This technology relates to co-manipulation robotic systems, such as those designed to be coupled to clinician-selected surgical instruments to permit movement of the robot arm(s) via movement at the handle of the surgical instrument(s), along with enhanced features for setup and automatic intraoperative movements.
Managing vision and access during a laparoscopic procedure is a challenge. The surgical assistant paradigm is inherently imperfect, as the assistant is being asked to anticipate and see with the surgeon's eyes, without standing where the surgeon stands, and similarly to anticipate and adjust how the surgeon wants the tissue of interest exposed, throughout the procedure. For example, during a laparoscopic procedure, one assistant may be required to hold a retractor device to expose tissue for the surgeon, while another assistant may be required to hold a laparoscope device to provide a field of view of the surgical space within the patient to the surgeon during the procedure, either one of which may be required to hold the respective tools in an impractical position, e.g., from between the arms of the surgeon while the surgeon is actively operating additional surgical instruments.
Various attempts have been made at solving this issue. For example, a rail-mounted orthopedic retractor, which is a purely mechanical device that is mounted to the patient bed/table, may be used to hold a laparoscope device in position during a laparoscopic procedure, and another rail-mounted orthopedic retractor may be used to hold a retractor device in position during the laparoscopic procedure. However, the rail-mounted orthopedic retractor requires extensive manual interaction to unlock, reposition, and lock the tool in position.
Complex robot-assisted systems such as the Da Vinci Surgical System (made available by Intuitive Surgical, Sunnyvale, California) have been used by surgeons to enhance laparoscopic surgical procedures by permitting the surgeon to tele-operatively perform the procedure from a surgeon console remote from the patient console holding the surgical instruments. Such complex robot-assisted systems are very expensive, and have a very large footprint and take up a lot of space in the operating room. Moreover, such robot-assisted systems typically require unique system-specific surgical instruments that are compatible with the system, and thus surgeons may not use standard off-the-shelf surgical instruments that they are used to. As such, the surgeon is required to learn an entirely different way of performing the laparoscopic procedure.
In view of the foregoing drawbacks of previously known systems and methods, there exists a need for a system that provides the surgeon with the ability to seamlessly position and manipulate various surgical instruments as needed, thus avoiding the workflow limitations inherent to both human and mechanical solutions.
The present disclosure overcomes the drawbacks of previously-known systems and methods by providing a co-manipulation surgical system to assist with laparoscopic surgery performed using a surgical instrument having a handle, an operating end, and an elongated shaft therebetween. The co-manipulation surgical system may include a robot arm having a proximal end, a distal end that may be removably coupled to the surgical instrument, a plurality of links, and a plurality of joints between the proximal end and the distal end. The co-manipulation surgical system further may include a controller operatively coupled the robot arm. The controller may be programmed to cause the robot arm to automatically switch between: a passive mode responsive to determining that movement of the robot arm due to movement at the handle of the surgical instrument is less than a predetermined amount for at least a predetermined dwell time period, wherein the controller may be programmed to cause the robot arm to maintain a static position in the passive mode; and a co-manipulation mode responsive to determining that force applied at the robot arm due to force applied at the handle of the surgical instrument exceeds a predetermined threshold, wherein the controller may be programmed to permit the robot arm to be freely moveable in the co-manipulation mode responsive to movement at the handle of the surgical instrument for performing laparoscopic surgery using the surgical instrument, and wherein the controller may be programmed to apply a first impedance to the robot arm in the co-manipulation mode to account for weight of the surgical instrument and the robot arm. The controller further may be programmed to cause the robot arm to automatically switch to a haptic mode responsive to determining that at least a portion of the robot arm is outside a predefined haptic barrier, wherein the controller may be programmed to apply a second impedance to the robot arm in the haptic mode greater than the first impedance, thereby making movement of the robot arm responsive to movement at the handle of the surgical instrument more viscous in the haptic mode than in the co-manipulation mode.
In accordance with one aspect of the present disclosure, a co-manipulation surgical system to assist with laparoscopic surgery performed using a surgical instrument having a handle, an operating end, and an elongated shaft therebetween is provided. The co-manipulation surgical system may include a robot arm comprising a proximal end, a distal end configured to be removably coupled to the surgical instrument, a plurality of links, and a plurality of joints between the proximal end and the distal end, and a controller operatively coupled to the robot arm and configured to permit the robot arm to be freely moveable responsive to movement at the handle of the surgical instrument for performing laparoscopic surgery. The controller programmed to: cause the robot arm to maintain a static position in a passive mode responsive to determining that movement of the robot arm due to movement at the handle of the surgical instrument is less than a predetermined amount for at least a predetermined dwell time period; identify, when the surgical instrument comprises a laparoscope having a field of view, a target surgical instrument within the field of view of the laparoscope based on image data from the laparoscope; and cause the robot arm to switch to an instrument centering mode where the robot arm moves the laparoscope to maintain the target surgical instrument within the field of view of the laparoscope.
The controller may be configured to cause the robot arm to automatically switch to a co-manipulation mode responsive to determining that force applied at the robot arm due to force applied at the handle of the surgical instrument exceeds a predetermined threshold. Accordingly, the controller may be configured to permit the robot arm to be freely moveable in the co-manipulation mode responsive to movement at the handle of the surgical instrument, while applying an impedance to the robot arm in the co-manipulation mode to account for weight of the surgical instrument and the robot arm. In addition, the controller may be configured to identify the target surgical instrument within the field of view of the laparoscope by detecting a predefined gestural pattern by the target surgical instrument within the field of view of the laparoscope. The predefined gestural pattern may comprise positioning of the target surgical instrument within a center portion of the field of view of the laparoscope and maintaining the position of the target surgical instrument within the center portion for at least a predetermined hold period. In some embodiments, the controller may be configured to identify the target surgical instrument within the field of view of the laparoscope based on user input identifying the target surgical instrument. Moreover, the controller may be configured to distinguish the target surgical instrument from one or more other surgical instruments within the field of view of the laparoscope. In the instrument centering mode, the controller may cause the robot arm to move the laparoscope to maintain the target surgical instrument within a predefined boundary region within the field of view of the laparoscope, such that the robot arm does not move the laparoscope unless the target surgical instrument moves outside of the predefined boundary region.
Moreover, in the instrument centering mode, the controller may cause the robot arm to move the laparoscope by executing a trajectory generation algorithm to generate a trajectory from a current position of the laparoscope to a desired position of the laparoscope, and causing the robot arm to move the laparoscope along the trajectory to maintain the target surgical instrument within the field of view of the laparoscope. Accordingly, the controller may be configured to: permit the robot arm to be freely moveable in a co-manipulation mode responsive to determining that force applied at the robot arm due to force applied at the laparoscope exceeds a predetermined threshold, while applying an impedance to the robot arm in the co-manipulation mode to account for weight of the laparoscope and the robot arm; record a trajectory of the freely moving robot arm when the movement of the robot arm deviates from the generated trajectory; and update the trajectory generation algorithm based the recorded trajectory. The generated trajectory may comprise moving the robot arm along a longitudinal axis of the laparoscope to maintain the target surgical instrument within the field of view of the laparoscope and within a predetermined resolution threshold. In addition, the generated trajectory may comprise moving the robot arm along at least one of a longitudinal axis of the laparoscope or an axis perpendicular to the longitudinal axis of the laparoscope to maintain the target surgical instrument within the field of view of the laparoscope.
The trajectory may be generated by: measuring a current position of the distal end of the robot arm; determining a point of entry of the laparoscope into the patient; and calculating a distance required to move the distal end of the robot arm from its current position to a second position that causes a distal end of the laparoscope to move from its current position to the desired position based on the point of entry and a known length between the distal end of the robot arm and the distal end of the laparoscope. The controller may cause the robot arm to move the laparoscope along the trajectory by: calculating a force required to move the distal end of the robot arm the distance from its current position to the second position; and applying torque to the at least some joints of the plurality of joints of the robot arm based on the calculated force to move the distal end of the robot arm the distance from its current position to the second position to thereby move the distal end of the laparoscope from its current position to the desired position. Further, the controller may be configured to: detect an offset angle between a camera head of the laparoscope and the laparoscope; and calibrate the trajectory to correct the offset angle such that movement of the laparoscope along the calibrated trajectory maintains the target surgical instrument within the field of view of the laparoscope. For example, the controller may be configured to detect the offset angle by: causing the robot arm to move along a predetermined trajectory in a known direction in a robot arm coordinate frame; measuring an actual movement of a static object within the field of view of the laparoscope responsive to movement of the robot arm along the predetermined trajectory; and comparing the actual movement of the static object with an expected movement of the static object associated with the predetermined trajectory.
The controller further may be configured to cause the robot arm to switch to the instrument centering mode responsive to user input. In addition, the controller may be configured to: determine a phase of the laparoscopic surgery; estimate the target surgical instrument based on the phase of the laparoscopic surgery; and identify the target surgical instrument within the field of view of the laparoscope based on the estimation and the image data from the laparoscope. Moreover, the controller may be configured to: determine a phase of the laparoscopic surgery; and automatically switch to the instrument centering mode responsive to the phase of the laparoscopic surgery. Accordingly, the controller may be configured to: identify one or more anatomical structures within the field of view of the laparoscope based on image data from the laparoscope; determine the phase of the laparoscope surgery based on the identified one or more anatomical structures; and cause the robot arm, in the instrument centering mode, to move the laparoscope to maintain the identified one or more anatomical structures within the field of view of the laparoscope. Additionally, the controller may be configured to: generate an overlay indicative of the target surgical instrument; and cause the overlay to be displayed over the image data from the laparoscope via a graphical user interface.
The controller may be configured to: cause the robot arm to move the laparoscope in a predetermined trajectory; and compare an actual trajectory of the image data from the laparoscope during movement along the predetermined trajectory with an expected trajectory of the image data associated with the predetermined trajectory to determine an angle of a distal tip of the laparoscope. For example, the predetermined trajectory may comprise a circular pattern in a single plane. Moreover, the controller may be configured to identify the target surgical instrument within the field of view of the laparoscope based on image data from the laparoscope using machine learning algorithms executed at the controller. For example, the machine learning algorithms may be trained with a database of annotated image data of associated surgical instruments. Accordingly, the machine learning algorithms may be configured to evaluate pixels of the image data from the laparoscope and indicate if the pixels correspond to the target surgical instrument to identify the target surgical instrument. The controller may be configured to identify the target surgical instrument within the field of view of the laparoscope in real time. The controller may be configured to cause, in the instrument centering mode, the robot arm to move the laparoscope to track the target surgical instrument that is being manually held by a surgeon. In some embodiments, the system may include a second robot arm configured to be removably coupled to the target surgical instrument that is being manually held by the surgeon.
In accordance with another aspect of the present disclosure, a method for assisting with laparoscopic surgery is provided. The method may include: providing a robot arm comprising a proximal end, a distal end configured to be removably coupled a laparoscope, a plurality of links, and a plurality of joints between the proximal end and the distal end; permitting, via a controller operatively coupled to the robot arm, the robot arm to be freely moveable responsive to movement at the handle of the laparoscope for performing laparoscopic surgery; automatically causing, via the controller, the robot arm to maintain a static position in a passive mode responsive to determining that movement of the robot arm due to movement at the handle of the laparoscope is less than a predetermined amount for at least a predetermined dwell time period; identifying, via the controller, a target surgical instrument within a field of view of the laparoscope based on image data from the laparoscope; switching, via the controller, the robot arm to an instrument centering mode; and automatically causing, via the controller while in the instrument centering mode, the robot arm to move the laparoscope to maintain the target surgical instrument within the field of view of the laparoscope. For example, identifying the target surgical instrument within the field of view of the laparoscope may comprise detecting, via the controller, a predefined gestural pattern by the target surgical instrument within the field of view of the laparoscope, the predefined gestural pattern comprising positioning of the target surgical instrument within a center portion of the field of view of the laparoscope and maintaining the position of the target surgical instrument within the center portion for at least a predetermined hold period.
In accordance with another aspect of the present disclosure, a co-manipulation surgical system to assist with laparoscopic surgery performed using a surgical instrument is provided. The co-manipulation surgical system may include a robot arm comprising a plurality of links, a plurality of joints, a proximal end operatively coupled to a base of the robot arm, and a distal region having a distal end configured to be removably coupled to the surgical instrument, and a platform coupled to the base of the robot arm. The platform may comprise a stage assembly configured to independently move the base of the robot arm in at least two degrees of freedom relative to the platform. Accordingly, in a user guided setup mode, application of a force at the distal region of the robot arm in a first direction may cause the stage assembly to move the base of the robot arm in a first degree of freedom of the at least two degrees of freedom relative to the platform.
For example, in the user guided setup mode, the stage assembly may be configured to move the base of the robot arm in the first degree of freedom when the force applied at the distal region of the robot arm in the first direction exceeds a predetermined force threshold. Further, in the user guided setup mode, the stage assembly may be configured to stop moving the base of the robot arm in the first degree of freedom when the force applied at the distal region of the robot arm in the first direction falls below a predetermined release threshold. Moreover, in the user guided setup mode, the stage assembly may be configured to stop moving the base of the robot arm in the first degree of freedom upon application of a counter force at the robot arm in a second direction opposite to the first direction. In addition, in the user guided setup mode, application of a force at the distal region of the robot arm in a second direction may cause the stage assembly to move the base of the robot arm in a second degree of freedom of the at least two degrees of freedom relative to the platform. The system further may include an actuator configured to be actuated to switch the system to the user guided setup mode. In some embodiments, the system remains in the user guided setup mode only while the actuator is actuated. The actuator may be disposed on a collar rotatably coupled to a link of the plurality of links, such that actuation of the actuator permits rotation of the collar in a first direction to cause rotation of a distal link of the plurality of links adjacent to a setup joint of the plurality of joints in a corresponding first direction relative to a proximal link of the plurality of links adjacent to the setup joint, and permits rotation of the collar in a second direction to cause rotation of the distal link adjacent to the setup joint in a corresponding second direction relative to the proximal link adjacent to the setup joint.
The system further may include a graphical user interface operatively coupled to the stage assembly. The graphical user interface may be configured to display an actuator configured to be actuated to cause the stage assembly to move the base of the robot arm in at least one of the at least two degrees of freedom relative to the platform. For example, the actuator may comprise a slidable cursor configured to be moved relative to a neutral center point of a cursor pad, such that movement of the slidable cursor in a direction relative to the neutral center point within the cursor pad may cause the stage assembly to move the base of the robot arm in a corresponding direction relative to the platform. The stage assembly may be configured to move the base of the robot arm in the corresponding direction relative to the platform at a velocity that correlates with a distance of the slidable cursor from the neutral center point. In addition, the graphical user interface may be configured to display one or more indicators, the one or more indicators indicative of a configuration of the robot arm relative to the platform in real-time responsive to actuation of the actuator. Moreover, in a co-manipulation mode, the robot arm may be permitted to be freely moveable responsive to movement at the handle of the surgical instrument for performing laparoscopic surgery.
The system further may include a plurality of motors disposed within the base, the plurality of motors operatively coupled to at least some joints of the plurality of joints, and a controller operatively coupled to the plurality of motors. The controller may be programmed to: measure current of the plurality of motors, the measured current indicative of force applied at the distal region of the robot arm; and cause, in the user guided setup mode, the stage assembly to move the base of the robot arm in at least one of the at least two degrees of freedom based on the measured current. The controller further may be operatively coupled to a setup joint of the plurality of joints of the robot arm, such that the controller may be programmed to: determine if one or more objects are within a predetermined proximity threshold of the robot arm; and automatically rotate a distal link of the plurality of links adjacent to the setup joint relative to a proximal link of the plurality of links adjacent to the setup joint to avoid a collision with the one or more objects as the stage assembly moves the base of the robot arm in at least one of the at least two degrees of freedom relative to the platform in the user guided setup mode.
The system further may include one or more depth sensors configured to detect the one or more objects adjacent to the robot arm, and generate one or more signals indicative of a proximity of the one or more objects to the robot arm. Accordingly, the controller may be configured to determine if the one or more objects are within the predetermined proximity threshold of the robot arm based on the one or more signals. For example, the one or more depth sensors may comprise one or more proximity sensors disposed within the base of the robot arm, the one or more proximity sensors comprising at least one of electromagnetic, capacitive, ultrasonic, or infrared proximity sensors. Additionally, or alternatively, the one or more depth sensors may comprise one or more depth cameras. Accordingly, the controller may be configured to stop movement of the base of the robot arm via the stage assembly if the one or more objects are within the predetermined proximity threshold. The co-manipulation surgical system may not be teleoperated via user input received at a remote surgeon console.
In accordance with another aspect of the present disclosure, a method for assisting with laparoscopic surgery using a robot arm comprising a plurality of links, and a plurality of joints, a proximal end operatively coupled to a base of the robot arm, and a distal region having a distal end configured to be removably coupled to a surgical instrument is provided. The method may include: switching, via a controller operatively coupled to a stage assembly operatively coupled to the base of the robot arm, the system to a user guided setup mode; and causing, via the controller in the user guided setup mode, the stage assembly to move the base of the robot arm in a first degree of freedom of at least two degrees of freedom relative to a platform coupled to the stage assembly upon application of a force at the distal region of the robot arm in a first direction. For example, causing the stage assembly to move the base of the robot arm in the first degree of freedom may comprise causing, via the controller in the user guided setup mode, the stage assembly to move the base of the robot arm in the first degree of freedom when the force applied at the distal region of the robot arm in the first direction exceeds a predetermined force threshold.
The method further may include causing, via the controller in the user guided setup mode, the stage assembly to stop moving the base of the robot arm in the first degree of freedom when the force applied at the distal region of the robot arm in the first direction falls below a predetermined release threshold. In addition, the method may include causing, via the controller in the user guided setup mode, the stage assembly to stop moving the base of the robot arm in the first degree of freedom upon application of a counter force at the robot arm in a second direction opposite to the first direction. Further, the method may include causing, via the controller in the user guided setup mode, the stage assembly to move the base of the robot arm in a second degree of freedom of the at least two degrees of freedom relative to the platform upon application of a force at the distal region of the robot arm in a second direction. Moreover, switching the system to the user guided setup mode may comprise switching the system to the user guided setup mode responsive to actuation of an actuator operatively coupled to the controller, such that the system may remain in the user guided setup mode only while the actuator is actuated.
The method further may include causing, via the controller in the user guided setup mode, the stage assembly to move the base of the robot arm in at least one of the at least two degrees of freedom relative to the platform responsive to actuation of an actuator displayed on a graphical user interface operatively coupled to the controller. Accordingly, the method further may include causing, via the controller in the user guided setup mode, the graphical user interface to display one or more indicators indicative of a configuration of the robot arm relative to the platform in real-time responsive to actuation of the actuator. The method further may include determining, via the controller in the user guided setup mode, if one or more objects are within a predetermined proximity threshold of the robot arm; and stopping, via the controller if the one or more objects are within the predetermined proximity threshold, movement of the base of the robot arm via the stage assembly to avoid a collision with the one or more objects as the stage assembly moves the base of the robot arm in at least one of the at least two degrees of freedom relative to the platform. Moreover, the method may include switching, via the controller, the system to a co-manipulation mode; and permitting, via the controller in the co-manipulation mode, the robot arm to be freely moveable responsive to movement at a handle of the surgical instrument for performing laparoscopic surgery.
In accordance with another aspect of the present disclosure, a co-manipulation surgical system to assist with laparoscopic surgery performed using a surgical instrument having a handle, an operating end, and an elongated shaft therebetween is provided. The co-manipulation surgical system may include a robot arm comprising a plurality of links, a plurality of joints comprising one or more motorized joints, a setup joint, and one or more passive joints, a proximal end operatively coupled to a base of the robot arm, and a distal region having a distal end configured to be removably coupled to the surgical instrument, and a plurality of motors operatively coupled to the one or more motorized joints and to the setup joint. In addition, the system may include an actuator operatively coupled to the setup joint and configured to be actuated to cause rotation of a distal link of the plurality of links adjacent to the setup joint relative to a proximal link of the plurality of links adjacent to the setup joint from a first setup configuration to a second setup configuration responsive to actuation of the actuator. Accordingly, when the actuator is in an unactuated state, the robot arm may be permitted to be freely moveable responsive to movement at the handle of the surgical instrument for performing laparoscopic surgery via the one or more motorized joints and the one or more passive joints while the distal link adjacent to the setup joint and the proximal link adjacent to the setup joint remain in the second setup configuration.
The actuator may comprise a collar rotatably coupled to a link of the plurality of links, the collar configured to be rotated in a first direction relative to the link of the plurality of links to cause rotation of the distal link adjacent to the setup joint in a corresponding first direction relative to the proximal link adjacent to the setup joint, and rotated in a second direction relative to the link of the plurality of links to cause rotation of the distal link adjacent to the setup joint in a corresponding second direction relative to the proximal link adjacent to the setup joint. Moreover, the collar may comprise a setup mode actuator, the setup mode actuator configured to be actuated to permit the rotation of the distal link adjacent to the setup joint in the corresponding first and second directions relative to the proximal link adjacent to the setup joint responsive to rotation of the collar. The collar may be spring-enforced such that upon release of the collar in any position, the collar is configured to return to a neutral position relative to the link of the plurality of links.
The system further may include a graphical user interface operatively coupled to the setup joint, such that the actuator may be configured to be displayed on the graphical user interface. For example, the actuator may comprise a slidable cursor configured to be moved relative to a neutral center point, such that movement of the slidable cursor in a first direction relative to the neutral center point causes rotation of the distal link adjacent to the setup joint in a first direction relative to the proximal link adjacent to the setup joint, and movement of the slidable cursor in a second direction relative to the neutral center point causes rotation of the distal link adjacent to the setup joint in a second direction relative to the proximal link adjacent to the setup joint. In some embodiments, the distal link adjacent to the setup joint may be configured to rotate in the corresponding direction relative to the proximal link adjacent to the setup joint a velocity that correlates with a distance of the slidable cursor from the neutral center point. In addition, the graphical user interface may be configured to display an indicator, the indicator indicative of a configuration of the distal link adjacent to the setup joint relative to the proximal link adjacent to the setup joint in real-time responsive to actuation of the actuator. Additionally, the graphical user interface may be configured to display graphical representations of a plurality of configurations of the distal link adjacent to the setup joint relative to the proximal link adjacent to the setup joint, such that a position of the indicator relative to the graphical representations of the plurality of configurations may be indicative of the configuration of the distal link adjacent to the setup joint relative to the proximal link adjacent to the setup joint in real-time responsive to actuation of the actuator.
The system further may include a controller operatively coupled to the robot arm, the controller programmed to cause the robot arm to be freely moveably responsive to movement at the handle of the surgical instrument for performing laparoscopic surgery during an operating stage. The controller may be configured to switch from the operating stage to a setup stage upon actuation of a setup mode actuator, such that actuation of the actuator only causes rotation of the distal link adjacent to the setup joint relative to the proximal link adjacent to the setup joint when the setup mode actuator is in an actuated state. When the actuator is in an actuated state, application of a force at the distal region of the robot arm in a first direction may cause rotation of the distal link adjacent to the setup joint in a first direction relative to the proximal link adjacent to the setup joint, and application of a force at the distal region of the robot arm in a second direction causes rotation of the distal link adjacent to the setup joint in a second direction relative to the proximal link adjacent to the setup joint. Moreover, when the actuator is in the unactuated state, the setup joint may be configured to cause the distal and proximal links adjacent to the setup joint to be fixed relative to each other in the second setup configuration. In addition, all motors of the plurality of motors operatively coupled to the one or more motorized joints may be disposed within the base of the robot arm. Moreover, a shoulder link of the plurality of links may comprise a distal shoulder link rotatably coupled to a proximal shoulder link via the setup joint, and the motor of the plurality of motors operatively coupled to the setup joint may not back-drivable. For example, the motor of the plurality of motors operatively coupled to the setup joint may be disposed on the shoulder link adjacent to the setup joint.
The system further may include a platform operatively coupled to the base of the robot arm, the platform comprising a stage assembly configured to independently move the base of the robot arm in a horizontal direction and in a vertical direction relative to the platform. Accordingly, in a user guided setup mode, application of a force at the distal region of the robot arm in a first direction may cause the stage assembly to move the base of the robot arm in the horizontal direction relative to the platform, and application of a force at the distal region of the robot arm in a second direction may cause the stage assembly to move the base of the robot arm in the vertical direction relative to the platform. The system further may include a setup mode actuator configured to be actuated to switch the system to the user guided setup mode, such that the system may remain in the user guided setup mode only while the setup mode actuator is actuated. In some embodiments, the actuator may comprise a collar rotatably coupled to a link of the plurality of links, such that the setup mode actuator may be disposed on the collar. Accordingly, actuation of the setup mode actuator may permit rotation of the collar in a first direction to cause rotation of the distal link adjacent to the setup joint in a corresponding first direction relative to the proximal link adjacent to the setup joint, and may permit rotation of the collar in a second direction to cause rotation of the distal link adjacent to the setup joint in a corresponding second direction relative to the proximal link adjacent to the setup joint. The co-manipulation surgical system may not be teleoperated via user input received at a remote surgeon console.
