Matrix assisted laser desorption/ionization time-of-fight mass (MALDI-TOF) spectrometry is an established technique for analyzing a variety of nonvolatile molecules including proteins, peptides, oligonucleotides, lipids, glycans, and other molecules of biological importance. While this technology has been applied to many applications, widespread acceptance has been limited by many factors including cost and complexity of the instruments, relatively poor reliability, and insufficient performance in terms of speed, sensitivity, resolution, and mass accuracy.
In the art, different types of TOF analyzers are required depending on the properties of the molecules to be analyzed. For example, a simple linear analyzer is preferred for analyzing high mass ions such as intact proteins, oligonucleotides, and large glycans, while a reflecting analyzer is required to achieve sufficient resolving power and mass accuracy for analyzing peptides and small molecules. Determination of molecular structure by MS-MS techniques requires yet another analyzer. In some commercial instruments all of these types of analyzers are combined in a single instrument. This has the benefit of reducing the cost somewhat relative to three separate instruments, but the downside is a substantial increase in complexity, reduction in reliability, and compromises are required that make the performance of all of the analyzers less than optimal.
Many areas of science require accurate determination of the molecular masses and relative intensities of a variety of molecules in complex mixtures and while many types of mass spectrometers are known in the art, each has well-known advantages and disadvantages for particular types of measurements. Time-of-flight (TOF) with reflecting analyzers provides excellent resolving power, mass accuracy, and sensitivity at lower masses (up to 5-10 kda), but performance is poor at higher masses primarily because of substantial fragmentation of ions in flight. At higher masses, simple linear TOF analyzers provide satisfactory sensitivity, but resolving power is limited. An important advantage of TOF MS is that essentially all of the ions produced are detected, unlike scanning MS instruments.
Many applications in the study of metabolics and proteomics require measurements on peptides and small molecules with high resolving power and mass accuracy. These are often present in complex mixtures and sensitivity over a relatively broad mass range, speed of analysis, reliability, and ease of use are also very important. An objective of the invention is a mass spectrometer providing optimum performance for these and similar applications that is reliable, easy to use, and relatively inexpensive.
The mass spectrometer or analyzer of the present invention comprises a MALDI sample plate and pulsed ion source located in an evacuated source ion housing; an analyzer vacuum housing isolated from the evacuated source ion housing by a gate valve containing an aperture and maintained at ground potential; a vacuum generator that maintains high vacuum in the analyzer; a pulsed laser beam that enters the evacuated source housing through the aperture in the gate valve when the valve is open and strikes the surface of a sample plate within the source producing ions that enter the analyzer through the aperture; an ion detector in close proximity to the gate valve and the ion beam; a field-free drift space at ground potential; an ion mirror at the opposite end of the drift space from the gate valve and ion detector; and high voltage supplies for supplying electrical potential to the ion mirror.
In one embodiment, the analyzer further comprises an ion lens in close proximity to the ion source and aligned with the ion beam passing through the aperture in the gate valve.
In one embodiment, the analyzer further comprises ion deflectors (deflection electrodes) in close proximity to the ion lens for deflecting the ions to reach the detector. At least one of the deflection electrodes is energized by a time dependent voltage that causes ions in one or more selected mass ranges to be deflected away from the detector.
In one embodiment the transverse distance from the laser beam to the center line of the detector is not more than 25 mm.
In one embodiment the length of the field-free region, the lengths of each of the stages of the mirror, and the voltages applied to the mirror are chosen to provide both first and second order velocity focusing from the source focus to the detector.
The present invention provides a method for optimizing the ion source operating conditions to give the optimum resolving power possible for a given set of initial conditions, ion energy, and overall size of the analyzer.
A high voltage pulse generator supplies a voltage pulse to the MALDI sample plate at a predetermined time after the laser pulse. In one embodiment a digital delay generator determines the delay with an uncertainty and jitter of less than 1 nanaosecond between the laser pulse and the voltage pulse. The time between the voltage pulse and the time that ions are detected at the detector is recorded by the digitizer to produce a time-of-flight spectrum that may be interpreted as a mass spectrum by techniques well known in the art.
An object of the present invention is to provide the optimum practical performance within limitations imposed by the length of the analyzer, the accelerating voltage, and the initial conditions including the width of the initial velocity distribution of the ions produced by MALDI and the uncertainty in initial position due, for example, due to the size of the matrix crystals.
