The use or focused ion-beam (FIB) microscopes has become common for the preparation of specimens for later analysis in a transmission electron microscope (TEM) or scanning transmission electron microscope (STEM). The in-situ lift-out technique has become the method of choice for the preparation of a tiny specimen for TEM inspection. TEM and STEM inspection offer fine image resolution (<0.1 nm), but require election-transparent (<100 nm thick) sections of the bulk sample. TEM and STEM inspection usually takes place in a separate TEM or STEM device, which requires the transfer of the fragile TEM specimen to another location.
Dual-beam (DB-FIB) instruments having both an ion beam and an electron beam are being more widely used for TEM specimen preparation and inspection. The DB-FIB instrument combines the high-resolution imaging of the SEM with precision, site-specific ion milling of the FIB. The combination of SEM and FIB in the same chamber allows for the location, preparation, and inspection of specimens in the same microscope.
After a specimen is excised from a larger sample, it is preferably moved away from the larger sample by a nano-manipulator system and attached to a TEM specimen grid fur further investigation. A suitable nano-manipulator system is the Omniprobe AutoProbe™, manufactured by Omniprobe, Inc. of Dallas, Tex. Such a nano-manipulator system will typically have a probe part mat is inserted into the vacuum chamber of the FIB instrument.
Existing holders for TEM grids either provide only one orientation of the TEM grid relative to the electron and ion beams, or else allow a limited range of orientations with respect thereto. It is desirable to vary the orientation of an excised TEM specimen to allow better thinning or viewing. It may be advantageous, for example, to perform backside milling of the specimen or shape a MEMS structure in the specimen. Also, existing systems typically rely on tilting the stage of the DB-FIB to vary the angle of the TEM grid holder with respect to the charged-particle beams. Such stages have a limited range of tilt, usually no more than the angle between the electron beam and the ion beam in the instrument. Often, a TEM specimen will require adjustment to an angle with respect to the DB-FIB horizontal that is outside the range of movement of the instrument's tilt stage. This situation can occur, for example, when a probe tip bearing a specimen is attached to a TEM grid, and the attachment operation causes an undesirable rotation of one or more axis of the specimen.
There is a need for a reliable and simple means and method for varying the orientation of the TEM specimen with respect to the charged-particle beams inside the DB-FIB beyond the range available by the tilt stage of the instrument itself, while also securely holding the TEM grid in place during all changes in orientation and maintaining the vacuum inside the instrument chamber.
The figures are not necessarily to scale.
The embodiments of the variable-tilt TEM grid holder disclosed and claimed in this application allow variable orientation of TEM specimens inside a DB-FIB. The field of application is not limited to dual-beam FIB systems or to semiconductor samples. Applications could include, for example, nano-mechanical systems or biological samples, or samples examined in other vacuum or charged particle beam instruments.
The apparatus and methods disclosed in this application provide for improved orientation of a TEM specimen relative to both focused ion and electron beams and higher throughput TEM specimen thinning and preparation within the DB-FIB, because the orientation of the specimen can be adjusted beyond the range possible with the typical DB-FIB tilt stage (360), and further because the specimen does not have to be removed from the microscope to adjust the angle of the specimen relative to the energetic beams.
The pivot plate (140) comprises slots (210) for TEM grids (230). The TEM grids (230) are held against the slots (210) by clamps (150) that are normally biased open by springs (170). The clamps (150) allow easy and secure holding and loading of TEM grids (230). As the damp screw (160) is loosened, the clamp (150) is raised by the compression spring (170) and rotates counterclockwise with the loosening of the right-threaded clamp screw (160) until it rests against the stopper pin (200). This leaves clear access to the TEM grid slot (210) for TEM grid (230) loading or unloading.
The hard stop (130) depicted in the figures is desirable to hold the pivoting plate (140) in a fixed position while the slots (210) are loaded with TEM grids (230) and the clamps (150) are tightened against the TEM grids (230). The height of the hard stop (130) thus determines the maximum angle to which the pivoting plate (140) can be inclined in the direction of the hard stop (130). In other embodiments, the variable-tilt TEM grid holder (100) may be constructed without the hard stop (130), thus allowing a range of movement of the pivoting plate (140) of ±90 degrees or more from its vertical.
Although the embodiment of the variable-tilt TEM grid holder (100) just described is shown holding a standard TEM grid (230), it can hold grids or assemblies of other shapes having approximately the same size as the standard TEM grid (230), or the shape of the slots (210) and the clamp (150) could be modified to hold specimens or assemblies of different sizes.
