Aspects of this invention are related to teleoperated minimally invasive surgical systems, and more particularly are related to controlling orientation of master and slave surgical instrument tips in a teleoperated minimally invasive surgical system.
The da Vinci® surgical system, manufactured by Intuitive Surgical, Inc., Sunnyvale, California, is a minimally invasive, teleoperated robotic system that offers patients many benefits, such as reduced trauma to the body, faster recovery and shorter hospital stay. One component of the da Vinci® Surgical System is a master tool manipulator that a surgeon uses to manipulate a surgical instrument, referred to as a slave surgical instrument.
The master grip of the master tool manipulator is specially designed to be both ergonomic and intuitive for controlling the slave surgical instrument. The surgeon holds the master grip in a particular way using his/her forefinger and thumb, so that targeting and grasping involves intuitive pointing and pinching motions.
To enable intuitive control of the slave surgical instrument, the master grip must be aligned in orientation with the slave surgical instrument tip in the view reference frame of the stereoscopic viewer. The motions of the slave surgical instrument tip follow master motions via teleoperation and are consistent in both directions of motion as well as absolute orientation. If orientation alignment is not achieved, the slave surgical instrument tip may still rotate in the desired direction, but the slave surgical instrument tip neither points in the same absolute direction nor rolls along the same axis as the surgeon is pointing.
The master tool manipulator uses motors in a gimbal assembly to actively align the orientation axes of the master grip with the associated slave surgical instrument tip in view coordinates. This alignment happens automatically before the surgeon engages teleoperation. Moreover, the system automatically preserves this alignment during manipulation of the camera or instrument outer axes.
Specifically, when entering following on the da Vinci® surgical system, the master grip must be aligned with the orientation of the slave surgical instrument tip before the da Vinci® surgical system operates properly in following. The present system performs a master alignment whenever the system transitions from a mode where this orientation alignment may have been compromised (after a tool change, camera clutch, slave clutch, swapping of arms in a 4th arm system etc.).
A master alignment calculates a set of master wrist joint angles that cause the orientation of the master grip to match the orientation of the slave surgical instrument tip, without changing the master grip position. The master wrist joints are then commanded to match the calculated angles using the motors.
The da Vinci® surgical system checks that the master and slave orientations match before allowing the user to enter following. If the orientations don't match (presumably because the user has over powered the master and not allowed the master to complete the alignment) a warning message is displayed and the master alignment is attempted again. This often slows down the surgeon's entry into following and requires a powered master tool manipulator with motors in the gimbal assembly to move master wrist joints into the proper orientation.
A minimally invasive surgical system includes a slave surgical instrument having a slave surgical instrument tip and a master grip. The master grip is coupled to the slave surgical instrument tip by a teleoperation servo control system. In one aspect, a ratcheting system, within the teleoperation servo control system, seamlessly and continuously improves the alignment, in a common coordinate frame of reference, of the master grip with respect to the slave surgical instrument tip as the master grip is moved.
The ratcheting system results in intuitive alignment between the master grip and the slave surgical instrument tip. Also, the ratcheting system provides a direct association between what the surgeon is doing, manipulating the master grip, and what the surgeon is seeing at a surgeon's console, movement of the slave surgical instrument tip in the display. This is achieved without the delays associated with the conventional powered alignments before following could be entered. Also, the ratcheting system, in one aspect, eliminates the need for the powered master grip and permits use of lower cost readily available components as the master grip.
In one aspect, the slave surgical instrument tip and the master grip have Cartesian position components and orientation components in a common coordinate reference frame. The ratcheting system can achieve alignment for all components, a subset of components, e.g., the orientation components or the Cartesian components, or in a component-wise fashion for a particular set of components.
A ratcheting system, in the teleoperation servo control system, is (i) coupled to the master grip to receive the alignment of the master grip in a common coordinate frame, and (ii) coupled to the slave surgical instrument. The ratcheting system controls the motion of the slave by continuously reducing the alignment error in the common coordinate frame, as the master grip moves. This is done without autonomous motion of the slave surgical instrument tip and without autonomous motion of the master grip.
In one aspect, the alignment error in the common coordinate frame is an orientation error. The orientation error comprises an angle. In one aspect, the angle is an angle θe defined as:
θe=cos−1[0.5*(RΔ11+RΔ22+RΔ33−1)],
where RΔ11, RΔ22, RΔ33 are diagonal elements of a relative rotation matrix RΔ. In another aspect, the angle is one Euler angle in a set of Euler angles.
In another aspect, the ratcheting system includes an introduce control loop error controller that introduces a control system error in the alignments of the slave surgical instrument tip and the master grip in the common frame of reference. For example, an angular velocity of the slave surgical instrument is penalized based upon an amount of misalignment in the common reference frame between the slave surgical instrument tip and the master grip. In still another aspect, the ratcheting system includes a roll-joint limit controller for determining whether motion of the slave surgical instrument has reached a joint limit and constraining motion of the slave surgical instrument tip upon reaching the joint limit. In still yet another aspect, the ratcheting system includes multiple orientation solutions, a best orientation controller selects one solution from the multiple orientation solutions that is closest to the orientation of the master grip.
A method using the above minimally surgical system includes controlling motion of a slave surgical instrument tip in the minimally invasive surgical system with motion of a master grip in the minimally invasive surgical system. In this method, a ratcheting system in the minimally invasive surgical system receives an alignment of the slave surgical instrument tip and an alignment of the master grip, both in a common frame of reference. The ratcheting system ratchets the alignment of the slave surgical instrument tip to the alignment of the master grip by continuously reducing an alignment error between the alignments, as the master grip moves, without autonomous motion of the slave surgical instrument tip and without autonomous motion of the master grip.
In one aspect, the ratcheting includes moving the slave surgical instrument tip and the master grip in a same relative way upon engagement of following by the minimally invasive surgical system irrespective of the alignments in the common reference frame. The ratcheting also includes bleeding off the alignment error in the common frame of reference by using a new alignment error when the motion of the master grip reduces the alignment error in the common reference frame, and by using a current alignment error in the common reference frame when the motion of the master grip does not reduces the alignment error thereby continuously reducing the alignment error between the alignments in the common reference frame.