In TOF mass spectrometry the performance can generally be improved by increasing the length of the analyzer and, for higher masses, by increasing the accelerating voltage, but these tend to increase the cost and reduce the reliability. The initial conditions are determined by the ionization process and are independent of the TOF analyzer design. In one embodiment of the present invention the accelerating voltage is 10 kilovolts, and the effective length of the analyzer is 3200 mm.
In one embodiment deflection electrodes are provided in a field-free region adjacent to the extraction electrode and energized to deflect ions in either of two orthogonal directions. At least one of the deflection electrodes may be energized by a time dependent voltage that causes ions in one or more selected mass ranges to be deflected away from the detector.
In one embodiment, the present invention provides a reflecting time-of-flight mass spectrometer comprising an ion source vacuum housing configured to receive a MALDI sample plate; a pulsed ion source located within the ion source housing; an analyzer vacuum housing; a gate valve located between and operably connecting said ion source vacuum housing and said analyzer vacuum housing and maintained at or near ground potential; a field-free drift space at ground potential located within said analyzer vacuum housing; an ion mirror located at the end of the field-free space in said analyzer vacuum housing opposite said gate valve; and an ion detector located in the field-free space within the analyzer vacuum housing in close proximity to the gate valve and having an input surface to receive ions reflected by the ion mirror. The reflecting time-of-flight mass spectrometer of the present invention may further comprise a pulsed laser beam directed to strike the MALDI sample plate and produce a pulse of ions; a high voltage pulse generator operably connected to the pulsed ion source; and a time delay generator providing a predetermined time delay between the laser pulse and the high voltage pulse.
In one embodiment, the reflecting time-of-flight mass spectrometer of the present invention has a predetermined time delay comprising an uncertainty which is not more than 1 nanosecond.
In one embodiment, the reflecting time-of-flight mass spectrometer of the invention is configured to contain one or more ion optical elements for spatially focusing an ion beam. The optical elements consist of at least an extraction electrode at ground potential in close proximity to the MALDI sample plate and a first ion lens located between the pulsed ion source and the gate valve.
According to the present invention, the ion lenses may comprise either an einzel lens or a cathode lens.
In one embodiment the high voltage pulse is between 10-30 kilovolts relative to ground potential. In one embodiment, the high voltage pulse is at about 10 kilovolts relative to ground potential. As used herein, “about” includes values which are within 10% of the value stated. For example, “about 10 kilovolts” includes voltages of between 9 and 11 kilovolts.
In one embodiment the distance between the MALDI sample plate and a grounded extraction electrode is between 1-5 mm. In one embodiment, the distance between the MALDI sample plate and a grounded extraction electrode is between 1-3 mm. In one embodiment, the distance between the MALDI sample plate and a grounded extraction electrode is less than 3 mm.
In a preferred embodiment the high voltage pulse is 10 kilovolts relative to ground potential and the distance between the MALDI sample plate and a grounded extraction electrode is less than 3 mm.
In certain embodiments, the ion mirrors are two-stage ion mirrors. The two-stage ion mirrors of the present invention may comprise two substantially uniform fields, wherein the field boundaries are defined by grids that are substantially parallel or may comprise two substantially uniform fields, wherein the field boundaries are defined by substantially parallel conducting diaphragms having small apertures aligned with an incident and reflected ion beam.
In one embodiment, the electrical field strength in the first stage of the two-stage ion mirror adjacent to the field-free region is substantially greater than the electrical field strength in the second stage of the two-stage ion mirror.
In one embodiment, the electrical field strength in the first stage of the two-stage ion mirror adjacent to the field-free region is between two and four times greater than the electrical field strength in the second stage of the two-stage ion mirror.
According to the present invention, the reflecting time-of-flight mass spectrometer may further comprise one or more pairs of deflection electrodes located in the field-free region at ground with any pair energized to deflect ions in either of two orthogonal directions. In one embodiment, at least one of the deflection electrodes of any pair of deflection electrodes is energized by a time-dependent voltage resulting in the deflection of ions in one or more selected mass ranges.
In one embodiment, the transverse distance from the pulsed laser beam to the center line of the ion detector is between 5 and 20 mm. In one embodiment the transverse distance from the pulsed laser beam to the center line of the ion detector is between 10 and 15 mm.
In a preferred embodiment, the transverse distance from the pulsed laser beam to the center line of the ion detector is not more than 25 mm.
In one embodiment, the input surface of the ion detector is perpendicular to the axis of the ion mirror with a maximum error of +/−0.05 degrees.
