The section headings used herein are for organizational purposes only and should not be construed as limiting the subject matter described in the present application in any way.
Many mass spectrometer applications require an accurate determination of the molecular masses and relative intensities of metabolites, peptides, and intact proteins in complex mixtures. Tandem mass spectrometry provides information on the structure and sequence of many biological polymers and allows unknown samples to be accurately identified. Tandem mass spectrometers employ a first mass analyzer to produce, separate and select a precursor ion, and a second mass analyzer to fragment the selected ions and record the fragment mass spectrum from the selected precursor. A wide variety of mass analyzers and combinations thereof for use in tandem mass spectrometry are known in the literature.
An important advantage of TOF Mass Spectrometry (MS) is that essentially all of the ions produced are detected, which is unlike scanning MS instruments. This advantage is lost in conventional MS-MS instruments where each precursor is selected sequentially and all non-selected ions are lost. This limitation can be overcome by selecting multiple precursors following each laser shot and recording fragment spectra from each can partially overcome this loss and dramatically improve speed and sample utilization without requiring the acquisition of raw spectra at a higher rate.
Several approaches to matrix assisted laser desorption/ionization (MALDI)-TOF MS-MS are described in the prior art. All of these approaches are based on the observation that at least a portion of the ions produced in the MALDI ion source may fragment as they travel through a field-free region. Ions may be energized and fragment as the result of excess energy acquired during the initial laser desorption process, or by energetic collisions with neutral molecules in the plume produced by the laser, or by collisions with neutral gas molecules in the field-free drift region. These fragment ions travel through the drift region with approximately the same velocity as the precursor, but their kinetic energy is reduced in proportion to the mass of the neutral fragment that is lost. A timed ion selector may be placed in the drift space to transmit a small range of selected ions and to reject all others. In a TOF mass analyzer employing a reflector, the lower energy fragment ions penetrate less deeply into the reflector and arrive at the detector earlier in time than the corresponding precursors. Conventional reflectors focus ions in time over a relatively narrow range of kinetic energies. Thus, only a small mass range of fragments are focused for given potentials applied to the reflector.
In work by Spengler and Kaufmann, the limitation in mass range was overcome by taking a series of spectra at different mirror voltages and piecing them together to produce the complete fragment spectrum. An alternate approach is to use a “curved field reflector” that focuses the ions in time over a broader energy range. The TOF-TOF approach employs a pulsed accelerator to re-accelerate a selected range of precursor ions and their fragments so that the energy spread of the fragments is sufficiently small that the complete spectrum can be adequately focused using a single set of reflector potentials. All of these approaches have been used to successfully produce MS-MS spectra following MALDI ionization, but each suffers from serious limitations that have stalled widespread acceptance. For example, each approach involves relatively low-resolution selection of a single precursor, and generation of the MS-MS spectrum for that precursor, while ions generated from other precursors present in the sample are discarded. Furthermore, the sensitivity, speed, resolution, and mass accuracy for the first two techniques are inadequate for many applications.
The first practical time-of-flight (TOF) mass spectrometer was described by Wiley and McClaren more than 50 years ago. TOF mass spectrometers were generally considered to be only a tool for exotic studies of ion properties for many years. See, for example, “Time-of-Flight Mass Spectrometry: Instrumentation and Applications in Biological Research,” Cotter R J., American Chemical Society, Washington, D.C. 1997, for review of the history, development, and applications of TOF-MS in biological research.
Early TOF mass spectrometer systems included ion sources with electron ionization in the gas phase where a beam of electrons is directed into the ion source. The ions produced have a distribution of initial positions and velocities that is determined by the intersection of the electron beam with the neutral molecules present in the ion source. The initial position of the ions and their velocities are independent variables that affect the flight time of the ions in a TOF-MS. Wiley and McLaren developed and demonstrated methods for minimizing the contribution of each of these distributions. Techniques for minimizing the contribution of initial position are called “space focusing” techniques. Techniques for minimizing the contribution of initial velocity are called “time lag focusing” techniques. One important conclusion made by Wiley and McLaren is that it is impossible to simultaneously achieve both space focusing and velocity focusing. According to Wiley and McLaren, optimization of these TOF mass spectrometers requires finding the optimum compromise between the space focusing and velocity focusing distributions.
The advent of naturally pulsed ion sources such as CF plasma desorption ions source, static secondary ion mass spectrometry (SIMS), and matrix-assisted laser desorption/ionization (MALDI) ion sources has led to renewed interest in TOF mass spectrometers. Recent work in TOF mass spectrometry has focused on developing new and improved TOF instruments and software that take advantage of MALDI and electrospray (ESI) ionization sources that have removed the volatility barrier for mass spectrometry and that have facilitated applications of important biological applications.
The ion focusing techniques used with MALDI and electrospray (ESI) ion sources reflect the practical limits on the position and velocity distributions that can be achieved with these techniques. Achieving optimum performance with electrospray ionization and MALDI ionization methods requires finding the best compromise between space and velocity focusing. Electrospray ionization methods have been developed to improve space focusing. Electrospray ionization forms a beam of ions with a relatively broad distribution of initial positions and a very narrow distribution in velocity in the direction that ions are accelerated.
In contrast, MALDI ionization methods have been developed to improve velocity focusing. MALDI ionization methods use samples deposited in matrix crystals on a solid surface. The variation in the initial ion position is approximately equal to the size of the crystals, which is small. However, the velocity distribution is relatively broad because the ions are energetically ejected from the surface by the incident laser irradiation.