In accordance with another aspect of the present disclosure, a method for assisting with laparoscopic surgery using a robot arm comprising a plurality of links, a plurality of joints comprising one or more motorized joints, a setup joint, and one or more passive joints, a proximal end operatively coupled to a base of the robot arm, and a distal region having a distal end configured to be removably coupled to a surgical instrument is provided. The method may include: actuating an actuator operatively coupled to a motor operatively coupled to the setup joint to cause rotation of a distal link of the plurality of links adjacent to the setup joint relative to a proximal link of the plurality of links adjacent to the setup joint from a first setup configuration to a second setup configuration responsive to actuation of the actuator; and moving, when the actuator is in an unactuated state, the robot arm responsive to movement at the handle of the surgical instrument for performing laparoscopic surgery via the one or more motorized joints and the one or more passive joints while the distal link adjacent to the setup joint and proximal link adjacent to the setup joint remain in the second setup configuration. For example, actuating the actuator to cause rotation of the distal link adjacent to the setup joint relative to the proximal link adjacent to the setup joint may comprise rotating a collar rotatably coupled to a link of the plurality of links in a first direction to cause rotation of the distal link adjacent to the setup joint in a corresponding first direction relative to the proximal link adjacent to the setup joint, and rotating the collar in a second direction to cause rotation of the distal link adjacent to the setup joint in a corresponding second direction relative to the proximal link adjacent to the setup joint. Moreover, actuating the actuator to cause rotation of the distal link adjacent to the setup joint relative to the proximal link adjacent to the setup joint further may comprise actuating a setup mode actuator disposed on the collar to permit the rotation of the distal link adjacent to the setup joint in the corresponding first and second directions relative to the proximal link adjacent to the setup joint responsive to rotation of the collar.
In addition, actuating the actuator to cause rotation of the link distal to the setup joint relative to the link proximal to the setup joint may comprise actuating the actuator displayed on a graphical user interface. For example, actuating the actuator displayed on the graphical user interface may comprise moving a slidable cursor relative to a neutral center point, such that movement of the slidable cursor in a first direction relative to the neutral center point causes rotation of the distal link adjacent to the setup joint in a first direction relative to the proximal link adjacent to the setup joint, and movement of the slidable cursor in a second direction relative to the neutral center point causes rotation of the distal link adjacent to the setup joint in a second direction relative to the proximal link adjacent to the setup joint. Accordingly, the method further may include displaying, via the graphical user interface, an indicator indicative of a configuration of the distal link adjacent to the setup joint relative to the proximal link adjacent to the setup joint in real-time responsive to actuation of the actuator.
In addition, the method may include displaying, via the graphical user interface, graphical representations of a plurality of configurations of the distal link adjacent to the setup joint relative to the proximal link adjacent to the setup joint, such that a position of the indicator relative to the graphical representations of the plurality of configurations is indicative of the configuration of the distal link adjacent to the setup joint relative to the proximal link adjacent to the setup joint in real-time responsive to actuation of the actuator. Moreover, actuating the actuator to cause rotation of the link distal to the setup joint relative to the link proximal to the setup joint may comprise applying, when the actuator is in an actuated state, a force at the distal region of the robot arm in a direction to cause rotation of the distal link adjacent to the setup joint in a corresponding direction relative to the proximal link adjacent to the setup joint. The method further may include applying, in a user guided setup mode, a force at the distal region of the robot arm in a direction to cause a stage assembly operatively coupled to the base of the robot arm to move the base of the robot arm in a corresponding direction relative to a platform coupled to the stage assembly.
In accordance with another aspect of the present disclosure, a co-manipulation surgical system for providing adaptive gravity compensation to a robot arm comprising a plurality of links, a plurality of joints, and a distal end configured to be removably coupled to a surgical instrument is provided. The co-manipulation surgical system may comprise at least one processor configured to: apply an initial gravity compensation to the robot arm to compensate for gravity of the surgical instrument based on an estimated instrument parameter associated with the surgical instrument; calculate, during application of the initial gravity compensation, a hold force required to maintain the distal end of the robot arm in a static position in a passive mode; and determine a calibrated instrument parameter for the surgical instrument based on the hold force, the calibrated instrument parameter selected to adjust the hold force required to maintain the distal end of the robot arm in the static position in the passive mode during application of an adjusted gravity compensation to the robot arm based on the calibrated instrument parameter.
The at least one processor further may be configured to apply torque to one or more motorized joints of the plurality of joints of the robot arm to apply the initial gravity compensation to the robot arm to compensate for gravity of the surgical instrument. The estimated instrument parameter and the calibrated instrument parameter may comprise at least one of a mass or a center of mass associated with the surgical instrument. In addition, the at least one processor may be configured to: load a calibration file associated with a known parameter of the surgical instrument, such that the calibration file may comprise the estimated instrument parameter. For example, the known parameter may comprise a diameter of an elongated shaft of the surgical instrument. Moreover, the at least one processor may be configured to determine the known parameter upon coupling of the surgical instrument to the distal end of the robot arm via a coupler body removably coupled to the surgical instrument and to the distal end of the robot arm. In some embodiments, the at least one processor may be configured to determine the known parameter based on the coupler body. The system further may include an optical sensor configured to collect depth data, such that the at least one processor may be configured to determine the known parameter based on the depth data. Additionally, or alternatively, the system may include a user interface operatively coupled to the at least one processor, such that the at least one processor is configured to determine the known parameter via user input received by the user interface.
The calibrated instrument parameter may be selected to adjust the hold force during application of the adjusted gravity compensation based on the calibrated instrument parameter within a predetermined range associated with a known parameter of the surgical instrument. Moreover, when the distal end of the robot arm is not subjected to any external forces other than gravity on the robot arm and the surgical instrument in the static position, the calibrated instrument parameter may be selected to adjust the hold force to or near zero upon application of the adjusted gravity compensation based on the calibrated instrument parameter. In addition, when the distal end of the robot arm is subjected to one or more external forces in addition to gravity on the robot arm and the surgical instrument in the static position, the calibrated instrument parameter may be selected to adjust the hold force within a predetermined range associated with a known parameter of the surgical instrument.
The at least one processor further may be configured to: calculate the adjusted gravity compensation of the surgical instrument based on the calibrated instrument parameter; and apply the adjusted gravity compensation to the robot arm to compensate for gravity of the surgical instrument. For example, the at least one processor may be configured to apply torque to one or more motorized joints of the plurality of joints of the robot arm to apply the adjusted gravity compensation to the robot arm to compensate for gravity of the surgical instrument. Moreover, the at least one processor may be configured to cause the robot arm to automatically switch to a co-manipulation mode responsive to determining that force applied at the robot arm due to force applied at a handle of the surgical instrument exceeds a predetermined force threshold. Additionally, the at least one processor may be configured to permit the robot arm to be freely moveable in the co-manipulation mode responsive to movement at the handle of the surgical instrument, while applying the adjusted gravity compensation to the robot arm to compensate for gravity of the surgical instrument in the co-manipulation mode.
The at least one processor further may be configured to calculate the adjusted hold force to maintain the distal end of the robot arm in the static position in the passive mode upon application of the adjusted gravity compensation. Accordingly, the at least one processor may be configured to: establish a baseline hold force based on the adjusted hold force after a predetermined time period upon initiation of the passive mode; and apply a predetermined constant breakaway force threshold to the robot arm based on the baseline hold force, such that the at least one processor may not maintain the distal end of the robot arm in the static position if the hold force exceeds the predetermined constant breakaway force threshold. In addition, the at least one processor may be configured to apply a predetermined high breakaway force threshold during the predetermined time period, such that the at least one processor may not maintain the distal end of the robot arm in the static position if the hold force exceeds the predetermined high breakaway force threshold during the predetermined time period. Moreover, the at least one processor may be configured to cause the robot arm to automatically switch to the passive mode responsive to determining that movement of the robot arm due to movement at a handle of the surgical instrument is less than a predetermined amount for at least a predetermined dwell time period. The at least one processor further may be configured to record the calibrated instrument parameter in a calibration file associated with the surgical instrument.
In accordance with another aspect of the present disclosure, a method for assisting with laparoscopic surgery using a robot arm comprising a proximal end, a distal end configured to be removably coupled to a surgical instrument, a plurality of links, and a plurality of joints between the proximal end and the distal end is provided. The method may include: applying, via a controller operatively coupled to the robot arm, an initial gravity compensation to the robot arm to compensate for gravity of the surgical instrument when the surgical instrument is coupled to the distal end of the robot arm based on an estimated instrument parameter associated with the surgical instrument; calculating, via the controller during application of the initial gravity compensation, a hold force required to maintain the distal end of the robot arm in a static position in a passive mode; and determining, via the controller, a calibrated instrument parameter for the surgical instrument based on the hold force, the calibrated instrument parameter selected to adjust the hold force required to maintain the distal end of the robot arm in the static position in the passive mode during application of an adjusted gravity compensation to the robot arm based on the calibrated instrument parameter. The estimated instrument parameter and the calibrated instrument parameter may comprise at least one of a mass or a center of mass associated with the surgical instrument.
The method further may include loading, via the controller, a calibration file associated with a known parameter of the surgical instrument, such that the calibration file may comprise the estimated instrument parameter. For example, the known parameter may comprise a diameter of an elongated shaft of the surgical instrument. In addition, the method may include: coupling the surgical instrument to the distal end of the robot arm via a coupler body removably coupled to the surgical instrument; and determining, via the controller, the known parameter based on the coupler body. Additionally, the method may include determining, via the controller, the known parameter via user input received by a user interface operatively coupled to the controller. Moreover, determining the calibrated instrument parameter based on the hold force may comprise determining the calibrated instrument parameter selected to adjust the hold force upon application of the adjusted gravity compensation within a predetermined range associated with a known parameter of the surgical instrument. The method further may include: calculating, via the controller, the adjusted gravity compensation of the surgical instrument based on the calibrated instrument parameter; and applying, via the controller, torque to one or more motorized joints of the plurality of joints of the robot arm to apply the adjusted gravity compensation to the robot arm to compensate for gravity of the surgical instrument.
In addition, the method may include: automatically switching, via the controller, to a co-manipulation mode responsive to determining that force applied at the robot arm due to force applied at the handle of the surgical instrument exceeds a predetermined force threshold; and permitting, via the controller, the robot arm to be freely moveable in the co-manipulation mode responsive to movement at the handle of the surgical instrument, while applying the adjusted gravity compensation to the robot arm to compensate for gravity of the surgical instrument in the co-manipulation mode. The method further may include: calculating, via the controller, the adjusted hold force to maintain the distal end of the robot arm in the static position in the passive mode upon application of the adjusted gravity compensation; establishing, via the controller, a baseline hold force based on the adjusted hold force after a predetermined time period upon initiation of the passive mode; and applying, via the controller, a predetermined constant breakaway force threshold to the robot arm based on the baseline hold force, wherein the controller does not maintain the distal end of the robot arm in the static position if the hold force exceeds the predetermined constant breakaway force threshold.
In accordance with another aspect of the present disclosure, a co-manipulation surgical system for operating a robot arm comprising a plurality of links, a plurality of joints, and a distal end configured to be removably coupled to a surgical instrument is provided. The co-manipulation surgical system may comprise at least one processor configured to: cause the robot arm to switch to a passive mode responsive to determining that movement of the robot arm due to movement at a handle of the surgical instrument is less than a predetermined amount for at least a predetermined dwell time period, the at least one processor configured to cause the robot arm to maintain a static position in the passive mode; apply gravity compensation to the robot arm to compensate for gravity of the surgical instrument; calculate, during application of the gravity compensation, a hold force required to maintain the distal end of the robot arm in the static position in the passive mode; establish a baseline hold force based on the hold force; and apply a breakaway force threshold to the robot arm based on the baseline hold force, the breakaway force threshold being a predetermined amount of force required to be applied to the robot arm to cause the robot arm to exit the passive mode. A magnitude of the breakaway force threshold may be equal in every direction relative to the baseline hold force. For example, a total amount of force required to be applied to the robot arm in a direction to cause the robot arm to exit the passive mode may be a sum of the baseline hold force and the breakaway force threshold in the direction.
The hold force required to maintain the distal end of the robot arm in the static position may be continuously calculated in the passive mode. Accordingly, the at least one processor may be configured to determine that the surgical instrument is in contact with one or more anatomical structures in the passive mode if the hold force gradually increases over time. In addition, the at least one processor may be configured to calculate the hold force required to maintain the distal end of the robot arm in the static position in the passive mode when one or more external forces are applied to the surgical instrument by one or more anatomical structures having an unknown mass. Moreover, the at least one processor may be configured to determine a force required to be applied to the distal end of the robot arm to move the distal end of the robot arm from a current position to the static position to calculate the hold force required to maintain the distal end of the robot arm in the static position in the passive mode. The at least one processor may further be configured to cause the robot arm to automatically switch to a co-manipulation mode responsive to determining that the hold force required to maintain the distal end of the robot arm in the static position exceeds the breakaway force threshold, such that the at least one processor may be configured to permit the robot arm to be freely moveable in the co-manipulation mode responsive to movement at the handle of the surgical instrument, while applying the gravity compensation to the robot arm to compensate for gravity of the surgical instrument in the co-manipulation mode.
The at least one processor may be configured to sense a force applied at the distal end of the robot arm to calculate the hold force required to maintain the distal end of the robot arm in the static position in the passive mode. For example, the at least one processor may be configured to measure current of a plurality of motors operatively coupled to at least some joints of the plurality of joints to sense the force applied at the distal end of the robot arm. Moreover, the at least one processor may be configured to apply torque to at least some joints of the plurality of joints of the robot arm to apply the gravity compensation to the robot arm to compensate for gravity of the surgical instrument.
In addition, the at least one processor may be configured to establish the baseline hold force after a predetermined time period upon initiation of the passive mode. Accordingly, the at least one processor may be configured to: apply a high breakaway force threshold to the robot arm during the predetermined time period, the high breakaway force threshold greater than the breakaway force threshold, such that the at least one processor may be configured to cause, if the hold force required to maintain the distal end of the robot arm in the static position exceeds the high breakaway force threshold during the predetermined time period, the robot arm to exit the passive mode. For example, the high breakaway force threshold may be selected to prevent inadvertent disengagement of the robot arm from passive mode in response to inadvertent forces applied at the distal end of the robot arm during the predetermined time period. The at least one processor may further be configured to: apply an initial breakaway force threshold to the robot arm during the predetermined time period; and apply, if force applied at the distal end of the robot arm exceeds the initial breakaway force threshold during the predetermined time period, a high breakaway force threshold during the predetermined time period, the high breakaway force threshold greater than the breakaway force threshold, such that the at least one processor may be configured to cause, if the hold force required to maintain the distal end of the robot arm in the static position exceeds the high breakaway force threshold during the predetermined time period, the robot arm to exit the passive mode.
Moreover, the at least one processor may be configured to: apply, if the hold force fluctuates after the predetermined time period upon initiation of the passive mode such that the baseline hold force cannot be established based on the calculated hold force, a default breakaway force threshold to the robot arm, such that the at least one processor may be configured to cause, if the hold force required to maintain the distal end of the robot arm in the static position exceeds the default breakaway force threshold, the robot arm to exit the passive mode. For example, the at least one processor may be configured to select the default breakaway force threshold from between a default high breakaway force threshold and a default low breakaway force threshold based on user input via a graphical user interface operatively coupled to the at least one processor. Additionally, the at least one processor may be configured to adjust at least one of the default high breakaway force threshold or the default low breakaway force threshold based on user input via the graphical user interface.
The at least one processor may further be configured to: apply the gravity compensation to the robot arm to compensate for gravity of the surgical instrument based on an estimated instrument parameter associated with the surgical instrument; determine a calibrated instrument parameter for the surgical instrument based on the hold force; and apply an adjusted gravity compensation to the robot arm based on the calibrated instrument parameter, such that the baseline hold force may be established based on the hold force required to maintain the distal end of the robot arm in the static position in the passive mode upon application of the adjusted gravity compensation to the robot arm. Moreover, the calibrated instrument parameter may be selected such that, during application of the adjusted gravity compensation, the hold force is adjusted within a predetermined range associated with a known parameter of the surgical instrument.
In accordance with another aspect of the present disclosure, a method for assisting with laparoscopic surgery using a robot arm comprising a proximal end, a distal end configured to be removably coupled a surgical instrument, a plurality of links, and a plurality of joints between the proximal end and the distal end is provided. The method may include: causing, via a controller operatively coupled to the robot arm, the robot arm to switch to a passive mode responsive to determining that movement of the robot arm due to movement at a handle of the surgical instrument is less than a predetermined amount for at least a predetermined dwell time period, the controller configured to cause the robot arm to maintain a static position in the passive mode; applying, via the controller, gravity compensation to the robot arm to compensate for gravity of the surgical instrument; calculating, via the controller during application of the gravity compensation, a hold force required to maintain the distal end of the robot arm in a static position in the passive mode; establishing, via the controller, a baseline hold force based on the hold force; and applying, via the controller, a breakaway force threshold to the robot arm based on the baseline hold force, the breakaway force threshold being a predetermined amount of force required to be applied to the robot arm to cause the robot arm to exit the passive mode. A magnitude of the breakaway force threshold may be equal in every direction relative to the baseline hold force, and a total amount of force required to be applied to the robot arm in a direction to cause the robot arm to exit the passive mode may be a sum of the baseline hold force and the breakaway force threshold in the direction.
Calculating the hold force required to maintain the distal end of the robot arm in the static position in the passive mode may comprise continuously calculating, via the controller, the hold force required to maintain the distal end of the robot arm in the static position in the passive mode. Moreover, the method may include determining, via the controller, that the surgical instrument is in contact with one or more anatomical structures in the passive mode if the hold force gradually increases over time. In addition, calculating the hold force required to maintain the distal end of the robot arm in the static position in the passive mode may comprise calculating, via the controller, the hold force required to maintain the distal end of the robot arm in the static position in the passive mode when one or more external forces are applied to the surgical instrument by one or more anatomical structures having an unknown mass. Additionally or alternatively, calculating the hold force required to maintain the distal end of the robot arm in the static position in the passive mode may comprise determining, via the controller, a force required to be applied to the distal end of the robot arm to move the distal end of the robot arm from a current position to the static position. The method further may include causing, via the controller, the robot arm to automatically switch to a co-manipulation mode responsive to determining that the hold force required to maintain the distal end of the robot arm in the static position exceeds the breakaway force threshold, such that the robot arm may be permitted to be freely moveable in the co-manipulation mode responsive to movement at the handle of the surgical instrument, while the gravity compensation is applied to the robot arm to compensate for gravity of the surgical instrument in the co-manipulation mode.
Moreover, calculating the hold force required to maintain the distal end of the robot arm in the static position in the passive mode may comprise sensing, via the controller, a force applied at the distal end of the robot arm. For example, sensing the force applied at the distal end of the robot arm may comprise measuring, via the controller, current of a plurality of motors operatively coupled to at least some joints of the plurality of joints. Additionally, applying gravity compensation to the robot arm to compensate for gravity of the surgical instrument may comprise applying, via the controller, torque to at least some joints of the plurality of joints of the robot arm. Establishing the baseline hold force based on the hold force may comprise establishing, via the controller, the baseline hold force after a predetermined time period upon initiation of the passive mode. Accordingly, the method further may include: applying, via the controller, a high breakaway force threshold to the robot arm during the predetermined time period, the high breakaway force threshold greater than the breakaway force threshold; and causing, via the controller if the hold force required to maintain the distal end of the robot arm in the static position exceeds the high breakaway force threshold during the predetermined time period, the robot arm to exit the passive mode. For example, the high breakaway force threshold may be selected to prevent inadvertent disengagement of the robot arm from passive mode in response to inadvertent forces applied at the distal end of the robot arm during the predetermined time period.
The method further may include: applying, via the controller, an initial breakaway force threshold to the robot arm during the predetermined time period; applying, via the controller if force applied at the distal end of the robot arm exceeds the initial breakaway force threshold during the predetermined time period, a high breakaway force threshold during the predetermined time period, the high breakaway force threshold greater than the breakaway force threshold; and causing, via the controller if the hold force required to maintain the distal end of the robot arm in the static position exceeds the high breakaway force threshold during the predetermined time period, the robot arm to exit the passive mode. The method further may include: applying, via the controller, if the hold force fluctuates after the predetermined time period upon initiation of the passive mode such that the baseline hold force cannot be established based on the calculated hold force, a default breakaway force threshold to the robot arm; and causing, via the controller if the hold force required to maintain the distal end of the robot arm in the static position exceeds the default breakaway force threshold, the robot arm to exit the passive mode. In addition, the method may include selecting, via the controller, the default breakaway force threshold from between a default high breakaway force threshold and a default low breakaway force threshold based on user input via a graphical user interface operatively coupled to the controller. The method further may include adjusting, via the controller, at least one of the default high breakaway force threshold or the default low breakaway force threshold based on user input via the graphical user interface.
Moreover, applying gravity compensation to the robot arm to compensate for gravity of the surgical instrument may comprise applying, via the controller, the gravity compensation to the robot arm to compensate for gravity of the surgical instrument based on an estimated instrument parameter associated with the surgical instrument. Accordingly, the method further may include: determining, via the controller, a calibrated instrument parameter for the surgical instrument based on the hold force; and applying, via the controller, an adjusted gravity compensation to the robot arm based on the calibrated instrument parameter, such that establishing the baseline hold force based on the hold force may comprise establishing, via the controller, the baseline hold force based on the hold force required to maintain the distal end of the robot arm in the static position in the passive mode upon application of the adjusted gravity compensation to the robot arm. For example, the calibrated instrument parameter may be selected such that, during application of the adjusted gravity compensation, the hold force is adjusted within a predetermined range associated with a known parameter of the surgical instrument.
In accordance with another aspect of the present disclosure, another co-manipulation surgical system to assist with laparoscopic surgery performed using a surgical instrument having a handle, an operating end, and an elongated shaft therebetween is provided. The system may include a robot arm comprising a proximal end, a distal end configured to be removably coupled to the surgical instrument, a plurality of links, and a plurality of joints, and a controller operatively coupled to the robot arm and configured to permit the robot arm to be freely moveable responsive to movement at the handle of the surgical instrument for performing the surgical procedure using the surgical instrument. The controller may be programmed to: identify a type of the surgical instrument coupled to the distal end of the robot arm; apply a first impedance to the robot arm to account for weight of the surgical instrument and the robot arm; and apply a second impedance to the robot arm based on the type of the surgical instrument to adjust viscosity at the distal end of the robot arm to thereby guide a movement of the surgical instrument by the user during a predetermined phase of the surgical procedure.
For example, the identified type of the surgical instrument may comprise a suturing device, and the predetermined phase of the surgical procedure may comprise a suturing phase, such that the second impedance may be sufficient to provide more viscous control of the suturing device during the suturing phase of the surgical procedure. Additionally, or alternatively, the identified type of the surgical instrument may comprise a stapling device, and the predetermined phase of the surgical procedure may comprise a stapling phase, such that the second impedance may be sufficient to provide stiff grounding to facilitate force application of the stapling device during the stapling phase of the surgical procedure. The controller may further be configured to identify the predetermined phase of the surgical procedure based on the type of the surgical instrument. Moreover, the type of the surgical instrument may be selected from a list comprising at least one of a wristed instrument, a stapling device, a dissection device, a suturing device, a retraction device, a tissue removal device, or a clip applier device. The controller may be configured to apply the second impedance to the robot arm based on the type of the surgical instrument to adjust viscosity at the distal end of the robot arm to thereby guide the movement of the surgical instrument by the user during the predetermined phase of the surgical procedure without actively causing movement of the robot arm.
In accordance with another aspect of the present disclosure, a computer implemented system for providing image registration to a robot arm comprising a plurality of links, a plurality of joints, and a distal end configured to be removably coupled to a laparoscope having a rotatable camera sensor module is provided. The system may comprise at least one processor configured to: retrieve a plurality of images from the laparoscope during movement of a field of view of the laparoscope; compute motion of individual pixels between consecutive images of the plurality of images via a computer vision technique, the motion of individual pixels indicative of image motion; calculate an average of the motion of individual pixels in an x and y direction of the plurality of images to obtain an image motion direction; and compute an angular offset between the camera sensor module and the distal end of the robot arm based on the image motion direction.
The at least one processor further may be configured to: synchronize the image motion and movement of the distal end of the robot arm associated with the movement of the field of view of the laparoscope; and compare the image motion direction with the movement of the distal end of the robot arm to compute the angular offset between the camera sensor module and the distal end of the robot arm. Moreover, the at least one processor may be configured to cause, in a foreground mode, the robot arm to move the laparoscope along a predetermined trajectory, such that the image motion may be synchronized with movement of the distal end of the robot arm associated with movement of the laparoscope along the predetermined trajectory. Additionally, or alternatively, in a background mode, the image motion may be synchronized with movement of the distal end of the robot arm responsive to movement of the field of view of the laparoscope by a user. The at least one processor may be configured to retrieve data indicative of the movement of the distal end of the robot arm via one or more sensors operatively coupled to at least some joints of the plurality of joints of the robot arm.
Moreover, the at least one processor may be configured to validate the image motion direction. For example, the at least one processor may be configured to: calculate a norm of a vector of the image motion direction to determine a magnitude of the image motion; and compare the magnitude of the image motion with a magnitude threshold. Accordingly, the image motion direction may be validated if the magnitude of the image motion exceeds the magnitude threshold. Additionally, or alternatively, the at least one processor may be configured to: calculate a percentage of image pixels that moved between consecutive images based on the motion of individual pixels; and compare the percentage with a percentage threshold. Accordingly, the image motion direction may be validated if the percentage exceeds the percentage threshold. The at least one processor may be configured to determine whether the image motion is due to at least one of movement of the field of view of the laparoscope or local motion of one or more tools or tissue within the plurality of images based on the comparison of the percentage with the percentage threshold. Additionally, or alternatively, the at least one processor may be configured to: calculate a relative angle between each motion of the individual pixels and the image motion direction to determine whether each motion of the individual pixels are in agreement with the image motion direction; and compare a percentage of individual pixels motion that are in agreement with the image motion direction with an agreement threshold. Accordingly, the image motion direction may be validated if the percentage exceeds the agreement threshold.