The present invention provides a method for designing a high-resolution MALDI-TOF mass spectrometer with predetermined limits on overall size and uncertainty in the time measurement comprising determining or estimating the uncertainties in the initial velocity and position of the ions produced in the ion source; calculating values for the critical distance parameters defining the analyzer geometry; calculating the optimum time lag between laser pulse and high-voltage extraction pulse as a function of focus mass; calculating the optimum accelerating voltages and mirror voltages as functions of focus mass and calculating the theoretical resolving power as a function of m/z.
The present invention provides a method for designing a high-resolution MALDI-TOF mass spectrometer to achieve a specified resolving power at a specified mass with specified values of the uncertainties in the initial velocity and position of ions produced in the ion source and the uncertainty in the time measurement comprising calculating the minimum overall length and values for the critical distance parameters defining the analyzer geometry; calculating the optimum accelerating voltages and mirror voltages; and calculating the optimum time lag between laser pulse and high-voltage extraction pulse.
In one embodiment, the pulsed laser beam operates at a frequency of 5 khz.
The foregoing and other objects, features and advantages of the invention will be apparent from the following more particular description of preferred embodiments of the invention, as illustrated in the accompanying drawings in which like reference characters refer to the same parts throughout the different views. The drawings are not necessarily to scale, emphasis instead being placed upon illustrating the principles of the invention.
A description of preferred embodiments of the invention follows. Referring now to
The ion beam 85 then passes through a field-free drift tube (or space) 80 and into an ion mirror 200 and are reflecting back through the field-free drift tube (or space) 80. The ion mirror comprises electrodes 214 and 224, each having an electrode feedthrough 212 and 222, respectively, through the analyzer vacuum housing. The mirror 200 also comprises a mirror grid 210 and a mirror electrode 220. The ions then pass through a grid 112 built into the detector and strike the input surface 92 of the detector 90 which is housed in housing 110. The detector comprises a dual channel plate electron multiplier. Each ion impinging on the input surface 92 produces a large number of electrons (ca. 1 million) in a narrow pulse at the output surface 94. The gain of the dual channel plate electron multiplier is determined by the bias voltage Vd applied across the dual channel plate. The electrons are accelerated by the electric field between the output surface 94 and the anode 100 at ground potential, and strike the anode producing an electrical pulse 102 that is coupled through an electrical feedthrough 104 in the wall of the analyzer vacuum housing 25 and connected to the input of a digitizer (not shown).
In one embodiment, the ion mirror dimensions and operating voltages are chosen so that the time required for ions to travel from a predetermined focal point 81 in the field-free drift tube (or space) 80, be reflected by the ion mirror, and reach the detector 90 is independent of the energy of the ions to both first and second order. First and second order focusing in a reflector requires satisfying the following equations:
4d3/Dm=1−3/w (1)
4d4/Dm=w−3/2+(4d3/Dm)/(w+w1/2) (2)
where Dm is the total length of the ion path from the focal point 81 to the mirror entrance plus the path from the mirror exit to the detector input surface 92, d3 is the length of a first region of the mirror, d4 is the distance that an ion with initial energy V penetrates into a second region of the mirror and w=V/(V−V1) is the ratio of the ion energy at the entrance to the mirror to that at the entrance to the second region with the intermediate electrode at potential V1.
Thus, first and second order focusing can be achieved for any value of w>3, and the corresponding distance ratios are uniquely determined by equations (1) and (2). For predetermined values of d3 and Dm, voltage V1 applied to mirror grid 210 is adjusted to satisfy equation (1) and voltage V2 applied to mirror electrode 220 is adjusted to satisfy equation (2), where
d4=d40(V−V1)/(V2−V1) (3)
In a preferred embodiment w=3.66, V1/V=0.7268, d3=d40=100 mm, V2/V=1.008, and d4=97.2 mm and the focal length Dm=2218.5 mm for first and second order focusing.
The effective length of the ion mirror is given by
Dem=4d4w1/2+4d3[w/(w−1)][1−w−1/2]=1020.6 mm (4)
The first order mass dependent focal length of a single-stage ion source is given by
D1=2d1+(2d1)2/vnΔt (5)
And the second order focal length is given by
D2=6d1 (6)
And these are satisfied simultaneously if
vnΔt=d1 (7)
If the energy is 10 kV and the focus mass is 3 kDa, this requires that Δt=118 nsec. The total effective flight distance is then
De=2218.5+1020.6+18+4=3263.1 (8)
and the effective length of the lens is included in Dm.
The effective length, De, of a time-of-flight analyzer may be defined as the length of a field-free region for which the flight time of an ion with kinetic energy corresponding to that in the field free drift tube (or space) 80 is equal to that of the same ion in the analyzer including accelerating and decelerating fields.
