The present invention relates to systems and methods for quantitative measurement of phase in low-light microscopy of optically thick media, particularly unlabeled cells.
It has become increasingly clear that understanding morphogenesis and disease requires three-dimensional tissue cultures and models. Effective 3D imaging techniques, capable of reporting on subcellular as well as multicellular scales, in a time-resolved manner, are crucial for achieving this goal. While the light microscope has been the main tool of investigation in biomedicine for four centuries, the current requirements for 3D imaging pose new, difficult challenges. Due to their insignificant absorption in the visible spectrum, most living cells exhibit very low contrast when imaged by visible light microscopy. Consequently, fluorescence microscopy has become the main tool of investigation in cell biology. Due to extraordinary progress in designing fluorescence tags, structures in the cell can be imaged with high specificity.
More recently, super-resolution microscopy methods based on fluorescence have opened new directions of investigation, toward the nanoscale subcellular structure. However, fluorescence imaging is subject to several limitations. Absorption of the excitation light may cause the fluorophore to irreversibly alter its molecular structure and stop fluorescing. This process, known as photobleaching, limits the time interval over which continuous imaging can be performed. The excitation light is typically toxic to cells, a phenomenon referred to as phototoxicity.
Addressing cell cultures in particular, transmitted light modalities appear to be ideal for studying cell growth and proliferation due to low phototoxicity, an absence of photobleaching and easy sample preparation. Yet, such assays are most frequently conducted with the aid of labels. While specificity granted by external markers is crucial for certain applications, quantifying cell growth over longer time scales remains a grand challenge. It has been known for some time that indicators of cell proliferation do not have equal growth. More recent methods using vibrating hollow cantilevers to weigh cell passing through are limited to non-adherent cells. Methods based on vibrating pedestals have also been demonstrated, but at the expense of mass sensitivity.
Label-free microscopy provides a solution to overcoming these limitations. Two classical methods are phase contrast (PC) microscopy and differential interference contrast (DIC) microscopy. Both PC and DIC microscopy indicate modifications to the wavefront of incident light propagating through a sample. However, to date, neither of them has provided a quantitative measure of wavefront deformation in three dimensions.
Along these lines, Cogswell et al., “Quantitative DIC microscopy using a geometric phase shifter”, Proc. SPIE 2984, Three-Dimensional Microscopy: Image Acquisition and Processing IV, (1997) proposed DIC with geometrically-induced phase shifting, applied for two-dimensional (2D) imaging. DIC with two orthogonal shear directions has been used to obtain 2D quantitative phase images, as described, for example by King et al., “Quantitative phase microscopy through differential interference imaging,” Journal of Biomedical Optics, vol. 13(2), 024020 (2008). Both of these references are incorporated herein by reference.
Quantitative phase imaging (QPI) is an approach that quantitatively ascertains the phase shift in a wavefront propagating through a refractive medium such as a cell. QPI is described in Popescu, Quantitative Phase Imaging of Cells and Tissues, (McGraw-Hill, 2011), which is incorporated herein by reference. However, imaging optically thick, multiple scattering specimens is challenging for any optical method, including QPI. The fundamental obstacle is that multiple scattering generates an incoherent background, which ultimately degrades the image contrast. An imaging method dedicated to imaging optically thick specimens must include a mechanism to subdue the multiple scattering backgrounds and exhibit strong spatial sectioning to suppress the out of focus light.
The invention described below in directed toward solving the problem of quantitatively measuring phase shift as a function of position in three-dimensions within a sample, even if it is optically thick and gives rise to multiple scattering.
In accordance with embodiments of the invention, a method is provided for quantitative optical phase imaging of a sample characterized by a surface. The method has steps of:
generating a first and a second replica field of an image field, each of the image field and the first and second replica fields characterized by a respective optical phase, the second replica field cross-polarized and shifted in a shift direction transverse to a normal to the surface of the sample by a transverse offset;
spatially Fourier transforming the first and second replica fields;
retarding the second replica field by four successive phase shifts;
inverse Fourier transforming the first and second replica fields;
passing the first and second replica field through an analyzer polarizer and superposing the first and second replica fields on a detector array to create four successive detector signals; and
solving the four successive detector signals to derive a gradient of the optical phase of the image field.
