Time-of-flight (TOF) mass spectrometry (MS) has become a widely used analytical technique. Two important metrics of mass spectrometry instrumentation performance are resolving power and sensitivity. In mass spectrometry, the mass resolving power of a measurement is related to the ability to separate ions of differing mass-to-charge ratio (m/z) values. The sensitivity of a mass spectrometry instrument is related to the efficiency of ion transmission from source to detector, and the efficiency of ion detection. In various mass spectrometers, including TOF instruments, it is possible to improve the resolving power at the expense of sensitivity, and vice versa.
There are several aspects of TOF MS that can inherently limit the resolution of a TOF mass analyzer. Specifically, ions can be formed in the source region at different times, at different positions, and with different initial velocities. These spreads in ion formation time, position and velocity can result in some ions with the same m/z achieving different kinetic energies (and some ions with different m/z achieving the same kinetic energy) due to differences in the length of time they spend in the extracting electrical field, differences in the strength of the electrical field where they are formed, and/or different initial kinetic energies. As a result, the resolving power and performance of the TOF mass spectrometer instrument can be degraded.
The mass resolving power of a mass spectrometer may be expressed as a ratio m/δm, where m is the mass of a particular singly charged ion and /δm is the width of the peak in mass units. In traditional TOF instruments, ions are separated according to their flight time, t, to a detector, and in most cases the mass/charge ratio is proportional to the square of the flight time. Thus, the resolving power, R, can be expressed as,
R=m/δm,
and as
R=t/2δt
in a TOF instrument.
In a simple linear TOF instrument comprising an ion source where the ions are formed and accelerated to a final energy that is substantially independent of the m/z ratio of the ions, the flight time is proportional to the effective flight distance, inversely proportional to the square root of the ion energy, and directly proportional to the square root of the mass/charge ratio. Any variation in the kinetic energy or effective flight distance for an ion of a particular m/z causes a variation in the flight time and corresponding reduction in resolving power.
In many cases a major factor limiting resolving power can be the spread in kinetic energy of the ions. In these cases an ion mirror is often employed to compensate for, to first or second order, the effect of kinetic energy on flight time, thereby improving the resolving power of the TOF instrument. One property of prior art ion mirrors, however, is that they produce energy dispersion whereby ions of differing kinetic energies may be time focused at a particular focal plane, but are displaced in a direction parallel to the plane according to their kinetic energies. In many applications this may not be a problem, but in others it can limit both the resolving power and the sensitivity of the mass analyzer. For example, in a single stage TOF instrument this energy dispersion can cause ions of different kinetic energies to strike different spots on the detector, but if the detector is sufficiently large, and the plane of the detector is accurately aligned with the focal plane, then no loss in either resolving power or sensitivity substantially occurs. However, applications where the ion mirror is used in the first stage of a TOF-TOF system, energy dispersion in the first stage can cause significant losses in both sensitivity and resolving power in the second stage of the instrument.
The present teachings relate to ion optics systems for mass analyzer systems.
An ion mirror can be used to reflect ions from a first focal plane (an object plane) to a second focal plane (an image plane) such that ions at the first focal plane reach the second focal plane at substantially the same time despite differences in kinetic energy that existed between these ions at the first focal plane. Herein we refer to the process whereby an ion mirror can be used to bring ions with different kinetic energies to a particular plane in space at substantially the same time as “energy focusing.” However, although ions can be made to arrive substantially simultaneously at an image plane despite differences in kinetic energy between them at the object plane, ions with differing kinetic energy do not arrive at the same spatial location on the image plane. Rather, the exit trajectories of ions with different kinetic energy intersect the image plane (or a plane substantially parallel to the image plane) at different spatial locations, which are typically laterally dispersed across such a plane. This process has been referred to as “energy dispersion” because, for example, it refers to a spatial dispersion of the ion trajectories that is due to differences in ion kinetic energy.
The skilled artisan will recognize that the concepts described herein using the terms energy dispersion, energy focusing, object plane and image plane can be described using different terms. As an ion mirror can be used to bring ions with different kinetic energies to a particular plane in space at substantially the same time, this process has been referred to by several terms in the art including, “energy focusing,” “time focusing” and “temporal focusing.” In addition, for example, the terms “space focus,” “space focus plane,” “space focal plane,” “time focus,” and “time focus plane” have all been used in the art to refer to one or more of what are referred to herein as the object plane and image plane. Unfortunately, the terms “energy focusing,” “time focusing,” “temporal focusing,” “space focus,” “space focus plane,” “space focal plane,” “time focus,” and “time focus plane” have also been used in the art of time-of-flight mass spectrometry to describe processes that are fundamentally different from the energy focusing of an ion mirror. Accordingly, given the complex usage of terminology found in the mass spectrometry art, the terms “energy dispersion,” “energy focusing,”0 “object plane” and “image plane” used herein were chosen for conciseness and consistency in explanation only and should not be construed out of the context of the present teachings to limit the subject matter described in any way.
The present teachings provide ion optics systems comprising two or more ion mirrors. In various embodiments, the present teachings provide ion optics systems that can provide energy focusing of ions with substantially no spatial dispersion due to differences in kinetic energy the ions may have had on entering the ion optics system. It is to be understood that differences in ion kinetic energy due to other processes that might arise after ions enter the ion optics system (e.g., including, but not limited to, space charge effects, ion fragmentation, etc.) are not considered by the present teachings to be differences in kinetic energy the ions have on entering the ion optics system. In various embodiments, the ion mirrors of an ion optics system according to the present teachings are arranged substantially mirror-symmetric about a plane.
