The skilled person in the art will understand that the drawings, described below, are for illustration purposes only. The drawings are not intended to limit the scope of the applicants' teachings in anyway.
In the drawings, like reference numerals indicate like parts.
It should be understood that the phrase “a” or “an” used in conjunction with the applicants' teachings with reference to various elements encompasses “one or more” or “at least one” unless the context clearly indicates otherwise. Reference is first made to
In the present teachings, where there are regions of differential pressure, there are two physical forces available to accelerate the ions toward the inlet region 26 of the mass analyzer 20 namely, gas drag and electric fields. The two forces have distinct properties; most notably gas drag will move both charged and neutral particles whereas electric fields will move only charged particles. The combined effects of these two forces can be used to advantage to achieve high transfer efficiency for ions and maintain some degree of discrimination between ions and neutral particles.
Based on prior art imaging studies (Puretzky et al., Physical Review Letters, 1999, 83, 444-447), there is indication that some analyte ions occupy the central axis of the plume while a large portion of the matrix or other particles rapidly diverges in the radial direction. Gas drag forces immediately in front of the aperture of the inlet cone to the vacuum chamber are strong and will efficiently draw both ions and neutrals through the aperture. The pressure differential across the aperture isolating these two regions (upstream and downstream of the aperture) is sufficiently high that the mean free path between collisions is much smaller than the 4 mm orifice diameter of the aperture such that losses due to collisions with the walls defining the aperture are insignificant. The subsequent gas expansion injects both the analyte ions and neutral contaminants well into (e.g., several centimeters) the vacuum chamber at roughly sonic velocities. Positioning the sample plate close to the cone aperture assures efficient transfer of ions with the disadvantage of allowing neutral contaminants, along with analyte ions, into the mass analyzer. The target can be spaced outside the immediate influence of the gas drag region and electric fields can be used to direct the ions toward the gas flow streams, thereby presenting an opportunity to discriminate between ions and neutrals.
Optics and analyzer contamination is of particular concern for an instrument whose purpose is to conduct high sample loads. The opportunity to discriminate ions from neutral particles can be achievable in accordance with the applicants' teachings. Experiments were designed to see if a preferential transfer of ions over neutral contaminants could be achieved, taking advantage of the viscous gas drag forces, derived from the pressure differential between the ion source and the vacuum chamber (also often referred to as the Q0 optics region), and focusing electric fields. In various embodiments, a significant proportion of the neutral components of the rapidly diverging matrix plume can be pumped away by the vacuum system if positioned outside the influence of the gas drag region. The importance of these considerations is to find a reasonable balance between high transmission efficiency of analyte ions operating entirely within the gas drag region and reduced transmission efficiency with electric field focusing but maintaining some degree of discrimination between ions and neutral particles. Experiments were conducted to see if this discrimination between ions and neutrals could be achieved. Three experiments were conducted situating the target at varying distances from the cone aperture in order to move the ablation plume in and out of the gas drag region where gas drag forces exert a predominant influence. Absolute signal intensities were used as a measure of efficiency of ion transfer. Measuring the extent of neutral particle discrimination involved laser ablation of a large number of samples and tracking the deposition of neutral material on a lens located within the vacuum chamber by both visual inspection and by measuring a loss of signal intensity due to charging of the ion optics. This charging effect can create distorted potentials due to the build up of insulating layers on the ion optics over time. Defining “one sample” as the laser ablation of a portion of material from one location on the sample plate, samples were consecutively ablated until the absolute signal intensity fell to below 50% of starting conditions. The experiments were repeated on 2 different drug species using the standard CHCA matrix.