In accordance with other embodiments of the invention, retarding the second replica field may include employing a spatial light modulator. The four successive phase shifts may be integral multiples of π/2, and the transverse offset may be smaller than a diffraction spot characterizing an optical system whereby the first and second replica fields are obtained.
In accordance with further embodiments of the invention, the method may also have a step of integrating the gradient of the optical phase to obtain an image of the optical phase of the image field. Solving the four successive detector signals may also include deriving a mutual intensity function between the first and second replica fields.
In accordance with another aspect of the present invention, a system is provided for quantitative optical phase imaging of a sample. The system has a microscope for generating a first and a second replica field of an image field, where each of the image field and the first and second replica fields is characterized by an optical phase. The second replica field is cross-polarized and shifted in a shift direction transverse to a normal to the surface of the sample by a transverse offset. The system also has a first lens for spatially Fourier transforming the first and second replica fields, a spatial light modulator for retarding the second replica field by four successive phase shifts; and a second lens for inverse Fourier transforming the first and second replica fields. An analyzer polarizer is aligned at 45° with respect to respective polarizations of the first and second replica fields, A detector array creates four successive detector signals, and a processor is provided for solving the four successive detector signals to derive a gradient of the optical phase of the image field.
In accordance with other embodiments of the invention, the microscope may includes a Wollaston prism for separating the first and second replica fields.
In accordance with yet a further aspect of the present invention, a module is provided for receiving a first and a second replica field of an image field of a sample characterized by a surface and for quantitative optical phase imaging of the sample. The module has a first lens for spatially Fourier transforming the first and second replica fields, a spatial light modulator for retarding the second replica field by four successive phase shifts, a second lens for inverse Fourier transforming the first and second replica fields, and an analyzer polarizer aligned at 45° with respect to respective polarizations of the first and second replica fields.
In accordance with yet another aspect of the present invention, a method is provided for assessing viability of an embryo. The method has steps of:
generating a first and a second replica field of an image field containing the embryo, each of the image field and the first and second replica fields characterized by a respective optical phase, the second replica field cross-polarized and shifted in a shift direction transverse to a normal to the surface of the sample by a transverse offset;
spatially Fourier transforming the first and second replica fields;
retarding the second replica field by four successive phase shifts;
inverse Fourier transforming the first and second replica fields;
passing the first and second replica field through an analyzer polarizer and superposing the first and second replica fields on a detector array to create four successive detector signals;
solving the four successive detector signals to derive a gradient of the optical phase of the image field; and
applying morphokinetic criteria to a function of the gradient of the optical phase to characterize embryo viability.
The current patent or application file contains at least one drawing executed in color. Copies of this patent or patent application publication with color drawing(s) will be provided by the Office upon request and payment of the necessary fee.
The foregoing features of the invention will be more readily understood by reference to the following detailed description, taken with reference to the accompanying drawings, in which:
The following terms shall have the meanings indicated, unless otherwise dictated by context:
As used herein, the term “detector” shall refer broadly to a focal plane detector array and any associated optics constituting a camera and bringing fields to focus on a detector array.
The term “image” shall refer to any multidimensional representation, whether in tangible or otherwise perceptible form, or otherwise, whereby a value of some characteristic (amplitude, phase, etc.) is associated with each of a plurality of locations corresponding to dimensional coordinates of an object in physical space, though not necessarily mapped one-to-one thereonto. Thus, for example, the graphic display of the spatial distribution of some field, either scalar or vectorial, such as brightness or color, constitutes an image. So, also, does an array of numbers, such as a 3D holographic dataset, in a computer memory or holographic medium. Similarly, “imaging” refers to the rendering of a stated physical characteristic in terms of one or more images.
The terms “object,” “sample,” and “specimen” shall refer, interchangeably, to a tangible, non-transitory physical object capable of being rendered as an image.
