The present teachings relate to methods and apparatus for mass spectrometry, and more specifically, the present teachings relate to methods and apparatus for time-of-flight mass spectrometry.
One application for mass spectrometry is directed to the study of biological samples, where sample molecules are converted into ions, in an ionization step, and then detected by a mass analyzer, in mass separation and detection steps. Various types of ionization techniques are presently known, which typically create ions in a region of nominal atmospheric pressure or within vacuum. Mass analyzers can be quadrupole analyzers where RF/DC ion guides are used for transmitting ions within a narrow slice of mass-to-charge ratio (m/z) values, magnetic sector analyzers where a large magnetic field exerts a force perpendicular to the ion motion to deflect ions according to their m/z and time-of-flight (“TOF”) analyzers where measuring the flight time for each ion allows the determination of its m/z.
Time-of-flight mass spectrometers, TOF MS, are advantageous because they are instruments with virtually unlimited mass-to-charge ratio range and with potentially higher sensitivity than scanning instruments because they can record all the ions generated from each ionization step. Time-of-flight mass spectrometers measure the mass of an ion indirectly by accelerating the ion in a vacuum to a fixed energy and measuring the time of flight over a fixed distance to a detector. Variations of the energy, the distance or the measurement of time, however, may produce errors in measured mass. Some of these variations may result from components of the system with parameters that may vary with changes in temperature.
There are certain techniques that can be applied, for example, to power supplies and clocks for reducing the affects due to the temperature coefficient of various components within the TOF MS. These include temperature compensation, where components with equal and opposite temperature coefficients are employed, and oven control to directly regulate the temperature of sensitive components. Typically, TCXO and OCXO are the common terms used to represent temperature controlled crystal oscillators and oven controlled crystal oscillators respectively. Oven control of voltage references has been used in voltage calibrators and in integrated circuit voltage references. Close matching of temperature coefficients of the high voltage resistors used in feedback dividers has been used in high voltage power supplies to compensate for a significant source of drift. Oven control of critical components in high voltage power supplies can be an extension of these techniques. After these techniques are applied, the remaining challenge is to take into account the drift caused by thermal expansion effects on the flight distance of the ions from source to detector.
It has been proposed to control the temperature of the materials in the TOF MS that undergo expansion and that have an impact on the time-of-flight measurement, but this can be costly and ineffective due to thermal time constants of the affected materials.
Another technique for dealing with the problems of mass drift errors is to run periodic calibrations using known mass standards to tune the TOF MS and eliminate these errors. Yet running calibrations too frequently when such calibration may be unnecessary results in downtime from sample analysis which has an undesirable effect for high throughput analyses. Improvements to reduce the drift rate will in turn reduce the frequency of calibrations required to maintain a given error limit.
In view of the foregoing, the present teachings provide improved methods and apparatus for conducting time-of-flight mass spectrometry. In various embodiments, the method comprises establishing a reference flight distance for a pulse of ions corresponding to a reference temperature of one or more components of an ion flight path assembly; obtaining a temperature measurement of the one or more components of the ion flight path assembly; correlating a thermal expansion of the flight path assembly with the temperature measurement; compensating for the thermal expansion of the flight path assembly by adjusting the flight times of the ions to correspond with the reference flight distance; and obtaining a mass spectrum using the adjusted flight times. In various embodiments, a mass spectrometer comprises a flight path assembly comprising one or more ion optic components for providing a transmission path configured for obtaining the time-of-flight for ions to be analyzed, a temperature sensor mounted on one or more components of the assembly for obtaining a temperature measurement of the flight path assembly, a power supply system connected to one or more of the ion optic components, the power supply system being adjustable in response to signal that is a function of the temperature measurement for providing one or more adjusted flight time parameters, and wherein the mass spectrum is obtained using one or more of the adjusted flight-time parameters to compensate for the thermal expansion of the flight path assembly. In various embodiments, the temperature signal is used with pre-determined thermal expansion correction factors for the flight path assembly to calculate a correction factor to control another component of the TOF MS, such as a power supply system. The power supply system can be controlled by a system controller where the controller applies the correction factor to adjust the voltage to one or more ion optic components within the flight path assembly to compensate for the ion flight times.