The at least one processor further may be configured to: cause the robot arm to move the laparoscope along a predetermined axial trajectory; compare the image motion direction with a direction threshold; and determine whether the laparoscope has a flat or angled tip based on the comparison of the image motion direction with the direction threshold. In some embodiments, movement of the field of view of the laparoscope may be due to zooming of the camera sensor module, such that the at least one processor may be configured to: compare the image motion direction with a direction threshold; and determine whether the laparoscope has a flat or angled tip based on the comparison of the image motion direction with the direction threshold. The controller further may be configured to cause the robot arm to automatically switch to a co-manipulation mode responsive to determining that force applied at the robot arm due to force applied at the laparoscope exceeds a predetermined threshold. Accordingly, the controller may be configured to permit the robot arm to be freely moveable in the co-manipulation mode responsive to movement at the laparoscope, while applying an impedance to the robot arm in the co-manipulation mode to account for weight of the laparoscope and the robot arm.
In accordance with another aspect of the present disclosure, a system for robotic surgery is provided. The system may include a robot arm comprising a proximal end operatively coupled to a base of the robot arm, a distal end, a plurality of links, and a plurality of joints between the proximal end and the distal end, and the robot arm may be configured to be positioned adjacent to a bed for holding a patient during surgery. The system further may include a platform coupled to the base of the robot arm, and the platform may comprise a stage assembly configured to independently move the base of the robot arm in at least two degrees of freedom relative to the platform. In addition, the system may include a graphical user interface comprising a plurality of predetermined, selectable surgical procedures, and a controller operatively coupled to the robot arm. The controller may be programmed to: during a surgery setup phase, automatically position the robot arm in a first position specific to a first surgical procedure relative to the bed in response to selection of the first surgical procedure of the plurality of predetermined, selectable surgical procedures; and during the surgery setup phase, automatically position the robot arm in a second position specific to a second surgical procedure relative to the bed in response to selection of the second surgical procedure of the plurality of predetermined, selectable surgical procedures. For example, the first position specific to the first surgical procedure may be different than the second position specific to the second surgical procedure.
Moreover, the controller may be configured to, upon selection of the cholecystectomy: cause a shoulder link of the plurality of links of the robot arm to rotate in a leftward direction relative to the platform; and cause the stage assembly to move the base of the robot arm in a downward direction of a first degree of freedom of the at least two degrees of freedom and in an outward direction of a second degree of freedom of the at least two degrees of freedom. In addition, the system may include a second robot arm, such that the controller may be configured to, upon selection of the cholecystectomy, cause a stage assembly of the second robot arm to move a base of the second robot arm in an upward direction of a first degree of freedom of at least two degrees of freedom and in an inward direction of a second degree of freedom of the at least two degrees of freedom. At least one of the first or second surgical procedures may comprise a cholecystectomy, gastric sleeve, hiatal hernia repair, Nissen fundoplication, inguinal hernia repair (TEP), right, left, and/or complete colectomy, gastric bypass, sigmoid colectomy, umbilical hernia repair, or incisional hernia repair.
In accordance with another aspect of the present disclosure, a co-manipulation surgical system to assist with laparoscopic surgery performed using a surgical instrument is provided. The system may include a plurality of robot arms, each robot arm comprising a plurality of links, a plurality of joints, a proximal region operatively coupled to a base, and a distal region configured to be removably coupled to the surgical instrument. The system further may include a platform coupled to the base of each robot arm, wherein, in a surgical mode, each of the robot arms may be configured to extend away from the platform and towards a surgical site, and a controller configured to, upon actuation of a stow mode, cause each of the robot arms to transition to a retracted stow configuration via the robot arm's respective plurality of links and joints and rotate about the robot arm's respective base such that each of the robot arms is within a footprint of the platform in the retracted stow configuration. For example, in the retracted stow configuration, each of the robot arms may be entirely within the footprint of the platform.
Upon actuation of the stow mode, the controller may cause a first robot arm of the plurality of robot arms to rotate in a first direction about its base, and may cause a second robot arm of the plurality of robot arms to rotate in a second direction about its base, the second direction opposite to the first direction. Alternatively, the second direction may be the same as the first direction. In the retracted stow configuration, a first robot arm of the plurality of robot arms may have a same configuration and orientation as a second robot arm of the plurality of robot arms relative to the platform. In addition, in the retracted stow configuration, each of the robot arms may extend in a direction towards a rear of the platform. Upon actuation of the stow mode, the controller may cause at least one robot arm of the plurality of robot arms to move to a position via the at least one robot arm's respective plurality of links and joints, such that, upon rotation of the at least one robot arm about the at least one robot arm's respective base, collision between the at least one robot arm and the platform is avoided.
The platform may comprise a stage assembly configured to independently move the base of each robot arm in at least two degrees of freedom relative to the platform. Accordingly, upon actuation of the stow mode, the controller may cause the stage assembly to move the base of each robot arm in an inward direction along a first degree of freedom of the at least two degrees of freedom, such that, upon rotation of each of the robot arms about the robot arm's respective base, each of the robot arms is within the footprint of the platform in the retracted stow configuration. In addition, upon actuation of the stow mode, the controller may cause the stage assembly to move the base of each robot arm in a downward direction along a second degree of freedom of the at least two degrees of freedom, such that the robot arms have a minimal height relative to the platform for transportation and/or storage of the co-manipulation surgical system in the retracted stow configuration. Moreover, upon actuation of the stow mode, the controller may cause the stage assembly to move the base of each robot arm in a downward direction along a second degree of freedom of the at least two degrees of freedom, such that the robot arms have a minimal height relative to the platform for transportation and/or storage of the co-manipulation surgical system in the retracted stow configuration. Additionally, upon actuation of the stow mode, the controller may cause the stage assembly to move the base of at least one robot arm in an outward direction along a first degree of freedom of the at least two degrees of freedom and/or an upward direction along a second degree of freedom of the at least two degrees of freedom, such that, upon rotation of at least one robot arm of the plurality of robot arms about the at least one robot arm's respective base, collision between the at least one robot arm and the platform is avoided. In some embodiments, upon actuation of the stow mode, the controller may cause the stage assembly to rotate the base of each robot arm in a direction towards one another along a horizontal plane relative to the platform, such that, upon rotation of each of the robot arms about the robot arm's respective base, each of the robot arms is within the footprint of the platform in the retracted stow configuration.
The system further may include a graphical user interface operatively coupled to the controller. The graphical user interface may be configured to display a stow mode actuator configured to be actuated by a user to actuate the stow mode. Moreover, upon actuation of a compact mode, the controller may cause each of the robot arms to transition to a semi-retracted configuration via the robot arm's respective plurality of links and joints and rotate about the robot arm's respective base such that each of the robot arms extend away from the platform to facilitate transportation of the co-manipulation surgical system, wherein the semi-retracted configuration is less retracted than the retracted stow configuration of stow mode. In addition, upon actuation of a drape mode, the controller may cause each of the robot arms to transition to an extended configuration via the robot arm's respective plurality of links and joints and rotate about the robot arm's respective base such that each of the robot arms extend away from the platform to facilitate draping of the robot arms. In the extended configuration, distal links of the plurality of links of each robot arm may be substantially parallel to the ground.
In accordance with another aspect of the present disclosure, a method for assisting with laparoscopic surgery is provided. The method may include providing a plurality of robot arms, each robot arm comprising a plurality of links, a plurality of joints, a proximal region operatively coupled to a base, and a distal region configured to be removably coupled to a surgical instrument, each base coupled to a platform such that, in a surgical mode, each of the robot arms may be configured to extend away from the platform and towards a surgical site; and automatically causing, upon actuation of a stow mode, each of the robot arms to transition to a retracted stow configuration via the robot arm's respective plurality of links and joints and rotate about the robot arm's respective base such that each of the robot arms is within a footprint of the platform in the retracted stow configuration. Automatically causing each of the robot arms to transition to the retracted stow configuration may comprise automatically rotating a first robot arm of the plurality of robot arms in a first direction about its base, and automatically rotating a second robot arm of the plurality of robot arms in a second direction about its base, the second direction opposite to the first direction. Alternatively, the second direction may be the same as the first direction. In addition, automatically causing each of the robot arms to transition to the retracted stow configuration may comprise automatically causing at least one robot arm of the plurality of robot arms to move to a position via the at least one robot arm's respective plurality of links and joints, such that, upon rotation of the at least one robot arm about the at least one robot arm's respective base, collision between the at least one robot arm and the platform is avoided.
The platform may comprise a stage assembly configured to independently move the base of each robot arm in at least two degrees of freedom relative to the platform. Accordingly, automatically causing each of the robot arms to transition to the retracted stow configuration may comprise automatically causing the stage assembly to move the base of each robot arm in an inward direction along a first degree of freedom of the at least two degrees of freedom, such that, upon rotation of each of the robot arms about the robot arm's respective base, each of the robot arms is within the footprint of the platform in the retracted stow configuration. Automatically causing each of the robot arms to transition to the retracted stow configuration may comprise automatically causing the stage assembly to move the base of each robot arm in a downward direction along a second degree of freedom of the at least two degrees of freedom, such that the robot arms have a minimal height relative to the platform for transportation and/or storage in the retracted stow configuration. In addition, automatically causing each of the robot arms to transition to the retracted stow configuration may comprise automatically causing the stage assembly to move the base of at least one robot arm in an outward direction along a first degree of freedom of the at least two degrees of freedom and/or an upward direction along a second degree of freedom of the at least two degrees of freedom, such that, upon rotation of at least one robot arm of the plurality of robot arms about the at least one robot arm's respective base, collision between the at least one robot arm and the platform is avoided.
The method further may include displaying, via a graphical user interface, a stow mode actuator configured to be actuated by a user to actuate the stow mode. Moreover, the method may include automatically causing, upon actuation of a compact mode, each of the robot arms to transition to a semi-retracted configuration via the robot arm's respective plurality of links and joints and rotate about the robot arm's respective base such that each of the robot arms extend away from the platform to facilitate transportation of the platform, wherein the semi-retracted configuration is less retracted than the retracted stow configuration of stow mode. In addition, the method may include automatically causing, upon actuation of a drape mode, each of the robot arms to transition to an extended configuration via the robot arm's respective plurality of links and joints and rotate about the robot arm's respective base such that each of the robot arms extend away from the platform to facilitate draping of the robot arms. In the extended configuration, distal links of the plurality of links of each robot arm may be substantially parallel to the ground. The method further may include covering the plurality of robot arms with a sterile drape in the extended configuration.
In accordance with another aspect of the present disclosure, a co-manipulation surgical system to assist with laparoscopic surgery performed using a surgical instrument is provided. The co-manipulation surgical system may comprise a robot arm comprising a proximal region having a base, a plurality of links, an elbow joint, a shoulder joint coupled to the base, and a distal region having a distal end configured to be removably coupled to the surgical instrument, and a platform coupled to the base of the robot arm, the platform configured to be moveable on a floor. Accordingly, in a drape mode, a link of the plurality of links between the elbow joint and the shoulder joint may be configured to extend away from the base at a first angle and a link of the plurality of links distal to the elbow joint may be configured to extend away from the base at a second angle so as to be more aligned with the floor, to thereby facilitate draping of the robot arm. For example, the first angle may be selected such that the link between the elbow joint and the shoulder joint is configured to be substantially vertical in the drape mode. In addition, the second angle may be selected such that the link distal to the elbow joint is configured to be substantially horizontal in the drape mode. Additionally or alternatively, the second angle may be selected such that the link distal to the elbow joint is configured to be substantially parallel with the floor in the drape mode.
The platform may comprise a plurality of wheels configured to permit mobility of the platform on the floor. In addition, the robot arm may not be teleoperated via user input received at a remote surgeon console. The co-manipulation surgical system further may comprise a controller operatively coupled to the robot arm. The controller may be configured to, upon actuation of the drape mode, automatically cause the link between the elbow joint and the shoulder joint to extend away from the base at the first angle and the link distal to the elbow joint to extend away from the base at the second angle. For example, upon actuation of the drape mode, the controller may automatically cause the robot arm to rotate about the base such that the link between the elbow joint and the shoulder joint extends away from the platform at the first angle and the link distal to the elbow joint extends away from the platform at the second angle. The controller may further be configured to apply a virtual haptic boundary about the robot arm such that, when the robot arm is within the virtual haptic boundary, the controller applies a high level of impedance to the robot arm such that movement of the robot arm is more viscous within the virtual haptic boundary. Additionally or alternatively, the controller may further be configured to apply a temporary localized virtual haptic boundary at the distal region of the robot arm in the drape mode such that movement of the robot arm is more viscous, thereby stabilizing the robot arm in the drape mode.
The co-manipulation surgical system further may comprise a coupler configured to be removably coupled to the distal end of the robot arm after application of a sterile drape over the robot arm, the coupler configured to be removably coupled to the surgical instrument, e.g., to thereby removably couple the surgical instrument to the distal end of the robot arm. Accordingly, the controller may further be configured to remove the temporary localized virtual haptic boundary at the distal region of the robot arm upon detection of the coupler being removably coupled to the distal end of the robot arm. Moreover, the platform may be configured to move the base of the robot arm in at least two degrees of freedom relative to the floor, such that, upon actuation of the drape mode, the controller may automatically cause the platform to move the base of the robot arm in at least one of the at least two degrees of freedom relative to the floor to align the link distal to the elbow joint with a height of a user to facilitate draping of the robot arm by the user. Accordingly, the controller may further be configured to store data indicative of the height of the user in a user-specific profile, and execute the user-specific profile responsive to user input received at a user interface operatively coupled to the controller. Additionally or alternatively, the co-manipulation surgical system further may comprise an optical sensor operatively coupled to the controller and configured to collect depth data, such that the controller may be configured to determine the height of the user based on depth data collected by the optical sensor.
In some embodiments, the co-manipulation surgical system further may comprise a second robot arm having a second base coupled to the platform, such that, in the drape mode, a plurality of links of the second robot arm may be parallel with the corresponding plurality of links of the robot arm. The co-manipulation surgical system further may comprise a sterile drape having a first portion sized and shaped to cover at least the robot arm and a second portion sized and shaped to cover at least the second robot arm. For example, the first and second portions of the sterile drape may each comprise seal end portions configured to contact distal ends of the respective robot arms, such that the robot arms are fully enclosed within the sterile drape. Moreover, a proximal end of the sterile drape may comprise an opening sized and shaped to receive at least a portion of the platform therethrough. The proximal end of the sterile drape may comprise an elastic band along the opening configured to facilitate wrapping of the sterile drape over the at least the portion of the platform. In addition, the sterile drape may comprise one or more straps configured to secure the first and second portions of the sterile drape to the respective robot arm. Further, the sterile drape may comprise one or more rigid guides configured to be grabbed by a user to guide the sterile drape over the robot arms. The co-manipulation surgical system further may comprise first and second couplers configured to be removably coupled to the distal ends of the respective robot arms after application of the sterile drape over the robot arms, the first and second couplers each configured to be removably coupled to a surgical instrument. Accordingly, the controller may further be configured to apply a temporary localized virtual haptic boundary at the distal regions of the robot arms in the drape mode such that movement of the robot arms are more viscous, thereby stabilizing the robot arms in the drape mode, and remove the temporary localized virtual haptic boundary at the distal regions of the robot arms upon detection that both the first and second couplers are removably coupled to the distal ends of the robot arms.
In accordance with another aspect of the present disclosure, a method for assisting with laparoscopic surgery is provided. The method may comprise: providing a robot arm comprising a proximal region having a base coupled to a platform configured to be moveable on a floor, a plurality of links, an elbow joint, a shoulder joint coupled to the base, and a distal region having a distal end configured to be removably coupled to the surgical instrument; and transitioning the robot arm to a drape mode where a link of the plurality of links between the elbow joint and the shoulder joint extends away from the base at a first angle and a link of the plurality of links distal to the elbow joint extends away from the base at a second angle so as to be more aligned with the floor, to thereby facilitate draping of the robot arm. For example, when the link distal to the elbow joint extends away from the base at the second angle, the link distal to the elbow joint may be substantially horizontal.
Transitioning the robot arm to the drape mode may comprise automatically causing, upon actuation of the drape mode via a controller operatively coupled to the robot arm, the link between the elbow joint and the shoulder joint to extend away from the base at the first angle and the link distal to the elbow joint to extend away from the base at the second angle. Moreover, transitioning the robot arm to the drape mode further may comprise automatically causing, upon actuation of the drape mode via the controller, the robot arm to rotate about the base such that the link between the elbow joint and the shoulder joint extends away from the platform at the first angle and the link distal to the elbow joint extends away from the platform at the second angle. The method further may comprise applying, via the controller, a temporary localized virtual haptic boundary at the distal region of the robot arm in the drape mode such that movement of the robot arm is more viscous, thereby stabilizing the robot arm in the drape mode. Additionally, the method may comprise: applying a sterile drape over the robot arm; removably coupling a coupler to the distal end of the robot arm such that the sterile drape is disposed between the sterile drape and the distal end of the robot arm; and removably coupling the surgical instrument to the coupler. Moreover, the method may comprise: providing a second robot arm having a second base coupled to the platform; and transitioning the second robot arm to the drape mode where a plurality of links of the second robot arm are parallel with the corresponding plurality of links of the robot arm in the drape mode.
Disclosed herein are co-manipulation surgical robot systems for assisting an operator, e.g., a surgeon, in performing a surgical procedure, e.g., a laparoscopic procedure, and methods of use thereof. Currently, laparoscopic procedures typically require a surgeon and one or more assistants. For example, as shown in
As shown in
The co-manipulation surgical robot systems described herein provide superior control and stability such that the surgeon and/or assistant may seamlessly position various off-the-shelf surgical instruments as needed, thus avoiding the workflow limitations inherent to both human and mechanical solutions. For example, the robot arms of the co-manipulation surgical robot system may provide surgical assistance by holding a first surgical instrument, e.g., a laparoscope, via a first robot arm, and a second surgical instrument, e.g., a retractor, via a second robot arm, stable throughout the procedure to provide an optimum view of the surgical site and reduce the variability of force applied by the surgical instruments to the body wall at the trocar point. As will be understood by a person having ordinary skill in the art, the robots arms of the co-manipulation surgical robot systems described herein may hold any surgical instrument, preferably having a long and thin instrument shaft, used for surgical procedures such as laparoscopic procedures including, e.g., endoscopes/laparoscopes, retractors, graspers, surgical scissors, needle holders, needle drivers, clamps, suturing instruments, cautery tools, staplers, clip appliers, hooks, etc.
The co-manipulation surgical robot system further allows the surgeon to easily maneuver both tools when necessary, providing superior control and stability over the procedure and overall safety. Any implementations of the systems described herein enable a surgeon to directly co-manipulate instruments while remaining sterile at the patient bedside. For example, the system may include two robot arms that may be used by the surgeon to hold both a laparoscope and a retractor. During a surgical procedure, the system may seamlessly reposition either instrument to provide optimal visualization and exposure of the surgical field. Both instruments may be directly coupled to the robot arms of the system and the system may constantly monitor and record the position of the two instruments and/or the two robot arms throughout the procedure. Moreover, the system may record information such as the position and orientation of surgical instruments attached to the robot arms, sensor readings related to force(s) applied at proximal and distal ends of the surgical instruments attached to robot arms, force required to hold each instrument in position, endoscopic video streams, algorithm parameters, operating room 3D stream captured with an optical scanning device, including, e.g., position(s) of surgical entry port(s), position and movements of the surgeon's hands, surgical instrument(s) position and orientation, whether or not attached to robot arms, patient position, and patient table orientation and height.
Such data may be used to develop a database of historical data that may be used to develop the algorithms used in some implementations to control one or more aspects of an operation of the system. In addition, such data may be used during a procedure to control of one or more aspects of an operation of the system per one or more algorithms of the system. For example, the data may be used to assess a level of fatigue of a user of the system as described in U.S. Pat. No. 11,504,197, the entire contents of which is incorporated herein by reference.
As the operator manipulates a robot arm of the co-manipulation surgical robot system by applying movement to the surgical instrument coupled to the robot arm, the system may automatically transition the robot arm between various operational modes upon determination of predefined conditions. For example, the system may transition the robot arm to a passive mode responsive to determining that movement of the robot arm due to movement at the handle of the surgical instrument is less than a predetermined amount for at least a predetermined dwell time period, such that in the passive mode, the robot arm maintains a static position, e.g., to prevent damage to the equipment and/or injury to the patient. Additionally, the system may transition the robot arm to a co-manipulation mode responsive to determining that force applied at the robot arm due to force applied at the handle of the surgical instrument exceeds a predetermined threshold, such that in the co-manipulation mode, the robot arm is permitted to be freely moveable responsive to movement at the handle of the surgical instrument for performing laparoscopic surgery using the surgical instrument, while a first impedance is applied to the robot arm in the co-manipulation mode to account for weight of the surgical instrument and the robot arm. Moreover, the system may transition the robot arm to a haptic mode responsive to determining that at least a portion of the robot arm is outside a predefined haptic barrier, such that in the haptic mode, a second impedance greater than the first impedance is applied to the robot arm, thereby making movement of the robot arm responsive to movement at the handle of the surgical instrument more viscous in the haptic mode than in the co-manipulation mode. The system further may transition the robot arm to a robotic assist mode responsive to detecting various conditions that warrant automated movement of the robot arm to guide the surgical instrument attached thereto, e.g., along a planned trajectory or to avoid a collision with another object or person in the surgical space. For example, in an instrument centering mode of the robotic assist mode, a robot arm coupled to a laparoscope may automatically move the laparoscope along a planned trajectory to track an identified surgical instrument and maintain the instrument within the field of view of the laparoscope to provide assisted instrument centering. As described in further detail below, the system further may transition the robot arm to one or more setup modes for manual and/or automatic reconfiguration of the robot arm to an optimized position for a given surgical procedure.
Referring now to
As shown in
In some embodiments, platform 200 may move robot arm 300a and robot arm 300b in the same direction simultaneously, and/or further may cause rotational movement of robot arm 300a and robot arm 300b. For example, the stages may be rotatable relative to platform 200 to thereby cause robot arm 300a and robot arm 300b to rotate towards or away from each other, e.g., along a horizontal plane defined by the x-axis and the y-axis. The center of rotation of each stage may be spaced apart from one another or, alternatively, may be coaxial with one another. Accordingly, in some embodiments, platform 200 may have a smaller footprint, such that robot arms 300a, 300b may be entirely within the footprint of platform 200 in a stow mode, and the stages may rotate and extend base portions 302a, 302b of robot arms 300a, 300b away from each other along the horizontal plane to achieve a spaced apart configuration generally provided by a wider footprint platform, as described in further detail below with regard to
Referring again to
Additionally or alternatively, wheels 104 may be electrically powered such that they may be actuated to electrically engage/disengage the respective braking mechanism. When ready for operation, platform 200 may be moved to a desired position at the side of the patient bed and locked in place via wheels 204, and the vertical and horizontal positions of robot arms 300a and 300b may be adjusted to an optimum position relative to the patient for the procedure via vertical extenders 206a, 206b and horizontal extenders 208a, 208b, responsive to user input received by graphical user interface display 210, and/or via user guided stage control as described in further detail below. As described in further detail below, platform 200 may automatically move robot arm 300a and robot arm 300b responsive to detection of, e.g., potential collisions with other objects and/or persons within the operating room and/or user input applied via the robot arms, during a laparoscopic procedure and/or during setup of the robot arms.
Moreover, system 100 may include a plurality of depth sensors, e.g., proximity sensors 212, disposed on platform 100. Proximity sensors 212 may be, e.g., a depth camera, a stereo RGB camera, a LIDAR device, and/or an electromagnetic, capacitive, ultrasound, or infrared proximity sensor, etc. For example, a first set of proximity sensors 212 may be positioned on robot arm 300a, e.g., at a lower portion of base portion 302a, and a second set of proximity sensors 212 may be positioned on robot arm 300b, e.g., at a lower portion of base portion 302b, to thereby enhance detection of objects approaching the vicinity of robot arms 300a, 300b. For example, as shown in
As base portions 302a, 302b are generally lower than the more distal components of robot arms 300a, 300b, they may be more prone to collision with, e.g., the patient bed, as the stages of platform 200 move robot arms 300a, 300b horizontally and vertically relative to platform 200. Accordingly, the system may generate an alert, e.g., via indicators 334 as described in further detail below, when the proximity sensors detect that the proximity between the robot arms and one or more objects within the operating room falls below a predetermined distance threshold. For example, indicators 334 may illuminate in a predetermined color and/or pattern, e.g., blinking, to indicate proximity with the one or more objects, and the frequency of the blinking may increase as the proximity gets closer. Moreover, the system may cause the stage assembly of platform 200 to stop movement of robot arms relative to platform 200 when the proximity between the robot arms and the one or more objects within the operating room falls below the predetermined distance threshold. In addition, the system further may display, e.g., via GUI 210, an indication that an object is within a predetermined proximity of the robot arm, as determined by forward proximity sensors 212a and/or bottom proximity sensors 212b.