The first experiment used laser pulses striking the sample plate at an angle of incidence of 25 degrees relative to the center line of the first ion optical axis of the mass analyzer, an inlet aperture diameter of 4 mm, and a 2 mm spacing between the sample plate and the inlet aperture, where the entire plume would be within the gas drag region and showed the highest efficiency of ion transfer. However, this configuration was also the most susceptible to contamination, showing distinctly visible deposits surrounding the inlet aperture and the quadrupole rods of the Q0 region of the vacuum chamber, and heavy contamination surrounding the exit aperture of the Q0 region after only 30,000 sample ablations at which point a 50% signal reduction was observed. Evidence of this contamination is shown in
Referring to
E=V/d, (Equation 1)
wherein E represents an electric field strength, V represents a first electrical potential applied on the sample plate 34 measured relative to the electrical potential applied on the cone 54 and d represents the distance between the sample plate 34 and the aperture 54a of the mass analyzer 40.
As previously mentioned, the gas drag and electric field forces available to accelerate the ions toward the inlet of the mass analyzer each have distinct properties with respect to particle movement within the analyzer system, that is, gas drag will move both charged and neutral particles whereas electric fields will move charged particles alone. Gas drag forces can have a stronger influence on charged particles than can electric field forces for the electrical potentials applied in these experiments; neutral particles are not charged and therefore are not influenced by electric field forces. In the various embodiments of the applicants' teachings, instrument geometry including spacing between the ion source and the inlet aperture can be configured such that the ablation plume can remain at a sufficient distance away from the inlet region and thus gas drag forces do not substantially draw the charged and the neutral contaminant particles into the mass analyzer. At this sufficient distance where the plume does not substantially intersect the gas drag region and thus the plume is not substantially influenced by the gas drag forces produced in the gas drag region, an electric field can be applied to separate the charged analyte ions from the neutral contaminant particles which can be present in the plume. The electric field can direct only the charged analyte ions into the vicinity of the gas drag region such that the gas drag forces can draw the analyte ions into the mass analyzer, thereby reducing contamination of the system. The combined effects of these two forces can be used to determine appropriate instrument geometry to achieve high transfer efficiency for ions while maintaining some degree of discrimination between ions and neutral particles. For example, the influence of electric field forces depends on the distance of the sample plate 34 from the inlet region 46 of the mass analyzer 40 and the potential applied on the sample plate 34 with respect to the potential applied to the cone 54. The applied electric field can be used to direct the ions toward the inlet region 46 of the mass analyzer 40. In various embodiments, the electrical potential can be from 10 V to 250 V. In various embodiments, the sample plate 34 can be positioned at least 16 mm from the inlet cone 54a of the mass analyzer 40, and the electrical potential on the sample plate 34 can be 60 V, providing an electric field strength of 60 V/16 mm which is 3.75 V/mm or 3,750 V/m.
The influence of gas drag on the plume is also a consideration in the determination of appropriate instrument geometry. Without being held to a particular theory, it is believed that a qualitative guide for determining instrument geometry with respect to gas drag forces can be estimated from the surface area of the aperture (e.g., 28a, 54a) in relation to the surface area of a semi-sphere centered about the inlet aperture. At any given distance from the aperture (radius “r”), the gas “pull through” effect can be approximated by the gas velocity reduction further away from the inlet 5 aperture. This reduction can be approximated by the ratio of the area of the inlet aperture to the area of the semi-sphere at a given distance r. The surface area of the semi-sphere can be given by 2πr2.
At a distance r away from the inlet aperture, the relative gas velocity reduction can be determined as follows:
where D is the diameter of the inlet aperture and r is the distance from the inlet aperture.
If r is one aperture diameter then velocity is reduced by ⅛ (D/D)2 or ⅛ of the velocity through the inlet aperture. The velocity of the 15 gas flow at this distance is not substantially reduced (as is the case in a prior art system with an inlet aperture diameter of 4 mm and a distance of 4 mm from the sample plate to the inlet aperture), only falling off to ⅛ of the velocity through the inlet aperture and as such the amount of contamination that would enter the mass analyzer over time is significant enough to degrade 20 performance of the mass analyzer.
If r is 1.5 aperture diameters, then r=1.5 D and the gas velocity is reduced by ⅛ (D/1.5D)2 or ⅛× 1/2.25= 1/18 of the velocity through the inlet aperture, an improvement over the case of r being equal to one aperture diameter.