The ratcheting system also generates a new alignment for the slave surgical instrument using the new alignment error, and sends a command to the slave surgical instrument based on the new alignment.
In another aspect, the ratcheting system introduces a control error between the slave surgical instrument alignment and the master grip alignment. The introduction of a control error further includes penalizing a commanded angular velocity for the slave surgical instrument tip based on the current alignment error.
In the drawings, the first digit of a figure number indicates the figure in which the element with that figure number first appeared.
Aspects of this invention replace the powered master tool manipulator, such as that used in the da Vinci® Surgical Robot System manufactured by Intuitive Surgical, Inc. of Sunnyvale, California with an unpowered master tool manipulator that includes at least one master grip. In minimally invasive surgical system 100, a surgeon, at a console 114, grasps a master grip (not shown) between the thumb and forefinger so that targeting and grasping still involves intuitive pointing and pinching motions. The motion of the master grip is used by control system 150, as described more completely below, to move an end-effector of slave surgical instrument 112.
Unlike the conventional system that required the surgeon to wait until the alignment of the master grip and the slave surgical instrument end-effector, in a common frame of reference, were positioned so that following could be entered, a ratcheting system 130, within a teleoperation servo control system 160 of control system 150, is activated when the surgeon starts to move the master grip. Irrespective of the alignment between the master grip and the end-effector of slave surgical instrument 112 in the common frame of reference, teleoperation servo control system 160 enters following between the master grip and the surgical instrument end-effector, sometimes called a slave surgical instrument tip, and activates ratcheting system 130.
Ratcheting system 130 seamlessly and continuously improves the alignment of the master grip with respect to the slave surgical instrument tip, in the common frame of reference, as the master grip is moved. Ratcheting system 130 ratchets the movement of the slave surgical instrument tip to continuously and seamlessly reduce any alignment error, in the common frame of reference, between the slave surgical instrument tip and the master grip. Ratcheting system 130 achieves the alignment without autonomous motion of either the master grip, or the slave surgical instrument tip.
Ratcheting system 130 results in intuitive alignment between the master grip and the slave surgical instrument tip as viewed by the surgeon. Also, ratcheting system 130 provides a direct association between what the surgeon is doing, manipulating the master grip, and what the surgeon is seeing at console 114, movement of the slave surgical instrument tip in the display. This is achieved without the delays associated with the conventional powered alignments before following could be entered. Also, ratcheting system 130 eliminates the need for the powered master grip and permits use of lower cost readily available components as the master grip.
Console 114 (
Console 114 (
The surgeon sits comfortably and looks into the master display on console 114 throughout surgery. The surgeon performs a medical procedure by manipulating at least master grip 230 (
The master display is positioned on console 114 (
The real-time image from the endoscope is projected into perspective image 210 such that the surgeon can manipulate a surgical instrument end-effector of tool 112, through its associated master grip 230, as if viewing the workspace in substantially true presence. By true presence, it is meant that the presentation of an image is a true perspective image simulating the viewpoint of an operator that is physically manipulating the surgical instrument. Thus, control system 150 transforms the coordinates of surgical instrument 112 to a perceived position so that the perspective image is the image that the surgeon would see if the endoscope were looking directly at surgical tool 112 from the surgeon's eye-level during an open cavity procedure.
Control system 150 performs various functions in system 100. Control system 150 receives the images from an endoscope and generates the stereoscopic image that the surgeon sees. In a conventional manner, control system 150 maps the slave alignment with respect to tip of the endoscope and maps the master alignment with respect to an aspect of the surgeon into the common frame of reference, sometimes called common reference frame, which is used by ratcheting system 130. See for example, U.S. Pat. No. 6,424,885, entitled “Camera Referenced Control in a Minimally invasive Surgical Apparatus,” of Niemeyer et al. issued on Jul. 23, 2002, which is incorporated herein by reference in its entirety.
Control system 150 uses teleoperation servo control system 160 to translate and to transfer the mechanical motion of master grip 230 to an associated robotic arm 113 through control commands 135 so that the surgeon can effectively manipulate slave surgical instrument 112. The functions performed by teleoperation servo control system 160 are equivalent to the conventional functions when considered in conjunction with ratcheting system 130 that is also included in control system 150.
The number of surgical tools used at one time and consequently, the number of robotic arms being used in system 100 generally depends on the medical procedure to be performed and the space constraints within the operating room, among other factors. If it is necessary to change one or more of the tools being used during a procedure, an assistant may remove the tool no longer being used from its robot arm, and replace the tool with another tool from a tray in the operating room.
Although described as a control system 150, it is to be appreciated that control system 150 may be implemented in practice by any combination of hardware, software that is executed on a processor, and firmware. Also, its functions, as described herein, may be performed by one unit, or divided up among different components, each of which may be implemented in turn by any combination of hardware, software that is executed on a processor, and firmware. When divided up among different components, the components may be centralized in one location or distributed across system 100 for distributed processing purposes.
When there are only two master grips in system 100, and when the surgeon wants to control movement of a slave surgical instrument different from the two slave surgical instruments coupled to the two master grips, the surgeon may lock one or both of the two slave surgical instruments in place. The surgeon then associates one or both of the master grips with other slave surgical instruments held by other of the robotic arms and ratcheting system 130 becomes active with respect to those instruments.
Master grip 230 provides an alignment that is mapped into the common frame of reference, but is not powered, or at least does not include a powered wrist. However, the features of master grip 230, as described more completely below, can be used with a master grip that includes a powered wrist. As used herein, a powered wrist means a wrist that includes at least one motor to control the positioning of the wrist. The motor can be used to compensate for gravity and friction and to provide range of motion force-feedback, but any powered alignment capability is turned-off.
Master grip 230 includes two levers 231, 232, which the surgeon typically grasps between the thumb and forefinger. As the surgeon moves master grip 230, in one aspect, master grip alignment information 132 is provided to ratcheting system 130 in the common reference frame(See
Thus, in one aspect of method 300 (
A receive new alignment of master and slave operation 310, associated with receive orientation module 371, receives the master grip alignment and slave surgical instrument tip alignment in the common frame of reference. Ratcheting system 130 waits in following check operation 320 until following is initiated. Following check operation 320 should not be interpreted as requiring continuous polling, but rather simply interpreted as nothing is done with respect to the alignments in the common frame of reference until following is initiated.