Surgical robot system 100 is configured for co-manipulation, such that system 100 may assist the user or operator, e.g., a surgeon and/or surgical assistant, by permitting the user to freely move robot arm 300a and/or robot arm 300b due to manipulation of one or more surgical instruments coupled with the robot arms in response to force applied by the user to the surgical instruments. Accordingly, system 100 may be configured so that it is not controlled remotely, such that robot arms 300 move directly responsive to movement of the surgical instrument coupled thereto by the operator, while compensating for the mass of the surgical instrument and of the respective robot arm and providing localized impedance along the robot arm, thereby increasing the accuracy of the movements or actions of the operator as the operator manipulates the surgical instrument.
System 100 may be particularly useful in laparoscopic surgical procedures and/or other surgical procedures that utilize long and thin instruments that may be inserted, e.g., via cannulas, into the body of a patient to allow surgical intervention. As will be understood by a person having ordinary skill in the art, system 100 may be used for any desired or suitable surgical operation. Moreover, system 100 may be used in conjunction or cooperation with video monitoring provided by one or more cameras and/or one or more endoscopes so that an operator of system 100 may view and monitor the use of the instrument coupled with robot arms 300a, 300b via respective coupler interfaces 400a, 400b. For example, robot arm 300a may be removeably coupled with and manipulate an endoscope, while robot arm 300b may be may be removeably coupled with and manipulate a surgical instrument.
As shown in
Optical scanners 202, and any other electronics, wiring, or other components of the system, may be supported via platform 200 such that optical scanners 202 are mounted in a fixed location relative to the other objects in the surgical space, and the position and orientation of optical scanners 202 are known or may be determined with respect to the global coordinate system of the system, and accordingly, the robot arms. This allows all data streams to be transformed into a single coordinate system for development purposes. Moreover, telemetry data captured by optical scanners 202, e.g., indicative of the movements of the surgeon's hands, other body parts, the patient bed, the cut-out in a sterile drape over the patient on the surgical bed, the exposed skin through the cut-out in the sterile drape, the trocar(s), the surgical instruments, and other components of the system, may be recorded to provide a rich and detailed dataset describing the precise movements and forces applied by the surgeon throughout the procedure.
As shown in
As shown in
As shown in
The data obtained by the optical scanners may be used to optimize the procedures performed by the system including, e.g., automatic servoing (i.e., moving) of one or more portions of robot arms 300. By tracking the tendency of the surgeon to keep the tools in a particular region of interest and/or the tendency of the surgeon to avoid moving the tools into a particular region of interest, the system may optimize the automatic servoing algorithm to provide more stability in the particular region of interest. In addition, the data obtained may be used to optimize the procedures performed by the system including, e.g., automatic re-centering of the field of view of the optical scanning devices of the system. For example, if the system detects that the surgeon has moved or predicts that the surgeon might move out of the field of view, the system may cause the robot arm supporting the optical scanning device, e.g., a laparoscope, to automatically adjust the laparoscope to track the desired location of the image as the surgeon performs the desired procedure, as described in further detail below. This behavior may be surgeon-specific and may require an understanding of a particular surgeon's preference for an operating region of interest. Additionally or alternatively, this behavior may be procedure-specific. Thus, the system may control the robot arms pursuant to specific operating requirements and/or preferences of a particular surgeon. Moreover, if the system detects that the robot arms are in an extended position for a period of time exceeding a predetermined threshold, the system may cause the stages coupled to the base portions of the robot arms to move the robot arms in a manner to case extension of the robot arms, and thereby provide additional range for extension of the robot arms by the user.
Referring now to
Robot arm 300 further may include shoulder link 305, which includes proximal shoulder link 306 rotatably coupled to distal shoulder link 308. A proximal end of proximal shoulder link 306 may be rotatably coupled to shoulder portion 304 of the base at shoulder joint 318, such that proximal shoulder link 306 may be rotated relative to shoulder portion 304 about axis Q2 at shoulder joint 318. As shown in
In addition, robot arm 300 may include actuator 330, e.g., a collar, lever, button, or switch, operatively coupled to a motor operatively coupled to distal shoulder link 308 and/or proximal shoulder link 306 at joint 320, such that distal shoulder link 308 may only be rotated relative to proximal should link 306 upon actuation of actuator 330. Actuator 330 may be configured to permit dual actuation, e.g., a first actuation to cause distal shoulder link 308 to rotate in a first direction relative to shoulder link 306, and a second actuation to cause distal shoulder link 308 to rotate in a second direction opposite to the first direction. For example, as shown in
As shown in
Accordingly, axis Q3 may be a “setup” axis, such distal shoulder link 308 may be rotated and fixed relative to proximal shoulder link 306 during a setup stage prior to an operating stage where robot arm 300 is used in a surgical procedure, as described in further detail with regard to
Robot arm 300 further may include elbow link 310. A proximal end of elbow link 310 may be rotatably coupled to a distal end of distal shoulder link 308 at elbow joint 322, such that elbow link 310 may be rotated relative to distal shoulder link 308 about axis Q4 at elbow joint 322. Robot arm 300 further may include wrist portion 311, which may include proximal wrist link 312 rotatably coupled to the distal end of elbow link 310 at wrist joint 324, middle wrist link 314 rotatably coupled to proximal wrist link 312 at joint 326, and distal wrist link 316 coupled to/extending from middle wrist link 314, which may be rotatably coupled to surgical instrument coupler interface 400 (not shown) at joint 328, as further shown in
Referring again to
As shown in
Motor M4 may be operatively coupled to setup joint 320 to thereby apply a torque to joint 320 to actuate rotation of distal shoulder link 308 relative to proximal shoulder link 306 about axis Q3. Unlike the other motorized joints described herein, e.g., base joint 303, shoulder joint 318, and elbow joint 322, motorized joint 320 is preferably not “back-drivable,” in that the user cannot actuate motorized joint 320, e.g., via movement of the surgical instrument coupled to the robot arm when the system is in co-manipulation mode. Instead, as described above, actuation of motorized joint 320 may be conducted via one or more actuators, e.g., actuator 330 and/or an actuator displayed on GUI 210, that may be actuated to automatically cause rotation of distal shoulder link 308 relative to proximal shoulder link 306.
Axis Q6 and axis Q7 may each be a “passive” axis, such that middle wrist link 314 may be rotated relative to proximal wrist link 312 at passive joint 326 without any applied impedance from system 100, and surgical instrument coupler interface 400 may be rotated relative to distal wrist link 316 at passive joint 328 without any applied impedance from system 100. The distal end of distal wrist link 316 may be rotatably coupled to surgical instrument coupler interface 400 for removably coupling with a surgical instrument, e.g., via coupler body 500 as shown in
Referring again to
Prior to attachment with a surgical instrument, robot arm 300 may be manually manipulated by a user, e.g., to position robot arm 300 is a desired position for coupling with the surgical instrument. For example, the user may manually manipulate robot arm 300 via wrist portion 311, actuator 330, and/or actuator 332. Upon actuation of actuator 330, the user may automatically rotate distal shoulder link 308, and upon actuation of actuator 332, the user may manually manipulate proximal wrist portion 312. Moreover, robot arm 300 may further be manually moved by application of a force directly on the other links and/or joints of robot arm 300.
In some embodiments, in a user guided setup mode, responsive to force applied to a distal region of robot arm 300, e.g., at any location distal to Q4 such as at wrist portion 311, wrist joint 324, elbow link 310, the surgical instrument, etc., by the user, e.g., exceeding a predetermined force threshold or in a predetermined pattern, in a given direction, e.g., in/out and/or up/down, the processor of the co-manipulation robot platform may cause the stages of platform 200 coupled to base portion 302 of robot arm 300 to move robot arm 300 in the same/corresponding direction, e.g., via vertical extenders 206a, 206b and horizontal extenders 208a, 208b, until the force applied to robot arm 300 by the user is detected by the system to drop below a predetermined threshold, e.g., when the user releases robot arm 300. Due to the lever arm effect, forces applied to robot arm 300 farther from Q4 will be greater than forces applied closer to Q4, and therefore may be preferable during user guided setup mode to move the stages of platform 200. In some embodiments, the system may cause movement of the base of the robot arm via the stage assembly upon application of force at the distal region of the robot arm in the user guided setup mode at a velocity corresponding to the amount of force applied at the distal region of the robot arm. Accordingly, the velocity of movement of the stage assembly may be controlled by adjusting the amount of force applied to the distal region of the robot arm in the user guided setup mode, and further may slow down as the stage assembly reaches or nears its maximum extension range.
As described above, in some embodiments, the processor of the co-manipulation robot platform also may cause the distal shoulder link to rotate relative to the proximal shoulder link responsive to force applied to the distal region of robot arm 300 by the user, e.g., exceeding a predetermined force threshold or in a predetermined pattern, in a given direction, e.g., left/right. In some embodiments, the system may stop movement of robot arm 300 in the same direction as the force applied by the user when the user applies a counter force to robot arm 300, e.g., in a direction opposite to the direction of movement of robot arm 300, to facilitate setup of robot arm 300 relative to the patient. This feature may be initiated/stopped via user actuation, e.g., by actuating actuator 336 on collar 330, voice command, etc. In a preferred embodiment, the system only switches to the user guided setup mode when actuator 336 is in an actuated state, e.g., actively being pressed by a user. Accordingly, the user may actuate actuator 336 with one hand, while simultaneously applying force to the distal region of the robot arm with the other hand to cause movement of the stages of platform 200 while actuator 336 is actuated.
For example, upon actuation of the user guided setup mode, the user may apply a force that exceeds a predetermined force threshold on wrist portion 311 in a first direction, e.g., by applying a pulling or pushing force, which causes the stages of platform 200 to move robot arm 300 in that same direction until the user stops movement of wrist portion 311, e.g., by letting go of robot arm 300 or by applying a counter force to robot arm 300, and/or a maximum extension of the stage assembly is reached, such that the system stops movement of the stages of platform 200. For example, a subsequent pushing force may be counter to an initial pulling force, and a subsequent pulling force may be counter to an initial pushing force. Moreover, the stages of platform 200 may stop moving robot arm 300 when force applied at the distal region of robot arm 300 falls below a predetermined release threshold, which may include letting go of robot arm 300. Accordingly, force applied at the distal region of the robot arm, e.g., wrist portion 311, wrist joint 324, elbow link 310, etc., may serve as an input for motion generated in particular directions of the robot arms via the stages coupled thereto. Such automated movement of the stages of platform 200 responsive to force applied to the distal end of robot arm 300 by the user may be limited to when the system is in a predefined operating mode, e.g., a user guided setup mode, which may be entered in during setup and/or during a surgical procedure, e.g., upon actuation of actuator 336, GUI 210, and/or via voice control.
Similarly, when the user applies a counter force exceeding a predetermined threshold in a predefined direction distinct from the directions that cause horizontal (x-axis) and vertical (z-axis) movement of the stages of platform 200, the system may automatically actuate motorized joint 320 to cause rotation of distal shoulder link 308 relative to proximal shoulder link 306 to facilitate movement of robot arm 300 in the predefined direction. For example, similar to how the system may cause the stages of platform 200 to move robot arm 300 responsive to movement of the distal region of robot arm 300 by the user, e.g., back/forth along the x-axis or up/down along the z-axis, as described above, the system may cause motorized joint 320 to rotate distal shoulder link 308 relative to proximal shoulder link 306 to move robot arm 300 along the y-axis responsive to movement of the distal end of robot arm 300 by the user along the y-axis. Accordingly, the system may stop actuation of motorized joint 320 when the force applied by the user to the distal region of robot arm 300 drops below a predetermined threshold. M4 may be controlled by a processor of the co-manipulation robot platform.
Upon attachment to the surgical instrument, robot arm 300 may still be manipulated manually by the user exerting force, e.g., one or more linear forces and/or one or more torques, directly to robot arm 300; however, during the laparoscopic procedure, the operator preferably manipulates robot arm 300 only via the handle of the surgical instrument, which applies force/torque to the distal end of the robot arm 300, and accordingly the links and joints of robot arm 300. As the operator applies a force to the surgical instrument attached to robot arm 300, thereby causing movement of the surgical instrument, robot arm 300 will move responsive to the movement of the surgical instrument to provide the operator the ability to freely move surgical instrument relative to the patient. As described in further detail below, robot arm 300 may apply an impedance to account for weight of the surgical instrument and of robot arm 300 itself, e.g., gravity compensation, as the operator moves the surgical instrument, thereby making it easier for the operator to move the instrument despite gravitational forces and/or inertial forces being exerted on the robot arm and/or the surgical instrument. As will be understood by a person having ordinary skill in the art, robot arm 300 may include less or more articulation joints than is shown in
In addition, each of robot arms 300 further may include indicators 334 for visually indicating the operational mode associated with the respective robot arm in real-time. For example, indicators 334 may be positioned on at least one of shoulder portion 304, shoulder link 305, elbow link 310, platform 200, lighthouse 203, display 210, etc. As shown in
Moreover, indicators 334, 334a, 334b may include lights, e.g., LED lights, that may illuminate in a variety of distinct colors and in distinct patterns, e.g., solid on or blinking. For example, each operational mode of system 100 may be associated with a uniquely colored light, such as red, yellow, blue, green, purple, white, orange, etc., as described in, for example, U.S. Pat. No. 11,504,197, the contents of which are incorporated herein by reference. Accordingly, indicators 334, 334a, 334b may indicate a transition from one operational mode to another operational mode. Additionally or alternatively, transitions from one operational mode to another operational mode may be indicated to a user via haptic feedback, e.g., a vibration delivered to the distal end of the robot arm, and accordingly to the surgical instrument coupled thereto. For example, the distal end of the robot arm may vibrate as the robot arm transitions from co-manipulation mode to static mode to assure the user that the robot arm is in static/passive mode and will remain in position upon release by the user and/or after the system identifies a hold as part of the instrument detection phase of the instrument centering mode described below. Additionally or alternatively, an audible alert may be emitted to indicate to the user when the robot arm transitions from one operational mode to another operational mode.
Referring now to
Moreover, coupler interface 400 may include an extended portion configured to be inserted within link 316. Coupler interface 400 may be rotatably coupled to the distal end of distal wrist link 316 using any suitable fasteners or connectors, e.g., magnets, screws, pins, clamps, welds, adhesive, rivets, and/or any other suitable faster or any combination of the foregoing. In addition, as described in U.S. Patent Appl. Publ. No. 2023/0114137, coupler interface 400 may include a repulsion magnet disposed within protrusion 404. The repulsion magnet is configured to apply a magnetic force to a magnet slidably disposed within coupler body 500 to facilitate determination of when coupler body 500 is coupled to coupler interface 400 and no surgical instrument is coupled to coupler body 500, e.g., by causing the magnet to move to a position within coupler body 500 with a maximum distance from coupler interface 400, and/or to facilitate coupling of the surgical instrument to coupler body 500, as described in further detail below. Moreover, as described above, robot arm 300 may include one or more encoders E7 for measuring angulation of between distal wrist link 316 and surgical instrument coupler interface 400 may be disposed on or adjacent to joint 328, e.g., within link 316. For example, encoders E7 may include two or more encoders positioned circumferentially around the extended portion of coupler interface 400.
Referring now to 7A to 7C, coupler body 500 is provided. Coupler body 500 may be configured to be removably coupled to a surgical instrument having a predefined shaft diameter, e.g., a 10 mm surgical instrument. Coupler body 500 is preferably designed to be locked to the distal end of the robot arm with a sterile drape therebetween such that the robot arm remains covered and sterile throughout a procedure. Further, coupler body 500 also has a separate portion for locking to a surgical instrument (e.g., a commercially available laparoscopic instrument) to permit the clinician to perform the surgeries with the robot arm(s) as described herein. As shown in
Additionally, coupler interface connection portion 504 may include a pair of locking arms 506 configured to facilitate securing of coupler body 500 to coupler interface 400 when protrusion 404 is disposed within groove 505. Each of locking arms 506 may include handle portion 510 sized and shaped to be actuated by the user's fingers, and connection portion 508 sized and shaped to engage with locking portions 406 of protrusion 404. For example, connection portion 508 may have a tapered profile for securely engaging with locking portion 406. Locking arms 506 may be pivotally coupled to coupler interface connection portion 504, such that locking arms 506 may be transitionable between an unlocked state and a locked state. Moreover, locking arms 506 may be pivotally coupled to coupler interface connection portion 504 via a spring, e.g., a torsion spring, an extension spring, a compression spring, etc., such that locking arms 506 are biased toward the locked state. Accordingly, handle 510 may be actuated to transition locking arms 506 from the locked state to the unlocked state.
Accordingly, prior to coupling coupler body 500 to coupler interface 400, a sterile drape may be positioned between coupler body 500 and coupler interface 400, such that the sterile drape may be draped over robot arm 300, as described above. Moreover, an elastic band of the sterile drape may be hooked onto a hook disposed on lighthouse 203 to secure the drape over lighthouse 203. The sterile drape may be marked and secured with, e.g., peel-off labels, to facilitate efficient application of the drape. The user may then apply a force to handle portions 510 of locking arms 506, e.g., pinch handle portions 510 toward each other, to thereby cause connection portions 508 to move away from each other towards the unlocked state and out of groove 505, and provide clearance for protrusion 404 to be received within groove 505. When locking arms 506 are in their unlocked state, coupler body 500 may be coupled to coupler interface 400 such that protrusion 404 is disposed within groove 505. Once protrusion 404 is disposed within groove 505, the user may release handle portions 510, such that locking arms 506 move back towards their locked state and connection portion 508 engages with locking portion 406 of protrusion 406. Accordingly, the engagement of connection portion 508 and locking portion 406 due to the corresponding geometries of connection portion 508 and locking portion 406 may prevent movement between coupler body 500 and coupler interface 400, to thereby securely couple coupler body 500 to coupler interface 400.
As shown in
In addition, surgical instrument connection portion 502 may include clamp 518 pivotally coupled to surgical instrument connection portion 502 about axis 512, such that clamp 518 may be transitionable between an unlocked state and a locked state. Moreover, clamp 518 may be pivotally coupled to surgical instrument connection portion 502 via a torsion spring, such that clamp 518 is biased toward the locked state. Clamp 518 may include locking portion 520 configured to secure the surgical instrument within opening 516 when clamp 518 is in its locked state. For example, a lower surface of locking portion 520 may define the upper surface of opening 516 when clamp 518 is in its locked state, such that locking portion 520 prevents upward movement of the surgical instrument when the surgical instrument is positioned within opening 516 and clamp 518 is in its locked state.
The upper surface of locking portion 520 may be tapered to facilitate guidance of the surgical instrument into opening 516 along with tapered portions 514. Accordingly, the tapered angle of locking portion 520 may be alone sufficient to permit a surgical instrument to be inserted into opening 516, such that insertion of the surgical instrument towards opening 516 applies a force against the tapered upper surface of locking portion 520, thereby causing clamp 518 to rotate about axis 512 from the locked state to the unlocked state to permit the surgical instrument to be received by opening 516. Clamp 518 further may include handle 522 sized and shaped to be actuated by the user's fingers to transition clamp 518 from the locked state to the unlocked state. For example, handle 522 may be actuated to transition clamp 518 to the unlocked state for insertion of the surgical instrument into opening 516, and/or for removal of the surgical instrument from opening 516.
Moreover, coupler body 500 further may include switch 524 pivotally coupled to surgical instrument connection portion 502, and configured to facilitate securement of the surgical instrument within opening 516. For example, switch 524 may include one or more surgical instrument engagement portions 526, each having a geometry that corresponds with the outer diameter of the shaft of the surgical instrument to be inserted within opening 516. In addition, switch 524 may include handle portion 528 sized and shaped to be actuated by the user's fingers to transition switch 524 between an unlocked state and a locked state where surgical instrument engagement portion 526 engages with the surgical instrument shaft within opening 516 and applies a friction force to the surgical instrument shaft.
Moreover, in its locked state, surgical instrument engagement portion 526 further defines opening 516. Surgical instrument engagement portion 526 may have a coefficient of friction, such that when the surgical instrument is disposed within opening 516 and switch 524 is in its locked state, surgical instrument engagement portion 526 applies a friction force against the surgical instrument that prevents longitudinal movement of the surgical instrument relative to coupler body 500, while permitting rotational movement of the surgical instrument within opening 516. For example, the friction force applied to shaft 10a by surgical instrument engagement portion 526 facilitates securement of shaft 10a within coupler body 500, such that longitudinal movement of surgical instrument 10 is prevented unless the longitudinal force applied to surgical instrument 10 exceeds at least the friction force applied to shaft 10a by surgical instrument engagement portion 526, while the rotational force required to overcome the friction force and cause rotational of shaft 10a within opening 516 is minimized. Accordingly, when the surgical instrument is disposed within opening 516, switch 524 may be actuated to its unlocked state to permit the user to readjust/move the surgical instrument longitudinally relative to coupler body 500 within opening 516, and back to its locked state to prevent longitudinal movement of the surgical instrument relative to coupler body 500. Preferably, both switch 524 and clamp 518 must be in their unlocked states to permit removal of the surgical instrument from coupler body 500.
Alternatively, the coupler interface and the coupler body may be constructed as described in U.S. Patent Appl. Publ. No. 2023/0114137, as shown in
Moreover, protrusion 604 may include one or more locking portions 606 disposed on the outer surface of the sidewall of protrusion 604. For example, locking portions 606 may be indentations/grooves extending along the outer surface of protrusion 604, and sized and shaped to engage with locking arms 660 of connection portion 650, as described in further detail below, for securing the coupler body to coupler interface 600, and for securing the sterile drape between connection portion 650 and coupler interface 600. Preferably, protrusion 604 includes a pair of locking portions 606. For example, as shown in
As shown in
As shown in
As shown in
Magnet 540 may have a magnetic force such that when coupler body 500 is coupled to coupler interface 400, magnet 540 induces a magnetic field, which may be detected by one or more magnetic field sensors, e.g., disposed within link 316 and/or coupler interface 400. Accordingly, the strength of the induced magnetic field will be proportional to the distance between magnet 540 and coupler interface 400 such that the magnetic field detected by the magnetic field sensors may be indicative of the position of magnet 540, and accordingly holder 530, within coupler body 500. Similarly, when no magnetic field is induced via magnet 540, the magnetic field sensors may detect that coupler body 500 is not coupled to coupler interface 400. Moreover, the repulsion magnet of coupler interface 400 may have a magnetic force such that when coupler body 500 is coupled to coupler interface 400, the repulsion magnet applies a magnetic force to magnet 540 to thereby cause magnet 540, and accordingly holder 530, to move away from coupler interface connection portion 504. The position of holder 530 relative to coupler body 500 may be indicative of whether a surgical instrument is or is not coupled to coupler body 500 when coupler body 500 is coupled to coupler interface 400. For example, as shown in
As shown in
As shown in
Moreover, the position of magnet 540 within channel 503 will depend on the diameter size of the surgical instrument disposed within opening 516 when coupler body 500 is coupled to coupler interface 400, such that the induced magnetic field will vary based on the surgical instrument shaft size disposed within opening 516. Accordingly, the system may identify the precise size of the surgical instrument shaft based on the strength of the magnetic field induced by magnet 540, as detected by the magnetic field sensors. Based on the identified type of surgical instrument coupled to coupler body 500, the system may load the calibration file associated with the identified surgical instrument as described above. Moreover, based on the identified make of the surgical instrument, provided that each specific make has a distinguishable shaft diameter size, the system may determine whether the attached surgical instrument is authorized for use with the system.
Referring now to
Referring now to
In some embodiments, upon actuation of the drape mode, the system may automatically cause the stages of platform 200 to move base portions 302a, 302b, e.g., in the vertical direction relative to the ground, to thereby align wrist portions 311a, 311b and elbow links 310a, 310b with the height of the user to facilitate applying the sterile drape over robot arms 300a, 300b. The height of the user may be automatically determined by the system based on depth data collected by the one or more optical scanners. Additionally or alternatively, the height of the user and/or the preferred height of wrist portions 311a, 311b and elbow links 310a, 310b of the user for draping in drape mode may be stored in a memory of the system in a user-specific profile that may be uploaded for execution by the system responsive to user input at GUI 210. System 100 may store the predetermined drape-ready configuration, e.g., within the user-specific profile, such that upon actuation of the system in the “drape mode” during setup by the user, system 100 may cause robot arms 300 to automatically move to the predetermined drape-ready configuration.
Moreover, in the extended position in drape mode, the robot arms may be within the virtual haptic boundary, such that the robot arms are in the haptic mode and a high level of impedance is applied to the robot arms thereby making movement of the robot arms more viscous, which makes it easier for the operator to drape the robot arms, yet provide movement thereto if necessary. In addition, if the robot arms are not within the predefined virtual haptic boundary around the workspace in the drape-ready configuration, the system may apply a temporary localized virtual haptic boundary at least the distal end of the robot arms, as described in further detail below. The system may remove the temporary localized virtual haptic boundary upon detection that the coupler bodies are coupled to the distal ends of the respective robot arms. For example, the system may remove the temporary localized virtual haptic boundary from both robot arms upon determination that both coupler bodies have been coupled to both robot arms. Alternatively, the system may remove the temporary localized virtual haptic boundary from each robot arm upon determining that a coupler body has been coupled to the respective robot arm.