Similarly, even more improvement can be achieved if r is 2 aperture diameters. In that case, r=2 D and the gas velocity is reduced by ⅛ (D/2D)2 or ⅛×¼= 1/32 of the velocity through the inlet aperture.
Likewise, if r is 3 aperture diameters, then r=3 D and the gas velocity is reduced even further, that is, by ⅛ (D/3D)2 or ⅛× 1/9= 1/72 of the velocity through the inlet aperture.
Even more improvement can be achieved in the case where r is 4 aperture diameters, then r=4 D and the gas velocity is reduced by ⅛ (D/4D)2 or ⅛× 1/16= 1/128 of the velocity through the inlet aperture. At a distance where the center of the plume does not substantially intersect the gas drag region, an electric field can then be placed to separate the charged analyte ions from the neutral contaminant particles present in the plume. The electric field can direct the charged analyte ions into the vicinity of the gas drag region such that the gas drag forces can draw the analyte ions into the mass analyzer while neutrals are unaffected by the electric field and do not enter the mass analyzer, thereby reducing contamination of the system.
As will be appreciated by those of skill in the art, the foregoing analysis represents a qualitative approach in configuring instrument geometry to determine how to minimize the effect of neutrals entering the mass analyzer from the ablation plume. A more rigorous analysis of the geometry required for the plume to substantially remain outside of the gas drag region takes into effect other parameters including the angle of incidence of the laser pulses, the pressure of the source gas and the diameter of the inlet aperture and is provided below.
In the following equations, cc represents proportionality.
The velocity of the gas flowing from the source to the inlet aperture, VGAS, is reduced, as discussed above, with the square of the radial distance, r, from the tip of the inlet aperture, and can be determined as follows:
where {dot over (m)} is the mass flow of source gas into the inlet aperture and r is the distance from the inlet aperture.
The mass flow can be determined from the following equation:
where {dot over (m)} is the mass flow of source gas into the orifice, M is the molecular weight of the source gas, T0 is the temperature of the source gas, p0 is the pressure of the source gas, and D is the diameter of the inlet aperture.
Substituting Equation 4 into Equation 3 gives the following equation:
Neutral contaminants contained within the laser ablation plume move at a certain velocity, VPLUME, and beyond a given radial distance, r=R (see
AVPLUME>VGAS (Equation 6)
where A is a constant of proportionality with an approximate value of 1.
Substituting Equation 5 into Equation 6 gives the following equation:
which is similar to Equation 2, where p0, T0, and M are constants.
Rearranging Equation 7 by lumping the constants A, M, T0, and VPLUME into the constant B and solving for the distance r=R yields the following mathematical relation for determining the radius of the gas flow dominated region:
R>BD√{square root over (p0)} (Equation 8)
Referring to
sin(θ)>R/L (Equation 9)
where θ is the angle between the center line of the ion path of the instrument and the plume, R is a given radius of the gas flow dominated region, and L is the distance between the sample plate and the inlet aperture.
Substituting Equation 8 into Equation 9 gives the following equation:
sin(θ)>B√{square root over (p0)}D/L (Equation 10)
where constant B is a measure of system contamination by neutrals; higher values of B indicate lower contamination per sample.
For example, in the commercially available QSTAR® system (Applied Biosystems/MDS Sciex), p0=1 Torr, D=4 mm, L=4 mm, and θ=62 degrees. To configure a system that can reduce contamination better than the QSTAR system, B is chosen to be 1 Torr−1/2 or 1, which gives the following equation:
sin(θ)>√{square root over (p0)}D/L (Equation 11)
where the equation can be considered unitless if p0 is in Torr.
Therefore, contamination can be reduced, compared to the prior art, by increasing the angle θ or by increasing the distance between the sample plate and the inlet aperture, L, or by decreasing the diameter of the inlet aperture, D, or decreasing the source pressure, p0.