In one embodiment, the effective length, De, is approximately 3263.1 mm and ion energy is 10 kV, corresponding to a high-voltage pulse 12 of 10 kV in amplitude applied to MALDI sample plate 10. In this embodiment the flight time is approximately
t=(3263.1/0.0139)(m/10)1/2=74,240 m1/2 (9)
where t is in nsec and m in kDa. For a repetition rate of 5 khz the maximum flight time is 200,000 nsec thus the maximum mass is 7.2 kDa starting from mass zero.
The low mass region is dominated by ions from the MALDI matrix that are generally not useful for the analysis of samples. Also, if ions of masses higher than 7.2 kDa are produced, these will arrive following the next laser pulse and will be recorded at an incorrect mass. Therefore, in one embodiment an ion gate is provided that limits the mass range of ions exiting the ion source following each laser pulse so that only ions within a predetermined mass range are transmitted and detected.
In one embodiment, the length D of the field free drift tube (or space) 80 is large compared to the sum of the other distances, and d1 to is small as practical without initiating electrical discharge within the vacuum system.
Voltage may be applied to one or more of the electrodes, 27A, 27B, 28A, and 28B to deflect ions in the ion beam 85 produced by the pulsed laser beam 60 striking sample 29 deposited on the surface of the MALDI plate 10. A voltage difference between 27A and 27B deflects the ions in a direction perpendicular to the plane of the drawing, and a voltage difference between 28A and 28B deflects ions in the plane of the drawing.
Voltages can be applied as necessary to correct for misalignments in the ion optics and to direct ions along a preferred path to the detector. Also, a time dependent voltage can be applied to one or more of the deflection electrodes to deflect ions within predetermined mass ranges so that they cannot reach the detector and to allow ions in other predetermined mass ranges to pass through undeflected.
Electrodes 50 and 51 together with the extraction electrode 20 comprise an einzel lens that may be energized by applying voltage VL to electrode 50.
The effective length of the lens is given by
DeL=2d2Z[1−(1−Z−1)1/2] where Z=V/VL (10)
In one embodiment Z=2, and the DeL is approximately equal to 1.17 d2. The effective length of the lens is included in the field-free space between the exit from the source and the ion mirror.
In one embodiment, ion mirror 200 comprises a two-stage gridless mirror shown schematically in
In operation an ion beam enters the reflector through aperture 203 in first mirror plate 202 at a small angle θ 250 relative to a perpendicular 260 to plate 202. Potentials are applied to plates 204 and 206 causing the ions to pass through aperture 205 in plate 204 and be reflected back through aperture 207 in plate 204 and 209 in plate 202 and exits reflector 200 along a trajectory at an angle 251 relative to perpendicular 260 that is equal in degree but opposite in direction to angle 250. A set of substantially identical electrodes 230 and insulators 240 are stacked as illustrated in
Aperture diameters are chosen sufficiently large to allow a substantial fraction of the unfragmented ions to pass through the mirror. It is within the skill in the art to select an appropriate aperture size for the application. Ions that have lost significant energy due to fragmentation in flight follow a different path and are prevented from passing through the exit aperture 209.
The arrangement employed to insure that the fields are substantially uniform in the region that the ion beam passes through is illustrated in
The time required for an ion to travel from the ion source to a deflection electrode (i.e., deflector) following application of the high-voltage accelerating pulse to the MALDI plate 10 is essentially proportional to the square root of the mass-to-charge ratio, and this time can be calculated with sufficient accuracy from a knowledge of the applied voltage V and the distances involved. To transmit ions within a specified mass range, for example from m1 to m2, voltage is applied to the deflector at or before the laser pulse occurs and continues until the time that m1 arrives at the entrance to the deflector, and is turned off until the time that m2 exits the deflector. After m2 exits the deflector, the voltage is turned back on. For example, mass ranges such as 0.5-11.5 kDa or 0.1-9 kDa can be acquired at 5 khz by using the mass gate to select a portion of the spectrum corresponding to arrival times at the detector within a 200 microsecond window corresponding to the time between laser pulses. Any ions outside the selected range are removed by the mass gate and the possibility of high masses overlapping into the spectrum produced by the next laser pulse is removed. The mass gate can also be employed to limit the mass range to a narrower window when required by the application.
The limit on resolving power set by time resolution is given by
Rt−1=t/2δt (11)
Where δt is the uncertainty of the time measurement.
