Wireless sensors and miniature electronics are important in monitoring the vital parameters of the human body and the surrounding environment. These devices are particularly important to the elderly, children, pregnant women and the disabled. Wireless sensor systems can and do provide lifesaving assistance, particularly to these categories of patients.
Recent advances in nanoelectronics and semiconductor technology have spurred rapid progress in the development of and integration of nanosensors into application specific integrated circuit (ASIC) devices for biomedical, chemical, and other sensor applications Nanotechnology has been applied, for example, in integrated RF-powered contact lenses, and smart lens technology capable of sensing glucose level from tears. It is desirable to further adapt nanoelectronic systems to other types of biosensors.
There is, for example, a global epidemic in diabetes with a rising incident rate for both type 2 and type 1 diabetes T1D. One in ten healthcare dollars in the U.S. is spent on costs directly attributable to diabetes, with over half of these costs directly or indirectly resulting from poor maintenance of blood glucose (BG) levels. Persons with T1D require very tight monitoring of BG levels to avoid complications from not only hyperglycemia but also hypoglycemia (HYPO). An electronic sensor that monitors the volatile organic compounds corresponding to changes in human breath, therefore, is desirable for detecting HYPO.
Successful implementation of these sensor devices, however, requires low-power readout systems with robust and reliable output that is independent of environmental conditions such as temperature to operate effectively and over a reasonable time frame. Available general purpose analog to digital converters (ADC), however, consume a relatively large amount of energy that limits the lifetime of wireless sensors. Energy efficient, small foot-print, high sensitivity ring oscillator-based sensor read-out systems, particularly complementary metal-oxide-semiconductor (CMOS) technology, have been proposed to meet this need. Traditional ring oscillators, however, are temperature-dependent, and have a limited tuning frequency range. These deficiencies have been major impediments on the path to the realization of low-power sensing systems. The present disclosure addresses these and other issues.
The present disclosure describes a low power miniaturized wireless sensor system equipped with multiple sensors to monitor vital parameters and communicate the information with gateway device for healthcare and other applications. The system employs a ring oscillator design that provides a reduction in power consumption, a wide tuning frequency range, and temperature stable operation for advanced miniature sensing systems.
The present disclosure provides an application specific integrated circuit (ASIC) consisting of analog and digital sub-systems forming a system on chip (SOC) using complementary metal-oxide-semiconductor (CMOS) technology. A low power and wide tuning range current-starved-ring-oscillator design drives capacitance and resistance-based sensors using an arrangement of delay elements with two levels of control voltages. A bias unit provides these two levels of control voltages and consist of CMOS cascade current mirror to maximize voltage swing which give the oscillator wider tuning range and lower temperature induced variations.
The ASIC design uses an efficient method of analog to digital conversion and a novel sequential sensor monitoring for power management. The conversion of analog sensor input to digital is achieved by counting the number of pulses of a sensor-driver in one clock cycle.
A novel method of power management and sequential monitoring of several sensors with CMOS technology is also disclosed. A power efficient digital subsystem design includes a system management unit (SMU) that enables or disables a sensor id. The disclosed design captures the pulse waves from a sensor for a portion of clock cycles, 3 clocks out of a 16-clock cycle, for example, and transmits the signal to a counter module. As a result, the analog sub-system is at ‘on-state’ for only 3/16th fraction (18%) of the time, leading to reduced power consumption.
The system also includes a ring oscillator based temperature sensor supported with symmetrical load that detects temperature from −50° C. to 100° C. with resolution of 0.1° C.
In one embodiment of the invention, a wireless sensor system is provided comprising a sensor for producing pulse signals corresponding to a monitored biometric parameter, a ring oscillator for driving the sensor, a counter for counting the pulses produced by the sensor, a shift register for converting the pulses acquired by the counter to a serial data package, and a system management unit, the system management unit in communication with each of an output of the sensor, an input to drive the ring oscillator, a start and a stop control of the counter, and a start and a stop control of the shift register, the system management unit programmed to drive the ring oscillator to cause the sensor to produce the pulse signal, enable the counter to count pulses corresponding to the sensor data, and to start and stop the shift register to produce the serial data package, wherein the serial data package is adapted to be transmitted wirelessly through an RF transmitter.
In another aspect, a ring oscillator for use in a sensor-driver is provided. The ring oscillator includes a plurality of odd stages, each comprising a current-starved inverter, and a plurality of even stages, each comprising an inverter connected in parallel with the current starved inverters. A capacitive element connected between subsequent odd and even stages of current starved inverters and inverters and adapted to selectively produce a delay, wherein the delay through the ring oscillator is controlled by adjusting the current applied through the CMOS capacitors.