Check operation 320 is used for illustration only and should not be viewed as limiting. The particular technique used to determine whether following is initiated could be based on an interrupt, an event, a particular flag or bit changing state, etc.
Upon engagement of following between master grip 230 and slave surgical instrument tip 212T, ratchet alignment process 330, which is associated with ratchet alignment module 372, generates commands that result in slave surgical instrument tip 212T moving in the same relative way as master grip 230 is moved irrespective of any alignment error in the common frame of reference between master grip 230 and slave surgical instrument tip 212T.
Note that for convenience, slave surgical instrument tip 212T is used in this description, as this is what the surgeon sees moving. The movement of this image corresponds directly to the movement of the slave surgical instrument tip itself. One knowledgeable in the field understands that movement of the image is a direct result of movement of the tip itself by the robot arm in response to a command from control system 150, as described herein.
As both master grip 230 and slave surgical instrument tip 212T move, ratchet alignment process 330 bleeds off any alignment error in the common reference frame between master grip 230 and slave surgical instrument tip 212T. For example, when the surgeon moves master grip 230 in a way that reduces the alignment error, ratchet alignment process 330 uses the reduced alignment error in the following between master grip 230 and slave surgical instrument tip 212T. Conversely, when the surgeon moves master grip 230 in a way that increases the alignment error, ratchet alignment process 330 uses the current alignment error, and not the increased alignment error, in the following between master grip 230 and slave surgical instrument tip 212T.
Thus, ratchet alignment process 330 seamlessly and continuously improves the absolute alignment of master grip 230 with respect to slave surgical instrument tip 212T. Ratchet alignment process 330 achieves the continuous improvement in absolute alignment without causing autonomous motion of either master grip 230 or slave surgical instrument tip 212T.
Ratcheting system 130 can ratchet an alignment error in the common reference frame that is either a position error or an orientation error. In the following examples, ratcheting of an orientation error in a common frame of reference is considered. However, the orientation error examples are illustrative only and are not intended to be limiting to the specific aspects described. In view of this disclosure, one knowledgeable in the field can apply the same principles to ratchet a position error.
In the following examples, as the surgeon moves master grip 230, in one aspect, a master grip orientation 235 in the common frame of reference is provided to ratcheting system 130. In this aspect, slave surgical instrument tip 212T has a slave orientation 225 in the common frame reference, which is equivalent to the orientation of the slave surgical instrument tip itself. Ratcheting system 130 also receives this slave orientation in the common frame of reference. In this aspect, the z-axis of master grip 230 is on the roll axis and points away from the hand. The corresponding z-axis on the slave is, for example, along the centerline of the jaws and indicates where the jaws are pointing.
In one aspect, orientation 235 of master grip 230 in the common frame of reference is defined in teleoperation servo control system 160 by a master grip rotation matrix Rm, which in this example is a three-by-three matrix. Similarly, orientation 225 of slave surgical instrument tip 212T in the common frame of reference is defined in teleoperation servo control system 160 by a slave surgical instrument tip rotation matrix Rs, which in this example also is a three-by-three matrix. Master grip rotation matrix Rm and slave surgical instrument tip rotation matrix Rs are orthonormal rotation matrices.
With these matrix representations of orientations 235 and 225, the orientation error in the common frame of reference is a relative rotation matrix RΔ. Relative rotation matrix RΔ is defined as:
RΔ=RST*R (1)
where matrix RST is the transpose of slave surgical instrument tip rotation matrix Rs. It follows from the definitions of matrix RST and matrix Rm, taken with definition (1) that relative rotation matrix RΔ is a three-by-three matrix:
When orientation 235 of master grip 230 and orientation 225 of slave surgical instrument tip 212T are aligned in the common frame of reference, relative rotation matrix RΔ is an identity matrix, i.e., diagonal elements RΔ11, RΔ22, RΔ33 have a value of one and all other elements have a value of zero.
With these definitions of master grip rotation matrix Rm, slave surgical instrument tip rotation matrix Rs, and relative rotation matrix RΔ, subsequent master grip orientations are mapped to corresponding slave orientation commands, in the common frame of reference, via:
Rs=Rm*RΔT (2)
where matrix RΔT is the transpose of relative rotation matrix RΔ.
As described above, with a current relative rotation matrix RΔ−Current and a new relative rotation matrix RΔ−New, a decision must be made on which of the two relative rotation matrices to use in the mapping to the corresponding slave orientation command. New relative rotation matrix RΔ−New is used when
RΔ−New<RΔ−Current
However, it has been recognized that a scalar quantity can be used to determine which relative rotation matrix to use. Thus, a technique for comparing multi-dimension matrices is not needed to determine when new relative rotation matrix RΔ−New is less than current relative rotation matrix RΔ−Current.
As illustrate in
Rotation angle error θe is defined as:
θe=cos−1[0.5*(RΔ11+RΔ22+RΔ33−1)] (3)
Recall, as described above, when master grip orientation 235 and slave surgical instrument tip orientation 225 are aligned, diagonal elements RΔ11, RΔ22, RΔ33 have a value of one. Thus, when the two orientations are aligned, rotation angle error θe is zero.
In this aspect, ratchet alignment process 330 for ratcheting system 130 is implemented as ratchet orientation process 330A (
Upon entering following, generate new orientation error process 410, which is associated with orientation error generation module 481, accesses master grip rotation matrix Rm and slave surgical instrument tip rotation matrix Rs.
Next, process 410 uses master grip rotation matrix Rm and slave surgical instrument tip rotation matrix Rs to generate a new relative rotation matrix RΔ−New based on definition (1) above. Here, slave surgical instrument tip rotation matrix Rs is the current orientation of slave surgical instrument tip 212T and is stored for example, in slave orientation element 131 in memory 404 (
Orientation error reduced check operation 420 determines whether new rotation angle error θe−New is less than current rotation angle error θe−Current, e.g., check operation 420 is implemented as a comparator. If new rotation angle error θe−New is equal to or greater than current rotation angle error θe−Current, check operation 420 passes to set orientation operation 440. Conversely, if new rotation angle error θe−New is less than current rotation angle error θe−Current, check operation passes processing to set orientation operation 430.