A wide variety of arrangements of ion mirrors exists within the present teachings. For example, the ion mirrors can be arranged such that the ion trajectory exiting the ion optics system is substantially parallel, substantially anti-parallel, or at almost any angle in between, relative to the corresponding ion trajectory entering the ion optics system. The ion trajectory entering an ion optics system and the ion trajectory exiting the ion optics system can be on opposite sides of a symmetry plane.
In various embodiments, the ion mirrors can be arranged to provide a select lateral displacement, or substantially no lateral displacement between an incoming ion trajectory and the corresponding outgoing ion trajectory. For example, in various embodiments, the ion mirrors can be arranged such that the ion trajectory exiting an ion optics system is substantially coincident with the corresponding ion trajectory entering the ion optics system and either parallel or anti-parallel thereto.
In various aspects, the present teachings provide an ion optics system comprising an even number of ion mirrors arranged such that a trajectory of an ion exiting the ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the ion optics system at a position that is substantially independent of the kinetic energy the ion had on entering the ion optics system. In various embodiments, the ion mirrors are arranged in pairs where the first member and second member of each pair are disposed on opposite sides of a first plane such that the first member of the pair has a position that is substantially mirror-symmetric about the first plane relative to the position of the second member of the pair.
In various aspects, the present teachings provide an ion optics system comprising a first ion mirror and a second ion mirror, where the first ion mirror and second ion mirror are arranged such that a trajectory of an ion exiting the second ion mirror can be provided that intersects a surface substantially parallel to a focal surface of the second ion mirror at a position that is substantially independent of the kinetic energy the ion had on entering the first ion mirror. In various embodiments, the first ion mirror and the second ion mirror are disposed on opposite sides of a first plane such that the first ion mirror and the second ion mirror are arranged substantially mirror-symmetric about the first plane. Accordingly, in various embodiments, the electrical fields of the first ion mirror are substantially mirror-symmetric about the first plane with respect to the electrical fields of the second ion mirror.
In various aspects, the present teachings provide an ion optics system comprising two or more pairs of ion mirrors where the members of each pair of ion mirrors are disposed on opposite sides of a first plane such that the first member of a pair of ion mirrors has a position that is substantially mirror-symmetric about the first plane relative to the position of the second member of the pair. In various embodiments, the ion mirrors are arranged such that a trajectory of an ion exiting the ion optics system can be provided that intersects a surface substantially parallel to a focal surface of the ion optics system at a position that is substantially independent of the kinetic energy of the ion had on entering the ion optics system.
In various aspects, the present teachings provide an ion optics system comprising four ion mirrors where the first ion mirror and the second ion mirror disposed on opposite sides of a first plane such that the first ion mirror has a position that is substantially mirror-symmetric about the first plane relative to the position of the second ion mirror and where the third ion mirror and the fourth ion mirror are disposed on opposite sides of the first plane such that the third ion mirror has a position that is substantially mirror-symmetric about the first plane relative to the position of the fourth ion mirror. In various embodiments, the ion mirrors are arranged such that a trajectory of an ion exiting the fourth ion mirror can be provided that intersects a surface substantially parallel to a focal surface of the fourth ion mirror at a position that is substantially independent of the kinetic energy the ion had on entering the first ion mirror.
In various embodiments of an ion optics system of the present teachings, the ion optics systems comprises one or more of an ion source, ion selector, ion fragmentor, and ion detector. The ion optics systems can further comprise one or more ion guides (e.g., RF multipole guide, guide wire), ion-focusing elements (e.g., an einzel lens), and ion-steering elements (e.g., deflector plates). In various embodiments, an ion selector is positioned between two ion mirrors of an ion optics system to prevent the transmission of ions with select kinetic energies. Such placement can take advantage of the energy dispersion that can exist between at least two ion mirrors of the ion optics system. Suitable ion selectors include any structure that can prevent the transmission of ions based on ion position.
In various embodiments, an ion optics system of the present teachings comprises a first ion optics system and a second ion optics system. In various embodiments, the first ion optics system comprises an even number of ion mirrors arranged such that a trajectory of an ion exiting the first ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the first ion optics system at a position that is substantially independent of the ion kinetic energy; and the second ion optics system comprises an even number of ion mirrors arranged such that a trajectory of an ion exiting the second ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the second ion optics system at a position that is substantially independent of the ion kinetic energy. The ion mirrors of the first ion optics system, the second ion optics system, or both, can be arranged in pairs where the first member and second member of each pair are disposed on opposite sides of a first plane such that the first member of the pair has a position that is substantially mirror-symmetric about the first plane relative to the position of the second member of the pair.
In various embodiments, an ion fragmentor is disposed between the first ion optics system and the second ion optics system. The ion fragmentor is disposed, in some embodiments, such that the entrance to the ion fragmentor substantially coincides with the image surface (e.g., image plane) of the first ion optics system. In some embodiments, the ion fragmentor is disposed such that the exit of the ion fragmentor substantially coincides with a focal surface (e.g., an object focal surface) of the second ion optics system. In various embodiments, an ion selector can disposed between ion mirrors of the first ion optics system to prevent, for example, the transmission of ions with select kinetic energies between two ion mirrors of the first ion optics system, and thereby, select the range of ion kinetic energies transmitted by the first ion optics system. Accordingly, in various embodiments, the first ion optics system selects a primary ion, with a kinetic energy in a selected energy range, for introduction into an ion fragmentor and the second ion optics system is configured to transmit at least a portion of the fragment ions.