In accordance with embodiments of the present invention, a new QPI method is described herein, referred to as gradient light interference microscopy (GLIM). GLIM combines DIC microscopy with low-coherence interferometry and holography. In GLIM, two interfering fields are identical except for a small transverse spatial shift. This geometry ensures that the two fields suffer equal degradation due to multiple scattering. By accurately controlling the phase shift between the two waves, multiple intensity images are acquired that share the same incoherent background, but differ with respect to coherent contributions. As a result, GLIM can reject much of the multiple scattering contributions and yield remarkable contrast of thick objects.
Furthermore, GLIM is performed, as described below, with a fully-open condenser aperture (not shown), providing, as will be described, a small numerical aperture {Inventors: right?} and, thus, very strong optical sectioning. GLIM may advantageously provide tomographic imaging of both thin samples, e.g., single cells, and thick specimens, such as multicellular systems. The principles of GLIM operation are provided below, along with validation results on test samples, and time-resolved tomography of cells in culture, as well as embryo development.
GLIM enjoys the benefits of common-path white-light methods including nanometer path length stability, substantially speckle-free, and diffraction-limited resolution. At the smallest condenser aperture, GLIM gives exact values of the quantitative phase for thin samples. At the largest condenser aperture, GLIM can be used as a tomography method, providing three-dimensional information of thick imaging samples. GLIM has been demonstrated on various samples, including beads, Hela cells and bovine embryos.
With reference, first, to
An analyzer polarizer (not shown) that would ordinarily be used in DIC to render the two polarizations parallel, is not used in the present invention. Instead, both replicas of the image field enter the GLIM module 102. These replica fields are spatially Fourier transformed by the lens L1 at its back focal plane, the plane of a spatial light modulator (SLM 116) in
Un(r)=U(r)+U(r+δr)eiϕ
where δr=δx{circumflex over (x)} is the spatial offset between the two replica fields and U is the image field.
The intensity for each phase shift, In(r)=|Un(r)|2, can be written as
In(r)=I(r)+I(r+δr)+2|γ(r,δr)|cos [ϕ(r+δr)−ϕ(r)+ϕn], (2)
where I(r) and ϕ(r) are, respectively, the intensity and phase of the image field, and γ is the mutual intensity, i.e., the temporal cross-correlation function between these two fields, evaluated at zero delay, γ(r,δr)=(U*(r)U(r+δR)t. The phase ϕn=nπ/2 is the modulated phase offset between the two fields, externally controlled by a spatial light modulator (SLM) 116, disposed. It is to be understood that phase offsets between the two fields that are not multiples of π/2 are similarly encompassed within the scope of the present invention as claimed.
Imaging the fields interfering at detector 116 at four respective phase offsets gives rise to four intensity images, In, =n=1, . . . , 4, as shown in
Given the phase gradient, ∇xϕ, one can integrate along the gradient direction to get phase value, ϕ(r), using
where ϕ(0, y) is the initial value, which can be obtained with prior knowledge with respect to aspects of the specimen. For example, if (0, y) is a background location, the phase ϕ(0, y) is set to 0 radians.
Growth and proliferation of large populations of adherent cells over extended periods of time is demonstrated with reference to
In order to measure growth rates, the phase values are obtained by integrating the phase gradient at each frame in a time lapse series. These phase values are used to calculate the cell dry mass, using the linear relationship between optical pathlength map of a cell and its dry mass density, as first described by Barrer, “Interference microscopy and mass determination,” Nature, vol. 169, pp. 366-67 (1952), incorporated herein by reference. Quantitative phase imaging may be used to extract cell growth data based on this principle. A mass doubling period of approximately 36 hours was observed, which is 50% longer than the typical cell number doubling time, as cells can divide without doubling in mass.
Due to the high numerical aperture of the illumination, excellent sectioning capabilities may be advantageously achieved in accordance with embodiments of the present invention, yielding tomographic imaging of both thin and thick samples. Referring now to
Each FOV was scanned every 22 minutes. For each time point, the sample was scanned over a total depth of 28 μm with a step size of Δz=0.07 μm.
Compared to the phase gradient image, ∇xϕ, this cross-section has no diffraction streaks or “shadow” artifacts and clear cell boundaries.