Known TOF mass spectrometers use delayed pulsed acceleration in the ion source to achieve first order velocity focusing for a single selected ion mass-to-charge ratio. Delayed pulsed acceleration was referred to as “time lag focusing” by Wiley and McLaren and more recently is referred to as “delayed extraction” or “delayed pulsed extraction.” Although time lag focusing provides first order velocity focusing for a selected mass, it is not suitable for focusing a broad range of masses as described above. Furthermore, time lag focusing does not correct for variations in the initial ion position.
The present teachings, in accordance with preferred and exemplary embodiments, together with further advantages thereof, is more particularly described in the following detailed description, taken in conjunction with the accompanying drawings. The skilled person in the art will understand that the drawings, described below, are for illustration purposes only. The drawings are not necessarily to scale, emphasis instead generally being placed upon illustrating principles of the teaching. The drawings are not intended to limit the scope of the Applicant's teachings in any way.
The following variables are used in the Description of Various Embodiments section:
Reference in the specification to “one embodiment” or “an embodiment” means that a particular feature, structure, or characteristic described in connection with the embodiment is included in at least one embodiment of the teaching. The appearances of the phrase “in one embodiment” in various places in the specification are not necessarily all referring to the same embodiment.
It should be understood that the individual steps of the methods of the present teachings may be performed in any order and/or simultaneously as long as the teaching remains operable. Furthermore, it should be understood that the apparatus and methods of the present teachings can include any number or all of the described embodiments as long as the teaching remains operable.
The present teachings will now be described in more detail with reference to exemplary embodiments thereof as shown in the accompanying drawings. While the present teachings are described in conjunction with various embodiments and examples, it is not intended that the present teachings be limited to such embodiments. On the contrary, the present teachings encompass various alternatives, modifications and equivalents, as will be appreciated by those of skill in the art. Those of ordinary skill in the art having access to the teachings herein will recognize additional implementations, modifications, and embodiments, as well as other fields of use, which are within the scope of the present disclosure as described herein.
The present teaching relates to tandem time-of-flight mass spectrometer apparatus and methods of operating tandem time-of-flight mass spectrometer apparatus that employ a first stage time-of-flight analyzer which provides simultaneous space and velocity focusing for an ion of predetermined mass-to-charge ratio. In addition, the present teaching relates to tandem time-of-flight mass spectrometer apparatus and methods of operating tandem time-of-flight mass spectrometer apparatus that provide high mass resolution performance for a broad range of ions.
One aspect of the present teaching is that it has been discovered that pulsed acceleration in the ion source is not required to achieve velocity focusing. Another aspect of the present teaching is that it has been discovered that pulsed acceleration can be used for initiating time-of-flight measurements when a continuous beam of ions is generated. Another aspect of the present teaching is that it has been discovered that higher mass resolution can be achieved by using pulsed acceleration for initiating TOF measurements. Yet another aspect of the present teaching is that it has been discovered that using a first stage time-of-flight mass analyzer with simultaneous space and velocity focusing allows high resolution precursor selection to be achieved and also reduces the velocity spread of selected ions, thereby allowing high resolution fragment spectra to be generated and recorded in a second stage time-of-flight mass analyzer. These and other aspects of the present teaching are described in more detail below.
The first time-of-flight mass analyzer 12 comprises an ion source 102 that generates a pulse of ions, a pulsed ion accelerator 108, a low resolution timed ion selector 110, a first field-free drift space 114, a high resolution timed ion selector 116, and a second field-free drift space 118. The ion source 102 generates a pulse of ions. The pulsed ion accelerator 108 accelerates the pulse of ions. The low resolution timed ion selector 110 transmits a range of masses accelerated in pulsed accelerator 108 and rejects all others. The high resolution timed ion selector 116 transmits a predetermined set of precursor ions accelerated by pulsed ion accelerator 108. Selected precursor ions and fragments thereof produced in either field-free drift space 114 or 118 are transmitted to the second stage time-of-flight analyzer 20 where fragment ions from each selected precursor are separated according to the mass-to-charge ratio of the fragment and detected and recorded to produce mass spectra of the fragment ions.
The first time-of-flight analyzer 12 provides simultaneous space and velocity focusing for an ion of predetermined mass-to-charge ratio at the timed ion selector 116. In addition, the first time-of-flight analyzer 12 minimizes the focusing error for ions within a predetermined mass range including the focused mass.
In some embodiments, field-free drift spaces 114 and 118 comprise fragmentation chambers wherein ions may fragment spontaneously as the result of internal excitation in the ion source or as the result of excitation by collisions with neutral molecules in field-free spaces 114 or 118. In some embodiments, the pressure in at least one of the field-free regions 114 or 118 is increased to enhance excitation by collisions with neutral molecules. In some embodiments, at least one of the field-free regions 114 or 118 may be enclosed and differential pumped employed to allow the pressure in these regions to be increased without increasing the pressure in other regions of the tandem mass spectrometer. In general, in various embodiments, the pressure in each of the regions of the first time-of-flight analyzer 12 can be optimized separately.