In various aspects, the present teachings provide mass analyzer systems comprising an ion optics system and one or more mass analyzers. The one or more mass analyzers comprising, for example, at least one of a time-of-flight, quadrupole, RF multipole, magnetic sector, electrostatic sector, ion trap, and an ion mobility spectrometer. The mass analyzer systems can further comprise one or more ion guides (e.g., RF multipole guide, guide wire), ion-focusing elements (e.g., an einzel lens), ion-steering elements (e.g., deflector plates), ion sources, ion selectors, ion fragmentors, and ion detectors. In various embodiments, the mass analyzer systems the present teachings can provide include, but are not limited to: a first time-of-flight (TOF) mass selector for a tandem TOF-TOF mass spectrometer system; and a TOF-TOF mass spectrometer system.
In various embodiments, the present teachings provide mass analyzer systems comprising a first ion optics system and a first mass analyzer. The first ion optics system comprising an even number of ion mirrors arranged such that a trajectory of an ion exiting the first ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the first ion optics system at a position that is substantially independent of the kinetic energy the ion had on entering the first ion optics system; and the first mass analyzer comprising at least one of a time-of-flight, quadrupole, RF multipole, magnetic sector, electrostatic sector, ion trap, and an ion mobility spectrometer. In various embodiments, the first ion optics system selects a primary ion for introduction into an ion fragmentor and a mass analyzer is configured to analyze at least a portion of the fragment ion spectrum.
In various embodiments, a mass analyzer system further comprises one or more ion selectors. In various embodiments, an ion selector is disposed between: an ion optics system and a mass analyzer, two ion mirrors of an ion optics system to prevent the transmission of ions with select kinetic energies, or both. For example, in various embodiments, an ion selector is disposed between a ion optics system and a mass analyzer such that the location of the ion selector substantially coincides with the image surface (e.g., image plane) of the first ion optics system. Suitable ion selectors include, e.g., timed-ion-selectors. In various embodiments, the trajectory of ions from the first ion optics system is substantially coaxial with an axis of the ion selector. In some embodiments, the ion selector is energized to transmit only ions within a selected m/z range to, for example, an ion fragmentor disposed between the ion selector and the mass analyzer. Accordingly, in various embodiments, an ion selector selects a primary ion (from the ions transmitted by the ion optics system) for introduction into an ion fragmentor and a mass analyzer is configured to analyze at least a portion of the fragment ions.
In various embodiments, an ion selector is positioned between two ion mirrors of the first ion optics system to prevent the transmission of ions with select kinetic energies. Such placement can take advantage of the energy dispersion that can exist between at least two ion mirrors of the ion optics system. Suitable ion selectors include any structure that can prevent the transmission of ions based on ion position.
Accordingly, in various embodiments, an ion optics system with an ion selector selects a primary ion, with a kinetic energy in a selected energy range, for introduction into an ion fragmentor and a mass analyzer is configured to analyze at least a portion of the fragment ions. In various embodiments, a first ion optics system with an ion selector selects a primary ion, with a kinetic energy in a selected energy range, for introduction into an ion fragmentor, a second ion optics system is configured select at least a portion of the fragment ions with a kinetic energy in a selected energy range for transmittal, and a mass analyzer is configured to analyze at least a portion of the selected fragment ions.
In various embodiments, the present teachings provide mass analyzer systems comprising a first mass analyzer, a first ion optics system, and a second mass analyzer, where the first ion optics system comprises an even number of ion mirrors arranged such that a trajectory of an ion exiting the first ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the first ion optics system at a position that is substantially independent of the kinetic energy the ion had on entering the first ion optics system. The first mass analyzer comprising, for example, at least one of a time-of-flight, quadrupole, RF multipole, magnetic sector, electrostatic sector, ion trap, and an ion mobility spectrometer; and the second mass analyzer comprising, e.g., at least one of a time-of-flight, quadrupole, RF multipole, magnetic sector, electrostatic sector, ion trap, and an ion mobility spectrometer. In various embodiments the first and second mass analyzers each comprise a time-of-flight (e.g., a substantially electrical field free region).
In various embodiments, an ion selector can be disposed between ion mirrors of an ion optics system of the present teachings to prevent, for example, the transmission of ions with select kinetic energies between two ion mirrors of the ion optics system, and thereby, select the range of ion kinetic energies transmitted by the ion optics system. Accordingly, in various embodiments, an ion optics system with an ion selector selects a primary ion, with a kinetic energy in a selected energy range, for introduction into an ion fragmentor and a mass analyzer is configured to analyze at least a portion of the fragment ions.
In various embodiments, an ion selector (e.g., a timed-ion selector) is disposed between the first ion optics system and the mass analyzer. The ion selector is disposed, in some embodiments, such that the location of the ion selector substantially coincides with the image surface (e.g., image plane) of the first ion optics system. In various embodiments, the trajectory of ions from the first ion optics system is substantially coaxial with an axis of the ion selector. In some embodiments, the ion selector is energized to transmit only ions within a selected m/z range. Accordingly, in various embodiments, an ion selector selects a primary ion (from the ions transmitted by the ion optics system) for introduction into an ion fragmentor and a mass analyzer is configured to analyze at least a portion of the fragment ions.