The use of two set orientation operations 430, 440 is for ease of illustration only and is not intended to be limiting. A single set orientation operation could be used for the instances in which rotation angle error θe is reduced, for example. Also, note that in the first pass through ratchet orientation process 330A, check operation 420 may simply pass processing to set orientation error operation 430. Alternatively, in initialization, current rotation angle error θe−Current can be set to a large value so that new rotation angle error θe−New is always less than current rotation angle error θe−Current in the first pass. The particular aspect used to handle the first pass through process 330A is not critical so long as no autonomous motion is introduced.
Set orientation error operation 440 is effectively a no-op, because current rotation angle error θe−Current and current relative rotation matrix RΔ−Current are maintained as the current rotation angle error and the current relative rotation matrix, respectively. Thus, operation 440 can be removed and processing simply transfers from check operation 420 directly to generate new slave orientation 450.
Set orientation error operation 430, which is associated with comparator module 482, sets current rotation angle error θe−Current to new rotation angle error θe−New. Operation 430 also sets current relative rotation matrix RΔ−Current to new relative rotation matrix RΔ−New and then transfers to generate new slave orientation operation 450.
Generate new slave orientation 450, which is associated with slave generation orientation module 483, first generates a transpose matrix RΔT of current relative rotation matrix RΔ−Current. Next, a new slave orientation matrix Rs is obtained by combining transpose matrix RΔT and master grip rotation matrix Rm according to definition (2), above. Thus, operation 450 transforms the master orientation into a slave orientation.
Generate new slave orientation 450 transfers processing to send slave orientation command operation 460, which is associated with slave command generation module 484. Using new slave orientation Rs, operation 460 sends a command, via slave input/output (I/O) module 403, which results in slave surgical instrument tip 212T being moved. The movement depends upon which relative rotation matrix RΔ was used in the current iteration of process 330A.
In process 330A, When the surgeon moves master grip 230 in a way that reduces the alignment error in the common reference frame between master grip 230 and slave surgical instrument tip 212T, e.g., reduces rotation angle error θe, ratchet alignment process 330A (
Thus, ratchet alignment process 330A seamlessly and continuously improves the absolute alignment of master grip 230 with respect to slave surgical instrument tip 212T. Ratchet alignment process 330A achieves the continuous improvement in absolute alignment without autonomous motion of either master grip 230 or slave surgical instrument tip 212T.
If master grip 230 and the slave surgical instrument tip were rigidly connected, relative rotation matrix RΔ would be a constant. In reality, master grip 230 and the slave surgical instrument tip are not rigidly connected, but are coupled through teleoperation servo control system 160. Whenever master grip 230 and the slave surgical instrument tip are in motion, there is always a small amount of control error. This control error can be absorbed using process 330A and alignment obtained between master grip 230 and slave surgical instrument tip 212T.
However, this approach is highly dependent on the stiffness of the servo control loop. Under normal conditions in a teleoperated minimally invasive surgical system with a stiff servo control loop, the incremental improvements may be too small and thus take a long time to reach satisfactory alignment. Even in such systems, ratcheting system 130 is still used, but an introduce control system error controller 560 (
Techniques for introducing such a subtle control error by introduce control system error controller 560 include, but are not limited to:
In this example, introduce control system error operation 510 in controller 560 introduces the control system error after generation of new rotation angle error θe−New in process 410. This location in process 330B is illustrative only and is not intended to be limiting to this specific location in process 330B. Upon selection of a particular technique to introduce control system error, introduce control system error operation 510 can be placed at an appropriate position within process 330B. Also, an introduce control system error module is included within ratchet orientation module 472 (
In one aspect, operation 510 introduces control error by artificially adjusting the commanded angular velocity of the slave surgical instrument tip based on current rotation angle error θe−Current. Many different techniques can be used for penalizing the commanded angular velocity.
In one aspect of operation 510, current rotation angle error θe−Current is used in combination with an angular velocity penalty profile to generate an angular velocity scale factor for penalizing the commanded angular velocity of the slave instrument tip. In this aspect the angular velocity scale factor varies between zero for unacceptable rotation angle errors and one for a rotation angle error of zero.
The angular velocity may be penalized using an angular velocity penalty profile which smoothly varies between zero and one as rotation angle error θe changes. More generally, any continuous and monotonic angular velocity penalty profile can be used so long as the profile does not introduce autonomous motion and does not produce unexpected movement of the slave tip. As used here, continuous means that the first derivative of the profile exists at any point in the profile.
In one aspect, an angular velocity penalty profile is selected based on the experience of the surgeon. In another aspect, a single angular velocity penalty profile is used for all surgeons. In still yet another aspect, the angular velocity penalty profile is surgeon specific.
One angular velocity penalty profile suitable for use is a sigmoid shape curve.
In this example, first inflection point θe1 is placed at 30 degrees and second inflection point θe2 is placed at 90 degrees. Angle θe0 is the angle at which the weight, more specifically referred to as the angular velocity scale factor, is a one, while angle θe3 is the angle at which the weight becomes zero and remains zero for any larger misalignments.
The misalignment angle associated with inflection point θe1 is selected to represent the user tolerance to misalignment, whereas the misalignment angle associated with inflection point θe2 is selected to represent a maximum usable misalignment angle so that angles larger than inflection point θe2 are considered unusable alignments. Whenever rotation angle error θe is greater than inflection point θe2, the commanded angular velocity of the slave surgical instrument tip is scaled towards zero, which results in little or no motion of the slave surgical instrument tip. This allows relative rotation matrix RΔ to be readily updated as the operator rotates master grip 230 towards slave surgical instrument tip 212T.
As rotation angle error θe decreases towards inflection point θe1, the commanded angular velocity is gradually increased, allowing for more responsive motion of the slave surgical instrument tip while still inducing enough control error to further improve relative rotation matrix RΔ. Once rotation angle error θe becomes less than inflection point θe1, slave surgical instrument tip 212T is perceived by the surgeon to be well aligned with master grip 230. Therefore, the angular motion of the slave surgical instrument tip 212T is allowed to be more responsive to the motion of master grip 230. Additional small alignment corrections continue to be made until relative rotation matrix RΔ becomes the identity matrix and rotation angle error θe goes to zero.