In another embodiment, a sensor device for evaluating hypoglycemia based on the breath of a patient is disclosed. The sensor device comprises a plurality of sensors forming an array, each of the sensors selected to identify a volatile organic compound (VOCs) corresponding to a component of human breath indicative of hypoglycemia, a microcontroller in communication with the plurality of sensors, a user interface in communication with the microcontroller, and a memory in communication with the microcontroller. The microcontroller is programmed to receive an input signal from the user interface indicating a request to breathe into the device to evaluate hypoglycemia, activate the sensor array to detect VOCs corresponding to a users breath, evaluate output of the array to determine a level of hypoglycemia, and store the evaluated level of hypoglycemia in memory.
The sensor device as recited can include a wireless communications system for wirelessly transmitting the evaluated level of hypoglycemia, or a display that can display the level of hypoglycemia. The sensor device can also include an alert system for providing an alert that indicates hypoglycemia has been detected.
The sensors in the sensor device can be field effect transistors, including a channel comprising a sensor material. The sensors in the sensor device can also be resistive or capacitive sensors having a sensor material comprising a nanomaterial selected from the group consisting of gold nanoparticles, carbon nanotubes, graphene, fullerene, carbon black, and combinations thereof. The nanomaterials can be coated with a surface coating or one or more functional groups selected from the group consisting of C1-C9 thiol-alkanes, C10-C20 thiol-alkanes, C2-C9 thiol-aromatics, C10-C20 thiol-aromatics, and combinations thereof to detect ketones, aldehydes, alcohols, and nitrates; can comprise a material selected from the group consisting of polypyrrole, low-density polyethylene (LDPE), poly(ethylene-block-ethylene oxide) (PE-b-PEO), polyethylene glycol (PEG), poly methyl methacrylate (PMMA), poly(vinylidene fluoride-hexafluoropropylene) (PVDF-HFP), and combinations thereoffor acetone detection; or can comprise aromatic or aliphatic surface coatings for detection of aromatic and aliphatic carbon compounds.
These and other aspects of the invention will become apparent from the following description. In the description, reference is made to the accompanying drawings which form a part hereof, and in which there is shown a preferred embodiment of the invention. Such embodiment does not necessarily represent the full scope of the invention and reference is made therefore, to the claims herein for interpreting the scope of the invention.
Referring now to
Referring now also to
In operation, the sensors 26 in the analog subsystem 14 generates pulse waves 34. The output sensor voltage is amplified, and converted to digital pulses 34. The pulses output from the analog subsystem 14 are directed to the digital subsystem 12. The digital subsystem 12 converts the analog sensor output to digital by counting the number of pulses produced by a sensor-driver with the counter 32. The SMU 28 enables or disables the sensors 26 in the analog system 14, and controls sensor drivers to capture sensor output voltage. The counter 32 and corresponding shift register 30 convert output from the sensors 26 from parallel to serial data which can be transmitted via the RF transmitter 16 (
Referring still to
Referring still to
Referring again to
Referring again to
In operation the on-board counter 50 counts clock pulses to produce system control signals. The decoder 54 generates the enable signal, and the multiplexer 52 produces the pulse signal to be provided to the pulse counter 32 though output 64. Although other configurations are possible, here the on-board counter 50 is a four bit counter for counting clock signals, while the pulse counter 32 is an 8 bit counter.
To generate the enabling signal and process the final signal, the multiplexer 52 and decoder 54 sample the CLR260 and SYS 56 level signals. During the three clocks including CLR3 falling, between CLR3 falling and SYS rising, and SYS rising, the decoder 54 generates the enable signal 36 to activate a sensor driver 26 in the analog subsystem. The multiplexer 52 transmits the pulse signal to the pulse counter 32 along the “final” output line 64. Each sensor 25 is therefore monitored for 3 clocks during 64 clocks. During the clock cycle, the counter 32 counts the number of pulses in the received signal 64 to produce eight bit parallel data output that shows the number of the carried pulses in one clock, and the counter keeps the outputs until next zero in CLR3. After pulse counting is complete, the SMU 28 sends the shift 58 signal to the shift register 30 to begin conversion of the parallel data acquired by the pulse counter 32 to serial data. The SMU 28 produces the CLR3 signal to stop the conversion and data packing During conversion, the shift register 30 produces a data package produced of 11 bits. Therefore, 11 clocks after the zero state in the Shift signal 58, the CLR3 signal 62 fall to zero from high logic level. The time interval between counting finishing and resetting is 14 clocks. Therefore, during the time interval, the parallel data is shifted and outputted in serial form.