In various embodiments, variation of other parameters that influence the time of flight measurement can be used to control a different component of the analyzer to compensate for errors in ion flight times. For example, control of clock frequencies can be used to correct for errors due to thermal expansion of the ion flight path assembly.
These and other features of the present teachings are set forth herein.
The skilled person in the art will understand that the drawings, described below, are for illustration purposes only. The drawings are not intended to limit the scope of the present teachings in any way.
In the accompanying drawings:
In the drawings, like reference numerals indicate like parts.
It should be understood that the phrase “a” or “an” used in conjunction with the present teachings with reference to various elements encompasses “one or more” or “at least one” unless the context clearly indicates otherwise. Reference is first made to
In various embodiments, in a time of flight mass spectrometer, ions can be produced in the ion source 21 and a pulse of ions 30 can be accelerated through an electric field presented by the accelerator 26 through the application of an electrostatic potential between the sample support 25 and a second electrode 27. The pulse of ions 30 fly a fixed distance, commonly referred as the flight distance, to the detector 24 and the detector produces corresponding signals at the times that the ions arrive. In various embodiments, the flight distance can be the distance defined by the path from the sample support 25 to the electrostatic mirror 28 and from the electrostatic mirror 28 to the detector 24, such as in a reflector TOF MS. In various embodiments, the flight distance can be the distance defined by the path from the sample support 25 to the detector 52 with no voltage applied to the mirror 28, such as in a linear TOF MS. It will also be apparent that the detector can be positioned at intermediate locations along the path. The detector signal can be sampled using a fixed frequency clock starting at or near the time when the pulse of ions 30 is accelerated by ion accelerator 26. Time can be measured by the count of clock ticks divided by the clock frequency. This clock tick count, interpolated to a fraction of a tick, represents the measured flight time. The measured flight time in clock ticks is proportional to the flight time in seconds assuming a fixed and stable clock frequency. Centroids of signal pulses can be computed producing time measurements to a resolution that is finer than the clock period. The energy given to the ions can be determined by the power supplies 36, 38 and the flight distance can be determined by the assembly of mechanical and optical components that comprise the ion flight path assembly, which can be an assembly of various materials having different physical properties. Each of the above mentioned parameters contribute to the final determination of the mass of each ion in the pulse of ions 30.
A basic equation relating the parameters of energy, time and distance is the equation for the kinetic energy of a moving mass. The heavier ions fly slower than the lighter ones so they arrive later. The mass of the ion is then calculated from the measured time. From basic Newtonian physics, the energy (E) of a moving object is related to its mass (m) and velocity (v) by:
In equation (1), l is the length of the flight path or simply the flight distance and t is the time. The energy of a charged ion accelerated through an electric field is equal to the voltage (V) times the number of charges (z), so:
Equation (3) represents a basic equation for TOF MS. Solving this equation for mass (actually m/z) gives:
As more ion optic elements are added for added functionality and performance improvements, the equation of motion for the ions can become more complex. For example, there can be a delay time or time offset t0, from the measured start of the flight to the actual start due to signal propagation delays inherent in cables and in the electronic components. This delay time must be subtracted from the measured flight time to get the actual flight time. The accelerating voltage and the length of the flight path are held constant so the form of equation (4) becomes:
From equation (4), we see that the measured mass is a function of voltage, of time and of flight distance. It can be useful to know how a change or a drift of any one of these three parameters will affect the mass accuracy of the TOF MS. As used herein, drift refers to a mass error that is changing over time. To show this, the first partial derivative of equation (4) can be taken with respect to each of the three parameters, while holding the others constant. For simplicity, m/z can be replaced by m:
In practice, drift can be expressed in generic terms by ‘parts per million’ or ppm. This is one million times the difference in a parameter divided by the value of the parameter. For example, the ppm of mass drift, for a mass difference Δm, can be expressed as:
So, rewriting equations (11), (12) and (13) in the form of dm/m according to equation (7) and then in to the ‘parts per million’ format (the same as Δm/m) gives three basic sensitivity equations, differentially coupled to the mass drift (Δm/m):
In practice, the differential coupling coefficients found in equations (15), (16) and (17), expressed as constants 1, 2 and −2 respectively, can vary. Techniques for time focusing of ions of the same mass but different energies such as through use of delayed ion extraction and ion mirrors can reduce the coupling coefficients expected from the above equations that describe simpler TOF MS systems. Ion optics components such as Einzel lenses used for spatial focusing and deflectors for ion beam steering have small coupling coefficients for their applied voltages because they influence the ions over a short distance and do not change the net energy of the ions. For example, the coupling coefficient for the voltage on an ion mirror electrode in a time of flight mass spectrometer system such as the 4800 MALDI TOF/TOF™ Analyzer (Applied Biosystems/MDS Sciex) may be measured empirically to be 0.732 rather than 1.000 as might be expected from equation (15) above. These coupling coefficients determined by empirical measurements can be used for mass calibration or compensation purposes as will be described subsequently.