As shown in
As shown in
As shown in
Next, standing on the outer side of a robot arm, e.g., robot arm 300b, the user may pull drape portion 801b over elbow link 310b and shoulder link 305b via guide 804b until guide 804b is positioned at least partially over shoulder portion 304b. Preferably, the user pulls drape portion 801b over robot arm 300b while holding onto guide 804b from inside folded sterile drape 800 to maintain sterility. Then, standing on the outer side of the other robot arm, e.g., robot arm 300a, the user may pull drape portion 801a over elbow link 310a and shoulder link 305a via guide 804a until guide 804a is positioned at least partially over shoulder portion 304b, preferably while holding onto guide 804a from inside folded sterile drape 800 to maintain sterility. Guides 804a, 804b may rest on base portions 302a, 302b of robot arms 300a, 300b, when sterile drape 800 is completely draped over the robot arms. The user may then wrap the proximal portion of sterile drape 800 around base portions 302a, 302b and the stages of platform 200 via clastic band 806, and hook elastic band 806 to hook 209 of lighthouse 203. As shown in
Referring now to
As described above, M4 must be actuated, e.g., via actuator 330, to automatically rotate distal shoulder link 308 relative to proximal shoulder link 306 at joint 320. As shown in
As described in further detail below, system 100 may store one or more predetermined robot arm configurations including a predetermined degree of rotation of distal shoulder link 308 relative to proximal shoulder link 306 for one or more known surgical procedures, such that upon actuation of the system to an “operation-ready mode” during setup, system 100 may cause robot arms 300 to automatically move to the predetermined robot arm configuration. Moreover, as the robot arm is moved, either manually by the user or automatically during setup, based on depth data obtained from the one or more optical scanners, the system may detect when either the stages of platform 200 or the robot arm approaches a predetermined distance threshold relative to an object in the operating room, e.g., the surgical bed. Accordingly, the system may automatically reconfigure the robot arm to avoid a collision with the object, e.g., by automatically actuating motorized joint 320 to rotate distal shoulder link 308 relative to proximal shoulder link 306. Similarly, system 100 may automatically reconfigure the robot arm to avoid a collision with an object in the operating room by automatically actuating motorized joint 320 during a surgical procedure.
For example, the system may measure and record any of the following within the coordinate space of the system: motion of the handheld surgical instruments manipulated by the surgeon (attached to or apart from a robot arm); the presence/absence of other surgical staff (e.g., scrub nurse, circulating nurse, anesthesiologist, etc.); the height and angular orientation of the surgical table; patient position and volume on the surgical table; presence/absence of the drape on the patient; presence/absence of trocar ports, and if present, their position and orientation; gestures made by the surgical staff; tasks being performed by the surgical staff; interaction of the surgical staff with the system; surgical instrument identification; attachment or detachment “action” of surgical instruments to the system; position and orientation tracking of specific features of the surgical instruments relative to the system (e.g., camera head, coupler, fiducial marker(s), etc.); measurement of motion profiles or specific features in the scene that allow for the phase of the surgery to be identified; position, orientation, identity, and/or movement of any other instruments, features, and/or components of the system or being used by the surgical team.
The system may combine measurements and/or other data described above with any other telemetry data from the system and/or video data from the laparoscope to provide a comprehensive dataset with which to improve the overall usability, functionality, and safety of the co-manipulation robot-assisted surgical systems described herein. For example, as the system is being setup to start a procedure, optical scanner 202 may detect the height and orientation of the surgical table. This information may allow the system to automatically configure the degrees of freedom of platform 200 supporting robot arms 300 to the desired or correct positions relative to the surgical table. Specifically, optical scanner 202 may be used to ensure that the height of platform 200 is optimally positioned to ensure that robot arms 300 overlap with the intended surgical workspace. In addition, as described above, the system may automatically reconfigure the degrees of freedom of platform 200 as well as the arrangement of robot arms 300 responsive to movement of the surgical table, and accordingly the trocar(s), to maintain relative position between the distal end of the robot arms and the trocar(s).
In addition, optical scanner 202 may identify the specific surgeon carrying out the procedure, such that the system may use the surgeon's identity to load a system profile associated with the particular surgeon into the system. The system profile may include information related to a surgeon's operating parameter and/or preferences, a surgeon's patient list having parameters for each patient, the desired or required algorithm sensitivity for the surgeon, the degree of freedom positioning of the support platform, etc. Examples of algorithm sensitivities that may be surgeon-specific include: adapting/adjusting the force required to transition from passive mode to co-manipulation mode (e.g., from low force to high force), adapting/adjusting the viscosity felt by the surgeon when co-manipulating the robot arm (e.g., from low viscosity to high viscosity), preferred surgical instrument trajectories when performing specific laparoscopic procedures, etc. Moreover, the surgeon's preferences may include preferred arrangements of robot arm 300, e.g., the positioning of the links and joints of robot arm 300 relative to the patient, with regard to specific surgical instruments, e.g., the preferred arrangement may be different between a laparoscope and a retractor.
Based on the data captured by optical scanner 202, the system may generate a virtual model of the pieces of capital equipment and/or other objects in an operating room that are within a range of movement of the robot arms in the same co-ordinate space as the robot arms and surgical instruments coupled thereto, such that the virtual model may be stored and monitor, e.g., to detect potential collisions. Additionally, the system may track the position and orientation of each virtual model, and the objects within the virtual models as the objects move relative to each other, such that the system may alert the user if the proximity of (i.e., spacing between) any of the virtual models or objects falls below a predefined threshold, e.g., within 50 mm, 75 mm, from 30 mm or less to 100 mm, or more. The system may use this information to recommend a repositioning of platform 200 and/or other components of the system, the surgical table, and/or patient, and/or prevent the robot arm from switching to the co-manipulation mode as a result of the force applied to the robot arm by the collision with the staff member, even if the force exceeds the predetermined force threshold of the robot arm. Moreover, the system may stop or inhibit (e.g., prevent) further movement of a robot arm, e.g., freeze the robot arm, if the proximity of any of the virtual models or objects, e.g., a robot arm reaches or falls below the predefined threshold relative to another objects within the surgical space.
Moreover, based on the data captured by optical scanner 202, the system may track the motion of the handheld surgical instruments that are directly and independently controlled by the surgeon, that are not coupled with the robot arm. For example, the optical scanner 202 may track a clearly defined feature of the instrument, a fiducial marker attached to the instrument or to the gloves (e.g., the sterile gloves) of the surgeon, the coupler between the robot arm and the instrument, a distal tip of the instrument, and/or any other defined location on the instrument. The following are examples of uses and purposes of the motion data: (i) closing a control loop between a handheld instrument and the robot arm holding the camera, thus allowing the surgeon to servo (i.e., move) the camera by “pointing” with a handheld instrument; (ii) tracking information that may be used independently or in combination with other data streams to identify the phase of the surgical procedure; (iii) to identify the dominant hand of the surgeon; (iv) to monitor metrics associated with the experience of the surgeon; (v) to identify which tools the surgeon is using and when to change them for other tools; and/or (vi) tracking of the skin surface of the patient, as well as the number, position and orientation of the trocar ports. This data and information also may be used and computed by the system as part of the co-manipulation control paradigm. As will be understood by a person having ordinary skill in the art, the location/movement of a surgical instrument coupled to a robot arm of the system will be known by the system based on the known robot telemetry and current kinematics of the robot arm, without the need of data captured by optical scanner 202.
Based on the data captured by optical scanner 202, the system further may track the which instrument is being used in a respective port, how often instruments are swapped between ports, which ports have manually held instruments versus instruments coupled to the robot arm, to monitor and determine if additional trocar ports are added, if the system is holding the instruments in place while the patient or surgical table is moving (in which case, the system may change the operational mode of the robot arms to a passive mode and accommodate the movement by repositioning robot arm 300 and/or platform 200), and/or other conditions or parameters of the operating room or the system. The knowledge of the position and orientation of the skin surface and trocar ports relative to the robot arms may facilitate the implementation of “virtual boundaries” as described in further detail below.
Referring now to
As each feature is registered, its position and orientation may be assigned a local co-ordinate system and transformed into the global co-ordinate system the system using standard transformation matrices. Once all features are transformed into a single global co-ordinate system, an optimization algorithm, e.g., least squares and gradient descent, may be used to identify the most appropriate vertical and horizontal positions of robot arms 300a, 300b, which may be adjusted via platform 200, to maximize the workspace of the robot arms with respect to the insertion point on the patient. The optimal workspace may be dependent on the surgical operation to be performed and/or the surgeon's preferred position. Moreover, the system may generate and a display a virtual map, e.g., via GUI 210, graphically depicting the identified features within the operating room based on the depth and proximity data to guide the user when moving platform 200, as described in further detail below with regard to
Referring again to
Referring now to
Platform 1400 may contain memory and/or be coupled, via one or more buses, to read information from, or write information to, memory. Memory 1410 may include processor cache, including a multi-level hierarchical cache in which different levels have different capacities and access speeds. The memory also may include random access memory (RAM), other volatile storage devices, or non-volatile storage devices. Memory 1410 may be RAM, ROM, Flash, other volatile storage devices or non-volatile storage devices, or other known memory, or some combination thereof, and preferably includes storage in which data may be selectively saved. For example, the storage devices can include, for example, hard drives, optical discs, flash memory, and Zip drives. Programmable instructions may be stored on memory 1410 to execute algorithms for, e.g., calculating desired forces to be applied along robot arm 300 and/or the surgical instrument coupled thereto and applying impedances at respective joints of robot arm 300 to effect the desired forces.
Platform 1400 may incorporate processor 1402, which may consist of one or more processors and may be a general purpose processor, a digital signal processor (DSP), an application specific integrated circuit (ASIC), a field programmable gate array (FPGA) or other programmable logic device, discrete gate or transistor logic, discrete hardware components, or any suitable combination thereof designed to perform the functions described herein. Platform 1400 also may be implemented as a combination of computing devices, e.g., a combination of a DSP and a microprocessor, a plurality of microprocessors, one or more microprocessors in conjunction with a DSP core, or any other such configuration. Platform 1400, in conjunction with firmware/software stored in the memory may execute an operating system (e.g., operating system 1446), such as, for example, Windows, Mac OS, QNX, Unix or Solaris 5.10. Platform 1400 also executes software applications stored in the memory. For example, the software may be programs in any suitable programming language known to those skilled in the art, including, for example, C++, PHP, or Java.
Communication circuitry 1404 may include circuitry that allows platform 1400 to communicate with an image capture devices such as optical scanner and/or endoscope. Communication circuitry 1404 may be configured for wired and/or wireless communication over a network such as the Internet, a telephone network, a Bluetooth network, and/or a WiFi network using techniques known in the art. Communication circuitry 1404 may be a communication chip known in the art such as a Bluetooth chip and/or a WiFi chip. Communication circuitry 1404 permits platform 1400 to transfer information, such as force measurements on the body wall at the trocar insertion point locally and/or to a remote location such as a server.
Power supply 1406 may supply alternating current or direct current. Power supply 1406 may be a port to allow platform 1400 to be plugged into a conventional wall socket, e.g., via a cord with an AC to DC power converter and/or a USB port, for powering components within platform 1400. Power supply 1406 may be operatively coupled to an emergency switch, such that upon actuation of the emergency switch, power stops being supplied to the components within platform 1400 including, for example, the braking mechanism disposed on at least some joints of the plurality of joints of robot arm 300. For example, the braking mechanisms may require power to disengage, such that without power supplied to the braking mechanisms, the braking mechanisms engage to prevent movement of robot arm 300 without power. In direct current embodiments, power supply may include a suitable battery such as a replaceable battery or rechargeable battery and apparatus may include circuitry for charging the rechargeable battery, and a detachable power cord. For example, the battery may be an uninterruptable power supply (UPS) that may be charged when the system is plugged in, and which is only operatively coupled to certain computing components of the system, e.g., processor 1402, such that the battery may automatically provide power to the computing components when the system is temporarily unplugged from the electrical power source, e.g., to move the system to another side of a patient table during a multi-quadrant procedure. In some embodiments, the braking mechanism of wheels 204 of platform 200 also may be operatively coupled to the battery, such that they may be engaged/disengaged while the system is unplugged and moved around the operating room.
User interface 1408 may be used to receive inputs from, and/or provide outputs to, a user. For example, user interface 1408 may include a touchscreen display (e.g., GUI 210), switches, dials, lights, etc. Accordingly, user interface 1408 may display information such as selected surgical instrument identity and force measurements observed during operation of robot arm 300. Moreover, user interface 1408 may receive user input including adjustments to the predetermined amount of movement at the handle of the surgical instrument or the predetermined dwell time period to cause the robot arm to automatically switch to the passive mode, the predetermined threshold of force applied at the handle of the surgical instrument to cause the robot arm to automatically switch to the co-manipulation mode, a position of the predefined haptic barrier, an identity of the surgical instrument coupled to the distal end of the robot arm, a vertical height of the robot arm, a horizontal position of the robot arm, etc., such that platform 1400 may adjust the information/parameters accordingly. In some embodiments, user interface 1408 is not present on platform 1400, but is instead provided on a remote, external computing device communicatively connected to platform 1400 via communication circuitry 1404.
Memory 1410, which is one example of a non-transitory computer-readable medium, may be used to store operating system (OS) 1446, surgical instrument identification module 1412, surgical instrument calibration module 1414, encoder interface module 1416, robot arm position determination module 1418, trocar position detection module 1420, force detection module 1422, impedance calculation module 1424, motor interface module 1426, optical scanner interface module 1428, gesture detection module 1430, passive mode determination module 1432, co-manipulation mode determination module 1434, haptic mode determination module 1436, robotic assist mode determination module 1438, trajectory generation module 1440, fault detection module 1442, and indicator interface module 1444. The modules are provided in the form of computer-executable instructions/algorithms that may be executed by processor 1402 for performing various operations in accordance with the disclosure.
For example, during a procedure, the system may continuously run the algorithms described herein based on the data collected by the system. That data may be collected and/or recorded using any of the components and methods disclosed herein, including, e.g., from sensors/encoders within the robots, from optical scanning devices in communication with the other components of the robotic system, and/or from manual inputs by an operator of the system. Accordingly, the algorithms, the data, and the configuration of the system may enable the user to co-manipulate the robot arms with minimal impact and influence from the weight of the robot arms and/or surgical instruments coupled thereto, force of gravity, and other forces that traditional robot arms fail to compensate for. Some of the parameters of the algorithms described herein may control an aspect of the behavior of the system including, e.g., robustness of detected features, sensitivity to false positives, robot control gains, number of features to track, dead zone radius, etc.
Surgical instrument identification module 1412 may be executed by processor 1402 for identifying the surgical instrument coupled to each of the robot arms, and loading the appropriate calibration file into the controller system. For example, the calibration file for each surgical instrument may be stored in a database accessible by surgical instrument identification module 1412, and may include information associated with the surgical instrument such as, e.g., instrument type, make, weight, center of mass, length, instrument shaft diameter, etc. Accordingly, when the appropriate calibration file is loaded, and the associated surgical instrument is coupled to robot arm 300, the system will automatically account for the mass of the surgical instrument, e.g., compensate for gravity on the surgical instrument, when the surgical instrument is attached to robot arm 300 based on the data in the calibration file, such that robot arm 300 may hold the surgical instrument in position after the surgical instrument is coupled to the robot arm and the operator lets go of the surgical instrument. For example, surgical instrument identification module 1412 may identify the surgical instrument based on user input via user interface 1408, e.g., the operator may select the surgical instrument from a database of surgical instruments stored in memory 1410.
Moreover, in some embodiments, the system may be configured such that only pre-approved, verified surgical instruments, e.g., of a certain make, are authorized to be used with the system. A list of authorized instruments may be stored within a database in memory 1410 and/or uploaded from a remote database, e.g., a cloud database. Surgical instrument identification module 1412 may determine that a surgical instrument is authorized for use with the system via, e.g., user input by the user via user interface 1408 indicating that the surgical instrument is among a list of pre-approved instruments, the calibration file loaded for the surgical instrument either automatically when the surgical instrument is attached to the robot arm or manually loaded by a user, and/or real-time surgical instrument identification by the system. For example, surgical instrument identification module 1412 may identify the make of the surgical instrument based on image data observed and generated via optical scanners 202 and/or a laparoscope, before or during a procedure. Specifically, surgical instrument identification module 1412 may identify distinctive features of the surgical instrument, e.g., manufacture logo, handle design, instrument packaging, etc., from the image data to determine the type/make of the instrument. For example, many surgical instruments include an identification marker, e.g., brand logo, etched into or otherwise labeled at or near the distal tip of the instrument, and thus, the make of a surgical instrument may be identified via the video feed of a laparoscope received by optical scanner interface module 1428 having the distal tip of the surgical instrument within the field of view of the laparoscope. Additionally or alternatively, the image data obtain by optical scanner 202 may include measurement data associated with the specific instrument, such that surgical instrument identification module 1412 may compare such data with information contained within the database to identify the instrument and load the appropriate calibration file into the controller system.
Moreover, provided that each specific make of a laparoscope may have distinguishable output video feed quality, e.g., x-y pixel count, frame rate, noise signature, codec, etc., surgical instrument identification module 1412 may identify the make of the instrument by comparing the metadata acquired via the output video feed quality with those expected from an authorized laparoscope. Further, provided that each specific make of a surgical instrument may have a distinguishable and precise mass, surgical instrument identification module 1412 may identify the make of the instrument based on the mass of the surgical instrument, e.g., during surgical instrument calibration as described in further detail below. Additionally, provided that each specific make of a surgical instrument may have a distinguishable and precise shaft diameter, surgical instrument identification module 1412 may identify the surgical instrument based on the specific magnetic field strength measured by sensor 414 induced by the displaced magnet within the coupler body due to the diameter of the surgical instrument when the surgical instrument is coupled to the coupler body and the coupler body is coupled to coupler interface 400, as described above. Provided that each specific make of a surgical instrument may have distinguishable and precise impedance properties, the system may send a vibration pulse down the surgical instrument when the surgical instrument is coupled to the robot arm, such that surgical instrument identification module 1412 may identify the make of the surgical instrument based on the response data. Moreover, surgical instrument identification module 1412 may identify the make of the surgical instrument based on other measurable properties such as electrical resistance of the surgical instrument and/or magnetism of the surgical instrument, provided that such properties are distinguishable for each make of the surgical instrument.
In some embodiments, surgical instrument identification module 1412 may automatically identify the surgical instrument coupled with the robotic arm via the coupler body and the coupler interface using, e.g., an RFID transmitter chip and reader or receiver (e.g., placing an RFID sticker or transmitter on the surgical instrument that may transmit information about the surgical instrument to a receiver of the system), an near field communication (“NFC”) device such as a near field magnetic induction communication device, a barcode and scanner or other optical device, a magnet based communication system, reed switches, a Bluetooth transmitter, the weight of the instrument and/or data gathered from the optical scanner and a lookup table, an activation code associated with an authorized surgical instrument, and/or any other features or mechanisms described herein or suitable for identification of the surgical instrument. Surgical instrument identification module 1412 further may confirm that a surgical instrument is authorized by checking for a license and/or a hospital inventory.
In some embodiments, authorized surgical instruments may include indicators such as invisible ink on the tool shaft or handle that may be illuminated and detected via optical sensor 202, e.g., infra-red illumination that may be illuminated/detected via an IR-sensitive sensor of optical scanner 202, a unique reflective marking that may be illuminated and detected at a specific wavelength of light, a unique feature on the tool and/or coupling mechanism, e.g., shape, profile, indent, latching feature, etc., that facilitates a unique kinematic engagement between the tool and the coupling mechanism, a unique feature built into the sterile drape coupled between the coupler body and the coupler interface. In some embodiments, the system may be operatively coupled to a docking station configured to receive the surgical instrument therein, and to record measurements and detect identity indicators of the surgical instrument, to thereby update the calibration file and determine whether the surgical instrument is authorized.
Accordingly, upon coupling of an unauthorized surgical instrument to the robot arm, the system may generate an audible, visual, and/or haptic alert to inform the user of such unauthorized use, such that corrective action may be taken, e.g., replacing the unauthorized tool with an authorized tool. In some embodiments, the system may apply an increased level of viscosity to the robot arms when an unauthorized tool is coupled to the robot arm to inform the user via haptic feedback, and/or prevent motion of the system by engaging the braking mechanisms of the robot arm and applying impedance via the motors of the system. Moreover, some advanced features of the system such as instrument centering may be disabled until an authorized tool is used. In some embodiments, prior to the start of a procedure, upon attachment of an unauthorized tool, the system may lock the robot arm via the braking mechanisms and motors until the unauthorized tool is replaced with an authorized tool.
Moreover, based on the data obtained by optical scanner 202, e.g., tracked movements of the distal end of a laparoscope coupled to robot arm 300, and/or robot telemetry data obtained by system 100, e.g., known positions/movements of robot arm 300 based on the current kinematics of robot arm 300 calculated by system 100, in addition to image data captured by the laparoscope, the system may identify the type of laparoscope coupled to robot arm 300. For example, laparoscopes commonly used during laparoscopic procedures include flat-tipped laparoscopes and angled-tipped laparoscopes, e.g., a laparoscope having a 30 degree angled tip. The system may determine which laparoscope type is currently coupled to robot arm 300 by comparing the image data obtained by optical scanner 202 of a predefined pattern of movement of the laparoscope and/or known kinematic data of robot arm 300 during the predefined pattern of movement of the laparoscope, e.g., moving the distal end of the laparoscope in a circular pattern in a plane perpendicular to the longitudinal axis of the laparoscope, with the image data obtained by the laparoscope as the laparoscope is being moved in the predefined pattern of movement. For example, for a flat-tipped laparoscope, the image data captured by the laparoscope as the distal end of the laparoscope is moved in a circular pattern in the plane perpendicular to the longitudinal axis of the laparoscope should move along a circular planar path, e.g., there will be no change in depth of the field of view of the laparoscope; whereas, for an angled-tipped laparoscope, the image data captured by the laparoscope as the distal end of the laparoscope is moved in a circular pattern in the plane perpendicular to the longitudinal axis of the laparoscope will observe a change of depth of the field of view of the laparoscope.
Surgical instrument calibration module 1414 may be executed by processor 1402 for calibrating a surgical instrument, e.g., a surgical instrument that does not currently have an associated calibration file in the database stored in memory 1410. Accordingly, surgical instrument calibration module 1414 may calculate measurements and specifications of a surgical instrument when it is coupled to robot arm 300 and the system is in calibration mode, as described in further detail below with regard to
Similarly, memory 1410 may include an additional module, e.g., a system calibration module, which may be executed by processor 1402 for calibrating a new robot arm when a current robot arm is replaced, e.g., during a surgical procedure, based on the data obtained by optical scanner 202, with or without utilizing a tracker at the distal end of the new robot arm, to ensure the system is accurately aware of the kinematics of the new robot arm. Specifically, the system may calibrate optical scanner 202 to platform 200, calibrate the new robot arm with respect to the base portion of the new robot arm, and calibrate the new robot arm with respect to platform 200 when the new robot arm is coupled to platform 200. For example, based on the telemetry data obtained by optical scanner 202, the system calibration module may compare the actual real-time movements of the new robot arm as captured by optical scanner 202 to the movements expected based on commands sent to the new robot arm by the system, e.g., to execute a preprogrammed routine intended to move the new robot arm in specific positions, and generate a degree of error indicative of a deviation between the actual real-time movements of the new robot arm and the expected movements of the robot arm based on the preprogrammed routine. Surgical the system calibration module further may execute an optimization algorithm to reduce or eliminate the degree of error between the actual real-time movements and the expected movements, e.g., until the degree of error falls below a predetermined threshold. This calibration process may occur when the system is in a predefined calibration mode, or alternatively, in real-time during a surgical procedure after the new robot arm is coupled to platform 200.
Encoder interface module 1416 may be executed by processor 1402 for receiving and processing angulation measurement data from the plurality of encoders of robot arm 300, e.g., encoders E1-E7, in real time. For example, encoder interface module 1416 may calculate the change in angulation over time of the links of robot arm 300 rotatably coupled to a given joint associated with the encoder. As described above, the system may include redundant encoders at each joint of robot arm 300, to thereby ensure safe operation of robot arm 300. Moreover, additional encoders may be disposed on platform 100 to measure angulation/position of each robot arm relative to platform 200, e.g., the vertical and horizontal position of the robot arms relative to platform 200. Accordingly, an encoder may be disposed on platform 200 to measure movement of the robot arms along the vertical axis of platform 200 and another encoder may be disposed on platform 200 to measure movement of the robot arms along the horizontal axis of platform 200.
Robot arm position determination module 1418 may be executed by processor 1402 for determining the position of robot arm 300 and the surgical instrument attached thereto, if any, in 3D space in real time based on the angulation measurement data generated by encoder interface module 1416. For example, robot arm position determination module 1418 may determine the position of various links and joints of robot arm 300 as well as positions along the surgical instrument coupled to robot arm 300. Based on the position data of robot arm 300 and/or the surgical instrument, robot arm position determination module 1418 may calculate the velocity and/or acceleration of movement of robot arm 300 and the surgical instrument attached thereto in real time. For example, by determining the individual velocities of various joints of robot arm 300, e.g., via the encoder associated with each joint of the various joints, robot arm position determination module 1418 may determine the resultant velocity of the distal end of robot arm 300, which may be used by passive mode determination module 1432 to determine whether movement of the distal end of robot arm 300 is within a predetermined threshold for purposes of transitioning system 100 to passive mode, as described in further detail below.