The ion source 202 generates a pulse of ions 206. In one embodiment the ion source 202 includes a sample plate 208 that positions a sample 210 for analysis. An energy source, such as a laser, is positioned to provide a beam of energy 212 to the sample 210 positioned on the sample plate 208 that ionizes sample material and generates a pulse of ions 206. The beam of energy 212 can be a pulsed beam of energy, such as a pulsed beam of light. In another embodiment, a continuous source of ions is transmitted to ion source 202 and an accelerating pulse is applied periodically to ion source 202 to produce a pulse of ions.
The pulse of ions 206 is accelerated by ion accelerator 204 that includes a first 214 and second electrode 216 positioned adjacent to the sample plate 208. A pulsed ion accelerator 220 is positioned adjacent to the second electrode 216. In some embodiments, a first field-free ion drift space 218 is positioned between the electrode 216 and the pulsed ion accelerator 220. The pulsed ion accelerator 220 includes an entrance plate 222. A timed ion selector 224 is positioned adjacent to the pulsed ion accelerator 220. A field-free ion drift space 232 is positioned adjacent to the timed ion selector 224. A high resolution timed ion selector 228 is positioned at the end of the field-free ion drift space 232.
In operation, a beam of energy 212, which can be a pulsed beam of energy or a continuous beam is generated and directed to sample 210. Sample 210 may be deposited on the surface of sample plate 208 or may be present in the gas phase adjacent to sample plate 208. The pulsed beam of energy 212 can be a pulsed laser beam that produces ions from samples present either on sample plate 208 or in the gas phase proximate to the sample plate 208. A pulse of ions can also be produced by either a pulsed or continuous beam of ions to produce ions from samples present either on sample plate 208 or in the gas phase proximate to the sample plate 208 by a method known as secondary ionization mass spectrometry (SIMS). In some methods of operation, the sample 210 includes a UV absorbing matrix and ions are produced by matrix assisted laser desorption ionization (MALDI). In another method of operation, a continuous source of ions is produced by electrospray ionization and transmitted to ion source 202 and an accelerating pulse is applied periodically to ion source 202 to produce a pulse of ions.
The ion accelerator 204 is biased with a voltage to accelerate the pulse of ions into the pulsed ion accelerator 220. The pulsed ion accelerator 220 accelerates the pulse of ions. The timed ion selector 224 transmits ions accelerated by the pulsed ion accelerator 220 into the field-free drift space 226 and rejects other ions by directing the ions along trajectory 230. The accelerated ions transmitted by the timed ion selector 224 are then transmitted to high resolution timed ion selector 228.
The entrance plate 222 of the pulsed ion accelerator 220 is positioned adjacent to the second electrode 216. In some embodiments, the entrance plate 222 of the pulsed ion accelerator 220 is at a distance dc from the second electrode 216, which is at grounded potential. When an ion of predetermined mass-to-charge ratio reaches a predetermined point 312 in the pulsed accelerator 220, a pulsed voltage Vp 314 is applied to the entrance plate 222 of the pulsed ion accelerator 220. The pulsed voltage Vp focuses the ions through the second field-free drift space 226 to the high resolution timed ion selector 228, thereby removing (to first order) the effect of both initial position and initial velocity of the ions on the flight time from the pulsed accelerator 220 to the high resolution timed ion selector 228. The low resolution timed ion selector 224 located adjacent to the exit 223 of the pulsed accelerator 220 is activated to transmit only ions accelerated by the pulsed accelerator 220 and to also prevent all other ions from reaching the high resolution selector 228.
To illustrate this aspect of the present teaching, an analysis of a two-field ion accelerator for a first time-of-flight mass spectrometer is presented to show that both spatial and velocity focusing can be achieved simultaneously. The space focusing distance for a two-field ion accelerator is given by
Ds=2day3/2[1−(db/da)/(y+y1/2)]
where da is the length of the first accelerating field, db is the length of the second accelerating field and y is the ratio of the total accelerating potential V to the accelerating potential in the first field V−Vg, and where Vg is the potential applied the electrode intermediate to the two fields. The total effective length of the source is given by
Des=2day1/2[1+(da/db)/(y1/2+1)].
Thus, the time for ions to travel to point Dv from the exit 223 of the pulsed accelerator 220 is independent of the perturbation in velocity if
Dv=2d1(Va+V)/Vp
where Vp is the amplitude of the pulsed voltage, Va is the acceleration given to a predetermined precursor mass, and d1 is the length of the pulsed accelerating field. If the predetermined mass is at the center of the pulsed accelerating field, then it follows that
(Va/V)=q0=Vp/2V and
Dv=2d1(1+q0)/2q0.
The spatial focusing error also contributes to an increase in the mass-to-charge ratio peak width. The kinetic energy of ions with the spatial focusing error is given by zV(1−p2), where the perturbation in spatial focusing is given by
p2=(δx/2day).
At the space focus point, the ions with higher energy overtake the ions with lower energy. If the space focus is located at a greater distance than the pulsed accelerator, for example, in the vicinity of the detector, then the lower energy ions arrive at the pulsed accelerator before those with higher energy. The later arriving ions with relatively high energy are accelerated by the pulsed ion accelerator more than the ions with relatively low energy, which effectively increases their space focal distance. Thus, the change in spatial focal point due to the pulsed accelerator to first order is approximately
ΔD/Dv=(q0/2).
It has been discovered that the space focus and velocity focus can be made to coincide by adjusting the value of y so that
Ds=Dv−ΔD=Dv(1−q0/2).