By way of example, in the time of flight equations for simplified linear ion optics geometry a +20 ppm drift in acceleration voltage (ΔV/V) produces a +20 ppm drift in measured mass (Δm/m) according to equation (15), a +20 ppm drift in the frequency of the clock (Δt/t) measuring the flight time produces a +40 ppm drift in mass (Δm/m) according to equation (16), and a +20 ppm change in the length (Δl/l) of the flight path produces a −40 ppm change in mass (Δm/m) according to equation (17). In the example of an empirically measured coupling coefficient for an ion mirror of 0.732, a +20 ppm drift of the voltage applied would produce a +14.64 ppm drift in mass. An estimate of mass drift (Δm/m) based on measurements of other parameters can be compared to a predetermined mass error limit imposed on the TOF MS.
Despite efforts to reduce mass drift and associated mass accuracy errors using the foregoing techniques, a residual drift error that exceeds mass error limits required for certain applications can remain due to the complex nature of the components that interconnect to form the TOF MS system. In various embodiments, a mass calibration step can be performed by using one or more mass standards containing ions of known mass to essentially eliminate, for subsequent analyses, the effect of mass drift. For brevity, the terms calibration mass and known mass can mean the same. A mass spectrum, obtained with the TOF MS of the mass standard, can be correlated with the m/z values of the calibration mass. Subsequently, the correlation between the measured and the known mass can be used to compute calibration factors to arrive at a mass spectrum (peak intensity versus m/z value) from a time-of-flight spectrum (peak intensity versus time) to thus align the measured mass with the calibration m/z values. The calibration factors, in addition to the other parameters previously discussed, can be incorporated into the time-of-flight equations (4) and (5) so that a general form of the equations becomes:
m=f(a0,a1,a3, . . . an,t) (18)
Where mass m can be a function of time t and of the parameters a0, . . . an, the parameters can be substantially constant but can be a function of temperature as described above.
Equation (18) can be expressed in the form of:
√{square root over (m)}=a0+a1t+a2t2+ . . . antn (19)
Where, for brevity, the generic parameter notations, a0, . . . an, have been used in both equations (18) and (19). These parameters a0, . . . an, can be general and do not necessarily imply that they are the same in each of equations (18) and (19). As will be appreciated by those of skill in the art, the calibration model, equation (19), is still a generalized form and the polynomial powers are not limited to positive integers. Fractional and negative powers can be used as well as other functions of t. In various embodiments, a calibration step can include providing a measurement of a time-of-flight spectrum from a mixture of known mass standards and from the calculation of a best fit of the parameters, an, according to equation (19). The values of the parameters, an, can be calculated by applying the mathematical method of least squares as known in the art. This method minimizes the sum of the squares of the residual errors for all of the calibration masses. Frequent calibration with standards, while limiting the effects of thermal drift in the mass spectrometer, takes time away from the analysis of samples. Reduction of thermal drift allows greater time intervals between calibrations for a given maximum error limit.