With respect to artificial joint limits, artificial joint limits on slave motion are defined as a function of the alignment error in the common frame of reference, e.g., as a function of rotation angle error θe. When the alignment error is large, e.g., greater than ninety degrees, artificial joint limits are imposed on motion of slave surgical instrument tip 212T that allow instrument tip 212T to make small motions about the current position of instrument tip 212T.
As master grip 230 moves towards slave surgical instrument tip 212T, slave surgical instrument tip 212T moves in the same relative direction until an artificial joint limit is encountered. When an artificial joint limit is encountered, slave surgical instrument tip 212T remains fixed in position at the artificial joint limit. This allows ratcheting system 130 to aggressively reduce the alignment error as master grip 230 continues to move in the direction of slave surgical instrument tip 212T.
In one aspect, the artificial joint limits are gradually widened as the alignment error decreases. This makes the alignment process more seamless and continuous.
If, for example, low-pass filtering of the sensed master grip rotation matrix Rm is used to introduce a control system error, master grip rotation matrix Rm changes at a lower frequency. For example, humans can make controlled motions up to around 20 Hz. The master orientation can be aggressively filtered down, e.g. to one to five Hertz, to make the orientation changes more sluggish when alignment is poor. Again, a monotonic penalty profile could be used to move between zero Hertz and twenty Hertz as a function of the current orientation error.
Conversely, if, for example, low-pass filtering of commanded slave surgical instrument tip rotation matrix Rs is used to introduce a control system error, slave surgical instrument tip rotation matrix Rs changes at a lower frequency. In view of these examples, one knowledgeable in the field can select an appropriate low-pass filter so that the ratcheting seamlessly and continuously improves the absolute alignment of master grip 230 with respect to slave surgical instrument tip 212T. Ratchet alignment process 330B achieves the continuous improvement in absolute alignment without autonomous motion of either master grip 230 or slave surgical instrument tip 212T.
In another aspect, method 600 (
Receive new orientation of master and slave operation 310A, associated with receive orientation module 471A, receives master grip orientation 235 and slave surgical instrument tip orientation 225. Operation 310A transfers to check operation 320A.
System following check operation 320A determines whether system following has been initiated. If system following has not been initiated, save master orientation operation 610 saves the received master orientation in old master orientation storage location 632 in memory 404A and processing is returned from the ratcheting system. Thus, until following is initiated, the ratcheting system maintains the most recent orientation of master grip 230 in the common frame of reference.
When system following is initiated, check operation 320A transfers to generate new orientation error matrix operation 615, which is associated with orientation error generation module 481A (
Best solution check operation 620 determines which of a set of multiple solutions for the orientation best matches the current orientation of master grip 230 in the common frame of reference. For example, with respect to the orientation of how the surgeon grasped master grip 230, two possible solutions are considered, right side up and upside down (USD). Thus, in one aspect, check operation 620 is implemented in a multiple orientation solution controller 621 that includes an upside down (USD) controller.
Assuming the surgeon used her/his right hand to grasp master grip 230, when the thumb is on lever 231 (
Thus, using definition (3) above, in one aspect, check operation 620 performs the following operations:
cos(θe_s)=[0.5*(RΔ11+RΔ22+RΔ33−1)]
cos(θe_f)=[0.5*(−1*RΔ11−RΔ22+RΔ33−1)]
The orientation in the two dimensions is upside down when cos(θe_f) is greater than cos(θe_s) and upside-down orientation flag USD is set to true. Conversely, when cos(θe_f) is less than or equal to cos(θe_s), upside-down orientation flag USD is set to false.
Generate new rotation angle error θe operation 625 and generate current rotation angle error θe operation 630 are associated with orientation error generation module 481A. Operations 625 and 630 generate a new rotation angle error θe−New and a current rotation angle error θe−Current using the state of upside down orientation flag USD, a new relative rotation matrix RΔ−New, and a current relative rotation matrix RΔ−Current.
Specifically, if upside-down orientation flag USD is set to true, operation 625 first rotates current relative rotation matrix RΔ−Current in the two dimensions, and then determines current rotation angle error θe−Current, as explained above. If upside-down orientation flag USD is set to false, operation 625 determines current rotation angle error θe−Current without any rotation. Current rotation angle error θe−Current is checked and (1) if the value is greater than pi, current rotation angle error θe−Current is set to pi, or (2) if the value is less than zero, current rotation angle error θe−Current is set to zero. Thus, current rotation angle error θe−Current is bounded between zero and pi.
Similarly, if upside-down orientation flag USD is set to true, operation 630 first rotates new relative rotation matrix RΔ−New in the two dimensions, and then determines new rotation angle error θe−New, as explained above. If upside-down orientation flag USD is set to false, operation 630 determines new rotation angle error θe−New without any rotation. New rotation angle error θe−New is checked and (1) if the value is greater than pi, new rotation angle error θe−New is set to pi, or (2) if the value is less than zero, new rotation angle error θe−New is set to zero. Thus, new rotation angle error θe−New also is bounded between zero and pi.
Following the determination of the rotation angle errors in operations 625 and 630 based on the best solution, a control system error is introduced in generate penalized angular velocity operation 510A, which is associated with control system error introduction module 682 in ratchet orientation module 672 (
In this aspect, operation 510A is implemented in an introduce control system error controller 560A, which in turn is associated with control system error introduction module 682. Penalizing the angular velocity is an example of a one type of control system error that can be introduced by introduce control system error controller 560A. In general, introduce control system error controller 560A introduces a control error, such as those described above.
In this aspect of generate penalized angular velocity operation 510A, the angular velocity is penalized by multiplying the angular velocity sensed from master grip 230 by the angular velocity scale factor from penalty curve 500 based on current rotation angle error θe−Current, as described above. Following completion of operation 510A, processing transfers to joint limit control process, which is implemented in a roll-joint limit controller 635. Roll-joint limit controller 635 is associated with joint limit control module 683 in ratchet orientation module 672.