After completing data packing, the SMU 28 sends an enable signal 36 for next sensor-driver 26 and repeats all process again for a new sensor-driver 26. System Management Unit (SMU) 28 manages all processes and allocates the power supply for the operating modules rather than all modules in the system to minimize power consumption. Low power consumption is achieved by activating a sensor 25 for only a fraction of the clock cycle. Here a sensor 25 is activated only during 3 clocks of 16 clocks cycles while all other sensors remain inactive. The result is that each sensor is activated once every 64-clocks (4×16). Therefore the analog sub-system 14 consumes the power for one sensor during 3 clocks of 16 clocks cycle. With this method, the power consumption is reduced by 82% compared to a configuration where all the sensors are continuously activated. The sensors sampling rate depends on the clock of the system. As a result a higher sampling rate can be achieved by using higher frequency.
Referring again to
A conventional ring oscillator consists of an odd number of inverters (N) connected in series that form a closed loop path. The frequency of oscillation is determined by the overall delay in the inverter loop, which in turn is dependent on the delay in each inverter. The delay in an inverter is controlled by the current through the transistors that make up the inverter (called as control current, ICTRL). In this model, an increase in the current reduces the delay. If VOSC is the amplitude of oscillating output signal, the dependence of the delay in each inverter is given by:
Where CG is the sum of gate-source parasitic capacitances of the MOSFETs. The MOSFET parasitic capacitance further depends on the width and length of the gate of the transistor. The frequency of oscillation is given by:
Combining the above equations, we get the frequency of oscillation as a function of the control current as shown below:
A ring oscillator 38 constructed in accordance with one embodiment of the present disclosure is shown in
In reference to
V
SG
=(VDD−VCTRL1) and VGS
Thus, the drain currents for these two transistors can be written in term of control voltages as:
Where μp and μn are average hole and electron mobility in the channel, COX is Oxide Capacitance, W is gate width, L is gate length, and VT is threshold voltage, respectively. Equations (5) and (6) show that the drain current varies as the square of the control voltage, ISDM1 decreases with VCTRL1 and IDSM5 increases with VCTRL2. As symmetrical load builds up more current, it increases these effects. From
I
CTRL
=I
SD
+I
SD
, and (7)
VSG
That leads to
MOSFET M2 is always in saturation region, and increase in VCTRL1 increases VSDM1, which leads to a nonlinear increase in the current at M2. It eventually increases the total generated current entering the MOSFETs M3 and M4. To further enhance the stability, except for the first stage, direct connections to control voltage for all current starved elements are removed. This was achieved by connecting the symmetrical load 72 to its preceding stage as shown in
Delay in an inverter constructed with CMOS technology is in the range of several picoseconds. To reduce the number of inverters in the oscillator, and also to increase the system stability, dummy transistors are used as capacitors, which increase the delay at each stage. Dummy transistor (DM), shown in
As described above with reference to
In one embodiment of the invention, the oscillator design described above shown in
Referring now to
Referring still to
The sensing material 84 is selected to transform chemical concentrations of analytes in human breath to electrical signals. Analytes interact with sensor materials 84 to cause a change in the electronic or physical property of the material, resulting in a change in conductivity (change in resistance) or a change in the permittivity (change in capacitance) of the sensor material 84. The sensing material 84 can include, for example, finely tuned polymeric materials or nanoparticles coated with organic surface coatings which are sensitive to different VOC analytes. These types of sensors can therefore detect changes in the composition of VOC's in human breath that are, for example, permanently altered by disease states. Sensors have been developed and used to identify disease states such as types of cancer, including lung, breast, colorectal, prostate, and gastric, and also to identify whether a person has diabetes. (See Gang Peng, Meggie Hakim, Yoav Y. Broza, Salem Billan, Roxolyana Abdah-Bortnyak, Abraham Kuten, Ulrike Tisch, Hossam Haick, Detection of lung, breast, colorectal, and prostate cancers from exhaled breath using a single array of nanosensors, British Journal of Cancer 2010, 103, 542-551 Jae Kwak, Michelle Gallagher, Mehmet Hakan Ozdener, Charles J. Wysocki, Brett R. Goldsmith, Amaka Isamah, Adam Faranda, Steven S Fakharzadeh, Meenhard Herlyn, A. T. Johnson, Volatile biomarkers from human melanoma cells, Journal of Chromatography B 2013; 40 J. Hofbauer, H. Dressel, J. Seissler, A. R. Koczulla, D. Nowak, R. A. Jones, Analyse der Ausatemluft mittels Elektronischer Nase bei Patienten mit Diabetes mellitus, Pneumologie 64, V266.) Each of the references cited herein are incorporated by reference for their description of sensor devices detecting VOC's. Descriptions of sensors for detecting VOC's in breath based sensing are also disclosed in “VOLATILE ORGANIC COMPOUND SENSORS, AND METHODS OF MAKING AND USING THE SAME,” attorney docket number 144578.00127, filed on even day herewith, which is also hereby incorporated by reference in its entirety.