The focus position as a function of mass can be expressed as
(Dv/2d)=(1+q)(V/Vp)
where q=qo[1+2(Dea/d1)(1−(m0/m)1/2}] and m0 is the mass of the ion focused to first order at the high resolution timed ion selector 228. Dea=Des+Da, where Des is the effective length of the first accelerating field and Da is the distance from the end of the first field to the center of the pulsed accelerating field. The relative focusing error as function of mass is then equal to
ΔD/Dv=(q−q0)/(1+q0).
The maximum mass accelerated in the pulsed accelerator 220 under these conditions corresponds to q=2q0, and the minimum mass accelerated in the pulsed accelerator 220 under these conditions corresponds to q=0. Thus, the mass range that can be accelerated and focused is given by
mmax/mmin=[(1+d1/2Dea)/(1−d1/2Dea)]2.
The width of the peak at the selector 228 relative to the flight time is then given to first order by
δt/t=pΔD/D=p(q−q0)/(1+q0).
Since p1 and p2 are independent variables, the total effective perturbation accounting for all of the initial conditions is given by
p=[p12+p22]1/2 where
p1=[q0/(1+q0)[day/d1](δv0/vn) and
p2=[(1+q0)−1](δx/2day).
In general, the contribution to peak width is dominated by the velocity spread. In this case, the peak width of a mass in the range of accelerated masses is given by
δm/m=4(Deaday/Dv2)[1−(m0/m)1/2](δv0/vn).
Thus, precursor ions covering the full range of ions accelerated by pulsed accelerator 220 can be selected with high resolving power. Furthermore, the velocity spread of selected ions is given by p1 and is reduced relative to the velocity spread from the ions source.
Referring to both
In one embodiment, a first fragmentation chamber 240 is positioned in first field-free drift space 232. Ions accelerated by the first pulsed accelerator 220 and selected by the low resolution timed ion selector 224 enter into fragmentation chamber 240 where some of the precursor ions are fragmented. Ions exiting from fragmentation chamber 240 are separated with higher resolution by the high resolution timed ion selector 228. In some embodiments, ions transmitted by the ion selector 228 are fragmented further in the fragmentation chamber 260 positioned in the field-free space 250. Selected ions and fragment thereof are transmitted through entrance aperture 290 for the second time-of-flight mass spectrometer 20 (
A high resolution timed ion selector 228 is positioned at the simultaneous velocity and space focus of first time-of-flight mass spectrometer 200. In one embodiment, the timed ion selector 228 is a Bradbury-Nielsen type ion shutter or gate. A Bradbury-Nielsen type ion shutter or gate is an electrically activated ion gate. Bradbury-Nielsen timed ion selectors include parallel wires that are positioned orthogonal to the path of the ion beam. High-frequency voltage waveforms of opposite polarity are applied to alternate wires in the gate. The gates only pass charged particles at certain times in the waveform cycle when the voltage difference between wires is near zero. At other times, the ion beam is deflected to some angle by the potential difference established between the neighboring wires. The wires are oriented so that ions rejected by the timed ion selector 228 are deflected away from the entrance aperture 290 for the second time-of-flight mass spectrometer 20 (
A first ion fragmentation chamber 240 is positioned in the field-free space 232 between the output of the low resolution timed ion selector 224 and the high resolution timed ion selector 228. A second fragmentation chamber 260 is positioned between the output from high resolution timed ion selector 228 and the entrance aperture 290 to second time-of-flight mass spectrometer 20 (
The deflection of ions is proportional to the distance of the ions from the plane of the entrance aperture at the time the polarity switches. The mass resolving power can be adjusted by varying the amplitude of the voltage applied to the wires and is only weakly affected by the speed of the transition. In one embodiment where precise measurements are required, a power supply provides the wires of the Bradbury-Nielsen ion selector with an amplitude of approximately +/−500 volts with a 7 nsec switching time.
In the embodiment depicted in
Referring also to
The equations for calculating the performance of a single Bradbury-Nielsen type timed ion selector are known. Deflection angle can be determined from the following equation assuming that the voltage is turned on when the ion is at position x0 and then turned off when the ion is at position x1 relative to the plane of the gate:
tan α(x0,x1)=k(Vp/V0)[(2/π)tan−1({exp((πx1/de)}−(2/π)tan−1{exp(πx0/de)}],
where k is a deflection constant given by k=π{2 ln [cot(πR/2d)]}−1, Vp is the deflection voltage (+Vp on one wire set, −Vp on the other), V0 is the accelerating voltage of the ions, and de is the effective wire spacing given by de=d cos [(π(d−2R)/4d], where d is the distance between wires and R is the radius of the wire. The angles are expressed in radians.
For this calculation, the origin for the ion travel along the x axis is located at the plane of the selector. Thus, ions approaching the Bradbury-Nielsen type timed ion selectors are located at a negative x position and ions leaving the Bradbury-Nielsen type timed ion selectors are located at a positive x position. For continuous application of the deflection voltage, x0 goes to negative infinity, and x1 goes to positive infinity. Thus, for a continuous deflection voltage, the deflection angle can be expressed by the following equation:
tan αmax=2k(Vp/V0).
High resolution selection using a dual Bradbury-Nielson gate as depicted in
tan α=2k(Vp/V0))[1−(2/π)tan−1({exp((πx1/d)}].
The deflection voltage for the second gate 328 (
tan α=−2k(Vp/V0)[(2/π)tan−1({exp((πx2/de)}−1].