In various embodiments, a first ion selector can be disposed between ion mirrors of an ion optics system to select the range of ion kinetic energies transmitted by the ion optics system. Accordingly, in various embodiments, an ion optics system with an ion selector selects an ion, with a kinetic energy in a selected energy range, and a second ion selector (e.g. a timed-ion selector), disposed between the ion optics system and a mass analyzer, selects a primary ion for introduction into an ion fragmentor and the mass analyzer is configured to analyze at least a portion of the fragment ions.
In various embodiments, a mass analyzer system of the present teachings further comprises a second ion optics system. In various embodiments, the second ion optics system comprises an even number of ion mirrors arranged such that a trajectory of an ion exiting the second ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the second ion optics system at a position that is substantially independent of the kinetic energy the ion had on entering the second ion optics system. The ion mirrors of the first ion optics system, the second ion optics system, or both, can be arranged in pairs where the first member and second member of each pair are disposed on opposite sides of a first plane such that the first member of the pair has a position that is substantially mirror-symmetric about the first plane relative to the position of the second member of the pair.
In various embodiments, an ion selector is disposed between the first ion optics system and the second ion optics system. The ion selector is disposed, in some embodiments, such that the location of the ion selector substantially coincides with the image surface (e.g., image plane) of the first ion optics system, and the trajectory of ions from the first ion optics system is substantially coaxial with an axis of the ion selector. In some embodiments, the ion selector is energized to transmit only ions within a selected m/z range. Accordingly, in various embodiments, an ion selector selects a primary ion (from the ions transmitted by the first ion optics system) for introduction into an ion fragmentor, a second ion optics system is configured transmit at least a portion of the fragment ions to a mass analyzer which is configured to analyze at least a portion of the selected fragment ions.
As an ion selector can be disposed, in various embodiments, between ion mirrors of an ion optics system of the present teachings to prevent, for example, the transmission of ions with select kinetic energies, any one or more ion optics systems of a mass analyzer system can be, in various embodiments, configured to substantially transmit only ions in a select range of ion kinetic energies.
The foregoing and other aspects, embodiments, and features of the present teachings can be more fully understood from the following description in conjunction with the accompanying drawings. In the drawings like reference characters generally refer to like features and structural elements throughout the various figures. The drawings are not necessarily to scale, emphasis instead being placed upon illustrating the principles of the teachings.
The skilled artisan will understand that the drawings, described herein, are for illustration purposes only. In the drawings the present teachings are illustrated using single-stage ion mirrors, but any ion mirror known in the art, including, but not limited to, gridded ion mirrors employing two or more stages with different fields applied at each stage, as well as gridless ion mirrors, can be used. The drawings are not intended to limit the scope of the present teachings in any way.
To better understand the present teachings, an example of the behavior of ions in a conventional single-stage ion mirror employing a uniform electrical field is provided; and an example of the behavior of ions in a conventional parallel arrangement of two ion mirrors employing uniform electrical fields is provided.
Single Ion Mirror
To better understand the present teachings, an example of the behavior of ions in a conventional single-stage ion mirror employing a uniform electrical field is schematically illustrated in
ax=−(zV/md) (1)
ay=0 (2)
vx=v0 cos α−axt=v0 cos α−(zV/md)t (3)
vy=v0 sin α (4)
x=v0t cos α−(zV/2md)t2 (5)
y=v0t sin α (6)
where the symbol m represents the mass of the ion; z the charge of the ion; V the potential difference between the entrance electrode 102 and end electrode 108; d the distance, along the direction x, between the entrance and end electrodes; α the angle of the entrance ion trajectory relative to the normal to the electrical field 106 at the entrance electrode 102 as illustrated in
Solving equation (3) for t when vx=0 gives the time, t1, corresponding to maximum penetration into the electrical field.
t1=(md/zV)v0 cos α. (7)
Substituting for t1 in equations (5) and (6) yields the ion position at time t1:
x(t1)=d(V0/V)cos2α, (8)
y(t1)=2d(V0/V)cos α sin α=d(V0/V)sin 2α, (9)
where,
V0=(m/2z)v02. (10)
At time 2t1 the ions exit from the ion mirror (for the ion mirror of
y(2t1)=2y(t1) (11)
The distance that an ion would travel in the x direction in time t1 in the absence of the electrical field in the ion mirror, x(eff), can be given by,
x(eff)=v0t1 cos α=2d(V0/V)cos2α=2x(t1). (12)
Ions in the single-stage ion mirror follow a parabolic trajectory as schematically illustrated in
d(eff)=x(eff)/cos α=2d(V0/V)cos α. (13)
As can be seen from equations (10) and (13), for ions with a given m/z value, those with a lower kinetic energy, e.g., E1, have a shorter d(eff) than those with a higher kinetic energy, e.g., E2, as illustrated in
For an ion traveling with constant velocity v0 the time required to travel d(eff) is given by
t(eff)=d(eff)/v0=2d(V0/V)(m/2zV0)1/2 cos α=(md/zV)v0cos α=t1. (14)
and the distance traveled in the y direction at time 2t1 can be given by,
y(2t1)=2y(t1)=4d(V0/V)cos α sin α=2d(eff)sin α=y(eff). (15)
Thus, both the residence time of the ions in the single-stage ion mirror and the final direction of the ions after exiting the single-stage ion mirror are substantially identical to the hypothetical case of an ion elastically reflected from a planar mirror. The latter is physically impossible because it would require infinite acceleration, but it can allow the effects of combinations of mirrors to be illustrated and examined without introducing errors or approximations.