Roll-joint limit controller 635 determines whether a slave surgical instrument joint has reached a limit in a given direction and whether the sensed motion from master grip 230 is trying to move the slave surgical instrument tip in that given direction. In this aspect, slave roll limit check operation 640, in roll-joint limit controller 635, determines whether the roll joint in the slave surgical instrument has reached a roll limit, and if the sensed motion from master grip 230 is trying to rotate slave surgical instrument tip 212T in that roll direction.
In check operation 640, the upper and lower roll limits are predefined. Old master grip rotation matrix Rm−old, which was stored in old master orientation location 632, is transposed. A relative master grip rotation matrix RmΔ is defined as:
RmΔ=Rm−oldT*R
Thus, relative master grip rotation matrix RmΔ is a three-by-three matrix:
The master grip roll direction is defined as:
Master_roll_dir=[RΔ21−RΔ12].
Next, check operation 640 determines first whether the position of slave surgical instrument tip 212T is less than the lower roll joint limit plus a tolerance and master roll direction Master_roll_dir is less than zero. If the result of this first determination is true, the lower roll joint limit has been reached and the direction of the master grip rotation is in the direction of that lower roll limit. Thus, the slave surgical instrument tip motion is constrained.
If the result of this first determination is false, check operation 640 determines second whether the position of the slave surgical instrument tip is greater than the upper roll joint limit minus a tolerance and master roll direction Master_roll_dir is greater than zero. If the result of this second determination is true, the upper roll joint limit has been reached and the direction of the master grip rotation is in the direction of that upper roll limit. Thus, the slave surgical instrument motion is constrained.
When the slave surgical instrument motion is constrained, slave roll limit check operation 640 transfers to set current rotation angle error and angular velocity operation 645, and otherwise to rotation angle error reduced check operation 420A. Set current rotation angle error and angular velocity operation 645 sets current rotation angle error θe−Current to a value larger than any possible physical value, and sets the commanded angular velocity to zero. Set current rotation angle error and angular velocity operation 645 transfers processing to orientation error reduced check operation 420A.
Orientation error reduced check operation 420A determines whether new rotation angle error θe−New is less than current rotation angle error θe−Current, e.g., check operation 420 is implemented as a comparator. If new rotation angle error θe−New is equal to or greater than current rotation angle error θe−Current, check operation 420A passes to save master orientation operation 650. Conversely, if new rotation angle error θe−New is less than current rotation angle error θe−Current, check operation 420A passes processing to set orientation error operation 430A.
Set orientation error operation 430A, which is associated with comparator module 482A, sets current rotation angle error θe−Current to new rotation angle error θe−New. Operation 430A also sets current relative rotation matrix RΔ−Current to new relative rotation matrix RΔ−New and then transfers to save master orientation operation 650, which saves master grip rotation matrix Rm. in element 632 (
Generate new slave orientation 450B, which is associated with slave generation orientation module 483A (
Generate new slave orientation 450B transfers processing to send slave orientation command operation 460B (
When the surgeon moves master grip 230 in a way that reduces the alignment error between master grip 230 and slave surgical instrument tip 212T, e.g., reduces rotation angle error θe, ratchet alignment process 330C (
Thus, ratchet alignment process 330C seamlessly and continuously improves the absolute alignment of master grip 230 with respect to slave surgical instrument tip 212T. Ratchet alignment process 330C achieves the continuous improvement in absolute alignment without autonomous motion of either master grip 230 or slave surgical instrument tip 212T.
In method 600, roll-joint limit controller 635 was used to handle a corner-case for a slave range of motion. This corner-case is considered further with respect to
If the surgeon continued to rotate master grip 230 in the counter clockwise direction back to zero degrees, slave tip 792 suddenly might move to the orientation of slave tip 790, e.g., there would be autonomous motion. However, with corner-case joint limits on the motion of the slave surgical instrument tip such motion is inhibited. For example, when the orientation of master grip 230 comes within a predetermined tolerance of a joint limit, the commanded angular velocity of the slave surgical instrument tip is set to zero and current rotation angle error θe−Current is set to an artificially large number so that new relative rotation matrix RΔ−New is used. This essentially allows the master orientation to slip relative to the orientation of the slave surgical instrument tip when the slave surgical instrument tip is at a roll limit.
With a conventional powered master, the user would receive force feedback to indicate when a slave has reached a range of motion limit. With an unpowered master, a roll-joint limit controller, in the ratcheting system, can be used to maintain continuity of slave motion as the master drives the slave in and out of a joint limit.
In the above methods, various modules and operations and/or processes associated with those modules were described. The modules may be implemented in hardware, software that is executed on a processor, firmware or any combination of hardware, software or firmware.
When the modules include one or more instructions stored on a storage medium, the described operations and/or processes are the result of retrieval and execution of the one or more instructions on at least one processor in processor module 401 to obtain the transformation described. The particular modules described are illustrative only and are not intended to be limiting. In view of the disclosure, one knowledgeable in the field can combine modules together or separate a module into one or more additional modules as may be desired. Moreover while the modules are shown grouped in a common location this also is illustrative only.
TABLE 1 is an example of software instructions written in the C-programming language that when executed on a processor perform elements 302A, 610, 615, 620, 621, 625, 630, 510A, 560A, 635, 640, 645, 420A, 430A and 650.
In the examples of
The above description and the accompanying drawings that illustrate aspects and embodiments of the present inventions should not be taken as limiting—the claims define the protected inventions. Various mechanical, compositional, structural, electrical, and operational changes may be made without departing from the spirit and scope of this description and the claims. In some instances, well-known circuits, structures, and techniques have not been shown or described in detail to avoid obscuring the invention.
Further, this description's terminology is not intended to limit the invention. For example, spatially relative terms—such as “beneath”, “below”, “lower”, “above”, “upper”, “proximal”, “distal”, and the like—may be used to describe one element's or feature's relationship to another element or feature as illustrated in the figures. These spatially relative terms are intended to encompass different positions and orientations of the device in use or operation in addition to the position and orientation shown in the figures. For example, if the device in the figures is turned over, elements described as “below” or “beneath” other elements or features would then be “above” or “over” the other elements or features. Thus, the exemplary term “below” can encompass both positions and orientations of above and below. The device may be otherwise oriented (rotated 90 degrees or at other orientations) and the spatially relative descriptors used herein interpreted accordingly.