Design of TOF Analyzers
The principal measures of performance are sensitivity, mass accuracy, and resolving power. Sensitivity is the most difficult of these since it generally depends on a number of factors some of which are independent of the attributes of the analyzer. These include chemical noise associated with the matrix or impurities in the sample, and details of the sample preparation. For the purpose of assessing the performance of the analyzer independent of these extraneous (although often dominant) factors the major components of sensitivity are the efficiency with which sample molecules are converted to ions providing measurable peaks in the mass spectrum, and the ion noise associated with ions detected that provide no useful information. The efficiency may be further divided into ionization efficiency (ions produced/molecule desorbed), transmission efficiency, and detection efficiency. A very important term that is often ignored is the sampling efficiency (sample molecules desorbed/molecule loaded).
The major sources of ion loss and ion noise are fragmentation and scattering. Fragmentation can occur spontaneously at any point along the ion path as a result of excitation received in the ionization process. Fragmentation and scattering can also occur as the result of collisions of the ions with neutral molecules in the flight path or with electrodes and grids. A vacuum in the low 10−7 torr range is sufficient to effectively limit collisions with neutral molecules, but grids and defining apertures required to achieve resolving power in some cases may reduce sensitivity both due to ion loss and production of ion noise.
In a linear TOF, fragmentation in the field-free region may produce some tails on the peaks, but generally has at most a small effect on sensitivity. The major loss and source of ion noise is fragmentation in the ion accelerator. If acceleration occurs between the end of the drift space and the detector, ghost peaks may occur as the result of low mass charged fragments arriving early and neutral fragments arriving late. No defining apertures or grids are required in the linear analyzer.
In reflecting analyzers ions that fragment between the source and mirror will appear as broad peaks at an apparent mass below the peak for the precursor mass, since the fragments spend less time in the ion mirror. Ions fragmenting in the mirror are randomly distributed in the space between the parent ion and the fragment. Grids are often used in the mirror to improve resolving power; these may cause a significant loss in ion transmission and a source of ion noise.
In MALDI-TOF the most obvious limitation on resolving power and mass accuracy is set by the initial velocity distribution that is at least approximately independent of the mass and charge of the ions. Time lag focusing can be employed to reduce the effect of initial velocity, and the distribution in initial position of the ions may become the limiting factor. Other limits are imposed by trajectory errors and the uncertainty in the measurement of ion flight times.
Reflecting Analyzer
Referring again to
=97.2 mm, Dm=2218.5, Dem=1020.6. In this case the total effective length is
De=Dm+Dem+6d1+2d1=2218.5+1020.6+18+4=3263.1
and the effective length of the lens is included in Dm.
The various contributions to peak width in TOF MS can be summarized as follows: (expressed as Δm/m)
First order dependence on initial position
Rs1=[(Dv−Ds)/De](δx/d1) (12)
Where De is the effective length of the analyzer, δx is the uncertainty in the initial position, d1 is the length of the first region of the ion accelerator, and Dv and Ds are the focal lengths for velocity and space focusing, respectively, and are given by
Ds=2d1 (13)
Dv=Ds+(2d1)2/(vn*Δt)=6d1 (14)
Where Δt is the time lag between ion production and application of the accelerating field, and vn* is the nominal final velocity of the ion of mass m* focused at Dv. vn* is given by
vn*=C1(V/m*)1/2 (15)
The numerical constant C1 is given by
C1=(2z0/m0)1/2=2×1.60219×10−19 coul/1.66056×1027 kg=1.38914×104 (16)
For V in volts and m in Da (or m/z) the velocity of an ion is given by
v=C1(V/m)1/2 m/sec (17)
and all lengths are expressed in meters and times in seconds. It is numerically more convenient in many cases to express distances in mm and times in nanoseconds. In these cases C1=1.38914×10−2.
The time of flight is measured relative to the time that the extraction pulse is applied to the source electrode. The extraction delay Δt is the time between application of the laser pulse to the source and the extraction pulse. The measured flight time is relatively insensitive to the magnitude of the extraction delay, but jitter between the laser pulse and the extraction pulse causes a corresponding error in the velocity focus. In cases where Δt is small, this can be a significant contribution to the peak width. This contribution due to jitter δj is given by
RΔ=2(δj/Δt)(δv0/vn*)(Dv−Ds)/De=2(δjδv0/De)[(Dv−Ds)/2d0y]2 (18)
and is independent of mass.
With time lag focusing the first order dependence on initial velocity is given by
Rv1=[(4d1y)/De](δv0/vn)[1−(m/m*)1/2]=Rv1(0)[1−(m/m*)1/2] (19)
Where δv0 is the width of the velocity distribution. At the focus mass, m=m*, the first order term vanishes.