In a single-stage ion mirror the energy dispersion, the spatial dispersion of the ions in the y direction due to differences in ion kinetic energy on entering the ion mirror, can be given by the derivative of y with respect to the energy V0. Differentiating equation (15) with respect to V0 yields:
∂y(eff)/∂V0=(4d/V)cos α sin α=(2d/V)sin 2α (16)
Referring again to
The focal distances for the single-stage ion mirror with uniform electric field of
t(ff)=dff/v0. (17)
For a mass analyzer system consisting of a field-free region and an ion mirror the total flight time, t(total), can be given by
t(total)=t(ff)+2t1=dff/v0+(2md/zV)v0 cos α. (18)
The condition for first-order time focusing is that the derivative of t(total) with respect to velocity must vanish, that is,
∂t(total)/∂v0=−dff/v02+(2md/zV)cos α=0. (19)
Substituting for v0 from equation (10) and solving for dff can give the time focus condition for a single-stage ion mirror,
dff=4d(V0/V)cos α=2d(eff). (20)
Accordingly, as illustrated in
d1+d2=4d(V0/V)cos α (21)
where d, as described above with respect to
Parallel Ion Mirrors
To better understand the present teachings, an example of the behavior of ions in a conventional parallel arrangement of two ion mirrors 200 employing a uniform electrical field is illustrated in
One effect of using two parallel ion mirrors back-to-back, as illustrated in
Ion Optics Systems
A wide variety of ion mirrors can be employed in the ion optics systems of the present teachings including, but not limited to, single-stage, two-stage, and multi-stage ion mirrors. The electrical potential in a suitable ion mirror can be linear or non-linear. It is to be understood that the ion mirrors in the figures are illustrated schematically. For example, ion mirrors typically comprise multiple electrodes for establishing the electrical fields therein, and can contain guard electrodes to prevent stray electrical fields from entering field-free regions. The electrodes of a suitable ion mirror can comprise grids, can be gridless, or a mixture of gridded and gridless electrodes. Further, it is to be understood that although the entrance electrode electrical potential is often noted as zero, this is purely for convenience of notation and conciseness in the equations appearing herein. One of skill in the art will readily recognize that it is not necessary to the present teachings that the potential at an entrance electrode be at a true earth ground electrical potential. For example, the potential at the entrance electrode can be a “floating ground” with an electrical potential significantly above (or below) true earth ground (e.g., by thousands of volts or more). Accordingly, the description of an electrical potential as zero or as ground herein should not be construed to limit the value of an electrical potential with respect to earth ground in any way.
It is to be understood that the ion trajectories schematically illustrated in
Referring to
The symmetric arrangement of two ion mirrors 302, 304, illustrated in
In various embodiments, the two ion mirrors are disposed on opposite sides of a first plane 313 (illustrated as the line-of-intersection of the first plane with the plane of the page) such that the first ion mirror 302 and the second ion mirror 304 are arranged substantially mirror-symmetric about the first plane 313. The angle 314 between the initial trajectory 310 and final trajectory 312 is equal to about four times the angle, α, of the initial trajectory 310 with respect the normal 318 to the entrance electrical field of the first ion mirror.
The angle, α, between the incident ion trajectory and the normal to the entrance electrode electrical field can be any angle. This incident angle can be selected, for example, based on a desired angle between the incident ion trajectory and the outgoing ion trajectory. For entrance electrode electrical fields which are not substantially planar, the plane tangent to the entrance electrode electrical field at the point or region of intersection of the incident ion trajectory can be taken as the plane of the entrance electrode electrical field. Although the angle between the incident trajectory and the normal can be any value, for practical reasons the minimum practical angle can be limited by structures used to shield the ion beam (whether a continuous or pulsed beam) in the field-free region from the ion mirror voltages. In general, the physical size of an ion mirror in relation to the field-free distance increases with increasing incident angle, while the applied voltage required for a given kinetic energy is generally decreases with increasing angle.
Referring to
In various embodiments, the first ion mirror is positioned so that the plane of the entrance electrode electrical field of the first ion mirror lies substantially in a plane that intersects the first plane at about an angle β, and the entrance electrode electrical field of the second ion mirror lies substantially in a plane the that intersects the first plane at about an angle β. For entrance electrode electrical fields which are not substantially planar, the plane tangent to the entrance electrode electrical field at the point or region of intersection of the incident ion trajectory can be taken as the plane of the entrance electrode electrical field.
For example, referring again to
Examples of ion trajectories for ions with two different ion kinetic energies E1 and E2 (where E1<E2) on entry to the first ion mirror 402 are also illustrated in
One or more of the incident trajectory angle α and the angle β can be greater than about 22.5 degrees. For example, referring to
Examples of ion trajectories for ions with two different ion kinetic energies E1 and E2 (where E1<E2) on entry to the first ion mirror 502 are also illustrated in
In various embodiments, an ion selector can be positioned between the first ion mirror and the second ion mirror to prevent, for example, the transmission of ions with select kinetic energies from the first ion mirror to the second ion mirror. Such placement can take advantage of the energy dispersion of trajectories between the two ion mirrors. Suitable ion selectors include any structure that can prevent the transmission of ions between the first ion mirror and the second ion mirror based on ion position. Examples of suitable ion selectors include, but are not limited to, ion deflectors, and structures containing one or more openings (e.g., a slit, aperture, etc.). The openings can be constant or variable. Examples of suitable structures containing one or more openings include, but are not limited to, apertured plates, shutters, and choppers (e.g., rotary choppers). In some embodiments, the ion selector is positioned in a symmetry plane passing between the first and second ion mirrors.