By virtue of the depth sectioning provided by GLIM, several parameters may be computed for each cell and their time evolution may be studied.
The percentage of live births from Assisted Reproductive technology (ART) procedures is still rather low. One reason is because of the lack of objective and accurate evaluation of the embryo quality and viability before transfer. Morphological assessment is currently the main method used to determine embryo viability during in-vitro fertilization (IVF) cycles. However, studies have shown that the predictive power of the typical day 2 and 3 assessment of morphological parameters has remained low.
GLIM may be advantageously employed to perform tomography on optically thick specimens such as embryos. In one experiment, 60 bovine embryos were imaged, starting at 12 hours after fertilization, sampling every 30 minutes, over a seven-day period, using a 40×/0.75 NA objective. The lipid droplets, prominent in bovine embryos can be clearly identified. Results obtained in accordance with the present invention show that the embryo internal dynamics changes completely when the embryo dies. Specifically, the internal mass transport halts almost entirely, which suggest either a great increase in viscosity of the material or that the dynamic transport is mostly due to molecular motors, which stop in dead cells.
In accordance with embodiments of the present invention, a dynamic index marker (DIM) is based on GLIM data. For each image, the phase difference Δϕ(r⊥, t) and the mutual intensity γ(r⊥, δr, t) are combined at each time point t. To measure morphological changes, the time derivative image of γ is computed, i.e., γ′t(r⊥)=dγ(r⊥, δr, t)/dt. Based on γ′t, the spatial Cumulative Distribution Function (CDF) of the time derivative images is computed, Ft(x)=P({γ′t(r⊥)}r
DIM=[Dt/maxt
During periods of inactivity, the spatial distribution of γ′t across the embryo is substantially uniform and the histogram is narrow compared to periods of higher activity. It has been found that between the point of apparent normal, dynamic behavior and the one complete lack of dynamics, there exists a continuous process that lasts many hours, as shown in
Three-dimensional GLIM stacks of bovine embryos were obtained at different development stages. A 63×/1.4 NA oil immersive objective was employed at a transverse sampling rate of 10 pixel/μm. The condenser aperture was fully opened NAcon=0.55 to maximize the depth sectioning and spatial resolution. The embryos were scanned in the axial dimension over an interval of [−120 μm, 120 μm] with a step of Δz=0.05 μm.
Procedures:
Phase Gradient Extraction from Intensity Images
The intensity image at modulation ϕm is given by
In(r)=I(r)+I(r+δr)+2|γ(r,δr)|cos [Δϕ(r)+ϕn],
where Δϕ=ϕ(r+δr)−ϕ(r)≈∇x(ϕ)δx, the phase difference of interest, and ∇xϕ is the gradient of the phase of the image field in the x-direction. The spatial shift δx is the transverse displacement introduced by the DIC prism (also referred to herein as the lateral offset between the two DIC beams), estimated experimentally from measurements of the test samples. The quantity γ(r,δr) is the mutual intensity, i.e., the temporal cross-correlation function between these two fields, evaluated at zero delay, γ(r,δr)=U*(r)U(r+δr)t. Combining the four intensity frames, the phase gradient is obtained as
∇xϕ(r)=arg {[I4(r)−I2(r)],[I3(r)−I1(r)]}/δx
DIC Shear Estimation
The lateral offset δx between the two DIC beams relates the phase difference image, Δϕ linearly to a quantitative phase gradient ∇xϕ via the relation Δϕ=∇xϕδx. This parameter is known to the microscope manufacturer (Zeiss, Olympus, etc.), however it may not be known to the user. To estimate the spacing between the two beams, the associated phase shift is matched to a known calibration sample. In a preferred procedure, a fine tomographic stack of a small object (300 nm bead) is acquired, and line integration is performed in the direction of the DIC gradient. It is to be expected that the phase is maximized at the plane of best focus. The peak of the integrated phase corresponds to the theoretical maximum phase shift due by the control structure of φ*=(2πdΔn/λo), where Δn is the refractive index mismatch between the sample and the background, λ0 is the central wavelength, and d is the diameter of the bead. However, given a pixel dimension of p (in units of μm), the following relationship obtains:
where xo, x* are locations of the background and the center of the bead. These locations correspond to pixel indices of k0, and k*, respectively. Combining these two relations yields:
Spatial Light Modulator (SLM) Calibration
SLM gray values are calibrated against modulating phase by imaging the sample plane without the sample, i.e., Δϕ(r)=0. First, intensity images, Ig are acquired, where g corresponds to the SLM grayscale values on an 8-bit dynamic range. As a side note, one can further extend the resolution of the SLM modulation to one more bit by mixing discrete values on the SLM using a “checkerboard” pattern. More specifically, the modulation value set to the pixel x is given as
The intensity image Ig recorded by the camera for the grayscale value g is
Ig=I1+I2+2√{square root over (I1I2)} cos [ϕ(g)],
where I1 is the intensity of the modulated field, I2 is the intensity of the un-modulated field, and ϕ(g) is the phase modulation of interest. Using an empirical I(g) curve for g∈[0, 255], we obtain cos [ϕ(g)]={[Ig−α]/β}, where +={[maxg[Ig]+ming[Ig]]/2}, and β={[maxg[Ig]−ming[Ig]]/2}. Next, a Hilbert transform is used to obtain the complex analytic signal associated with the cosine signal, the imaginary part of which is sin [ϕ(g)]=H{[Ig−α]/β}. Here, H(.) denotes the operator to compute the Hilbert transform from a real part of a complex analytic signal to get the imaginary part. Combining the sine(.) and the cosine(.) signals, one obtains:
ϕ(g)=arg [Ig−α]/β,H{[Ig−α]/β}.
The ϕ(g) curve represents the desired SLM calibration. Note that there are many sets of four points that can be chosen for the working phase shift. Preferably, the portion of the curve that is most linear is to be chosen. From this curve, the π/2 sequence is chosen to meet two complementary criteria: maximum visibility and minimal phase error. To maximize the visibility, we evaluate several points around each amplitude maxima and choose the calibration where the steps are nearest to multiples of π/2. When multiple peaks are present, the peak most closely resembling a sinusoid is preferred.
Two-Dimensional Image Formation in GLIM
Consider a thin sample with a transmission function T(r⊥), the total incident field on which consists of two cross-polarizations. The two cross-polarizations generate two sample fields on the camera plane of U1(r⊥)={[Ui(r⊥)T(r⊥)]{circle around (v)}r
Here, Γi(.) is the mutual intensity function of the illumination at the sample plane. When the numerical aperture of the object, NAo, is large enough so that the spatial resolution is finer than any structure of the interest in the sample, the point spread function (PSF) ho can be approximated as ho(r⊥)≈δ(2)(r⊥), simplifying the foregoing expression into
γ(r⊥,δr)=Γi(Δr=0)T(r⊥)T*(r⊥−δr).
Consequently,
Δϕ(r⊥)=arg[γ(r⊥,δr)]≈ϕ(r⊥)−ϕ(r⊥−{circumflex over (x)}δx).
with ϕ=arg(T), which means that GLIM gives the correct phase difference of the sample irrespective of the coherence of the illumination. This is different from other common-path interferometry methods where the measurement reduction is highly dependent upon NAcon, as discussed in Edwards, C. et al., “Effects of spatial coherence in diffraction phase microscopy,” Opt. Express, vol. 22, pp. 5133-5146 (2014), and Nguyen et al., “Quantitative phase imaging with partially coherent illumination,” Opt. Lett., vol. 39, pp. 5511-14 (2014), both of which are incorporated herein by reference.