With first order focusing the velocity dependence becomes
Rv2=2[(2d1y)/(Dv−Ds)]2(δv0/vn)2 (20)
And with first and second order velocity focusing the velocity dependence becomes
Rv3=4[(2d1y)/(Dv−Ds)]3(δv0/vn)3 (21)
The dependence on the uncertainty in the time measurement δt is given by
Rt=2δt/t=(2δtC1/De)(V/m)1/2 (22)
The dependence on trajectory error δL is given by
RL=2δL/De (23)
A major contribution to δL is often the entrance into the channel plates of the detector. If the channels have diameter d and angle a relative to the beam, the mean value of δL is d/2 sin α. Thus this contribution is
RL=d/(De sin α) (24)
Noise and ripple on the high voltage supplies can also contribute to peak width. This term is given by
RV=ΔV/V (25)
where ΔV is the variation in V in the frequency range that effects the ion flight time.
It is obvious from these equations that increasing the effective length of the analyzer increases the resolving power, but some of the other effects are less obvious. The total contribution to peak width due to velocity spread is given by
Rv=Rm+(ΔD12/De)Rv2+[(De−ΔD12)/De]Rv3 (26)
where ΔD12 is the absolute value of the difference between Dv1 and Dv2. Assuming that each of the other contributions to peak width is independent, the overall resolving power is given by
R−1=[RΔ2+Rs12+Rv2+Rt2+RL2+RV2]−1/2 (27)
Optimization of the Reflecting Analyzer
For a reflecting analyzer with first and second order focusing the terms limiting the maximum resolving power are Rs1, Rv3, and Rt. The variation of resolving power with mass is determined primarily by Rv1 and may also be affected by Rt. In terms of the dimensionless parameter K=2d1/(Dv−Ds) the major contributions can be expressed as
Rs1=2K−1[δx/De] (28)
Rv3=4K3(δv0/vn)3 (29)
And R2=4K−2[δx/De]2+16K6(δv0/vn)6 (30)
The minimum value of R2 corresponds to d(R2)dK=0
−8K−3[δx/De]2+96K5(δv0/vn)6=0
K8=(1/12)[δx/De]2(δv0/vn)−6
K=0.733{[δx/De]/(δv0/vn)3}1/4 (31)
For one embodiment [δx/De]=0.01/3263.1=3×10−6, (δv0/vn)3=(0.0004/0.0254)3 =3.9×10−6
K=0.69. For the embodiment described above K=0.5; very close to the optimum. In the more general case
K=12−1/8(De)−1/4{[δxC13(δv0)−3}1/4(V/m*)3/8 (32)
The contributions to peak width due to Rs1 and Rv3 are shown as a function of K in
Rt=Δm/m=2(δt)C1V1/2/(Dem1/2)=2(0.75)(0.0139)(101/2)/(De[3]1/2)=3.81×10−2/De (33)
Using the optimum value of K, and inserting Rt, Rs1, and Rv3 for each De into equation (25) the maximum resolving power for 3 kDa at V=10 KV can be calculated as a function of the effective length De of the analyzer.
Results over a broad range are illustrated in
Calculated resolving power as a function of m/z is summarized in
Dv1=2d1+2.89d1=4.89d1 and Dv2=6d1 (34)
It is important to adjust the ion mirror potentials to achieve overall first order focusing, and the mirror can be adjusted to independently correct the second order focus. However, a small discrepancy in the second order focus is negligible so long as the error δD/De is small compared to δv0/vn*. The first and second order focal lengths of the two-stage mirror are given by
Dm1=4d4w3/2+4d3[w/(w−1)][1−w1/2] (35)
3Dm2=4d4w5/2+4d3[w/(w−1)][1−w3/2] (36)
Equations (1) and (2) are derived by setting these focal distances equal, but these can be varied independently, for example by adjusting d4 by changing V2 according to equation (3).
Some applications of MALDI-TOF require a small analyzer, for example, for a field portable instrument. The methods of this invention can also be applied to the optimum design of smaller analyzers. For example one may choose De=400 mm, and V=4 kV. For focus at m*=3 kDa the optimum value of K in this case is 0.826 according to equation (32). The calculated resolving power as a function of m/z is illustrated in
An optimized reflecting analyzer comprises a single-stage source with the accelerating distance as short as practical without causing electrical discharges. In one embodiment, the accelerating distance is 3 mm and the accelerating voltage is 10 kV. The analyzer further comprises a two-stage ion mirror with the source and mirror adjusted to provide simultaneous first and second order focusing with the source focus at the optimum value. The optimum value of the source focus is determined as a function of focus mass, accelerating voltage, effective length, and initial velocity and spatial distributions using methods described herein. The ultimate resolving power is limited only by the overall length of the analyzer as restricted by the dimensions of the laboratory, but is otherwise unrestricted.