Similar systems can be used to detect the transitory changes that can occur in, for example, hypoglycemia (HYPO). Here, the sensors 25 are again constructed with sensing materials 84 to detect transient disease states from VOCs that are present in human breath, as well as significant changes in the relative levels of VOCs. VOCs indicative of HYPO have been identified based on analysis of the breath of individuals identified as experiencing HYPO by, for example, diabetes alert dogs (DADs) using gas chromatograph/mass spectrometry (GC/MS) data. Table I identifies VOCs that have been shown experimentally to correlate with hypoglecemia, in the approximate ranges shown:
Additional data concerning the concentration of VOC's in breath can be collected, for example, by correlating blood glucose level data with breath samples collected when diabetic patients experience HYPO and normoglycemia. The samples can be collected, for example, in Tedlar bags, and transferred from the Tedlar reusable bags to deactivated glass vials with sorbent materials. Subsequently, samples are transferred to a solid phase microextracion (SPME) matrix at 80° C. The SPME matrix is inserted into the GC/MS instrumentation (Agilent GC7890A (GC) and 5975C (MS)) directly. The elements of the VOC signature panel are run through the GC/MS to confirm the identity of analytes previously determined solely by their mass spec signature, and to determine specific analyte concentrations for the HYPO signature breath profile. The number of components found may, in some cases, be increased by utilizing SPME fibers with different matrices including polydimethylsiloxane, carboxen, and divinylbenzene coatings.
The GC/MS data can be analyzed using an Automated Mass Spectral Deconvolution and Identification System, and the results subjected to principal component analysis (PCA) and other multivariate statistics to determine the HYPO signature breath profile from specific changes in the concentration of VOCs. Distinct VOC signatures for Normal breath and HYPO breath can be produced and used in sensor fabrication to create sensor arrays for identifying HYPO.
VOC concentrations such as those shown in Table 1 and collected as described above can be used to construct a sensor array to include individual sensors 25 having sensing materials 84 selected to identify VOCs associated with hypoglycemia, thereby providing a readout of whether breath analyzed with the sensor exhibits characteristics consistent with hypoglycemia. Although a number of materials can be used as sensing material 84, one suitable material is gold nanoparticles coated with dodecanethiol, available from OceanNanoTech LLC, Springdale, Arkansas. Similar materials can be synthesized as described in “Diagnosing lung cancer in exhaled breath using gold nanoparticles,” Nat. Nanotechnol., vol. 4, no. 10, pp. 669-673, October 2009, G. Peng, U. Tisch, O. Adams, M. Hakim, N. Shehada, Y. Y. Broza, S. Billan, R. Abdah-Bortnyak, A. Kuten, and H. Haick. In alternate embodiments, the nanomaterials can be selected from the group consisting of gold nanoparticles, carbon nanotubes, graphene, fullerene, carbon black, and combinations thereof.
These nanomaterials are coated with a surface coating or one or more functional groups selected from the group consisting of C1-C9 thiol-alkanes, C10-C20 thiol-alkanes, C2-C9 thiol-aromatics, C10-C20 thiol-aromatics, and combinations thereofusing a ligand exchange method in order to be suitable to detect ketones, aldehydes, alcohols, and nitrates (See M. Badea, A. Amine, G. Palleschi, D. Moscone, G. Volpe, and A. Curulli, “New electrochemical sensors for detection of nitrites and nitrates,” J. Electroanal. Chem., vol. 509, no. 1, pp. 66-72, August 2001). Other coatings can include polycyclic aromatic hydrocarbons (PAH), carboxylic acid, decanethiol, dodecanethio, tert-dodecanethiol, 4-methoxy-toluenethiol, 2-nitro-4-trifluoro-methylbenzenethiol, and 2-mercaptobenzoxazole.