There are other aspects of the TOF MS instrument which can have an affect on drift, but in the present teachings, the model as exemplified by equations (15), (16), and (17) can be sufficient to describe the major contributors to mass drift. All three parameters, voltage, time and distance, have temperature coefficients and thermal time constants which can contribute to the drift characteristics of the instrument. For example, the power supply electronic components, which provide voltages for ion optics, and the time measurement clock, which provides the timing, can each have temperature coefficient properties affecting the corresponding voltage and time values. The power supply components and the clock, additionally, can each have a thermal time constant, which can attribute to delayed response to any temperature variations. The components of the ion flight path assembly can have thermal expansion coefficients, which can result in altering the flight distance as a response to any temperature variation.
Returning to
In various embodiments, the present teachings compensate for thermal drift of the measured masses by compensating for the thermal expansion of materials prior to the acquisition of the mass spectrum. Accordingly, when the temperature of the flight path assembly deviates from, for example, an arbitrary reference temperature corresponding to the reference flight distance, the length of the flight path assembly can change according to the linear thermal expansion coefficient of the materials that comprise the flight path assembly. Subsequently, a correction factor corresponding to the changed length of the flight path assembly at the measured temperature can be applied to correct the mass error. Prior to the acquisition of the mass spectrum, the correction factor can be used to adjust, for example, a power supply voltage or clock frequency which in turn adjusts the measured flight times of the ions flying in the changed length so that the mass of the ions obtained with the reference flight distance and the corresponding adjusted flight times can be compensated for the thermal expansion prior to acquisition of the mass spectrum.
To demonstrate various embodiments, reference is now made to
By way of example, the system controller 44 can store a table of values, such as a table of Δl, that correspond to the expansion of the flight path at various temperatures or equations governing the thermal expansion of the different materials used for the construction of the flight path assembly. For example, the mean thermal coefficient of linear expansion of type 304 stainless steel commonly used for components of the flight path assembly is equal to 17 μm/m/° C., and the table of Δl values, can be derived from this coefficient. The system controller can reference the various temperatures in the table with the actual temperature measurement and make any necessary interpolations to establish the expansion of the flight path assembly that would result from an increased temperature and then to compensate for the expansion by providing a corresponding adjustment to an applied voltage or a clock frequency prior to obtaining the mass spectrum.
In various embodiments, a change in the length of the flight path assembly Δl as a result of the thermal expansion can be compensated by a change in the voltage ΔV according to equations (15) and (17) or their empirical equivalents:
According to equation (20), the system controller 44 can compensate for the thermal expansion of the assembly by providing an appropriate signal to a power supply system 46 that comprises a power supply control 48 and the power supplies 46,48 thereby changing the voltage applied to one or more of the ion optic components within the TOF MS. In effect, the measured flight times of the ions can be altered or adjusted by the voltage applied to one or more of the ion optic components, such as the ion accelerator 26 or the electrostatic mirror 28.
In various embodiments, a change in the length of the flight path assembly Δl as a result of the thermal expansion can be compensated by a change in the clock frequency Δf(=l/Δt) according to equations (16) and (17) or their empirical equivalents:
According to equation (22), the system controller 44 can compensate for the thermal expansion of the assembly by providing an appropriate signal that changes the frequency of the clock used to measure the time of flight. In effect, the measured flight times of the ions can be altered by the frequency of this clock.
While the present teachings are described in conjunction with various embodiments, it is not intended that the present teachings be limited to such embodiments. On the contrary, the present teachings encompass various alternatives, modifications, and equivalents, as will be appreciated by those of skill in the art. For example, the function of the signal conditioner 42 can be incorporated within the temperature sensor 40 so that the temperature sensor 40 can be adapted to directly produce a temperature signal proportional to the temperature of the assembly. This is shown graphically in
Furthermore, as shown in
Number | Name | Date | Kind |
---|---|---|---|
6049077 | Franzen | Apr 2000 | A |
6465777 | Rache | Oct 2002 | B1 |
6700118 | Yefchak et al. | Mar 2004 | B2 |
7183543 | Overney et al. | Feb 2007 | B1 |
20030034448 | Yefchak et al. | Feb 2003 | A1 |
20040031918 | Schoen et al. | Feb 2004 | A1 |
20050023454 | Bateman et al. | Feb 2005 | A1 |
20060043283 | Davis et al. | Mar 2006 | A1 |
Number | Date | Country | |
---|---|---|---|
20080087810 A1 | Apr 2008 | US |