Two-Dimensional Real-Time Interferometric Reconstruction
A software platform capable of mechanical automation and real-time phase retrieval was developed in order to fully automate GLIM data acquisition. An image acquisition platform is designed to overlap the GLIM computation with the operation of the camera, SLM, and microscope. The software is developed in C++ using the Qt framework. The real-time reconstruction runs on three threads with the first thread responsible for triggering new camera frames and modulating the SLM. The second thread receives incoming images and transferred them to the graphics card. The third thread is used to display the GUI and render the resulting phase maps (see
To remove the background curvature in real-time, Fourier band-pass filtering is performed to eliminate any slow-varying oscillation in the GLIM images. To improve the SNR of the reconstruction, mean intensities of the four frames are matched by proportionally adjusting the exposure time. For example, the exposure time is eight times longer for the extinction frame (π modulation) compared to maximum brightness frame (0π modulation). The longer exposure is later compensated numerically in extracting the phase. This operation results in an increased Signal-to-Noise Ratio (SNR), as illustrated by the standard deviation of the phase noise.
A GLIM system fabricated in accordance with one of the embodiments of the present invention operates at 10 phase images per second with a rendering rate at 40 frames per second. Since the computational portion is overlapped with acquisition, the rate-limiting factor in that system is exposure time. Thus, the longer exposure can be replaced by illumination with a brighter source.
Assembling the Two Dimensional Image Mosaic
To assemble the time-lapse scans, an image alignment and registration algorithm is applied that is capable of efficiently handling very large FOV.
The phase correlation algorithm proceeds by generating a list of connections along the X, Y, and time axis. This list is then sorted to improve performance by ensuring that mosaic tiles are read sequentially and relieved from computationally expensive steps such as disk access and the Fourier transform. To align the images, given two tiles, a(r⊥) and b(r⊥), we find a displacement vector r⊥n∈2, such that a(r⊥)=b(r⊥−r⊥n). This vector is a maximizer of the correlation function between the two tiles, i.e., r⊥n(a,b)=arg maxr
(a⊗b)(r⊥)=F−1{F(a)F(b)*}(r⊥).
The correlation function (a⊗b)(r⊥) may be masked with a searching window to refine dimensions of the searching area. To reduce the chance for secondary peaks, a two-pass peak detection may be implemented. In the first pass, a large search radius is used to gather the errors associated with each category of motion, as depicted in
After estimating the optimal transverse displacement, r⊥n(a,b), between every two neighboring tiles (a,b) using the phase correlation algorithm, an additional global optimization step is performed to find their best configuration of all tiles while taking into account the initial locations of the tiles, po[k], which are provided by the microscope. Let p†[k] be the optimal location of kth tile and p†={p†[1], . . . , p†[N]} be the best configuration, i.e., location vector, of all the tiles. To obtain p†, the following objective function is minimized:
Here, N(k) is the set of all tiles that have connections with tile kth. After solving for p†, it is used it to generate floating point .tiff files for quantitative analysis and rendered into the mipmap format used by Zoomify or TrakEM2 for easy access.
Using this procedure, zoomed-in images of the stitching results from all images in the whole FOV at different scales may be obtained.
Intracellular Mass Transport Calculation
The solution to the dispersion phase relation was computed directly on the phase gradient, ∇xϕ, which is possible because the it relates linearly to the dry mass density ρ(r⊥) More specifically, taking the two-dimensional Fourier transform of the phase gradient yields
F[∇xϕ](k⊥)=kxϕ(k⊥)=ξkxρ(k⊥),
where ξ is a constant, defined as ξ=2πα/λo with λo is central wavelength of the illumination and α≈0.2 ml/g the refractive index increment. F is the two-dimensional Fourier transform operator. Also, for heuristic convenience, ϕ(k⊥), ρ(k⊥) have been used to denote the two-dimensional Fourier transform of the phase and the dry mass density evaluated at the spatial frequency vector k⊥, respectively. Ignoring spatial frequencies along the kx=0 line, the phase gradient ∇xϕ can be used to compute the autocorrelation function g(k⊥,τ) instead of the dry mass density ρ, thus:
Automatic Segmentation of Hela Cells from GLIM Filtered Data
Automatic segmentation is advantageous in obtaining high-throughput, consistent, objective metrics on the cells during their development cycle. In accordance with an automatic segmentation procedure, an input stack of raw GLIM data consists of three cells with one of them in mitosis. First, the input is thresholded, hole-filtered, and morphologically transformed with the opening and closing with a 3×3×3 structure element. These steps eliminate spurious background noise, reduce the surface roughness, fill out gaps due to internal structures of the cells, and most importantly, generate a 3D binary map where a value of 1 are assigned to voxels inside the cells, and 0, otherwise. Then, the watershed algorithm is used to produce separating barriers splitting the binary map into multiple regions corresponding to different cells. Details of the watershed algorithm may be found in Vincent et al., “Watersheds in digital spaces: an efficient algorithm based on immersion simulations,” IEEE Transactions on Pattern Analysis & Machine Intelligence, vol. 13, pp. 583-98 (1991), incorporated herein by reference. The watershed algorithm uses an inverted distance map and some seeds generated as local maxima of the distance map. Here, the distance map is a scalar field where the value at each voxel is 0 if its value of the binary map is 0. On the other hand, when the binary is 1, the value at the voxel corresponds to the closest distance between it and the nearest boundary voxel. Next, the separating barriers are subtracted from the binary map and labels are given to connecting volumes in the resulting map. Finally, regions with volume measurement smaller than 900 μm3 or 300,000 voxels, equivalently, are eliminated.