Referring to
In various aspects, the present teachings provide an ion optics system comprising two or more of pairs of ion mirrors where the members of each pair of ion mirrors are disposed on opposite sides of a first plane such that the first member of a pair of ion mirrors has a position that is substantially mirror-symmetric about the first plane relative to the position of the second member of the pair. Referring to
In various embodiments, the ion mirrors are arranged such that a trajectory 620, 720, 820 of an ion exiting the ion optics system (i.e., a focal surface of the last ion mirror 610, 710, 804 of the ion optics system exited by the ion) can be provided that intersects a surface substantially parallel to a focal surface 622, 722, 822 (e.g., a focal plane) of the fourth ion mirror 610, 710, 810 at a position that is substantially independent of the kinetic energy the ion had on entering the ion optics system (e.g., on entering the first ion mirror 602, 702, 802).
The ion mirrors can be arranged to provide a selected relative angle, α, between an incident ion trajectory and the normal to the entrance electrode electrical field. In
In various embodiments, the ion mirrors are arranged such that an ion trajectory exiting the ion optics system is substantially anti-parallel (180 degrees) to the corresponding ion trajectory entering the ion optics system. For example, in
In various embodiments, the ion optics systems of
Similarly, the ion optics system of
The outgoing ion trajectory, for the ion optics systems of
Examples of ion trajectories for ions with two different ion kinetic energies E1 and E2 (where E1<E2) on entry to the first ion mirror 602, 702 are also illustrated in
Referring again to
In various embodiments, an ion selector 660 (e.g., a timed ion selector) can be positioned between the third and fourth ion mirrors. The ion selector 660 is disposed, in some embodiments, such that the location of the ion selector substantially coincides with the image surface (e.g., image plane) of a first ion optics system (e.g., the first ion mirror 602 and the third ion mirror 608 taken together), the symmetry plane 606, or both. In various embodiments, the trajectory of ions from the first ion optics system is substantially coaxial with an axis of the ion selector. In some embodiments, the ion selector is energized to transmit only ions within a selected m/z range. Accordingly, in various embodiments, an ion selector selects a primary ion (from the ions transmitted by the ion optics system) for introduction into an ion fragmentor 640. In various embodiments, a second ion optics system (e.g., the second ion mirror 604 and the fourth ion mirror 610 taken together) is configured select at least a portion of the fragment ions with a kinetic energy in a selected energy range for transmittal.
Referring to
In various embodiments, an ion selector 760 (e.g., a timed ion selector) can be positioned between the second and third ion mirrors. The ion selector 760 is disposed, in some embodiments, such that the location of the ion selector substantially coincides with the image surface (e.g., image plane) of a first ion optics system (e.g., the first ion mirror 702 and the second ion mirror 704 taken together), the symmetry plane 706, or both. In various embodiments, the trajectory of ions from the first ion optics system is substantially coaxial with an axis of the ion selector. In some embodiments, the ion selector is energized to transmit only ions within a selected m/z range. Accordingly, in various embodiments, an ion selector selects a primary ion (from the ions transmitted by the ion optics system) for introduction into an ion fragmentor 740. In various embodiments, a second ion optics system (e.g., the third ion mirror 708 and the fourth ion mirror 710 taken together) is configured select at least a portion of the fragment ions with a kinetic energy in a selected energy range for transmittal.
Referring again to
Examples of ion trajectories for ions with two different ion kinetic energies E1 and E2 (where E1<E2) on entry to the first ion mirror 802 are illustrated in
In various embodiments, the ion mirrors can be arranged such that the ion trajectory exiting an ion optics system is substantially coincident with the corresponding ion trajectory entering the ion optics system and either substantially parallel or substantially anti-parallel thereto. For example, in
Referring to
In various aspects, the present teachings provide mass analyzer systems comprising a first ion optics system and one or more of an ion source, ion selector, ion fragmentor, and ion detector, ion guide, ion-focusing element, ion-steering element, and one or more mass analyzers (e.g., one or more of a time-of-flight, quadrupole, RF multipole, magnetic sector, electrostatic sector, ion trap, and ion mobility spectrometer). The first ion optics system comprising an even number of ion mirrors arranged such that a trajectory of an ion exiting the first ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the ion optics system at a position that is substantially independent of the kinetic energy the ion had on entering the first ion optics system. In various embodiments, the ion mirrors of the first ion optics system are arranged in pairs with the first member and second member of each pair are disposed on opposite sides of a first plane such that the first member of the pair has a position that is substantially mirror-symmetric about the first plane relative to the position of the second member of the pair. The mass analyzer systems can further comprise one or more ion guides (e.g., RF multipole guide, guide wire), ion-focusing elements (e.g., an einzel lens), and ion-steering elements (e.g., deflector plates).
Suitable ion sources include, but are not limited to, electron impact (EI) ionization, electrospray ionization (ESI), and matrix-assisted laser desorption ionization (MALDI) sources. Suitable ion detectors include, but are not limited to, electron multiplies, channeltrons, microchannel plates (MCP), and charge coupled devices (CCD).