Trocar position detection module 1420 may be executed by processor 1402 for determining the position and/or orientation of one or more trocar port inserted within the patient. The position and/or orientation of a trocar port may be derived based on data obtained from, e.g., inertial measurement units and/or accelerometers, optical scanners, electromechanical tracking instruments, linear encoders, the sensors and data as described above. For example, the position of the trocar ports on the patient may be determined using a laser pointing system that may be mounted on one or more of the components of the system, e.g., wrist portion 311 of the robot arm, and may be controlled by the system to point to the optimal or determined position on the patient's body to insert the trocar. Moreover, upon insertion of the surgical instrument that is attached to robot arm 300 through a trocar, virtual lines may continuously be established along the longitudinal axis of the surgical instrument, the alignment/orientation of which may be automatically determined upon attachment of the surgical instrument to coupler interface 400 via the coupler body via the magnetic connection as described above, in real time as the surgical instrument moves about the trocar point. Moreover, when the surgical instrument is inserted within the trocar port, it will be pointing toward the trocar point, and accordingly, distal wrist link 316 will also point toward the trocar point, the angle of which may be measured by an encoder associated therewith. Accordingly, the trocar point may be calculated as the intersection of the plurality of virtual lines continuously established along the longitudinal axis of the surgical instrument. In this manner, the calculated trocar point will remained fixed relative to the patient as the surgical instrument is maneuvered about the trocar port, e.g., rotated or moved in or out of the patient. In addition, the orientation of the trocar port and its position relative to robot arm 300 may be determined based on image data received from one or more optical scanners, e.g., a LiDAR camera and/or an RGBD camera. By measuring the true position and orientation of the trocar ports, the system may be provided an additional safety check to ensure that the system level computations are correct, e.g., to ensure that the actual motion of the robot arms or instrument matches a commanded motion of the robot arms or instrument in robotic assist mode.
Based on the known position and/or orientation of a trocar port in addition to the known position of the distal end of robot arm 300 from robot arm position determination module 1418, the system may maintain the position of the distal end of robot arm 300 relative to the trocar point as robot arm 300 moves, e.g., via vertical or horizontal adjustment thereof by platform 200, or as the patient table height is adjusted, thereby causing the height of the patient's abdomen to move, thereby keeping the surgical instrument within the patient's body and coupled to robot arm 300 steady during these external movements. To achieve this, the known position of the distal end of robot arm 300 from robot arm position determination module 1418 is calculated in the global frame of the system by adding position of platform 200 to the kinematics calculations (e.g., the “forward kinematics” of robot arm 300 in the context of serial chain robotic manipulators).
With the position of the distal end of robot arm 300 known globally, the system may hold that position steady by applying appropriate forces to robot arm 300 during the external movements that minimize the error between its current and desired positions. Accordingly, for example, when a surgical instrument coupled to the distal end of robot arm 300 is inserted through a trocar port such that the tip of the instrument is inside of the patient, and a user adjusts the height of the patient table, the system may apply forces/torques to robot arm 300 to reconfigure robot arm 300 and/or cause movement of the stages of platform 200 to maintain the relative position between the distal end of robot arm 300, and accordingly the surgical instrument, and the trocar port. In some embodiments, the system may cause the distal end of robot arm 300 to retract slightly such that the tip of the surgical instrument is positioned within the trocar port and out of contact with anatomical structures within the patient's body prior to reconfiguring robot arm 300 to maintain the relative position between the surgical instrument and the trocar port.
Force detection module 1422 may be executed by processor 1402 for detecting forces applied on robot arm 300, e.g., at the joints or links of robot arm 300 or along the surgical instrument, as well as applied on the trocar, e.g., body wall forces. For example, force detection module 1422 may receive motor current measurements in real time at each motor, e.g., M1, M2, M3, disposed within the base of robot arm 300, which are each operatively coupled to a joint of robot arm 300, e.g., base joint 303, shoulder joint 318, elbow joint 322, wrist joint 332. The motor current measurements are indicative of the amount of force applied to the associated joint. Accordingly, the force applied to each joint of robot arm 300 as well as to the surgical instrument attached thereto may be calculated based on the motor current measurements and the position data generated by robot arm position determination module 1418 and/or trocar position detection module 1420.
Due to the passive axes at the distal end of robot arm 300, the force applied by the instrument coupled with the robot arm on the trocar may remain generally consistent throughout the workspace of the robot arm. The force on the trocar may be affected by the interaction of the distal tip of the instrument with tissue within the body. For example, if a tissue retractor advanced through the trocar is engaged with (e.g., grasping) bodily tissue or another object inside the body, the force exerted on the end of the instrument from the bodily tissue or other object may cause a change in the force applied to the trocar. In some aspects, the force on the trocar may be a function of how much weight is being lifted by the instrument being used.
Impedance calculation module 1424 may be executed by processor 1402 for determining the amount of impedance/torque needed to be applied to respective joints of robot arm 300 to achieve the desired effect, e.g., holding robot arm 300 in a static position in the passive mode, permitting robot arm 300 to move freely while compensating for gravity of robot arm 300 and the surgical instrument attached thereto in the co-manipulation mode, applying increased impedance to robot arm 300 when robot arm 300 and/or the surgical instrument attached thereto is within a predefined virtual haptic barrier in the haptic mode, etc.
For example, impedance calculation module 1424 may determine the amount of force required by robot arm 300 to achieve the desired effect based on position data of robot arm 300 generated by robot arm position determination module 1418 and the position data of the trocar generated by trocar position detection module 1420. For example, by determining the position of the distal end of robot arm 300, as well as the point of entry of the surgical instrument into the patient, e.g., the trocar position, and with knowledge of one or more instrument parameters, e.g., mass and center of mass of the surgical instrument stored by surgical instrument calibration module 1414, impedance calculation module 1424 may calculate the amount of force required to compensate for gravity of the surgical instrument (compensation force), as described in further detail below with regard to
Moreover, by determining the position of the distal end of robot arm 300, and accordingly, a change in position of the distal end of robot arm 300 over time, for example, due to an external force applied to the distal end of robot arm 300, e.g., by tissue held by the operating end of the surgical instrument, and with knowledge of one or more instrument parameters, e.g., mass, center of mass, and length of the surgical instrument stored by surgical instrument calibration module 1414, impedance calculation module 1424 may calculate the amount of force required to maintain the surgical instrument in a static position (hold force), as described in further detail below with regard to
Additionally or alternatively, by determining the forces applied on robot arm 300 via force detection module 1422, as well as the position/velocity/acceleration of the distal end of robot arm 300 in 3D space via robot arm position determination module 1418, the desired force/impedance to be applied to robot arm 300 to compensate for the applied forces may be calculated, e.g., for gravity compensation or to hold robot arm 300 in a static position in the passive mode. Accordingly, the desired force may be converted to torque to be applied at the joints of robot arm 300, e.g., by the motors operatively coupled to the joints of robot arm 300. For example, the robot Jacobian may be used for this purpose.
Motor interface module 1426 may be executed by processor 1402 for receiving motor current readings at each motor, e.g., M1, M2, M3, M4, disposed within the base of robot arm 300, and for actuating the respective motors, e.g., by applying a predetermined impedance to achieved the desired outcome as described herein and/or to cause the joints operatively coupled to the respective motors to move, such as in the robotic assist mode. For example, motor interface module 1426 may actuate M4 to cause rotation of distal shoulder link 308 relative to proximal shoulder link 306.
As described above, the data streams from the robot arms, the camera feed from the laparoscope, the data acquired from optical scanner 202 and/or proximity sensors 212, as well as data optionally captured from one or more imaging devices disposed on a structure adjacent to the robot arms, the walls, ceiling, or other structures within the operating room, may be recorded, stored, and used individually or in combination to understand and control the surgical system and procedures of the surgical system. The foregoing components, devices, and combinations thereof are collectively referred to herein as optical scanners or optical scanning devices.
Optical scanner interface module 1428 may be executed by processor 1402 for receiving depth data obtained by an optical scanning device, e.g., optical scanner 202, and processing the depth data to detect, e.g., predefined conditions therein. Moreover, optical scanner interface module 1428 may generate depth maps indicative of the received depth data, which may be displayed to the operator, e.g., via a monitor. Based on the depth map generated by the optical scanning devices, optical scanner interface module 1428 may cluster different groups of (depth) pixels into unique objects, a process which is referred to as object segmentation. Examples of such algorithms for segmentation may include: matching acquired depth map data to a known template of an object to segment; using a combination of depth and RGB color image to identify and isolate relevant pixels for the object; and/or machine learning algorithms trained on a real or synthetic dataset to objects to identify and segment. Examples of such segmentation on a depth map may include: locating the robot arms or determining the position of the robot arms; identifying patient ports (e.g., trocar ports) in 3D space and determining a distance from the instruments to the trocar ports; determining the relative distances between, e.g., the stages of platform 200, robot arm 300, any surgical instruments attached thereto, and objects/persons in the operating room such as the surgical table, drapes, etc.; identifying the surgeon and distinguishing the surgeon from other operators in the room; and/or identifying the surgeon in the sensor's field of view. Moreover, the system may use object segmentation algorithms to uniquely identify the surgeon and track the surgeon with respect to, for example, a surgical table, a patient, one or more robot arms, etc. In addition, the system may use object segmentation algorithms to determine if a surgeon is touching or handling either of the robot arms and, if so, identify which robot arm is being touched or handled by the surgeon.
Optical scanner interface module 1428 further may use object segmentation algorithms to analyze image data obtained from a laparoscope to locate and track one or more surgical instruments and/or anatomical structures and distinguish the tracked surgical instrument(s) and/or anatomical structure(s) from other objects and structures within the field of view of the laparoscope. For example, the object segmentation algorithms may include deep learning approaches. Specifically, a neural network may be trained for instrument/anatomical structure detection via a manually annotated video dataset sampled from multiple laparoscopic surgeries includes various surgical instruments and anatomical environments. For example, as shown in
Optical scanner interface module 1428 further may receive image data from additional optical scanning devices as defined herein, including for example, an endoscope operatively coupled to the system. Moreover, optical scanner interface module 1428 may receive depth data obtained by proximity sensors 212 coupled to platform 200 and process the depth data to generate a virtual map of the area surrounding platform 200, as described below with regarding to
Gesture detection module 1430 may be executed by processor 1402 for detecting predefined gestural patterns as user input, and executing an action associated with the user input. The predefined gestural patterns may include, for example, movement of a surgical instrument (whether or not attached to robot arm 300), movement of robot arm 300 or other components of the system, e.g., foot pedal, buttons, etc., and/or movement of the operator in a predefined pattern. For example, movement of the surgical instrument back and forth in a first direction (e.g., left/right, up/down, forward/backward, in a circle) may be associated with a first user input requiring a first action by the system and/or back and forth in a second direction (e.g., left/right, up/down, forward/backward, in a circle) that is different than the first direction may be associated with a second user input requiring a second action by the system. Similarly, pressing the foot pedal or a button operatively coupled with the system in a predefined manner may be associated with a third user input requiring a third action by the system, and movement of the operator's head back and forth or up and down repeatedly may be associated with a fourth user input requiring a fourth action by the system. Various predefined gestural patterns associated with different components or operators of the system may be redundant such that the associated user input may be the same for different gestural patterns. The predefined gestural patterns may be detected by, e.g., an optical scanning device such as a laparoscope or optical scanner 202 via optical scanner interface module 1428 or directly by force applied to robot arm 300 via force detection module 1422 or other components of the system.
Actions responsive to user input associated with predefined gestural patterns may include, for example, enabling tool tracking to servo (i.e., move) the laparoscope based on the motion of a handheld tool and/or automatically to maintain the handheld tool within a field of view of the laparoscope; engaging the brakes on (e.g., preventing further movement of) the robot arm; engaging a software lock on the robot arm; dynamically changing the length of time that the robot arm takes to transition between states from a default setting; loading a virtual menu overlay on the video feed whereby a surgical instrument in the field of view of the laparoscope functions as a pointer to trigger further actions available from the virtual menu; start/stop a recording of image data; and/or identifying which member of the surgical staff is touching the robot arm, if any. This information may be used to ensure that the system does not move if the surgeon is not touching the robot arm, e.g., to avoid the scenario where an external force is acting on the robot arm (e.g., a light cable or other wire being pulled across the robot arm) and the system perceives the force to be intentional from the surgeon. The same information may be used to detect the gaze direction of the surgeon, e.g., whether the surgeon is looking at the video feed or somewhere else in the room, such that the system may freeze the robot arm if the surgeon's gaze is not in the direction it should be. Additionally, the system may reposition a field of view of a camera based on, for example, the direction a surgeon is facing or based on the objects that the surgeon appears to be looking at, based on the data from the optical scanner 1100. Moreover, moving the distal tip of a surgical instrument to a center portion of the laparoscopic field of view, e.g., defined by a predetermined boundary region, and holding the position for more than a predetermined time threshold may be associated with a user input detected by gesture detection module 1430 to enable took tracking, as described in further detail below.
Moreover, a predefined gestural pattern such as double-tapping a distal portion of the robot arm and/or a predetermined sinusoidal movement of the camera head of the laparoscope about the trocar may be associated with a user input detected by gesture detection module 1430 to start and/or stop a recording of image/audio data by the optical scanning devices. Specifically, there may be key moments during a procedure that the user may want recorded, and which the user may want to be able to locate in a quick manner without having to go through an entire recording of the entire procedure to find the key moments. By providing an easy way for the user to initiate and stop a recording via simple predefined gestural patterns, such that the recording is saved to a folder with a timestamp associated with that particular procedure, the user may easily locate the recording for review and/or teaching purposes. This feature may be particularly useful for diagnostic procedures. In some embodiments, in response to detection of the predefined gestural pattern by gesture detection module 1430, the system may record and save a predetermined portion of the image data, e.g., ten seconds before and ten seconds after the predefined gestural pattern is detected. Moreover, the select recordings of key moments by the user may be used by the system to indicate key phase segmentation for a given procedure. The user further may generate case notes via the recordings, e.g., by indicating progression through different phases of a procedure when performing a procedure based on a template of the procedure accessible via the system.
As described above, responsive to detection of a predefined gestural pattern by the user, e.g., a predefined pattern of movement of the distal tip of the surgical instrument within the field of view of the laparoscope, gesture detection module 1430 may cause a virtual menu to overlay on the video feed, such that the surgical instrument within the field of view of the laparoscope functions as a pointer, as shown in
In some embodiments, initiation of the display of the virtual menu overlay on the video feed may be triggered by, e.g., actuation of an external actuator such as a foot pedal, a predefined pattern of force applied to the robot arm such double tapping wrist portion 311 and/or the surgical instrument coupled to the robot arm as detected by encoders at the distal end of the robot arm, voice activation, wireless buttons, hot buttons, etc. In some embodiments, the operator may actively switch the system to a command mode, e.g., via user interface 1408, where particular movements or gestures of the robot arm, surgical instrument, operator, or otherwise as described herein are monitored by gesture detection module 1430 to determine if they are consistent with a predefined gestural pattern associated with a predefined user input.
Passive mode determination module 1432 may be executed by processor 1402 for analyzing the operating characteristics of robot arm 300 to determine whether to switch the operational mode of robot arm 300 to the passive mode where the system applies impedance to the joints of robot arm 300 via motor interface module 1426 in an amount sufficient to maintain robot arm 300, and accordingly a surgical instrument attached thereto, if any, in a static position, thereby compensating for mass of robot arm 300 and the surgical instrument, and any other external forces acting of robot arm 300 and/or the surgical instrument. If robot arm 300 is moved slightly while in the passive mode, but not with enough force to switch out of the passive mode, the system may adjust the amount of impedance applied the robot arm 300 to maintain the static position, and continue this process until robot arm 300 is held in a static position. For example, passive mode determination module 1432 may determine to switch the operational mode of robot arm 300 to the passive mode if movement of the robot arm due to movement at the handle of the surgical instrument as determined by force detection module 1422 is less than a predetermined amount, e.g., no more than 1 to 5 mm, for at least a predetermined dwell time period associated with robot arm 300. The predetermined dwell time period refers to the length of time that robot arm 300 and/or the surgical instrument attached thereto, if any, are held in a static position. For example, the predetermined dwell time may range between, e.g., 0.1 to 3 seconds or more, and may be adjusted by the operator.
In some embodiments, passive mode determination module 1432 may determine to switch the operational mode of robot arm 300 to the passive mode if movement of the distal end of the robot arm due to movement at the handle of the surgical instrument as determined by force detection module 1422 has a velocity that is less than a predetermined dwell velocity/speed. For example, if passive mode determination module 1432 determines that the distal end of the robot arm 300 and/or the surgical instrument attached thereto, if any, moves at a speed that is lower than the predetermined dwell speed during an entire predetermined dwell period, then passive mode determination module 1432 may switch the operational mode of robot arm 300 to the passive mode.
Co-manipulation mode determination module 1434 may be executed by processor 1402 for analyzing the operating characteristics of robot arm 300 to determine whether to switch the operational mode of robot arm 300 to the co-manipulation mode where robot arm 300 is permitted to be freely moveable responsive to movement at the handle of the surgical instrument for performing laparoscopic surgery using the surgical instrument, while the system applies an impedance to robot arm 300 via motor interface module 1426 in an amount sufficient to account for mass of the surgical instrument and robot arm 300. Moreover, the impedance applied to robot arm 300 may provide a predetermined level of viscosity perceivable by the operator.
Moreover, the force exerted by the user on the surgical instrument and any external tissue forces applied to the surgical instrument may be directionally dependent. For example, if the force exerted by the user on the surgical instrument is in the same direction as an external tissue force applied to the surgical instrument, the two forces may be additive such that the amount of force exerted by the user on the surgical instrument needed to overcome the predefined force threshold may be reduced by the magnitude of the external tissue force such that a lower force than the predefined force threshold would be required to exit the passive mode and enter the co-manipulation mode. On the other hand, if the force exerted by the user on the surgical instrument is in a direction opposite to an external tissue force applied to the surgical instrument, than the necessary amount of force exerted by the user on the surgical instrument needed to overcome the predefined force threshold may be increased by the magnitude of the external tissue force such that a higher force than the predefined force threshold would be required to exit the passive mode and enter the co-manipulation mode.
In addition, if the force exerted by the user on the surgical instrument is in a direction that is perpendicular to an external tissue force applied to the surgical instrument, than the necessary amount of force exerted by the user on the surgical instrument needed to overcome the predefined force threshold may not be affected by the magnitude of the external tissue force such that the necessary force exerted by the user on the surgical instrument needed to exit the passive mode and enter the co-manipulation mode will equal the predefined force threshold. For other directions, the force vectors of the applied forces may be added to or offset by the force vectors of the external tissue forces to overcome predefined force threshold values for the system or the particular surgical instrument that is coupled with the robot arm, depending on the direction of the external tissue force, if any, and the force applied by the user. In some embodiments, co-manipulation mode determination module 1434 may determine to switch the operational mode of robot arm 300 to the co-manipulation mode based on the identity of the surgical instrument.
Haptic mode determination module 1436 may be executed by processor 1402 for analyzing the operating characteristics of robot arm 300 to determine whether to switch the operational mode of robot arm 300 to the haptic mode where the system applies an impedance to robot arm 300 via motor interface module 1426 in an amount higher than applied in the co-manipulation mode, thereby making movement of robot arm 300 responsive to movement at the handle of the surgical instrument more viscous in the co-manipulation mode. For example, haptic mode determination module 1436 may determine to switch the operational mode of robot arm 300 to the haptic mode if at least a portion of robot arm 300 and/or the surgical instrument attached thereto is within a predefined virtual haptic boundary. Specifically, a virtual haptic boundary may be established by the system, such that the robot arm or the surgical instrument coupled thereto should not breach the boundary. For example, a virtual boundary may be established at the surface of the patient to prevent any portion of the robot arms or the instruments supported by the robot arms from contacting the patient, except through the one or more trocars. Similarly, the virtual haptic boundary may include a haptic funnel to help guide the instrument into the patient as the operator inserts the instrument into a trocar port.
Moreover, a virtual haptic boundary, e.g., haptic shell, may be established at a predetermined distance surrounding the workspace to prevent over-extension of the robot arm away from the operation site, as well as to minimize “runaway” of the robot arm. For example, after an instrument is decoupled from the coupler body coupled to the robot arm, the magnet within the coupler body described above should return its maximum position away from the repulsion magnet within coupler interface 400 to thereby indicate that the surgical instrument has been removed; however, if the magnet does not return to that position, the system may think that the surgical instrument is still attached to the robot arm and continue to compensate for the mass of the surgical instrument, thereby causing the distal end of the robot arm to “runaway,” e.g., drift upward. Accordingly, the virtual haptic boundary may slow down the drifting robot arm to avoid potential collision with other objects or people. For example, the virtual haptic boundary may set at chest-level of a user to prevent the robot arm from colliding with the user's head.
Accordingly, based on position data of robot arm 300 and/or the surgical instrument coupled thereto, e.g., received by robot arm position determination module 1418 and/or trocar position detection module 1420, haptic mode determination module 1436 may determine if robot arm 300 and/or the surgical instrument is within the predefined virtual haptic boundary, and accordingly transition robot arm 300 to the haptic mode where processor 1402 may instruct associated motors to apply an effective amount of impedance to the joints of robot arm 300 perceivable by the operator to communicate to the operator the virtual haptic boundary. Accordingly, the viscosity of robot arm 300 observed by the operator will be much higher than in co-manipulation mode. In some embodiments, haptic mode determination module 1436 may determine to switch the operational mode of robot arm 300 to the haptic mode based on the identity of the surgical instrument.
Moreover, haptic mode determination module 1436 may generate temporary localized virtual haptic boundaries at the distal ends of the robot arms during predetermined phases of a procedure/clinical workflow to prevent “runaway” and further enhance safety of the system. The predetermined phases may include, e.g., during draping/tear-down of the drape, immediately after tool removal is detected, and/or immediately after coupler body removal is detected. For example, increasing viscosity of the robot arms during draping/tear-down may help stabilize the robot arms and prevent excessive movement/runaway thereof, and increasing viscosity during tool/coupler body removal may prevent “runaway” due to a force applied to the robot arm by the user during the removal process. The localized virtual haptic boundary may be temporary in that it may only be applied for a predetermined time period, e.g., a few seconds after the predetermined phase is identified. The predetermined phases of a procedure may be determined/estimated via, e.g., user input via GUI 210 and/or voice command, uploaded from a database stored within the system, and/or via telemetry of the robot arms and identification and/or positions of the surgical instruments. For example, the system may gather and analyze telemetry data regarding forces being applied to the robot arm to assess or estimate whether a user is attempting to remove a tool from the robot arm, and accordingly, haptic mode determination module 1436 may generate a temporary localized virtual haptic boundary at the distal end of the robot arm to facilitate tool removal.
In addition, haptic mode determination module 1436 may adjust the amount of viscosity, e.g., impedance, applied at the distal ends of the robot arms during predetermined phases of a surgical procedure to guide specific movements during the predetermined phase based on the type of surgical instrument coupled to the robot arm, e.g., wristed instruments such as a needle driver or grasper, stapling devices, dissection devices, suturing devices, retraction devices such as a fan retractor, tissue removal devices such as a gallbladder bag, clip applier devices, etc. For example, during a suturing phase of a procedure, viscosity at the distal end of the robot arm may be increased to provide more viscous control to the user during operation of the suture device. Additionally, during a stapling phase of a procedure, viscosity at the distal end of the robot arm may be increased to provide very stiff grounding for force application of the stapling device operated by the user. Accordingly, the increased viscosity may facilitate performance of a specific movement by the user without actively moving the robot arm to perform the specific movement.
Robotic assist mode determination module 1438 may be executed by processor 1402 for analyzing the operating characteristics of robot arm 300 to determine whether to switch the operational mode of robot arm 300 to the robotic assist mode where processor 1402 may instruct associated motors via motor interface module 1426 to cause movement of corresponding link and joints of robot arm 300 to achieve a desired outcome. For example, robotic assist mode determination module 1438 may determine to switch the operational mode of robot arm 300 to the robotic assist mode if a predefined condition exists based on data obtained from, e.g., optical scanner interface module 1428.
For example, robotic assist mode determination module 1438 may determine that a condition exists, e.g., that one or more trocars are not in an optimal position, for example, due to movement of the patient, such that robot arm 300 should be repositioned to maintain the trocar in the optimal position, e.g., in an approximate center of the movement range of robot arm 300, thereby minimizing the risk of reaching a joint limit of the robot arm during a procedure. Thus, in robotic assist mode, processor 1402 may instruct system to reposition robot arm 300, e.g., via vertical/horizontal adjustment by platform 100 or via the joints and links of robot arm 300, to better align the surgical instrument workspace.
Robotic assist mode determination module 1438 may determine that a condition exists, e.g., the distance between an object, e.g., capital equipment or a member of the surgical staff other than the surgeon, and robot arm 300 reaches or falls below a predetermined threshold, based on image data obtained from the laparoscope or optical scanner 202 via optical scanner interface module 1428, such that the robot arm should be frozen to avoid collision with the object. Thus, in robotic assist mode, processor 1402 may instruct robot arm 300 apply the brakes to slow down the robot arm or inhibit or prevent movement within a predetermined distance from the other object to thereby prevent the inadvertent movement of the robot arm that may otherwise result from such a collision or inadvertent force.
Robotic assist mode determination module 1438 further may determine that a condition exists, e.g., robot arm 300 is in an extended position for a period of time exceeding a predetermined threshold during a surgical procedure, such that the robot arm should be repositioned to provide the user more available workspace in the vicinity of the surgical instrument coupled to the extended robot arm. Thus, in robotic assist mode, processor 1402 may instruct the system to reposition robot arm 300, e.g., via vertical/horizontal adjustment by platform 100 and/or via the joints and links of robot arm 300, to move robot arm 300 closer to the surgical instrument.