The other major contributor to peak width is due to uncertainty in the time measurement due to the finite width of single ion pulses and the width of the bins in the digitizer. With standard 5 um dual channel plate detectors and digitizers with 0.5 nsec bins the uncertainty δt is about 1.5 nsec. Commercial detectors are now available that provide single ion peak widths less than 0.5 nsec and digitizers with 0.25 nsec bins are available. These may allow the uncertainty, δt, in the time measurement to be reduced to about 0.75 nsec. The limit on peak width is
Rt=Δm/m=2δtvn/De=2(δt/De)C1(V/m)1/2 (37)
The optimum value of V/m for given initial conditions and geometry can be determined by finding the minimum for R2 due to contributions from Rt and Rv3. Thus
R2=[2δtvn/De]2+16K6(δv0/vn)6 (38)
The minimum value of R2 corresponds to d(R2)/dvn=0
vn=121/8(Kδv0)3/4(De/δt)1/4=C1(V/m)1/2 (39)
The optimum value of K determined by optimizing between Rs1 and Rv3 is given by equation (32).
The overall optimum can be found by simultaneously satisfying both (32) and (37). This is satisfied for
vn=δx/Kδt (40)
This condition corresponds to Rs1=Rt. The resulting peak width R is then given by
R=[2Rs12+Rv32]1/2 (41)
The global optimum conditions can be determined by substituting vn as determined from equation (40) and determining the optimum value of K from the resulting equation. Thus from equation (30)
R2=8K−2[δx/De]2+16K6(δv0/vn)6=8K−2[δx/De]2+16K6(δv0)6(Kδt/δx)6 (42)
And the minimum value found for d(R2)/dK=0 is
K=(12)−1/14(δx/δtδv0)3/7(δx/De)1/7 (43)
The optimum value of V/m is given by
V/m=(δx/KC1δt)2 (44)
Equations (43) and (44) give the focusing parameter K and voltage V corresponding to maximum resolving power for a given mass m with an analyzer of effective length De, for time measurement uncertainty δt, initial velocity spread δv0, and initial position uncertainty δx.
The equations presented here provide the theoretical background for methods to design and optimize reflecting analyzers for generating spectra with high resolution and mass accuracy. The emphasis is on application to MALDI, but the techniques described can be applied to any TOF mass spectrometer. If the initial conditions including the initial velocity spread δv0, and initial position uncertainty δx are known or can be accurately estimated, and if the measurement uncertainty δt and the jitter in the delay δj are known, then for any size analyzer the optimum time lag Δt, the optimum mirror voltages, and optimum acceleration voltage can be determined accurately for any specified focus mass. Furthermore, the maximum resolving power possible can be accurately determined. Alternatively for any specified resolving power required the minimum analyzer size and optimum acceleration voltage can be determined.
Calibration for Accurate Mass Determination
With first and second order focusing the flight time is proportional to the square root of the mass except for the time spent in the ion source that depends on the initial velocity. Thus the total flight time for one embodiment is given by
t−t0=(De/vn)[1−2d1yv0/(Devn)]=Am1/2[1−Bm1/2]=X (45)
where t0 is the offset between the extraction pulse and the start time of the digitizer, and the default values of the constants are
A=De/CV1/2B=(2d1y/De)(v0/CV1/2) (46)
This equation can be inverted using the quadratic formula to give an explicit expression for mass as a function of flight time.
m1/2=(2B)−1[1−(1−4BX/A)1/2] (47)
Higher order terms may become important if a very wide mass range is employed. A higher order correction can be determined by the following procedure.
Z(m)=[(t−t0)/{Am1/2(1−Bm1/2)}]=1−C(m−m0) (48)
If a significant systematic variation of Z with m is observed, then the results are fitted to an explicit function, such as given in equation (48). This factor Z(m) is then applied to the value of m1/2 from equation (47) to determine the accurate mass. The value determined from equation (47) is divided by Z(m).
The values of t0, A, and B are determined by least squares fit from three or more peaks to equation (1). If a systematic variation of Z is observed, then the higher order term may be important, and the offset m0 may be necessary to compensate for the systematic error in the calibration.