Deflection by second gate 328 (
As illustrated in
The first time-of-flight analyzer 150 provides simultaneous space and velocity focusing for an ion of predetermined mass-to-charge ratio at the timed ion selector 178, and also minimizes the focusing error for ions within a predetermined mass range including the focused mass. In some embodiments, the field-free drift spaces 168 and 172 comprise fragmentation chambers wherein ions may fragment spontaneously as the result of internal excitation in the ion source or as the result of excitation by collisions with neutral molecules in field-free spaces 168 or 172. In some embodiments, the pressure in at least one of the field-free regions 168 or 172 is increased to enhance excitation by collisions with neutral molecules. In some embodiments, field-free regions 168 or 172 may be enclosed and differential pumped employed to allow the pressure in these regions to be increased without increasing the pressure in other regions of the tandem mass spectrometer.
The addition of the ion mirror 158 provides a longer flight path between the ion source 152 and the high resolution timed ion selector 178 relative to the flight time between the ion source 208 and the high resolution timed ion selector 228 in the embodiment illustrated in
When the ions selected by the first time-of-flight mass spectrometer substantially reach the center 403 of the pulsed accelerator 404, an accelerating voltage pulse Vp 432 is applied to the ion accelerator 404. In one embodiment, a timed ion selector 414 is positioned in the field-free region 416 between the exit 405 from the pulsed accelerator 404 and the static accelerating field 406. The timed ion selector 414 is energized to reject fragment ions within a predetermined mass range from each selected precursor ions.
The first time-of-flight mass analyzer 612 comprises an ion source 702, a pulsed ion accelerator 708, a low resolution timed ion selector 710, a first field-free drift space 714, a high resolution timed ion selector 716, and a second field-free drift space 718. The ion source 702 generates a pulse of ions. The pulsed ion accelerator 708 accelerates the pulse of ions. The low resolution timed ion selector 710 transmits a range of masses accelerated in pulsed accelerator 708 and rejects all others. The high resolution timed ion selector 716 transmits a predetermined set of precursor ions accelerated by pulsed ion accelerator 708.
The second stage time-of-flight mass spectrometer 620 according to the present teaching comprises a pulsed ion accelerator 804 positioned adjacent to the entrance 862 of the second stage time-of-flight mass spectrometer 620, a static electric field region 805, a field-free region 810, and an ion detector 808 at the end of region 810. In one embodiment, an ion mirror (not shown) is located in field-free region 810 between the exit from static electric field region 805 and detector 810. A pulsed potential Vp 832 is applied to the pulsed ion accelerator 804 and a static potential Va 834 is applied to the static electric field region 805. Both the pulsed potential Vp 832 and the static potential Va 834 are chosen such that ions are focused at the ion detector 808. The ion detector 808 can be electrically connected to a transient digitizer 830, which can perform signal averaging and other signal processing.
When the ions selected by the first time-of-flight mass spectrometer substantially reach the center of the pulsed accelerator 804, the accelerating voltage pulse Vp 832 is applied to the ion accelerator 804. In one embodiment a timed ion selector 814 is positioned between the exit of the pulsed accelerator 804 and the static accelerating field region 805. The timed ion selector 814 is energized to reject fragment ions within a predetermined mass range from each selected precursor ions.
The tandem time-of-flight mass spectrometer 600 according to the present teaching further comprises a static high voltage generator 900, a pulsed high voltage generator 910, and a multiplexed time delay generator 920. In one specific embodiment, the outputs of the generators 900 and 910, the transient digitizer 830, and the time delay generator 920 are controlled by a processor or by a computer 930. The static high voltage generator 900 provides static high voltages (including ground potential) to all the elements comprising the tandem time-of-flight mass spectrometer 600. The magnitude of these voltages is controlled by the computer 930 to an appropriate level that focuses the ions. The computer 930 executes algorithms that calculate the appropriate static and pulsed high voltages and time delays required to focus ions of predetermined mass-to charge ratio. The computer 930 also interfaces with and controls the high voltage generators 900 and 910 and the multiplexed time delay generator 920. The pulsed high voltage generator 910 provides pulsed voltages to the ion source 702, the pulsed accelerator 708, the low resolution timed ion selector 710, the high resolution timed ion selector 716, the pulsed accelerator 804, and the timed ion selector 814. The amplitudes of the pulsed voltages are controlled by computer 930. Computer 930 also programs the multiplexed time delay generator 920 to control the timing of the pulses produced by pulsed high voltage generator 910 as required to accelerate and focus the ions. Signals generated by the digitizer 830 are transmitted to the computer 930 for processing the ion intensities as a function of flight time into calibrated mass spectra. The computer 930 also controls the time and input voltage ranges of digitizer 830.
It should be understood by those skilled in the art that the schematic diagrams shown in the Figures are only schematic representations and that various additional elements would be necessary to complete a functional mass spectrometer according to the present teachings, including power supplies, delay generators, and a vacuum housing. In addition, a vacuum pumping arrangement is required to maintain the operating pressures in the vacuum chamber housing of the mass spectrometer at the desired operating levels. In various embodiments, differential vacuum pumping is employed.
The tandem time-of-flight mass spectrometer according to the present teaching provides high mass resolving power for precursor selection for both MS and MS-MS spectra. In various embodiments, the mass spectrometer can be configured for either positive or negative ions, and can be readily switched from one type of ion to the other type of ions.
