar.
Where the sample quantity is greater, a large sample container permits only a small proportion of the sample to be removed from it. In that case, the pressure in the sample container is only slightly affected.
Other than blowing of the analyte gas out of the exit towards the first supply valve 7 along a gas supply line, rather than sucking it along a gas supply line to that first supply valve 7, as in
(i) Continuous Carrier Gas Flow with Peak or Pulse Analyte Gas in it
The embodiments described above in connection with
As explained above, where the analyte gas flow is a gas pulse and is thus generally not constant in time, it does not have the same isotopic composition throughout the pulse. Thus it is desirable to ensure that the whole pulse is sampled, either through a transfer of the whole pulse into the variable volume reservoir 5, or by transferring only a part of the pulse, by splitting away part of the gas mixture, as described above, in Section I(ii), during the whole time of the analyte pulse.
Measurement then takes place in accordance with the description set out in I(iii) above.
In order to permit determination of when, in the carrier gas flow, the analyte gas pulse occurs—and hence, when measurement should be started—various techniques can be employed. For example, in some circumstances, the time of commencement of the pulse is already known. This is the case, for example, when an elemental analyzer is employed. It is known when the sample is placed within the elemental analyzer and the transfer time of the analyte through and out of the elemental analyzer is also known (and is a short period of time). Alternatively, where the time is not known, an additional detector such as a thermal conductivity detector (TCD) may be employed. Further, a technique similar to that described above in Section I(i)(b) above can be employed. In particular, with the gas inlet system 20 of
As an alternative to the arrangement of
Still a further example of a pulsed analyte gas supply for the gas inlet system of
As seen in
As a GC chromatogram consists of a number of peaks, it is necessary to isolate only interesting peaks by connecting the first source of analyte gas 9 to the variable volume reservoir 5, by appropriate switching of the first supply valve 7, and by commencing expansion of the variable volume reservoir 5. The time at which selection of an interesting peak takes place is determined through the use of an additional detector 29 such as a flame ionisation detector (FID). Part of the gas flow after the GC is split away through a split 30 to the FID detector 29. Alternatively, it may be possible to use an in line detector, located where the split 30 is shown in
There are several options to investigate several peaks within one gas chromatogram. Firstly, it may be accepted that only one peak per injection can be analyzed. Then the number of injections is increased to take that into account. Secondly, instead of a single variable volume reservoir 5, multiple reservoirs may be employed, each of which individually stores separate peaks. Still further, a stopped flow approach may be employed in the GC. After a first peak is stored in the variable volume reservoir 5, and whilst it is being processed, the carrier gas flows in the gas chromatograph 28 so that the other peaks are “parked” in that GC 28. The stopped flow approach, though rarely employed in practice, is known in the art—see for example Analytical Chemistry 2008, Volume 80, Pages 5481-5486 which describes a GC/NMR coupling. Peaks may be broadened through a diffusion process—see for example Chromatographia 14(12), 1981, 695-696. There are, however, columns where these effects are negligible: see, for example, (http://www.sigmaaldrich.com/content/dam/sigmaaldrich/countries/japan/analytical-chromatography/doc/j-t412094.pdf)
As still another option within the gas inlet system having a GC input coupling is to employ an injection loop valve. This is shown in
(ii) Uses of the Discrete Sampling Inlet Port 10 with a Second Source of Analyte Gas 11
In a variant of the implementation of
Although the foregoing describes various discrete sample injection arrangements as a second source of analyte gas 11, any of these arrangements alone or in combination may be combined using a multi-port device. A suitable arrangement is shown in
Before commencing measurement, each separate port of multi-port 22 may be flushed according to the procedure set out above in connection with
Although a number of embodiments have been described, the skilled reader will readily recognize that various alternative arrangements may be contemplated. Moreover, it is to be understood, in the foregoing, that the terms “first source of analyte gas” and “second source of analyte gas” are not intended to imply necessarily that the analyte gases themselves are different. As has been explained, the first source of analyte gas is, in the embodiments described, either a continuous or pulsed flow of analyte gas within a continuous flow of carrier gas, whilst the second source of analyte gas is exemplified as a discrete sampling arrangement. However, the terms “first” and “second”, as used in the claims in respect of the analyte gas sources, are not intended to imply one or other of the analyte supply types (continuous, pulsed, discrete . . . ), but are simply used as terms to distinguish between alternative analyte gas supplies that maybe present.
The gas inlet system preferentially comprises a variable volume reservoir, as described. However, the reservoir does not necessarily need to be of variable volume for the gas inlet to function. The gas analyte can be taken up and/or expelled from a reservoir of fixed volume, for example by control of the gas flow into a (fixed volume) reservoir and/or through the use of one or more pumps. Furthermore, the volume of the reservoir may be fixed during take up of the analyte, but then compressed to expel it, or of variable volume during uptake of the analyte gas, but of fixed volume when subsequently expelled.
The foregoing description and accompanying drawings illustrate that in various embodiments the invention has the following capabilities:
Number | Date | Country | Kind |
---|---|---|---|
1319766.0 | Nov 2013 | GB | national |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/EP2014/074205 | 11/10/2014 | WO | 00 |
Publishing Document | Publishing Date | Country | Kind |
---|---|---|---|
WO2015/067812 | 5/14/2015 | WO | A |
Number | Name | Date | Kind |
---|---|---|---|
5644130 | Raatz | Jul 1997 | A |
5976890 | Gehre et al. | Nov 1999 | A |
20030218131 | Fujita | Nov 2003 | A1 |
20070274845 | Ruprecht et al. | Nov 2007 | A1 |
20090159795 | Hatscher | Jun 2009 | A1 |
20100198736 | Marino | Aug 2010 | A1 |
Number | Date | Country |
---|---|---|
11083851 | Mar 1999 | JP |
2009-222613 | Oct 2009 | JP |
2009222613 | Oct 2009 | JP |
2006014555 | Feb 2006 | WO |
2010088657 | Aug 2010 | WO |
Entry |
---|
Benson, et al “Forensic applications of isotope ratio mass spectrometry—a review” Forensic Science International 157 (2006) pp. 1-22. |
Brenna, et al., “High-precision continuous-flow isotope ratio mass spectrometry”, Mass Spectrometry Reviews 16 (5), 1997, pp. 227-258. |
Number | Date | Country | |
---|---|---|---|
20160282263 A1 | Sep 2016 | US |