Suitable ion fragmentors include, but are not limited to, those operating on the principles of: collision induced dissociation (CID, also referred to as collisionally assisted dissociation (CAD)), photoinduced dissociation (PID), surface induced dissociation (SID), post source decay, or combinations thereof. Examples of suitable ion fragmentors include, but are not limited to, collision cells (in which ions are fragmented by causing them to collide with neutral gas molecules), photodissociation cells (in which ions are fragmented by irradiating them with a beam of photons), and surface dissociation fragmentors (in which ions are fragmented by colliding them with a solid or a liquid surface).
In various embodiments, an ion optics system, a mass analyzer system, or both, of the present teachings comprises an ion selector. Although in many applications of TOF mass spectrometry it is generally desired to transmit all of the ions within the energy range produced by the ion source, in some applications only select ranges of ion kinetic energies are of interest. In addition to the ions produced directly in the ion source with differing kinetic energies, there may be ions present with lower kinetic energy due to, for example, loss of energy due to fragmentation of the ion after production in, for example, an ion source accelerating field or a field-free space following the ion source. In various embodiments, these ions can be removed by using an ion selector as an energy filter in an ion optics system of the present teachings.
Examples of suitable ion selectors include, but are not limited to, ion deflectors, and structures containing one or more openings (e.g., a slit, aperture, etc.). The openings can be constant or variable. Examples of suitable structures containing one or more openings include, but are not limited to, apertured plates, shutters, and choppers (e.g., rotary choppers).
In various applications of various embodiments comprising an ion selector in an ion optics system, it can be desirable to determine the kinetic energy distributions of the ions of differing masses. This can be accomplished, in various embodiments, by placing a narrow slit or aperture between two of the ion mirrors where the ion trajectories are spatially dispersed due to differences in kinetic energy such that only ions within a small energy increment are transmitted. For example, by measuring the intensities of the ion signals at the ion detector as a function of the voltage applied to the ion mirrors, the energy distributions for all of the ions detected can be measured, with the ions of differing masses arriving at the ion detector at different times.
In various embodiments, a mass analyzer system comprises an ion source, an ion optics system, an ion detector, and one or more mass analyzers (e.g., a substantially electrical field free region which can serve as a time-of-flight) where the ion optics system comprises an even number of ion mirrors arranged such that a trajectory of an ion exiting the ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the ion optics system at a position that is substantially independent of the kinetic energy the ion had on entering the ion optics system.
For example, adding a pulsed ion source, an ion detector, and a mass analyzer (e.g., an electrical field free region) to any of the configurations illustrated in
Referring to
Referring to
In various aspects, the present teachings provide mass analyzer systems comprising a ion optics system and a mass analyzer. The ion optics system comprising an even number of ion mirrors arranged such that a trajectory of an ion exiting the ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the ion optics system at a position that is substantially independent of the kinetic energy the ion had on entering the first ion optics system; and the mass analyzer comprising, for example, at least one of a time-of-flight, ion trap, quadrupole, RF multipole, magnetic sector, electrostatic sector, and ion mobility spectrometer.
In various embodiments, an ion fragmentor is disposed between the ion optics system and the mass analyzer. The ion fragmentor is disposed, in some embodiments, such that the entrance to the ion fragmentor substantially coincides with the image surface (e.g., image plane) of the ion optics system. In some embodiments, ion fragmentor is disposed such that the exit of the ion fragmentor substantially coincides with a focal surface (e.g., an object focal surface) of the mass analyzer.
In various embodiments, an ion selector can be disposed between ion mirrors of the first ion optics system to prevent, for example, the transmission of ions with select kinetic energies between two ion mirrors of the first ion optics system, and thereby, select the range of ion kinetic energies transmitted by the first ion optics system. Accordingly, in various embodiments, the first ion optics system selects a primary ion, with a kinetic energy in a selected energy range, for introduction into an ion fragmentor and a mass analyzer is configured to analyze at least a portion of the fragment ion spectrum.
Referring again to
Referring to
In various embodiments, an ion optics system can be disposed in the field free-region of a mass spectrometer to provide an ion beam with substantially no energy dispersion. For example, adding any of the ion optics system configurations illustrated in
Selected ions and fragments thereof produced in the second field-free region (e.g. using an ion fragmentor) can be further accelerated after they travel an additional distance D3 by a second ion accelerator 1118 providing additional energy Vcc 1120. In various embodiments, selected ions and fragments thereof can be focused at a distance F from the entrance to second ion accelerator 1118. The accelerated ions and fragments can be separated and analyzed in a second mass analyzer 1122. The distance F can be the distance to a focal plane of the second mass analyzer 1122. The timed ion selector, together with the first field-free region 1108 and the second mass analyzer 1122 can comprises a tandem TOF-TOF mass analyzer in which the first stage of the analyzer for selecting ions is a linear TOF (first field-free region 1108) and where the second stage of the analyzer (second mass analyzer 1122), for fragments analysis, can be a linear or reflecting analyzer.