In
Both TEM grids in
An enlarged side view of an embodiment of the variable-tilt TEM grid holder (100) is shown in
In the figures, the slots (210) are shown as situated in the plane of the axes of the pivot shafts (190). This embodiment is desirable, since it is then easier to bring the slot (210) and therefore the TEM specimen (270) into parallel with the axis of the ion beam column (320), as later described. If, however, the central axis of the pivoting plate (140) is disposed in the plane of the pivot shafts (190), the same result can be achieved by additionally adjustments the horizontal (X-Y) motion of the DB-FIB stage.
The variable-tilt TEM grid holder (100) can be assembled and mounted on the sample plate (280) outside the DB-FIB. It can be placed into the DB-FIB chamber pre-loaded with specimen-laden TEM grids (230), or the TEM grids (230) can be loaded into the TEM grid slots (210) without TEM specimens (270) and the specimens (270) can be attached inside the DB-FIB The variable-tilt TEM grid holder (100) can be mounted on the sample plate (280) using mounting screws (300) through holes (120).
The orientation of the pivoting plate (140) can be changed manually using the gripper (250) shown in
The angles shown in
To prepare the variable-tilt TEM grid holder (100) for operation, the pivoting plate (140) is brought to the position where it hits the hard stop (130), if present. The clamping screws (160) are loosened and the spring-raised clamp (150) is rotated to expose the TEM grid slot (210). After the TEM grids (230), holding TEM specimens (270), are loaded into their slots (210), they can be secured by clamps (150) turned into the closed position, and the assembly will be ready for operation. Typically, the foregoing operations will be done outside the DB-FIB. The pivoting plate (140) can be rotated to set it at the desired angle using the gripper (250) or a motor (260).
The basic steps of the operating process of the variable-tilt TEM grid holder (100) are as follows. First, at least one variable-tilt TEM gild holder (100) is mounted on the sample plate (280) as shown in
If the TEM specimen (270) is not substantially vertical, the operator chooses an adjustment angle to correct the orientation of the specimen to be parallel to the electron beam, and rotates the pivoting plate (140) through that angle using the gripper (250) or the motor (260).
If the rotation angle is less than desired, an adjustment can be made by any of the means described earlier. If the rotation angle is larger than desired, the tilt stage (360) can be rotated at 180 degrees about the vertical axis of the DB-FIB, and the pivoting plate (140) adjustment can be made from the opposite side of the pivoting plate (140). The relationship between the distance the gripper (250) is moved along the axis of the probe shaft versus total tilt angle achieved can be established experimentally and can be used to determine this distance. The same angle estimation can be used for rotation of the pivoting plate shaft (190), if a motor or other actuator (260) is used. This relationship can be characterized in a lookup table or in an approximate equation with the angular adjustment being the input variable and the displacement of the gripper (250) being the output variable.
After the user finds the TEM specimen (270) to be substantially parallel to the electron beam, the DB-FIB sample stage (330) can be then inclined to the tilt angle estimated to make the preferred axis of thinning (275) of the TEM specimen (270) substantially parallel to the axis of the ion beam column (320). This tilt will generally be the angle between the electron beam column (310) and the ion-beam column (320). This angle is specific to the microscope manufacturer, but in the case of a DB-FIB manufactured by FEI Company, it would be approximately 52 degrees. The TEM specimen (270) can now be thinned by the ion-beam (350) acting substantially parallel to the face of the TEM specimen (270), along its preferred axis of thinning (275). After thinning, the TEM specimen (270) can be examined using the electron beam (340).
It may be the case that the TEM specimen (270) is not substantially parallel to the desired thinning angle, or that the angle of its preferred thinning axis (275) relative to the ion beam (350) is preferably adjusted to accomplish the desired thinning. In this case, the pivoting plate (140) can be rotated as earlier described to bring the preferred thinning axis (275) of the TEM specimen (270) to the desired angle with respect to the ion beam (350).
After the thinning process has proceeded for a time, an electron-beam (340) scan can again be made, followed by further pivoting plate (140) adjustment, if needed, and additional thinning. After thinning of all the TEM specimens (270) in a variable-tilt holder (100) is completed, the variable-tilt TEM grid holder (100) holding the TEM specimens can be transferred outside the DB-FIB for further analysis by TEM.
None of the description in this application should be read as implying that any particular element step, or function is an essential element, which must be included in the claim scope; the scope of patented subject matter is defined only by the allowed claims. Moreover, none of the claims is intended to invoke paragraph six of 35 USC section 112 unless the exact words “means for” are used, followed by a gerund. The claims as filed are intended to be as comprehensive as possible, and no subject matter is intentionally relinquished, dedicated, or abandoned.