Tandem mass spectrometry according to the present teaching provides information on the structure and sequence of many biological polymers and allows unknown samples to be accurately identified. Tandem mass spectrometers according to the present teaching employ a first mass analyzer to produce, separate and select a precursor ion, and a second mass analyzer to fragment the selected ions and record the fragment mass spectrum from the selected precursor. A wide variety of mass analyzers and combinations thereof for use in tandem mass spectrometry can be used with the present teaching. One aspect of the present teaching employs simultaneous space and velocity focusing in a time-of-flight mass spectrometer which allows simultaneous high resolution selection of multiple precursor ions and rapid and accurate determination of masses of fragment ions from selected precursors.
For example, one method for identifying an unknown sample, such as a biological polymer, using a tandem mass spectrometer according to the present invention includes generating an ion beam comprising a plurality of ions. In some methods, the ion beam is generated with MALDI. At least one monoisotopic precursor ion is then selected from the plurality of ions using a first time-of-flight mass spectrometer configured to perform simultaneous space and velocity focusing. In some embodiments, a predetermined portion of the fragment ions from each monoisotopic precursor are selected. At least one of the selected monoisotopic precursor ions is then fragmented. The fragmented selected monoisotopic precursor ions are separated with a second time-of-flight mass analyzer so that a flight time of precursor ions and fragments thereof to a detector is dependent on a mass-to-charge ratio of the selected precursor ions and fragments thereof and is nearly independent of a velocity distribution of the selected precursor ions and fragments thereof. The separated fragmented ions are then detected with a detector and the fragment ion mass spectra are recorded for at least one selected precursor ion. Some methods for identifying an unknown sample according to the present teaching elucidate at least one of a structure and a sequence of the unknown sample.
In one embodiment, single isotopes can be selected and fragmented up to m/z 2500 with no detectable loss in ion transmission and less than 1% contribution from adjacent masses. In some cases ten or more monoisotopic precursor ions can be selected simultaneously and fragmented to produce fragment ions. This allows generation of very high quality MS-MS spectra at unprecedented speed. For example, all of the peptides present in a complex peptide mass fingerprint containing a hundred or more peaks can be fragmented and identified without exhausting the sample by using a mass spectrometer according to the present teaching. Thus, speed and sensitivity of the MS-MS measurements can keep pace with the MS results, and high-quality, interpretable MS-MS spectra can be generated on detected peptides at very low concentrations.
The present teaching employing simultaneous space and velocity focusing provides a method for accurate and sensitive quantization of low levels of selected samples in complex mixtures. Quantitative mass spectrometry generally requires using labeled standards, but unlike other instruments, the method of the present teaching allows simultaneous measurement of multiple components, and the entire fragment spectrum for each can be recorded to improve sensitivity and accuracy. Furthermore, both sample and standard can be acquired at the same time in the same spectrum, and all of the labeled fragments show up as doublets. Quantization is accomplished by measuring the relative intensities of the doublets, thus improving both the accuracy and precision of the measurements since potential interferences are drastically reduced.
For example, a method for quantifying an unknown sample using a tandem mass spectrometer according to the present teaching includes generating an ion beam comprising a plurality of ions and then selecting at least two monoisotopic precursor ion from the plurality of ions using a first time-of-flight mass spectrometer configured to perform simultaneous space and velocity focusing. At least one of the selected precursor ions can be a molecular ion of a known molecule present at a predetermined concentration in the sample. At least two of the selected monoisotopic precursor ions are then fragmented. The fragmented selected monoisotopic precursor ions are separated with a second time-of-flight mass analyzer so that a flight time of precursor ions and fragments thereof to a detector is dependent on a mass-to-charge ratio of the selected precursor ions and fragments thereof and is nearly independent of a velocity distribution of the selected precursor ions and fragments thereof. The separated fragmented ions are detected with a detector and then the fragment ion mass spectra for at least two selected precursor ion is recorded.
While the Applicant's teachings are described in conjunction with various embodiments, it is not intended that the Applicant's teachings be limited to such embodiments. On the contrary, the Applicant's teachings encompass various alternatives, modifications, and equivalents, as will be appreciated by those of skill in the art, which may be made therein without departing from the spirit and scope of the teaching.
The present application is a continuation-in-part of U.S. patent application Ser. No. 12/549,076, filed on Aug. 27, 2009. The present application is also a continuation-in-part of U.S. patent application Ser. No. 12/968,254, filed on Dec. 14, 2010. The present application is also a continuation-in-part of U.S. patent application Ser. No. 13/034,525, filed on Feb. 24, 2011, which is a continuation-in-part of U.S. patent application Ser. No. 12/968,254, filed on Dec. 14, 2010. The entire contents of U.S. patent application Ser. Nos. 12/549,076, 12/968,254, and 13/034,525 are all herein incorporated by reference.