In addition, robotic assist mode determination module 1438 may determine that a condition exists, e.g., the field of view of a laparoscope coupled to robot arm 300 or optical scanner 202 is not optimal for a given surgical procedure, e.g., due to blocking by the surgeon or assistant or another component of the system, based on image data obtained from the laparoscope or optical scanner 202 via optical scanner interface module 1428, such that the robot arm coupled to the laparoscope or optical scanner 202 should be repositioned or zoom in/out to optimize the field of view of the surgical site for the operator. Thus, in robotic assist mode, processor 1402 may instruct robot arm 300, either automatically/quasi-automatically or responsive to user input by the operator, to move to reposition the laparoscope and/or cause the laparoscope to zoom in or zoom out, or to increase a resolution of an image, or otherwise. For example, the user input by the operator may be determined by gesture detection module 1430, as described above, such that movement of the robot arm or a surgical instrument in a predefined gestural pattern in a first direction causes the endoscope to increase resolution or magnification and in a second direction causes the endoscope to decrease resolution or magnification, and movement in another predefined gestural pattern causes the robot arm holding the laparoscope to retract away from the patient's body.
In some embodiments, robotic assist mode determination module 1438 may determine that a condition exists, e.g., initiation of an “instrument centering” mode by the user and identification of a target surgical instrument such as a handheld tool within the field of view of the laparoscope attached to the robot arm, such that the robot arm should switch to robotic assist mode to provide assisted scope control to center the instrument within the field of view of the laparoscope. For example, as described above, positioning and maintaining the distal tip of a surgical instrument at a center portion of the laparoscopic field of view, e.g., defined by a predetermined boundary region, for more than a predetermined time threshold may be associated with a user input detected by gesture detection module 1430 to enable tool tracking of the surgical instrument, such that robotic assist mode determination module 1438 switches the robot arm attached to the laparoscope to the robotic assist mode to provide automated instrument centering. For example, robotic assist mode determination module 1438 may execute one or more machine learning algorithms on the image data received from the laparoscope to identify the target surgical instrument within the field of view of the laparoscope, e.g., by evaluating pixels of the image data received by the laparoscope and indicating if the pixels correspond to the target surgical instrument, to thereby identify the target surgical instrument. The machine learning algorithms may be trained with a database of annotated image data of associated surgical instruments using trained modules, such as convolution kernels. For example, the machine learning algorithms may pass convolution kernels across the images and attribute a score indicative of the likelihood that a given pixel, or group of pixels, is correctly identified/classified. The score may be derived from a feature map generated by the kernels. By training the machine learning algorithms on more data, the algorithms may be improved by updating the weights in the kernels. As will be understood by a person having ordinary skill in the art, other machine learning algorithms may be used, e.g., pattern matching.
In the robotic assist mode, robotic assist mode determination module 1438 may determine one or more conditions exist, e.g., the target surgical instrument within the field of view of the laparoscope moves out of a predefined boundary region, e.g., a predefined rectangular or circular region about a center point of the field of view, within the field of view indicating that the robot arm should reposition the laparoscope to maintain the target surgical instrument within the predefined boundary region within the field of view of the laparoscope, and/or the resolution of the target surgical instrument in the laparoscope feed falls below a predetermined resolution threshold indicating that the robot arm should move the laparoscope to zoom out, and/or the detected size of the surgical instrument falls below a predetermined size threshold indicating that the robot arm should move the laparoscope to zoom in. In some embodiments, the predefined boundary region may be a selected fraction of the field of view, e.g., the inner two-thirds of the field of view of the laparoscope. Thus, movement of the target surgical instrument within the predefined boundary region may not cause movement of the robot arm, and accordingly movement of the laparoscope; whereas, movement of the target surgical instrument outside of the predetermined boundary region causes the robot arm to move the laparoscope to maintain the target surgical instrument within the predefined boundary region within the field of view of the laparoscope. Further, upon initiation of the instrument centering mode by the system to track a handheld surgical instrument within the field of view of the laparoscope, using object segmentation to distinguish surgical instruments when more than one surgical instrument is within the field of view of the laparoscope, the system may disregard non-target surgical instruments, e.g., surgical instrument(s) coupled to a robot arm within the field of view of the laparoscope, and track only the target surgical instrument, e.g., the handheld surgical instrument within the field of view of the laparoscope.
Moreover, robotic assist mode determination module 1438 may determine one or more conditions exist, e.g., the surgical procedure being performed by the operator enters a known phase of the surgical procedure indicating that the laparoscope should focus on a specific surgical instrument and/or a specific anatomical structure. As described above, the system may use object segmentation to identify anatomical structures within the field of view of the laparoscope during a procedure, such that upon detection of a specific surgical instrument/anatomical structure, the system may determine that the surgical procedure has entered a known phase. For example, during a cholecystectomy procedure, upon detection and identification of the cystic duct and target artery, e.g., via object segmentation and an stored/online database, robotic assist mode determination module 1438 may determine that the laparoscope should track the cystic duct and the artery, e.g., such that surgical clips may be applied, and therefore maintain the cystic duct and the artery within the field of view of the laparoscope during this known phase of the cholecystectomy procedure. Accordingly, the system may cause the robot arm to cause movement of the laparoscope to center its field of view on the cystic duct and the artery. As will be understood by a person having ordinary skill in the art, robotic assist mode determination module 1438 may be trained to identify various phases of various surgical procedures, such that the system may provide instrument centering to focus the field of view of the laparoscope on key surgical instruments and/or anatomical structures based on the phase of the surgical procedure. Moreover, the system may provide automated centering to focus the field of view of the laparoscope on key anatomical structures rather than a surgical instrument based on the type of surgical instrument identified within the field of view of the laparoscope using any of the surgical instrument identification methods described herein. For example, if the surgical instrument is identified as a suture device, the system may provide assisted scope control to focus the field of view of the laparoscope on the anatomical structure(s) being sutured rather than the suture device.
Trajectory generation module 1440 may be executed by processor 1402 for generating a trajectory from the current position of the distal end of the robot arm to a desired position of the robot arm, which will result in moving the distal end of the laparoscope from its current position to a desired position to maintain the target surgical instrument within the field of view of the laparoscope to thereby provide instrument centering in the robotic assist mode. The generated trajectory is configured to provide the robot arm precise control over the speed and smoothness of the motion of the laparoscope as it is moved along the trajectory. Specifically, as shown in
In some embodiments, the surgeon's preferences may be learned based on data from past procedures and/or sensors collecting information about current procedure including a surgeon's current pose, a surgeon's height, a surgeon's hand preference, and other similar factors. For example, the system may record when a user interacts with the system and also record what the user does with the system, such that the dataset may allow for surgeon preferences to be “learned” and updated over time, as described below with regard to
For example, the system may provide a predetermined override time period where the user may manually move the laparoscope, e.g., to change the field of view, clean the laparoscope, adjust the zoom, etc., and return the laparoscope to a stationary position before the system exits instrument centering mode. For example, the predetermined override time period may be, e.g., 6 to 12 seconds, or preferably 8 seconds. During the override time period, the user may adjust the reference distance, e.g., the distance between the tip of the laparoscope and the tip of the target instrument, by bringing the instrument tip to the desired distance from the laparoscope for the laparoscope to follow once the target instrument is detected in the instrument centering mode. Upon exiting of the instrument centering mode, instrument centering mode may be reentered via, e.g., actuation at GUI 210. Moreover, the overriding motion of the robot arm may be recorded as the operator's preferred trajectory given the current position of the target surgical instrument, such that the trajectory generation module associated with the given operator's surgeon profile may be updated to include the operator's preferred trajectory. Moreover, as shown in
In addition, trajectory generation module 1440 may account for an angular offset between a camera sensor module, e.g., the camera head of the laparoscope, and the laparoscope coupled to the robot arm, e.g., distal end of the robot arm, when generating the trajectory. For example, the camera head of the laparoscope may be attached to the laparoscope at an arbitrary location relative to the robot coordinate system, with an unknown rotational position relative to the laparoscope. Thus, for an angled-tip laparoscope, movement of the robot arm along the generated trajectory will not result in the desired movement of the field of view of the laparoscope when the laparoscope is not in the appropriate orientation relative to the laparoscope. For example, if the camera head is upside-down relative to the laparoscope, movement of the robot arm that moves the laparoscope in the up direction will result in a shift of the field of view of the laparoscope in the down direction. Trajectory generation module 1440 may detect the angular offset between the camera head and the distal end of the robot arm, as described in further detail below with regard to
For example, trajectory generation module 1440 may detect the angular offset by causing the robot arm to execute a predefined movement pattern, e.g., moving forward/backward and/or side-to-side, and comparing the expected movement of a static object within the field of view of the laparoscope, e.g., an anatomical structure or a surgical instrument, with the actual movement of the static object within the field of view of the laparoscope. As will be understood by a person having ordinary skill in the art, a “static object” may be any object within the field of view of the laparoscope which is generally considered “static” with respect to the movement of the robot arm, and thus, may include objects in the background scene that move slightly due to, e.g., respiratory motion, cardiac motion, etc. Based on the detected offset, trajectory generation module 1440 may calibrate the trajectory generation module to account for the offset and generate a calibrated trajectory, such that movement of the laparoscope along the calibrated trajectory will result in the expected shift of the field of view of the laparoscope to maintain the target surgical instrument within the field of view of the laparoscope. In some embodiments, the system may include one or more rotation sensors configured to detect the angular position of the camera head relative to the laparoscope, and accordingly an angular offset, and thus would not need to execute the predefined movement pattern. Moreover, in some embodiments, additional image data, e.g., from optical scanner 202, indicative of movement of the proximal end of the laparoscope external to the patient may be used to facilitation determination of the offset.
Similarly, trajectory generation module 1440 may account for positional offset of the attachment point of the coupler body along the shaft of the surgical instrument. For example, as described above, the coupler body is preferably coupled to the instrument shaft at the proximal-most point along the shaft, such that the distance between the distal end of the robot arm and the instrument tip may be known for the force measurements described in further detail below. However, in the event that the coupler body is not attached at the proximal-most point along the instrument shaft, movement of the distal end of the robot arm may not result in the desired movement of the instrument tip. Accordingly, the system may detect the positional offset of the coupler body attachment point, if any, and calibrate its force measurement algorithms to account for the offset, such that movement of the distal end of the robot arm will result in the expected movement of the instrument tip.
Fault detection module 1442 may be executed by processor 1402 for analyzing the data indicative of the operating characteristics of the system, e.g. position data generated by robot arm position determination module 1418 and/or trocar position detection module 1420 and/or force measurement calculated by force detection module 1422, to detect whether a fault condition is present. For example, fault detection module 1442 may a fault condition of the system and determine whether the fault condition is a “minor fault,” a “major fault,” or a “critical fault,” wherein each category of fault condition may be cleared in a different predefined manner.
For example, fault detection module 1442 may detect a minor fault condition such as robot arm 300 being moved with a velocity exceeding a predetermined velocity threshold, which may be cleared, e.g., by slowing down the movement of robot arm 300. In some embodiments, the system may automatically apply additional impedance to robot arm 300 when robot arm 300 is moving too fast, e.g., a temporary localized virtual boundary, to thereby force the operator to slow down movement of robot arm 300. The temporary localized virtual boundary may be applied whether or not an instrument is attached to the robot arm, but that the velocity of the distal end of the robot arm exceeds a predetermined velocity threshold. Moreover, fault detection module 1442 may detect a major fault condition such as an inadvertent bump of robot arm 300 as indicated by a large force applied to robot arm 300 by a person other than the operator. In response to detection of a major fault condition, fault detection module 1442 may actuate the braking mechanism associate with each motorized joint of robot arm 300 (or at least the joints associated with the major fault condition), to thereby freeze robot arm 300 and inhibit further movement of robot arm 300. Such a major fault condition may be cleared by the operator actuating a “clear” option displayed on user interface 1408. Fault detection module 1442 may detect a critical fault condition such as redundant encoders associated with a given joint of robot arm 300 generating different angulation measurements with a delta exceeding a predetermined threshold. In response to detection of a critical fault condition, fault detection module 1442 may actuate the braking mechanism associate with each motorized joint of robot arm 300 to thereby freeze robot arm 300 and inhibit further movement of robot arm 300. Such a critical fault condition may be cleared by the operator restarting the system. Upon restart of the system, if the critical fault condition is still detected by fault detection module 1442, robot arm 300 will remain frozen until the critical fault condition is cleared.
Indicator interface module 1444 may be executed by processor 1402 for causing indicators 334 to communicate the state of the system, e.g., the operational mode of robot arm 300, to the operator or other users, based on, for example, determinations made by passive mode determination module 1432, co-manipulation mode determination module 1434, haptic mode determination module 1436, and/or robotic assist mode determination module 1438. For example, indicator interface module 1444 may cause indicators 334 to illuminate in specific color light associated with a specific state of the system. For example, indicator interface module 1444 may cause indicators 334 to illuminate in a first color (e.g., yellow) to indicate that no surgical instrument is attached to the robot arm, and that the robot arm may be moved freely such that the system compensates for the mass of the robot arm; in a second color (e.g., purple) to indicate that a surgical tool is attached to the robot arm, and that the robot arm may be moved freely such that the system compensates for the mass of the robot arm and the mass of the surgical instrument coupled to the robot arm; in a third color (e.g., blue) to indicate that a surgical instrument is attached to the robot arm, and that the robot arm is in the passive mode as determined by passive mode determination module 1432; in a fourth color (e.g., pulsing orange) to indicate that at least a portion of the robot arm and/or the surgical instrument attached thereto is within the virtual haptic boundary, e.g., 1.4 m or more above the ground; in a fifth color (e.g., pulsing red) to indicate that a fault has been detected by the system by fault detection module 1442. As will be understood by a person having ordinary skill in the art, different colors and patterns may be communicated by indicators 334 to indicate the states of the system described above.
Additionally, indicators 334 may be illuminated in other distinct colors and/or patterns to communicate additional maneuvers by robot arm 300, e.g., when robot arm 300 retracts the surgical arm in the robotic assist mode, or performs another robotically-assisted maneuver in the robotic assist mode. As described above, indicators 334 further may include devices for emitting other alerts such as an audible alert or text alert. Accordingly, indicator interface module 1444 may cause indicators 334 to communicate the state of the system to the operator using audio or text, as well as or instead of light. For example, indicator interface module 1444 may cause one or more speakers to emit an audible alert that changes in, e.g., amplitude and/or frequency, as robot arm 300 approaches a potential collision with one or more objects/persons within the operating room.
Additionally or alternatively, indicator interface module 1444 may communicate the state of the system, e.g., transition from co-manipulation mode to passive mode, via haptic feedback at the distal end of robot arm 300, and accordingly on the surgical instrument coupled thereto. For example, when the surgical instrument is held in a position for the predetermined dwell time such that the system switches to passive mode, the user may feel a vibration at the surgical instrument indicating that the system has transitioned to passive mode and that the user may let go of the surgical instrument. As another example, the user may feel a vibration after the surgical instrument is coupled to the coupler body to indicate that the surgical instrument is successfully coupled to the robot arm. The vibration may be strong enough to be felt by the user, but weak enough such that any movement at the distal tip of the surgical instrument resulting therefrom is negligible.
The co-manipulation surgical robot systems described herein may include additional modules within memory 1410 of platform 1400 for executing additional tasks based on the data obtained. For example, the system may determine if the surgical instrument has been detached from robot arm 300 based on data indicative of the position of the distal end of robot arm 300 relative to the trocar point generated by trocar position detection module 1420, as well as the direction of an instrument shaft and/or an orientation of the distal-most link of robot arm 300, e.g., distal wrist link 316. For example, if the instrument is pointing directly at the trocar, then there is a higher probability that a tool is attached to the robot arm. Moreover, axis Q7 of robot arm 300 may indicate the pointing direction of the instrument and, if the instrument is passing through the trocar port, the distal wrist link 316 will point in a direction of the trocar port. Therefore, if distal wrist link 316 is not pointing toward the trocar port, then the system may determine that the robot arm is not supporting an instrument or the instrument is not advanced through the trocar port. For example, when an instrument is detached from robot arm 300 and robot arm 300 is moved, the computed direction of the instrument shaft (e.g., the direction that the instrument would point if attached to robot arm 300) may no longer point to the trocar entry point and likely will not point to the trocar entry point. Accordingly, the may alert a user if the system determines that no tool is coupled with robot arm 300, e.g., via indicators 334, and/or apply a localized virtual haptic boundary to slow down the robot arm if the robot arm is moving in a single direction and there is no movement at Q7.
Referring now to
Alternatively, as described above, the system may automatically identify the surgical instrument upon attachment to the robot arm, and accordingly, may automatically load the corresponding calibration file. For example, as described above, the system may identify at least the shaft diameter of the surgical instrument coupled to the coupler body (or that will be coupled to the coupler body) when the coupler body is coupled to the coupler interface based on the specific magnetic field strength measured by sensor 414 induced by the displaced magnet within the coupler body, which may be indicative of whether a 5 mm or 10 mm coupler body is coupled to the coupler interface, with or without the surgical instrument attached. Accordingly, the system may automatically load a calibration file associated with a surgical instrument having the identified shaft diameter, which may be used to automatically calibrate a surgical instrument in real-time via adaptive gravity compensation, as described in further detail below with regard to
If the calibration file for the selected surgical instrument is not available in the database, the operator may self-calibrate the surgical instrument using the system. For example,
At step 1905, the system compensates for the gravity of the surgical instrument and the force applied by the hand of the operator, e.g., by measuring the force applied to the distal end of robot arm 300 due to the mass of the surgical instrument. As described above, the force applied to the distal end of robot arm 300 may be measured by measuring the motor current across the motors disposed in the base of robot arm 300. If the system overcompensates for the gravity of the surgical instrument, at step 1906, robot arm 300 may “runaway”, e.g., drift upward. The runaway effect may be detected at step 1907, and at step 1908, indicators 334 may blink to indicate to the operator of the runaway. At step 1909, the system may identify the runaway as a minor fault, and accordingly apply additional impedance to robot arm 300 and freeze robot arm 300 when robot arm 300 slows down before removing the additional impedance. Once the minor fault is addressed, calibration process 1900 may return to step 1903.
After step 1905, when the system compensates for the gravity of the surgical instrument, if the surgical instrument is detached, either accidentally or manually by the operator at step 1911, at step 1910, the system detected the detachment of the surgical instrument from robot arm 300. As a result, the system will stop compensating for the gravity of the surgical instrument, and calibration process 1900 may return to step 1903. Moreover, as described above, the system may apply a temporary localized virtual haptic boundary at the distal end of the robot arm upon detection of the detachment of the surgical instrument from robot arm 300. After step 1905, when the system compensates for the gravity of the surgical instrument, calibration process 1900 is ready to enter calibration mode at step 1912. For example, the operator may initiate calibration mode via user interface 1408 at step 1913. At step 1914, the system may indicate to the operator, e.g., via user interface 1408 and/or blinking of indicators 334, that it is safe to let go of surgical instrument, such that the operator may let go of the surgical instrument at step 1916. At step 1915, the system calibrates the surgical instrument. As described above, provided that each specific make of a surgical instrument may have a distinguishable, precise mass, the make of the surgical instrument may be determined by comparing the calibrated mass with a stored or online database of surgical instruments. Accordingly, a surgical instrument may be labeled as an unauthorized surgical instrument during calibration.
Referring again to
For example, as shown in
Referring again to
Moreover, as the operator freely moves the retractor in the co-manipulation mode, e.g., prior to inserting the tip of the retractor through the trocar within the patient, if the operator moves the tip of the retractor too close to the patient's skin away from the trocar port, and a virtual haptic boundary has been established by the system on the skin of the patient outside the trocar ports, the system may automatically switch to the haptic mode. Accordingly, the system may apply an impedance to the second robot arm that is much higher than the impedance applied to the second robot arm in co-manipulation mode to indicate to the operator that they are approaching or within the virtual haptic boundary. For example, movement of the retractor by the operator may feel much more viscous in the haptic mode. The system may remain in the haptic mode until the operator moves the retractor out of the virtual haptic boundary. In some embodiments, in the haptic mode, the second robot arm may reduce the effects of gravity, eliminate tremor of the instrument tip, and apply force feedback to avoid critical structures as defined by the virtual haptic boundary. Accordingly, the system does not replace the operator, but rather augments the operator's capabilities through features such as gravity compensation, tremor removal, haptic barriers, force feedback, etc.
In some embodiments, the system may switch the second robot arm to the robotic assist mode. For example, as the operator attempts to retract the tissue, if more force is required to retract the tissue than the operator is able or willing to apply to the retractor, the operator may provide user input to the system indicating that the operator wants the second robot arm to assist in the retraction of the tissue. For example, as described above, the operator may perform a predefined gestural pattern that may be detected by, e.g., optical scanner 202, such that the system switches the second robot arm to the robotic assist mode and causes the motors of the second robot arm to move the second robot arm, and accordingly the retractor, to provide the additional force required to retract the tissue.
In addition, instead of manually manipulating the laparoscope coupled to the first robot arm as described, the operator may provide another user input to the system indicating that the operator wants the system to reposition the laparoscope. For example, if the operator is actively manipulating a surgical scissor, which may or may not be coupled to a robot arm of the system, such that the tip of the surgical scissor is within the field of view of the laparoscope coupled to the first robot arm, the operator may perform a predefined gestural pattern with the tip of the surgical scissor, e.g., moving the surgical scissor quickly back in forth in a particular direction. The predefined gestural pattern of the surgical scissor may be captured as image data by the laparoscope, and based on the data, the system may detect and associate the predefined gestural pattern with a predefined user input requiring that the system switch the first robot arm from the passive mode to the robotic assist mode, and cause the first robot arm to reposition itself, and accordingly the laparoscope, to adjust the field of view in the direction of the pattern motion of the surgical scissor, or alternatively, reposition itself to adjust the field of view to ensure that the tip of the surgical scissors remain within an optimum position within the field of view of the laparoscope during the procedure. As described above, additional gestural patterns may be performed via the surgical scissor within the field of view of the laparoscope to cause the first robot arm to retract the laparoscope and/or to cause the laparoscope itself to zoom in or zoom out or improve resolution.
In some embodiments, as described above, based on the image data captured by the laparoscope, using object tracking of the additional tools in the field of view of the laparoscope, e.g., the surgical scissors actively operated by the operator, the system may cause the first robot arm coupled to the laparoscope to switch to the robotic assist mode and cause the first robot arm to automatically reposition itself to adjust the field of view to ensure that the tip of the surgical scissors remains within an predefined optimum position, e.g., a boundary region, within the field of view of the laparoscope during the procedure.
For example,
If the system determines that a surgical instrument has been selected to be followed at step 2104 and/or if the system identifies that there is a primary tool for the given phase of the surgical procedure, at step 2106, the system may determine whether the selected surgical instrument is the primary tool for the given phase of the surgical procedure. If the surgical instrument to be followed is not the primary tool, at step 2107, the system determines whether it can successfully track the selected surgical instrument across consecutive images of the laparoscope video feed. If the system determines that it can track the selected surgical instrument, the system proceeds to provide instrument centering at step 2108. If the system determines that the surgical instrument to be followed is the primary tool at step 2106, then at step 2108, the system detects the selected surgical instrument within the image data, e.g., a single image of the laparoscope video feed, and proceeds to provide instrument centering at step 2109.
At step 2110, the system continuously detects/follows/tracks the surgical instrument within the field of view of the image data. For example, at step 2111, the system checks the scale, e.g., resolution/size, of the surgical instrument within the image data, and determines at step 2112 whether or not the scale of the surgical instrument within the image data has changed. If the scale has not changed, the system continues to detect/follow/track the surgical instrument within the image data. If the scale has changed, at 2113, the system may send a command to the robot arm to move the laparoscope, e.g., along the longitudinal axis of the laparoscope, to zoom in/out based on the change of scale of the surgical instrument within the image data. For example, if the scale changes such that the size of the surgical instrument increase, and thus, the resolution decreases, the command may be to zoom out, e.g., retract the laparoscope.
Simultaneously, at step 2114, the system checks the position of the surgical instrument within the image data, and determines at step 2115 whether or not the position of the surgical instrument within the image data has changed, e.g., moved outside of a predefined virtual boundary region within the field of view of the laparoscope. If the position has not changed, such that the surgical instrument, though moving, has not been detected to have moved outside of the predefined boundary region, the system continues to detect/follow/track the surgical instrument within the image data. If the position has changed such that the surgical instrument is detected to have moved outside of the predefined boundary region, at 2116, the system may generate a trajectory, as described above, and send a command to the robot arm to move the laparoscope along the trajectory to a desired position to maintain the surgical instrument within the predefined boundary region. For example, the command may cause the robot arm to move the laparoscope to a position where the surgical instrument is at the center of the predefined boundary region. Moreover, as described above, the system may correct for a detected angular offset between the camera head of the laparoscope and the laparoscope in robotic assist mode to provide accurate instrument centering.
As shown in
As described above, once the trajectory is determined, the system calculates the force required to apply at the joints of the robot arm to move the robot arm, and accordingly the laparoscope, along the trajectory.
As shown in
F
eff
+W+F
tr=0=>Ftr=−W−Feff
Where Feff is the force at the distal end of robot arm 300 (e.g., the “end-effector force” of robot arm 300), W is the weight vector of the surgical instrument (=−mgz), and Ftr is the trocar force. Accordingly, Feff is the desired force sent to the system, which is the sum of all the forces generated in the algorithm pipeline including, e.g., gravity compensation, hold, etc.