Additional sensing materials that can be used in these applications include polypyrrole, low-density polyethylene (LDPE), poly(ethylene-block-ethylene oxide) (PE-b-PEO), polyethylene glycol (PEG), poly methyl methacrylate (PMMA), poly(vinylidene fluoride-hexafluoropropylene) (PVDF-HFP), and combinations thereoffor acetone detection (J.-B. Yu, H.-G. Byun, M.-S. So, and J.-S. Huh, “Analysis of diabetic patient's breath with conducting polymer sensor array,” Sens. Actuators B Chem., vol. 108, no. 1-2, pp. 305-308, July 2005.) 2,3-diaminonapthalene which is sensitive to nitrates (A. K. Nussler, M. Glanemann, A. Schirmeier, L. Liu, and N. C. Nüssler, “Fluorometric measurement of nitrite/nitrate by 2,3-diaminonaphthalene,” Nat. Protoc., vol. 1, no. 5, pp. 2223-2226, December 2006.) and aromatic and aliphatic surface coatings for aromatic and aliphatic carbon compounds. (G. Peng, M. Hakim, Y. Y. Broza, S. Billan, R. Abdah-Bortnyak, A. Kuten, U. Tisch, and H. Haick, “Detection of lung, breast, colorectal, and prostate cancers from exhaled breath using a single array of nanosensors,” Br. J. Cancer, vol. 103, no. 4, pp. 542-551, August 2010.)
The sensors can be FET sensors, as described above, resistive, or capacitive sensors. To fabricate the sensors in a resistor or capacitor model, the sensing material is 84 placed between two conducting electrodes 86, 88 and change in resistance or capacitance when breath samples are exposed is measured. Fabrication techniques can include photolithography, self-assembly, spin-casting, drop-casting, spray-coating, chemical-bath-deposition, hot-pressing, evaporation, and sputtering. The layer-by-layer (LbL) self-assembly technique can be advantageously used to with a wide choice of materials and a precise control of film properties at the molecular level to create thin, sensitive channels of sensing material 84.
To test operation, fabricated sensor devices including sensing material 84 are placed in an air-tight chamber. Individual VOCs (with appropriate concentrations) and breath samples are passed through the chamber. Resistance or capacitance can be measured using a Keithley semiconductor characterization instrument. The temperature, pressure, and humidity of the chamber are also measured.
As the VOCs are injected into the chamber, and come in contact with the sensing material 84, they are adsorbed on the material. The attachment of the target analyte molecule to the sensing material 84, through reversible chemical bonding, intermolecular interactions, or physical adsorption will alter the electronic or physical properties of the sensing material 84. For a transistor-based device, the change in electronic property of the channel film can be measured through the drain current and the gate and drain voltages. The relationship between the saturation region drain current and gate voltage of a FET is shown in equation (1).
Where, ID=drain current, μ, =mobility, C0=gate capacitance, W and L=width and length of channel, VG=gate voltage, and VT=turn on voltage.
The drain current in the sub-threshold region is given by equation (2).
Where, VD=drain voltage. As the analytes attach to the sensing material 84, the change in drain current and threshold voltage will be measured. The layer-by-layer (LbL) self-assembly can be used to enable the deposition of films as thin as a few molecules thick, which enables the construction of thinner channel layers and consequently more sensitive FETs to changes caused by adsorbed VOCs. Although FETs are shown and described, other electronic devices can be constructed as sensors. For example, diode models can be constructed to include sensing materials where adsorption of the target analyte causes changes in charge mobility.
Referring now to
In operation, the individual current levels from the sensor array 90 can be collected by readout circuit 92 and analyzed by microcontroller 94 for the corresponding breath status, as described more fully below. The sensor readout circuit 92 manages the sequence of reading the sensor array 90, and input the acquired data to the microcontroller 94, and performing sensor resetting or clearing functions after the read-out or when it receives such an instruction from the microcontroller 94. The microcontroller 94 process the data using a set of instructions provided to compute a breath status value, and can save a date and time stamp in memory 96, where the data can be used as a comparator to evaluate changes in the breath. The breath status value can also be displayed on a local display 98, and wirelessly transmitted the wireless communications system 104 to the patient, designated caregivers, or both. The microcontroller 94 can also activate the alert system 100, which can provide audio or visual signals to a user. The alert system 100 can include, for example, light emitting diodes (LEDs), circuitry for producing audio alerts, and other types of devices.