Number | Name | Date | Kind |
---|---|---|---|
4960992 | Vestal et al. | Oct 1990 | A |
5087815 | Schultz et al. | Feb 1992 | A |
5144127 | Williams et al. | Sep 1992 | A |
5160840 | Vestal | Nov 1992 | A |
5166518 | Freedman | Nov 1992 | A |
5510613 | Reilly et al. | Apr 1996 | A |
5625184 | Vestal et al. | Apr 1997 | A |
5627369 | Vestal et al. | May 1997 | A |
5847385 | Dresch | Dec 1998 | A |
5861623 | Park | Jan 1999 | A |
6057543 | Vestal et al. | May 2000 | A |
6300627 | Koster et al. | Oct 2001 | B1 |
6489610 | Barofsky et al. | Dec 2002 | B1 |
6512225 | Vestal et al. | Jan 2003 | B2 |
6534764 | Verentchikov et al. | Mar 2003 | B1 |
6621074 | Vestal | Sep 2003 | B1 |
6661001 | Park | Dec 2003 | B2 |
6872941 | Whitehouse et al. | Mar 2005 | B1 |
7214320 | Gregori et al. | May 2007 | B1 |
7223966 | Weiss et al. | May 2007 | B2 |
7355169 | McLuckey et al. | Apr 2008 | B2 |
7498588 | Gordon et al. | Mar 2009 | B1 |
7564026 | Vestal | Jul 2009 | B2 |
7564028 | Vestal | Jul 2009 | B2 |
7589319 | Vestal | Sep 2009 | B2 |
7663100 | Vestal | Feb 2010 | B2 |
7667195 | Vestal | Feb 2010 | B2 |
7932491 | Vestal | Apr 2011 | B2 |
20020158194 | Vestal et al. | Oct 2002 | A1 |
20030141447 | Verentchikov et al. | Jul 2003 | A1 |
20030180807 | Hess et al. | Sep 2003 | A1 |
20040119012 | Vestal | Jun 2004 | A1 |
20040203175 | Li et al. | Oct 2004 | A1 |
20040213721 | Arno et al. | Oct 2004 | A1 |
20050035285 | Tan et al. | Feb 2005 | A1 |
20050116162 | Vestal | Jun 2005 | A1 |
20050242279 | Verentchikov | Nov 2005 | A1 |
20050269505 | Ermer | Dec 2005 | A1 |
20050279933 | Appelhans et al. | Dec 2005 | A1 |
20060266941 | Vestal | Nov 2006 | A1 |
20080067349 | Moskovets et al. | Mar 2008 | A1 |
20080078931 | Vestal et al. | Apr 2008 | A1 |
20080272291 | Vestal | Nov 2008 | A1 |
20090294658 | Vestal et al. | Dec 2009 | A1 |
20100181473 | Blenkinsopp et al. | Jul 2010 | A1 |
20100276585 | Lehmann | Nov 2010 | A1 |
20100301202 | Vestal | Dec 2010 | A1 |
20110049350 | Vestal | Mar 2011 | A1 |
20110155901 | Vestal | Jun 2011 | A1 |
20110266431 | Vestal | Nov 2011 | A1 |
Number | Date | Country |
---|---|---|
00-77823 | Dec 2000 | WO |
2004-030025 | Apr 2004 | WO |
2006-064280 | Jun 2006 | WO |
2006064280 | Jun 2006 | WO |
2010-138781 | Dec 2010 | WO |
Entry |
---|
“Notification of Transmittal of the International Search Report and the Written Opinion of the International Searching Authority, or the Declaration” for PCT/US2011/063855, Jul. 27, 2012, 11 pgs., International Searching Authority/Korea, Korean Intellectual Property Office, Daejeon Metropolitan City, Republic of Korea. |
“Notification of Transmittal of the International Search Report and the Written Opinion of the International Searching Authority, or the Declaration” for PCT/US2012/025761, Sep. 25, 2012, 9 pages, International Searching Authority/KR, Daejeon Metropolitan City, Republic of Korea. |
“Notification Concerning Transmittal of International Preliminary Report on Patentability (Chapter I of the Patent Cooperation Treaty” for PCT/US2010/060902, Jul. 12, 2012, 7 pages, The International Bureau of WIPO, Geneva, Switzerland. |
Beavis, Ronald C., et al., Factors Affecting the Ultraviolet Laser Desorption of Properties, Rapid Communications in Mass Spectrometry, 1989, pp. 233-237, vol. 3 No. 9, Heyden & Son Limited. |
Bergmann, T., et al., High-Resolution Time-of-Flight Mass Spectrometer, Rev. Sci. Instrum., Apr. 1989, pp. 792-793, vol. 60, No. 4, American Institute of Physics. |
Beussman, Douglas J., et al., Tandem Reflectron Time-of-Flight Mass Spectrometer Utilizing Photodissociation, Analytical Chemistry, Nov. 1, 1995, pp. 3952-3957, vol. 67, No. 21, American Chemical Society. |
Colby, Steven M., et al., Space-Velocity Correlation Focusing, Analytical Chemistry, Apr. 15, 1996, pp. 1419-1428, vol. 68, No. 8, American Chemical Society. |
Cornish, Timothy J., et al., A Curved Field Reflectron Time-of-Flight Mass Spectrometer for the Simultaneous Focusing of Metastable Product Ions, Rapid Communication in Mass Spectrometry, 1994 pp. 781-785, vol. 8, John Wiley & Sons. |
Cornish, Timothy J., et al., Tandem Time-of-Flight Mass Spectrometer, Analytical Chemistry, Apr. 15, 1993, pp. 1043-1047, vol. 65, No. 8. |
Hillenkamp, F., Laser Desorption Mass Spectrometry: Mechanisms, Techniques and Applications, 1989, pp. 354-362, vol. 11A, Heyden & Son, London. |
Kaufmann, R., et al., Mass Spectrometric Sequencing of Linear Peptides by Product-Ion Analysis in a Reflectron Time-of-Flight Mass Spectrometer Using Matrix Assisted Laser Desorption Ionization, Rapid Communications in Mass Spectrometry, 1993, pp. 902-910, vol. 7, John Wiley & Sons, Ltd. |