As shown in
W+F
eff
+F
tr
+F
tt=0
F
eff
×D1+Ftr×D2+Ftt×D3=0
Here, distances D1 and D3 are known as described above, and D2 may be derived based on the known position of the distal end of robot arm 300 and the calculated position of trocar Tr. As shown in
As described above, the system may alert the operator if the forces, e.g., force Ftt applied to the tip of the instrument and/or force Ftr applied by the instrument at trocar Tr, are greater than the respective threshold forces, and accordingly freeze the system if the calculated force is greater than the threshold force, and/or reduce the force exerted at the trocar point at the body wall or at the tip of the instrument by automatically applying brakes or stopping forces to robot arm 300, by slowing or impeding further movement of the instrument in the direction that would increase forces applied at the tip of the instrument or the trocar, and/or automatically moving the robotic arm in a direction that reduces the force being exerted at the instrument tip and/or at the trocar point at the body wall.
Accordingly, with knowledge of, e.g., the position of the trocar, the distance from the coupler body to the instrument tip, the current position of the distal end of the robot arm, the current position of the instrument tip, the desired position of the distal end of the robot arm that provides the desired position of the instrument tip, the system may calculate the force required to apply to the distal end of the robot arm to move it from its current position to its desired position to thereby move the instrument tip from its current position to its desired position. For example, when a robot arm is in passive mode, the desired position of the distal end of the robot arm/instrument tip will be the static position of the distal end of the robot arm/instrument tip when passive mode was initiated, and any forces below the breakaway force applied to the distal end of the robot arm, e.g., due to perturbations, may cause a change in the current position of the distal end of the robot arm/instrument tip. Accordingly, the system may calculate the force required to apply to the distal end of the robot arm to move it from its current position to the desired static position (i.e., the “hold force”), and apply the requisite torque to the robot arm to effectively maintain the robot arm in the desired static position in passive mode.
Similarly, as shown in
Moreover, as described above, the system may automatically calibrate a surgical instrument in real-time via adaptive gravity compensation, e.g., based on the hold forces required to maintain the robot arm in a static position in passive mode. For example, an instrument's gravity characteristics may change dynamically at different locations and/or with different attachments, e.g., scope cables, forces from the patient's body wall, during operation of the robot arm. Accordingly, the system may update a calibration file of the surgical instrument in real-time, e.g., in passive mode, to thereby dynamically adjust the gravity compensation applied by the system.
At step 2402, the system may automatically load a calibration file associated one or more known parameters of a surgical instrument coupled to the robot arm. For example, as described above, at least the shaft diameter of the surgical instrument may be determined upon coupling of the coupler body to the coupler interface (with or without the surgical instrument attached), e.g., based on the specific magnetic field strength induced by the displaced magnet within the coupler body. Additionally, or alternatively, image data captured by the optical scanner(s) may be used to identify one or more known parameters of the surgical instrument, e.g., the length and/or instrument type, and/or measured motor currents may be used to estimate the mass of the instrument and/or a gravity category of the instrument such as “light,” “medium,” or “heavy.” Accordingly, the system may load a calibration file associated with the known parameter(s), e.g., a calibration file associated with a 5 mm or 10 mm surgical instrument. As will be understood by a person having ordinary skill in the art, various surgical instruments having the same shaft diameter also may have other similar parameters, such as mass, center of gravity, etc. Accordingly, the loaded calibration file may include one or more instrument parameters that deviate from the actual parameters of the attached surgical instrument within an acceptable range. Moreover, the loaded calibration file may be used to inform acceptable gravity compensation adjustments, as described in further detail below.
At step 2404, the system may apply gravity compensation to the robot arm coupled to the surgical instrument based on the instrument parameters within the loaded calibration file. For example, based on the mass of the surgical instrument stored in the calibration file (whether or not it is accurate), the system may calculate the amount of force to apply to the robot arm to compensate for the presumed weight of the surgical instrument and maintain the robot arm in a static position in passive mode, and apply the requisite torque to the robot arm to provide the gravity compensation.
At step 2406, the system may calculate the hold force required to maintain the robot arm in a static position in passive mode. Preferably, the system continuously calculates the hold force while in passive mode; however, in some embodiments, the system may begin calculating the hold force after a predetermined time period to ensure that the robot arm is steady in a static position, as described in further detail below with regard to
At step 2408, the system may determine one or more calibrated instrument parameters for the surgical instrument, e.g., the mass and/or center of mass of the surgical instrument, based on the calculated hold force, and update the calibration file accordingly. For example, the system may determine and update the mass of the surgical instrument within the calibration file to a mass that corresponds with a gravity compensation force that would result in no hold force being required to be applied to the distal end of the robot arm to maintain the surgical instrument in the desired static position, e.g., the updated mass is equal to the actual mass of the surgical instrument. As will be understood by a person having ordinary skill in the art, the system may update and save the loaded calibration file as a new calibration file associated with the specific surgical instrument coupled to the robot arm and/or create a new calibration file for the specific surgical instrument.
At step 2410, the system may calculate and apply a subsequent, adjusted gravity compensation force to the distal end of the robot arm based on the calibrated instrument parameter, e.g., the updated mass of the surgical instrument, which will minimize the hold force required to maintain the surgical instrument in the desired static position within a predetermined condition, e.g., by reducing the hold force to or near zero. Moreover, the system may be programmed to only permit gravity compensation adjustments within a predetermined range for a particularly sized instrument, e.g., based on the known shaft diameter upon coupling of the coupler body to the coupler interface. For example, if the system knows that the instrument coupled to the robot arm is a 5 mm diameter instrument, which is generally associated with a known range of masses, the mass of the instrument within the calibration file may be updated to a mass only within a predetermined range of the presumed mass within the loaded calibration file, such that gravity compensation also may only be adjusted within a predetermined range. Moreover, additional known parameters, e.g., gravity category, instrument type, instrument length, etc., may also be associated with a respective known range of mass for that particular parameter, to thereby limit the range of gravity compensation adjustment for the particular surgical instrument. In addition, adjustments to the gravity compensation applied to the robot arm by the system may be applied gradually so as to avoid jumpy motion of the robot arm.
At step 2408, the system may determine and update, e.g., the mass and/or center of mass Lcg, of the instrument within the loaded calibration file to a mass and/or center of mass that corresponds with an increased gravity compensation force that would result in a reduction of the hold force required to be applied to the distal end of the robot arm to maintain the surgical instrument in the desired static position. Accordingly, as shown in
Referring now to
Moreover, the rate/pattern of the change in the hold force over time upon initiation of passive mode may further be indicative of whether the perturbations are due to, e.g., the instrument holding tissue or the user repositioning the instrument. For example, a slow gradual increase of hold force may indicate that the instrument is holding tissue in passive mode; whereas, a change in hold force having a fluctuating profile may indicate that the user is still interacting with the instrument and has not released the instrument, e.g., with forces less than the breakaway force. In addition, the system may temporarily change the breakaway force of the robot arm to a predetermined, high value that would require a large amount of force to transition the robot arm from passive mode to co-manipulation mode for at least the duration of the predetermined time period before which the hold force is expected to become steady. Moreover, the system may set the hold force calculated after the predetermined time period as a baseline hold force, and adjust the breakaway force based on the baseline hold force to compensate for otherwise large hold forces required to maintain a surgical instrument in a desired static position in passive mode, as described in further detail with regard to
As shown in
As described above, the system may continuously calculate the hold force required to maintain the distal end of the robot arm, and accordingly the surgical instrument coupled thereto, in the desired static position while in passive mode, and may calculate the average hold force after the predetermined time period upon initiation of passive mode, and establish this value as the baseline hold force. As further described above, gravity compensation may be dynamically adjusted to thereby adjust the gravity compensation force to the robot arm and reduce the baseline hold force within acceptable limits associated the surgical instrument. Accordingly, the baseline hold force may be the hold force calculated after the predetermined time period minus any reductions due to increased gravity compensation. Moreover, the system may then establish the breakaway force based on the baseline hold force, e.g., as a value having a predetermined delta from the baseline hold force, as shown in
As shown in
As shown in
For example,
As shown in
As described above, the systems described herein may detect the angular offset between a camera head of a laparoscope and the distal end of the robot arm coupled to the laparoscope. For example,
Referring now to
Image motion calculation may be used to quantify image motion, e.g., the changes across consecutive images acquired from the laparoscopic camera resulting from movement of the laparoscope during operation, e.g., during the robotic instrument centering mode. For example, to quantify image motion for image registration, the motion of individual pixels of the laparoscopic images between consecutive images may be computed at image motion computation 2902, and the motion results may then be combined to obtain the image motion direction on the image space at image motion direction 2904.
At image motion computation 2902, a plurality of 2D images may be received from the laparoscope device, and a computer vision technique such as, for example, optical flow, may be used to compute the motion of individual pixels of the laparoscopic images between consecutive images, which provides displacements of each individual pixel between consecutive images in the x and y directions within the 2D plane of the images. With the x and y displacements of the individual pixels between consecutive images, at image motion direction 2904, the averages, e.g., means/median, of both the x and y displacements may be calculated to obtain the motion vector in the 2D image space, to thereby calculate the image motion direction. For example, the image motion direction (in angle) may be calculated as:
√{square root over (xmean2+ymean2)}
Moreover, motion percentage may be calculated as the percentage of image pixels that moved between consecutive images, which may be indicative of whether the motion is global motion introduced by movement of the laparoscope device, or local motions caused by, e.g., tissue-tool interaction. In addition, the percentage of moved pixels that agree on the computed image motion direction may be counted to calculate motion consensus. The agreement may be checked by calculating the relative angle between the motion vector of each individual pixel between consecutive images and the image motion direction vector. For example, the relative angle may be calculated as:
The relative angle may be compared against a predetermined relative angle threshold, such that the individual pixel motion vector is determined to agree with the computed image motion vector is the relative angle is greater than the predetermined relative angle threshold. Accordingly, based on the validation metrics above, the pixels in the black boarders and the pixels indicating local tissue motion may be filtered from the optical flow displayed in
Image-to-robot synchronization may be used to differentiate the image motion caused by the movement of the laparoscope device from the image motion caused by local tissue/instrument movement. For example, the motion of the laparoscope device, e.g., the motion of the distal end of the robot arm, may be retrieved from the robot arm sensors, e.g., encoders, during operation. Moreover, the corresponding motions of the image and the laparoscope device may be synchronized based on the available timestamps of robot arm motion and the laparoscopic images. As shown in
As shown in
In background mode, at robot pivoting motion identification 2912, movement of the distal end of the robot arm, e.g., passive robot arm motion, responsive to movement of the laparoscope device by the user may be retrieved from the robot arm sensors. At synchronization 2914, based on the timestamps of the retrieved robot arm movements, corresponding timestamps of the computed image motion may be extracted. Accordingly, at orientation offset comparison 2910, a comparison between the robot arm motion and the computed image motion may be performed to calculate the angular offset between the orientation of the camera sensor module and the distal end of the robot arm, e.g., by calculate the angle between the computed image motion direction vector and the retrieved robot arm motion vector.
Moreover, manual rotation of the camera sensor module relative to the laparoscope device by the user during operation may be detected in real-time. For example, when the robot arm is stationary while the camera sensor module is being rotated by the user, the motion of each individual pixel of the images may be computed between consecutive images responsive to the rotation, and aggregated in the angle space to obtain the rotation change. The computer vision techniques described herein further may be used to identify the laparoscope device type, e.g., whether the laparoscope device is has a flat tip or an angled tip. For example, the system may cause the robot arm to move the laparoscope device back/forth along the longitudinal axis of the laparoscope device. Accordingly, based on the validation metrics described above, if the computed image motion direction is greater than a predetermined threshold indicating a major direction on the image space, the system may determine that the laparoscope device is not a flat tip laparoscope device. In contrast, if the computed image motion direction is less than the predetermined threshold indicating a minor direction on the image space, e.g., zoom in/out, the system may determine that the laparoscope device is a flat tip laparoscope device.
Referring again to
Referring now to
As shown in
In some embodiments, the controller may only cause display 210 to display the virtual map while platform 200 is being moved within the operating room. For example, platform 200 may include one or more actuators, e.g., a button, lever, or handlebar, that may be operatively coupled to the braking mechanism of the wheels of platform 200, such that upon actuation of the actuator, the braking mechanism is disengaged such that mobility of platform 200 is permitted. Accordingly, when the actuator is not actuated, the braking mechanism is engaged such that mobility of platform 200 is prevented. Thus, upon actuation of the actuator, the controller may automatically cause display 210 to display the virtual map, such that operator O can view the area surrounding platform 200 before, during, or after movement of platform 200 while the braking mechanism is disengaged. Once the actuator is released, such that the braking mechanism is reengaged, display 210 may stop displaying the virtual map. In some embodiments, when the virtual map indicates that platform 200 and/or robot arms 300a, 300b are approaching or within the predetermined distance from the one or more objects/persons within the operating room, the controller may override actuation of the actuator by the operator and reengage the braking mechanism to thereby prevent further movement of platform 200. Accordingly, the actuator may need to be released and re-actuated by the operator to disengage the braking mechanism and permit further movement of platform 200.
Moreover, the system may process color and/or depth data obtained from optical scanners 202 and proximity sensors 212 to identify objects within the operating room, e.g., the patient bed or the trocar, as well as the planes associated with the identified objects. With knowledge of the location platform 200 and robot arms 300a, 300b relative to the identified objects, the system may cause the stages coupled to the base portions of robot arms 300a, 300b to automatically move (or stop movement of) robot arms 300a, 300b to avoid collision with the identified objects during setup, e.g., when robot arms 300a, 300b approaches a predetermined distance threshold relative to the identified objects. In addition, the system may generate and emit, e.g., an audible alert indicative of the proximity of the stages of platform 200 and/or robot arms 300a, 300b relative to the identified objects. For example, the audible alert may change in amplitude and/or frequency as the distance between the stages of platform 200 and/or robot arms 300a, 300b and the identified objects decreases, as perceived by the system based on the depth data.
In addition, with knowledge of the location platform 200 and robot arms 300a, 300b relative to the trocar, if the system detects that the position of the patient bed, and accordingly the trocar, is changing, e.g., via adjustment by a user, the system may automatically adjust the arrangement of the robot arm to accommodate the movement of the patient bed and maintain relative position between the distal end of the robot arm and the trocar. In some embodiments, upon detection of movement of the patient bed, the system may automatically move the robot arm to retract the surgical instrument coupled thereto within the trocar, prior to automatically adjusting the arrangement of the robot arm to maintain relative position between the distal end of the robot arm and the trocar, such that the distal end of the surgical instrument is positioned within the trocar and away from anatomical structures within the patient.
Referring to
Referring now to
In addition,
Referring now to
Non-real-time computer 3602 further may provide user feedback 3612 to the user via user interface 3614. User feedback may include, e.g., collision notifications, positioning information and/or recommendations regarding the various components of the system, the operational mode that has been detected by the system, etc. Non-real-time computer 3602 further may provide commands 3618, e.g., high level commands, to real-time computer 3608. High-level commands may include, e.g., mode changes, trajectories, haptic barriers, user configurations, etc. Real-time computer 3608 may include robot controller 3620 programmed to provide robot commands 3622, e.g., motion or force commands, to the one or more robot arms 3624, e.g., robot arms 300. Robot controller 3620 may receive robot feedback data 3626, e.g., motion, force, and/or touchpoint data, etc., from the one or more robotic arms 3624.
Referring now to
Moreover, centralizing procedure data may enable the running of large data analytics on a wide range of clinical procedures coming from different users. Analysis of data may result in optimized settings for a specific procedure, including, e.g., optimized system positioning, optimal ports placement, optimal algorithms settings for each robot arm and/or detection of procedure abnormalities (e.g., excessive force, time, bleeding, etc.). These optimal settings or parameters may depend on patient and tool characteristics. As described above, a surgeon may load and use optimal settings from another surgeon or group of surgeons. This way, an optimal setup may be achieved depending on, e.g., the surgeon's level of expertise. To keep track of the various users in the distributed network of cobot systems, it may be beneficial to identify each user. As such, the user may log into the cobot system and access their profile online as necessary. This way the user may have access to their profile anywhere and will be able to perform a clinical procedure with their settings at a different hospital location.
An example user profile may contain the user's specific settings and information, including, e.g., username; level of expertise; different procedures performed, and/or region of clinical practice. In addition, the clinical procedure may require a user to store specific settings such as clinical procedure (e.g., cholecystectomy, hernia, etc.), table orientation and height, preferred port placement, settings per assistant arm for each algorithm, patient characteristics (e.g., BMI, age, sex), and/or surgical tools characteristics and specifications (e.g., weights, length, center of gravity, etc.). The user may be able to enable his own profile, and optionally may enable another user's profile, such as the profile of a peer, the most representative profile of a surgeon of the user's area of practice, the most representative profile of a surgeon with a specific level of expertise, and/or the recommended profile according to patient characteristics.
The identification of a user may be performed via password, RFID key, facial recognition, etc. Learning from a large number of procedures may result in a greater level of optimization of the cobot system setup for a given procedure. This may include, e.g., cart position, individual robot arm position, surgical table height and orientation, port placement, setup joints position, laparoscope trajectories during instrument centering. These settings may be based on patient height, weight, and sex, and further may be interdependent. For example, the optimal port placement may depend on patient table orientation.
Additionally, a clinical procedure may be described as a sequence of clinical procedures steps. Learning these different steps may allow the cobot system to infer in real time the actual step for a given procedure. For example, learning clinical steps from procedures may allow or enable: adjustment of algorithm settings, adjustment of robot arm configuration to facilitate user action in a given phase, adjustment of a laparoscope position based on the phase of the procedure, the system to give the practical custom reminders, the system to notify staff of an estimate procedure end time, the system to alert staff if necessary equipment is not available in the room, and/or the system to alert staff of the occurrence of an emergency situation.
During a clinical procedure, the surgeon will often realize simple and routine surgical tasks such as grasping, retracting, cutting etc. Learning these different tasks may allow the cobot system to infer in real time preferences and habits of the surgeon regarding a sequence of a procedure in real time. Some algorithms of the cobot system may be tuned (i.e., adjusted and optimized) during the procedure based on this sequence recognition and help the user to be better at this simple surgical task. An example of such a task is the automated retraction of a liver during a gall bladder procedure. By aggregating the information over many cases, the optimized force vectors may be developed.
Further, some complications may occur during a clinical procedure that may result in unexpected steps or surgical acts. Learning how to discriminate these unexpected events would help the cobot system to enable some specific safety features. In case of emergency, the robot arms may be stopped or motion restricted depending on the level of emergency detected by the system.
Referring now to
As shown in
Similarly, as shown in
As shown in
Additionally, as shown in
The system may cause the platform and/or robot arms to move to each preset configuration by causing movement of the platform and/or robot arms in a limited number of degrees of freedom. For example, as shown in
For example, in the case of the laparoscopic cholecystectomy in the upper right quadrant of the patient's body, Arm 1 may hold a laparoscope and Arm 2 may hold a grasper, the surgical bed angle may be steep reverse Trendelenberg (head up), and the system may be positioned on the right side of the patient. Moreover, the grasper held by Arm 2 will be used to push tissue superiorly (in the direction of the patient's right shoulder), and the laparoscope would be best positioned at the umbilicus, also pointing superiorly in the direction of the patient's right shoulder. Thus, to optimize the surgeon's workspace for a cholecystectomy surgical procedure, as shown in Table 1, the preset configuration is such that the Arm 1 stages are down and out, and the Arm 1 shoulder is rotated to the left about the Q3 axis; whereas, the Arm 2 stages are up and in, and the Arm 2 shoulder is in the neutral configuration, as shown in
Moreover, in the case of a laparoscopic gastric sleeve from the midline to the upper left quadrant of the patient's body, Arm 1 may hold a laparoscope and Arm 2 may hold a grasper, the surgical bed angle may be steep reverse Trendelenberg (head up), and the system may be positioned on the right side of the patient. However, because the area of the operation is larger, e.g., from the midline to upper left quadrant, and the procedure involves retracting tissue inferiorly (towards the patient's feet), to optimize the surgeon's workspace for a gastric sleeve surgical procedure, as shown in Table 1, the preset configuration is such that the Arm 1 stages are down and in, and the Arm 1 shoulder is in the neutral configuration; whereas, the Arm 2 stages are up and in, and the Arm 2 shoulder is rotated to the right about the Q3 axis. Accordingly, the surgeon may hold active instruments and reach underneath Arm 2, with more space to retract towards the feet. In addition, in the case of a laparoscopic left colectomy in the upper and lower left quadrants of the patient's body, the surgical bed angle may be Trendelenberg tilt right, and the system may be positioned on the left side of the patient. Thus, to optimize the surgeon's workspace for a left colectomy surgical procedure, as shown in Table 1, the preset configuration is such that the Arm 1 stages are up and out, and the Arm 1 shoulder is rotated to the right about the Q3 axis; whereas, the Arm 2 stages are up and out, and the Arm 2 shoulder is rotated to the left about the Q3 axis. In this manner, the robot arm(s) may be automatically positioned in a configuration specific to the selected surgical procedure to assist with arm setup for that specific procedure.
Referring now to
Referring now to
Referring again to
As described above, in some embodiments, the stages may be rotatable relative to the platform to thereby cause the robot arms to rotate towards or away from each other, e.g., along a horizontal plane defined by the x-axis and the y-axis. Accordingly, in some embodiments, the platform 200 may have a smaller footprint, such that the robot arms may be entirely within the footprint of the platform in the stow mode, and the stages may be actuated to rotate and extend the base portions of the robot arms away from each other along the horizontal plane to achieve a spaced apart configuration generally provided by the wider footprint platform.
Referring now to
Some implementations of the systems described herein may be configured to be controlled or manipulated remotely, e.g., via joystick or other suitable remote control device, computer vision algorithm, force measuring algorithm, and/or by other means. However, in a preferred embodiment, the systems described herein operate without any telemetry, e.g., the robot arm is not teleoperated via a remote surgeon console separate from the robot arm, but instead the robot arm moves in response to movement applied to the surgical instrument coupled thereto. Any robot-assisted movements applied to the surgical instrument by the system, e.g., in the robotic assist mode, are not responsive to user input received at a remote surgeon console.
While various illustrative embodiments of the invention are described above, it will be apparent to one skilled in the art that various changes and modifications may be made therein without departing from the invention. The appended claims are intended to cover all such changes and modifications that fall within the true scope of the invention.
Number | Date | Country | Kind |
---|---|---|---|
21305417.4 | Mar 2021 | EP | regional |
21305929.8 | Jul 2021 | EP | regional |
21306904.0 | Dec 2021 | EP | regional |
21306905.7 | Dec 2021 | EP | regional |
22306496.5 | Oct 2022 | EP | regional |
23305026.9 | Jan 2023 | EP | regional |
This application is a continuation-in-part application of U.S. patent application Ser. No. 18/331,054, filed Jun. 7, 2023, now U.S. Pat. No. 11,844,583, which is a continuation-in-part application of U.S. patent application Ser. No. 18/297,489, filed Apr. 7, 2023, which is a continuation application of U.S. patent application Ser. No. 17/816,958, filed Aug. 2, 2022, now U.S. Pat. No. 11,622,826, which is a continuation application of PCT Patent Appl. No. PCT/IB2022/056159, filed Jul. 1, 2022, and a continuation-in-part application of PCT Patent Appl. No. PCT/IB2022/052989, filed Mar. 30, 2022, and claims priority to EP Patent Appl. No. 21306904.0, filed Dec. 22, 2021, EP Patent Appl. No. 21306905.7, filed Dec. 22, 2021, and EP Patent Appl. No. 21305929.8, filed Jul. 5, 2021, and EP Patent Appl. No. 21305417.4, filed Mar. 31, 2021, the entire contents of each of which are incorporated herein by reference. U.S. patent application Ser. No. 18/331,054, filed Jun. 7, 2023, is also a continuation-in-part application of U.S. patent application Ser. No. 18/057,191, filed Nov. 18, 2022, now U.S. Pat. No. 11,812,938, which is a continuation-in-part application of U.S. patent application Ser. No. 17/815,885, filed Jul. 28, 2022, now U.S. Pat. No. 11,504,197, and claims priority to U.S. Provisional Patent Appl. No. 63/378,434, filed Oct. 5, 2022, and EP Patent Appl. No. 22306496.5, filed Oct. 5, 2022, the entire contents of each of which are incorporated herein by reference. This application is also a continuation-in-part application of U.S. patent application Ser. No. 18/480,360, filed Oct. 3, 2023, which claims the benefit of priority of U.S. Provisional Patent Appl. No. 63/495,527, filed Apr. 11, 2023, U.S. Provisional Patent Appl. No. 63/479,142, filed Jan. 9, 2023, and EP Patent Appl. No. 23305026.9, filed Jan. 9, 2023, the entire contents of each of which are incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
63378434 | Oct 2022 | US | |
63495527 | Apr 2023 | US | |
63479142 | Jan 2023 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 17816958 | Aug 2022 | US |
Child | 18297489 | US | |
Parent | PCT/IB2022/056159 | Jul 2022 | WO |
Child | 17816958 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 18331054 | Jun 2023 | US |
Child | 18540710 | US | |
Parent | 18297489 | Apr 2023 | US |
Child | 18331054 | US | |
Parent | PCT/IB2022/052989 | Mar 2022 | WO |
Child | PCT/IB2022/056159 | US | |
Parent | 18057191 | Nov 2022 | US |
Child | 18331054 | US | |
Parent | 17815885 | Jul 2022 | US |
Child | 18057191 | US | |
Parent | 18480360 | Oct 2023 | US |
Child | 18540710 | US |