Decreasing the source pressure is not an efficient or practical design consideration due to the non-linear (square root) relation given by Equation 10. For example, to achieve the same effect of increasing the distance between the sample plate and the inlet aperture from 4 mm to 16 mm, one would need to reduce the source pressure by a factor of 16 (in various embodiments to 60 mTorr), which is far outside the normal region of operation. Changing the diameter of the of the inlet aperture has other deleterious effects on instrument design. For example, increasing the diameter increases the pumping load to maintain the vacuum pressure at desired levels thereby necessitating the use of large, costly pumps, while decreasing the diameter effects ion transmission efficiency and hence sensitivity. Consequently, practical considerations tend to favor making adjustments in either the angle 0 or the distance between the sample plate and the inlet aperture.
For example, when the inlet aperture diameter, D, is 4 mm, p0 is 1 Torr, and B is 1 Torr−1/2, and the angle between the center line of the ion path of the instrument and the plume, angle θ, can be from 0 to 80 degrees, the distance between the sample plate and the inlet aperture, L, can be calculated, as shown in Table 1, to reduce contamination entering the system,
Neutral particles and matrix molecules, if a matrix is used, can also be present in the plume 44. It is generally known that the plume 44 tends to travel in a direction towards the incoming laser beam 42, (Rapid Communications in Mass Spectrometry, 1995, 9, 515-518; International Journal of Mass Spectrometry, 1998, 177, 111-118; Rapid Communications in Mass Spectrometry, 1999, 13, 792-797). As shown above, contaminants entering the mass analyzer can be reduced substantially since the geometry of various embodiments of the applicants' teachings can sufficiently separate the plume 44 and the gas drag region 56. In this manner neutral particles and matrix molecules that can be present in the plume 44 and which tend to contaminate the system are not substantially influenced by the electric field and continue to follow a trajectory directed away from the inlet region 46 of the mass analyzer 40 and towards the laser beam 42, which can be at an angle of incidence from about 0 to about 80 degrees to the center line of the first ion optical axis 38 of the mass analyzer 40.
As shown in
Referring to
In a further example, samples were analyzed under the same settings for a system configured in accordance with the applicants' teachings. The following parameters were used: the pressure in the vacuum chamber was 8 mTorr, laser repetition rate was 1 kHz, pulse energy was 2.5-3.5 μJ, the voltage on the cone was 0V and the voltage on the sample plate was 60 V. The compound that was analyzed was Prazosin.
The following describes a general use of the applicants' teachings which is not limited to any particular embodiment, but can be applied to any embodiment. In operation, a laser, which can be, but is not limited to, a high-repetition rate solid state laser, strikes a portion of a sample, comprising an analyte, on a tilted sample plate. A plume comprising analyte ions can be generated and moves toward the incoming laser beam and away from the inlet region of the mass analyzer. The analyte ions are typically produced at pressures of about 0.2 to about 2 Torr in high pressure MALDI applications, as known in the art. Neutral and matrix particles which can be present in the plume can contaminate the system if they enter the inlet region of the mass analyzer and deposit in locations that can influence the performance of the system, generally close to ion trajectories. In the applicants' teachings, both ions and neutral and matrix particles, if present, follow an initial trajectory towards the incoming laser beam and away from the inlet region of the mass analyzer. This allows the neutrals and contaminants to deposit in places where they do not influence the ion trajectories, which can dramatically improve the signal integrity of a system that has processed a large number of samples. However, since the initial trajectories of the ions point away from the inlet region of the mass analyzer as well, the sensitivity of this system is lower. An electric field, generated by applying a first electrical potential, can direct the analyte ions into the inlet region of the mass analyzer. A second electrical potential can be applied either independently or in conjunction with the first electrical potential to direct the analyte ions in the vicinity of the gas drag region where gas drag forces can draw the analyte ions into the inlet aperture of the mass analyzer.
While the applicants' teachings are described in conjunction with various embodiments, it is not intended that the applicants+ teachings be limited to such embodiments. On the contrary, the applicants' teachings encompass various alternatives, modifications, and equivalents, as will be appreciated by those skilled in the art.