Dry Mass Estimation Using GLIM
Barer, “Interference microscopy and mass determination,” Nature, vol. 169, pp. 366-67 (1952), incorporated herein by reference, showed that in two cases of serum albumin and serum globulin the following relation holds with good accuracy: α=Δn/C=[(np−ns)/C], Here, α is the refractive index increment, C is the number of grams of dry protein per 100 ml, and np, ns are the refractive indices of the protein and the solvent, respectively. Note that this relation holds in the 3D settings where these quantities are functions of the 3D coordinate r. The quantity C is the mass density of dry protein, calculated as (dm/dV)(r), where m is the total dry mass of the cell, and V is its volume. Therefore,
where the variation of α(r) is relatively small α≈0.18-0.21 ml/g, with a common value of
where φ(r⊥)=βoΔn(r⊥)h(r⊥), and S the effective projected area of the cell. Unfortunately, this assumption is not applicable to optically thick sample.
The total dry mass may be derived directly, as now shown, given that V is accurately provided by GLIM. First, the foregoing mass equation may be cast in terms of the susceptibility using the following approximation, Δn(r)≈[(np2(r)−ns2)/(2ns)]=[χ(r)/(2ns)]. This yields:
In the foregoing expression, the refractive index of the surrounding media ns can be approximated as 1.33. To obtain the average susceptibility
Time-Lapse Measuring of Mass, Surface Area, Volume and Sphericity of Hela Cells
Combining various fields of view, several metrics may be obtained, in accordance with the present invention, over cells during their development cycles.
Three Dimensional Image Formation in GLIM
As shown above, small values of NAcon give a precise value of the phase gradient. The present discussion now focuses on the other regime, that of large NAcon, where the depth sectioning is best, thanks to the maximum angular coverage. Under the 1st order Born approximation, two sample fields coming to the camera plane can be written as
U1(r)=Ui(r)−βo2{[Uiχ]{circle around (v)}rg}(r),
U2(r)=Ui(r)−βo2{[Uiχ(r+δr)]{circle around (v)}rg}(r).
Here, g(r) is the propagation kernel, defined by the microscope's objective and given by g(r)≈iFk
where Γi is the mutual intensity function of the illumination, which only depends on the coordinate difference under the statistically homogeneous assumption, namely, Γi(r1,r2;0)=Γi(r1−r2)=∫Sc(k⊥)ei[k
γ(r,δr)=Γi(0;0)−βo2{χ{circle around (v)}r[Γig*)(r)+(Γi*g)(r−{circumflex over (x)}δx)]}.
Assuming a non-absorbing object with a real χ, with a uniform unit-amplitude intensity distribution at the condenser aperture, with g(r)≈iFk
∇xϕ(r)=arg[γ(r,δr)]/δx=βo2 arctan {χ{circle around (v)}rIm[(μig*)(r)−(μig*)(r−{circumflex over (x)}δx)]}/δx≈βo2(χ{circle around (v)}rIm[(μig*)(r)−(μig*)(r−{circumflex over (x)}δx)])/δx,
where μi(r)=Γi(r;0)/Γi(0;0), the complex degree of spatial coherence of the incident field.