The singular forms “a”, “an”, and “the” are intended to include the plural forms as well, unless the context indicates otherwise. The terms “comprises”, “comprising”, “includes”, and the like specify the presence of stated features, steps, operations, elements, and/or components but do not preclude the presence or addition of one or more other features, steps, operations, elements, components, and/or groups.
All examples and illustrative references are non-limiting and should not be used to limit the claims to specific implementations and embodiments described herein and their equivalents. The headings are solely for formatting and should not be used to limit the subject matter in any way, because text under one heading may cross reference or apply to text under one or more headings. Finally, in view of this disclosure, particular features described in relation to one aspect or embodiment may be applied to other disclosed aspects or embodiments of the invention, even though not specifically shown in the drawings or described in the text.
While the memory is illustrated as a unified structure, this should not be interpreted as requiring that all memory is at the same physical location. All or part of the memory can be in a different physical location than a processor. Memory refers to a volatile memory, a non-volatile memory, or any combination of the two.
A processor is coupled to a memory containing instructions executed by the processor. This could be accomplished within a computer system, or alternatively via a connection to another computer via modems and analog lines, or digital interfaces and a digital carrier line.
Herein, a computer program product comprises a medium configured to store computer readable code needed for any one or any combination of methods 300, 330, 330A, 330B, 600, 330C or in which computer readable code for any one or any combination of methods 300, 330, 330A, 330B, 600, 330C is stored. Some examples of computer program products are CD-ROM discs, DVD discs, flash memory, ROM cards, floppy discs, magnetic tapes, computer hard drives, servers on a network and signals transmitted over a network representing computer readable program code. A tangible computer program product comprises a medium configured to store computer readable instructions for any one of, or any combination of methods 300, 330, 330A, 330B, 600, 330C or in which computer readable instructions for any one of, or any combination of methods 300, 330, 330A, 330B, 600, 330C is stored. Tangible computer program products are CD-ROM discs, DVD discs, flash memory, ROM cards, floppy discs, magnetic tapes, computer hard drives and other physical storage mediums.
In view of this disclosure, instructions used in any one of, or any combination of methods 300, 330, 330A, 330B, 600, 330C can be implemented in a wide variety of computer system configurations using an operating system and computer programming language of interest to the user.
Further, various different minimally invasive systems and methods can be implemented in view of this disclosure.
In one aspect, a minimally invasive surgical system comprises:
The ratcheting system further includes an introduce control system error controller for introducing a control system error with respect to the alignments of the slave surgical instrument tip and the master grip.
The introduce control system error controller includes, but is not limited to, any one of or any combination of:
In another aspect, the ratcheting system includes a multiple orientation solution controller for determining an orientation solution closest to an orientation of the master grip. The multiple orientation controller can be included in any combination of controllers, including but not limited to the introduce control system error controller. In one aspect, the multiple orientation solution controller includes an upside down controller.
In another aspect, a method for controlling alignment of a slave surgical instrument tip in a minimally invasive surgical system with alignment of a master grip in the minimally invasive surgical system comprises:
In one aspect, the ratcheting further comprises:
The introducing a control system error includes, but is not limited to, any one of or any combination of:
In still another aspect, the ratcheting further comprises selecting an orientation solution from a group of orientation solutions based on an orientation, in the common frame of reference, of the master grip. In one aspect, the selecting an orientation solution is in combination with the introducing a control system error. In another aspect, the selecting an orientation solution includes selecting an upside down solution.
This application is a continuation of U.S. patent application Ser. No. 17/107,862 (filed Nov. 30, 2020), which is a continuation of U.S. patent application Ser. No. 16/862,412 (filed Apr. 29, 2020) and now 10,881,473, which is a continuation of U.S. patent application Ser. No. 16/353,932 (filed Mar. 14, 2019) and now U.S. Pat. No. 10,675,109, which is a continuation of U.S. patent application Ser. No. 15/706,883 (filed Sep. 18, 2017 and now U.S. Pat. No. 10,278,783), which is a continuation of U.S. patent application Ser. No. 15/399,600 (filed Jan. 5, 2017 and now U.S. Pat. No. 9,814,537), which is a continuation of U.S. patent application Ser. No. 14/996,073 (filed Jan. 14, 2016 and now U.S. Pat. No. 9,579,164), which is a continuation of U.S. patent application Ser. No. 14/534,526 (filed Nov. 6, 2014 and now U.S. Pat. No. 9,265,584), which is a continuation of U.S. patent application Ser. No. 13/839,438 (filed Mar. 15, 2013 and now U.S. Pat. No. 8,903,549), which is a continuation of U.S. patent application Ser. No. 12/495,213 (filed Jun. 30, 2009 and now U.S. Pat. No. 8,423,186), the full disclosures of each of which are incorporated by reference herein for all purposes.