Mamyrin, B.A. et al., The Mass-Reflectron, A New Nonmagnetic Time-of-Flight Mass Spectrometer With High Resolution, Sov. Phys. 1973, pp. 45-48, vol. 37, No. 1, American Institute of Physics. |
Matsuda, H., et al., Particle Flight Times Through Electrostatic and Magnetic Sector Fields and Quadrupoles to Second Order, International Journal of Mass Spectrometry and Ion Physics, 1982, pp. 157-168, vol. 42, Elsevier Scientific Publishing Company, Amsterdam, The Netherlands. |
Neuser, H. J., et al., High-Resolution Laser Mass Spectrometry, International Journal of Mass Spectrometry and Ion Process, 1984, pp. 147-156, vol. 60, Elsevier Science Publishers B.V., Amsterdam, The Netherlands. |
Vestal, M. L., et al., Delayed Extraction Matrix-Assisted Laser Desorption Time-of-Flight Mass Spectrometry, Rapid Communications in Mass Spectrometry, 1995, pp. 1044-1050, vol. 9, John Wiley & Sons, Ltd. |
Vestal, M. L., et al., Resolution and Mass Accuracy in Matrix Accuracy in Matrix-Assisted Laser Desorption Ionization-Time-of-Flight, American Society for Mass Spectrometry, 1998, pp. 892-911, Elsevier Science, Inc. |
Vestal, M., High Performance MALDI-TOF Mass Spectrometry for Proteomics, International Journal of Mass Spectrometry, 2007, pp. 83-92. |
Wiley, W. C., et al., Time-of-Flight Mass Spectrometer With Improved Resolution, The Review of Scientific Instruments, Dec. 1955, pp. 1150-1157, vol. 26, No. 13. |
Zhou, J. Kinetic Energy Measurements of Molecular Ions Ejected Into an Electric Field by Matrix-Assisted Laser Desorption, Rapid Communications in Mass Spectrometry, Sep. 1992, pp. 671-678, vol. 6, John Wiley & Sons, Ltd. |
Vestal, M., Linear Time-of-Flight Mass Spectrometry With Simultaneous Space and Velocity Focusing, U.S. Appl. No. 12/968,254, filed Dec. 14, 2010. |
“Notification of Transmittal of the International Search Report and the Written Opinion of the International Searching Authority, or the Declaration” for PCT/US2010/022122, Aug. 16, 2010, 9 pages, International Searching Authority, Korean Intellectual Property Office, Seo-gu, Daejeon, Republic of Korea. |
“Notification Concerning Transmittal of International Preliminary Report on Patentability (Chapter I of the Patent Cooperation Treaty)” for PCT/US2009/045108, Dec. 9, 2010, 9 pages, The International Bureau of WIPO, Geneva, Switzerland. |
“Notification of Transmittal of the International Search Report and the Written Opinion of the International Searching Authority, or the Declaration” for PCT/US2010/046074, Apr. 15, 2011, 8 pages, International Searching Authority, Korean Intellectual Property Office, Seo-gu, Daejeon, Republic of Korea. |
“Notification of Transmittal of the International Search Report and the Written Opinion of the International Searching Authority, or the Declaration” for PCT/US2010/036501, Jan. 4, 2011, 9 pages, International Searching Authority, Korean Intellectual Property Office, Seo-gu, Daejeon, Republic of Korea. |
“Notification Concerning Transmittal of International Preliminary Report on Patentability (Chapter I of the Patent Cooperation Treaty” for PCT/US2010/046074, Mar. 8, 2012, 5 pages, The International Bureau of WIPO, Geneva, Switzerland. |
“Office Action” for U.S. Appl. No. 12/651,070, Dec. 9, 2011, 31 pages. |
“Office Action” for U.S. Appl. No. 12/651,070, May 24, 2011, 50 pages. |
“Office Action” for U.S. Appl. No. 12/968,254, May 14, 2012, 23 pages. |
“Notification Concerning Transmittal Of International Preliminary Report On Patentability” for PCT/US2012/025761, Sep. 6, 2013, 6 pages, The International Bureau of WIPO, Geneva, Switzerland. |
“Notification Of Transmittal Of The International Search Report And The Written Opinion Of The International Searching Authority, Or The Declaration” for PCT/US2013/028953, Jun. 27, 2013, 13 pages, Korean Intellectual Property Office, Daejeon Metropolitan City, Republic of Korea. |
“Notification Concerning Transmittal Of International Preliminary Report On Patentability” for PCT/US2011/063855, Jun. 27, 2013, 8 pages, The International Bureau of WIPO, Geneva, Switzerland. |
Number | Date | Country | |
---|---|---|---|
20120168618 A1 | Jul 2012 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 12549076 | Aug 2009 | US |
Child | 13415802 | US | |
Parent | 12968254 | Dec 2010 | US |
Child | 12549076 | US | |
Parent | 13034525 | Feb 2011 | US |
Child | 12968254 | US | |
Parent | 12968254 | US | |
Child | 13034525 | US |