Use of a linear analyzer in the first stage of such an instrument can, however, reduce resolution in situations where the ion source provides ions with the same m/z value but with differing kinetic energies. For example, the energy distribution of ions produced by a MALDI source is typically dependent on the laser fluence, properties of the MALDI matrix and other variables, so that the arrival time distribution of ions of a particular m/z value at the timed ion selector can vary in an uncontrolled fashion. Although a conventional reflecting analyzer could be used for the first stage to improve resolution, the outgoing trajectories of conventional reflecting analyzers are dependent on the kinetic energies of the incoming ions even though the incoming ions may be confined to a beam of very small diameter. Such energy dispersion creates an ion beam that cannot be focused effectively to allow high transmission efficiency through the remainder of a typical TOF-TOF instrument. In various embodiments, use of an ion optics system according to the present teachings inserted in the first field-free region can facilitate overcoming this problem.
For example, a first ion optics system 1106 (comprising an even number of ion mirrors arranged such that a trajectory of an ion exiting the first ion optics system can be provided that intersects a surface substantially parallel to the image focal surface of the first ion optics system at a position that is substantially independent of the kinetic energy the ion had on entering the first ion optics system) can be inserted into the first field-free region 1108 of the TOF-TOF system. In this configuration the time focus for the ion source plus the time focus for the first ion optics system 1106 is chosen so that ions of a selected mass may be focused in time at the timed ion selector (TIS). Normally, the focal length for the first ion optics system 1106 is chosen to be significantly longer than that for the ion source so that effects of source conditions on focus can be reduced.
Aspects, embodiments, and features of the present teachings may be further understood from the following examples, which should not be construed as limiting the scope of the present teachings in any way.
Examples 1 and 2 present results obtained with an Applied Biosystems® 4700 Proteomics Analyzer (sold by Applied Biosystems, 850 Lincoln Centre Drive, Foster City, Calif. 94404, U.S.A.) modified to include in the first field-free region an ion optics system substantially similar to that illustrated in
Referring to
Referring to
This example presents experimental data obtained with the above modified 4700 Proteomics Analyzer operated as a TOF mass analyzer in “unmodified 4700 Proteomics Analyzer” operational mode and in a mode utilizing the inserted ion optics system 1200. In
The vertical lines in
This example presents experimental data obtained with the above modified 4700 Proteomics Analyzer operated as a TOF-TOF mass analyzer in a “4700 Proteomics Analyzer” utilizing the inserted ion optics system 1200. In TOF-TOF operational mode (or MS/MS mode) ions are selected for the second stage of analysis using the timed ion selector of the 4700 Proteomics Analyzer.
The fragment ions in
All literature and similar material cited in this application, including, but not limited to, patents, patent applications, articles, books, treatises, and web pages, regardless of the format of such literature and similar materials, are expressly incorporated by reference in their entirety. In the event that one or more of the incorporated literature and similar materials differs from or contradicts this application, including but not limited to defined terms, term usage, described techniques, or the like, this application controls.
The section headings used herein are for organizational purposes only and are not to be construed as limiting the subject matter described in any way.
While the present teachings have been described in conjunction with various embodiments and examples, it is not intended that the present teachings be limited to such embodiments or examples. On the contrary, the present teachings encompass various alternatives, modifications, and equivalents, as will be appreciated by those of skill in the art.
The claims should not be read as limited to the described order or elements unless stated to that effect. It should be understood that various changes in form and detail may be made without departing from the scope of the appended claims. By way of example, any of the disclosed features can be combined with any of the other disclosed features to provide an ion optics system or mass analyzer system in accordance with the present teachings. For example, any of the various disclosed embodiments of an ion optics system can be combined with one or more of an ion source, ion selector, ion fragmentor, and ion detector, ion guide, ion-focusing element, ion-steering element, another ion optics system and one or more mass analyzers (e.g. one or more of a time-of-flight, ion trap, quadrupole, RF multipole, magnetic sector, electrostatic sector, and ion mobility spectrometer), to provide a mass analyzer an mass analyzer system in accordance with the present teachings. Therefore, all embodiments that come within the scope and spirit of the following claims and equivalents thereto are claimed.
Number | Name | Date | Kind |
---|---|---|---|
4418280 | Matsuda | Nov 1983 | A |
5126565 | Rose | Jun 1992 | A |
5128543 | Reed et al. | Jul 1992 | A |
5202563 | Cotter et al. | Apr 1993 | A |
5449914 | Rose et al. | Sep 1995 | A |
5464985 | Cornish et al. | Nov 1995 | A |
5637879 | Schueler | Jun 1997 | A |
5955730 | Kerley et al. | Sep 1999 | A |
6013913 | Hanson | Jan 2000 | A |
6274866 | Li | Aug 2001 | B1 |
6365892 | Cotter et al. | Apr 2002 | B1 |
6469295 | Park | Oct 2002 | B1 |
6489610 | Barofsky et al. | Dec 2002 | B1 |
6580070 | Cornish et al. | Jun 2003 | B2 |
6770878 | Uhlemann et al. | Aug 2004 | B2 |
6781121 | Davis et al. | Aug 2004 | B1 |
6867414 | Buttrill, Jr. | Mar 2005 | B2 |
6949736 | Ishihara | Sep 2005 | B2 |
Number | Date | Country |
---|---|---|
19633496 | Feb 1996 | DE |
0427532 | May 1991 | EP |
2080021 | Jan 1982 | GB |
2339958 | Feb 2000 | GB |
198707762 | Dec 1987 | WO |
200169648 | Sep 2001 | WO |
Number | Date | Country | |
---|---|---|---|
20060163469 A1 | Jul 2006 | US |