In various embodiments, the distance between the sample plate and the inlet aperture can be 5 mm or greater. In various embodiments, the distance between the sample plate and the inlet aperture can be 6 mm or greater. In various embodiments, the distance between the sample plate and the inlet aperture can be 10 mm or greater. In various embodiments, the distance between the sample plate and the inlet aperture can be between 5 and 20 mm. In various embodiments, the distance between the sample plate and the inlet aperture can be between 12 and 20 mm. In various embodiments, the distance between the sample plate and the inlet aperture can be 16 mm.
In various aspects, the use of a high repetition laser can allow data to be acquired rapidly and a speed of approximately one second or less for each sample point on the sample plate can be achieved. With a high repetition laser, several thousand laser shots per sample can be obtained in a few seconds, which can dramatically improve overall precision and obtain high-throughput screening of samples. Typically, in such a high-throughput system, samples can be analyzed quicker and build up of contaminants can occur in a short period of time. For example, after analysis of approximately 30,000 samples in a prior art system, performance can be noticeably reduced such that cleaning of the system is necessary. When cleaning of the system is required, vacuum needs to be broken which can result in significant down time of the system. The applicants' teachings can reduce the frequency of system cleaning, and therefore down time, which is especially important for high-throughput applications.
In various embodiments, the sample can comprise a MALDI matrix, but a matrix-free sample can also be used, especially for samples in which the mass range is limited by matrix interference. An example of a matrix-free technique is desorption/ionization on silicon (DIOS) where analyte molecules are trapped within a porous silicon surface from which they are laser desorbed and ionized (Nature, 1999, 399, 243-246). The absence of matrix interference can allow for the analysis of small molecules below 300 m/z. Although a matrix-free sample can be used, neutrals as well as other contaminants can still be present and can form deposits that can contaminate the system.
The sample plate illustrated in the figures is representative of a typical plate, but the applicants' teachings are not limited to such a configuration. Other configurations of the sample plate as known in the art can be used. For example, the sample plate can contain topological features, as known in the art. The sample plate can be a curved, disc shape or can comprise other materials such as tape.
In various embodiments, an electrode can be a conducting element on which a potential is provided. An electrode can include, but is not limited to, a plate, ring, rod or tube.
In various embodiments, the mass analyzer can be, but is not limited to, a mass spectrometric instrument which can employ single MS, tandem (MS/MS) or multi-dimensional (MSn) mass spectrometry. Mass spectrometers can include, but are not limited to, a triple quadrupole, an ion trap, a hybrid linear ion trap, a time-of-flight, quadrupole time-of-flight, an RF multipole, a magnetic sector, an electrostatic sector, and an ion mobility spectrometer. Mass analyzers can include, but are not limited to, mass filters, mass selectors, ion focusing and/or ion steering elements, for example, ion guides. Mass analyzers also can include, but are not limited to ion reflectors and/or ion fragmentors, for example, collision cells, photodissociation cells, and surface dissociation fragmentors.
All literature and similar material cited in this application, including, but not limited to, patents, patent applications, articles, books, treatises, and web pages, regardless of the format of such literature and similar materials, are expressly incorporated by reference in their entirety. In the event that one or more of the incorporated literature and similar materials differs from or contradicts this application, including but not limited to defined terms, term usage, described techniques, or the like, this application controls.
While the applicants' teachings have been particularly shown and described with reference to specific illustrative embodiments, it should be understood that various changes in form and detail may be made without departing from the spirit and scope of the teachings. Therefore, all embodiments that come within the scope and spirit of the teachings, and equivalents thereto, are claimed. The descriptions and diagrams of the methods of the applicants' teachings should not be read as limited to the described order of elements unless stated to that effect.
While the applicants' teachings have been described in conjunction with various embodiments and examples, it is not intended that the applicants' teachings be limited to such embodiments or examples. On the contrary, the applicants' teachings encompass various alternatives, modifications, and equivalents, as will be appreciated by those of skill in the art, and all such modifications or variations are believed to be within the sphere and scope of the invention.
This application claims to the benefit of U.S. Provisional Patent Application Ser. No. 60/779,818 filed Mar. 7, 2006, incorporated herein by reference
Number | Date | Country | |
---|---|---|---|
60779818 | Mar 2006 | US |