The wireless communications system 104 can include various types of wireless communications devices, including, for example, a radiofrequency, Bluetooth, or GSM communication system that is in communication with the microcontroller 94, and one or more corresponding antenna 108. Although individual blocks are shown in the block diagram illustrated to represent circuitry having specific functions, it will be apparent that circuitry for performing functions shown in the blocks can be constructed using one or more electronic component, and that blocks can be separated and combined. For example, the wireless system 104 may consist of more than one component, or more than one integrated circuit (IC) chip. Further, various blocks can be combined into a single IC chip or other device.
Referring now to
Referring still to
Referring now also to
Referring still to
Alternatively, the results can be transmitted to a cell phone 106, 110. In some applications, the results can be both locally displayed on display 98 and wirelessly transmitted to a smart phone, cell phone, computer, or other type of personal computing device. For example, the user can observe the result on a local sensor display 98, acknowledge the result if a HYPO breath status is indicated (step 122), and optionally activate a pushbutton or other device associated with user interface 102 to transmit the information to an external device of the type described above. (step 124) The transmitted data can be transmitted to the user and/or the user's selected caregiver. The communication between the sensor device 91 and other wireless devices will be encrypted to ensure patient privacy. Each sensor device 91 will be encrypted with a PGP key that can be selected by the user at the time of device calibration, and must be similarly programmed into the application corresponding to the receiving cell phone, smart phone, or other communications device. The PGP key will be transmitted by the sensor device 91 along with the raw sensor data.
Referring again to
The smart phone application can also download primary and secondary caregiver phone numbers and setup information to the sensor device 91 through wireless communications system 104. The patient and caregivers' smart phones can then receive and integrate breath status data into an ongoing historical log, and the log data can be provided, for example, on a data visualization dashboard. The smart phone interface can display, for example, patient name, date and time of most recent breath status readings, the current breath status, with the corresponding color code for that status, and the time-line breath status data visualization.
The smart phone application can also include menus. The menus can include, for example, a “Home” icon for accessing the other menu functions; a “Sensor” menu to access the setup tools for downloading caregiver phone numbers into the sensor device 91 and to perform sensor calibration; a “Security” menu for preventing unauthorized access to all transmitted personal health data, such as GSM encrypted algorithms including secure encryption protocols for blocking invaders through its peer-to-peer authentication; a “Dashboard” menu that provides an interface for viewing past and current readings of breath status; a “Social” menu enabling access to tools for sending or receiving messages between the patient and the caregiver; a “Profile” menu enabling patients to input their personal information, name, age, etc.; and “Setup” menu enabling access to general system configuration.
After the patient blows into the device, a breath status alert will be sent automatically to the electronic devices identified in memory 96. The patient and the caregiver will be able to review test results, which can be, for example, identified with words such as “NORMAL,” “PROBABLE HYPO,” or “HYPO”. The status can also be color-coded, and provided with a date and time stamp. In some applications, the display can also provide access to historical data. For example, the user can be provided with a function to scroll left and right to see current and past readings, allowing comparison of levels throughout the day. Alerts can be evaluated by the patient in coordination with their caregiver to apply corrective measures to bring the status level back to Normal.
It should be understood that the methods and apparatuses described above are only exemplary and do not limit the scope of the invention, and that various modifications could be made by those skilled in the art that would fall within the scope of the invention. For example, although specific analog and digital configurations are shown, it will be apparent that components of the systems can be combined or structured in different formats. Programmable devices can also be used to provide similar functions. The sensors described herein can be used with various monitoring applications in addition to medical applications. Although specific hardware elements are described, it will be apparent to those of ordinary skill in the art that equivalent elements can be used, and that the construction can be re-configured to reduce the number of components in the system. To apprise the public of the scope of this invention, the following claims are made:
This application claims the benefit of U.S. Provisional patent application Ser. No. 61/978,567 filed on Apr. 11, 2014 and entitled “Low Power Wireless Sensor System with Ring Oscillator” and claims the benefit of U.S. Provisional patent application Ser. No. 62/008,127 filed on Jun. 5, 2014, and claims the benefit of U.S. Provisional patent application Ser. No. 61/978,490, filed Apr. 11, 2014, each of which are hereby incorporated herein by reference in their entireties.
Number | Date | Country | |
---|---|---|---|
61978567 | Apr 2014 | US | |
62008127 | Jun 2014 | US | |
61978490 | Apr 2014 | US |