Since the PSF is odd, the TF is purely imaginary. The absolute value of the TF vanishes at kx=0, since a constant signal in the x-direction is filtered out by the gradient operator. Therefore, there is a missing area around kx=0 not covered by the TF. This region of missing frequencies is similar to the “missing cone” problem known in diffraction tomography. Resolving this area requires rotating the sample or using additional priors, e.g., smoothness constraint. Larger NAcon reduces the size of the missing cone, allowing the system to record more frequency components. The larger the NAcon, the more transverse frequencies are captured by the TF. More importantly, around NAcon/2 more kz frequency bandwidth is captured with NAcon=0.55 compared to those with NAcon=0.09, which essentially means that depth sectioning improves when NAcon increases. The axial elongation of the PSF decreases for larger NAcon The improvement in depth sectioning for increasing NAcon also reduces the ringing effects since less non-specific information from one z-plane is propagated into neighboring planes. Any diffraction ringing phenomenon is suppressed when NAcon increases.
Filtering Method to Improve Depth Sectioning of GLIM
In order to improve the optical sectioning, low-frequency components are removed from the data using a high-pass filter. Steps of our methods are summarized in
where ε is a small number, set to be 10−4 to avoid amplifying frequency components with small signal-to-noise ratios (SNRs). To further improve the axial resolution, it is necessary to significantly suppress the low-frequency components in χWeiner(r). This may be achieved by applying high-pass filtering in the x-y domain for each recorded z-image. In each dimension (x and y), a convolution with a finite-length impulse response (FIR), chosen as hhp(x)=[0.25, −0.25, 0, −0.25, 0.25], is applied. As a result of this high pass filtering, χhp(r) has most of the small transverse frequencies suppressed and, as a result, yields very good depth sectioning. Note that this high-pass filtering step can be combined with the Wiener deconvolution step since both are linear operators. Also, there is no need to perform any z-processing in our proposed method. This allows the processing to be done effectively by interlacing with image acquisition.
After filtering, a log [abs(.)] transform may be applied to increase the contrast of the retained high-frequency components in the output image. To further suppress the background noise, only signals with amplitude within [−5.0, 0.0] are retained. Finally, to smooth the image and remove high-frequency oscillations in the image due to missing small transverse frequencies, bilateral filtering may be applied to the transformed results, as described by Tomasi et al., “Bilateral Filtering for Gray and Color Images,” in Sixth International Conference on Computer Vision (IEEE), pp. 839-846 (1998), which is incorporated herein by reference.
The embodiments of the invention described above are intended to be merely exemplary; numerous variations and modifications will be apparent to those skilled in the art. All such variations and modifications are intended to be within the scope of the present invention as defined in any appended claims.
The present application claims the priority of U.S. Provisional Patent Application 62/425,268, filed Nov. 22, 2016, and incorporated herein by reference.
This invention was made with government support under CBET-0939511 STC and IIP-1353368, awarded by the National Science Foundation. The Government has certain rights in the invention.
Number | Name | Date | Kind |
---|---|---|---|
7990543 | Mello | Aug 2011 | B1 |
20090276188 | Cui | Nov 2009 | A1 |
20160259158 | Girshovitz | Sep 2016 | A1 |
Entry |
---|
Gogswell et al., “Quantitative DIC microscopy using a geometric phase shifter”, Proc. SPIE 2984, Three-Dimensional Microscopy: Image Acquisition and Processing IV, pp. 72-82, (1997). |
King et al., “Quantitative phase microscopy through differential interference imaging,” Journal of Biomedical Optics, vol. 13(2), 024020, 31 pages (2008). |
Edwards et al., “Effects of spatial coherence in diffraction phase microscopy,” Opt. Express, vol. 22, pp. 5133-5146, (2014). |
Number | Date | Country | |
---|---|---|---|
20180143001 A1 | May 2018 | US |
Number | Date | Country | |
---|---|---|---|
62425268 | Nov 2016 | US |