While this invention has been particularly shown and described with references to preferred embodiments thereof, it will be understood by those skilled in the art that various changes in form and details may be made therein without departing from the scope of the invention encompassed by the appended claims.
Number | Name | Date | Kind |
---|---|---|---|
3727047 | Janes | Apr 1973 | A |
4730111 | Vestal et al. | Mar 1988 | A |
4731533 | Vestal | Mar 1988 | A |
4766312 | Fergusson et al. | Aug 1988 | A |
4814612 | Vestal et al. | Mar 1989 | A |
4861989 | Vestal et al. | Aug 1989 | A |
4883958 | Vestal | Nov 1989 | A |
4902891 | Vestal | Feb 1990 | A |
4958529 | Vestal | Sep 1990 | A |
4960992 | Vestal et al. | Oct 1990 | A |
5015845 | Allen et al. | May 1991 | A |
5160840 | Vestal | Nov 1992 | A |
5498545 | Vestal | Mar 1996 | A |
5625184 | Vestal et al. | Apr 1997 | A |
5627369 | Vestal et al. | May 1997 | A |
5654545 | Holle et al. | Aug 1997 | A |
5760393 | Vestal et al. | Jun 1998 | A |
6002127 | Vestal et al. | Dec 1999 | A |
6057543 | Vestal et al. | May 2000 | A |
6175112 | Karger et al. | Jan 2001 | B1 |
6274866 | Li | Aug 2001 | B1 |
6281493 | Vestal et al. | Aug 2001 | B1 |
RE37485 | Vestal | Dec 2001 | E |
6326615 | Syage et al. | Dec 2001 | B1 |
6348688 | Vestal | Feb 2002 | B1 |
6414306 | Mayer-Posner et al. | Jul 2002 | B1 |
6441369 | Vestal et al. | Aug 2002 | B1 |
6504150 | Verentchikov et al. | Jan 2003 | B1 |
6512225 | Vestal et al. | Jan 2003 | B2 |
6534764 | Verentchikov et al. | Mar 2003 | B1 |
6541765 | Vestal | Apr 2003 | B1 |
6570152 | Hoyes | May 2003 | B1 |
6621074 | Vestal | Sep 2003 | B1 |
6670609 | Franzen et al. | Dec 2003 | B2 |
6674070 | Karger et al. | Jan 2004 | B2 |
6770870 | Vestal | Aug 2004 | B2 |
6825463 | Karger et al. | Nov 2004 | B2 |
6831270 | Furuta et al. | Dec 2004 | B2 |
6844545 | Hutchins et al. | Jan 2005 | B1 |
6900061 | Smirnov et al. | May 2005 | B2 |
6918309 | Brock et al. | Jul 2005 | B2 |
6933497 | Vestal | Aug 2005 | B2 |
6952011 | Brown et al. | Oct 2005 | B2 |
6953928 | Vestal et al. | Oct 2005 | B2 |
6995363 | Donegan et al. | Feb 2006 | B2 |
7030373 | Vestal et al. | Apr 2006 | B2 |
7064319 | Hashimoto et al. | Jun 2006 | B2 |
7109480 | Vestal et al. | Sep 2006 | B2 |
RE39353 | Vestal | Oct 2006 | E |
7176454 | Hayden et al. | Feb 2007 | B2 |
20030057368 | Franzen et al. | Mar 2003 | A1 |
20030116707 | Brown et al. | Jun 2003 | A1 |
20050031496 | Laurell et al. | Feb 2005 | A1 |
20050087685 | Bouvier et al. | Apr 2005 | A1 |
20050116162 | Vestal | Jun 2005 | A1 |
20050130222 | Lee et al. | Jun 2005 | A1 |
20050178959 | Lopez-Avila et al. | Aug 2005 | A1 |
20050269505 | Ermer | Dec 2005 | A1 |
20060266941 | Vestal | Nov 2006 | A1 |
20060273252 | Hayden et al. | Dec 2006 | A1 |
20070038387 | Chen et al. | Feb 2007 | A1 |
20070054416 | Regnier et al. | Mar 2007 | A1 |
20070187585 | Verentchikov | Aug 2007 | A1 |
20080272287 | Vestal | Nov 2008 | A1 |
20080272289 | Vestal | Nov 2008 | A1 |
20080272293 | Vestal | Nov 2008 | A1 |
Number | Date | Country |
---|---|---|
2 370 114 | Jun 2002 | GB |
WO 2004018102 | Mar 2004 | WO |
WO 2005061111 | Jul 2005 | WO |
Number | Date | Country | |
---|---|---|---|
20080272290 A1 | Nov 2008 | US |