Number | Name | Date | Kind |
---|---|---|---|
6364888 | Niemeyer | Apr 2002 | B1 |
6424885 | Niemeyer et al. | Jul 2002 | B1 |
6594552 | Nowlin et al. | Jul 2003 | B1 |
6630993 | Hedges et al. | Oct 2003 | B1 |
6766204 | Niemeyer et al. | Jul 2004 | B2 |
6786896 | Madhani et al. | Sep 2004 | B1 |
6853879 | Sunaoshi | Feb 2005 | B2 |
7126303 | Farritor et al. | Oct 2006 | B2 |
7339341 | Oleynikov et al. | Mar 2008 | B2 |
7492116 | Oleynikov et al. | Feb 2009 | B2 |
7656106 | Iwashita | Feb 2010 | B2 |
8423186 | Itkowitz | Apr 2013 | B2 |
8903549 | Itkowitz et al. | Dec 2014 | B2 |
8924021 | Dariush et al. | Dec 2014 | B2 |
9265584 | Itkowitz et al. | Feb 2016 | B2 |
9579164 | Itkowitz et al. | Feb 2017 | B2 |
9795453 | Tierney | Oct 2017 | B2 |
9814537 | Itkowitz et al. | Nov 2017 | B2 |
10278783 | Itkowitz et al. | May 2019 | B2 |
10675109 | Itkowitz et al. | Jun 2020 | B2 |
10881473 | Itkowitz et al. | Jan 2021 | B2 |
20020128552 | Nowlin | Sep 2002 | A1 |
20020133173 | Brock | Sep 2002 | A1 |
20030120283 | Stoianovici et al. | Jun 2003 | A1 |
20040111183 | Sutherland et al. | Jun 2004 | A1 |
20040254680 | Sunaoshi | Dec 2004 | A1 |
20050024331 | Berkley et al. | Feb 2005 | A1 |
20050166413 | Crampton et al. | Aug 2005 | A1 |
20050222554 | Wallace | Oct 2005 | A1 |
20050251110 | Nixon | Nov 2005 | A1 |
20060030840 | Nowlin | Feb 2006 | A1 |
20060178556 | Hasser et al. | Aug 2006 | A1 |
20060241414 | Nowlin | Oct 2006 | A1 |
20070013336 | Nowlin | Jan 2007 | A1 |
20070055291 | Birkmeyer et al. | Mar 2007 | A1 |
20070080658 | Farritor et al. | Apr 2007 | A1 |
20070197896 | Moll et al. | Aug 2007 | A1 |
20070299427 | Yeung | Dec 2007 | A1 |
20080009697 | Haider et al. | Jan 2008 | A1 |
20080046122 | Manzo et al. | Feb 2008 | A1 |
20080111513 | Farritor et al. | May 2008 | A1 |
20080114494 | Nixon | May 2008 | A1 |
20080132913 | Brock et al. | Jun 2008 | A1 |
20080154246 | Nowlin et al. | Jun 2008 | A1 |
20080235970 | Crampton | Oct 2008 | A1 |
20080319557 | Summers et al. | Dec 2008 | A1 |
20090000136 | Crampton | Jan 2009 | A1 |
20090088634 | Zhao | Apr 2009 | A1 |
20090088773 | Zhao et al. | Apr 2009 | A1 |
20090088774 | Swarup et al. | Apr 2009 | A1 |
20090088897 | Zhao et al. | Apr 2009 | A1 |
20090163929 | Yeung et al. | Jun 2009 | A1 |
20090192524 | Itkowitz | Jul 2009 | A1 |
20090259340 | Umemoto et al. | Oct 2009 | A1 |
20090326552 | Diolaiti | Dec 2009 | A1 |
20100225209 | Goldberg | Sep 2010 | A1 |
20100300230 | Helmer | Dec 2010 | A1 |
20100332031 | Itkowitz | Dec 2010 | A1 |
20110118748 | Itkowitz | May 2011 | A1 |
20110257653 | Hughes et al. | Oct 2011 | A1 |
20110275957 | Bhandari | Nov 2011 | A1 |
20110304819 | Juhasz et al. | Dec 2011 | A1 |
20110306986 | Lee et al. | Dec 2011 | A1 |
20110319714 | Roelle et al. | Dec 2011 | A1 |
20110319910 | Roelle et al. | Dec 2011 | A1 |
20130010081 | Tenney et al. | Jan 2013 | A1 |
20130238127 | Ohta et al. | Sep 2013 | A1 |
20140163664 | Goldsmith | Jun 2014 | A1 |
20160210882 | Gulasy et al. | Jul 2016 | A1 |
20160256223 | Haimerl et al. | Sep 2016 | A1 |
20170095301 | Brisson | Apr 2017 | A1 |
20210077210 | Itkowitz et al. | Mar 2021 | A1 |
Number | Date | Country |
---|---|---|
1533745 | Oct 2004 | CN |
WO-2007120353 | Oct 2007 | WO |
WO-2008133956 | Nov 2008 | WO |
WO-2009023801 | Feb 2009 | WO |
Entry |
---|
Cunningham, Steve, “3D Viewing and Rotation Using Orthonormal Bases,” Graphic Gems, Andrew S. Glassner, ed., 1990, pp. 516-521, Academic Press, Inc., Boston, MA, USA. |
Extended European Search Report for Application No. 16160639.7, dated Aug. 12, 2016, 8 pages. |
Extended European Search Report for Application No. 18203548.5 dated May 27, 2019, 12 pages. |
Hekstra, Gerben J. and Ed F.A. Deprettere, “Fast Rotations: Low-cost Arithmetic Methods for Orthonormal Rotation,” Proceedings of the 13th IEEE Symposium on Computer Arithmetic, Jul. 1997, pp. 116-125, IEEE. |
PCT/US10/38256 International Search Report and Written Opinion of the International Searching Authority, dated Aug. 20, 2010, 15 pages. |
Stroz, Kazimierz, “Derivation of the Rotation Matrix in General Rectilinear Systems by Means of Vector and Matrix Formalism,” Journal of Applied Crystallography, Dec. 1996, pp. 736-737, vol. 29, Part 6, International Union of Crystallography, GB. |
Vertut, J, and Coiffet, P., “Robot Technology: Teleoperation and Robotics Evolution and Development,” English translation, Prentice-Hall, Inc., Inglewood Cliffs, NJ, USA 1986, vol. 3A, 332 pages. |
Number | Date | Country | |
---|---|---|---|
20230329817 A1 | Oct 2023 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 17107862 | Nov 2020 | US |
Child | 18309110 | US | |
Parent | 16862412 | Apr 2020 | US |
Child | 17107862 | US | |
Parent | 16353932 | Mar 2019 | US |
Child | 16862412 | US | |
Parent | 15706883 | Sep 2017 | US |
Child | 16353932 | US | |
Parent | 15399600 | Jan 2017 | US |
Child | 15706883 | US | |
Parent | 14996073 | Jan 2016 | US |
Child | 15399600 | US | |
Parent | 14534526 | Nov 2014 | US |
Child | 14996073 | US | |
Parent | 13839438 | Mar 2013 | US |
Child | 14534526 | US | |
Parent | 12495213 | Jun 2009 | US |
Child | 13839438 | US |